Prospective clinical study specimens; Routine clinical blood culture results
The prospective clinical study compared the T2Candida Panel results against standard-of-care blood culture results to determine clinical specificity and sensitivity in a real-world patient population.
Prospective clinical study; Standard of care; Clinical specificity; Candidemia
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Prospective clinical study; Prospective comparison of T2Candida Panel results with standard-of-care blood culture results; Study Period: August 2013 - October 2013 and April 2014
Adult patients referred for a diagnostic blood culture per routine standard of care; Sample Size: 1501; Number of Sites: 9
Standard blood culture (aerobic and anaerobic bottle set)
Clinical sensitivity and specificity
Indications for Use
The T2Candida Panel and T2Dx® Instrument is a qualitative T2 Magnetic Resonance (T2MR®) assay for the direct detection of Candida species in EDTA human whole blood specimens from patients with symptoms of, or medical conditions predisposing the patient to, invasive fungal infections. The T2Candida Panel identifies five species of Candida and categorizes them into the following three species groups: 1. Candida albicans and/or Candida tropicalis, 2. Candida parapsilosis 3. Candida glabrata and/or Candida krusei The T2Candida Panel does not distinguish between C. albicans and C. tropicalis. The T2Candida Panel does not distinguish between C. glabrata and C. krusei. The T2Candida Panel is indicated for the presumptive diagnosis of candidemia. The T2Candida Panel is performed independent of blood culture. Concomitant blood cultures are necessary to recover organisms for susceptibility testing or further identification. The T2Candida positive and negative External Controls are intended to be used as quality control samples with the T2Candida Panel when run on the T2Dx® instrument system. These controls are not intended for use with other assays or systems.
Device Story
T2Dx Instrument uses T2 Magnetic Resonance (T2MR) technology to detect Candida species directly from EDTA whole blood; bypasses traditional blood culture. Instrument processes blood samples to amplify and identify specific Candida nucleic acid sequences; categorizes into three groups: C. albicans/tropicalis, C. parapsilosis, and C. glabrata/krusei. Used in clinical laboratory settings; operated by trained laboratory personnel. Output provides qualitative detection results to clinicians; aids in rapid presumptive diagnosis of candidemia. Enables faster clinical decision-making regarding antifungal therapy compared to culture-based methods. Concomitant blood cultures remain necessary for definitive identification and susceptibility testing.
Clinical Evidence
Not explicitly detailed in the provided document; however, the order mandates that premarket submissions must include detailed documentation from clinical studies comparing device performance to well-accepted reference methods.
Technological Characteristics
T2 Magnetic Resonance (T2MR) assay; qualitative nucleic acid-based amplification and detection. Analyzes EDTA whole blood. System includes T2Dx instrument and T2Candida Panel reagents. Requires specific primer/probe sequences for Candida identification. Software-controlled instrument operation.
Indications for Use
Indicated for presumptive diagnosis of candidemia in patients with symptoms or predisposing conditions for invasive fungal infections. Requires concomitant blood cultures for organism recovery/susceptibility testing.
Regulatory Classification
Identification
A nucleic acid-based device for the amplification, detection, and identification of microbial pathogens directly from whole blood specimens is a qualitative in vitro device intended for the amplification, detection, and identification of microbial-associated nucleic acid sequences from patients with suspected bloodstream infections. This device is intended to aid in the diagnosis of bloodstream infection when used in conjunction with clinical signs and symptoms and other laboratory findings.
Special Controls
A nucleic acid-based device for the amplification, detection and identification of microbial pathogens directly from whole blood specimens must comply with the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed device description documentation, including the device components, ancillary reagents required but not provided, and a detailed explanation of the methodology, including primer/probe sequence, design, and rationale for sequence selection.
(2) Premarket notification submissions must include detailed documentation from the following analytical and clinical performance studies: Analytical sensitivity (limit of detection), reactivity, inclusivity, precision, reproducibility, interference, cross reactivity, carryover, and cross contamination.
(3) Premarket notification submissions must include detailed documentation from a clinical study. The study, performed on a study population consistent with the intended use population, must compare the device performance to results obtained from well-accepted reference methods.
(4) Premarket notification submissions must include detailed documentation for device software, including, but not limited to, software applications and hardware-based devices that incorporate software.
(5) The device labeling must include limitations regarding the need for culture confirmation of negative specimens, as appropriate.
(6) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling.
(7) Premarket notification submissions must include details on an end user device training program that will be offered while marketing the device, as appropriate.
(8) As part of the risk management activities performed under 21 CFR 820.10(c) design and development, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument.
Submission Summary (Full Text)
{0}------------------------------------------------
# EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR T2Candida Panel and T2Dx® Instrument
## DECISION SUMMARY
### A. DEN Number:
DEN140019
### B. Purpose for Submission:
De Novo request for evaluation of automatic class III designation for the T2Candida Panel and T2Dx® Instrument
### C. Measurands:
The assay amplifies and detects nucleic acids of the following species: Candida albicans and/or Candida tropicalis Candida parapsilosis Candida krusei and/or Candida glabrata
### D. Type of Test:
The T2Candida Panel, performed on the T2Dx® Instrument, is a molecular diagnostic assay for the detection of the above listed Candida species from whole blood specimens obtained from patients with signs and symptoms of invasive Candida infection.
### E. Applicant:
T2 Biosystems, Inc.
## F. Proprietary and Established Names:
T2Candida Panel and T2Dx® Instrument
## G. Regulatory Information:
- 1. Regulation section:
21 CFR 866.3960
- 2. Classification:
Class II
- 3. Product code(s):
{1}------------------------------------------------
PII
NSU
- 4. Panel:
83- Microbiology
# H. Intended Use:
- 1. Intended use(s):
The T2Candida Panel and T2Dx® Instrument is a qualitative T2 Magnetic Resonance (T2MR®) assay for the direct detection of Candida species in EDTA human whole blood specimens from patients with symptoms of, or medical conditions predisposing the patient to, invasive fungal infections. The T2Candida Panel identifies five species of Candida and categorizes them into the following three species groups:
- 1. Candida albicans and/or Candida tropicalis,
- 2. Candida parapsilosis
- 3. Candida glabrata and/or Candida krusei
The T2Candida Panel does not distinguish between C. albicans and C. tropicalis. The T2Candida Panel does not distinguish between C. glabrata and C. krusei.
The T2Candida Panel is indicated for the presumptive diagnosis of candidemia. The T2Candida Panel is performed independent of blood culture. Concomitant blood cultures are necessary to recover organisms for susceptibility testing or further identification.
The T2Candida positive and negative External Controls are intended to be used as quality control samples with the T2Candida Panel when run on the T2Dx® instrument system. These controls are not intended for use with other assays or systems.
- 2. Indication(s) for use:
Same as Intended Use
- 3. Special conditions for use statement(s):
For prescription use only.
- 4. Special instrument requirements:
The T2Candida Panel is performed on the T2Dx® Instrument.
{2}------------------------------------------------
# I. Device Description:
The T2Candida panel and T2Dx® Instrument is comprised of the T2Candida Panel performed on the T2Dx® Instrument. The T2Candida Panel is a qualitative molecular diagnostic assay that employs a whole blood compatible PCR amplification followed by T2 magnetic resonance (T2MR) detection. The T2Candida Panel is performed on the T2Dx Instrument which executes all steps after specimen loading. A K2 EDTA whole blood specimen is loaded onto the T2Candida Sample Inlet, which is then placed on the T2Candida Base along with the T2Candida Reagent Pack. The Reagent Pack contains the internal control, amplification reagent, enzyme and the probe-coupled superparamagnetic particles for each Candida target. Three milliliters of the blood specimen is transferred to the T2Dx® Instrument where lysis of the red blood cells, concentration and lysis of the Candida cells and amplification of the Candida DNA takes place. Amplification products are detected by T2MR detection using species-specific probes which are attached to the superparamagnetic particles. At the end of each assay, the T2Dx® Instrument uses a bleach solution to neutralize all liquids on the cartridge to mitigate the risk of amplicon contamination. The assay provides an identification of Candida albicans and/or Candida tropicalis, Candida parapsilosis, and Candida glabrata and/or Candida krusei. The test does not distinguish between C. albicans and C. tropicalis. The test does not distinguish between C. glabrata and C. krusei.
