Archived clinical specimens (leftover samples from standard of care testing)
Archived clinical specimens were used to supplement the prospective clinical study to evaluate the clinical sensitivity and specificity of the device for analytes not encountered in sufficient numbers during the prospective study.
Archived clinical specimens; Clinical sensitivity; Clinical specificity; Standard of care
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Clinical Study with preselected, archived specimens; Retrospective analysis of archived clinical specimens
Patients presenting with signs and symptoms of respiratory tract infection or pharyngitis; Sample Size: 562 nasopharyngeal swabs and 136 throat swabs; Number of Sites: 4 U.S. study sites
FDA-cleared Multi-Analyte Panels and/or PCR with bidirectional sequencing
Clinical sensitivity and specificity (Positive/Negative Percent Agreement)
Indications for Use
The BIOFIRE® SPOTFIRE® Respiratory/Sore Throat (R/ST) Panel is a multiplexed polymerase chain reaction (PCR) test intended for use with the BIOFIRE® System for the simultaneous, qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swab (NPS) specimens obtained from individuals with signs and symptoms of respiratory tract infection, including COVID-19; (Respiratory menu) or in throat swab (TS) specimens from individuals with signs and symptoms of pharyngitis (Sore Throat menu). The following organism types and subtypes are identified and differentiated using the SPOTFIRE R/ST Panel: Respiratory Menu Viruses Adenovirus Coronavirus SARS-CoV-2 Coronavirus (seasonal) Human metapneumovirus Human rhinovirus/enterovirus Influenza A virus Influenza A virus/H1-2009 Influenza A virus/ H3 Influenza B virus Parainfluenza virus Respiratory syncytial virus - Bacteria Bordetella parapertussis Bordetella pertussis Chlamydia pneumoniae Mycoplasma pneumoniae Sore Throat Menu Viruses Adenovirus Coronavirus (seasonal) Human metapneumovirus Human rhinovirus/enterovirus Influenza A virus Influenza A virus/H1-2009 Influenza A virus/H3 Influenza B virus Parainfluenza virus Respiratory syncytial virus Bacteria Chlamydia pneumoniae Mycoplasma pneumoniae Streptococcus dysgalactiae (Group C/G Strep) Streptococcus pyogenes (Group A Strep) Nucleic acids from the viral and bacterial organisms identified by this test are generally detectable in NPS/TS specimens during the acute phase of infection and identification of specific viral and bacterial nucleic acids from individuals exhibiting signs and symptoms of respiratory infection and/or pharyngtis are indicative of the identified microorganism and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Negative results in the setting of a respiratory illness and/or pharyngitis may be due to infection with pathogens that are not detected by this test, or a respiratory tract infection that may not be detected by an NPS or TS specimen. Positive results do not rule out coinfection with other organisms. The agent(s) detected by the SPOTFIRE R/ST Panel may not be the definite cause of disease. Additional laboratory testing (e.g., bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible respiratory tract infection and/or pharyngitis.
Device Story
The BIOFIRE SPOTFIRE R/ST Panel is a closed-system, automated, multiplexed PCR-based diagnostic test. It processes nasopharyngeal or throat swab samples to detect viral and bacterial respiratory pathogens. The user adds the sample to a reagent pouch, which is then inserted into the BIOFIRE SPOTFIRE System. The system performs automated nucleic acid extraction, nested multiplex PCR, and DNA melting analysis. Results are reported in approximately 15 minutes. The device is intended for use in clinical settings by healthcare providers to aid in the diagnosis of respiratory tract infections and pharyngitis. The output provides qualitative detection of specific pathogens, which clinicians use alongside other clinical and epidemiological data to inform patient management decisions.
Clinical Evidence
Clinical performance was evaluated using prospectively collected (1,120 NPS, 877 TS), archived, and contrived specimens. Comparator methods included FDA-cleared multi-analyte panels, culture, and PCR sequencing. PPA for prospectively collected NPS specimens ranged from 96.3-100% and NPA from 90.6-100%. For TS specimens, PPA ranged from 87.5-100% and NPA from 93.5-100%. Results support the clinical sensitivity and specificity for the identified pathogens.
Technological Characteristics
The device uses a closed-system disposable pouch containing freeze-dried reagents for nucleic acid extraction, reverse transcription, and nested multiplex PCR. Detection is via DNA melting analysis using LC Green Plus fluorescent dye. The system is automated, uses a touchscreen-equipped control station, and is factory-calibrated. It is designed for professional use in clinical settings.
Indications for Use
Indicated for individuals with signs/symptoms of respiratory tract infection (NPS specimens) or pharyngitis (TS specimens) to detect/identify specific viral and bacterial nucleic acids. Not for sole diagnostic use; results should be used with clinical/epidemiological data.
Regulatory Classification
Identification
A device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test is an in vitro diagnostic device intended for the detection and identification of SARS-CoV-2 and other microbial agents when in a multi-target test in human clinical respiratory specimens from patients suspected of respiratory infection who are at risk for exposure or who may have been exposed to these agents. The device is intended to aid in the diagnosis of respiratory infection in conjunction with other clinical, epidemiologic, and laboratory data or other risk factors.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The intended use in the labeling required under § 809.10 of this chapter must include a description of the following: Analytes and targets the device detects and identifies, the specimen types tested, the results provided to the user, the clinical indications for which the test is to be used, the specific intended population(s), the intended use locations including testing location(s) where the device is to be used (if applicable), and other conditions of use as appropriate.
(2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(3) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens;
(ii) Detailed descriptions of the performance characteristics of the device for each specimen type claimed in the intended use based on analytical studies including the following, as applicable: Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, precision, reproducibility, and clinical studies;
(iii) Detailed descriptions of the test procedure(s), the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing;
(iv) A warning statement that viral culture should not be attempted in cases of positive results for SARS-CoV-2 and/or any similar microbial agents unless a facility with an appropriate level of laboratory biosafety (
*e.g.,* BSL 3 and BSL 3+, etc.) is available to receive and culture specimens; and(v) A prominent statement that device performance has not been established for specimens collected from individuals not identified in the intended use population (
*e.g.,* when applicable, that device performance has not been established in individuals without signs or symptoms of respiratory infection).(vi) Limiting statements that indicate that:
(A) A negative test result does not preclude the possibility of infection;
(B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) There is a risk of incorrect results due to the presence of nucleic acid sequence variants in the targeted pathogens;
(D) That positive and negative predictive values are highly dependent on prevalence;
(E) Accurate results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(F) When applicable (
*e.g.,* recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer-reviewed literature), that the clinical performance may be affected by testing a specific clinical subpopulation or for a specific claimed specimen type.(4) Design verification and validation must include:
(i) Detailed documentation, including performance results, from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, as appropriate, additional characterized clinical samples. The clinical study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained using a comparator that FDA has determined is appropriate. Detailed documentation must include the clinical study protocol (including a predefined statistical analysis plan), study report, testing results, and results of all statistical analyses.
(ii) Risk analysis and documentation demonstrating how risk control measures are implemented to address device system hazards, such as Failure Modes Effects Analysis and/or Hazard Analysis. This documentation must include a detailed description of a protocol (including all procedures and methods) for the continuous monitoring, identification, and handling of genetic mutations and/or novel respiratory pathogen isolates or strains (
*e.g.,* regular review of published literature and periodic in silico analysis of target sequences to detect possible mismatches). All results of this protocol, including any findings, must be documented and must include any additional data analysis that is requested by FDA in response to any performance concerns identified under this section or identified by FDA during routine evaluation. Additionally, if requested by FDA, these evaluations must be submitted to FDA for FDA review within 48 hours of the request. Results that are reasonably interpreted to support the conclusion that novel respiratory pathogen strains or isolates impact the stated expected performance of the device must be sent to FDA immediately.(iii) A detailed description of the identity, phylogenetic relationship, and other recognized characterization of the respiratory pathogen(s) that the device is designed to detect. In addition, detailed documentation describing how to interpret the device results and other measures that might be needed for a laboratory diagnosis of respiratory infection.
(iv) A detailed device description, including device components, ancillary reagents required but not provided, and a detailed explanation of the methodology, including molecular target(s) for each analyte, design of target detection reagents, rationale for target selection, limiting factors of the device (
*e.g.,* saturation level of hybridization and maximum amplification and detection cycle number, etc.), internal and external controls, and computational path from collected raw data to reported result (*e.g.,* how collected raw signals are converted into a reported signal and result), as applicable.(v) A detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review.
(vi) For devices intended for the detection and identification of microbial agents for which an FDA recommended reference panel is available, design verification and validation must include the performance results of an analytical study testing the FDA recommended reference panel of characterized samples. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(vii) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens, the design verification and validation must include a detailed description of the identity, phylogenetic relationship, or other recognized characterization of the Influenza A and B viruses that the device is designed to detect, a description of how the device results might be used in a diagnostic algorithm and other measures that might be needed for a laboratory identification of Influenza A or B virus and of specific Influenza A virus subtypes, and a description of the clinical and epidemiological parameters that are relevant to a patient case diagnosis of Influenza A or B and of specific Influenza A virus subtypes. An evaluation of the device compared to a currently appropriate and FDA accepted comparator method. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(5) When applicable, performance results of the analytical study testing the FDA recommended reference panel described in paragraph (b)(4)(vi) of this section must be included in the device's labeling under § 809.10(b) of this chapter.
(6) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens in addition to detection of SARS-CoV-2 and similar microbial agents, the required labeling under § 809.10(b) of this chapter must include the following:
(i) Where applicable, a limiting statement that performance characteristics for Influenza A were established when Influenza A/H3 and A/H1-2009 (or other pertinent Influenza A subtypes) were the predominant Influenza A viruses in circulation.
(ii) Where applicable, a warning statement that reads if infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to State or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
(iii) Where the device results interpretation involves combining the outputs of several targets to get the final results, such as a device that both detects Influenza A and differentiates all known Influenza A subtypes that are currently circulating, the device's labeling must include a clear interpretation instruction for all valid and invalid output combinations, and recommendations for any required followup actions or retesting in the case of an unusual or unexpected device result.
(iv) A limiting statement that if a specimen yields a positive result for Influenza A, but produces negative test results for all specific influenza A subtypes intended to be differentiated (
*i.e.,* H1-2009 and H3), this result requires notification of appropriate local, State, or Federal public health authorities to determine necessary measures for verification and to further determine whether the specimen represents a novel strain of Influenza A.(7) If one of the actions listed at section 564(b)(1)(A) through (D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those influenza viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized influenza viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's labeling required under § 809.10(b) of this chapter that accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that accompanies the device, prominently providing a hyperlink to the manufacturer's public website where the analytical reactivity testing data can be found. The manufacturer's website, as well as the primary part of the manufacturer's website that discusses the device, must provide a prominently placed hyperlink to the website containing this information and must allow unrestricted viewing access.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
# DECISION SUMMARY
# ASSAY AND INSTRUMENT
# I Background Information:
A 510(k) Number
K232954
B Applicant
BioFire Diagnostics, LLC.
