← Product Code [PCH](/submissions/MI/subpart-d%E2%80%94serological-reagents/PCH) · K242367

# BIOFIRE FILMARRAY Gastrointestinal (GI) Panel (K242367)

_Biofire Diagnostics, LLC · PCH · Nov 7, 2024 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PCH/K242367

## Device Facts

- **Applicant:** Biofire Diagnostics, LLC
- **Product Code:** [PCH](/submissions/MI/subpart-d%E2%80%94serological-reagents/PCH.md)
- **Decision Date:** Nov 7, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3990
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

The BIOFIRE® FILMARRAY® Gastrointestinal (GI) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with BIOFIRE® FILMARRAY® Systems. The BIOFIRE GI Panel is capable of the simultaneous detection and identification of nucleic acids from multiple bacteria, viruses, and parasites directly from stool samples in Cary Blair transport media obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following bacteria (including several diarrheagenic E. coli/Shigella pathotypes), parasites, and viruses are identified using the BIOFIRE GI Panel: Campylobacter (C. jejuni/C. coli/C. upsaliensis) ● . Clostridiodes (Clostridium) difficile (C. difficile) toxin A/B Plesiomonas shigelloides . Salmonella . Vibrio (V. parahaemolyticus/V. vulnificus/ V. cholerae), including specific identification of Vibrio cholerae Yersinia enterocolitica ● Enteroaggregative Escherichia coli (EAEC) Enteropathogenic Escherichia coli (EPEC) . Enterotoxigenic Escherichia coli (ETEC) It/st ● . Shiga-like toxin-producing Escherichia coli (STEC) stx1/stx2 (including specific identification of the E. coli 0157 serogroup within STEC) Shigella/ Enteroinvasive Escherichia coli (EIEC) Cryptosporidium ● . Cyclospora cayetanensis . Entamoeba histolytica . Giardia lamblia (also known as G. intestinalis and G. duodenalis) Adenovirus F 40/41 Astrovirus Norovirus GI/GII Rotavirus A Sapovirus (Genogroups I, II, IV, and V) ● The BIOFIRE GI Panel is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness and results are meant to be used in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule out co-infection with organisms not included in the BIOFIRE GI Panel. The agent detected may not be the definite cause of the disease. Concomitant culture is necessary for organism recovery and further typing of bacterial agents. This device is not intended to monitor or guide treatment for C. difficile infection. Due to the small number of positive specimens collected for certain organisms during the prospective clinical study, performance characteristics for E. coli O157, Plesiomonas shigelloides, Yersinia enterocolitica, Astrovirus A were established primarily with retrospective clinical specimens. Performance characteristics for Entamoeba histolytica, and Vibrio (V. parahaemolyticus, V. vulnificus, and Vibrio cholerae) were established primarily using contrived clinical specimens. Negative BIOFIRE GI Panel results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease. A gastrointestinal microorganism multiplex nucleic acid-based assay also aids in the detection and identification of acute gastroenteritis in the context of outbreaks.

## Device Story

Multiplexed nucleic acid-based in vitro diagnostic test; uses stool samples in Cary Blair transport media. Operates on BIOFIRE FILMARRAY 2.0 or Torch systems; performs automated sample preparation, nested multiplex PCR, and high-resolution melting (HRM) analysis. Provides qualitative detection of 22 gastrointestinal pathogens. Used in clinical laboratories; results interpreted by automated software. Assists clinicians in diagnosing gastrointestinal illness; results used alongside clinical/epidemiological data. Post-market update addresses potential false-positive Norovirus GI/GII results due to cross-reactivity with commensal bacteria (e.g., Mediterraneibacter gnavus, Parabacteroides merdae).

## Clinical Evidence

Supplemental clinical study of 872 fresh stool specimens compared BIOFIRE GI Panel results against CDC CaliciNet Norovirus GI/GII assay. Norovirus GI/GII assay showed 97.1% PPA (34/35) and 96.5% NPA (808/837). Discrepant analysis identified cross-reactivity with non-viral targets (e.g., Parabacteroides, Prevotella, Ruminococcus) in 20 false-positive specimens. Results support updated labeling regarding potential cross-reactivity.

## Technological Characteristics

Nested multiplex PCR; high-resolution melting (HRM) analysis for target identification. Automated sample processing and analysis on BIOFIRE 2.0 or Torch systems. Qualitative detection of DNA/RNA. Reagents stored at room temperature; 12-month shelf life. No changes to hardware or software.

## Regulatory Identification

A gastrointestinal microorganism multiplex nucleic acid-based assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple gastrointestinal microbial nucleic acids extracted from human stool specimens. The device detects specific nucleic acid sequences for organism identification as well as for determining the presence of toxin genes. The detection and identification of a specific gastrointestinal microbial nucleic acid from individuals exhibiting signs and symptoms of gastrointestinal infection aids in the diagnosis of gastrointestinal infection when used in conjunction with clinical evaluation and other laboratory findings. A gastrointestinal microorganism multiplex nucleic acid-based assay also aids in the detection and identification of acute gastroenteritis in the context of outbreaks.