# J. Standard/Guidance Document Referenced (if applicable):
- . IEC 61010-1:2001, (Second Edition). Safety requirements for electrical equipment for measurement control and laboratory use - General requirements. 2001
- . IEC61010-2-010:2003 (Second Edition). Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-010. Particular requirements for laboratory equipment for the heating of materials. 2003
- IEC 61010-2-081:2001 (First Edition) + A1:2003. Safety requirements for electrical . equipment for measurement, control and laboratory use - Part 2-081: Particular requirements for automatic and semiautomatic laboratory equipment for analysis and other purposes. 2001
- . IEC 61010-2-101:2002 (Second Edition). Safety requirements for electrical equipment for measurement, control, and laboratory use. Particular requirements for in vitro diagnostic (IVD) medical equipment. 2002.
- . IEC 61326-1:2005. Electrical equipment for measurement, control and laboratory use -EMC requirements - Part 1: General requirements. 2005
- . IEC 61326-2-6:2005. Electrical equipment for measurement, control and laboratory use -EMC requirements - Part 2-6: Particular requirements - In vitro diagnostic (IVD) medical equipment. 2005
- . CISPR 11:2003 Group 1 Class A. Industrial, scientific and medical equipment -Radiofrequency disturbance characteristics – Limits and methods of measurement. 2003
- . CLSI MM03-A2. Molecular Diagnostic Methods for Infectious Diseases: Approved Guideline. 2008
- CLSI EP17-A2. Protocols for Determination of Limits of Detection and Limits of . Quantitation: Approved Guideline. 2013.
{3}------------------------------------------------
- CLSI EP07-A2. Interference Testing in Clinical Chemistry. 2007 ●
- . CLSI EP05-A2. Evaluation of Precision Performance of Quantitative Measurement Methods. 2005
- CLSI EP25-A. Evaluation of Stability of In Vitro Diagnostic Reagents. 2013.
- LIS01-A2. Standard Specification for Low Level Protocol to Transfer Messages Between Clinical Laboratory Instruments and Computer Systems; Approved Standard - Second Edition. Maintained by CLSI. 2009
- . LIS02-A2. Standard Specification for Transferring Information Between Clinical Laboratory Instruments and Information Systems; Approved Standard - Second Edition. Maintained by CLSI. 2008
- . ISTA 7E. Testing Standard for Thermal Transport Packaging Used in Parcel Delivery System Shipment. 2010
- ISTA 2A. Partial Simulation Performance Test (ASTM D5276 Drop Test). 2011. ●
- CLSI EP12-A2. User Protocol for Evaluation of Qualitative Test Performance: Approved . Guideline - Second Edition. 2014.
# K. Test Principle:
In the T2Candida Panel and the T2Dx® Instrument a single pair of pan-Candida primers hybridizes to conserved sequences within the 5.8S and 26S ribosomal RNA operon of Candida species and amplify the intervening transcribed spacer 2 sequence (ITS2) released from the lysed Candida cells contained in the specimen. The primers are mixed in a ratio such that an asymmetric produces a predominantly single stranded nucleic acid after amplification. The primers are designed to also amplify the internal control. Speciesspecific and internal control-specific probes provide the specificity of the assay. The instrument detects the amplified PCR product directly in the whole blood matrix by amplicon-induced agglomeration of superparamagnetic particles to which the speciesspecific and internal control-specific probes are attached. The detection method measures the spin-spin relaxation signals of water molecules. For these measurements, the T2Dx Instrument utilizes a small permanent magnet and a specialized radio frequency coil to measure the T2MR signal from the water molecules. T2MR measures the disorder of the nuclear spins of the water molecules contained in the sample and this disorder is directly proportional to the superparamagnetic particle clustering state. The T2Dx® Instrument reports a positive or negative result for each detection channel [C. albicans/C. tropicalis (A/T), C. parapsilosis (P), C. krusei/ C. glabrata (K/G) and internal control (IC)].
# L. Performance Characteristics:
# 1. Analytical performance:
## a. Precision/Reproducibility
A multicenter reproducibility study was performed to determine the run to run, reagent lot, day to day, and site to site reproducibility. Testing was performed at three sites (two external and one internal) with a panel of three Candida species (C. albicans, C. parapsilosis and C. glabrata), each tested at two concentrations (1 - 2X
{4}------------------------------------------------
LoD. 3 - 4X LoD) using two reagent lots. Testing was performed for six nonconsecutive days with two runs and two operators per day. Organisms were tested in triplicate. A total of 108 data points were determined for each analyte at each concentration.
The organisms were prepared in fresh negative human blood samples and the concentration of organisms that were spiked in the samples was confirmed using colony count testing. Spiked samples were prepared by T2 Biosystems, deidentified and shipped to the testing sites under refrigerated conditions and stored at 2 - 8° C until testing. Reproducibility results were acceptable with a range of 95.4 to 100% agreement with expected results for each Candida species tested. A summary of the results is shown in Table 1 below.
Of the total 830 tests obtained, an analysis was conducted to assess the rate of invalid results. instrument failure or false positives. Those were as follows:
Invalid Results: 13/830 tests (1.6%) Instrument Failures: 17/830 tests (2.1%) False Positive: 3/830 (0.4%)
| Organism | Concentration | Test site | No.<br>Detected | No. Not<br>Detected | %<br>Agreement<br>with<br>Expected |
|----------------------------------|---------------|-----------|-----------------|---------------------|------------------------------------|
| <i>C.</i><br><i>parapsilosis</i> | 1 – 2 X LoD | Site 1 | 36 | 0 | |
| | | Site 2 | 36 | 0 | 108/108 |
| | | Site 3 | 36 | 0 | 100% |
| | | All sites | 108 | 0 | |
| <i>C.</i><br><i>parapsilosis</i> | 3-4 X LoD | Site 1 | 36 | 0 | |
| | | Site 2 | 36 | 0 | 108/108 |
| | | Site 3 | 36 | 0 | 100% |
| | | All sites | 108 | 0 | |
| <i>C. glabrata</i> | 1-2 X LoDa | Site 1 | 35 | 1 | |
| | | Site 2 | 34 | 2 | 105/108 |
| | | Site 3 | 36 | 0 | 97.2% |
| | | All sites | 105 | 3 | |
| <i>C. glabrata</i> | 3-4 X LoD | Site 1 | 36 | 0 | |
| | | Site 2 | 35 | 1 | 106/108 |
| | | Site 3 | 35 | 1 | 98.1% |
| | | All sites | 106 | 2 | |
| <i>C. albicans</i> | 1-2 X LoDb | Site 1 | 35 | 1 | |
| | | Site 2 | 35 | 1 | 103/108 |
| | | Site 3 | 33 | 3 | 95.4% |
| | | All sites | 103 | 5 | |
### Table 1. Summary of Reproducibility Results Across Sites, Reagents and Operators
{5}------------------------------------------------
| | 3-4 X LoD | Site 1 | 35 | 1 | 107/108<br>99.1% |
|----------|-----------|-----------|-----|-----|------------------|
| | | Site 2 | 36 | 0 | |
| | | Site 3 | 36 | 0 | |
| | | All sites | 107 | 1 | |
| Negative | N/A | Site 1 | 0 | 36 | 108/108<br>100% |
| | | Site 2 | 0 | 36 | |
| | | Site 3 | 0 | 36 | |
| | | All sites | 0 | 108 | |
ª Two false positive results in the P channel
b One false positive result in the P channel
- b. Linearity/assay Reportable Range:
Not Applicable
- Traceability, Stability, Expected Values (controls, calibrators, or methods): C.
The T2Candida Panel and T2Dx® Instrument require two types of controls, the internal control and the external controls. The Internal Control (IC) is introduced into each specimen during sample processing on the T2Dx® instrument; the pan-Candida primers are designed to amplify the target Candida species and the internal control DNA. The IC is designed to report on the presence of any inhibitors in the clinical specimen and on any instrument-related errors. The signal cut-off for the IC is a 66(4) (b)(4) If the IC signal is below its cutoff and if the patient specimen is negative for all Candida targets, the specimen result cannot be determined and the IC result will be flagged as "Invalid" by the software resulting in an invalid test. If any of the Candida targets is determined to be positive. the specimen will be reported as positive for that target, regardless of the performance of the IC.
The external controls are provided in a kit and include both positive and negative controls prepared in a whole blood matrix. The T2 positive external control includes two blends of Candida species which can interrogate each T2Candida detection channel. The T2Candida A/P/G positive external control is comprised of a blend of C. albicans, C. parapsilosis and C. glabrata. The T/P/K positive external control is comprised of a blend of C. tropicalis, C. parapsilosis and C. krusei. The Candida concentration in the positive external controls is at least 3-5X LoD. The T2 negative external control is a whole blood specimen free of any Candida cells.
