VERIGENE GRAM POSITIVE BLOOD CULTURE NUCLEIC ACID TEST (BC-GP)

DEN120014 · Nanosphere, Inc. · PAM · Jun 26, 2012 · Microbiology

Device Facts

Record IDDEN120014
Device NameVERIGENE GRAM POSITIVE BLOOD CULTURE NUCLEIC ACID TEST (BC-GP)
ApplicantNanosphere, Inc.
Product CodePAM · Microbiology
Decision DateJun 26, 2012
DecisionDENG
Submission TypePost-NSE
Regulation21 CFR 866.3365
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
DEN120014 · Jun 26, 2012VERIGENE GRAM POSITIVE BLOOD CULTURE NUCLEIC ACID TEST (BC-GP)Nanosphere, Inc.Routine clinical blood culture specimensThe device performance was evaluated by comparing results from the BC-GP test against conventional laboratory reference methods (culture, biochemical, and sequencing) using 1,251 fresh and 175 frozen clinical blood culture specimens collected from patients receiving routine care.Method comparison study; Routine clinical care; Blood culture specimens

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Method comparison study; Method comparison study using clinical and simulated specimensIndividuals receiving routine care requiring blood culture testing; Sample Size: 1767 total (1251 fresh, 175 frozen, 216 contrived); Number of Sites: 5Conventional biochemical, culture, and bidirectional sequencing reference methodsPercent agreement between BC-GP test and reference methods for identification of gram-positive bacteria and resistance markers

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Staphylococcus aureus identificationGold nanoparticle probe-based microarray detection99.1% (332/335) positive agreement; 100% (1091/1091) negative agreementClinical study (n=1767): 1251 fresh, 175 frozen, and 216 contrived blood culture specimens.
Staphylococcus epidermidis identificationGold nanoparticle probe-based microarray detection93.1% (296/318) positive agreement; 98.9% (1096/1108) negative agreementClinical study (n=1767): 1251 fresh, 175 frozen, and 216 contrived blood culture specimens.
Staphylococcus lugdunensis identificationGold nanoparticle probe-based microarray detection95.0% (19/20) positive agreement; 100% (1406/1406) negative agreementClinical study (n=1767): 1251 fresh, 175 frozen, and 216 contrived blood culture specimens.
Staphylococcus species identificationGold nanoparticle probe-based microarray detection98.0% (925/944) positive agreement; 99.4% (479/482) negative agreementClinical study (n=1767): 1251 fresh, 175 frozen, and 216 contrived blood culture specimens.
mecA resistance marker detectionGold nanoparticle probe-based microarray detection94.2% (375/398) positive agreement; 98.2% (1009/1028) negative agreementClinical study (n=1767): 1251 fresh, 175 frozen, and 216 contrived blood culture specimens.
Enterococcus faecalis identificationGold nanoparticle probe-based microarray detection96.9% (93/96) positive agreement; 99.9% (1329/1330) negative agreementClinical study (n=1767): 1251 fresh, 175 frozen, and 216 contrived blood culture specimens.
Enterococcus faecium identificationGold nanoparticle probe-based microarray detection97.1% (66/68) positive agreement; 100% (1358/1358) negative agreementClinical study (n=1767): 1251 fresh, 175 frozen, and 216 contrived blood culture specimens.
vanA resistance marker detectionGold nanoparticle probe-based microarray detection94.2% (65/69) positive agreement; 99.8% (1354/1357) negative agreementClinical study (n=1767): 1251 fresh, 175 frozen, and 216 contrived blood culture specimens.
vanB resistance marker detectionGold nanoparticle probe-based microarray detection100% (3/3) positive agreement; 100% (1423/1423) negative agreementClinical study (n=1767): 1251 fresh, 175 frozen, and 216 contrived blood culture specimens.
Streptococcus species identificationGold nanoparticle probe-based microarray detection93.6% (206/220) positive agreement; 99.6% (1201/1206) negative agreementClinical study (n=1767): 1251 fresh, 175 frozen, and 216 contrived blood culture specimens.
Streptococcus pneumoniae identificationGold nanoparticle probe-based microarray detection100% (38/38) positive agreement; 99.6% (1383/1388) negative agreementClinical study (n=1767): 1251 fresh, 175 frozen, and 216 contrived blood culture specimens.
Streptococcus agalactiae identificationGold nanoparticle probe-based microarray detection98.6% (70/71) positive agreement; 100% (1355/1355) negative agreementClinical study (n=1767): 1251 fresh, 175 frozen, and 216 contrived blood culture specimens.
Streptococcus pyogenes identificationGold nanoparticle probe-based microarray detection95.8% (23/24) positive agreement; 100% (1402/1402) negative agreementClinical study (n=1767): 1251 fresh, 175 frozen, and 216 contrived blood culture specimens.
Streptococcus anginosus group identificationGold nanoparticle probe-based microarray detection100% (12/12) positive agreement; 99.8% (1411/1414) negative agreementClinical study (n=1767): 1251 fresh, 175 frozen, and 216 contrived blood culture specimens.
Listeria species identificationGold nanoparticle probe-based microarray detection100% (3/3) positive agreement; 100% (1423/1423) negative agreementClinical study (n=1767): 1251 fresh, 175 frozen, and 216 contrived blood culture specimens.