C Proprietary and Established Names
BIOFIRE SPOTFIRE Respiratory/Sore Throat (R/ST) Panel
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QOF | Class II | 21 CFR 866.3981 - Device To Detect And Identify Nucleic Acid Targets In Respiratory Specimens From Microbial Agents That Cause The SARS-Cov-2 Respiratory Infection And Other Microbial Agents When In A Multi-Target Test | MI - Microbiology |
| OEM | Class II | 21 CFR 866.3980 - Respiratory viral panel multiplex nucleic acid assay | MI - Microbiology |
| OOU | Class II | 21 CFR 866.3980 - Respiratory viral panel multiplex nucleic acid assay | MI - Microbiology |
| OTG | Class II | 21 CFR 866.3980 - Respiratory viral panel multiplex nucleic acid assay | MI - Microbiology |
| OZE | Class II | 21 CFR 866.3980 - Respiratory viral panel multiplex nucleic acid assay | MI - Microbiology |
| OZX | Class II | 21 CFR 866.3980 - Respiratory viral panel | MI - Microbiology |
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| | | multiplex nucleic acid assay | |
| OZY | Class II | 21 CFR 866.3980 - Respiratory viral panel multiplex nucleic acid assay | MI - Microbiology |
| OZZ | Class II | 21 CFR 866.3980 - Respiratory viral panel multiplex nucleic acid assay | MI - Microbiology |
| OCC | Class II | 21 CFR 866.3980 - Respiratory viral panel multiplex nucleic acid assay | MI - Microbiology |
| NSU | Class II | 21 CFR 862.2570 - Instrumentation for clinical multiplex test systems | CH - Clinical Chemistry |
| PGX | Class II | 21 CFR 866.2680 - Streptococcus spp. nucleic acid-based assay | MI - Microbiology |
**Note:** The primary regulation for device classification is 21 CFR 866.3981 (Product Code QOF). Other regulations and associated product codes are cited for tracking purposes.
## II Submission/Device Overview:
### A Purpose for Submission:
To obtain a substantial equivalence determination for the BIOFIRE SPOTFIRE Respiratory/Sore Throat (R/ST) Panel performed on the BIOFIRE SPOTFIRE System.
### B Measurand:
The BIOFIRE SPOTFIRE Respiratory/Sore Throat (R/ST) Panel detects and identifies nucleic acids from the following pathogens: Adenovirus, Seasonal Coronavirus (229E, HKU1, OC43 and NL63, not differentiated), Severe Acute Respiratory Syndrome (SARS)-Coronavirus-2, Human Metapneumovirus, Human Rhinovirus/Enterovirus (not differentiated), Influenza A virus with subtyping of A/H1-2009 and A/H3 (reported separately), Influenza B virus, Parainfluenza Virus 1-4 (not differentiated), Respiratory Syncytial Virus, *Bordetella parapertussis*, *Bordetella pertussis*, *Chlamydia pneumoniae*, *Mycoplasma pneumoniae*, *Streptococcus dysgalactiae* (Group C/G *Streptococcus*) and *Streptococcus pyogenes* (Group A *Streptococcus*).
### C Type of Test:
Multiplex nucleic acid assay for use with the BIOFIRE SPOTFIRE System for the qualitative detection and identification of viral and/or bacterial pathogens in nasopharyngeal swabs from patients with signs and symptoms of respiratory tract infection (BIOFIRE SPOTFIRE R Panel) or in throat swabs from patients with signs and symptoms of pharyngitis (BIOFIRE SPOTFIRE ST Panel).
K232954 - Page 2 of 75
{2}
### III Intended Use/Indications for Use:
#### A Intended Use(s):
See Indications for Use below.
#### B Indication(s) for Use:
The BIOFIRE SPOTFIRE Respiratory/Sore Throat (R/ST) Panel is a multiplexed polymerase chain reaction (PCR) test intended for use with the BIOFIRE SPOTFIRE System for the simultaneous, qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swab (NPS) specimens obtained from individuals with signs and symptoms of respiratory tract infection, including COVID-19; (Respiratory menu) or in throat swab (TS) specimens from individuals with signs and symptoms of pharyngitis (Sore Throat menu).
The following organism types and subtypes are identified and differentiated using the SPOTFIRE R/ST Panel:
| Respiratory Menu | Sore Throat Menu |
| --- | --- |
| **Viruses** | **Viruses** |
| Adenovirus | Adenovirus |
| Coronavirus SARS-CoV-2 | Coronavirus (seasonal) |
| Coronavirus (seasonal) | Human metapneumovirus |
| Human metapneumovirus | Human rhinovirus/enterovirus |
| Human rhinovirus/enterovirus | Influenza A virus |
| Influenza A virus | Influenza A virus/H1-2009 |
| Influenza A virus/H1-2009 | Influenza A virus/H3 |
| Influenza A virus/H3 | Influenza B virus |
| Influenza B virus | Parainfluenza virus |
| Parainfluenza virus | Respiratory syncytial virus |
| Respiratory syncytial virus | |
| **Bacteria** | **Bacteria** |
| *Bordetella parapertussis* | *Chlamydia pneumoniae* |
| *Bordetella pertussis* | *Mycoplasma pneumoniae* |
| *Chlamydia pneumoniae* | *Streptococcus dysgalactiae* (Group C/G Strep) |
| *Mycoplasma pneumoniae* | *Streptococcus pyogenes* (Group A Strep) |
Nucleic acids from the viral and bacterial organisms identified by this test are generally detectable in NPS/TS specimens during the acute phase of infection. The detection and identification of specific viral and bacterial nucleic acids from individuals exhibiting signs and symptoms of respiratory infection and/or pharyngitis are indicative of the presence of the identified microorganism and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.
Negative results in the setting of a respiratory illness and/or pharyngitis may be due to infection with pathogens that are not detected by this test, or a respiratory tract infection that may not be detected by an NPS or TS specimen. Positive results do not rule out coinfection with other
K232954 - Page 3 of 75
{3}
organisms. The agent(s) detected by the SPOTFIRE R/ST Panel may not be the definite cause of disease.
Additional laboratory testing (e.g., bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible respiratory tract infection and/or pharyngitis.
**C Special Conditions for Use Statement(s):**
Rx - For Prescription Use Only
**D Special Instrument Requirements:**
BIOFIRE SPOTFIRE System
**IV Device/System Characteristics:**
**A Device Description:**
The BIOFIRE SPOTFIRE Respiratory/Sore Throat (R/ST) Panel is a multiplexed nucleic acid-amplification-based test that is intended to detect and identify multiple respiratory viral and bacterial pathogens in nasopharyngeal swabs (NPS) and throat swabs (TS) from individuals with signs and symptoms of respiratory tract infection or pharyngitis, respectively. The BIOFIRE SPOTFIRE R/ST Panel is performed using the BIOFIRE SPOTFIRE System, which is an automated polymerase chain reaction (PCR)-based *in vitro* diagnostic instrument system for use with reagent pouches for specific indications.
The BIOFIRE SPOTFIRE System automates nucleic acid extraction and nested multiplex PCR in closed system. The resulting PCR products are evaluated using assay-specific DNA melting analysis. The BIOFIRE SPOTFIRE Systems Software executes the BIOFIRE SPOTFIRE R/ST Panel test according to parameters specified in the BIOFIRE SPOTFIRE R/ST Panel Software. The same protocol is used for sample preparation, nucleic acid extraction, reverse transcription (RT)-PCR amplification and analysis for both NPS and TS specimens but with different result reporting according to the specimen type selected. The sample is added to the reagent pouch and the pouch is loaded into the BIOFIRE SPOTFIRE System, after which results are reported in approximately 15 minutes without further user intervention.
The BIOFIRE SPOTFIRE System is comprised of between one and four modules that are connected to a single SPOTFIRE Control Station equipped with the SPOTFIRE System Software. The first module is placed on top of the Control Station and up to three additional modules may be stacked on top as required. Each module can be accessed at random to perform a test, independent of the other modules attached to the same Control Station.
The BIOFIRE SPOTFIRE R/ST Panel Reagent Kit includes the following components:
- BIOFIRE SPOTFIRE R/ST Panel Pouches (30 ea.)
- Each packed under vacuum in a metal canister and outer bag
- BIOFIRE Sample Preparation Reagent Kit (SPRK) (32 ea.)
- Fixed volume transfer pipette for addition of the test sample to the Sample Injection Vial
K232954 - Page 4 of 75
{4}
○ BIOFIRE Sample Buffer ampoule for addition to the Sample Injection Vial
○ BIOFIRE Sample Injection Vial for mixing of the test sample and Sample Buffer
○ BIOFIRE Hydration Injection Vial containing Hydration Solution for pouch rehydration.
# B Principle of Operation:
The BIOFIRE SPOTFIRE Panel Pouch is a closed-system disposable that contains all the reagents needed for sample preparation, reverse transcription, PCR amplification and fluorescent detection, including melt curve analysis of amplified products. The rigid plastic “fitment” of the pouch contains reagents in freeze-dried form while the flexible plastic portion of the pouch is divided into discrete “blisters” in which different chemical processes are carried out. After sample collection, the user adds the sample to the Panel Pouch, selects the applicable specimen type from the SPOTFIRE System touchscreen and then places the pouch into the instrument to start the run. All other operations are automated.
The SPOTFIRE Systems Software is designed to run on the SPOTFIRE Control Station that consists of an integrated computer and touchscreen. The SPOTFIRE Systems Software interfaces with the user via the touchscreen to operate the instrument, collect, save, analyze and retrieve data, report test results and view stored information. The BIOFIRE SPOTFIRE R/ST Panel Software works with the Systems Software and specifies the assay-specific parameters to perform the test, interpret the data and display the reported results.
To perform a test with the BIOFIRE SPOTFIRE R/ST Panel, the patient sample is first mixed with a denaturing Sample Buffer to inactivate nucleases. The user then loads the diluted sample into the BIOFIRE SPOTFIRE R/ST Panel Pouch via the injection port. The vacuum in the pouch draws the liquid into the sample well of the pouch fitment and in the process rehydrates the cell-based Process Control. Upon inserting into the instrument, the sample mixture is moved to the lysis blister where mechanical agitation with zirconia-silica beads is used to complete cell lysis. The lysate is then mixed with glass-coated magnetic beads which adsorb the nucleic acids liberated in the lysis step. The extracted nucleic acids are washed to remove potential inhibitors and then eluted into buffer that is used to rehydrate the contents of the first stage PCR blister, which includes all the components needed for reverse transcription (RT) and first stage PCR. RT-PCR is initiated upon achievement of the requisite temperature within the amplification chamber. After completion of first-stage amplification, the products of the reaction are diluted and combined with fresh reagents that are distributed over a second stage PCR array. The second stage primers are internal (“nested”) with respect to the first stage primers and each analyte or control target is represented by three discrete amplification reactions. The second stage PCR array includes an assay control, the PCR2 Control, failure of which indicates a failure in the second stage PCR and invalidates the run.
Included in the PCR master mix for the second stage PCR is a fluorescent dye (LC Green Plus) that is specific for double-stranded DNA. The dye intercalates with the double stranded-PCR products and fluoresces at a specific wavelength that is detected by the BIOFIRE SPOTFIRE System. At the end of the second stage PCR, the temperature of the array is gradually raised at a controlled ramp rate and the fluorescence is monitored. Fluorescence decreases at the melting temperature of the amplicons, and the temperature at which this occurs is characteristic of each amplification product. The Panel Software automatically evaluates the calculated melting temperature (Tm) in comparison to the predefined range for each specific analyte. The results
K232954 - Page 5 of 75
{5}
from each of the three amplification reactions for each analyte are combined by the software to determine the final test result. For an assay to be called positive, at least two of the three wells must have a PCR product with a Tm that is within the specified range for that analyte. Assays that do not meet these criteria are reported as negative.
Positive and negative results for the assay controls (Process Control and PCR2 Control) are compared to their expected values and assigned a “pass” or “fail” result. Panel-specific rules define how control failures affect interpretation of test results. For the BIOFIRE SPOTFIRE R/ST Panel, failure of either the RNA Process Control or PCR2 Control invalidates the results for all analytes. The final sample result is determined by the interpretation rules defined in the Panel Software.
For most bacteria and viruses detected by the BIOFIRE SPOTFIRE R/ST Panel, a “positive” result is reported if two or more of the wells corresponding to a single assay contain detectable PCR product of the appropriate Tm. However, to ensure inclusivity, certain analytes are targeted by multiple assays, the results of which are interpreted in combination, as described below in **Tables 1 and 2**.