As noted in sections 3a and 3b below, clinical studies were conducted using prospectively collected specimens and contrived specimens. During the clinical and analytical studies for the T2Candida assay, fresh whole blood external controls were used; a positive and negative control was run each day of the contrived and prospective arms of the study; positive controls alternated between the APG and the TPK controls. A summary of the quality control test results across all sites is listed in Table 2 below.
{6}------------------------------------------------
| Target<br>Species | Site | No. Tested | No. Positive | No. Negative | No. Invalid |
|---------------------|-----------|------------|--------------|--------------|-------------|
| A/P/G<br>controla | Site 1 | 67 | 67 | 0 | 0 |
| | Site2 | 68 | 66 | 1c | 1 |
| | Site 3 | 56 | 56 | 0 | 0 |
| | Site 4 | 49 | 46 | 0 | 3 |
| | Site 5 | 99 | 97 | 0 | 2 |
| | Site 6 | 16 | 15 | 1c | 0 |
| | Site 7 | 11 | 10 | 0 | 1 |
| | All Sites | 366 | 357 (97.5%) | 2 (0.6%) | 7 (1.9%) |
| T/P/K<br>Controlb | Site 1 | 69 | 67 | 0 | 2 |
| | Site2 | 73 | 72 | 0 | 1 |
| | Site 3 | 58 | 56 | 0 | 2 |
| | Site 4 | 60 | 58 | 0 | 2 |
| | Site 5 | 102 | 98 | 0 | 4 |
| | Site 6 | 13 | 12 | 0 | 1 |
| | Site 7 | 14 | 13 | 0 | 1 |
| | All Sites | 389 | 376 (96.7) | 0 | 13 (3.3%) |
| Negative<br>Control | Site 1 | 138 | 0 | 136 | 2 |
| | Site2 | 140 | 2d | 137 | 1 |
| | Site 3 | 115 | 3d | 109 | 3 |
| | Site 4 | 109 | 2d | 105 | 2 |
| | Site 5 | 200 | 4d | 194 | 2 |
| | Site 6 | 29 | 0 | 25 | 4 |
| | Site 7 | 26 | 0 | 24 | 2 |
| | All Sites | 757 | 11 (1.4%) | 730 (96.4%) | 16 (2.1%) |
Table 2. Summary of Quality Control Testing, Including Testing Performed During the Clinical (Prospective and Contrived) Testing Arms.
a A/P/G control contains: C. albicans, C. parapsilosis and C. glabrata.
b T/P/K control contains C. tropicalis, C. parapsilosis and C. krusei.
C One of three channels negative.
d One of three channels positive.
Fresh vs. Frozen Control Equivalence. Due to the limited shelf life of the fresh whole blood external controls, the positive and negative controls will be provided in a separate package as frozen whole blood external controls containing the same organism blends as is found in the fresh whole blood controls. These controls are intended to be stored at -20° C, thawed and equilibrated to room temperature prior to testing. In order to evaluate the equivalence of the frozen external controls with the fresh whole blood external controls, a study was conducted with two lots each of fresh and frozen controls using 65 samples (each, fresh and frozen) of the A/P/G positive control, 65 samples (each, fresh and frozen) of the T/P/K positive control and 65 samples (each, fresh and frozen) of the negative control. Results confirmed the equivalence of the fresh and frozen controls. Results for each of the three channels (A/T, P, K/G) are shown in Table 3 below.
{7}------------------------------------------------
| Control | Lot | No tested | A/T<br>Channel<br>Pos (%) | P<br>Channel<br>Pos (%) | K/G<br>Channel<br>Pos (%) | IC Pos<br>(%) | Invalid |
|--------------------|-----|-----------|---------------------------|-------------------------|---------------------------|---------------|---------|
| A/P/G<br>Fresh | 1 | 65 | 65 (100) | 65 (100) | 65 (100) | 65<br>(100) | 0 |
| A/P/G<br>Fresh | 2 | 65 | 65 (100) | 65 (100) | 65 (100) | 65<br>(100) | 0 |
| A/P/G<br>Frozen | 1 | 65 | 65 (100) | 65 (100) | 64 (98.5) | 65<br>(100) | 0 |
| A/P/G<br>Frozen | 2 | 65 | 65 (100) | 65 (100) | 64 (98.5) | 65<br>(100) | 0 |
| T/P/K<br>Fresh | 1 | 65 | 64 (98.5) | 65 (100) | 64(98.5) | 65<br>(100) | 0 |
| T/P/K<br>Fresh | 2 | 65 | 64 (98.5) | 65 (100) | 64(98.5) | 65<br>(100) | 0 |
| T/P/K<br>Frozen | 1 | 65 | 65 (100) | 65 (100) | 65 (100) | 65<br>(100) | 0 |
| T/P/K<br>Frozen | 2 | 65 | 65 (100) | 64 (98.5) | 65 (100) | 65<br>(100) | 0 |
| Negative<br>fresh | 1 | 65 | 0 | 0 | 0 | 65<br>(100) | 9* |
| Negative<br>fresh | 2 | 65 | 0 | 0 | 0 | 65<br>(100) | 0 |
| Negative<br>frozen | 1 | 65 | 0 | 0 | 0 | 65<br>(100) | 3* |
| Negative<br>frozen | 2 | 65 | 0 | 0 | 0 | 65<br>(100) | 3* |
Table 3. Control Lots 1 and 2, Fresh vs. Frozen Equivalence
*Invalid tests were repeated and provided negative results.
External Control Reproducibility. The lot-to-lot reproducibility of the frozen external controls was further assessed by evaluating the T2 signal obtained from four production lots of the frozen controls. Ten replicates of each of the three frozen controls (A/P/G positive, T/P/K positive and negative) from each of the four lots were evaluated. The average T2 signal, %CV and % positive was evaluated for each lot. Results indicated that the four lots provided equivalent results.
Frozen Control Stability. The stability of the frozen positive and negative controls was also assessed. Three lots of each of the three controls (A/P/G positive, T/P/K positive and negative) were tested using the T2Candida Panel after 0 days, 30 days and 60 days of storage at -20° C. For each time point, the mean of the measured T2 signals from the A/T, P and K/G channels for each of the controls was expected to be statistically equivalent to the mean T2 signals measured in those channels on Day 0. At the time of clearance, all controls showed acceptable performance for at least 60 days.
{8}------------------------------------------------
### d. Detection Limit:
LoD testing was performed using two strains of each species targeted by the T2Candida Panel and performed on the T2Dx® Instrument. LoD testing consisted of an initial screening phase and a confirmatory phase. In the initial screening phase triple or double spiked samples (A/G/P or T/K) were prepared at 5 concentrations (8, 4, 2, <1 CFU/mL and negative) and stored at 2 - 8° C until testing. (The applicability of using multiple-spiked samples was evaluated in a separate study, see below). Seven replicates per spike level/Candida strain were tested to determine the putative LoD, which was the lowest concentration of Candida in which the positive detection rate met or exceeded 95%. These putative concentrations were used in the confirmatory phase to determine the exact LoD for each species. In the confirmatory phase, a total of at least 20 replicates of each of two strains per species were tested using single-organism-spiked samples. The LoD was confirmed at the putative LoD concentration if no fewer than 19 of the 20 replicates gave a positive signal for the appropriate organism (95.0% detection rate). For one strain each of two species, C. albicans and C. parapsilosis, the confirmatory testing did not validate the putative LoD determined in the initial studies. Those strains were retested at the next higher dilution, resulting in a 100% positive detection rate: this concentration was determined to be the LoD for the species. Results of the confirmatory LoD testing are shown in Table 4 below.
| | Strain 1 | | Strain 2 | | |
|-----------------|----------------------|--------|----------------------|--------|------------------|
| Species | # Positive/Total (%) | CFU/mL | # Positive/Total (%) | CFU/mL | Final LoD CFU/mL |
| C. albicans | 19/20 (95) | 1 | 21/21 (100) | 2 | 2 |
| C. tropicalis | 20/20 (100) | 1 | 21/21 (100) | 1 | 1 |
| C. parapsilosis | 20/20 (100) | 2 | 20/20 (100) | 3 | 3 |
| C. glabrata | 20/20 (100) | 2 | 20/20 (100) | 2 | 2 |
| C. krusei | 19/20 (95) | 1 | 19/20 (95) | 1 | 1 |
### Table 4. LoD Results
Invalid results: 23/845 (2.7%) False positives: 5/845 tests (0.6%)
Single Species Spike, Multi-species Spike Equivalence. e.