Indications for Use

The Verigene® Gram Positive Blood Culture Nucleic Acid Test (BC-GP) performed using the sample-to-result Verigene System is a qualitative, multiplexed in vitro diagnostic test for the simultaneous detection and identification of potentially pathogenic gram-positive bacteria which may cause bloodstream infection (BSI). BC-GP is performed directly on positive blood culture using BACTEC™ Plus Aerobic/F and BacT/ALERT FA FAN® Aerobic blood culture bottles, which contain gram positive bacteria. BC-GP is indicated for use in conjunction with other clinical and laboratory findings, such as culture, to aid in the diagnosis of bacterial bloodstream infections; however, it is not used to monitor bloodstream infections. BC-GP detects and identifies the following bacterial genera and species: Staphylococcus spp. (S. aureus, S. epidermidis, S. lugdunensis), Streptococcus spp. (S. pneumoniae, S. pyogenes, S. agalactiae, S. anginosus group), Enterococcus (E. faecalis, E. faecium), and Listeria spp. In addition, BC-GP detects the mecA resistance marker, inferring mecA-mediated methicillin resistance, and the vanA and vanB resistance markers, inferring vanA/vanB-mediated vancomycin resistance. In mixed growth, BC-GP does not specifically attribute van-mediated vancomycin resistance to either E. faecium, or mecA-mediated methicillin resistance to either S. aureus or S. epidermidis. BC-GP is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial bloodstream infections; however, is not to be used to monitor these infections. Sub-culturing of positive blood cultures is necessary to recover organisms for susceptibility testing. identification of organisms not detected by BC-GP, differentiation of mixed growth, association of antimicrobial resistance marker genes to a specific organism, or for epidemiological typing. For prescription use only.

Device Story

The Verigene BC-GP is a sample-to-result, automated molecular diagnostic system used in clinical laboratories. It processes positive blood culture bottles to identify gram-positive bacteria and resistance markers (mecA, vanA, vanB). The system uses a disposable test cartridge for automated cell lysis and magnetic bead-based DNA isolation. Target detection occurs via a microarray format using Capture and Mediator oligonucleotides and gold nanoparticle probes; silver enhancement of these probes creates light-scattering aggregates. The Verigene Reader images these spots to determine the presence or absence of specific nucleic acid sequences. The system provides results in approximately 2.5 hours post-Gram stain. Clinicians use these results alongside culture and other findings to guide early, targeted antibiotic therapy for bloodstream infections, potentially improving patient survival by reducing time to effective treatment. It does not replace traditional culture/susceptibility testing.

Clinical Evidence

Clinical study (n=1767) compared BC-GP to conventional culture, biochemical, and sequencing methods. Included 1251 fresh, 175 frozen, and 216 contrived specimens. Performance showed high concordance: S. aureus (99.1% positive agreement), S. epidermidis (93.1%), Enterococcus faecalis (96.9%), and E. faecium (97.1%). Resistance marker agreement: mecA (94.2%), vanA (94.2%), vanB (100%). Initial no-call rate 4.7%, final 1.1%.

Technological Characteristics

Molecular assay using microarray-based detection of nucleic acid targets. Employs gold nanoparticle probe-based technology with silver enhancement. Automated sample preparation (cell lysis, magnetic bead DNA isolation). System components: Verigene Reader and Verigene Processor SP. Connectivity: Barcode scanner for sample tracking. No sterility requirements. Software: Verigene Reader 1.8.2b1, Processor SP 3.3b2.

Indications for Use

Indicated for patients with suspected bacterial bloodstream infections. Used on positive blood culture samples (BACTEC Plus Aerobic/F or BacT/ALERT FA FAN Aerobic) containing gram-positive bacteria to identify specific pathogens (Staphylococcus, Streptococcus, Enterococcus, Listeria spp.) and resistance markers (mecA, vanA, vanB). Not for monitoring infections.

Regulatory Classification

Identification

A multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures is a qualitative in vitro device intended to simultaneously detect and identify microorganism nucleic acids from blood cultures that test positive by Gram stain or other microbiological stains. The device detects specific nucleic acid sequences for microorganism identification as well as for antimicrobial resistance. This device aids in the diagnosis of bloodstream infections when used in conjunction with other clinical and laboratory findings. However, the device does not replace traditional methods for culture and susceptibility testing.