**Table 1.** Interpretation of results for analytes targeted by multiple assays
| Analyte | Number of Independent Assays | Positive Result Interpretation |
| --- | --- | --- |
| Adenovirus | 4 | ≥ 1 assay positive |
| Coronavirus (seasonal) | 4 | ≥ 1 assay positive |
| Influenza A Virus | 4 | ≥ 2 assays positive ^{1} |
| Parainfluenza Virus | 4 | ≥ 1 assay positive |
| SARS-CoV-2 | 2 | ≥ 1 assay positive |
$^{1}$ At least one of the two pan-influenza A assays and *either* the H1-2009- *or* the H3-specific assay must produce positive results to report a positive subtype result (**Table 2**).
K232954 - Page 6 of 75
{6}
**Table 2.** Possible result permutations for detection of influenza A virus in patient samples
| Assay | | | | Result Reported | Action Required |
| --- | --- | --- | --- | --- | --- |
| FluA-pan1 | FluA-pan2 | FluA-H1-2009 | FluA-H3 | | |
| Positive | Positive | Positive | Negative | POSITIVE Influenza A Virus Subtype H1-2009 | Report |
| Positive | Negative | Positive | Negative | | |
| Negative | Positive | Positive | Negative | | |
| Positive | Positive | Negative | Positive | POSITIVE Influenza A Virus Subtype H3 | Report |
| Positive | Negative | Negative | Positive | | |
| Negative | Positive | Negative | Positive | | |
| Positive | Positive | Positive | Positive | POSITIVE Influenza A Virus Multiple Subtypes | Report |
| Positive | Negative | Positive | Positive | | |
| Negative | Positive | Positive | Positive | | |
| Positive | Positive | Negative | Negative | POSITIVE Influenza A Virus No Subtype Identified | Retest^{1} |
| Positive | Negative | Negative | Negative | UNCERTAIN Influenza A Virus | |
| Negative | Positive | Negative | Negative | | |
| Negative | Negative | Positive | Negative | | |
| Negative | Negative | Negative | Positive | | |
| Negative | Negative | Positive | Positive | | |
| Negative | Negative | Negative | Negative | NEGATIVE | Report |
$^{1}$ If the retest result is the same, the user is instructed to contact public health authorities for confirmatory testing.
### C Instrument Description Information:
#### 1. Instrument Name:
BIOFIRE SPOTFIRE System
#### 2. Specimen Identification:
Specimen identification can be entered manually or by scanning a barcode.
K232954 - Page 7 of 75
{7}
### 3. Specimen Sampling and Handling:
Use of the BIOFIRE SPOTFIRE R/ST Panel requires collection of either a nasopharyngeal swab (NPS) or throat swab (TS) according to standard procedures. Nasopharyngeal swabs should be placed in 1-3 mL of viral transport medium, while throat swabs should be placed in 1 mL liquid Amies Medium. The minimum sample volume required to perform a test is 300 μL. Specimens should be tested as soon as possible following collection but may be stored for up to 4 hours at room temperature (15-25 °C) or for up to 3 days at 2-8 °C or up to 30 days at ≤ -15 °C.
The SPOTFIRE System Software includes step-by-step on-screen instructions that guide the user through the process of starting a run on the instrument. After cleaning the work area and Pouch Loading Station, the user removes a BIOFIRE SPOTFIRE R /ST Panel pouch from its vacuum packaging and places it into the Pouch Loading Station. They then hydrate the pouch using the Hydration Injection Vial by injecting the contents through the Hydration Solution Injection Port and transfer a fixed volume of sample to the Sample Injection Vial, together with the entire contents of the Sample Buffer ampoule. After mixing the Sample Injection Vial by inversion, the user injects the mixture into the pouch via the Sample Injection Port, enters the sample identity into the instrument (manually or using the barcode scanner), selects the applicable specimen type (NPS or TS) and initiates the run by inserting the pouch into the instrument. The test will then proceed to completion, including interpretation of results, without further user intervention.
### 4. Calibration:
The BIOFIRE SPOTFIRE System is factory calibrated. User calibration is not required.
### 5. Quality Control:
*Internal Controls*
Two process controls are included in each BIOFIRE SPOTFIRE R/ST Panel pouch:
1) *RNA Process Control*
The RNA Process Control comprises freeze-dried *Schizosaccharomyces pombe* which is rehydrated when the sample is loaded. The control material is carried through all stages of the test process, including nucleic acid extraction and purification, reverse transcription, first and second stage PCR amplification and DNA melt analysis. A positive result indicates that all process steps carried out in the BIOFIRE SPOTFIRE R/ST Panel pouch were completed successfully.
2) *PCR2 Control*
The PCR2 control comprises a DNA target sequence that is dried into specific wells used for the second stage amplification and melting curve analysis. A positive result indicates that these processes were successful.
A positive result is required for both the RNA Process Control and the PCR2 Control for the results for an individual pouch to be considered valid. If either control fails, the sample should be retested using a new pouch.
K232954 - Page 8 of 75
{8}
# External Controls
BioFire Diagnostics recommends routine testing of External Positive and Negative Controls that are commercially available from bioMérieux or Maine Molecular Quality Controls, Inc.:
- SPOTFIRE RSP Positive Control (M42638)
- SPOTFIRE RSP Negative Control (M42738)
These controls are designed to monitor the performance of each of the individual assays within the BIOFIRE SPOTFIRE R/ST Panel and should be tested as Quality Controls in accordance with their FDA-cleared Instructions For Use (K233611) and the labeling for the BIOFIRE SPOTFIRE R/ST Panel.
In general, External Controls should be tested when training a new operator, or whenever a new shipment of BIOFIRE SPOTFIRE Panels or a new SPOTFIRE Control Station is received. Additional Negative Controls should also be run at least monthly in each testing location to monitor for environmental contamination.
A total of 351 External Positive and Negative Controls were tested during the Prospective Clinical Study described in Section VII (C). Of these, 11 were associated with instrument errors or internal control failures and 24 were tested using an incorrect or expired reagent pouch. As a result, there were 316 External Controls with valid results (141 Positive and 175 Negative), including 4 (1.3%) that were incorrectly designated by the operator as patient samples, rather than control materials. Of the 141 valid External Positive Controls, 98.6% (139/141) returned the expected results for all analytes as did all 175 (100%) of the valid Negative External Controls.
# V Substantial Equivalence Information:
# A Predicate Device Name(s):
BIOFIRE SPOTFIRE Respiratory (R) Panel
# B Predicate 510(k) Number(s):
K213954
# C Comparison with Predicate(s):
| Device & Predicate Device(s): | K232954 | K213954 |
| --- | --- | --- |
| Device Trade Name | BIOFIRE SPOTFIRE Respiratory/Sore Throat (R/ST) Panel | BIOFIRE SPOTFIRE Respiratory (R) Panel |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The BIOFIRE SPOTFIRE Respiratory/Sore Throat (R/ST) Panel is a multiplexed polymerase chain reaction (PCR) test intended for use with the BIOFIRE SPOTFIRE System for the simultaneous, qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal | The BIOFIRE SPOTFIRE Respiratory (R) Panel (SPOTFIRE R Panel) is a multiplexed polymerase chain reaction (PCR) test intended for use with the BIOFIRE SPOTFIRE System for the simultaneous, qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in |
K232954 - Page 9 of 75
{9}
| Device & Predicate Device(s): | K232954 | K213954 |
| --- | --- | --- |
| | swab (NPS) specimens obtained from individuals with signs and symptoms of respiratory tract infection, including COVID-19; (Respiratory menu) or in throat swab (TS) specimens from individuals with signs and symptoms of pharyngitis (Sore Throat menu). The following organism types and subtypes are identified and differentiated using the SPOTFIRE R/ST Panel: **Respiratory Menu** *Viruses:* Adenovirus Coronavirus SARS-CoV-2 Coronavirus (seasonal) Human metapneumovirus Human rhinovirus/enterovirus Influenza A virus ○ Influenza A virus/H1-2009 ○ Influenza A virus/H3 Influenza B virus Parainfluenza virus Respiratory syncytial virus *Bacteria:* *Bordetella parapertussis* *Bordetella pertussis* *Chlamydia pneumoniae* *Mycoplasma pneumoniae* **Sore Throat Menu** *Viruses:* Adenovirus Coronavirus (seasonal) Human metapneumovirus Human rhinovirus/enterovirus Influenza A virus ○ Influenza A virus/H1-2009 ○ Influenza A virus/H3 Influenza B virus Parainfluenza virus Respiratory syncytial virus *Bacteria:* *Chlamydia pneumoniae* *Mycoplasma pneumoniae* *Streptococcus dysgalactiae* (Group C/G Strep) *Streptococcus pyogenes* (Group A Strep) Nucleic acids from the viral and bacterial organisms identified by this test are generally detectable in | nasopharyngeal swab (NPS) specimens obtained from individuals with signs and symptoms of respiratory tract infection, including COVID-19. The following organism types and subtypes are identified and differentiated using the SPOTFIRE R Panel: *Viruses:* Adenovirus Coronavirus (seasonal) Coronavirus SARS-CoV-2 Human metapneumovirus Human rhinovirus/enterovirus Influenza A virus ○ Influenza A virus A/H1-2009 ○ Influenza A virus A/H3 Influenza B virus Parainfluenza virus Respiratory syncytial virus *Bacteria:* *Bordetella parapertussis* *Bordetella pertussis* *Chlamydia pneumoniae* *Mycoplasma pneumoniae* Nucleic acids from the viral and bacterial organisms identified by this test are generally detectable in NPS specimens during the acute phase of infection. The detection and identification of specific viral and bacterial nucleic acids from individuals exhibiting signs and/or symptoms of respiratory infection are indicative of the presence of the identified microorganism and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Negative results in the setting of a respiratory illness may be due to infection with pathogens that are not detected by this test, or lower respiratory tract infection that may not be detected by an NPS specimen. Positive results do not rule out coinfection with other organisms. The agent(s) detected |
K232954 - Page 10 of 75
{10}
| Device & Predicate Device(s): | K232954 | K213954 |
| --- | --- | --- |
| | NPS/TS specimens during the acute phase of infection. The detection and identification of specific viral and bacterial nucleic acids from individuals exhibiting signs and symptoms of respiratory infection and/or pharyngitis are indicative of the presence of the identified microorganism and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Negative results in the setting of a respiratory illness and/or pharyngitis may be due to infection with pathogens that are not detected by this test, or a respiratory tract infection that may not be detected by an NPS or TS specimen. Positive results do not rule out coinfection with other organisms. The agent(s) detected by the SPOTFIRE R/ST Panel may not be the definite cause of disease. Additional laboratory testing (e.g., bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible respiratory tract infection and/or pharyngitis. | by the SPOTFIRE R Panel may not be the definite cause of disease. Additional laboratory testing (e.g., bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible respiratory tract infection. |