A comparison of the equivalence of results obtained for samples spiked with a single species of Candida vs. samples spiked with three species of Candida was performed. This study was performed to determine the applicability of testing multiple species in a single blood specimen during the analytical studies.
A single strain of each of the five Candida species was tested at concentrations of 2-3X LoD. Twenty replicates of each single spike and multi-spike sample was tested; multi-spiked samples contained the combinations of either C. albicans, C. parapsilosis, and C. glabrata (each at 2-3 X their specific LoD) or C. tropicalis, C. parapsilosis, and C. krusei (each at 2-3 X their specific LoD). The combinations
{9}------------------------------------------------
were chosen to provide a specific result for each detection channel of the instrument.
Results from samples containing three Candida species were positive in each channel and were determined to be equivalent to the results for the same species when tested using a blood specimen spiked with only a single species.
Invalid Results: 5/281 (1.8%) Instrument failure: 4/281 (1.4%) False positive: 2/281 (0.7%)
### f. Analytical Sensitivity:
Fifteen human strains of each target species were tested the T2Candida Panel and T2Dx Instrument. The identification of all isolates was confirmed by sequence analysis of the ITS2 region of the ribosomal operon. Isolates were tested in triplicate at 2-3X LoD; testing was repeated for strains which were not detected. Results of the analytical sensitivity study are shown in Table 5 below.
| Species | No. strains tested/no.<br>positive (%) |
|-----------------|----------------------------------------|
| C. albicans | 15/15 (100%) |
| C. tropicalis | 14/15 (93.3%)* |
| C. krusei | 15/15 (100%) |
| C. glabrata | 15/15 (100%) |
| C. parapsilosis | 15/15 (100%) |
### Table 5. Analytical Sensitivity Results
*Repeat testing of 20 replicates of the C. tropicalis isolate that was not detected gave positive results in all replicates.
False positives: 2/225 (0.9%)
#### Co-infection Studies: g.
A competitive inhibition study was performed to evaluate the sensitivity of the T2Candida Panel and T2Dx® Instrument to detect Candida present at a concentration of 1-2X LoD in the presence of other clinically relevant organisms that may be present in a co-infection. Three combinations of organisms were tested in whole blood samples: 1) samples containing two Candida target species, both in concentrations at 1-2X LoD; 2) samples containing a Candida species at 1-2X LoD in combination with a different Candida species at a concentration of 100 CFU/ML: 3) samples containing a Candida species at [b](4) TS/CC! in combination with members of other genera at a concentration of 100 CFU/mL. The high concentration of 100 CFU/mL was considered to be clinically relevant as two mL of blood is used in the assay resulting in an organism concentration of 200 CFU/assay.
{10}------------------------------------------------
Thirty one pairs of different Candida species, both at a concentration of 1-2X LoD, were tested in replicates of four for a total of 124 tests. Of these, 118 tests were positive for both Candida species (95.2%).
Sixty three pairs of two Candida species were tested in replicates of four with one species spiked at 1-2X LoD and the other species spiked at 100 CFU/mL for a total of 252 tests. The Candida species spiked at 100 CFU/mL were detected in all samples. Two hundred forty four tests were positive for the Candida species spiked at 1-2X LoD (96.8%).
Fifty samples including Candida species (spiked at 1–2X LoD) in combination with representatives of other genera (Pseudomonas aeruginosa, Staphylococcus aureus, Staphylococcus epidermidis, Staphylococcus saprophyticus and Streptococcus pneumoniae, spiked at 100 CFU/mL) were tested in replicates of four for a total of 200 tests. Of these, the Candida species was detected in 189 (94.5%) tests. Results of the co-infection study are shown in Table 6 below.
Invalid Results: 24/1597 (1.5%) False Positives: 5/1597 (0.3%)
### Table 6. Results of Competitive Inhibition Studies
| Organism<br>combinations | Concentration | No.<br>Pairs<br>Tested | Total No. of<br>Tests (4<br>Replicates<br>Per<br>Organism<br>Combination) | No. of Positive<br>Tests/Total No. of<br>Tests (%) | 95% CI |
|-------------------------------|-------------------------|------------------------|---------------------------------------------------------------------------|----------------------------------------------------|----------------|
| Candida<br>sp./Candida<br>sp. | Both at 1 - 2X<br>LoD | 31 | 124 | 118/124 (95.2%) | 89.8 -<br>97.8 |
| | 1-2X LoD/<br>100 CFU/mL | 63 | 252 | 244/252 (96.8%) | 93.8 -<br>98.4 |
| Candida sp./<br>other genus | 1-2X LoD/<br>100 CFU/mL | 50 | 200 | 189/200 (94.5%) | 90.4 -<br>96.9 |
#### f. Analytical Specificity:
A cross reactivity study was performed using 80 non-target, clinically relevant or environmental organisms including 21 yeast species, nine viruses, 25 fungi and 25 species of bacteria. Isolates were initially tested in triplicate at a concentration of 100 CFU/mL for yeast, molds and bacteria, and viruses were tested at a concentration of 102 PFU/mL. Any strain which showed cross reactivity or gave an invalid result was further evaluated at lower, more clinically relevant concentrations of organisms in blood (100, 33 and 10 CFU/mL).
Cross reactivity was defined as an increase in the T2 signal above the established cutoff for the Candida detection channels when tested at clinically relevant
{11}------------------------------------------------
concentrations. Cross reactivity required both amplification of the organism with the pan-Candida PCR primers and subsequent detection with any of the capture probes.
Forty one species tested at a concentration of 106 CFU/mL (10° PFU/mL for viruses) showed no cross reactivity and had valid IC values. These species are listed below in Table 7.
| Bacteria | |
|--------------------------|------------------------------|
| Acinetobacter baumannii | Staphylococcus aureus MRSA |
| Bacteroides fragilis | Staphylococcus auricularis |
| Clostridium perfringens | Staphylococcus epidermidis |
| Enterobacter cloacae | Staphylococcus haemolyticus |
| Klebsiella oxytoca | Staphylococcus hominis |
| Klebsiella pneumoniae | Staphylococcus intermedius |
| Morganella morganii | Staphylococcus saprophyticus |
| Pseudomonas aeruginosa | Staphylococcus warneri |
| Serratia marcescens | Streptococcus mutans |
| Enterococcus faecalis | Streptococcus pneumoniae |
| Staphylococcus aureus | Streptococcus pyogenes |
| Fungi | Viruses |
| Acremonium kiliense | Adenovirus |
| Malassezia furfur | Cytomegalovirus |
| Malassezia pachydermatis | Enterovirus |
| Mucor oblongiellipticus | Epstein-Barr Virus |
| Phialophora richardsiae | Hepatitis A |
| Rhizomucor microsporous | Hepatitis B |
| Rhizopus pusillus | Herpes simplex Virus 1 |
| Rhizopus oryzae | Herpes simplex Virus 2 |
| Scedosporium prolificans | Varicella zoster Virus |
Table 7. Species Providing Valid IC Values and No Cross Reactivity
Thirty species tested at a concentration of 106 CFU/mL gave an invalid result in at least one replicate due to a low IC value but gave no positive signals in any of the Candida channels. Further testing with these strains at clinically relevant concentrations (100, 33 and 10 CFU/mL) provided negative T2Candida results with acceptable IC values and were determined to be non-cross reacting (See Table 8 below).
Five species (Candida rugosa, Candida sojae, Acinetobacter lwoffii, E. coli and E. faecalis) gave positive Candida results (and either valid or invalid IC results) in one or more replicates when tested at a concentration of 10° CFU/mL. Repeat testing at clinically relevant organism concentrations (100, 33 or 10 CFU/mL) showed no cross reactivity (Table 8 below).