## Special Controls

*Classification.* Class II (special controls). The special controls are set forth in FDA's guideline document entitled: “Class II Special Controls Guideline: Gastrointestinal Microorganism Multiplex Nucleic Acid-Based Assays for Detection and Identification of Microorganisms and Toxin Genes from Human Stool Specimens.” For availability of the guideline document, see § 866.1(e).

## Predicate Devices

- BIOFIRE FILMARRAY Gastrointestinal (GI) Panel ([K230404](/device/K230404.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

ASSAY ONLY

## I Background Information:

A 510(k) Number

K242367

B Applicant

BioFire Diagnostics, LLC

C Proprietary and Established Names

BIOFIRE FILMARRAY Gastrointestinal (GI) Panel

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  PCH | Class II | 21 CFR 866.3990 - Gastrointestinal Microorganism Multiplex Nucleic Acid-Based Assay | MI - Microbiology  |

## II Submission/Device Overview:

### A Purpose for Submission:

To update labeling of the cleared BIOFIRE FILMARRAY Gastrointestinal Panel to include new clinical data and analytical cross-reactivity testing evaluating potential risk of false positive results for the Norovirus GI/GII assay target.

### B Measurand:

Campylobacter (C. jejuni/C. coli/C. upsaliensis), Clostridium difficile (C. difficile) toxin A/B, Plesiomonas shigelloides, Salmonella, Vibrio (V. parahaemolyticus/V. vulnificus/V. cholerae), Yersinia enterocolitica, Enteroaggregative Escherichia coli (EAEC), Enteropathogenic Escherichia coli (EPEC), Enterotoxigenic Escherichia coli (ETEC) lt/st, Shiga-like toxin-producing Escherichia coli (STEC) stx1/stx2, Shigella/ Enteroinvasive Escherichia coli (EIEC), Cryptosporidium, Cyclospora cayetanensis, Entamoeba histolytica, Giardia lamblia, Adenovirus F 40/41, Astrovirus, Norovirus GI/GII, Rotavirus A, Sapovirus (Genogroups I, II, IV, and V)

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

{1}

C Type of Test:

A multiplexed nucleic acid-based test intended for use with the FilmArray 2.0 System or the Torch System for the qualitative *in vitro* detection and identification of multiple bacteria, viruses, and parasites. The BIOFIRE FILMARRAY GI Panel assay is performed directly from stool samples in Cary Blair transport media.

III Intended Use/Indications for Use:

A Intended Use(s):

See Indications for Use below.

B Indication(s) for Use:

The BIOFIRE FILMARRAY Gastrointestinal (GI) Panel is a qualitative multiplexed nucleic acid-based *in vitro* diagnostic test intended for use with BIOFIRE FILMARRAY Systems. The BIOFIRE GI Panel is capable of the simultaneous detection and identification of nucleic acids from multiple bacteria, viruses, and parasites directly from stool samples in Cary Blair transport media obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following bacteria (including several diarrheagenic *E. coli*/Shigella pathotypes), parasites, and viruses are identified using the BIOFIRE GI Panel:

- *Campylobacter* (*C. jejuni*/C. coli/C. upsaliensis)
- *Clostridium difficile* (*C. difficile*) toxin A/B
- *Plesiomonas shigelloides*
- *Salmonella*
- *Vibrio* (*V. parahaemolyticus*/V. vulnificus/V. cholerae), including specific identification of *Vibrio cholerae*
- *Yersinia enterocolitica*
- Enteroaggregative *Escherichia coli* (EAEC)
- Enteropathogenic *Escherichia coli* (EPEC)
- Enterotoxigenic *Escherichia coli* (ETEC) lt/st
- Shiga-like toxin-producing *Escherichia coli* (STEC) stx1/stx2 (including specific identification of the *E. coli* O157 serogroup within STEC)
- *Shigella*/Enteroinvasive *Escherichia coli* (EIEC)
- *Cryptosporidium*
- *Cyclospora cayetanensis*

K242367 - Page 2 of 16

{2}

- Entamoeba histolytica
- Giardia lamblia (also known as G. intestinalis and G. duodenalis)
- Adenovirus F 40/41
- Astrovirus
- Norovirus GI/GII
- Rotavirus A
- Sapovirus (Genogroups I, II, IV, and V)

The BIOFIRE GI Panel is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness and results are meant to be used in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule out co-infection with organisms not included in the BIOFIRE GI Panel. The agent detected may not be the definite cause of the disease.

Concomitant culture is necessary for organism recovery and further typing of bacterial agents.

This device is not intended to monitor or guide treatment for C. difficile infection.

Due to the small number of positive specimens collected for certain organisms during the prospective clinical study, performance characteristics for E. coli O157, Plesiomonas shigelloides, Yersinia enterocolitica, Astrovirus, and Rotavirus A were established primarily with retrospective clinical specimens.

Performance characteristics for Entamoeba histolytica, and Vibrio (V. parahaemolyticus, V. vulnificus, and Vibrio cholerae) were established primarily using contrived clinical specimens.