| Technology | Same | Highly multiplexed PCR with DNA melting analysis |
| Reagents | Same | Sealed reagent pouch containing all reagents necessary for sample processing (nucleic acid extraction), reverse transcription (RT)-PCR amplification and melt curve analysis. |
| Internal Controls | Same | In-built controls for sample processing/nucleic acid extraction, amplification and detection. |
| Instrument System | Same | BIOFIRE SPOTFIRE System |
| Time to Result | Same | ~15 minutes |
| **General Device Characteristic Differences** | | |
| Disease Syndrome | Same *plus* signs and symptoms of | Signs and symptoms of respiratory |
K232954 - Page 11 of 75
{11}
| Device & Predicate Device(s): | K232954 | K213954 |
| --- | --- | --- |
| | pharyngitis | tract infection |
| Specimen Types | Same *plus* throat swabs | Nasopharyngeal swabs |
| Organisms Detected | Same for Respiratory Menu Sore Throat Menu does not report results for ○ Coronavirus SARS-CoV-2 ○ *Bordetella parapertussis* ○ *Bordetella pertussis* but reports additional results for: ○ *Streptococcus dysgalactiae* (Group C/G Strep) ○ *Streptococcus pyogenes* (Group A Strep) | *Viruses:* Adenovirus Coronavirus (seasonal) Coronavirus SARS-CoV-2 Human metapneumovirus Human rhinovirus/enterovirus Influenza A virus ○ Influenza A virus A/H1-2009 ○ Influenza A virus A/H3 Influenza B virus Parainfluenza virus Respiratory syncytial virus *Bacteria:* *Bordetella parapertussis* *Bordetella pertussis* *Chlamydia pneumoniae* *Mycoplasma pneumoniae* |
## VI Standards/Guidance Documents Referenced:
### *Guidance Documents*
- • Guidance for Clinical Laboratories, Commercial Manufactures, and FDA Staff – Policy for Coronavirus Disease-2019 Tests During the Public Health Emergency (2020)
- • Guidance for Industry - Part 11, Electronic Records; Electronic Signatures - Scope and Application (August 2003)
- • Guidance for Industry - Computerized Systems Used in Clinical Investigations (May 2007)
- • Guidance for Industry - Oversight of Clinical Investigations - A Risk-Based Approach to Monitoring (August 2013)
- • Guidance for Industry - Electronic Source Data in Clinical Investigations (September 2013)
- • Guidance for IRBs, Clinical Investigators, and Sponsors – Informed Consent Information Sheet (July 2014)
- • Guidance for Industry and FDA Staff - Acceptance of Clinical Data to Support Medical Device Applications and Submissions - Frequently Asked Questions (February 2018)
- • Guidance for Industry and FDA Staff - Recommendations for Clinical Laboratory Improvement Amendments of 1988 (CLIA) Waiver Applications for Manufacturers of *In Vitro* Diagnostic Devices (February 2020)
- • Guidance for Industry and FDA Staff - Recommendations for Dual 510(k) and CLIA Waiver by Application Studies (February 2020)
- • Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay (October 2009)
- • Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays (October 2009)
- • Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays (October 2009)
K232954 - Page 12 of 75
{12}
- Guidance for Industry and FDA Staff - Statistical Guidance on Reporting Results from Studies Evaluating Diagnostic Tests, (March 2007)
- Guidance for Industry and FDA Staff, Evaluation and Reporting of Age-, Race-, and Ethnicity-Specific Data in Medical Device Clinical Studies (September 2017)
- Guidance for Sponsors, Investigators, and IRBs - Impact of Certain Provisions of the Revised Common Rule on FDA-Regulated Clinical Investigations (October 2018)
- Guidance for Sponsors, Institutional Review Boards, and FDA Staff - Guidance on Informed Consent for *In Vitro* Diagnostic Device Studies Using Leftover Human Specimens that are Not Individually Identifiable (April 2006)
- Guidance for Industry and Food and Drug Administration Staff - Highly Multiplexed Microbiological/Medical Countermeasure *In Vitro* Nucleic Acid Based Diagnostic Devices (August 2014)
- Guidance for Industry and FDA Staff - Statistical Guidance on Reporting Results from Studies Evaluating Diagnostic Tests (March 2007)
- CLSI. User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline - Second Edition. CLSI document EP12-A2. Wayne, PA: Clinical and Laboratory Standards Institute; 2008.
- CLSI. Evaluation of Stability of *In Vitro* Diagnostic Reagents; Approved Guideline. CLSI document EP25-A. Wayne, PA: Clinical and Laboratory Standards Institute; 2009.
- CLSI. Interference Testing in Clinical Chemistry. 3rd ed. CLSI guideline EP07. Wayne, PA: Clinical and Laboratory Standards Institute; 2018.
- ICH E6(R1): Guideline for Good Clinical Practice (June 1996)
- ICH Harmonised Guideline: Integrated Addendum to ICH E6(R1): Guideline for Good Clinical Practice E6(R2) (November 2016)
- GHTF: Clinical Evidence for IVD Medical Devices - Clinical Performance Studies for *In Vitro* Diagnostic Medical Devices (November 2012)
- WMA Declaration of Helsinki, Ethical Principles for Medical Research Involving Human Subjects
- Guidance for Industry and FDA Staff - Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 2005)
- Guidance for Industry and FDA Staff - Off-The-Shelf Software Use in Medical Devices (September 2019)
- Final Guidance for Industry and FDA Staff - General Principles of Software Validation (January 2002)
- Guidance for Industry and FDA Staff - Content of Premarket Submissions for Management of Cybersecurity in Medical Devices (October 2014)
# Standards
- ISO 14971:2019 Medical devices - Application of risk management to medical devices
- IEC 62366-1:2015 Medical devices - Part 1: Application of usability engineering to medical devices
- ISO 62304 Edition 1.1b:2015 / EN 62304:2006 + A1:2015 - Medical device software - software lifecycle processes
- ISO 15223-1:2016: Medical Devices - Symbols to be used with medical device labels, labeling and information to be supplied - Part 1: General requirements
- ISO 13485:2016/EN ISO 13485:2016; Medical devices - Quality Management System - Requirements for regulatory purposes
- ISO 20916:2019; *In vitro* diagnostic medical devices. Clinical performance studies using
K232954 - Page 13 of 75
{13}
specimens from human subjects. Good study practice
- EN 13612:2002; Performance evaluation of in vitro diagnostic medical devices
- EN ISO 18113-1:2011; In vitro diagnostic medical devices - Information supplied by the manufacturer (labeling). Terms, definition and general requirements
- EN ISO 18113-2:2011; In vitro diagnostic medical devices - Information supplied by the manufacturer (labeling) – Part 2: In vitro diagnostic reagents for professional use
- EN ISO 23640:2015; In vitro diagnostic medical devices - Evaluation of stability of in vitro diagnostic reagents
- EN 13641:2002; Elimination or reduction of risk of infection related to in vitro diagnostics reagents
# VII Performance Characteristics (if/when applicable):
# A Analytical Performance:
# 1. Precision/Reproducibility:
The reproducibility and precision of the BIOFIRE SPOTFIRE R/ST Panel on different BIOFIRE SPOTFIRE Systems was evaluated with multiple operators who tested positive and negative samples over multiple days at three study sites, in addition to BioFire Diagnostics. All analytes detected by the panel were evaluated using contrived positive specimens at concentrations equivalent to ≤ 3X their respective analytical limit of detection (LoD) (Table 3). Negative samples were prepared using artificial matrix only.
Table 3. Contrived samples used to evaluate the precision/reproducibility of the BIOFIRE SPOTFIRE R/ST Panel
| Sample | Analyte/Strain | Source Identity | Concentration (per mL) |
| --- | --- | --- | --- |
| 1 | None | -- | -- |
| 2 | *Bordetella pertussis* A639 | ZeptoMetrix 0801459 | 990 CFU |
| | *Mycoplasma pneumoniae* M129 | ZeptoMetrix 0801579 | 10 CCU |
| | Adenovirus Species B Type 3 | ZeptoMetrix 0810062CF | 2.4 TCID_{50} |
| | Coronavirus NL63 | ZeptoMetrix 0810228CF | 0.0025 TCID_{50} |
| | Influenza B B/Florida/02/06 | ZeptoMetrix 0810037CF | 0.099 TCID_{50} |
| | Parainfluenza virus 4a | ZeptoMetrix 0810060CF | 200 TCID_{50} |
| 3 | *Bordetella parapertussis* E595 | ZeptoMetrix 0801462 | 120 CFU |
| | *Chlamydia pneumoniae* AR-39 | ATCC 53592 | 20 IFU |
| | Human metapneumovirus 3 Type B1 | ZeptoMetrix 0810156CF | 0.75 TCID_{50} |
| | Parainfluenza virus 1 | ZeptoMetrix 0810014CF | 4.6 TCID_{50} |
K232954 - Page 14 of 75
{14}
| Sample | Analyte/Strain | Source Identity | Concentration (per mL) |
| --- | --- | --- | --- |
| | Influenza A H1N1pdm Michigan/45/15 | ZeptoMetrix 0810538CF | 2.5 TCID_{50} |
| | Coronavirus SARS-CoV-2 2019-nCoV/USA-WA1/2020 | ATCC VR-1986HK | 250 copies |
| 4 | *Streptococcus dysgalactiae* subsp. *equisimilis* X068 | ZeptoMetrix 0810516 | 330 CFU |
| | Human enterovirus D68 US/MO/14-18947 | ATCC VR-1823 | 11 TCID_{50} |
| | Parainfluenza virus 2 | ZeptoMetrix 0810015CF | 42 TCID_{50} |
| | Respiratory syncytial virus A 2006 | ZeptoMetrix 0810040ACF | 0.062 TCID_{50} |
| | Coronavirus 229E | ATCC VR-740 | 2.0 TCID_{50} |
| 5 | Parainfluenza virus 3 | ZeptoMetrix 0810016CF | 8.8 TCID_{50} |
| | Influenza A H3N2 Hong Kong/4801/14 | ZeptoMetrix 0810526CF | 2.6 TCID_{50} |
| | Coronavirus OC43 | ZeptoMetrix 0810024CF | 1.6 x 10^{-2} TCID_{50} |
| | *Streptococcus pyogenes* Typing strain T1 | ATCC 12344 | 1,400 cells |
CCU: Color Changing Units; CFU: Colony Forming Units; IFU: Inclusion Forming Units; TCID$_{50}$: Tissue Culture Infectious Dose-50%
At BioFire Diagnostics, testing was performed on three BIOFIRE SPOTFIRE Systems, using three lots of reagents over five consecutive days. On each day of testing, two operators each tested three replicates of each sample (one per reagent lot) on each BIOFIRE SPOTFIRE System for a total of 90 replicates per sample overall (2 operators x 3 samples x 3 instruments x 5 days = 90 replicates).
At the external study sites, testing was performed with a single BIOFIRE SPOTFIRE System at each site and occurred over five non-consecutive days using a single lot of reagents that was common across each of the sites. On each day, two operators each tested two replicates of each sample for a total of 60 replicates per sample overall (2 operators x 2 samples x 3 sites x 5 days = 60 replicates).
For each analyte there were 150 positive and 600 negative samples. Samples that produced an invalid result on initial testing due to process control failure (n = 12) or instrument error (n = 2) were retested. Precision and reproducibility for each analyte were calculated by assessing the agreement between the reported test results and the expected results for each sample.
A summary of the results of the study, stratified by analyte and BIOFIRE SPOTFIRE System/site is provided in Table 4. Positive agreement for all analytes was > 95% and there was negligible difference between systems, sites, reagent lots or operators. Negative agreement was 100% in all cases.
K232954 - Page 15 of 75
{15}
Overall, the reproducibility and precision of the BIOFIRE SPOTFIRE R/ST Panel for detection of the target analytes was determined to be acceptable.