Special Controls

In combination with the general controls of the FD&C Act, the Verigene® Gram Positive Blood Culture Nucleic Acid Test is subject to the following special controls: The special controls for the BC-GP Assay are contained in the guideline document entitled "Class II Special Controls Guideline: Multiplex Nucleic Acid Assay for Identification of Microorganisms and Resistance Markers from Positive Blood Cultures."

*Classification.* Class II (special controls). The special control for this device is FDA's guideline document entitled “Class II Special Controls Guideline: Multiplex Nucleic Acid Assay for Identification of Microorganisms and Resistance Markers from Positive Blood Cultures.” For availability of the guideline document, see § 866.1(e).

Submission Summary (Full Text)

{0}------------------------------------------------ ## EVALUATION OF AUTOMATIC CLASS III DESIGNATION (DE NOVO) FOR VERIGENE® GRAM POSITIVE BLOOD CULTURE NUCLEIC ACID TEST (BC-GP) ### REGULATORY INFORMATION FDA identifies this generic type of device as: Multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures A multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures is a qualitative in vitro device intended to simultaneously detect and identify microorganism nucleic acids from blood cultures that test positive by Gram stain or other microbiological stains. The device detects specific nucleic acid sequences for microorganism identification as well as for antimicrobial resistance. This device aids in the diagnosis of bloodstream infections when used in conjunction with other clinical and laboratory findings. However, the device does not replace traditional methods for culture and susceptibility testing. ### NEW REGULATION NUMBER: 21 CFR 866.3365 CLASSIFICATION: II PRODUCT CODE: PAM DEVICE NAME: VERIGENE® GRAM POSITIVE BLOOD CULTURE NUCLEIC ACID TEST (BC-GP) ## BACKGROUND : DEVICE NAME: VERIGENE® GRAM POSITIVE BLOOD CULTURE NUCLEIC ACID TEST (BC-GP) 510(K): K113450 DATE OF 510(K) NSE DECISION: JUNE 12, 2012 DATE OF DE NOVO PETITION: JUNE 14, 2012 PETITIONER CONTACT: NANOSPHERE, INC. - MR. MARK DEL VECHIO PETITIONER'S RECOMMENDED CLASSIFICATION: II ### PETITIONER'S RECOMMENDED CONTROLS: - · Class II Special Controls Guidance Document: Multiplex Nucleic Acid Assay for Identification of Microorganism and Resistance Markers from Positive Blood Cultures - General controls {1}------------------------------------------------ ### INDICATIONS FOR USE The Verigene® Gram Positive Blood Culture Nucleic Acid Test (BC-GP) performed using the sample-to-result Verigene System is a qualitative, multiplexed in vitro diagnostic test for the simultaneous detection and identification of potentially pathogenic gram-positive bacteria which may cause bloodstream infection (BSI). BC-GP is performed directly on positive blood culture using BACTEC™ Plus Aerobic/F and BacT/ALERT FA FAN® Aerobic blood culture bottles, which contain gram positive bacteria. BC-GP is indicated for use in conjunction with other clinical and laboratory findings, such as culture, to aid in the diagnosis of bacterial bloodstream infections; however, it is not used to monitor bloodstream infections. BC-GP detects and identifies the following bacterial genera and species: | Staphylococcus spp. | Streptococcus spp. | Enterococcus | |-----------------------|--------------------------|---------------| | Staphylococcus aureus | Streptococcus | faecalis | | Staphylococcus | pneumoniae | Enterococcus | | epidermidis | Streptococcus pyogenes | faecium | | Staphylococcus | Streptococcus agalactiae | | | lugdunensis | Streptococcus anginosus | Listeria spp. | | | group | | In addition, BC-GP detects the mecA resistance marker, inferring mecA-mediated methicillin resistance, and the vanA and vanB resistance markers, inferring vanA/vanB-mediated vancomycin resistance. In mixed growth, BC-GP does not specifically attribute van-mediated vancomycin resistance to either E. faecium, or mecA-mediated methicillin resistance to either S. aureus or S. epidermidis. BC-GP is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial bloodstream infections; however, is not to be used to monitor these infections. Sub-culturing of positive blood cultures is necessary to recover organisms for susceptibility testing. identification of organisms not detected by BC-GP, differentiation of mixed growth, association of antimicrobial resistance marker genes to a specific organism, or for epidemiological typing. ### LIMITATIONS For prescription use only PLEASE REFER TO THE BC-GP LABELING FOR A MORE COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS. ### DEVICE DESCRIPTION The Verigene® Gram Positive Blood Culture Nucleic Acid Test (BC-GP) is a molecular assav which relies on detection of specific nucleic acid targets in a microarray format. For each of the {2}------------------------------------------------ bacterial