The Applied Biosystems™ TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel is a multiplex, real-time reverse transcription polymerase chain reaction (RT-PCR) in vitro diagnostic test for the qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A virus, influenza B virus, and respiratory syncytial virus (RSV). Nucleic acids are isolated and purified from nasopharyngeal (NP) swab and anterior nasal (AN) swab specimens obtained from individuals exhibiting signs and symptoms of a respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2, influenza, and RSV can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza A, influenza B, and RSV A/B (undifferentiated) infections in humans and is not intended to detect influenza C virus infections. Nucleic acids from the viral organisms identified by this test are generally detectable in NP and AN swab specimens during the acute phase of infection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infection are indicative of the presence of the identified virus and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out coinfection with other organisms. The organism(s) detected by the Applied Biosystems™ TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A virus, influenza B virus, or RSV infections. The Applied Biosystems™ TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel is intended for use by qualified and trained clinical laboratory personnel specifically instructed and trained in the techniques of real-time PCR and in vitro diagnostic procedures.
Device Story
Multiplex real-time RT-PCR assay for qualitative detection of SARS-CoV-2, influenza A, influenza B, and RSV. Input: RNA extracted from nasopharyngeal or anterior nasal swabs using MagMAX Dx Viral/Pathogen NA Isolation Kit on KingFisher Apex Dx system. Processing: Reverse transcription and PCR amplification on QuantStudio 5 Dx instrument using gene-specific primers/probes (SARS-CoV-2: Orf1a, Orf1b, N; Flu A: PB1, M; Flu B: M, NS; RSV: NP, M, L). Output: Fluorescent signal monitored by instrument; Diomni software interprets results based on Assay Definition File (ADF) and Cq cutoffs. Used in clinical laboratories by trained personnel. Results aid differential diagnosis in conjunction with clinical/epidemiological data. Benefits patient by enabling rapid, accurate identification of common respiratory viruses to guide management.
Clinical Evidence
Prospective clinical study (N=1,840) compared subject device to FDA-cleared molecular comparator. Evaluated 1,620 NP and 1,541 AN swabs. PPA for SARS-CoV-2 was 96.6% (NP) and 95.3% (AN); NPA was 96.47% (NP) and 96.46% (AN). Flu A/B and RSV performance showed high agreement. Enrichment study (N=69) supplemented data for Flu A/B and RSV. Bench testing confirmed LoD, inclusivity, interference, and stability.
Technological Characteristics
Multiplex real-time RT-PCR. Targets: SARS-CoV-2 (3 genes), Flu A (2 genes), Flu B (2 genes), RSV (3 genes), RNase P (internal control). Instrumentation: QuantStudio 5 Dx. Sample Prep: Magnetic bead-based (MagMAX Dx). Connectivity: Standalone instrument with Diomni software. Software: Rule-based interpretation via Assay Definition File (ADF).
Indications for Use
Indicated for qualitative detection and differentiation of SARS-CoV-2, influenza A, influenza B, and RSV in nasopharyngeal and anterior nasal swab specimens from symptomatic individuals. Intended to aid in differential diagnosis of respiratory infections. Not intended to detect influenza C.
Regulatory Classification
Identification
A device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test is an in vitro diagnostic device intended for the detection and identification of SARS-CoV-2 and other microbial agents when in a multi-target test in human clinical respiratory specimens from patients suspected of respiratory infection who are at risk for exposure or who may have been exposed to these agents. The device is intended to aid in the diagnosis of respiratory infection in conjunction with other clinical, epidemiologic, and laboratory data or other risk factors.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The intended use in the labeling required under § 809.10 of this chapter must include a description of the following: Analytes and targets the device detects and identifies, the specimen types tested, the results provided to the user, the clinical indications for which the test is to be used, the specific intended population(s), the intended use locations including testing location(s) where the device is to be used (if applicable), and other conditions of use as appropriate.