Table 4. Summary of results from the Precision and Reproducibility Study with the BIOFIRE SPOTFIRE R/ST Panel, stratified by BIOFIRE SPOTFIRE System and study site
| Assay/Analyte | Expected Result | Agreement (%) Stratified by Site and BIOFIRE SPOTFIRE System | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | BioFire Diagnostics | | | External Site | | | Total | |
| | | A | B | C | 1 | 2 | 3 | | |
| **Adenovirus** Adenovirus Species B Type 3 | Positive | 30/30 (100) | 30/30 (100) | 30/30 (100) | 20/20 (100) | 20/20 (100) | 20/20 (100) | 150/150 (100) | |
| | | 90/90 (100) | | | 60/60 (100) | | | | |
| | | Negative | 120/120 (100) | 120/120 (100) | 120/120 (100) | 80/80 (100) | 80/80 (100) | 80/80 (100) | 600/600 (100) |
| | | | 360/360 (100) | | | 240/240 (100) | | | |
| **Coronavirus (seasonal)** | Coronavirus 229E | Positive | 30/30 (100) | 30/30 (100) | 29/30 (96.7) | 20/20 (100) | 20/20 (100) | 20/20 (100) | 149/150 (99.3) |
| | | | 89/90 (98.9) | | | 60/60 (100) | | | |
| | Coronavirus OC43 | Positive | 30/30 (100) | 30/30 (100) | 30/30 (100) | 20/20 (100) | 20/20 (100) | 20/20 (100) | 150/150 (100) |
| | | | 90/90 (100) | | | 60/60 (100) | | | |
| | Coronavirus NL63 | Positive | 27/30 (90.0) | 30/30 (100) | 30/30 (100) | 18/20 (90.0) | 20/20 (100) | 18/20 (90.0) | 143/150 (95.3) |
| | | | 87/90 (96.7) | | | 56/60 (93.3) | | | |
| | None | Negative | 60/60 (100) | 60/60 (100) | 60/60 (100) | 40/40 (100) | 40/40 (100) | 40/40 (100) | 300/300 (100) |
| | | | 180/180 (100) | | | 120/120 (100) | | | |
| **Coronavirus SARS-CoV-2** SARS-CoV-2 2019-nCoV/USA-WA1/2020 | Positive | 30/30 (100) | 30/30 (100) | 30/30 (100) | 20/20 (100) | 20/20 (100) | 20/20 (100) | 150/150 (100) | |
| | | 90/90 (100) | | | 60/60 (100) | | | | |
| | | Negative | 120/120 (100) | 120/120 (100) | 120/120 (100) | 80/80 (100) | 80/80 (100) | 80/80 (100) | 600/600 (100) |
| | | | 360/360 (100) | | | 240/240 (100) | | | |
| **Human Metapneumovirus** hMPV 3 Type B1 | Positive | 30/30 (100) | 30/30 (100) | 30/30 (100) | 19/20 (95.0) | 20/20 (100) | 19/20 (95.0) | 148/150 (98.7) | |
| | | 90/90 (100) | | | 58/60 (96.7) | | | | |
| | | Negative | 120/120 (100) | 120/120 (100) | 120/120 (100) | 80/80 (100) | 80/80 (100) | 80/80 (100) | 600/600 (100) |
| | | | 360/360 (100) | | | 240/240 (100) | | | |
| **Human Rhinovirus/Enterovirus** Enterovirus D68 US/MO/14-18947 | Positive | 30/30 (100) | 30/30 (100) | 30/30 (100) | 20/20 (100) | 20/20 (100) | 20/20 (100) | 150/150 (100) | |
| | | 90/90 (100) | | | 60/60 (100) | | | | |
| | | Negative | 120/120 (100) | 120/120 (100) | 120/120 (100) | 80/80 (100) | 80/80 (100) | 80/80 (100) | 600/600 (100) |
| | | | 360/360 (100) | | | 240/240 (100) | | | |
| **Influenza A Virus** | Influenza A H1N1pdm | Positive | 30/30 (100) | 30/30 (100) | 30/30 (100) | 20/20 (100) | 20/20 (100) | 20/20 (100) | 150/150 (100) |
| | | | 90/90 (100) | | | 60/60 (100) | | | |
| | Influenza A H3N2 | Positive | 30/30 (100) | 30/30 (100) | 30/30 (100) | 19/20^{3} (95.0) | 19/20 (95.0) | 20/20 (100) | 148/150 (98.7) |
| | | | 90/90 (100) | | | 58/60 (96.7) | | | |
| | None | Negative | 90/90 (100) | 90/90 (100) | 90/90 (100) | 60/60 (100) | 60/60 (100) | 60/60 (100) | 450/450 (100) |
| | | | 270/270 (100) | | | 180/180 (100) | | | |
K232954 - Page 16 of 75
{16}
| Assay/Analyte | Expected Result | Agreement (%) Stratified by Site and BIOFIRE SPOTFIRE System | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | BioFire Diagnostics | | | External Site | | | Total | |
| | | A | B | C | 1 | 2 | 3 | | |
| **Influenza A Virus Subtype H1-2009**^{1} Influenza A H1N1pdm Michigan/45/15 | Positive | 30/30 (100) | 30/30 (100) | 30/30 (100) | 20/20 (100) | 20/20 (100) | 20/20 (100) | 150/150 (100) | |
| | | 90/90 (100) | | | 60/60 (100) | | | | |
| | Negative | 120/120 (100) | 120/120 (100) | 120/120 (100) | 80/80 (100) | 80/80 (100) | 80/80 (100) | 600/600 (100) | |
| | | 360/360 (100) | | | 240/240 (100) | | | | |
| **Influenza A Virus Subtype H3**^{2} Influenza A H3N2 Hong Kong/4801/14 | Positive | 30/30 (100) | 30/30 (100) | 30/30 (100) | 19/20^{3} (95.0) | 19/20 (95.0) | 20/20 (100) | 148/150 (98.7) | |
| | | 90/90 (100) | | | 58/60 (96.7) | | | | |
| | Negative | 120/120 (100) | 120/120 (100) | 120/120 (100) | 80/80 (100) | 80/80 (100) | 80/80 (100) | 600/600 (100) | |
| | | 360/360 (100) | | | 240/240 (100) | | | | |
| **Influenza B Virus** Influenza B B/Florida/02/06 | Positive | 29/30 (96.7) | 30/30 (100) | 30/30 (100) | 20/20 (100) | 20/20 (100) | 20/20 (100) | 149/150 (99.3) | |
| | | 89/90 (98.9) | | | 60/60 (100) | | | | |
| | Negative | 120/120 (100) | 120/120 (100) | 120/120 (100) | 80/80 (100) | 80/80 (100) | 80/80 (100) | 600/600 (100) | |
| | | 360/360 (100) | | | 240/240 (100) | | | | |
| **Parainfluenza Virus** | Parainfluenza Virus 1 | Positive | 30/30 (100) | 30/30 (100) | 30/30 (100) | 20/20 (100) | 20/20 (100) | 20/20 (100) | 150/150 (100) |
| | | | 90/90 (100) | | | 60/60 (100) | | | |
| | Parainfluenza Virus 2 | Positive | 30/30 (100) | 29/30 (96.7) | 28/30 (93.3) | 20/20 (100) | 20/20 (100) | 20/20 (100) | 147/150 (98.0) |
| | | | 87/90 (96.7) | | | 60/60 (100) | | | |
| | Parainfluenza Virus 3 | Positive | 28/30 (93.3) | 30/30 (100) | 30/30 (100) | 19/20 (95.5) | 18/20 (90.0) | 20/20 (100) | 145/150 (96.7) |
| | | | 88/90 (97.8) | | | 57/60 (95.0) | | | |
| | Parainfluenza Virus 4 | Positive | 29/30 (96.7) | 30/30 (100) | 30/30 (100) | 20/20 (100) | 20/20 (100) | 20/20 (100) | 149/150 (99.3) |
| | | | 89/90 (98.9) | | | 60/60 (100) | | | |
| | None | Negative | 30/30 (100) | 30/30 (100) | 30/30 (100) | 20/20 (100) | 20/20 (100) | 20/20 (100) | 150/150 (100) |
| | | | 90/90 (100) | | | 60/60 (100) | | | |
| **Respiratory Syncytial Virus** RSV A 2006 | Positive | 30/30 (100) | 30/30 (100) | 30/30 (100) | 20/20 (100) | 19/20 (95.0) | 20/20 (100) | 149/150 (99.3) | |
| | | 90/90 (100) | | | 59/60 (98.3) | | | | |
| | Negative | 120/120 (100) | 120/120 (100) | 120/120 (100) | 80/80 (100) | 80/80 (100) | 80/80 (100) | 600/600 (100) | |
| | | 360/360 (100) | | | 240/240 (100) | | | | |
| **Bordetella parapertussis** *B. parapertussis* E595 | Positive | 30/30 (100) | 30/30 (100) | 30/30 (100) | 20/20 (100) | 20/20 (100) | 20/20 (100) | 150/150 (100) | |
| | | 90/90 (100) | | | 60/60 (100) | | | | |
| | Negative | 120/120 (100) | 120/120 (100) | 120/120 (100) | 80/80 (100) | 80/80 (100) | 80/80 (100) | 600/600 (100) | |
| | | 360/360 (100) | | | 240/240 (100) | | | | |
| **Bordetella pertussis** *B. pertussis* A639 | Positive | 29/30 (96.7) | 29/30 (96.7) | 30/30 (100) | 20/20 (100) | 20/20 (100) | 20/20 (100) | 148/150 (98.7) | |
| | | 88/90 (97.8) | | | 60/60 (100) | | | | |
| | Negative | 120/120 (100) | 120/120 (100) | 120/120 (100) | 80/80 (100) | 80/80 (100) | 80/80 (100) | 600/600 (100) | |
| | | 360/360 (100) | | | 240/240 (100) | | | | |
K232954 - Page 17 of 75
{17}
| Assay/Analyte | Expected Result | Agreement (%) Stratified by Site and BIOFIRE SPOTFIRE System | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | BioFire Diagnostics | | | External Site | | | Total |
| | | A | B | C | 1 | 2 | 3 | |
| *Chlamydia pneumoniae* *Chlamydia pneumoniae* AR-39 | Positive | 30/30 (100) | 30/30 (100) | 30/30 (100) | 20/20 (100) | 20/20 (100) | 20/20 (100) | 150/150 (100) |
| | | 90/90 (100) | | | 60/60 (100) | | | |
| | Negative | 120/120 (100) | 120/120 (100) | 120/120 (100) | 80/80 (100) | 80/80 (100) | 80/80 (100) | 600/600 (100) |
| | | 360/360 (100) | | | 240/240 (100) | | | |
| *Mycoplasma pneumoniae* *M. pneumoniae* M129 | Positive | 29/30 (96.7) | 30/30 (100) | 30/30 (100) | 20/20 (100) | 20/20 (100) | 20/20 (100) | 149/150 (99.3) |
| | | 89/90 (98.9) | | | 60/60 (100) | | | |
| | Negative | 120/120 (100) | 120/120 (100) | 120/120 (100) | 80/80 (100) | 80/80 (100) | 80/80 (100) | 600/600 (100) |
| | | 360/360 (100) | | | 240/240 (100) | | | |
| *Streptococcus dysgalactiae* *S. dysgalactiae* subsp. *equisimilis* Z068 | Positive | 30/30 (100) | 29/30 (96.7) | 27/30 (90.0) | 20/20 (100) | 20/20 (100) | 20/20 (100) | 146/150 (97.3) |
| | | 86/90 (95.6) | | | 60/60 (100) | | | |
| | Negative | 120/120 (100) | 120/120 (100) | 120/120 (100) | 80/80 (100) | 80/80 (100) | 80/80 (100) | 600/600 (100) |
| | | 360/360 (100) | | | 240/240 (100) | | | |
| *Streptococcus pyogenes* *S. pyogenes* Typing strain T1 | Positive | 30/30 (100) | 30/30 (100) | 30/30 (100) | 20/20 (100) | 19/20 (95.0) | 20/20 (100) | 149/150 (99.3) |
| | | 90/90 (100) | | | 59/60 (98.3) | | | |
| | Negative | 120/120 (100) | 120/120 (100) | 120/120 (100) | 80/80 (100) | 80/80 (100) | 80/80 (100) | 600/600 (100) |
| | | 360/360 (100) | | | 240/240 (100) | | | |
| **Positive Agreement By System/Site** | **681/690 (98.7)** | **687/690 (99.6)** | **684/690 (99.1)** | **454/460 (98.7)** | **454/460 (98.7)** | **457/460 (99.3)** | **3417/ 3450 (99.0)** | |
| | | **2052/2070 (99.1)** | | | **1365/1380 (98.9)** | | | |
1 1 positive sample was initially reported as Positive: Influenza A Virus (No Subtype Identified) and was therefore retested. The retest result was as expected. The retest results from this sample for each analyte are included in the summary table.
