← Product Code [PEN](/submissions/MI/subpart-d%E2%80%94serological-reagents/PEN) · K181493

# FilmArray Blood Culture Identification (BCID) Panel (K181493)

_Biofire Diagnostics, LLC · PEN · Jul 5, 2018 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PEN/K181493

## Device Facts

- **Applicant:** Biofire Diagnostics, LLC
- **Product Code:** [PEN](/submissions/MI/subpart-d%E2%80%94serological-reagents/PEN.md)
- **Decision Date:** Jul 5, 2018
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 866.3365
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

The FilmArray Blood Culture Identification (BCID) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with FilmArray systems. The FilmArray BCID Panel is capable of simultaneous detection and identification of multiple bacterial and yeast nucleic acids and select genetic determinants of antimicrobial resistance. The FilmArray BCID Panel assay is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system.-Results are intended to be interpreted in conjunction with Gram stain results.

## Device Story

Multiplex nucleic acid test for positive blood culture samples; uses FilmArray system. Pouch contains freeze-dried reagents for automated nucleic acid purification and nested, multiplex PCR. Instrument uses inflatable bladders, seal points, and pneumatic pistons to move liquids; Peltier devices control thermal cycling. First stage: large volume multiplex PCR; second stage: singleplex nested PCR in array wells. Detection via DNA melt curve analysis using fluorescent dye. Digital camera captures images; software interprets results. Used in clinical laboratories; operated by trained personnel. Provides rapid identification (approx. 1 hour) of 24 pathogens and 3 resistance markers. Results aid clinicians in targeted antimicrobial therapy selection, potentially improving patient outcomes by reducing time to appropriate treatment.

## Clinical Evidence

Bench testing only. Validation of new raw materials for blood culture bottles demonstrated negative results across 30 tests (90% confidence of <5% false positive rate). Interference testing for BacT/ALERT FA Plus and FN Plus media using 21 contrived samples showed equivalent performance to control (BD BACTEC Plus Aerobic/F). Analytical specificity and cross-reactivity updated via in-house exclusivity testing and literature review of organism re-classifications.

## Technological Characteristics

Nested multiplex PCR; high-resolution melting analysis. Instrumentation: FilmArray System, FilmArray 2.0, or FilmArray Torch. Reagents stored at room temperature. Automated sample preparation and reagent hydration via injection vial. Moderate complexity. Software performs automated test interpretation and report generation.

## Regulatory Identification

A multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures is a qualitative in vitro device intended to simultaneously detect and identify microorganism nucleic acids from blood cultures that test positive by Gram stain or other microbiological stains. The device detects specific nucleic acid sequences for microorganism identification as well as for antimicrobial resistance. This device aids in the diagnosis of bloodstream infections when used in conjunction with other clinical and laboratory findings. However, the device does not replace traditional methods for culture and susceptibility testing.

## Special Controls

In combination with the general controls of the FD&C Act, the Verigene® Gram Positive Blood Culture Nucleic Acid Test is subject to the following special controls: The special controls for the BC-GP Assay are contained in the guideline document entitled "Class II Special Controls Guideline: Multiplex Nucleic Acid Assay for Identification of Microorganisms and Resistance Markers from Positive Blood Cultures."

*Classification.* Class II (special controls). The special control for this device is FDA's guideline document entitled “Class II Special Controls Guideline: Multiplex Nucleic Acid Assay for Identification of Microorganisms and Resistance Markers from Positive Blood Cultures.” For availability of the guideline document, see § 866.1(e).

## Predicate Devices

- FilmArray Blood Culture Identification (BCID) Panel ([K160457](/device/K160457.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

SPECIAL 510(K): DEVICE MODIFICATION OIR DECISION MEMORANDUM

510(k) Number: K181493

This 510(k) submission contains information/data on modifications made to the applicant’s own class II or class I devices requiring 510(k). The following items are present and acceptable:

1. The name and 510(k) number of the applicant’s previously cleared device. (For a preamendments device, a statement to this effect has been provided.)

2. Applicant’s statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use).

3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user’s and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.