(2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(3) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens;
(ii) Detailed descriptions of the performance characteristics of the device for each specimen type claimed in the intended use based on analytical studies including the following, as applicable: Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, precision, reproducibility, and clinical studies;
(iii) Detailed descriptions of the test procedure(s), the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing;
(iv) A warning statement that viral culture should not be attempted in cases of positive results for SARS-CoV-2 and/or any similar microbial agents unless a facility with an appropriate level of laboratory biosafety (
*e.g.,* BSL 3 and BSL 3+, etc.) is available to receive and culture specimens; and(v) A prominent statement that device performance has not been established for specimens collected from individuals not identified in the intended use population (
*e.g.,* when applicable, that device performance has not been established in individuals without signs or symptoms of respiratory infection).(vi) Limiting statements that indicate that:
(A) A negative test result does not preclude the possibility of infection;
(B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) There is a risk of incorrect results due to the presence of nucleic acid sequence variants in the targeted pathogens;
(D) That positive and negative predictive values are highly dependent on prevalence;
(E) Accurate results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(F) When applicable (
*e.g.,* recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer-reviewed literature), that the clinical performance may be affected by testing a specific clinical subpopulation or for a specific claimed specimen type.(4) Design verification and validation must include:
(i) Detailed documentation, including performance results, from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, as appropriate, additional characterized clinical samples. The clinical study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained using a comparator that FDA has determined is appropriate. Detailed documentation must include the clinical study protocol (including a predefined statistical analysis plan), study report, testing results, and results of all statistical analyses.
(ii) Risk analysis and documentation demonstrating how risk control measures are implemented to address device system hazards, such as Failure Modes Effects Analysis and/or Hazard Analysis. This documentation must include a detailed description of a protocol (including all procedures and methods) for the continuous monitoring, identification, and handling of genetic mutations and/or novel respiratory pathogen isolates or strains (
*e.g.,* regular review of published literature and periodic in silico analysis of target sequences to detect possible mismatches). All results of this protocol, including any findings, must be documented and must include any additional data analysis that is requested by FDA in response to any performance concerns identified under this section or identified by FDA during routine evaluation. Additionally, if requested by FDA, these evaluations must be submitted to FDA for FDA review within 48 hours of the request. Results that are reasonably interpreted to support the conclusion that novel respiratory pathogen strains or isolates impact the stated expected performance of the device must be sent to FDA immediately.(iii) A detailed description of the identity, phylogenetic relationship, and other recognized characterization of the respiratory pathogen(s) that the device is designed to detect. In addition, detailed documentation describing how to interpret the device results and other measures that might be needed for a laboratory diagnosis of respiratory infection.
(iv) A detailed device description, including device components, ancillary reagents required but not provided, and a detailed explanation of the methodology, including molecular target(s) for each analyte, design of target detection reagents, rationale for target selection, limiting factors of the device (
*e.g.,* saturation level of hybridization and maximum amplification and detection cycle number, etc.), internal and external controls, and computational path from collected raw data to reported result (*e.g.,* how collected raw signals are converted into a reported signal and result), as applicable.(v) A detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review.
(vi) For devices intended for the detection and identification of microbial agents for which an FDA recommended reference panel is available, design verification and validation must include the performance results of an analytical study testing the FDA recommended reference panel of characterized samples. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(vii) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens, the design verification and validation must include a detailed description of the identity, phylogenetic relationship, or other recognized characterization of the Influenza A and B viruses that the device is designed to detect, a description of how the device results might be used in a diagnostic algorithm and other measures that might be needed for a laboratory identification of Influenza A or B virus and of specific Influenza A virus subtypes, and a description of the clinical and epidemiological parameters that are relevant to a patient case diagnosis of Influenza A or B and of specific Influenza A virus subtypes. An evaluation of the device compared to a currently appropriate and FDA accepted comparator method. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(5) When applicable, performance results of the analytical study testing the FDA recommended reference panel described in paragraph (b)(4)(vi) of this section must be included in the device's labeling under § 809.10(b) of this chapter.
(6) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens in addition to detection of SARS-CoV-2 and similar microbial agents, the required labeling under § 809.10(b) of this chapter must include the following:
(i) Where applicable, a limiting statement that performance characteristics for Influenza A were established when Influenza A/H3 and A/H1-2009 (or other pertinent Influenza A subtypes) were the predominant Influenza A viruses in circulation.