Negative BIOFIRE GI Panel results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.

A gastrointestinal microorganism multiplex nucleic acid-based assay also aids in the detection and identification of acute gastroenteritis in the context of outbreaks.

## C Special Conditions for Use Statement(s):

Rx - For Prescription Use Only

For in vitro diagnostic use only

## D Special Instrument Requirements:

The BIOFIRE FILMARRAY Gastrointestinal (GI) Panel is to be used with the BIOFIRE 2.0 System or BIOFIRE Torch System.

## IV Device/System Characteristics:

K242367 - Page 3 of 16

{3}

K242367 - Page 4 of 16

# A Device Description:

The BIOFIRE FILMARRAY Gastrointestinal Panel (BIOFIRE FILMARRAY GI Panel) was originally cleared in K140407.

The BIOFIRE FILMARRAY GI Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with BIOFIRE FILMARRAY systems. The BIOFIRE FILMARRAY GI Panel is capable of the simultaneous detection and identification of nucleic acids from multiple bacteria, viruses, and parasites directly from stool samples in Cary Blair transport media obtained from individuals with signs and/or symptoms of gastrointestinal infection.

# B Principle of Operation:

The principle of operation remains unchanged from the original clearance (K140407). Refer to the original published decision summary for specific details on the principle of operation of the device.

In response to a perceived elevated rate of BIOFIRE FILMARRAY GI Panel Norovirus GI/GII false positive results from customers, post-market analytical and clinical investigations were conducted to assess potential cross-reactivities of the Norovirus GI/GII assay and to provide an updated assessment of the Norovirus assay clinical performance.

# V Substantial Equivalence Information:

# A Predicate Device Name(s):

BIOFIRE FILMARRAY Gastrointestinal (GI) Panel

# B Predicate 510(k) Number(s):

K230404

# C Comparison with Predicate(s):

|  Device & Predicate Device(s): | K242367 | K230404  |
| --- | --- | --- |
|  Device Trade Name | BIOFIRE FILMARRAY Gastrointestinal (GI) Panel | BIOFIRE FILMARRAY Gastrointestinal (GI) Panel  |
|  General Device Characteristic Similarities |  |   |
|  Intended Use/Indications For Use | The BIOFIRE FILMARRAY Gastrointestinal (GI) Panel is a qualitative multiplexed nucleic | Same  |

{4}

K242367 - Page 5 of 16

|   | acid-based in vitro diagnostic test intended for use with BIOFIRE FILMARRAY Systems. The BIOFIRE GI Panel is capable of the simultaneous detection and identification of nucleic acids from multiple bacteria, viruses, and parasites directly from stool samples in Cary Blair transport media obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following bacteria (including several diarrheagenic E. coli/Shigella pathotypes), parasites, and viruses are identified using the BIOFIRE GI Panel: • Campylobacter (C. jejuni/C. coli/C. upsaliensis) • Clostridium difficile (C. difficile) toxin A/B • Plesiomonas shigelloides • Salmonella • Vibrio (V. parahaemolyticus/V. vulnificus/V. cholerae), including specific identification of Vibrio cholerae • Yersinia enterocolitica • Enteroaggregative Escherichia coli (EAEC) • Enteropathogenic Escherichia coli (EPEC)  |
| --- | --- |

{5}

K242367 - Page 6 of 16
|   | • Enterotoxigenic Escherichia coli (ETEC) lt/st
• Shiga-like toxin-producing Escherichia coli (STEC) stx1/stx2 (including specific identification of the E. coli O157 serogroup within STEC)
• Shigella/ Enteroinvasive Escherichia coli (EIEC)
• Cryptosporidium
• Cyclospora cayetanensis
• Entamoeba histolytica
• Giardia lamblia (also known as G. intestinalis and G. duodenalis)
• Adenovirus F 40/41
• Astrovirus
• Norovirus GI/GII
• Rotavirus A
• Sapovirus (Genogroups I, II, IV, and V)

The BIOFIRE GI Panel is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness and results are meant to be used in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule out co-infection with organisms not included in the BIOFIRE GI Panel. The agent detected may not be the definite cause of  |
| --- | --- |

{6}

K242367 - Page 7 of 16

|   | the disease. |   |
| --- | --- | --- |
|   | Concomitant culture is necessary for organism recovery and further typing of bacterial agents. |   |
|   | This device is not intended to monitor or guide treatment for C. difficile infection. |   |
|   | Due to the small number of positive specimens collected for certain organisms during the prospective clinical study, performance characteristics for E. coli O157, Plesiomonas shigelloides, Yersinia enterocolitica, Astrovirus, and Rotavirus A were established primarily with retrospective clinical specimens. |   |
|   | Performance characteristics for Entamoeba histolytica, and Vibrio (V. parahaemolyticus, V. vulnificus, and Vibrio cholerae) were established primarily using contrived clinical specimens. |   |
|   | Negative BIOFIRE GI Panel results in the setting of clinical illness compatible with gastroenteritis may be |   |

{7}

K242367 - Page 8 of 16
|   | due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn’s disease.