Submitter states in the 510k Summary and Submission that modifications to the intended use and labeling does not alter the test principle or product. Submitted changes to the device include;

- Labeling updated to removal of the Manufacturer-Specific False Positive Blood Culture Bottle Warnings added as a result of recall.
- Labeling updated to expand the limitation on blood bottle contamination to address discrepant results and require concordant FilmArray BCID Panel and Grams stain results before patient reporting.
- Labeling updated to add two bioMerieux blood culture media bottle types (BacT/ALERT FA Plus and BacT/ALERT FN Plus), validated by Interference testing, for use on the FilmArray BCID Panel.
- Labeling updated to add new organism cross-reactivity identified thru post-market surveillance, customer investigations, organism re-classifications (literature), and in-house Exclusivity validation testing.

a. Comparison Information (similarities and differences) to applicant’s legally marketed predicate device including, labeling, intended use, and physical characteristics is shown in the table below.

{1}

|  Element | Modified Device: FilmArray BCID Panel (with modified intended use and labeling) | Predicate: FilmArray BCID Panel (K160457)  |
| --- | --- | --- |
|  Indications for Use | The FilmArray Blood Culture Identification (BCID) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with FilmArray systems. The FilmArray BCID Panel is capable of simultaneous detection and identification of multiple bacterial and yeast nucleic acids and select genetic determinants of antimicrobial resistance. The FilmArray BCID Panel assay is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system.-Results are intended to be interpreted in conjunction with Gram stain results. | Same  |
|  Organisms Detected | Enterococci, Listeria monocytogenes, Staphylococci (including specific differentiation of Staphylococcus aureus), Streptococci (with specific differentiation of Streptococcus agalactiae, Streptococcus pneumoniae, and Streptococcus pyogenes), Acinetobacter baumannii, Enterobacteriaceae (including specific differentiation of the Enterobacter cloacae complex, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Proteus, and Serratia marcescens), Haemophilus influenzae, Neisseria meningitidis (encapsulated), Pseudomonas aeruginosa, Candida albicans, Candida glabrata, Candida krusei, Candida parapsilosis, Candida tropicalis, and resistance markers mecA, vanA, vanB, and blaKPC (KPC) | Same  |
|  Analyte | DNA | Same  |
|  Specimen Types | Positive blood culture samples containing gram- positive or gram-negative bacteria and/or yeast. | Same  |
|  Technological Principles | Nested multiplex PCR followed by high resolution melting analysis to confirm identity of amplified product. | Same  |
|  Instrumentation | Single instrument FilmArray System, FilmArray 2.0 System, or FilmArray Torch System | Same  |
|  Time to result | About 1 hour | Same  |
|  Test Interpretation | Automated test interpretation and report generation. User cannot access raw data. | Same  |
|  Reagent Hydration and Sample Loading | FilmArray Injection Vial-based loading procedure | Same  |
|  Sample Preparation Method | Sample Processing is automated in the FilmArray BCID pouch. | Same  |
|  Reagent Storage | Reagents are stored at room temperature. | Same  |
|  Controls | Two controls are included in each reagent pouch to control for sample processing and both stages of PCR and melt analysis. | Same  |
|  User Complexity | Moderate | Same  |

The FilmArray BCID Panel Instruction Booklet was updated to Remove False Positive Warnings for Manufacturer-Specific Blood Culture Bottles

{2}

In 2014, BioFire reported an increased risk of false positive results (specifically for Enterococcus and Pseudomonas aeruginosa) when the FilmArray BCID Panel is used with bioMérieux BacT/ALERT Standard Anaerobic (SN) blood culture bottles. The source of these false positive results was identified, corrected, and removed. Therefore, the limitation and similar warnings regarding this throughout the Instruction Booklet were removed.

4. A Design Control Activities Summary which includes:

a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis.

Risk analysis was performed to identify risks, their possible causes, and appropriate control mechanisms. All risks were evaluated in the context of 21 CFR 807.81(a)(3) and FDA’s guidance document ‘510(k) Device Modifications: Deciding When to Submit a 510(k) for a Change to an Existing Device’. Upon analysis, the following risks were found:

- The assessment identified no new user risks for False Negative Results, False Negative Results, or user Injuries.
- The firm identified two false-positive associated user risks.