(ii) Where applicable, a warning statement that reads if infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to State or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
(iii) Where the device results interpretation involves combining the outputs of several targets to get the final results, such as a device that both detects Influenza A and differentiates all known Influenza A subtypes that are currently circulating, the device's labeling must include a clear interpretation instruction for all valid and invalid output combinations, and recommendations for any required followup actions or retesting in the case of an unusual or unexpected device result.
(iv) A limiting statement that if a specimen yields a positive result for Influenza A, but produces negative test results for all specific influenza A subtypes intended to be differentiated (
*i.e.,* H1-2009 and H3), this result requires notification of appropriate local, State, or Federal public health authorities to determine necessary measures for verification and to further determine whether the specimen represents a novel strain of Influenza A.(7) If one of the actions listed at section 564(b)(1)(A) through (D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those influenza viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized influenza viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's labeling required under § 809.10(b) of this chapter that accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that accompanies the device, prominently providing a hyperlink to the manufacturer's public website where the analytical reactivity testing data can be found. The manufacturer's website, as well as the primary part of the manufacturer's website that discusses the device, must provide a prominently placed hyperlink to the website containing this information and must allow unrestricted viewing access.
K252387 — Simplexa COVID-19/ Flu A/B & RSV Direct (MOL4450); Simplexa COVID-19/ Flu A/B & RSV Positive Control Pack (MOL4460) · Diasorin Molecular, LLC · Oct 30, 2025
K241573 — Alinity m Resp-4-Plex · Abbott Molecular, Inc. · Feb 14, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 8, 2025
Life Technologies Corporation Stacey Moltchanoff Regulatory Affairs Manager 6055 Sunol Boulevard Pleasanton, California 94566
Re: K241806
Trade/Device Name: Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel Regulation Number: 21 CFR 866.3981 Regulation Name: Device To Detect And Identify Nucleic Acid Targets In Respiratory Specimens From Microbial Agents That Cause The Sars-Cov-2 Respiratory Infection And Other Microbial Agents When In A Multi-Target Test Regulatory Class: Class II Product Code: QOF Dated: June 21, 2024
Received: June 21, 2024
Dear Stacey Moltchanoff:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Anna M. Mielech -S
Anna Mielech, PhD. Deputy Branch Chief (Acting) Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241806
#### Device Name
Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel
#### Indications for Use (Describe)
The Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel is a multiplex, real-time reverse transcription polymerase chain reaction (RT-PCR) in vitro diagnostic test for the qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza B virus, and respiratory syncytial virus (RSV). Nucleic acids are isolated and purified from nasopharyngeal (NP) swab and anterior nasal (AN) swab specimens obtained from individuals exhibiting signs and symptoms of a respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2, influenza, and RSV can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza B, and RSV A/B (undifferentiated) infections in humans and is not intended to detect influenza C virus infections.
Nucleic acids from the viral organisms identified by this test are generally detectable in NP and AN swab specimens during the acute phase of infection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infective of the presence of the identified virus and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings.
The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out coinfection with other organism(s) detected by the Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A virus, influenza B virus, or RSV infections.
The Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel is intended for use by qualified and trained clinical laboratory personnel specifically instructed and trained in the techniques of real-time PCR and in vitro diagnostic procedures.
