# PRI · Procalcitonin Assay

_Microbiology · 21 CFR 866.3215 · Class 2_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PRI

## Overview

- **Product Code:** PRI
- **Device Name:** Procalcitonin Assay
- **Regulation:** [21 CFR 866.3215](/submissions/MI/subpart-d%E2%80%94serological-reagents/866.3215)
- **Device Class:** 2
- **Review Panel:** [Microbiology](/submissions/MI)

## Identification

A device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis is identified as an in vitro device intended for the detection and qualitative and/or quantitative measurement of one or more non-microbial analytes in human clinical specimens to aid in the assessment of patients with suspected sepsis when used in conjunction with clinical signs and symptoms and other clinical and laboratory findings.

## Classification Rationale

Class II (special controls). The special controls for this device are:

## Special Controls

A device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis must comply with the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the device's detailed Indications for Use statement describing what the device detects and measures, the results provided to the user, whether the measure is qualitative and/or quantitative, the clinical indications for which the test is to be used, and the specific population(s) for which the device use is intended.
(2) Premarket notification submissions must include detailed documentation of the device description, including (as applicable), all device components, software, ancillary reagents required but not provided, explanation of the device principle and methodology, and for molecular devices include detailed documentation of the primer/probe sequence, design, and rationale for sequence selection.
(3) Premarket notification submissions must include detailed documentation of applicable analytical studies, such as, analytical sensitivity (Limit of Detection, Limit of Blank, and Limit of Quantitation), precision, reproducibility, analytical measuring range, interference, cross-reactivity, and specimen stability.
(4) Premarket notification submissions must include detailed documentation of a prospective clinical study or, if appropriate, results from an equivalent sample set. This detailed documentation must include the following information:
(i) Results must demonstrate adequate device performance relative to a well-accepted comparator.
(ii) Clinical sample results must demonstrate consistency of device output throughout the device measuring range likely to be encountered in the Intended Use population.
(iii) Clinical study documentation must include the original study protocol (including predefined statistical analysis plan), study report documenting support for the Indications for Use(s), and results of all statistical analyses.
(5) Premarket notification submissions must include evaluation of the level of the non-microbial analyte in asymptomatic patients with demographic characteristics (
*e.g.,* age, racial, ethnic, and gender distribution) similar to the Intended Use population.(6) As part of the risk management activities performed under 21 CFR 820.10(c) design and development, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument.
(7) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling, and a detailed explanation of the interpretation of the limitations of the samples (
*e.g.,* collected on day of diagnosis) must be included in the device's 21 CFR 809.10(b)(10) compliant labeling.

## Recent Cleared Devices (8 of 8)

| Record | Device Name | Applicant | Decision Date | Decision |
| --- | --- | --- | --- | --- |
| [K220262](https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PRI/K220262.md) | Dimension EXL LOCI BRAHMS Procalcitonin (PCT) | Siemens Healthcare Diagnostics, Inc. | Aug 26, 2022 | SESE |
| [K192815](https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PRI/K192815.md) | Elecsys BRAHMS PCT | Roche Diagnostics | Mar 9, 2020 | SESE |
| [K173927](https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PRI/K173927.md) | Elecsys BRAHMS PCT | Roche Diagnostics | Jul 6, 2018 | SESE |
| [K173683](https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PRI/K173683.md) | LIAISON BRAHMS PCT II GEN, LIAISON Control BRAHMS PCT II GEN, LIAISON BRAHMS PCT II GEN Verifiers | DiaSorin, Inc. | Feb 27, 2018 | SESE |
| [K172713](https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PRI/K172713.md) | Lumipulse G BRAHMS PCT Immunoreaction Cartridges, Lumipulse G BRAHMS PCT Calibrators set | Fujirebio Diagnostics,Inc. | Dec 10, 2017 | SESE |
| [K170652](https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PRI/K170652.md) | ARCHITECT B.R.A.H.M.S PCT,  ARCHITECT B.R.A.H.M.S PCT Calibrators, ARCHITECT B.R.A.H.M.S PCT Controls | Fisher Diagnostics | Jun 1, 2017 | SESE |
| [K171338](https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PRI/K171338.md) | B R A H M S PCT sensitive KRYPTOR | Brahms GmbH | May 31, 2017 | SESE |
| [K162827](https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PRI/K162827.md) | VIDAS B.R.A.H.M.S. PCT (PCT) | bioMerieux, Inc. | Feb 23, 2017 | SESE |

## Top Applicants

- Roche Diagnostics — 2 clearances
- Brahms GmbH — 1 clearance
- DiaSorin, Inc. — 1 clearance
- Fisher Diagnostics — 1 clearance
- Fujirebio Diagnostics,Inc. — 1 clearance

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PRI](https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PRI)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).