← Product Code [PMN](/submissions/MI/subpart-d%E2%80%94serological-reagents/PMN) · K191168

# Cepheid Xpert SA Nasal Complete Control Panel (K191168)

_Microbiologics, Inc. · PMN · Jun 25, 2019 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PMN/K191168

## Device Facts

- **Applicant:** Microbiologics, Inc.
- **Product Code:** [PMN](/submissions/MI/subpart-d%E2%80%94serological-reagents/PMN.md)
- **Decision Date:** Jun 25, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3920
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

The Cepheid Xpert® SA Nasal Complete Control Panel is intended for use as an external assayed positive and negative quality control to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Methicilin-Resistant Staphylococcus aureus (MRSA) and Staphylococcus aureus (SA) performed with the Cepheid Xpert® SA Nasal Complete Assay on the GeneXpert® Dx System. The controls comprise cultured and inactivated Methicillin-Resistant Staphylococcus aureus as the positive control 1: Staphylococcus aureus as the positive control 2; Staphylococcus epidermidis as the negative control. The Cepheid Xpert® SA Nasal Complete Control Panel is not intended to replace manufacturer controls provided with the Cepheid Xpert® SA Nasal Complete Assay.

## Device Story

External assayed quality control material; monitors DNA extraction, amplification, and detection processes of Cepheid Xpert SA Nasal Complete Assay. Contains heat-inactivated microorganisms on swabs: MRSA (positive control 1), MSSA (positive control 2), and MSSE (negative control). Used in clinical laboratory settings; processed like patient samples on GeneXpert Dx System. Provides reproducible performance above assay limit of detection. Healthcare providers use results to verify assay performance; ensures reliability of patient test results.

## Clinical Evidence

Precision and reproducibility study conducted across 3 sites with 6 operators and 3 lots over 5 days. Total of 90 tests per positive analyte and 90 tests for negative analyte. Results showed 100% agreement (90/90) for MRSA and SA detection. Mean Ct values and %CV reported for target genes (spa, mecA, SCC) and sample processing control. No clinical diagnostic data; bench testing only.

## Technological Characteristics

Lyophilized, heat-inactivated microorganisms on swabs. Positive control 1: MRSA; Positive control 2: MSSA; Negative control: MSSE. Designed for use with GeneXpert Dx System. Standards referenced: CLSI EP25-A (stability), CLSI EP05-A3 (precision), ANSI/ASQ Z1.4-2003 (sampling).

## Regulatory Identification

An assayed quality control material for clinical microbiology assays is a device indicated for use in a test system to estimate test precision or to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. This type of device consists of single or multiple microbiological analytes intended for use with either qualitative or quantitative assays.

## Special Controls

An assayed quality control material for clinical microbiology assays must comply with the following special controls: (1) Premarket notification submissions must include detailed device description documentation and information concerning the composition of the quality control material, including, as appropriate: (i) Analyte concentration; Expected values: (ii) Analyte source: (iii) (iv) Base matrix; (v) Added components; (vi) Safety and handling information; and, (vii) Detailed instructions for use. (2) Premarket notification submissions must include detailed documentation, including line data as well as detailed study protocols and a statistical analysis plan used to establish performance, including: (i) Description of the process for value assignment and validation. (ii) Description of the protocol(s) used to establish stability. (iii) Line data establishing precision/reproducibility. (iv) Where applicable, assessment of matrix effects and any significant differences between the quality control material and typical patient samples in terms of conditions known to cause analytical error or affect assay performance. (v) Where applicable, identify or define traceability or relationship to a domestic or international standard reference material and/or method. (vi) Where applicable, detailed documentation related to studies for surrogate controls. (3) Premarket notification submissions must include an adequate mitigation (e.g., realtime stability program) to the risk of false results due to potential modifications to the assays specified in the device's 21 CFR 809.10 compliant labeling. (4) Your 21 CFR 809.10 compliant labeling must include the following: (i) The intended use in your 21 CFR 809.10(a)(2) and 21 CFR 809.10(b)(2) compliant labeling must include the following: (A) Assayed control material analyte(s); (B) Whether the material is intended for quantitative or qualitative assays: (C) Stating if the material is a surrogate control; (D)The system(s), instrument(s), or test(s) for which the quality control material is intended. (ii) The intended use in your 21 CFR 809.10(a)(2) and 21 CFR 809.10(b)(2) compliant labeling must include the following statement: "This product is not intended to replace manufacturer controls provided with the device." (iii)A limiting statement that reads "Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements."