2 2 positive samples were initially reported as Positive: Influenza A Virus (No Subtype Identified) and 1 positive sample was initially reported as Uncertain: Influenza A Virus. All 3 samples were retested and the retest results were as expected. The retest results from these samples for each analyte are included in the summary table.
3 1 sample was reported as Uncertain Influenza A Virus but was not retested.
# 2. Linearity:
Not applicable.
# 3. Analytical Specificity/Interference:
# Specificity (Exclusivity)
The analytical specificity (exclusivity) of the BIOFIRE SPOTFIRE R/ST Panel was evaluated through a combination of in silico analysis and laboratory testing of on- and off-panel species, phylogenetically related viruses and bacteria and other microorganisms that may be found in the respiratory tract.
# In silico Analysis
In silico analysis of assay specificity/exclusivity was performed by conducting searches with whole amplicon and paired inner primer sequences using the NCBI Basic Local Alignment Search Tool (BLAST) "blastn" tool. The location and orientation of regions of
K232954 - Page 18 of 75
{18}
homology with the primer sequences were evaluated to determine the potential for amplification. A summary of the results from the analysis is provided in Table 5. The BIOFIRE SPOTFIRE R/ST panel assays were shown to be specific for their respective analytes with the following exceptions, which are noted in the device labeling:
- Cross-reaction of the SARS-CoV-2 assays with closely-related bat and pangolin coronavirus sequences
- Cross-reaction of the Human Rhinovirus/Enterovirus assay with B. pertussis, B. parapertussis and B. bronchiseptica
- Cross-reaction of the B. parapertussis assay with strains of B. bronchiseptica that carry IS1001
Table 5. Summary of in silico BLAST analysis to evaluate exclusivity
| Viruses | |
| --- | --- |
| Analyte | Predicted Cross-reaction |
| Adenovirus | None |
| Coronavirus (seasonal) | None^{1} |
| Coronavirus SARS-CoV-2 | Bat coronavirus Pangolin coronavirus Bat SARS-like coronavirus |
| Human Metapneumovirus | None |
| Human Rhinovirus/Enterovirus | Bovine picornavirus Canine picornavirus B. pertussis, B. parapertussis, B. bronchiseptica |
| Influenza A Virus | None^{2} |
| Influenza A H1-2009 | Swine Influenza A/H1 sequences Some sequences of avian Influenza A/H1N2 |
| Influenza A H3 | None |
| Influenza B Virus | None |
| Parainfluenza Virus | None |
| Respiratory Syncytial Virus | None |
| Bacteria | |
| Analyte | Predicted Cross-reaction |
| Bordetella parapertussis | Strains of B. bronchiseptica carrying IS1001 |
| Bordetella pertussis | None |
| Chlamydia pneumoniae | Chlamydia gallinacea |
| Mycoplasma pneumoniae | None |
| Streptococcus dysgalactiae subsp. equisimilis | None |
| Streptococcus pyogenes | None |
$^{1}$ Some zoonotic coronaviruses may cross-react but these are not expected to be found in human clinical specimens.
$^{2}$ A sequence from Protopolystoma xenopodis (a parasitic worm found in frogs) exhibited significant homology to the pan-influenza A PCR primers but is not expected to be found in human clinical specimens.
### Laboratory Testing of Analytical Specificity
The analytical specificity (exclusivity) of the BIOFIRE SPOTFIRE R/ST Panel was also evaluated by testing high concentrations of on- and off-panel viruses, bacteria and other microorganisms that are phylogenetically related or may be found in nasopharyngeal and/or throat swab specimens. Analysis of on-panel analytes was performed to evaluate the potential
K232954 - Page 19 of 75
{19}
for intra-panel cross-reactivity (i.e., detection of an analyte by an assay other than the target-specific assay for that analyte). Analysis of off-panel analytes was performed to evaluate the potential for cross-reaction with other respiratory flora that are not intended to be detected by the panel. Each on-panel species was tested in triplicate and the results are shown in Tables 6 and 7. The expected results were obtained with each on-panel virus and no cross-reaction was observed (Table 6). Based on in silico analysis, it was predicted that B. parapertussis and B. pertussis would produce positive results with the Human Rhinovirus/Enterovirus assay. B. pertussis produced false-positive results at 1.3 x 10¹⁰ CFU/mL but was reported negative when tested at 1.3 x 10⁹ CFU/mL (Table 7). False-positive results for Human Rhinovirus/Enterovirus were not observed with B. parapertussis at the concentration tested (4.6 x 10⁹ CFU/mL). However, the potential for cross-reaction of the human Rhinovirus/Enterovirus assay with both B. parapertussis and B. pertussis based on the in silico prediction is noted in the device labeling.
Table 6. Results from cross-reactivity testing with “on-panel” viruses
| Species | Source Identity | Concentration (per mL) | Observed Cross-reaction |
| --- | --- | --- | --- |
| Adenovirus A | ZeptoMetrix 0810073CF | 1.4 x 10⁵ TCID₅₀ | None |
| Adenovirus B | ZeptoMetrix 0810062CF | 1.2 x 10⁷ TCID₅₀ | None |
| Adenovirus C | ZeptoMetrix 0810110CF | 2.2 x 10⁶ TCID₅₀ | None |
| Adenovirus D | ZeptoMetrix 0810119CF | 1.7 x 10⁵ TCID₅₀ | None |
| Adenovirus E | ZeptoMetrix 081070CF | 1.5 x 10⁵ TCID₅₀ | None |
| Adenovirus F | ZeptoMetrix 0810085CF | 1.1 x 10⁶ TCID₅₀ | None |
| Coronavirus 229E | ATCC VR-740 | 8.9 x 10⁶ TCID₅₀ | None |
| Coronavirus HKU1 | Clinical specimen-1 ¹ | 4.5 x 10⁷ copies | None |
| | Clinical specimen-2 ² | | None ³ |
| Coronavirus NL63 | ZeptoMetrix 0810228CF | 5.0 x 10⁵ TCID₅₀ | None |
| Coronavirus OC43 | ZeptoMetrix 0810024CF | 3.6 x 10⁵ TCID₅₀ | None |
| Coronavirus SARS-CoV-2 | ATCC VR-1986HK | 7.6 x 10⁷ copies | None |
| Enterovirus D68 | ATCC VR-1823 | 1.6 x 10⁷ TCID₅₀ | None |
| Human Metapneumovirus A1 | ZeptoMetrix 0810161CF | 2.5 x 10⁵ TCID₅₀ | None |
| Human Metapneumovirus A2 | ZeptoMetrix 0810164CF | 3.6 x 10⁵ TCID₅₀ | None |
| Human Metapneumovirus B1 | ZeptoMetrix 0810156CF | 1.6 x 10⁴ TCID₅₀ | None |
| Human Metapneumovirus B2 | ZeptoMetrix 0810162CF | 1.3 x 10⁶ TCID₅₀ | None |
| Rhinovirus A1 | ZeptoMetrix 0810012CFN | 1.3 x 10⁶ TCID₅₀ | None |
| Influenza A H1N1pdm09 | ZeptoMetrix 08100538CF | 1.4 x 10⁵ TCID₅₀ | None |
| Influenza A H3N2 | ZeptoMetrix 010526CF | 7.2 x 10⁵ TCID₅₀ | None |
| Influenza B (Victoria) | BEI NR-44023 | 2.8 x 10⁸ CEID₅₀ | None |
| | ZeptoMetrix 0810037CF | 2.5 x 10⁵ TCID₅₀ | None |
| Influenza B (Yamagata) | ZeptoMetrix 0810256CF | 2.1 x 10⁴ TCID₅₀ | None |
| Parainfluenza Virus 1 | ZeptoMetrix 0810014CF | 4.2 x 10⁵ TCID₅₀ | None |
| Parainfluenza Virus 2 | ZeptoMetrix 0810015CF | 1.2 x 10⁷ TCID₅₀ | None |
| Parainfluenza Virus 3 | ZeptoMetrix 0810016CF | 3.4 x 10⁷ TCID₅₀ | None |
| Parainfluenza Virus 4a | ZeptoMetrix 0810060CF | 3.4 x 10⁷ TCID₅₀ | None |
| Respiratory Syncytial Virus A | ZeptoMetrix 0810040ACF | 4.2 x 10⁵ TCID₅₀ | None |
| Respiratory Syncytial Virus B | ZeptoMetrix 0810479CF | 4.2 x 10⁵ TCID₅₀ | None |
CEID₅₀: Chicken Egg Infectious Dose-50%; TCID₅₀: Tissue Culture Infectious Dose-50%
¹ Specimen previously reported Positive for Parainfluenza Virus 4 in addition to Coronavirus HKU1.
² Specimen previously reported Positive for Adenovirus in addition to Coronavirus HKU1.
³ 1/3 replicates reported Positive for Adenovirus (not considered to indicate cross-reactivity based on previous test results for the source specimen).
K232954 - Page 20 of 75
{20}
Table 7. Results from cross-reactivity testing with “on-panel” bacteria
| Species | Source Identity | Concentration (per mL) | Observed Cross-reaction |
| --- | --- | --- | --- |
| Bordetella parapertussis | ZeptoMetrix 0801462 | 4.6 x 10⁹ CFU | None¹ |
| Bordetella pertussis | ZeptoMetrix 0801459 | 1.3 x 10¹⁰ CFU | Human Rhinovirus/Enterovirus² |
| | | 1.3 x 10⁹ CFU | None² |
| Chlamydia pneumoniae | ATCC 53592 | 2.9 x 10⁷ IFU | None |
| Mycoplasma pneumoniae | ZeptoMetrix 0801579 | 2.5 x 10⁷ CCU | None |
| Streptococcus dysgalactiae subsp. equisimilis | ZeptoMetrix 0801516 | 7.8 x 10⁸ CFU | None |
| Streptococcus pyogenes | ATCC 12344 | 3.4 x 10⁸ cells | None |
CCU: Color Changing Units; CFU: Colony Forming Units; IFU: Inclusion Forming Units
¹ In silico analysis predicted detection by the Human Rhinovirus/Enterovirus assay but none was observed.
² Cross-reaction predicted by in silico analysis and confirmed by laboratory testing at ≥ 1.3 x 10¹⁰ CFU/mL.
The potential for cross-reaction with off-panel analytes was evaluated by testing 87 strains of bacteria, 2 species of yeast and 7 viruses that are either phylogenetically related to the on-panel analytes or which are commonly found in the respiratory tract (Table 8). Included in the study were isolates of S. dysgalactiae subsp. equisimilis of animal origin and S. dysgalactiae subsp. dysgalactiae, both of which lack the region of the S. dysgalactiae genome targeted by the BIOFIRE SPOTFIRE R/ST Panel. No cross-reaction was observed other than the expected detection of one strain of B. bronchiseptica by the assay for B. parapertussis.
