← Product Code [PSZ](/submissions/MI/subpart-d%E2%80%94serological-reagents/PSZ) · K053146

# QUICKVUE INFLUENZA A + B TEST (K053146)

_Quidel Corp. · PSZ · Dec 14, 2005 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PSZ/K053146

## Device Facts

- **Applicant:** Quidel Corp.
- **Product Code:** [PSZ](/submissions/MI/subpart-d%E2%80%94serological-reagents/PSZ.md)
- **Decision Date:** Dec 14, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3328
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Intended Use

The QuickVue Influenza A + B test allows for the rapid, qualitative detection of influenza type A and type B antigens directly from nasal swab, nasopharyngeal swab, nasal wash and/or nasal aspirate specimens. The test is intended for use as an aid in the rapid differential diagnosis of acute influenza type A and type B viral infections. The test is not intended to detect influenza C antigens. Negative test results should be confirmed by cell culture. The test is intended for professional and laboratory use.

## Device Story

The QuickVue Influenza A + B test is a lateral-flow immunoassay used in professional and laboratory settings to detect influenza A and B viral antigens. The device utilizes monoclonal antibodies specific to influenza A and B. The user places a clinical specimen (nasal/nasopharyngeal swab, wash, or aspirate) into an extraction reagent tube to disrupt viral particles and expose antigens. The test strip is then inserted into the tube for 10 minutes. If antigens are present, a pink-to-red test line appears for the respective type (A or B), alongside a blue procedural control line. The absence of the control line renders the test invalid. The output provides a rapid differential diagnosis, aiding clinicians in identifying the specific influenza type to inform patient management and epidemiological tracking.

## Clinical Evidence

Performance validated via multi-center field clinical study comparing results to viral culture (with RT-PCR for discrepant resolution). Additional studies included clinical performance with frozen nasal washes, transport media selection, and detection of H5N1 and embryonated egg-cultivated influenza A subtypes. Results demonstrated excellent sensitivity and specificity for the qualitative detection of influenza A and B antigens.

## Technological Characteristics

Lateral-flow immunoassay; utilizes monoclonal antibodies specific to influenza A and B antigens. One-step extraction using buffer and detergent. Form factor: test strip with two test lines (A and B) and one procedural control line. Manual, point-of-care/laboratory diagnostic. No electronic components or software.

## Regulatory Identification

An influenza virus antigen detection test system is a device intended for the qualitative detection of influenza viral antigens directly from clinical specimens in patients with signs and symptoms of respiratory infection. The test aids in the diagnosis of influenza infection and provides epidemiological information on influenza. Due to the propensity of the virus to mutate, new strains emerge over time which may potentially affect the performance of these devices. Because influenza is highly contagious and may lead to an acute respiratory tract infection causing severe illness and even death, the accuracy of these devices has serious public health implications.

## Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) The device's sensitivity and specificity performance characteristics or positive percent agreement and negative percent agreement, for each specimen type claimed in the intended use of the device, must meet one of the following two minimum clinical performance criteria:
(i) For devices evaluated as compared to an FDA-cleared nucleic acid based-test or other currently appropriate and FDA accepted comparator method other than correctly performed viral culture method:
(A) The positive percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent.
(B) The negative percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent.
(ii) For devices evaluated as compared to correctly performed viral culture method as the comparator method:
(A) The sensitivity estimate for the device when testing for influenza A must be at the point estimate of at least 90 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 80 percent. The sensitivity estimate for the device when testing for influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent.
(B) The specificity estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent.
(2) When performing testing to demonstrate the device meets the requirements in paragraph (b)(1) of this section, a currently appropriate and FDA accepted comparator method must be used to establish assay performance in clinical studies.
(3) Annual analytical reactivity testing of the device must be performed with contemporary influenza strains. This annual analytical reactivity testing must meet the following criteria:
(i) The appropriate strains to be tested will be identified by FDA in consultation with the Centers for Disease Control and Prevention (CDC) and sourced from CDC or an FDA-designated source. If the annual strains are not available from CDC, FDA will identify an alternative source for obtaining the requisite strains.
(ii) The testing must be conducted according to a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(iii) By July 31 of each calendar year, the results of the last 3 years of annual analytical reactivity testing must be included as part of the device's labeling. If a device has not been on the market long enough for 3 years of annual analytical reactivity testing to have been conducted since the device received marketing authorization from FDA, then the results of every annual analytical reactivity testing since the device received marketing authorization from FDA must be included. The results must be presented as part of the device's labeling in a tabular format, which includes the detailed information for each virus tested as described in the certificate of authentication, either by:
(A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where the analytical reactivity testing data can be found; or
(B) In the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.
(4) If one of the actions listed at section 564(b)(1)(A)-(D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services (HHS) determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate. The procedure and location of testing may depend on the nature of the emerging virus.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.