*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed device description documentation and information concerning the composition of the quality control material, including, as appropriate:
(i) Analyte concentration;
(ii) Expected values;
(iii) Analyte source;
(iv) Base matrix;
(v) Added components;
(vi) Safety and handling information; and
(vii) Detailed instructions for use.
(2) Premarket notification submissions must include detailed documentation, including line data as well as detailed study protocols and a statistical analysis plan used to establish performance, including:
(i) Description of the process for value assignment and validation.
(ii) Description of the protocol(s) used to establish stability.
(iii) Line data establishing precision/reproducibility.
(iv) Where applicable, assessment of matrix effects and any significant differences between the quality control material and typical patient samples in terms of conditions known to cause analytical error or affect assay performance.
(v) Where applicable, identify or define traceability or relationship to a domestic or international standard reference material and/or method.
(vi) Where applicable, detailed documentation related to studies for surrogate controls.
(3) Premarket notification submissions must include an adequate mitigation (e.g., real-time stability program) to the risk of false results due to potential modifications to the assays specified in the device's 21 CFR 809.10 compliant labeling.
(4) Your 21 CFR 809.10 compliant labeling must include the following:
(i) The intended use of your 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include the following:
(A) Assayed control material analyte(s);
(B) Whether the material is intended for quantitative or qualitative assays;
(C) Stating if the material is a surrogate control; and
(D) The system(s), instrument(s), or test(s) for which the quality control material is intended.
(ii) The intended use in your 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include the following statement: “This product is not intended to replace manufacturer controls provided with the device.”
(iii) A limiting statement that reads “Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements.”

## Predicate Devices

- Cepheid Xpert® GBS LB Control Panel ([K182472](/device/K182472.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

ASSAY ONLY

## I Background Information:

A 510(k) Number

K191168

B Applicant

Microbiologics, Inc.

C Proprietary and Established Names

Cepheid Xpert SA Nasal Complete Control Panel

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  PMN | Class II | 21 CFR 866.3920 - Assayed Quality Control Material For Clinical Microbiology Assays | IM - Immunology & MI - Microbiology  |

## II Submission/Device Overview:

A Purpose for Submission:

To obtain a Substantial Equivalence determination for the Cepheid Xpert SA Nasal Complete Control Panel for use with the Cepheid Xpert SA Nasal Complete Assay on the GeneXpert Dx System.

B Measurand:

Nucleic acid quality control material from inactivated Methicillin-Resistant Staphylococcus aureus (MRSA, positive control 1), Staphylococcus aureus (SA, positive control 2) and Staphylococcus epidermidis (negative control).

C Type of Test:

The Cepheid Xpert SA Nasal Complete Control Panel is an external assayed positive and negative quality control material to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Staphylococcus aureus (MRSA) and

K191168 - Page 1 of 8

{1}

Staphylococcus aureus (SA) performed with the Cepheid Xpert SA Nasal Complete Assay on the GeneXpert Dx System.

## III Intended Use/Indications for Use:

### A Intended Use(s):
See Indications for Use below.

### B Indication(s) for Use:
The Cepheid Xpert SA Nasal Complete Control Panel is intended for use as an external assayed positive and negative quality control to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Methicillin-Resistant Staphylococcus aureus (MRSA) and Staphylococcus aureus (SA) performed with the Cepheid Xpert SA Nasal Complete Assay on the GeneXpert Dx System. The controls comprise cultured and inactivated Methicillin-Resistant Staphylococcus aureus as the positive control 1; Staphylococcus aureus as the positive control 2; Staphylococcus epidermidis as the negative control.

The Cepheid Xpert SA Nasal Complete Control Panel is not intended to replace manufacturer controls provided with the Cepheid Xpert SA Nasal Complete Assay.

### C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only

For in vitro diagnostic use only.

This product is not intended to replace the manufacturer controls provided with the Cepheid Xpert SA Nasal Complete Assay.

### D Special Instrument Requirements:
The Cepheid Xpert SA Nasal Complete Control Panel is intended for use on the GeneXpert Dx System.