{12}------------------------------------------------
Table 8. Species Providing an Invalid IC or Positive Candida Results When Tested at 10° CFU/mL but Not When Tested at Clinically Relevant Concentrations; Not Considered to be Cross Reactive
| Organisms Giving Invalid IC Results at 106 CFU/mL | |
|---------------------------------------------------------|----------------------------|
| Candida albidus | Aspergillus flavus |
| Candida dubliniensis | Aspergillus fumigatus |
| Candida gigantensis | Aspergillus niger |
| Candida guilliermondii | Aspergillus terreus |
| Candida kefyr | Exophiala xenobiotica |
| Candida lunata | Fusarium proliferatum |
| Candida lusitaniae | Fusarium oxysporum |
| Candida nivariensis | Fusarium solani |
| Candida norvegensis | Kluyveromyces delphensis |
| Candida pelliculosa | Pichia anomala |
| Candida utilis | Paecilomyces variotii |
| Candida viswanathii | Scopulariopsis brevicaulis |
| Cryptococcus neoformans | Trichosporon asahii |
| Rhodotorula glutinis | Trichosporon inkin |
| | Trichosporon mucoides |
| | Trichoderma reesei |
| Organisms Giving Positive Candida Results at 106 CFU/mL | |
| Candida rugosa | Acinetobacter lwoffii |
| Candida sojae | Escherichia coli |
| | E. faecalis |
The following species were shown to cross react when tested at any concentration: C. bracarensis, S. cerevisiae, C. metapsilosis and C. orthopsilosis (Table 9 below).
In order to determine the effect of the presence of bacteria in the specimen on the IC value in actual clinical specimens, a review of the T2 signal data for 63 prospectively analyzed specimens that were positive for bacteria (as determined by the corresponding blood culture) was performed. The T2 signals for the internal control for all specimens was greater than the cutoff value, indicating that for samples containing clinically relevant organism concentrations, the internal control would be within acceptable levels and would not result in an invalid test result.
{13}------------------------------------------------
| | T2Candida Results at Various Concentrations | | | | | | | | | | | | | | | |
|---------------------------|---------------------------------------------|---|-----|---------|------------|---|-----|-------|-----------|---|-----|-------|-----------|---|-----|-------|
| | 106 CFU/mL | | | | 100 CFU/mL | | | | 33 CFU/mL | | | | 10 CFU/mL | | | |
| Organism | A/T | P | K/G | IC | A/T | P | K/G | IC | A/T | P | K/G | IC | A/T | P | K/G | IC |
| Candida bracarensis | N | N | P | Invalid | N | N | P | Valid | N | N | P | Valid | N | N | P | Valid |
| Saccharomyces cerevisiae* | N | N | P | Invalid | N | N | P | Valid | N | N | P | Valid | N | N | P | Valid |
| Candida orthopsilosis | N | P | N | Invalid | N | P | N | Valid | N | P | N | Valid | N | P | N | Valid |
| Candida metapsilosis | N | P | N | Invalid | N | P | N | Valid | N | P | N | Valid | N | P | N | Valid |
# Table 9. Species determined to be Cross Reactive
*Five strains evaluated
{14}------------------------------------------------
#### Interfering Substances g.
An interfering substances study was performed to determine and characterize the effects of potential endogenous and exogenous interfering substances on the performance of the T2Candida Panel and the T2Dx® Instrument. Interference testing was performed using a paired-difference format; the potentially interfering substance was added to a Candida-spiked sample at high concentration (to simulate worse case) and the bias relative to a Candida spiked control containing no interfering substances was determined. Substances were considered to be interfering if the T2 signal (in msec) was decreased by 30% or greater as compared to the sample containing no interferent. After testing average T2MR signal, standard deviation, %CV and % positive were evaluated for each sample.
A total of seven substances representing possible patient underlying conditions were tested at concentrations from 1X to 38X above clinical levels, with most at 3X clinical levels (Table 10).
| Underlying Source<br>or Condition | Endogenous<br>Interferent | Reference Level | Test Concentration |
|-----------------------------------|-----------------------------|------------------------------------------------------|-----------------------------------------------|
| Leukocytosis | Human DNA<br>(Buffy Coat) | WBC 3.5-11E7<br>cells/mL | 5E7 cells/mL |
| Icterus | Bilirubin<br>(conjugated) | 0-3.4 µmol/L | 50 mg/L |
| | Bilirubin<br>(unconjugated) | 5-21 µmol/L | 150 mg/L |
| | ALT | 17-50 U/L (plasma) | ≥ 3 times the upper<br>normal limit of 50 U/L |
| | AST | 9-48 U/L | ≥ 3 times the upper<br>normal limit of 48 U/L |
| Hemolysis | Hemoglobin | 1-2 g/L | 22.4 g/L |
| Lipemia | Intralipid | 0.34-3.7 mmol/L or<br>0.065 mg/dL<br>(triglycerides) | 1.3 mg/dL |
| Hyperproteinemia | Protein (albumin) | 39-51 g/L | 40 g/L |
| | Immunoglobulin G | 6-13 g/L | 15 g/L |
| Renal Failure | Creatinine | 6-17 mg/L | 50 mg/L |
| | Urea | 1.1-14.3 mmol/L | 42.9 mmol/L (2.58<br>mg/mL) |
| Multiple | Circulating human<br>DNA | N/A | 1 µg/mL |
Table 10. Endogenous Substances Evaluated to Represent Human Underlying Conditions
{15}------------------------------------------------
Twenty four exogenous substances were tested at concentrations from 1X to 38X above clinical levels (Table 11). Potential exogenous interferents were tested in sets of three for a total of eight interferent pools. All interferents were added to healthy donor blood; Candida species were spiked into the samples at 3X LoD using inocula containing either the multi-spike combinations of C. albicans/C. parapsilosis/C. glabrata or C. tropicalis/C. parapsilosis/C. krusei and tested in triplicate. If assay interference, as measured by a decrease in T2 signal by 30% or greater, was detected with any of the interferent pools, each interferent present in the pool was prepared as an individual test to determine which substance was acting as an interferent with the assay's performance.
Results indicated interference with the signal for Candida species or with the signal for the internal control for the following substances: Feraheme, Magnevist, EDTA, Ablavar, and samples with an interferent simulating lipemia (Intralipid).
Invalid Results: 12/1647 (0.7%) False Positives : 5/1647 (0.3%)
| Exogenous Interferent | Reference Level | Test Concentration |
|--------------------------------------------------------------------------|-------------------------|--------------------------|
| EDTA | 2 mg/mL | 6 mg/mL |
| Heparin | 350-1000 U/L | 3000 U/L |
| Calcium Hypochlorite | NA | 8% (80 mg/mL) |
| Fluconazole | 65.2-81.5 µmol/L | 245 µmol/L (75 µg/mL) |
| Micafungin | 100 mg dosage/day | 60 mg/L |
| Ferumoxytol (Feraheme) | 510 mg/dosage, 30 mg/mL | 306 mg/L |
| MRI Contrast Agent:<br>Magnevist (gadopentetate<br>dimeglumine, Gd-DTPA) | 0.01 mmol/kg | 5.4 mmol/L (6.86 mg/mL) |
| MRI Contrast Agent: Ablavar<br>(gadofosveset or Vasovist) | 0.03 mmol/kg | 1.6 mmol/L (1.56 mg/mL) |
| Amphotericin B Trihydrate | 1 mg/kg/day | 54.5 mg/L |
| Amphotericin B, liposomal<br>(Ambisome) | 3-5 mg/kg/day | 270 mg/L |
| Piperacillin/Pipril<br>(Piperacillin) | 6g | 3.6 g/L |
| Vancomycin | 12.4-27.6 µmol/L | 69.0 µmol/L (100 µg/mL) |
| Imipenem/Cilastatin<br>(Primaxin) | 1 g | 600 mg/L |
| Ciprofloxacin | 1.51-15.1 µmol/L | 30.2 µmol/L (10µg/mL) |
| Tazobactam (Tazobac) | 4.5 g | 8.1 g/L |
| Gentamycin sulfate | 10.5-20.9 µmol/L | 21 µmol/L (10 µg.mL) |
| Linezolid | 600 mg | 360 mg/L |
| Azithromycin (Zithromax) | 0.4-5.1 µmol/L | 15.3 µmol/L (11.5 µg/mL) |
| Clindamycin (Cleocin) | 4.0-29.7 µmol/L | 89.1 µmol/L (41.4 µg/mL) |
| Metronidazole | 35-234 µmol/L | 701 µmol/L (119.9 µg/mL) |
### Table 11. Exogenous Substances Evaluated
{16}------------------------------------------------
| Exogenous Interferent | Reference Level | Test Concentration |
|-----------------------|------------------|------------------------------------|
| Caspofungin | 70 mg | 42 mg/L |
| Lisinopril | 0.25 $\mu$ mol/L | 0.74 $\mu$ mol/L (326.7 $\mu$ g/L) |
| Cytarabine | 6 mg/kg | 324 mg/L |
#### h. Assay Cut-off:
To establish the cutoff values for the A/T, P, K/G and internal control detections, a limit of blank study was conducted using negative blood specimens from healthy (N=113) and unhealthy (N=39) patients, with two different reagent lots. The T2MR signal, in milliseconds (msec), was evaluated to establish the upper limit of T2 signal distribution on negative measurements and lower limit of T2 signal of positive internal control measurements. These limits encompassed 99% of the no target signals and 99% of the positive internal control signals, respectively. The mid-point of the interval between these upper and lower limits was 65 ms and was set as the clinical cutoff for A/T, P and K/G detection. The upper limit was 85 ms and was set as the cutoff for the internal control detection. Validation of the cutoff was completed by analysis of results obtained during the prospective clinical studies. Cutoff values are indicated in Table 12 below.