Type of Use (Select one or both, as applicable)
| <div><span style="font-size:14px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div><span style="font-size:14px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary
| Submitter: | Life Technologies Corporation (a legal entity of Thermo Fisher<br>Scientific) |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 6055 Sunol Blvd, Pleasanton, CA 94566, USA |
| Phone: | (1) 760-603-7200 |
| Fax: | N/A |
| Contact Person: | Stacey Moltchanoff |
| Contact email: | stacey.moltchanoff@thermofisher.com |
| Contact Phone Number: | 1 (707) 494-1756 |
| Date: | June 21, 2024 |
| Proprietary Name: | Applied Biosystems™ TaqPath™ COVID-19, Flu A, Flu B,<br>RSV Select Panel |
| Measurand: | The assay detects and identifies RNA from: SARS-CoV-2,<br>influenza A virus, influenza B virus, and respiratory syncytial<br>virus (RSV) |
| Regulation: | 866.3981 |
| Classification: | Class II |
| Product Code: | QOF |
| Panel: | MI - Microbiology |
| Predicate Device: | BioFire® Respiratory Panel 2.1 (DEN200031) |
The Applied Biosystems™ TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel is a multiplex, real-time reverse transcription polymerase chain reaction (RT-PCR) test. The severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A, influenza B, respiratory syncytial virus (RSV) A/B and RNase P primer and probe sets are designed to detect viral RNA in nasopharyngeal (NP) and anterior nasal (AN) swab specimens from individuals exhibiting signs and symptoms of a respiratory tract infection.
Each TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel includes the following components:
- TaqPath™ COVID-19, Flu A, Flu B, RSV Select Assay-Multiplex assays that contain . primer and probe sets specific to the following targets:
- Three SARS-CoV-2 targets (Orfla, Orf1b, and N genes) .
- . Two influenza A virus targets (PB1 & M genes)
- I Two influenza B virus targets (M & NS genes)
- . Three RSV targets (NP, M, and L protein genes)
- l RNase P (internal human sample collection control)
{5}------------------------------------------------
- TaqPath™ COVID-19, Flu A, Flu B, RSV Select Positive Control—Inactivated viral . control that contains SARS-CoV-2, influenza A, influenza B, and RSV.
- TaqPath™ COVID-19, Flu A, Flu B, RSV Select Negative Control—MS2 packaged RNA . control that contains targets specific to RNase P genomic regions targeted by the assay.
- TaqPath™ 1-Step Select Master Mix (No ROX)—Ready-to-use PCR mix, including . reverse transcriptase, polymerase, dNTPs, salts, and buffer.
- . Package Insert -- Provides the instructions and the link to download the instructions for use and other assets (including the ADF)
- An Assay Definition File (ADF) applicable to the instrument used in the workflow . (available via download).
In addition to the SARS-CoV-2, influenza A, influenza B and RSV viral assay targets, the assay portion of the panel includes RNase P, which serves as an endogenous internal process control to monitor extraction and amplification of each clinical sample. The TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel also contains external process positive and negative controls. The positive control (PC) component included is an inactivated viral control that contains SARS-CoV-2, influenza A, influenza B, and RSV viruses. The PC monitors extraction and real-time RT-PCR by demonstrating that each of the four viruses can be detected when present and that RNase P is not detected when absent. The negative control (NC) component included is an MS2 packaged RNA control that contains targets specific to RNase P genomic regions targeted by the assay. The NC also monitors extraction and real-time RT-PCR by demonstrating RNase P can be detected when present and that the four viruses are not detected when absent. The TaqPath™ 1-Step Select Master Mix (No ROX) included as a component of the kit is a ready-to-use PCR mix which contains a deoxyribonucleotide triphosphate mix (dNTPs), enzymes, and other components to permit reverse transcription and amplification of the assay targets. The TagPath™ 1-Step Select Master Mix (No ROX) also contains ribonuclease (RNase) inhibitors as well as deoxyuridine triphosphate (dUTP) and uracil N-glycosylase (known as UNG or UDG).
The TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel is provided in two overall kit configurations: either 200 reactions or 1.000 reactions. The materials provided within each TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel kit are shown in Table 1 (200 reactions, Cat. No. A56123) and Table 2 (1,000 reactions, Cat. No. A56124).