A gastrointestinal microorganism multiplex nucleic acid-based assay also aids in the detection and identification of acute gastroenteritis in the context of outbreaks. |   |
| --- | --- | --- |
|  Analyte | DNA/RNA | Same  |
|  Specimen Types | Human stool sample collected in Cary Blair transport media. | Same  |
|  Technological Principles | Nested multiplex PCR followed by high resolution melting analysis to confirm the identity of amplified product. | Same  |
|  Instrumentation | BIOFIRE 2.0 System or BIOFIRE Torch System | Same  |
|  Time to Result | About 1 hour | Same  |
|  Test Interpretation | Automated test interpretation and report generation. User cannot access raw data. | Same  |
|  Sample Preparation Method | Sample Processing is automated in the BIOFIRE System. | Same  |
|  Reagent Storage | Reagents are stored at room temperature. | Same  |
|  Shelf-Life | 12 months from Date of Manufacture | Same  |
|  Controls | Two controls are included in each reagent pouch to control for sample processing and | Same  |

{8}

K242367 - Page 9 of 16

|   | both stages of PCR and melt analysis. |   |
| --- | --- | --- |

## VI Standards/Guidance Documents Referenced:

### FDA-recognized Standards

- ISO 14971:2019 ‘Medical devices – Application of risk management to medical devices’
- ISO 15223-1:2021 ‘Medical devices – Symbols to be used with information to be supplied by the manufacturer – Part 1: General requirements’

## VII Performance Characteristics (if/when applicable):

### A Analytical Performance:

1. **Precision/Reproducibility:**

A reproducibility study was previously performed for the BIOFIRE FILMARRAY GI Panel performance in the original clearance (K140407) and in additional subsequent regulatory submissions (K143005, K160459, and K230404). Please refer to the published decision summaries for additional information. Since there is no change to the assay, instrument, or software, and a new reproducibility study was not conducted.

2. **Linearity:**

The BIOFIRE FILMARRAY GI Panel is a qualitative assay and therefore linearity studies are not applicable.

3. **Analytical Specificity/Interference:**

Analytical specificity and interfering substances studies were performed for the original BIOFIRE FILMARRAY GI Panel performance evaluation (K140407). Please refer to the published decision summary for additional information.

Additional analytical specificity testing was performed to evaluate the reactivity of the BIOFIRE FILMARRAY GI Panel Norovirus GI/GII assay (and other Panel assays) with multiple bacterial organisms identified in post-market investigations of false positive Norovirus GI/GII results in clinical specimens. Sequence data recovered from BIOFIRE FILMARRAY GI Panel pouch arrays suspected of false positive Norovirus GI/GII results were analyzed and aligned to sequences of a variety of commensal bacteria from the human gastrointestinal tract, including *Mediterraneibacter* (*Ruminococcus*) *gnavus*, *Enterobacter hormaechei*, *Anaerostipes hadrus*, *Parabacteroides merdae*, *Parabacteroides distasonis*, *Blautia wexlerae*, *Prevotella* sp. (*P. bivia*, *P. copri*, *P. histicola*, and *P. intermedia*) and *Bifidobacterium pseudocatenulatum*. Both wet-testing of isolates and *in silico* analysis approaches were used to evaluate cross reactivity of these organisms with the Norovirus GI/GII assay and other Panel assays.

{9}

# Isolate Wet Testing

Wet testing of organism isolates followed the general procedure used for Analytical Specificity testing in the original clearance for the BIOFIRE FILMARRAY GI Panel (K140407). Isolates were prepared and tested at concentrations &gt;1.0E+08 cells/mL in triplicate (one replicate using one of three different reagent lots). Initial positive results for any replicate were followed by at least two additional replicates to determine whether the result was reproducible. Norovirus GI/GII Detected results (Noro 1 assay positive) were reported in at least 1/3 or 2/5 replicates at the initial high concentration for *Mediterraneibacter* (*Ruminococcus*) *gnavus*, *Parabacteroides merdae*, *Anaerostipes hadrus* (one of two isolates), and *Enterobacter hormaechei* (one of two isolates). Norovirus GI/GII Not Detected results were reported for all replicates of *Parabacteroides distasonis*, *Bifidobacterium pseudocatenulatum* and *Blautia wexlerae*, all *Prevotella* sp., as well as one of two isolates of each *A. hadrus* and *E. hormaechei*.

## In silico Analysis

Amplicon sequences recovered from BIOFIRE FILMARRAY FILMARRAY GI Panel pouches with suspected false positive Norovirus GI/GII results were matched to a reference sequence for the best match species. The reference sequence was then used to search for regions flanking the amplicon which might be used as sites for the Norovirus GI/GII assay primer binding. Homology of assay primers with the predicted binding site was used to predict the likelihood of amplification and detection.