In the LoD studies, T2MR signals were typically (b)(4) Ts/CC msec for negative detections and ranged from (b)(4) TS/CCl msec to (b)(4) TS/CC| msec for positive detections.
| Detection<br>Channel | Lower limit of<br>valid signal (ms) | Clinical cutoff<br>(ms) |
|----------------------|-------------------------------------|-------------------------|
| A/T | 20 | 65 |
| P | 20 | 65 |
| K/G | 20 | 65 |
| IC | 85 | 85 |
## Table 12. Assay Cutoff Values
Application of assay cutoff to patient specimens. In the event that one or more of the A/T. P. and K/G particle detection signals is below the lower limit of valid signal (20ms) for a patient specimen, that entire specimen is deemed invalid because the specific detection reactions that were below the lower limit of valid signal had an associated failure.
#### i. Carryover/Cross Contamination Studies
A cross contamination study was performed to determine the rate of false positivity caused by organism carryover within the T2Dx Instrument. A checkerboard study (alternating high positives and negatives) was performed using two runs on five instruments using multi-spiked samples prepared to contain 100 CFU/mL (200 CFU/assay or 'test") interspaced with negative samples. An additional study was
{17}------------------------------------------------
performed using samples spiked with 1000 CFU/mL (2000 CFU/assay).
A possible total of 175 negative samples were run. Three of these samples were false positive (all with samples spiked at 100 CFU/mL) for a false positive rate of 1.7%.
#### j. Specimen Stability at 2-8° C
In order to support a labeling recommendation for storage of blood samples at 2 - 8°C prior to testing, a specimen stability study was performed. Ten replicates of whole blood specimens were spiked with each of the target Candida species at a concentration of 100 CFU/mL. The organism concentration was determined by colony count at Day 0, Day 1 and Day 3 of storage. Results indicated that the organism concentration in samples held for 1 day at 2 - 8 °C showed a change in organism concentration of -9% to +12.4%. Colony count results obtained after three days of storage showed a decrease in organism concentration of -4% to -17%. Device labeling indicates that blood specimens should be tested as soon as possible after collection and that specimens held for longer than one day at 2 - 8° C may result in a decrease of viable organisms in the specimen.
#### k. Lysis Tube and Reagent Stability
A stability study was performed to determine the shelf-life of the reagents and the lysis tubes at the recommended temperatures. Reagents were removed from the kits and each component was evaluated using quality control methods applied to the bulk reagents prior to kit fill. Results were compared to results obtained with recently manufactured reagents; all acceptance criteria were met indicating all reagents included in the kit were stable for at least seven months from date of manufacture. The following storage instructions will be recommended for the packaging and labeling: T2Candida Reagent Packs should be stored at 2-8°C; the T2Candida Cartridge should be stored at 15-30°C. The expiration date of each component will be indicated on the component label.
#### l. Shipping Stability
The T2Candida shipping containers containing Reagent Packs, the Cartridge Kit and the External Controls were tested for component, labeling and packaging damage according to the ISTA 2A drop test. Containers were dropped in several package orientations from a drop height of 38 inches. All shipping configurations passed the pre-specified acceptance criteria.
To evaluate the effects of thermal exposure, packaged components were exposed to temperatures between 24.5 and 30.4° C for 36 hours and between 4.3 and 14.2°C for 36 hours. After exposure the packaged T2Candida Panel components were functionally tested on the T2Dx® Instrument. Results were compared to results obtained by the same lot of T2Candida panel components stored at the recommended storage temperatures. All components passed the pre-specified acceptance criteria, that thermal cycle exposed T2Candida components were not statistically different
{18}------------------------------------------------
(p>0.5) than the control components.
- 2. Comparison Studies:
- a. Method Comparison with Predicate Device: Not Applicable
- b. Matrix Comparison: Not Applicable
### 3. Clinical Studies:
- a. Clinical Sensitivity:
Because of the low prevalence of Candida-positive blood cultures from clinical specimens (prevalence <0.5%), the sensitivity of the T2Candida Panel and T2Dx Instrument was determined using contrived samples. Contrived samples were prepared with representatives of each of the target Candida species and were tested with the T2Candida panel. Fifty unique Candida isolates of each of the five T2Candida targets were spiked into whole blood at specific concentrations for a total of 250 positive samples. In addition 50 unspiked blood specimens were included as negative samples and were tested randomly and in a blinded fashion throughout the studv.
# Isolates. Isolates were obtained from (b)(4) TS/CCI
(6)(4) TSCS), a reference laboratory and a clinical laboratory. The ITS region of each isolate was sequenced to confirm its identification; for all isolates the BLAST results of the determined sequence showed > 99.2% similarity to C. albicans, C. tropicalis, C. krusei, C. glabrata or C. parapsilosis, as appropriate.
Blood specimens. Three K2EDTA tubes containing 4 mL of blood in each tube were drawn from patients for whom blood cultures were ordered at three clinical sites. Thirty-four percent of specimens were obtained from patients determined to have some level of immunocompromise. Tubes were transported to T2Biosystems within 24 hours of collection under refrigeration conditions. In the event that a blood culture (from the same patient draw as the K2EDTA tubes) was determined to be Candida positive for the patient blood specimen. the contrived specimen's T2 analytical result was declared invalid and not analyzed as part of the T2 study.
Spiking. Contrived samples were spiked with a specific concentration of Candida based on Table 13 below to represent clinically relevant organism concentrations; a majority of samples contained Candida levels near the limit of detection.
{19}------------------------------------------------
| Organism (LoD*) | <1 CFU/mL | 1 – 10 CFU/mL | 11 – 30 CFU/mL | 31 – 100 CFU/mL |
|---------------------|-----------|---------------|----------------|-----------------|
| C. albicans (2) | 10 | 18 | 17 | 5 |
| C. tropicalis (1) | 10 | 18 | 17 | 5 |
| C. krusei (1) | 10 | 18 | 17 | 5 |
| C. glabrata (2) | 10 | 18 | 17 | 5 |
| C. parapsilosis (3) | 10 | 18 | 17 | 5 |
| Number (%) of Total | 50 (20) | 90 (36) | 85 (34) | 25 (10) |
### Table 13. Number of Contrived Specimens/Candida Species/Titer Level
*LoD in CFU/mL
Cell bullets were prepared for each isolate and stored frozen. The concentration of organism in the cell bullets was determined by (b)(4) CCI
The
(b)(4) CCI s was used to assign an organism concentration value (b)(4) CCI was used to guide contrived sample preparation. Cell viability and cell death was evaluated by comparing cell counts obtained before and after a single freeze/thaw cycle. Results were consistent with minimal cell death after a single freeze/thaw cycle.
Within 48 hours of specimen draw, contrived samples were prepared using a randomized organism list to determine the species and strain to be used each day of preparation. Cell bullets were diluted to appropriate levels and the concentration was verified by colony count testing: three replicates, each in 4 mL of K2EDTA blood, were prepared for each organism/inoculum concentration. For each three-tube set, one tube was retained by T2 Biosystems for quality control purposes and the remaining two tubes were distributed to a testing site. Duplicate samples were shipped to assure the availability of a sample in the event one of the sample tubes was damaged during transit or otherwise malfunctioned. Only one sample was tested by the site, the untested sample was discarded.
In addition to positive spiked samples, an additional 50 un-spiked blood specimens were sent to testing sites as blinded negative samples.
Contrived samples and negative samples were distributed across four clinical sites within 24 hours of preparation and analyzed with the T2Candida Panel and T2Dx Instrument within 48 hours of receipt. Instructions for performance of contrived sample testing specified that as part of the loading procedure, refrigerated (i.e. controls and contrived) samples should be allowed to equilibrate to room temperature for ~20 minutes prior to loading. In all cases, samples were tested within five hours of equilibration to room temperature.