## IV Device/System Characteristics:

### A Device Description:
The Cepheid Xpert SA Nasal Complete Control Panel is used to monitor the DNA extraction, amplification and detection processes of the Cepheid Xpert SA Nasal Complete Assay. The Cepheid Xpert SA Nasal Complete Control Panel contains cultured microorganisms inactivated by heat treatments. Each Cepheid Xpert SA Nasal Complete Control Panel consists of 6 individually packaged Methicillin-Resistant Staphylococcus aureus (MRSA) positive control swabs (positive control 1); 6 individually wrapped methicillin-susceptible Staphylococcus aureus (MSSA) positive control swabs (positive control 2); and 6 individually wrapped methicillin-susceptible Staphylococcus epidermidis (MSSE) negative control swabs. Each positive control 1 swab contains MRSA at a target level that is designed to provide reproducible performance above the limit of detection to produce positive results for each of the genes targeted by the Cepheid Xpert SA Nasal Complete Assay: Staphylococcal protein A gene spa (SPA), methicillin resistance gene mecA (mec), and the Staphylococcal cassette chromosome (SCC). Each positive control 2 swab contains MSSA at a target level designed to provide reproducible performance

K191168 - Page 2 of 8

{2}

above the limit of detection of the spa gene target of the Cepheid Xpert SA Nasal Complete Assay. The negative control swab contains MSSE which is not targeted/detected by the Cepheid Xpert SA Nasal Complete Assay. Each swab is individually wrapped with a desiccant in a heat-sealed foil pouch.

# B Principle of Operation:

The Cepheid Xpert SA Nasal Complete Control Panel is intended for use as external assayed quality control material for use in monitoring the DNA extraction, amplification and detection processes associated with the Cepheid Xpert SA Nasal Complete Assay on the GeneXpert Dx System.

# V Substantial Equivalence Information:

# A Predicate Device Name(s):

Cepheid Xpert GBS LB Control Panel

# B Predicate 510(k) Number(s):

K182472

# C Comparison with Predicate(s):

Table 1. Comparison with the Predicate

|  Device & Predicate Device(s): | K191168 | K182472  |
| --- | --- | --- |
|  Device Trade Name | Cepheid Xpert SA Nasal Complete Control Panel | Cepheid Xpert GBS LB Control Panel  |
|  General Device Characteristic Similarities  |   |   |
|  Intended Use/Indications for Use | External assayed positive and negative quality control to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Methicillin-Resistant Staphylococcus aureus (MRSA) and Staphylococcus aureus (SA) performed with the Cepheid Xpert SA Nasal Complete Assay on the GeneXpert Dx System. The controls comprise cultured and inactivated Methicillin-Resistant Staphylococcus aureus as the positive control 1; Staphylococcus aureus as the positive control 2; Staphylococcus epidermidis as the negative control. The Cepheid Xpert SA Nasal | External assayed positive and negative quality control materials to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Group B Streptococcus (GBS) performed with the Cepheid Xpert GBS LB Assay on the GeneXpert Instrument System. The controls comprise cultured and inactivated Streptococcus agalactiae as the positive control and Lactobacillus acidophilus as the negative control. The Cepheid Xpert GBS LB  |

K191168 - Page 3 of 8

{3}

|   | Complete Control Panel is not intended to replace manufacturer controls provided with the Cepheid Xpert SA Nasal Complete Assay. | Control Panel is not intended to replace the manufacturer controls provided with the device.  |
| --- | --- | --- |
|  Composition | Inactivated microorganisms | Same  |
|  Test System | Cepheid GeneXpert System | Same  |
|  Directions for Use | Process like a patient sample | Same  |
|  Assay Steps Monitored | Extraction, amplification, and detection | Same  |
|  Physical Format | Lyophilized swab | Lyophilized swab  |
|  General Device Characteristic Differences  |   |   |
|  Analytes | • Positive Control 1: (MRSA) Methicillin-resistant Staphylococcus aureus
• Positive Control 2: (SA) Staphylococcus aureus
• Negative Control: Staphylococcus epidermidis | • Positive Control: Streptococcus agalactiae
• Negative Control: Lactobacillus acidophilus  |
|  Assay Compatibility | Cepheid Xpert SA Nasal Complete Assay (K100822) | Cepheid Xpert GBS LB Assay (K121539)  |

VI Standards/Guidance Documents Referenced:

CLSI. Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline. CLSI Document EP25-A. Wayne, PA: Clinical and Laboratory Standards Institute; 2009.