Testing and *in silico* analysis concluded that *Mediterraneibacter* (*Ruminococcus*) *gnavus* and *Parabacteroides merdae* are potentially cross-reactive species and present a risk of false positive Norovirus GI/GII results if present in a specimen at very high concentration. Additionally, predicted cross-reactive sequences were identified in a subset of genomes of *Anaerostipes hadrus* and *Enterobacter hormaechei* (&lt;35% and &lt;25% of genomes, respectively). Sequence in approximately 50% of the *Parabacteroides distasonis* genomes can be amplified by a combination of outer and inner reaction primers, but *in silico* analyses suggest a false positive result by the BIOFIRE FILMARRAY GI Panel is unlikely due to exclusion by the assay Tm range. Tm range-based exclusion was also demonstrated for off-target amplification of sequence in *Bifidobacterium pseudocatenulatum* mitigating the risk of false results due to cross-reactivity with this organism. No risk of cross-reactivity/off-target amplification could be identified for *Blautia wexlerae* based on sequence analysis or testing.

The analytical specificity study to support the original clearance of the BIOFIRE FILMARRAY GI Panel evaluated *Prevotella melaninogenica* (a human commensal and opportunistic pathogen). No cross-reactivity with *P. melaninogenica* was observed, however internal investigations into potentially false positive Norovirus GI/GII results recovered sequence from non-specific amplicons which, when submitted for BLAST analysis, aligned to atypical *Prevotella* sp. *In silico* analyses of four additional *Prevotella* sp. (*P. bivia*, *P. copri*, *P. histicola*, and *P. intermedia*) did not indicate non-specific amplification or interaction with the Norovirus GI/GII assay. These results are consistent with the lack of any observed cross-reactivity from wet-testing of these isolates. Results from the additional cross-reactivity testing and *in silico* analyses (Tables 1 &amp; 2) will be included in the updated device labeling.

K242367 - Page 10 of 16

{10}

Table 1. Observed or Predicted Cross-Reactivity with Off-Panel Organisms

|  BIOFIRE GI Panel Test Result | Cross-Reactive Organism(s)  |
| --- | --- |
|  Entamoeba histolytica | Entamoeba dispar  |
|  Giardia lamblia | Bifidobacterium spp^{a}
Ruminococcus spp^{a}  |
|  Enterotoxigenic E.coli (ETEC)
It/st | Citrobacter koseri
Citrobacter sedlakii
Hafnia alvei^{a}
Cedecea davisiae^{a}  |
|  Norovirus GI/GII^{b} | Prevotella spp. (sequences from unculturable/uncharacterized species)^{c}
Mediterraneibacter (Ruminococcus) gnavus
Parabacteroides spp. (P. merdae, P. acidifaciens^{d}, P. distasonis^{e})
Anaerostipes hadrus (select sequences)^{f}
Enterobacter hormaechei (select sequences)^{g}  |
|  Salmonella | E. coli with variant type III secretion protein^{h}  |
|  Vibrio
(V. parahaemolyticus/V. vulnificus/
V. cholerae) | Vibrio alginolyticus
Vibrio fluvialis^{i}
Vibrio mimicus^{i}
Grimontia (formerly Vibrio) hollisae  |
|  Yersinia enterocolitica | Yersinia frederiksenii^{a,j}
Yersinia kristensenii^{j}  |

a Cross-reactivity was not observed when tested at high concentration (1.5×10⁹ cells/mL). However, cross-reactivity was suspected or confirmed in clinical specimens and/or the potential for cross-reactivity is supported by in silico predictions.
b Cross-reactivity was identified by post-market investigation of suspected false positive Norovirus GI/GII results in clinical specimens. Cross-reactivity with the species listed was confirmed by analytical testing at high concentration (&gt;2.4×10⁸ cells/mL) and/or is supported by sequence analysis.
c Cross-reactive sequences are inconsistent with other Prevotella sequence data, suggesting non-specific interaction with atypical or uncharacterized species and/or sequences.
d P. acidifaciens was not tested but was determined by sequence analysis to have a similar risk of cross-reactivity as P. merdae.
e Norovirus GI/GII Not Detected was reported when P. distasonis was tested at high concentration (3.1×10⁹ cells/mL). However, non-specific amplification products with Tm values close to the assay specific Tm range have been observed and the potential for false positive Norovirus GI/GII test results exists.
f The risk of false positive Norovirus GI/GII results due to cross-reactivity with A. hadrus is associated with only a subset of A. hadrus RefSeq genome sequences (&lt;35% as of June 2024).
g The risk of false positive Norovirus GI/GII results due to cross-reactivity with E. hormaechei is associated with only a subset of E. hormaechei RefSeq genome sequences (&lt;25% as of June 2024).
h Cross-reactivity resulting in false positive Salmonella results has not been observed in analytical or clinical testing. However, non-specific amplification products with Tm values close to the assay specific Tm range have been observed and the potential for false positive Salmonella test results exists.
i Detected at concentrations near the Vibrio assay LoD.
j Y. kristensenii and Y. fredericksenii are difficult to distinguish from Y. enterocolitica by standard laboratory methods.