{20}------------------------------------------------
Assay. Contrived samples were tested with the T2Candida Panel on a T2Dx instrument at four clinical sites per the testing protocol. For each specimen, T2Candida results for all four detection reactions (C. albicans/C. tropicalis, C. krusei/C. glabrata, C. parapsilosis and internal control) were collected yielding T2Candida sets for 300 different samples (250 spiked and 50 negative) for a total of 900 organism channel results. Positive percent agreement (PPA) across all organism concentrations was 94.0% for the A/T and P channels and 88.0% for the K/G channel. Results for the contrived samples by detection channel for all concentrations are shown in Table 14 below.
Results were also analyzed relative to Candida spike concentration (Table 15) and relative to species LoD (Table 16). PPA for samples spiked at concentrations > LoD showed 97.4 to 100.0% PPA across all detection channels.
All 50 negative samples gave negative results for all channels. A single false positive result (1/250, 0.4%) was seen in the parapsilosis channel with C. krusei.
| | Table 14. Contrived Specimen Performance by Detection Channel | | |
|--|---------------------------------------------------------------|--|--|
| | | | |
| Detection<br>Channel | PPA | 95 %CI | NPA | 95%CI |
|----------------------|----------------|-----------|------------------|------------|
| A/T | 94/100 (94.0%) | 87.5-97.2 | 200/200 (100.0%) | 98.1-100.0 |
| P | 47/50 (94.0%) | 83.8-97.9 | 249/250 (99.6%) | 97.8-99.9 |
| K/G | 88/100 (88.0%) | 80.2-93.0 | 200/200 (100.0%) | 98.1-100.0 |
Abbreviations: PPA, Positive percent agreement; NPA, Negative percent agreement; A/T, C. albicans/C. tropicalis channel; P, C. parapsilosis channel; K/G, C. krusei, C. glabrata channel
{21}------------------------------------------------
| Detection<br>Channel | <1 CFU/mL | | 1-10 CFU/mL | | 11-30 CFU/mL | | 31-100 CFU/mL | |
|----------------------|---------------|-------------|---------------|-------------|---------------|-------------|----------------|------------|
| | PPA | 95% CI | PPA | 95% CI | PPA | 95% CI | PPA | 95% CI |
| A/T | 16/20 (80.0%) | 58.4 - 91.9 | 34/36 (94.4%) | 81.9 - 98.5 | 34/34 (100%) | 89.8 - 100 | 10/10 (100.0%) | 72.2 - 100 |
| P | 8/10 (80.0%) | 49.0 - 94.3 | 17/18 (94.4%) | 74.2 - 99.0 | 17/17 (100%) | 81.6 - 100 | 5/5 (100%) | 56.6 - 100 |
| K/G | 11/20 (55.0%) | 34.2 - 74.2 | 34/36 (94.4%) | 81.9 - 98.5 | 33/34 (97.1%) | 85.1 - 99.5 | 10/10 (100%) | 72.2 - 100 |
Table 15. Contrived Specimen PPA by Detection Channel and Relative to Candida Spike Concentration
ªAbbreviations: A/T, C. albicans/C. tropicalis channel; P, C. parapsilosis channel; K/G, C. glabrata, C. krusei channel
| Detection<br>Channel | Species | LoD (CFU/mL) | PPA at ≥ LoDb | 95% CI | PPA at < LoDc | 95% CI |
|----------------------|-----------------|--------------|---------------|--------------|---------------|-------------|
| A/T | C. albicans | 2 | 77/79 (97.5%) | 91.2 – 99.3 | 17/21 (81.0%) | 60.0 - 92.3 |
| | C. tropicalis | 1 | | | | |
| P | C. parapsilosis | 3 | 32/32 (100%) | 89.3 - 100.0 | 15/18 (83.3%) | 60.8 - 94.2 |
| K/G | C. krusei | 1 | 75/77 (97.4%) | 91.0 - 99.3 | 13/23 (56.5%) | 36.8 - 74.4 |
| | C. glabrata | 2 | | | | |
ªAbbreviations: A/T, C. albicans/C. tropicalis channel; P, C. parapsilosis channel; K/G, C. glabrata, C. krusei channel b The calculations of PPA at > LoD included a portion of the samples spiked at 1-10 CFUmL (depending on the LoD of the species) and all samples spiked at 11-30 CFU/mL and 31-100 CFU/mL).
°The calculations of PPA at < LoD included all samples spiked at < 1 CFU/mL and a portion of the samples spiked at 1-10 CFU/mL (depending on the LoD of the species)
{22}------------------------------------------------
### b. Clinical specificity:
The specificity of the T2Candida Panel and T2Dx® Instrument was determined by a prospective comparison of the results of the T2Candida Panel with results from blood culture collected from the same draw at the same anatomical site.
Patients. A total of 1501 blood specimens were drawn from adult patients who had been referred for a diagnostic blood culture per routine standard of care. Informed consent was obtained. Forty-eight percent of specimens were obtained from patients determined to have some level of immunocompromise.
Study Sites. Specimens were collected at nine geographically diverse sites; testing was performed at seven sites.
Specimens. A standard blood culture set (aerobic and anaerobic bottle set) was obtained for each patient in accordance with standard institutional practice and blood culture system manufacturer's recommendations. In addition, three 4 mL specimens in K2EDTA vacutainers for the T2Candida Panel were obtained from the same anatomical collection site as the blood culture. The first tube drawn (Tube A) is used for the T2Candida assay and the remaining tubes were transferred to T2 Biosystems for -80 ° C storage and additional testing.
Results. Based on results from the 1501 prospective specificity of the T2Candida Panel and T2Dx® Instrument was determined to be 98.8% for the A/T channel, 99.2% for the P channel and 99.2% for the K/G channel (Table 17). A total of six blood cultures were positive for Candida species (Table 18). An additional blood culture was positive for C. parapsilosis, T2Candida results were positive in the P (C. parapsilosis) channel but the specimen was excluded from the study due to a protocol violation. A total of 63 of the 1501 prospective specimens were positive with bacteria only.
The T2Candida Panel and T2Dx® Instrument correctly detected 2/4 specimens that had corresponding blood cultures positive for C. albicans, 2/2 and 1/1 specimens that had corresponding blood cultures positive for C. parapsilosis and C. glabrata, respectively (Table 18).
## Table 17. Prospective Specimen Sensitivity and Specificity by Detection Channel
| Detection<br>Channel | Sensitivity | 95%CI | Specificity | 95%CI |
|----------------------|--------------|--------------|-------------------|-------------|
| A/T | 2/4* (50.0%) | 15.0 - 85.0 | 1479/1497 (98.8%) | 98.1 - 99.2 |
| P | 2/2 (100%) | 34.2 - 100.0 | 1487/1499 (99.2%) | 98.6 – 99.5 |
| K/G | 1/1 (100%) | 20.6 - 100.0 | 1499/1500 (99.9) | 99.6 - 99.9 |
* an additional specimen collected at the same time was positive for C. albicans
{23}------------------------------------------------
| Specimen | T2Candida Results | | | Blood Culture Result |
|----------|-------------------|----------|----------|---------------------------------|
| | A/T | P | K/G | |
| 1 | Positive | Positive | Negative | C. albicans and C. parapsilosis |
| 2 | Negative | Positive | Negative | C. parapsilosis |
| 3 | Negative | Negative | Positive | C. glabrata |
| 4 | Positive | Negative | Negative | C. albicans |
| 5 | Negative* | Negative | Negative | C. albicans |
| 6 | Negative | Negative | Negative | C. albicans |
### Table 18. T2Candida Results for Blood Cultures Positive for Candida Species.
*Repeat T2Candida results from an additional blood specimen drawn at the same time were positive in the A/T channel.
> False Positives. The T2Candida Panel indicated 31 positive detections in 29 patient specimens for which Candida was not detected in corresponding blood cultures. A chart and case history review was conducted for those patients. This review revealed that one sample was obtained from a patient who had proven intra-abdominal candidiasis with no positive blood cultures. In addition, six patients were receiving anti-fungal therapy and four patients were colonized with Candida species. Repeat testing of 26 of the 29 false positive specimens using a specimen drawn at the same time as the original specimen were negative for Candida by the T2Candida assay. Three specimens were unavailable for retesting.
- Other clinical supportive data (when a. and b. are not applicable): C. Not Applicable
- 4. Clinical cut-off:
## Table 19. Cut-off values for Candida and IC Channels
| Detection<br>Channel | Lower limit of<br>Valid Signal (msec) | Clinical Cutoff<br>(msec) |
|----------------------|---------------------------------------|---------------------------|
| A/T | 20 | 65 |
| P | 20 | 65 |
| K/G | 20 | 65 |
| IC | 85 | 85 |
## 5. Expected values/Reference range:
The overall incidence of Candida species as determined by the T2Candida Panel and T2Dx® Instrument in direct blood specimens in patients tested during this study was 0.3% (4/1501). All clinical specimens collected during this study were collected between August, 2013 and October 2013 and April 2014.