{6}------------------------------------------------
Table 1: TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel Components (200 reactions, Cat. No. A56123)
| Component | SKU (box<br>SKU) | Quantity | Volume per<br>tube or bottle |
|-----------------------------------------------------------------|------------------|----------|------------------------------|
| TaqPath™ COVID 19, Flu A, Flu B, RSV Select<br>Assay | A56440 | 2 tubes | 150 µL |
| TaqPath™ COVID 19, Flu A, Flu B, RSV Select<br>Positive Control | 963612 | 6 tubes | 250 µL |
| TaqPath™ COVID 19, Flu A, Flu B, RSV Select<br>Negative Control | 963609 | 6 tubes | 250 µL |
| TaqPath™ 1-Step Select Master Mix (No ROX) | A56518 | 2 tubes | 1 mL |
## Table 2: TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel Components (1,000 reactions, Cat. No. A56124)
| Component | SKU (box<br>SKU) | Quantity | Volume per<br>tube or bottle |
|-----------------------------------------------------------------|------------------|----------|------------------------------|
| TaqPath™ COVID 19, Flu A, Flu B, RSV Select<br>Assay | A56441 | 10 tubes | 150 µL |
| TaqPath™ COVID 19, Flu A, Flu B, RSV Select<br>Positive Control | 963613 | 12 tubes | 250 µL |
| TaqPath™ COVID 19, Flu A, Flu B, RSV Select<br>Negative Control | 963608 | 12 tubes | 250 µL |
| TaqPath™ 1-Step Select Master Mix (No ROX) | A56519 | 10 tubes | 1 mL |
Materials required but not provided with the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel:
- Applied Biosystems™ QuantStudio™ 5 Dx Real-Time PCR Instrument ●
- . MagMAX™ Dx Viral/Pathogen NA Isolation Kit
- KingFisher™ Apex Dx Purification System .
- . Additional reagents and materials listed in the instructions for use
A high-level overview of the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel workflow is shown in Figure 1 below.
{7}------------------------------------------------
Image /page/7/Figure/0 description: The image shows a diagram of a process with four steps: Sample Collection, Sample Preparation, RT-PCR, and Data Analysis. In the Sample Collection step, RNA is extracted from a patient sample. In the Sample Preparation step, nucleic acids are isolated and purified from the specimens using the MagMAX Dx Viral/Pathogen NA Isolation Kit with the KingFisher Apex Dx Purification System. In the RT-PCR step, RT-PCR is run on the QuantStudio 5 Dx Real-Time PCR Instrument. In the Data Analysis step, results are analyzed and interpreted by the Diomni Software according to the settings in the Assay Definition File.
#### Figure 1: TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel Workflow
The TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel workflow begins with nucleic acid extraction from nasopharyngeal and anterior nasal swab specimens that arrive at the testing site stored in viral transport media (VTM) or universal transport medium (UTM. Nucleic acids are isolated and purified from the specimens using the MagMAX™ Dx Viral/Pathogen NA Isolation Kit with the KingFisher™ Apex Dx Purification System.
The purified nucleic acid is reverse transcribed into cDNA and amplified using the TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel and an Applied Biosystems™ QuantStudio™ 5 Dx Real Time PCR Instrument. The Applied Biosystems™ Diomni™ Software, which is used for results interpretation, is supplied with the PCR instrument.
Once the reaction plate is loaded onto the real-time PCR instrument, RNA is reverse transcribed into cDNA using the same gene-specific primers used for PCR. After heat inactivation of the reverse transcriptase enzyme, thermal cycling for PCR amplification commences.
During the anneal/extension phase of a PCR cycle, primers and probes specific to the target genes bind to corresponding DNA sequences in the indicated targets. The 5' nuclease activity of Taq polymerase degrades the probe, causing the reporter dye to separate from the quencher dye, generating a fluorescent signal. With each cycle, additional reporter dye molecules are cleaved
{8}------------------------------------------------
from their respective probes, increasing the overall fluorescence intensity. Fluorescence intensity is monitored at each PCR cycle by the real-time PCR instrument.
The Diomni™ Software implements the thermal cycling program, secondary analysis and clinical interpretation of the results using assay-specific parameters from the Assay Definition File (ADF) that is installed in the software.
Following PCR, results are analyzed and interpreted by the Diomni™ Software according to the settings in the ADF. A detailed explanation of the calling rules, plate validity and Cq cutoff values for assay targets is provided in the instructions for use. A minimum of one negative control and one positive control must be present for each run for Diomni™ Software to call the run valid. Furthermore, all control wells must pass for the real-time RT-PCR plate to be considered valid.