CLSI. Evaluation of Precision of Quantitative Measurement Procedures: Approved Guideline – Third Edition. CLSI Document EP05-A3. Wayne, PA: Clinical and Laboratory Standards Institute; 2014.

ANSI. Sampling Procedures and Tables for Inspection by Attributes. ANSI/ASQ Z1.4-2003; 2013.

VII Performance Characteristics (if/when applicable):

A Analytical Performance:

1. Precision/Reproducibility:

A study was performed to determine the reproducibility of the Cepheid Xpert SA Nasal Complete Control Panel using the Cepheid Xpert SA Nasal Complete Assay on the GeneXpert Dx System. Testing was done at three different locations over five days. At each location, two operators each tested three different lots of control material for a total of at least 90 test results each for the positive control 1, positive control 2 and the negative control swabs (3 sites x 5 days x 2 operators x ≥3 replicates = ≥90 replicates in total). Summaries of the results for both the positive and negative controls are provided in Table 2 and Table 3. On initial testing, one positive control 1 and two positive control 2 swabs produced “NO RESULT” because insufficient data were collected to determine the sample status. After re

K191168 - Page 4 of 8

{4}

testing using new control swabs, all positive control swabs produced the expected positive results. Fourteen "INVALID" results due to failure of the Sample Processing Control, one "ERROR" due to system failure and one report of "NO RESULT" were observed with the negative control. After re-testing of new control swabs, all negative control swabs produced the expected negative results.

Table 2. Summary of results from the Reproducibility Study (qualitative)

|  Positive Control 1 (Methicillin-Resistant Staphylococcus aureus)  |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  Test location | Total Tests Performed | INVALID | NO RESULT1 | ERROR | Correct Result | Incorrect Result | Percent Correct2  |
|  1 | 30 | 0 | 0 | 0 | 30 | 0 | 100%  |
|  2 | 30 | 0 | 0 | 0 | 30 | 0 | 100%  |
|  3 | 31 | 0 | 1 | 0 | 30 | 0 | 100%  |
|  All sites | 91 | 0 | 1 | 0 | 90 | 0 | 100%  |
|  Positive Control 2 (Methicillin-Susceptible Staphylococcus aureus)  |   |   |   |   |   |   |   |
|  Test location | Total Tests Performed | INVALID | NO RESULT1 | ERROR | Correct Result | Incorrect Result | Percent Correct2  |
|  1 | 30 | 0 | 0 | 0 | 30 | 0 | 100%  |
|  2 | 30 | 0 | 0 | 0 | 30 | 0 | 100%  |
|  3 | 32 | 0 | 2 | 0 | 30 | 0 | 100%  |
|  All sites | 92 | 0 | 2 | 0 | 90 | 0 | 100%  |
|  Negative Control (Methicillin-Susceptible Staphylococcus epidermidis)  |   |   |   |   |   |   |   |
|  Test location | Total Tests Performed | INVALID3 | NO RESULT1 | ERROR4 | Correct Result | Incorrect Result | Percent Correct2  |
|  1 | 38 | 8 | 0 | 0 | 30 | 0 | 100%  |
|  2 | 33 | 2 | 0 | 1 | 30 | 0 | 100%  |
|  3 | 35 | 4 | 1 | 0 | 30 | 0 | 100%  |
|  All sites | 106 | 14 | 1 | 1 | 90 | 0 | 100%  |

1Number of NO RESULT responses observed; in each case new controls were tested, as indicated in the test protocol, and the expected results were obtained.
2Data from test runs that produced any type of error response were not included in the Percent Correct analysis.
3Number of INVALID responses observed; in each case new controls were tested, as indicated in the test protocol, and the expected results were obtained.
4Number of ERROR responses observed; a new control was tested, as indicated in the test protocol, and the expected results were obtained.