Table 8. Results from cross-reactivity testing with “off-panel” bacteria, viruses and fungi
| Bacteria | | | |
| --- | --- | --- | --- |
| Species | Source Identity | Concentration (per mL) | Observed Cross-reaction |
| Arcanobacterium bernardiae | ATCC BAA-441 | 1.6 x 10⁹ cells | None |
| Arcanobacterium haemolyticum | ATCC 9345 | 1.5 x 10⁸ cells | None |
| Arcanobacterium pyogenes | ATCC 49698 | 6.7 x 10⁹ cells | None |
| Bacillus cereus | ATCC 7064 | 8.3 x 10⁹ cells | None |
| Bordetella bronchiseptica | ATCC 10580 | 8.3 x 10⁹ cells | None |
| | ATCC 4617 | 7.9 x 10⁹ cells | None |
| | ATCC 19395 | 7.9 x 10⁹ cells | None |
| | NRRL B-59914 | 7.1 x 10⁹ cells | None |
| | NRRL B-59909 | 28 cells | B. parapertussis¹ |
| | | 2.8 cells | B. parapertussis¹ |
| Bordetella holmesii | ATCC 700052 | 8.3 x 10⁹ cells | None |
| Burkholderia cepacia | ATCC 51671 | 7.9 x 10⁹ cells | None |
| Campylobacter rectus | ATCC 33238 | 7.6 x 10⁷ cells | None |
| Chlamydia trachomatis | ZeptoMetrix 0801775 | 1.3 x 10⁸ IFU | None |
| Corynebacterium diphtheriae | ATCC 27010 | 8.0 x 10⁹ cells | None |
| Corynebacterium pseudodiphtheriticum | ATCC 10700 | 8.7 x 10⁹ cells | None |
| Enterococcus casseliflavus | ATCC 49605 | 8.0 x 10⁹ cells | None |
| Enterococcus faecalis | ZeptoMetrix 0801637 | 8.0 x 10⁹ cells | None |
| Escherichia coli | ATCC BAA-2196 | 7.2 x 10⁹ cells | None |
| Fusobacterium necrophorum subsp. funduliforme | ATCC 51357 | 4.4 x 10⁸ cells | None |
| Fusobacterium nucleatum | ATCC 25586 | 4.9 x 10⁸ cells | None |
| Fusobacterium varium | ATCC 27725 | 1.6 x 10⁸ cells | None |
| Gemella haemolysans | ATCC 10379 | 4.0 x 10⁹ cells | None |
| Gemella morbillorum | ATCC 27824 | 1.0 x 10⁸ cells | None |
| Granulicatella adiacens | ATCC 49175 | 1.3 x 10⁹ cells | None |
K232954 - Page 21 of 75
{21}
| Bacteria | | | |
| --- | --- | --- | --- |
| Species | Source Identity | Concentration (per mL) | Observed Cross-reaction |
| Haemophilus influenzae | ATCC 10211 | 8.3 x 10⁹ cells | None |
| Haemophilus parahaemolyticus | ATCC 49700 | 8.7 x 10⁹ cells | None |
| Klebsiella pneumoniae | CDC AR-BANK#0115 | 7.3 x 10⁹ cells | None |
| Lactobacillus rhamnosus | ATCC 7469 | 7.9 x 10⁹ cells | None |
| Lactobacillus lactis | ATCC 29146 | 6.2 x 10⁹ cells | None |
| Legionella pneumophila | ATCC 33215 | 7.0 x 10⁹ cells | None |
| Leptotrichia buccalis | ATCC 14201 | 4.4 x 10⁸ cells | None |
| Moraxella catarrhalis | ATCC 43627 | 7.2 x 10⁹ cells | None |
| Mycobacterium tuberculosis | ZeptoMetrix 0801660 | 6.1 x 10⁶ cells | None |
| Mycoplasma buccale | Mycoplasma Experience NC10136 | 1.4 x 10⁷ CFU | None |
| Mycoplasma faucium | Mycoplasma Experience NC10174 | 1.4 x 10⁶ CFU | None |
| Mycoplasma fermentans | Mycoplasma Experience NC10117 | 2.8 x 10⁷ CFU | None |
| Mycoplasma genitalium | Mycoplasma Experience NC10195 | 1.8 x 10⁶ CFU | None |
| Mycoplasma hominis | Mycoplasma Experience NC10111 | 1.2 x 10⁷ CFU | None |
| Mycoplasma lipophilum | Mycoplasma Experience NC10173 | 1.5 x 10⁶ CFU | None |
| Mycoplasma orale | Mycoplasma Experience NC10112 | 2.2 x 10⁷ CFU | None |
| Mycoplasma salivarium | Mycoplasma Experience NC10113 | 4.4 x 10⁶ CFU | None |
| Neisseria elongata | ATCC 25295 | 8.5 x 10⁹ cells | None |
| Neisseria gonorrhoeae | ZeptoMetrix 0801482 | 4.9 x 10⁷ CFU | None |
| Neisseria lactamica | ATCC 23971 | 2.7 x 10⁹ cells | None |
| Neisseria meningitidis | ATCC 13113 | 7.4 x 10⁹ cells | None |
| Neisseria sicca | ATCC 9913 | 7.2 x 10⁹ cells | None |
| Neisseria subflava | ATCC 49275 | 8.0 x 10⁹ cells | None |
| Parvimonas micra | ATCC 33270 | 6.0 x 10⁷ cells | None |
| Pneumocystis carinii | ATCC PRA-159 | 1.0 x 10⁷ nuclei | None |
| Porphyromonas endodontalis | ATCC 35406 | 1.6 x 10⁷ cells | None |
| Porphyromonas gingivalis | ATCC BAA-308 | 5.0 x 10⁸ cells | None |
| Prevotella histicola | BEI HM-471 | 9.0 x 10⁸ cells | None |
| Prevotella melaninogenica | ATCC 25845 | 6.9 x 10⁸ cells | None |
| Prevotella oralis | ATCC 33322 | 6.2 x 10⁸ cells | None |
| Pseudomonas aeruginosa | CDC AR-BANK#0092 | 8.3 x 10⁹ cells | None |
| Rhodococcus equi | ATCC 33706 | 7.3 x 10⁹ cells | None |
| Serratia marcescens | ATCC 27137 | 8.9 x 10⁹ cells | None |
| Staphylococcus aureus | ATCC BAA-1700 | 7.4 x 10⁹ cells | None |
| Staphylococcus epidermidis | ATCC 12228 | 8.0 x 10⁹ cells | None |
| Staphylococcus haemolyticus | ATCC 29968 | 8.0 x 10⁹ cells | None |
| Staphylococcus intermedius | ATCC 29663 | 8.2 x 10⁹ cells | None |
| Staphylococcus saprophyticus | ATCC 15305 | 8.1 x 10⁹ cells | None |
| Streptococcus agalactiae | ATCC 13813 | 6.0 x 10⁹ cells | None |
| Streptococcus anginosus | ATCC 700231 | 7.1 x 10⁹ cells | None |
| Streptococcus constellatus subsp. pharyngitis | NCTC 13122 | 5.6 x 10⁸ cells | None |
| Streptococcus dysgalactiae subsp. dysgalactiae ² | ATCC 43078 | 6.7 x 10⁹ cells | None |
| | NCTC 4669 | 7.4 x 10⁹ cells | None |
| | NCTC 4335 | 8.4 x 10⁹ cells | None |
| | NCTC 4670 | 6.6 x 10⁹ cells | None |
| | CCUG 27665 | 7.4 x 10⁹ cells | None |
| | CCUG 28112 | 6.7 x 10⁹ cells | None |
| | CCUG 28114 | 7.5 x 10⁹ cells | None |
| Streptococcus dysgalactiae subsp. equisimilis (pig) ² | CCUG 28117 | 7.1 x 10⁹ cells | None |
| Streptococcus dysgalactiae subsp. equisimilis (horse) ² | CCUG 27664 | 7.5 x 10⁹ cells | None |
| | ATCC 10009 | 6.9 x 10⁹ cells | None |
| Streptococcus gallolyticus | ATCC 43143 | 2.8 x 10⁹ cells | None |
| Streptococcus gordonii | ATCC 10558 | 4.5 x 10⁹ cells | None |
| Streptococcus intermedius | ATCC 27335 | 2.9 x 10⁹ cells | None |
| Streptococcus mitis | ATCC 15914 | 3.2 x 10⁹ cells | None |
| Streptococcus mutans | ATCC 25175 | 2.3 x 10⁹ cells | None |
| Streptococcus oralis | ATCC 10557 | 1.1 x 10⁹ cells | None |
K232954 - Page 22 of 75
{22}
| Bacteria | | | |
| --- | --- | --- | --- |
| Species | Source Identity | Concentration (per mL) | Observed Cross-reaction |
| Streptococcus parasanguinis | ATCC 15912 | 7.8 x 10⁹ cells | None |
| Streptococcus pneumoniae | ATCC 49619 | 2.5 x 10⁸ cells | None |
| Streptococcus salivarius | ATCC 13419 | 6.6 x 10⁹ cells | None |
| Streptococcus sanguinis | ATCC 10556 | 1.1 x 10⁹ cells | None |
| Tannerella forsythia | ATCC BAA-2717 | 2.6 x 10⁸ cells | None |
| Treponema denticola | ATCC 33520 | 2.2 x 10⁸ cells | None |
| Ureaplasma urealyticum | ATCC 27618 | 5.7 x 10⁷ cells | None |
| Veillonella parvula | ATCC 10790 | 4.7 x 10⁸ cells | None |
| Yeast/Fungi | | | |
| --- | --- | --- | --- |
| Species | Source Identity | Concentration (per mL) | Observed Cross-reaction |
| Candida albicans | ATCC MYA-2876 | 2.8 x 10⁸ cells | None |
| Saccharomyces cerevisiae | ATCC 18824 | 1.9 x 10⁸ cells | None |
| Viruses | | | |
| --- | --- | --- | --- |
| Species | Source Identity | Concentration (per mL) | Observed Cross-reaction |
| Cytomegalovirus | ZeptoMetrix 0810003CF | 1.9 x 10⁵ TCID₅₀ | None |
| Epstein Barr Virus | ZeptoMetrix 0810008CF | 5.9 x 10⁶ copies | None |
| Human Herpes Simplex Virus 1 | ATCC VR-260 | 8.9 x 10⁶ TCID₅₀ | None |
| Measles Virus | ZeptoMetrix 0810025CF | 2.5 x 10⁵ TCID₅₀ | None |
| Middle East Respiratory Syndrome Coronavirus | ZeptoMetrix 0810575CFHI | 1.2 x 10⁵ TCID₅₀ | None |
| Mumps virus | ZeptoMetrix 0810079CF | 2.0 x 10⁵ TCID₅₀ | None |
| Severe Acute Respiratory Syndrome Coronavirus | BEI NR-52346 | 5.3 x 10⁵ genomes | None |
CFU: Colony Forming Units; IFU: Inclusion Forming Units; TCID₅₀: Tissue Culture Infectious Dose-50%
¹ B. bronchiseptica NRRL B-59909 carries IS1001 which is the target for the BIOFIRE SPOTFIRE R/ST Panel B. parapertussis assay. Positive results were obtained with all target levels ≥ 28 cells/mL and with 1/3 replicates at 2.8 cells/mL.
² The BIOFIRE SPOTFIRE R/ST Panel assay for S. dysgalactiae is intended to detect strains of S. dysgalactiae subsp. equisimilis of human origin. Detection of S. dysgalactiae subsp. dysgalactiae or strains of S. dysgalactiae subsp. equisimilis of non-human origin is not expected.
Based on a combination of the in silico and laboratory testing data, the potential for cross-reaction of the BIOFIRE SPOTFIRE R/ST Panel with certain on- and off-panel analytes is noted in the device labeling, as summarized in Table 9.