Control well interpretive rules are show in Table 3 below.
| Negative Control (NC) | | | | | Well Call | Status |
|-----------------------|-------|-------|--------|---------|-----------|--------|
| C19 | Flu A | Flu B | RSV AB | RNase P | | |
| NEG | NEG | NEG | NEG | POS | Negative | Passed |
| All other scenarios | | | | | Invalid | Failed |
Table 3: Control Well Interpretive Rules
| Positive Control (PC) | | | | | Well Call | Status |
|-----------------------|-------|-------|--------|---------|-----------|--------|
| C19 | Flu A | Flu B | RSV AB | RNase P | | |
| POS | POS | POS | POS | NEG | Positive | Passed |
| All other scenarios | | | | | Invalid | Failed |
Results interpretation for patient samples is described in Table 4 below. A Positive call for RNase P is only required for samples that are Negative for all four viral targets; samples with no amplification are called Invalid.
{9}------------------------------------------------
| Target | | | | | Call | Result | |
|--------|-------|-------|--------|------------|----------|-------------------------------------------------|--|
| C19 | Flu A | Flu B | RSV AB | RNase P | Call | Result | |
| POS | POS | POS | POS | POS or NEG | Positive | SARS-CoV-2, Flu A, Flu B, RSVAB Detected | |
| POS | POS | POS | NEG | POS or NEG | Positive | SARS-CoV-2, Flu A, Flu B Detected | |
| POS | POS | NEG | POS | POS or NEG | Positive | SARS-CoV-2, Flu A, RSVAB Detected | |
| POS | POS | NEG | NEG | POS or NEG | Positive | SARS-CoV-2, Flu A Detected | |
| POS | NEG | POS | POS | POS or NEG | Positive | SARS-CoV-2, Flu B, RSVAB Detected | |
| POS | NEG | POS | NEG | POS or NEG | Positive | SARS-CoV-2, Flu B Detected | |
| POS | NEG | NEG | POS | POS or NEG | Positive | SARS-CoV-2, RSVAB Detected | |
| POS | NEG | NEG | NEG | POS or NEG | Positive | SARS-CoV-2 Detected | |
| NEG | POS | POS | POS | POS or NEG | Positive | Flu A, Flu B, RSVAB Detected | |
| NEG | POS | POS | NEG | POS or NEG | Positive | Flu A, Flu B Detected | |
| NEG | POS | NEG | POS | POS or NEG | Positive | Flu A, RSVAB Detected | |
| NEG | POS | NEG | NEG | POS or NEG | Positive | Flu A Detected | |
| NEG | NEG | POS | POS | POS or NEG | Positive | Flu B, RSVAB Detected | |
| NEG | NEG | POS | NEG | POS or NEG | Positive | Flu B Detected | |
| NEG | NEG | NEG | POS | POS or NEG | Positive | RSVAB Detected | |
| NEG | NEG | NEG | NEG | POS | Negative | SARS-CoV-2, Flu A, Flu B, RSVAB Not<br>Detected | |
| NEG | NEG | NEG | NEG | NEG | Invalid | RETEST[1] | |
Table 4: Patient Samples Interpretive Rules
[1] Retesting must be performed by re-extracting the original sample and repeating the real-time PCR. If the repeat result remains invalid, collection of a new specimen should be considered.
#### Intended Use
The Applied Biosystems™ TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel is a multiplex, real-time reverse transcription polymerase chain reaction (RT-PCR) in vitro diagnostic test for the qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A virus, influenza B virus, and respiratory syncytial virus (RSV). Nucleic acids are isolated and purified from nasopharyngeal (NP) swab and anterior nasal (AN) swab specimens obtained from individuals exhibiting signs and symptoms of a respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2,
{10}------------------------------------------------
influenza, and RSV can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza A, influenza B, and RSV A/B (undifferentiated) infections in humans and is not intended to detect influenza C virus infections.
