Browse hierarchy: [Microbiology (MI)](/submissions/MI) → [Subpart D — Serological Reagents](/submissions/MI/subpart-d%E2%80%94serological-reagents) → [21 CFR 866.3960](/submissions/MI/subpart-d%E2%80%94serological-reagents/866.3960) → PII — Candida Species Nucleic Acid Detection System

# PII · Candida Species Nucleic Acid Detection System

_Microbiology · 21 CFR 866.3960 · Class 2_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PII

## Overview

- **Product Code:** PII
- **Device Name:** Candida Species Nucleic Acid Detection System
- **Regulation:** [21 CFR 866.3960](/submissions/MI/subpart-d%E2%80%94serological-reagents/866.3960)
- **Device Class:** 2
- **Review Panel:** [Microbiology](/submissions/MI)

## Identification

A nucleic acid-based device for the amplification, detection, and identification of microbial pathogens directly from whole blood specimens is a qualitative in vitro device intended for the amplification, detection, and identification of microbial-associated nucleic acid sequences from patients with suspected bloodstream infections. This device is intended to aid in the diagnosis of bloodstream infection when used in conjunction with clinical signs and symptoms and other laboratory findings.

## Classification Rationale

Class II (special controls). The special controls for this device are:

## Special Controls

A nucleic acid-based device for the amplification, detection and identification of microbial pathogens directly from whole blood specimens must comply with the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed device description documentation, including the device components, ancillary reagents required but not provided, and a detailed explanation of the methodology, including primer/probe sequence, design, and rationale for sequence selection.
(2) Premarket notification submissions must include detailed documentation from the following analytical and clinical performance studies: Analytical sensitivity (limit of detection), reactivity, inclusivity, precision, reproducibility, interference, cross reactivity, carryover, and cross contamination.
(3) Premarket notification submissions must include detailed documentation from a clinical study. The study, performed on a study population consistent with the intended use population, must compare the device performance to results obtained from well-accepted reference methods.
(4) Premarket notification submissions must include detailed documentation for device software, including, but not limited to, software applications and hardware-based devices that incorporate software.
(5) The device labeling must include limitations regarding the need for culture confirmation of negative specimens, as appropriate.
(6) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling.
(7) Premarket notification submissions must include details on an end user device training program that will be offered while marketing the device, as appropriate.
(8) As part of the risk management activities performed under 21 CFR 820.10(c) design and development, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument.

## Recent Cleared Devices (3 of 3)

| Record | Device Name | Applicant | Decision Date | Decision |
| --- | --- | --- | --- | --- |
| [K234063](https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PII/K234063.md) | T2Candida 1.1 Panel | T2biosystems, Inc. | Sep 13, 2024 | SESE |
| [K173536](https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PII/K173536.md) | T2Candida 1.1 Panel | T2biosystems, Inc. | Dec 12, 2017 | SESE |
| [DEN140019](https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PII/DEN140019.md) | T2CANDIDA AND T2DX INSTRUMENT | T2biosystems, Inc. | Sep 22, 2014 | DENG |

## Top Applicants

- T2biosystems, Inc. — 3 clearances

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PII](https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PII)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