{24}------------------------------------------------
## M. Instrument Name:
T2Candida Panel and T2Dx® Instrument
## N. System Descriptions:
- 1. Modes of Operation:
The system operates in a fully automated mode with limited user intervention. Reagent packs, samples and cartridges are loading manually but all processing and analysis steps are controlled and monitored by the system.
# 2. Software:
FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:
Yes __X__ or No ______________________________________________________________________________________________________________________________________________________________
Level of Concern: Moderate
## Software Description:
The T2Dx instrument software controls the following:
- T2Dx instrument interactions with peripheral devices and systems (e.g. printer, ● barcode reader, portable memory device, service laptop computer and LIS system)
- T2Dx® instrument system diagnostics, including alarms and error notifications ●
- . T2Dx® instrument graphical user interface. including touch screen prompts that guide an operator through the steps associated with loading and unloading test cartridges/patient specimens on the T2Dx® instrument and obtaining test results
- T2Dx internal operating temperatures
- Operating parameters (e.g. time, temperature, etc.) of various T2Dx instrument . subsystems.
- Initiation of workflows, scheduling of specimens for processing and monitoring ● of workflow implementation
- Determination of the result
- Administrative functions (e.g. Require Operator ID, Positive Sample Counter, . Archive Results, Shutdown OS, and Restart Application)
This T2Dx® instrument software is designed and intended to interface with the T2Candida workflow software.
## Device Hazard Analysis:
A list of potential hazards was developed by conducting a bottom up analysis of potential failure modes. Potential failure modes were identified including those which could lead
{25}------------------------------------------------
to invalid or incorrect results. The RPN value for these potential hazards exceeded the threshold, which triggered a mitigation action. As a result, solutions were implemented to verify that risks were reduced.
### Architecture Design Chart:
The T2Dx instrument software and T2Candida workflow software architecture design chart was provided along with a summary of software architecture illustrations and references to description of each illustration.
## Software Requirements Specification (SRS):
The software requirements for T2Dx® instrument and the T2Candida panel were based on discussions with potential users, site visits, and a review of current, similar diagnostic systems. The Software Requirements Specifications (SRS) were developed using these findings.
### Software Design Specification (SDS):
Requirements defined in the SRS document are implemented according to the design specifications described.
## Traceability Analysis:
A traceability matrix which links requirements, specifications, hazards, mitigations and verification & validation testing for the software was acceptable.
## Software Development Environment Description:
Software development life cycle plan and software development configuration management plan for the T2Dx® instrument software and the T2Candida workflow software was acceptable.
## Verification and Validation Testing:
Based on the Software Requirements Specification ("SRS") and Software Design Specifications ("SDS"), software verification test plans were devised to verify that the software meets the requirements. Each element of the SRS was tested and found to meet the requirements. This was accomplished via a series of system, software integration, and code inspection activities:
- Unit level code inspection performed by an independent reviewer and associated ● unit/integration level automated or manual functional testing of code
- Module level (integrated software and hardware) functional testing ●
- System level performance verification
- . System level validation
## Revision Level History:
A software revision history record for the T2Dx® instrument software and T2Candida
{26}------------------------------------------------
workflow software was acceptable.
# Unresolved Anomalies:
The T2Dx software release version 1.0.0.9 contains three unresolved anomalies. Impact analysis of these unresolved anomalies on device safety or effectiveness was acceptable.
# EMC Testing:
The electrical safety of the T2Dx Instrument has been verified in accordance with the following standards:
Safety requirements for electrical equipment for measurement, control, and laboratory use (Safety).
- . IEC 61010-1:2001 (Second Edition): Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements
- IEC 61010-2-010: 2003 (2nd Ed): 100562844BOX-004: Safety requirements for ● electrical equipment for measurement, control, and laboratory use - Part 2-010: Particular requirements for laboratory equipment for the heating of materials
- IEC 61010- 2 - 081: 2001 (1st Edition) + A1:2003 : Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-081: Particular requirements for automatic and semi-automatic laboratory equipment for analysis and other purposes
- IEC 61010-2-101: 2002 (ed.1): Safety requirements for electrical equipment for . measurement, control and laboratory use - Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment
Electromagnetic Compatibility (EMC)
- IEC 61326-1:2005: Electrical equipment for measurement, control and laboratory ● use, control and laboratory use - EMC requirements - Part 1: General requirements
- IEC 61326-2-6:2005: Electrical equipment for measurement, control and ● laboratory use - EMC requirements - Part 2-6: Particular requirements - In vitro diagnostic (IVD) medical equipment
- 3. Specimen Identification:
The T2Dx® instrument interfaces with a handheld digital imager scanner. The barcode reader can be used to identify:
- the type of assay;
- assay component lot numbers; ●
- the specimen being analyzed; ●
- the T2Dx® instrument operator, if the operator uses barcode identification. ●
- 4. Specimen Sampling and Handling:
Blood should be collected in K2EDTA vacutainer tubes; the minimum volume for testing is 3 mL. Specimens should be tested as soon as possible after collection and should be
{27}------------------------------------------------
held at room temperature (15 - 25°C) for no longer than 12 hours or for up to one day at 2 - 8° C. Holding specimens longer than one day at 2 - 8° C may result in a decrease of viable organisms in the specimen. Blood specimens should be at room temperature at the time of testing and should be inverted a minimum of 8-10 times to ensure sample homogeneity. Samples that are not visually homogenous should not be tested. The vacutainer tubes are uncapped and placed on the T2Candida Sample Inlet. The specimen and sample inlet are inserted into the instrument.
- 5. Calibration:
Calibration routines are performed automatically and require no user intervention
- 6. Quality Control:
Users are recommended to follow all laboratory procedures, local, state, and/or federal requirements and accrediting organizations guidelines for the testing of all external positive and negative control materials regardless of source.
The following recommendations are provided which relate to external quality control testing at least once per month:
- A single POSITIVE (APG or TPK) control tube of either multiplex blend and a ● single NEGATIVE control tube from the T2Candida External Controls kit is run at least once every 30 days in order to verify the continuing performance of the T2Dx® instrument and the T2Candida panel reagents. Users should alternate the multiplex blend POSITIVE control tube with each OC check.
- . A POSITIVE APG, a POSITIVE TPK control tube and a NEGATIVE control tube from the T2Candida External Controls kit should be run when either of the following events occurs
- o A new reagent lot is received into the laboratory
- o Significant maintenance (including software upgrades) to the T2Dx Instrument
- . A single NEGATIVE control tube from the T2Candida External Controls kit should be run upon completion of the T2Dx® instrument decontamination procedure.
## O. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
Not Applicable
## P. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Parts 801 and 809 and the specials controls.
{28}------------------------------------------------
# Q. Identified Risks and Required Mitigations:
| Identified Risks | Required Mitigations |
|--------------------------------------------------------------------------------------------------------------|------------------------------------------|
| Incorrect identification of a pathogenic microorganism by the device can lead to improper patient management | Special Controls (1), (2), (3), (4), (5) |
| Failure to correctly interpret test results | Special Control (6) |
| Failure to correctly operate the instrument | Special Controls (7) and (8) |
# R. Benefit/Risk Analysis:
| Summary of<br>the Benefit(s) | The primary benefit provided by the T2Candida Panel and T2Dx® Instrument is the<br>earlier diagnosis of candidemia and early initiation of appropriate antifungal treatment.<br>Symptoms of candidemia may be nonspecific, and difficult to differentiate from other<br>infectious causes of disease. Early diagnosis of candidemia may be associated with<br>decreased morbidity and mortality for some patients. Additionally, negative testing<br>may provide additional clinician confidence when delaying empiric therapy or<br>discontinuing empiric antifungal therapy. Patients may experience fewer side effects<br>from antifungal therapy, and there may be additional benefits to antimicrobial<br>stewardship programs |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{29}------------------------------------------------
| Summary of<br>the Risk(s) | False positive and false negative results resulting from incorrect identification of a<br>pathogenic organism by the device, failure to correctly interpret test results, or failure<br>to correctly operate the instrument are the primary risks associated with use of the<br>T2Candida Panel and T2Dx® Instrument.<br><br>A false positive result may result in unnecessary antifungal therapy, with associa…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.