Table 9. Summary of known or predicted cross-reactivity
| SPOTFIRE R/ST Panel Assay | Cross-reactive Species | Description |
| --- | --- | --- |
| SARS-CoV-2 | Bat coronavirus Pangolin coronavirus Bat SARS-like coronavirus | Predicted cross-reaction with closely related sarbecoviruses isolated from bats and pangolin |
| Human Rhinovirus/Enterovirus | Bordetella bronchiseptica Bordetella parapertussis Bordetella pertussis Bovine and canine picornaviruses | Homologous sequences in the listed organisms may be detected and reported as Human Rhinovirus/Enterovirus |
| Influenza Virus A/H1-2009 | Influenza A H1N1 (swine) | The FluA-H1-2009 assay may detect influenza A H1 gene sequences of swine origin that will be reported as Influenza A Virus (subtype H1-2009) or as Influenza A (No Subtype Identified). |
K232954 - Page 23 of 75
{23}
| SPOTFIRE R/ST Panel Assay | Cross-reactive Species | Description |
| --- | --- | --- |
| | | depending on the strain and concentration present Certain strains of avian influenza A/H1N2 may also be detected and reported as influenza A/H1-2009 |
| *Bordetella parapertussis* | *Bordetella bronchiseptica* (isolates carrying IS1001) | Some strains of *B. bronchiseptica* carry IS1001, the target for the *B. parapertussis* assay, and may be detected and reported as *B. parapertussis* |
| *Chlamydia pneumoniae* | *Chlamydia gallinacea* | Homologous sequences in *C. gallinacea* may be detected and reported as *C. pneumoniae* |
### Interference Testing
A study was conducted to evaluate the potential for assay interference due to the presence of endogenous and exogenous substances that may be present in upper respiratory specimens. In addition, testing was also performed with representative on- and off-panel microorganisms at very high concentrations to evaluate the potential for competitive interference between analytes on the BIOFIRE SPOTFIRE R/ST Panel. Testing of each potentially interfering substance/microorganism was performed using contrived positive and negative specimens prepared in artificial throat swab matrix. The positive specimens contained representative on-panel analytes at levels close to their respective LoD (Table 10). In a separate study (refer to Sections VII (A)(6) and (B)(2), below) the analytical sensitivity of the BIOFIRE SPOTFIRE R/ST Panel was shown to be similar, irrespective of the specimen matrix (nasopharyngeal or throat swab) and therefore it was considered acceptable to use just one of the two matrices for the evaluation of assay interference. The concentration of each substance tested was chosen to represent a high level that may be found in natural clinical specimens. The results of the study are summarized in Tables 11-16.
Table 10. Representative on-panel analytes used to evaluate assay interference
| Assay | Strain/Source | Characteristics | Concentration (per mL)^{1} | Panel | |
| --- | --- | --- | --- | --- | --- |
| | | | | Respiratory | Sore Throat |
| Adenovirus | Adenovirus B Type 3 ZeptoMetrix 0810062CF | Non-enveloped DNA virus | 2.4 TCID_{50} | Yes | Yes |
| Coronavirus (seasonal) | Coronavirus NL63 ZeptoMetrix 0810228CF | Enveloped RNA virus | 0.099 TCID_{50} | Yes | Yes |
| Human Rhinovirus/Enterovirus | Human Rhinovirus A1 ZeptoMetrix 0810012CFN | Non-enveloped RNA virus | 0.63 TCID_{50} | Yes | Yes |
| Influenza B Virus | Influenza B virus B/Florida/02/06 (Victoria) ZeptoMetrix 0810255CF | Enveloped RNA virus | 60 IFU | Yes | Yes |
| *Chlamydia pneumoniae* | *Chlamydia pneumoniae* AR-39 ATCC 53592 | Gram negative bacterium | 0.0075 TCID_{50} | Yes | Yes |
| *Streptococcus pyogenes* | *Streptococcus pyogenes* SF-120 T-type 1 ATCC 12344 | Gram positive bacterium | 1,400 cells | Not applicable | Yes |
IFU: Inclusion Forming Units; TCID$_{50}$: Tissue Culture Infectious Dose-50%
$^{1}$ 3X LoD
Table 11. Results from testing the effect of endogenous substances on the performance of the BIOFIRE SPOTFIRE R/ST Panel
K232954 - Page 24 of 75
{24}
| Substance | Conc^{a} | Sample Type | Positive Results Reported/Replicates Tested | | | | | | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Viruses | | | | | | | | | | | Bacteria | | | | | |
| | | | Adenovirus | Coronavirus (seasonal) | Coronavirus SARS-CoV-2 | Human Metapneumovirus | Human Rhinovirus / Enterovirus | Influenza A Virus | Influenza A H1/2009 | Influenza A H3 | Influenza B Virus | Parainfluenza Virus | Respiratory Syncytial Virus | B. parapertussis | B. pertussis | C. pneumoniae | M. pneumoniae | S. dysgalactiae | S. pyogenes |
| Whole human blood (sodium citrate) | 10% v/v | Neg | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| | | Pos | 3/3 | 3/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 3/3 |
| Human sputum/mucus | 1% v/v | Neg | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| | | Pos | 3/3 | 3/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 3/3 |
| Human genomic DNA | 20 ng/μL | Neg | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| | | Pos | 5/5 | 5/5 | 0/5 | 0/5 | 4/5^{1} | 0/5 | 0/5 | 0/5 | 5/5 | 0/5 | 0/5 | 0/5 | 0/5 | 5/5 | 0/5 | 0/5 | 5/5 |
Neg: Expected result - Negative for all SPOTFIRE R/ST Panel analytes.
Pos: Expected result - Positive for representative SPOTFIRE R/ST Panel analytes highlighted in red.
Contrived Positive Samples contained the representative on-panel analytes at 3X LoD.
Contrived Negative Samples contained only the potentially interfering substance.
$^{1}$ Initially, 1/3 samples was reported Negative for Human Rhinovirus/Enterovirus; therefore, 2 additional replicates were tested, both of which produced the expected results for all analytes.
None of the endogenous substances tested caused reproducible interference with the BIOFIRE SPOTFIRE R/ST Panel.
Table 12. Results from testing the effect of exogenous substances on the performance of the BIOFIRE SPOTFIRE R/ST Panel
| Substance | Conc^{a} | Sample Type | Positive Results Reported/Replicates Tested | | | | | | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Viruses | | | | | | | | | | | Bacteria | | | | | |
| | | | Adenovirus | Coronavirus (seasonal) | Coronavirus SARS-CoV-2 | Human Metapneumovirus | Human Rhinovirus / Enterovirus | Influenza A Virus | Influenza A H1/2009 | Influenza A H3 | Influenza B Virus | Parainfluenza Virus | Respiratory Syncytial Virus | B. parapertussis | B. pertussis | C. pneumoniae | M. pneumoniae | S. dysgalactiae | S. pyogenes |
| Promethazine hydrochloride | 3.3 x 10^{-4} mg/mL | Neg | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| | | Pos | 3/3 | 3/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 3/3 |
| Acetaminophen (paracetamol) | 1.5 x 10^{-1} | Neg | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
K232954 - Page 25 of 75
{25}
| Substance | Conc* | Sample Type | Positive Results Reported/Replicates Tested | | | | | | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Viruses | | | | | | | | | | | Bacteria | | | | | |
| | | | Adenovirus | Coronavirus (seasonal) | Coronavirus SARS-CoV-2 | Human Metapneumovirus | Human Rhinovirus / Enterovirus | Influenza A Virus | Influenza A H1/2009 | Influenza A H3 | Influenza B Virus | Parainfluenza Virus | Respiratory Syncytial Virus | B. parapertussis | B. pertussis | C. pneumoniae | M. pneumoniae | S. dysgalactiae | S. pyogenes |
| | mg/mL | Pos | 3/3 | 3/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 3/3 |
| Acetylsalicylic acid (aspirin) | 3.0 x 10^{-2} mg/mL | Neg | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| | | Pos | 3/3 | 3/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 3/3 |
| Ibuprofen | 2.2 x 10^{-1} mg/mL | Neg | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| | | Pos | 5/5 | 5/5 | 0/5 | 0/5 | 4/5^{1} | 0/5 | 0/5 | 0/5 | 5/5 | 0/5 | 0/5 | 0/5 | 0/5 | 5/5 | 0/5 | 0/5 | 5/5 |
| Albuterol sulfate | 5.4 x 10^{-1} mg/mL | Neg | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| | | Pos | 3/3 | 3/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 3/3 |
| Triple antibiotic ointment (neomycin/polymyxin B/bacitracin) | 2% w/v | Neg | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| | | Pos | 3/3 | 3/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 3/3 |
| Mucinex Severe Nasal Congestion Relief Clear & Cool Spray | 1% v/v | Neg^{2} | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| | | Pos | 3/3 | 3/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 3/3 |
| Saline nasal spray (0.65% disodium phosphate phenylcarbinol, monosodium phosphate, benzalkonium chloride) | 1% v/v | Neg | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| | | Pos | 3/3 | 3/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 3/3 |
| Vicks VapoRub Cough Suppressant Topical Analgesic (4.8% camphor, 1.2% eucalyptus oil, 2.6% menthol) | 1% w/v | Neg | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| | | Pos | 3/3 | 3/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 3/3 |
| Vaseline petroleum jelly (100% white petrolatum) | 1% w/v | Neg | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| | | Pos | 3/3 | 3/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 3/3 |
| Orajel (0.13% benzalkonium chloride, 20% benzocaine, 0.5% menthol, 0.15% zinc chloride) | 2% w/v | Neg | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| | | Pos | 3/3 | 3/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 3/3 |
| Chloraseptic Sore Throat Spray (1.4% phenol) | 1% v/v | Neg | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| | | Pos | 3/3 | 3/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 3/3 |
| Vicks Formula 44 DM (0.67 mg/mL dextromethorphan hydrobromide, 13 mg/mL guaifenisin) | 1% v/v | Neg | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| | | Pos^{2} | 3/3 | 3/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 3/3 |
K232954 - Page 26 of 75
{26}
| Substance | Conc* | Sample Type | Positive Results Reported/Replicates Tested | | | | | | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Viruses | | | | | | | | | | | Bacteria | | | | | |
| | | | Adenovirus | Coronavirus (seasonal) | Coronavirus SARS-CoV-2 | Human Metapneumovirus | Human Rhinovirus / Enterovirus | Influenza A Virus | Influenza A H1/2009 | Influenza A H3 | Influenza B Virus | Parainfluenza Virus | Respiratory Syncytial Virus | B. parapertussis | B. pertussis | C. pneumoniae | M. pneumoniae | S. dysgalactiae | S. pyogenes |
| Phenylephrine hydrochloride | 1% w/v | Neg | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| | | Pos | 3/3 | 3/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 3/3 |
| Nasal spray (50 μg fluticasone propionate) | 1% v/v | Neg | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| | | Pos | 5/5 | 5/5 | 0/5 | 0/5 | 4/5^{1} | 0/5 | 0/5 | 0/5 | 5/5 | 0/5 | 0/5 | 0/5 | 0/5 | 5/5 | 0/5 | 0/5 | 5/5 |
| Sucrets Sore Throat (2.0 mg dyclonine hydrochloride/lozenge) | 1% w/v | Neg | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| | | Pos | 3/3 | 3/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 3/3 |
| Benadryl Allergy Liqui-gels (25 mg/capsule diphenhydramine hydrochloride) | 1% v/v | Neg | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| | | Pos^{2} | 3/3 | 3/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 3/3 |
| Zicam Cold Remedy (galphimia glauca, luffa operculata, sabadilla) | 1% v/v | Neg | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| | | Pos | 3/3 | 3/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 0/3 | 0/3 | 3/3 | 0/3 | 0/3 | 3/3 |
| Cold-eeze (2.3% zinc gluconate) | 1% w/v | Neg | 0/3 | 0/3 |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.