## Predicate Devices

- Quidel QuickVue® Influenza A+B test ([K031899](/device/K031899.md))
- Quidel QuickVue® Influenza test ([K991633](/device/K991633.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

| Submitted By:               | Quidel Corporation<br>10165 McKellar Court<br>San Diego, California 92121<br>Cell: 408-690-4938<br>Telephone: 408-616-4335<br>Fax: 408-616-4313                                                                                                                                                                                                                                                                                                                                                                       |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submission Contact:         | John D. Tamerius, Ph.D.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Date Prepared:              | November 9, 2005                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Device Trade Name:          | QuickVue® Influenza A + B test                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Common Name:                | Influenza test                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Predicate Devices:          | Quidel QuickVue® Influenza A+B test<br>(K031899)<br>Quidel QuickVue® Influenza test (K991633)                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Device Classification/Name: | 21 CFR 866.3330 / Influenza virus serological<br>reagents.                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                             | The device, the QuickVue Influenza A + B test,<br>is similar to other FDA-cleared devices used<br>for the qualitative detection of influenza type A<br>and B directly from clinical specimens. These<br>tests are used to aid in the diagnosis of<br>disease caused by influenza viruses A and B<br>and provide epidemiological information on<br>these diseases (21CFR 866.3330).<br>The Food and Drug Administration has<br>classified serological test systems for the<br>detection of influenza virus as Class I. |

{1}------------------------------------------------

- The Quick\/ue Influenza A + B test allows for Intended Use: the rapid, qualitative detection of influenza type A and type B antigens directly from nasal swab, nasopharyngeal swab, nasal wash and/or nasal aspirate specimens. The test is intended for use as an aid in the rapid differential diagnosis of acute influenza type A and type B viral infections. The test is not intended to detect influenza C antigens. Negative test results should be confirmed by The test is intended for cell culture. professional and laboratory use.
- Physiologic Basis of the Test: Influenza is a highly contagious, acute, viral infection of the respiratory tract. The causative agents of the disease are immunologically diverse, single-strand RNA viruses known as Influenza Viruses. There are three types of influenza viruses: A, B, and C. Type A viruses are the most prevalent and are associated with the most serious epidemics. Type B produces a disease that is generally milder than that caused by Type A. Type C has never been connected with a large epidemic of human disease. Both Type A and B viruses can circulate simultaneously, but usually one type is dominant during a given season.

Influenza antigens may be detected in clinical specimens by immunoassay. The QuickVue Influenza A + B test is a lateral-flow immunoassay highly sensitive using monoclonal antibodies that are specific for influenza antigens. The test is specific to influenza Types A and B antigen with no known cross-reactivity to normal flora or other known respiratory pathogens.

- Device Description: The QuickVue Influenza A + B test, has two Test Line indicators - one for type A and one for type B. The two Test Line indicators allow for the separate identification of type A and type B viral antigens from the same specimen. If either Test Line turns pink-to-red, the test is positive for influenza.

{2}------------------------------------------------

Device Description (cont.): Nasal swabs, nasopharyngeal swabs, nasal wash and/or nasal aspirates serve as specimens for this test. The patient specimen is placed in a tube containing Extraction Reagent, during which time the virus particles in the specimen are disrupted, exposing internal viral antigens. After extraction, the Test Strip is placed in the Extraction Tube for 10 minutes. During this time, the extracted specimen will react with the reagents in the Test Strip.