Nucleic acids from the viral organisms identified by this test are generally detectable in NP and AN swab specimens during the acute phase of infection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infection are indicative of the presence of the identified virus and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings.
The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out coinfection with other organisms. The organism(s) detected by the Applied Biosystems™ TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A virus, influenza B virus, or RSV infections.
The Applied Biosystems™ TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel is intended for use by qualified and trained clinical laboratory personnel specifically instructed and trained in the techniques of real-time PCR and in vitro diagnostic procedures.
#### Predicate Device Comparison
The similarities and differences between TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel and the predicate device, BioFire® Respiratory Panel 2.1 (DEN20031) are described in Table 5 and Table 6 below.
{11}------------------------------------------------
| Device & Predicate Device | New Device | DEN200031 |
|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade Name | TaqPath™ COVID-19, Flu A, Flu<br>B, RSV Select Panel | BioFire® Respiratory Panel 2.1<br>(RP2.1) |
| General Device<br>Characteristic Similarities | | |
| Organisms Detected | SARS-CoV-2, Influenza A,<br>Influenza B, Respiratory Syncytial<br>Virus | SARS-CoV-2, Influenza A,<br>Influenza B, Respiratory Syncytial<br>Virus |
| Analyte | RNA | RNA |
| Technological Principles | real-time reverse transcription<br>PCR | reverse transcription PCR |
| Specimen Types | nasopharyngeal swabs | nasopharyngeal swabs |
| Test Interpretation | Automated test interpretation | Automated test interpretation |
| Transport Media | Universal Transport Medium<br>(UTM) / Viral Transport Medium<br>(VTM) | Universal Transport Medium<br>(UTM) / Viral Transport Medium<br>(VTM) |
| General Device<br>Characteristic Differences | New Device | DEN200031 |
| Device Trade Name | TaqPath™ COVID-19, Flu A, Flu B, RSV Select<br>Panel | BioFire® Respiratory Panel 2.1 (RP2.1) |
| Intended Use/ Indications<br>For Use | The Applied Biosystems™ TaqPath™ COVID-19,<br>Flu A, Flu B, RSV Select Panel is a multiplex, real-<br>time reverse transcription polymerase chain reaction<br>(RT-PCR) in vitro diagnostic test for the qualitative<br>detection and differentiation of severe acute<br>respiratory syndrome coronavirus 2 (SARS-CoV-2),<br>influenza A virus, influenza B virus, and respiratory<br>syncytial virus (RSV). Nucleic acids are isolated and<br>purified from nasopharyngeal (NP) swab and anterior<br>nasal (AN) swab specimens obtained from individuals<br>exhibiting signs and symptoms of a respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2, influenza, and RSV can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza A, influenza B, and RSV A/B (undifferentiated) infections in humans and is not intended to detect influenza C virus infections.<br>Nucleic acids from the viral organisms identified by this test are generally detectable in NP and AN swab specimens during the acute phase of infection. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infection are indicative of the presence of the identified virus and aids in diagnosis if | The BioFire® Respiratory Panel 2.1 (RP2.1) is a<br>PCR-based multiplexed nucleic acid test intended<br>for use with the BioFire FilmArray® 2.0 or<br>BioFire FilmArray® Torch systems for the simultaneous qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swabs (NPS) obtained from individuals suspected of respiratory tract infections, including COVID-19.<br>The following organism types and subtypes are identified using the BioFire® RP2.1:<br>Adenovirus, Coronavirus 229E, Coronavirus HKU1, Coronavirus NL63, Coronavirus OC43, Severe Acute Respiratory Syndrome<br>Coronavirus (SARS-CoV-2), Human Metapneumovirus, Human Rhinovirus/Enterovirus, Influenza A, including subtypes H1, H1-2009,<br>and H3, Influenza B, |
| General Device<br>Characteristic Differences | New Device | DEN200031 |