Table 2. Off-Panel Organisms Evaluated with Wet-testing or by in silico Analyses

|  Bacteria  |   |   |
| --- | --- | --- |
|  Tested  |   |   |
|  Abiotrophia defectivia | Clostridium novyi | Morganella morganii  |
|  Acinetobacter baumannii | Clostridium perfringens | Parabacteroides distasonis^{f}  |
|  Acinetobacter lwoffii | Clostridium ramosum | Parabacteroides merdae^{g}  |
|  Aeromonas hydrophila | Clostridium septicum | Peptoniphilus asaccharolyticus  |

K242367 - Page 11 of 16

{11}

|  Alcaligenes faecalis | Clostridium sordellii | Peptostreptococcus anaerobius  |
| --- | --- | --- |
|  Anaerococcus tetradius | Clostridium tetani | Photobacterium damselae  |
|  Anaerostipes hadrus a,d | Collinsella aerofaciens | Porphyromonas asaccharolytica  |
|  Arcobacter butzleri | Corynebacterium genitalium | Prevotella bivia k  |
|  Arcobacter cryaerophilus | Desulfovibrio piger | Prevotella copri h  |
|  Bacillus cereus | Diffusely adherent E.coli | Prevotella intermedia h  |
|  Bacteroides fragilis | Escherichia blattae | Prevotella histicola h  |
|  Bacteroides thetaiotaomicron | Escherichia fergusonii | Prevotella melaninogenica h  |
|  Bacteroides vulgatus | Escherichia hermannii | Proteus mirabilis  |
|  Bifidobacterium adolescentis b | Escherichia vulneris | Proteus penneri  |
|  Bifidobacterium bifidum b | Edwardsiella tarda | Proteus vulgaris  |
|  Bifidobacterium longum b | Egglerhella lenta | Providencia alcalifaciens  |
|  Bifidobacterium pseudocatenulatum | Enterobacter cloacae | Pseudomonas aeruginosa  |
|  Blautia (Ruminococcus) obeum | Enterobacter hormaechei d,e | Ruminococcus bromii b  |
|  Blautia wexlerae | Enterococcus faecalis | Ruminococcus flavefaciens b  |
|  Campylobacter concisus | Enterococcus faecium | Selenomonas ruminantium  |
|  Campylobacter curvus | Eubacterium cylindroides | Serratia liquefaciens  |
|  Campylobacter fetus | Eubacterium rectale | Serratia marcescens  |
|  Campylobacter gracilis | Faecalibacterium prausnitzii | Shewanella algae  |
|  Campylobacter helveticus | Fusobacterium varium | Staphylococcus aureus  |
|  Campylobacter hominis | Gardnerella vaginalis | Staphylococcus epidermidis  |
|  Campylobacter hyointestinalis | Gemella morbillorum | Stenotrophomonas maltophilia  |
|  Campylobacter lari | Grimontia (Vibrio) hollisae | Streptococcus agalactiae  |
|  Campylobacter mucosalis | Haemophilus influenzae | Streptococcus intermedius  |
|  Campylobacter rectus | Hafnia alvei c | Streptococcus pyogenes  |
|  Campylobacter showae | Helicobacter fennelliae | Streptococcus salivarius  |
|  Campylobacter sputorum | Helicobacter pylori | Trabulsiella guamensis  |
|  Campylobacter ureolyticus | Klebsiella (Enterobacter) aerogenes | Veillonella parvula  |
|  Cedecea davisae c | Klebsiella oxytoca | Vibrio alginolyticus  |
|  Chlamydia trachomatis | Klebsiella pneumoniae | Vibrio fluvialis  |
|  Citrobacter amalonaticus | Lactobacillus acidophilus | Vibrio mimicus  |
|  Citrobacter freundii | Lactobacillus reuteri | Yersinia bercovieri  |
|  Citrobacter koseri d | Lactococcus lactis | Yersinia frederiksenii i  |
|  Citrobacter sedlakii | Leminorella grimonii | Yersinia intermedia  |
|  Clostridium acetobutylicum | Listeria monocytogenes | Yersinia kristensenii  |
|  Clostridium botulinum | Mediterraneibacter (Ruminococcus) gnavus | Yersinia mollaretii  |
|  Clostridium difficile non-toxigenic d | Megamonas hypermegale | Yersinia pseudotuberculosis  |
|  Clostridium histolyticum | Megasphaera elsdenii | Yersinia rohdei  |
|  Clostridium methylpentosum | Methanobrevibacter smithii |   |
|  Protozoa/Parasites |   | Fungi  |
| --- | --- | --- |
|  Tested | In silico Analysis Only | Tested  |
|  Babesia microti | Ancylostoma duodenale | Aspergillus fumigatus  |
|  Blastocystis hominis | Ascaris lumbricoides | Candida albicans  |
|  Conidiobolus lachnodes | Balantidium coli | Candida catenulate  |