Table 3. Summary of results from the Reproducibility Study (quantitative)

|   | Mean Ct (%CV)  |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  Test location | Positive Control 1 |   |   | Positive Control 2 | Negative Control  |
|   | SPA | mec | SCC | SPA | SPC  |
|  1 | 27.0(2.7) | 27.3(2.6) | 28.5(2.3) | 26.8(2.7) | 32.0(4.2)  |
|  2 | 28.1(2.3) | 28.4(2.0) | 29.6(1.9) | 27.8(2.5) | 31.8(3.5)  |
|  3 | 27.8(2.2) | 28.1(2.1) | 29.3(2.1) | 27.4(2.8) | 32.1(4.4)  |
|  All sites | 27.6(3.0) | 27.9(2.8) | 29.1(2.7) | 27.3(3.0) | 32.0(4.0)  |

SPC: Sample processing control
Ct: Cycle Threshold;  $\% \mathrm{CV}$  : Percent Coefficient of Variation

K191168 - Page 5 of 8

{5}

The reproducibility of the Cepheid Xpert SA Nasal Complete Control Panel within and between test locations, GeneXpert Instruments, operators, and lots was determined to be acceptable.

2. Linearity:
Not applicable.

3. Analytical Specificity/Interference:
Not applicable.

4. Assay Reportable Range:
Not applicable.

5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):

Traceability:
Not applicable.

Stability:
1. Shelf-life was established through an Accelerated Stability Study that was performed with three lots of each Positive and Negative Control from the Cepheid Xpert SA Nasal Complete Control Panel. The two types of positive controls and the negative control were placed at elevated temperatures of 43°C, 53°C and 63°C. The positive controls were tested in replicates of four at each of the four time points (Day 0, Day 14, Day 28 and Day 42). The negative controls were tested in duplicate at two time points (Day 0 and Day 42). The data provided no evidence of product degradation under the conditions tested. A shelf-life claim of 9 months at 2-25°C is acceptable. This shelf-life claim could be modified based on the results of the Real-Time Stability Study.

2. A Real-Time Stability, Shelf-Life Study is in process. Under a 10-point regression design, three lots of each control will be incubated at two temperatures (2-8°C and 25°C). Regression analysis will be performed at the end of the study to evaluate performance and determine if the acceptance criteria are met. At that time, the expiration date will be updated to the time point at which all criteria are met.

3. An In-Use Stability Study was done to determine how long the positive and negative controls would remain stable after reconstitution. Controls from a single lot were rehydrated and kept at room temperature for different time periods prior to testing. All results were as expected and indicate that Cepheid Xpert SA Nasal Complete Control Panel swabs may be used up to five hours following rehydration.

6. Detection Limit:
Not applicable.

7. Assay Cut-Off:
Not applicable.

K191168 - Page 6 of 8

{6}

# B Comparison Studies:

1. Method Comparison with Predicate Device: Not applicable.
2. Matrix Comparison: Not applicable.

# C Clinical Studies:

1. Clinical Sensitivity: Not applicable.
2. Clinical Specificity: Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): Not applicable.

# D Clinical Cut-Off:

Not applicable.

# E Expected Values/Reference Range:

The Cepheid Xpert SA Nasal Complete Control Panel is a qualitative control expected to produce the results described in Table 4 when tested with the Cepheid Xpert SA Nasal Complete assay using the GeneXpert Dx System.

Table 4: Expected values

|  Control | Analyte | Expected Assay Result | Interpretation  |
| --- | --- | --- | --- |
|  Positive Control 1 | Methicillin-Resistant S. aureus (MRSA) | MRSA POSITIVE; SA POSITIVE | MRSA target DNA detected; SA target DNA detected. All MRSA targets (spa, mecA and SCCmec) have a Ct within the valid range and endpoint above the threshold setting.  |
|  Positive Control 2 | Methicillin-Susceptible S. aureus | MRSA NEGATIVE; SA POSITIVE | MRSA target DNA not detected; SA target DNA detected. SA target (spa) has a Ct within the valid range and endpoint above the threshold setting. Target DNA for SCCmec and/or mecA is not detected.  |

K191168 - Page 7 of 8

{7}

|  Control | Analyte | Expected Assay Result | Interpretation  |
| --- | --- | --- | --- |
|  Negative Control | Methicillin-Susceptible S. epidermidis | MRSA NEGATIVE; SA NEGATIVE | SA target DNA not detected. SA target (spa) DNA is not detected. Target DNA for mecA and/or SCCmec may or may not be detected.  |

VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K191168 - Page 8 of 8

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PMN/K191168](https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PMN/K191168)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