> If the extracted specimen contains influenza Type A and/or B antigens, a pink-to-red Test Line along with a blue procedural Control Line will appear on the Test Strip. If influenza Type A and B viral antigens are not present, or present at very low levels, only a blue procedural Control Line will appear. If no blue procedural Control Line develops, the result is considered invalid.

| Features                           | QuickVue®<br>Influenza A+B<br>Test (proposed)                        | QuickVue®<br>Influenza A+B<br>test                              | QuickVue®<br>Influenza test                     |
|------------------------------------|----------------------------------------------------------------------|-----------------------------------------------------------------|-------------------------------------------------|
| Intended<br>Use                    | Differential detection<br>of influenza A and B<br>viral antigen      | Differential detection<br>of influenza A and B<br>viral antigen | Detection of influenza<br>A and B viral antigen |
| Specimen<br>Types                  | Nasal Swab<br>Nasopharyngeal<br>Swab<br>Nasal Wash<br>Nasal Aspirate | Nasal Swab<br>Nasal Wash<br>Nasal Aspirate                      | Nasal Swab<br>Nasal Wash<br>Nasal Aspirate      |
| Technology                         | Lateral-flow<br>immunoassay                                          | Lateral-flow<br>immunoassay                                     | Lateral-flow<br>immunoassay                     |
| Detection of<br>Influenza<br>Virus | Detection of Influenza<br>type A and type B                          | Detection of Influenza<br>type A and type B                     | Detection of Influenza<br>type A and type B     |
| Extraction                         | 1 step; buffer and<br>detergent                                      | 1 step; buffer and<br>detergent                                 | 1 step; buffer and<br>detergent                 |

#### Device Comparison:

{3}------------------------------------------------

## Summary of Performance Data:

Several additional studies were undertaken to validate the performance characteristics of the QuickVue® Influenza A + B test. These studies included the following:

- A multi-center field clinical study was conducted. Sensitivity, specificity 1. and overall accuracy when compared to viral culture and discrepant results resolved by RT-PCR were calculated.
- 2. A clinical performance study with frozen nasal washes demonstrated the QuickVue® Influenza A + B test gave excellent sensitivity and specificity.
- A transport media study to select suitable media for the transport of 3. clinical samples identified three transport media that performed optimally.
- A study confirmed the QuickVue® Influenza A + B test detects 4. Influenza A subtype H5N1 viruses that have been cultured in vitro.
- A study confirmed the QuickVue® Influenza A + B test detects 5. influenza A subtype viruses that had been cultivated in embryonated eggs.

## Conclusion:

These studies demonstrated the substantial equivalence of the QuickVue® Influenza A + B test to existing products already marketed. They further demonstrated the suitability of the product for laboratory and professional use. Such studies are a critical element in establishing the fundamental safety and effectiveness of the product and its appropriateness for commercial distribution.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human figures.

DEC I 4 2005

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

John D. Tamerius, Ph.D. Vice President, Clinical & Regulatory Affairs Ouidel Corporation 2981 Copper Road Santa Clara, CA 95051

k053146 Re:

Trade/Device Name: QuickVue® Influenza A + B Test Regulation Number: 21 CFR 866.3330 Regulation Name: Influenza Virus Serological Reagents Regulatory Class: Class I Product Code: GNX Dated: November 9, 2005 Received: November 10, 2005

Dear Dr. Tamerius:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for are may 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, alerty missions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act thany Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{5}------------------------------------------------

#### Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) I this letter will anow you to begin finding of substantial equivalence of your device to a legally promation noticate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the I va may overily Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours,

Sally, a Hogg

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

## Indications for Use

510(k) Number (if known):
K053146

Device Name:

QuickVue® Influenza A + B test

Indications for Use:

The QuickVue Influenza A + B test allows for the rapid, qualitative detection of influenza type A and type B antigens directly from nasal swab, nasopharyngeal swab, nasal wash and/or nasal aspirate specimens. The test is intended for use as an aid in the rapid differential diagnosis of acute influenza type A and type B viral infections. The test is not intended to detect influenza C antigens. Negative test results should be confirmed by cell culture. The test is intended for professional and laboratory use.

AND/OR Prescription Use X (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Sally a H
Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

|         | ﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪ |  |
|---------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Page __ |                                                                                                                                                                               |  |
|         | ot                                                                                                                                                                            |  |
|         |                                                                                                                                                                               |  |

K05 344

7

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PSZ/K053146](https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PSZ/K053146)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