K242367 - Page 12 of 16

{12}

K242367 - Page 13 of 16

|  Conidiobolus lobatus | Chilomastix mesnili | Penicillium marneffei  |
| --- | --- | --- |
|  Encephalitozoon hellem | Dientamoeba fragilis | Saccharomyces boulardi  |
|  Encephalitozoon intestinalis | Endolimax nana | Saccharomyces cerevisiae  |
|  Entamoeba dispar | Entamoeba coli |   |
|  Entamoeba gingivalis | Entamoeba hartmanni |   |
|  Entamoeba moshkovskii | Entamoeba polecki |   |
|  Giardia muris | Enterobius vermicularis |   |
|  Pentatrichomonas hominis | Enteromonas hominis |   |
|  Schistosoma mansoni | Isospora belli |   |
|  Toxoplasma gondii | Necator americanus |   |
|  Trichomonas tenax |  |   |
<h2>Viruses</h2>
|  Tested |   | In silico Analysis Only  |
| --- | --- | --- |
|  Adenovirus A:31 | Coronavirus 229E | Adenovirus G52  |
|  Adenovirus B:34 | Coxsackievirus B3 | Norovirus GIV  |
|  Adenovirus C:2 | Cytomegalovirus (CMV) | Rotavirus B  |
|  Adenovirus D:37 | Echovirus 6 | Rotavirus C  |
|  Adenovirus E:4a | Enterovirus 68 |   |
|  Astrovirus variant VA1 | Hepatitis A |   |
|  Astrovirus variant MLB | Herpes Simplex Type 2 |   |
|  Bocavirus Type 1 | Rhinovirus 1A |   |

a Anerostipes hadrus isolates (DSM 23942 and ATCC 29173) were tested at &gt;2.4×10⁸ cells/mL. Norovirus GI/GII Detected results were only observed with DSM 23942. The ATCC 29173 isolate does not carry the cross-reactive sequence.
b Though not observed in analytical testing, cross-reactivity of the Giardia lamblia assay with one or more Bifidobacterium and Ruminococcus species was observed in the clinical evaluation (see Table 1).
c Though not observed in analytical testing, possible cross-reactivity of the ETEC 2 assay with Hafnia alvei and Cedecea davisiae was observed in the clinical evaluation or predicted by in silico analysis (see Table 1).
d Two isolates of this species were tested for analytical specificity.
e Enterobacter hormaechei isolates (ATCC BAA-2082 and ATCC 49162) were tested at &gt;5.0×10⁸ cells/mL. Norovirus GI/GII Detected results were only observed with ATCC 49162. The ATCC BAA-2082 sequence is not predicted to be cross-reactive by in silico analysis.
f Though not observed in analytical testing, cross-reactivity of the Norovirus GI/GII assay with a sequence identified in roughly half (~50%) of the P. distasonis genomes evaluated could occur at high concentration.
g A similar risk of cross-reactivity was identified with sequences annotated as Parabacteroides sp. and P. acidifaciens.
h No cross-reactivity with high concentrations of various Prevotella species (commensal and pathogenic) was observed in analytical testing, but the potential for weak cross-reactivity between the Norovirus GI/GII assay and unique variant sequences annotated as unculturable Prevotella sp. has been identified via investigation of discrepant results in clinical specimens.
i Though not observed in analytical testing, in silico analysis indicates that, similar to Y. kristensenii, cross-reactivity between the Yersinia enterocolitica assay and Yersinia fredericksenii is possible at high concentrations (see Table 1).

## 4. Assay Reportable Range:

The BIOFIRE FILMARRAY GI Panel is a qualitative assay thus the assay reportable range is not applicable.

## 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):

No changes were made to the assay, assay reagents, instrument, or assay software. For additional information, please refer to the original published decision summary of K143005.

## 6. Detection Limit:

{13}

A limit of detection (LoD) study was performed to support the original BIOFIRE FILMARRAY GI Panel (K140407). Since no changes were made to the assay, assay reagents, instrument, or assay software, no new LoD study was conducted.

## 7. Assay Cut-Off:

The BIOFIRE system evaluates DNA melting curve data rather than real-time PCR curves or a measurement value (e.g. Ct values or RFU) to determine positive and negative results; hence, the system does not utilize cut-offs as traditionally defined. For additional information regarding the assay cut-offs refer to the published decision summary for the BIOFIRE FILMARRAY GI Panel (K140407).

## B Comparison Studies:

1. Method Comparison with Predicate Device:

Not applicable.

2. Matrix Comparison:

Not applicable.

## C Clinical Studies:

1. Clinical Sensitivity:

Not applicable.

2. Clinical Specificity:

The BIOFIRE FILMARRAY GI Panel Norovirus GI/GII negative percent agreement (NPA) was originally established in a prospective clinical study performed in 2013 to support the original clearance of the device (K140407). These clinical study results for the Norovirus GI/GII assay are summarized in Table 3 below.

Table 3. BIOFIRE FILMARRAY GI Panel Norovirus GI/GII Clinical Performance Summary (2013)

|  Analyte | Positive Percent Agreement |   |   | Negative Percent Agreement  |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |  TP/(TP+FN) | % | 95% CI | TN/(TN+FP) | % | 95% CI  |
|  Norovirus GI/GII | 52/55 | 94.5 | 84.9-98.9% | 1483/1501 | 98.8 | 98.1-99.3%  |

To evaluate potential risk of false positive results with the Norovirus GI/GII target, an additional supplemental clinical study was conducted. The study was conducted with 872 fresh stool specimens collected and tested at three geographically distinct clinical sites within the U.S. Patient demographic information from this study is summarized in Table 4 below.