- The firm determined the false positive risk was low based on testing of ‘false bottle rings’ in automated blood culture instruments. The firm determined that the false positive risk is mitigated through the updated labeling which states that the FilmArray BCID Panel ‘results are intended to be interpreted in conjunction with Gram stain results’.
- The firm determined that there is a risk that unpredictable and transient contamination of blood culture bottle media with nucleic acids from pathogens detected by the FilmArray BCID Panel affects assay performance and patient results. The firm determined that the risk is mitigated through the updated labeling which informs the user to not report conclusive findings unless FilmArray BCID Panel results are concordant with the clinical profile (i.e., other laboratory, epidemiological, or clinical findings).

b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied

In order to address the increased false positive risks for Enterococcus and Pseudomonas aeruginosa when the FilmArray BCID Panel is used with bioMérieux BacT/ALERT Standard Anaerobic (SN) blood culture bottles, BioFire identified the source of contamination (raw material, Gelatin Yeast PP90M DE), made corrections to the cleaning processes, manufacturing process, raw material specifications, and replaced the contaminating raw material with a similar but highly-filtered product.

Validation testing with three lots of the new raw material tested on the FilmArray BCID Panel demonstrated negative results for all 30 tests across all targets. BioFire believes that this test provides 90% confidence that SN bottles produced with the new PP90M MF raw material and manufacturing improvements will provide acceptable performance (i.e.,

{3}

lower than 5% False Positive rate). Additionally, the firm reported a reduction in false-positive related customer complaints related to Enterococcus and Pseudomonas aeruginosa organism.

Interference testing was conducted to add two bioMerieux blood culture media bottle types (BacT/ALERT FA Plus and BacT/ALERT FN Plus) for use with the FilmArray BCID Panel. In the validation study, seven contrived blood culture samples (six positive and one negative) were spiked with panel organism into (3) bottle types (BacT/ALERT FA Plus and BacT/ALERT FN Plus and the control BD BACTEC Plus Aerobic/F) and tested with the FilmArray BCID Panel. All 21 runs were valid with passed controls and no errors or unexpected results demonstrating that the resin-containing BacT/ALERT FA Plus and BacT/ALERT FN Plus bottle types do not interfere with the FilmArray BCID Panel and results are equivalent to those obtained with the BACTEC Plus Aerobic/F bottle.

Analytical Specificity and Crossreactivity testing was conducted to update labeling with newly identified organism not previously found during the initial premarket analytical and clinical studies. BioFire identified organism through thru post-market surveillance, customer investigations, organism re-classifications (literature). In-house Exclusivity testing was conducted on available organism using the protocol used to characterize the initial FilmArray Blood Culture Identification Panel assays. In-house validation testing and literature review identified multiple re-classifications and cross-reactant organisms to the current FilmArray BCID Panel assay. All identified crossreactants were added to the FilmArray BCID Panel Instruction Booklet Predicted and Observed Cross-Reactivity with On-Panel or Off-Panel Organisms table as crossreactants and noted as limitations in other appropriate locations in the FilmArray BCID Instruction Booklet. The new Cross-Reactive Organisms are **bolded** in table below.

Predicted and Observed Cross-Reactivity with On-Panel or Off-Panel Organisms

|  FilmArray BCID Panel Result | Cross-Reactive Organism(s)/Isolate(s)/Gene  |
| --- | --- |
|  Gram-positive Bacteria  |   |
|  Enterococcus | Some coagulase-negative Staphylococci^{a}  |
|  Gram-negative Bacteria  |   |
|  Acinetobacter baumannii | Acinetobacter calcoaceticus-baumannii (ACB) complex species:
Acinetobacter calcoaceticus (ssp. anitratus)^{b}
Acinetobacter pittii (formerly genomospecies 3)^{b}
Acinetobacter seifertii^{b,c}  |
|  Escherichia coli/
Enterobacteriaceae | Escherichia fergusonnii
Shigella species (S. boydii, S. dysenteriae, S. flexneri, S. sonnei)  |
|  Enterobacter cloacae complex/
Enterbacteriaceae | Pantoea (Enterobacter) agglomerans^{c}  |
|  Klebsiella oxytoca/
Enterobacteriaceae | Klebsiella michiganensis^{d}  |