| | used in conjunction with other clinical and<br>epidemiological information, and laboratory findings.<br>The results of this test should not be used as the sole<br>basis for diagnosis, treatment, or other patient<br>management decisions. Positive results do not rule out<br>coinfection with other organisms. The organism(s)<br>detected by the Applied Biosystems™ TaqPath™<br>COVID-19, Flu A, Flu B, RSV Select Panel may not<br>be the definite cause of disease. Negative results do<br>not preclude SARS-CoV-2, influenza A virus,<br>influenza B virus, or RSV infections.<br>The Applied Biosystems™ TaqPath™ COVID-19,<br>Flu A, Flu B, RSV Select Panel is intended for use by<br>qualified and trained clinical laboratory personnel<br>specifically instructed and trained in the techniques of<br>real-time PCR and in vitro diagnostic procedures. | Parainfluenza Virus 1, Parainfluenza Virus 2, Parainfluenza Virus 3, Parainfluenza Virus 4, Respiratory Syncytial Virus, Bordetella parapertussis (IS1001), Bordetella pertussis (ptxP), Chlamydia pneumoniae, and Mycoplasma pneumoniae Nucleic acids from the respiratory viral and<br>bacterial organisms identified by this test are<br>generally detectable in NPS specimens during the<br>acute phase of infection. The detection and<br>identification of specific viral and bacterial nucleic<br>acids from individuals exhibiting signs and/or<br>symptoms of respiratory infection is indicative of<br>the presence of the identified microorganism and<br>aids in the diagnosis of respiratory infection if<br>used in conjunction with other clinical and<br>epidemiological information. The results of this<br>test should not be used as the sole basis for<br>diagnosis, treatment, or other patient management<br>decisions.<br>Negative results in the setting of a respiratory<br>illness may be due to infection with pathogens that<br>are not detected by this test, or lower respiratory<br>tract infection that may not be detected by an NPS |
| General Device<br>Characteristic Differences | New Device | DEN200031 |
| | | specimen. Positive results do not rule out<br>coinfection with other organisms. The agent(s)<br>detected by the BioFire® RP2.1 may not be the<br>definite cause of disease. Additional laboratory<br>testing (e.g. bacterial and viral culture,<br>immunofluorescence, and radiography) may be<br>necessary when evaluating a patient with possible<br>respiratory tract infection. |
| Specimen Types | Anterior Nasal Swabs and Nasopharyngeal swabs | Nasopharyngeal swabs only |
| Instrumentation | Sample Prep/Nucleic Acid Purification:<br>• MagMAX™ Dx Viral/Pathogen NA Isolation Kit<br>(magnetic bead-based sample prep chemistry)<br>• KingFisher™ Apex Dx Purification System<br>(sample prep instrumentation)<br>Real-time PCR instrument:<br>• QuantStudio 5 Dx | Sample Prep/Nucleic Acid Purification:<br>• Magnetic bead-based chemistry is contained<br>within the BioFire® RP2.1 pouch<br>Instruments:<br>• FilmArray® 2.0<br>• FilmArray® Torch |
| Controls | One negative control and one positive control are run<br>for each plate (external process controls). The assay<br>detects endogenous RNase P present in the specimen<br>as an internal positive/process control. | Two internal controls are included in each reagent<br>pouch for quality control of sample processing and<br>both PCR stages and melt analysis. |
# Table 5: Predicate Device Similarities
{12}------------------------------------------------
# Table 6: Predicate Device Differences
{13}------------------------------------------------
{14}------------------------------------------------
{15}------------------------------------------------
## Analytical/Clinical Performance
Note that the terms “Cq” and “Ct” are used interchangeably throughout the documentation and refer to equivalent values. Quantification cycle (Cq) is an algorithm-agnostic term across different methods for determining the fractional cycle used for PCR reaction quantification, whereas "Ct", or threshold cycle, refers specifically to the baseline-threshold algorithm term used by the software for determining this same value.
## Analytical Performance Summaries
The analytical performance testing that was conducted to support substantial equivalence is summarized below.
## Limit of Detection Determination
The preliminary LoD was determined by probit analysis at a hit rate of 0.95. LoD was determined for two (2) strains of influenza A (Flu A) H1N1, two (…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.