Table 4. Demographic Data for Prospectively Collected Specimens

K242367 - Page 14 of 16

{14}

|  Total Specimens | 872  |
| --- | --- |
|  Sex | Number of Specimens (%)  |
|  Male | 394 (45.2%)  |
|  Female | 478 (54.8%)  |
|  Age Group | Number of Specimens (%)  |
|  < 1 year | 70 (8.0%)  |
|  1-5 years | 120 (13.8%)  |
|  6-12 years | 64 (7.3%)  |
|  13-21 years | 127 (14.6%)  |
|  22-64 years | 283 (32.5%)  |
|  65+ years | 208 (23.9%)  |

Frozen aliquots of each specimen were tested at bioMerieux using the CDC CaliciNet Norovirus GI/GII assay, the same Norovirus comparator assay used in the 2013 clinical study. The use of frozen specimens was supported by a fresh vs. frozen equivalency study conducted in the original submission (K140407). Of note, the current CDC CaliciNet Norovirus GI/GII assay has been updated from the version used as the comparator in the 2013 clinical study. The updated assay includes new probe designs and quantification cycle cut-off thresholds. Specimens used for comparator testing were randomized and blinded to test operators. Testing with the BIOFIRE FILMARRAY GI Panel resulted in an apparent detection rate for the Norovirus GI/GII assay of  $7.2\%$  (63/872). A comparison of the BIOFIRE FILMARRAY GI Panel Norovirus GI/GII assay results with the comparator method are summarized in Table 5.

Table 5. BIOFIRE FILMARRAY GI Panel Norovirus GI/GII Performance

|   | Comparator |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|   |   |  Pos | Neg | GI Panel Performance | 95% CI  |
|  GI Panel | Pos | 34 | 29 | 34/35 (97.1%) | 85.1-99.9%  |
|   |  Neg | 1 | 808 | 808/837 (96.5%) | 95.1-97.7%  |
|  Total |   | 35 | 837 |   |   |

Samples with discrepant results were further analyzed by an independent molecular method and bidirectional sequencing. Investigation of the single false negative result from Table 3 showed evidence of Norovirus using the independent molecular method with bidirectional sequencing. Additional investigation of the 29 false positives observed in the study indicated evidence of Norovirus in three specimens. Sequences from various non-viral targets (including Anaerostipes hadrus, Escherichia coli, Eubacterium sp., Homo sapiens, Jutongia huaianensis, Klebsiella pneumoniae, Parabacteroides distasonis, Parabacteroides merdae, Prevotella sp., and Ruminococcus gnavus) was observed in 20 false positive specimens. Several of these organisms were evaluated for cross-reactivity with BIOFIRE FILMARRAY GI Panel by wet testing or in silico methods (see Tables 1 &amp; 2). The cause of the six remaining false positive specimens was suspected to be variable detection near or below the LoD of the comparator assay or unexpected primer dimer amplification. While the observed NPA results of in this supplemental clinical study indicated a minor decrease in the lower

K242367 - Page 15 of 16

{15}

bound of the 95% confidence interval when compared to the original performance evaluation of the device, the results are acceptable.

## D Clinical Cut-Off:

Not applicable.

## E Expected Values/Reference Range:

Expected Values for all analytes on the BIOFIRE FILMARRAY GI Panel were obtained in the 2013 prospective clinical study and are summarized in the published decision summary for K140407. Expected Values for the supplemental study presented in this submission have been stratified by study site and are summarized in Table 6 below. The prevalence of positive Norovirus GI/GII assay results in this study stratified by age group are presented in Table 7.

Table 6. Expected Values by Clinical Site

|  GI Panel Result | Overall |   | Site 1 |   | Site 2 |   | Site 3  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  #/SA^{a} | EV^{b} | #/SA | EV | #/SA | EV | #/SA | EV  |
|  Norovirus | 63/872 | 7.2% | 8/253 | 3.2% | 43/337 | 12.8% | 12/282 | 4.3%  |

a #/SA = number of GI Panel positives divided by the number of specimens analyzed (i.e., with valid comparator results) for that analyte
b EV = Expected Value

Table 7. Expected Values by Age Group

|  GI Panel Result | Overall |   | <1 year |   | 1-5 years |   | 6-12 years |   | 13-21 years |   | 22-64 years |   | 65+ years  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  #/SA^{a} | EV^{b} | #/SA | EV | #/SA | EV | #/SA | EV | #/SA | EV | #/SA | EV | #/SA | EV  |
|  Norovirus | 63/872 | 7.2% | 13/70 | 18.6% | 16/120 | 13.3% | 7/64 | 10.9% | 11/127 | 8.7% | 12/283 | 4.2% | 4/207 | 1.9%  |

a #/SA = number of GI Panel positives divided by the number of specimens analyzed (i.e., with valid comparator results) for that analyte
b EV = Expected Value

## VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

## IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K242367 - Page 16 of 16

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PCH/K242367](https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/PCH/K242367)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