{4}

|  FilmArray BCID Panel Result | Cross-Reactive Organism(s)/Isolate(s)/Gene  |
| --- | --- |
|  Klebsiella pneumoniae/
Enterobacteriaceae | Klebsiella quasipneumoniae^{e}
Klebsiella variicola (aka Klebsiella pneumoniae variant 342)
Enterobacter aerogenes (reclassified as Klebsiella aerogenes)
Raoultella ornithinolytica^{f}  |
|  Serratia marcescens/
Enterobacteriaceae | Serratia species (S. entomophila^{g}, S. ficaria, S. odorifera^{g}, and S. rubidaea^{g})
Raoultella ornithinolytica^{f}
Pseudomonas aeruginosa (ATCC 25619)^{h}
Some Pseudomonas species (e.g. P. putida, P. poae, and P. veronii)^{i}
Some Pantoea species (e.g. P. calida, P. gavinia, and P. septica)^{i}  |
|  Haemophilus influenzae | Some Haemophilus species (H. haemolyticus, H. sputorum, and H. influenzae biotype IV (H. quentini))^{j}  |
|  Listeria monocytogenes | Listeria innocua^{c}  |
|  Yeast  |   |
|  Candida parapsilosis | Candida orthopsilosis (Group III Candida parapsilosis)^{i}
Candida multigemmis^{c}  |
|  Candida krusei | Some Streptococcus species(S. mitis, S. pneumoniae, S. dysgalactiae, S. equi, S. oralis, S. parauberis, S. pyogenes, S. salivarius, and S. thermophilus)^{i}  |
|  Antimicrobial Resistance Genes  |   |
|  vanA/B | vanM^{m}  |

a Cross-reactivity was not observed in analytical specificity testing, but is predicted by in silico analysis to occur only with some species (i.e., S. epidermidis, S. capitis and S. haemolyticus) when present in a sample at very high levels. The predicted cross-reactivity was observed infrequently in pre-analytical studies and the clinical evaluation (estimated occurrence of ~0.25% of all Staphylococcus positive patient samples).

b Acinetobacter calcoaceticus-baumannii (ACB) complex species are often mis-identified as A. baumannii by automated and manual microbial identification methods.

c Detection was not observed in analytical specificity testing, but testing and/or sequence analysis predicts that detection due to cross-reactivity may be possible at higher concentrations.

d Identified as a new species that is closely related to K. oxytoca.

e Identified as a new species (with subspecies) that is closely related to K. pneumoniae.

f Cross-reactivity was not observed when ATCC 31898 was tested at a concentration ~1×10⁸ CFU/mL, but cross-reactivity was observed in clinical positive blood cultures containing R. ornithinolytica.

g Cross-reactivity was observed only at high organism concentration (≥10⁹ CFU/mL); rare human pathogens.

h No cross-reactivity was observed with five other Pseudomonas aeruginosa isolates tested at ≥10⁸ CFU/mL.

i Primarily environmental microorganisms, may cause rare opportunistic infections in humans. The potential for cross-reactivity was identified in contrived samples and/or clinical blood culture specimens.

j Haemophilus spp. that are infrequently isolated from human blood culture and are difficult to distinguish from H. influenzae by automated and manual microbial identification methods. The potential for cross-reactivity with H. sputorum and H. quentini was identified in clinical blood culture specimens.

k Candida orthopsilosis is misidentified as C. parapsilosis by automated and manual microbial identification methods.

l Cross-reactivity was not observed in analytical specificity testing, but inefficient cross-reactivity with high levels of select streptococci (S. mitis and S. pneumoniae) has been identified via investigation of false results in contrived samples. In silico analysis also predicts potential cross-reactivity with S. dysgalactiae, S. equi, S. oralis, S. parauberis, S. pyogenes, S. salivarius, and S. thermophilus.

m Identified from a vancomycin-resistant Enterococcus faecium isolated in Asia, 2011; vanB resistance phenotype.

{5}

c) Declaration of Conformity to Design Controls

A “Declaration of Conformity” statement was submitted for the BioFire Diagnostics, LLC manufacturing facility. It was signed by the Vice President, Regulated Products and Clinical Affairs, and the Director of Quality Assurance. The statements indicate that:

- “To the best of my knowledge, the verification activities, as required for the risk analysis, for the modification were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met.”
- “The manufacturing facility, BioFire Diagnostics, LLC, is in conformance with the design control requirements”

The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the applicant’s description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The applicant has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.

6

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PEN/K181493](https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PEN/K181493)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