nucleic acid sequences detected by the BC-GP test, Capture and Mediator oligonucleotides are utilized for gold nanoparticle probe-based endpoint detection. The Capture oligonucleotides bind to a specific portion of the nucleic acid target and are themselves bound onto a substrate in the microarray. The Mediator oligonucleotides bind to a different portion of the same nucleic acid target and allow binding of a gold nanoparticle probe to a portion complementary to a gold nanoparticle probe. Specific silver enhancement of the bound gold nanoparticle probes at the capture sites results in gold-silver aggregates that scatter light with high efficiency. The BC-GP test is performed on the Verigene® System, a 'sample-to-result', fully automated, bench-top molecular diagnostics workstation. The Verigene System consists of two components: the Verigene Reader and the Verigene Processor SP. The BC-GP test utilizes single-use disposable test consumables and a self-contained Verigene Test Cartridge for each sample tested. For the BC-GP test, the Verigene System allows automated nucleic acid extraction from Gram-positive bacteria-containing blood culture specimens and target detection of bacteria-specific DNA. The Reader is the Verigene System's user interface, which serves as the central control unit for all aspects of test processing and results generation. The Reader's graphical user interface guides the user through test processing and test results using a barcode scanner. The user inserts the Test Cartridge into the Verigene Processor SP, which executes the test procedure. automating the steps of (1) Sample Preparation - Cell lysis and magnetic bead-based bacterial DNA isolation from blood culture samples and (2) Verigene Hybridization Test - Detection and identification of bacterial-specific DNA in a microarray format by using gold nanoparticle probe-based technology. After test processing is complete, to obtain the test results the user removes the Test Cartridge from the Processor SP, removes the reagent pack from the substrate holder, and inserts the substrate holder into the Reader for analysis. Light scatter from the capture spots is imaged by the Reader and intensities from the microarray spots are used to make decisions regarding the presence (Detected) or absence (Not Detected) of a bacterial nucleic acid sequence/analyte. ## SUMMARY OF NONCLINICAL/BENCH STUDIES - a. Limit of Detection Studies: The limit of detection (LoD) of the BC-GP assay was assessed and confirmed by using bacterial strains with established titers. By definition, the LoD is the lowest target concentration that can be detected approximately 95% of the time. For each strain, the dilution series began with the sample obtained at 'bottle positivity'. The dilution series were prepared by using a diluent matrix that was comprised of blood culture broth containing charcoal, human blood, and a common commensal skin bacterium at a minimum concentration of ~10 CFU/mL. Each dilution was tested in replicates of four. The putative LoD was the lowest concentration level where all the replicates for the analyte were {3}------------------------------------------------ 'Detected'. Once the putative LOD was established, 20 replicate samples were tested to confirm the LoD. Thirteen (13) strains of S. aureus (SA) and S. epidermidis (SE), both methicillin sensitive and resistant, ranged from 1.9 x 10 to 5.7 x 10 CFU/mL for S. aureus and 2.0 x 106 to 7.5 x 10° CFU/mL for S. epidermidis. Twelve (12) strains of E. faecalis and E. faecium, both vancomycin sensitive and resistant, ranged from 1.1 x 10' to 5.7 x 10' CFU/mL for E. faecalis and 2.4 x 10' to 3.7 x 10' CFU/mL for E. faecium. Sixteen (16) strains of Streptococcus spp., Staphylococcus spp. (other than S. aureus and S. epidermidis), and Listeria spp. ranged from (CFU/mL): | Streptococcus spp. | 1.8 x 106 to 1.2 x 108 | |---------------------|------------------------| | S. agalactiae | 1.2 x 107 to 2.2 x 107 | | S. pyogenes | 9.5 x 106 to 6.3 x 107 | | S. anginosus group | 1.4 x 107 to 1.2 x 108 | | S. lugdunensis | 3.4 x 106 to 4.0 x 106 | | S. pneumoniae | 1.8 x 106 to 9.9 x 106 | | Staphylococcus spp. | 2.9 x 106 to 4.0 x 106 | | Listeria spp. | 7.5 x 106 to 1.2 x 107 | ## b. Inclusivity Studies: In addition to the strains tested in the LoD studies, testing was performed on additional strains of each targeted organism/resistance marker. Bottles were spiked with each organism in blood culture bottles containing resins and whole blood. The inoculated blood culture bottles were placed in the automated blood culture system and incubated until positive by audible alarm (bottle ring). Positive specimens were plated to determine purity and quantitation (CFU/ml). Concentrations of tested positive blood culture specimens ranged from ~10° to ~10° CFU/mL at bottle ring. Testing was performed in duplicate with the BC-GP test. The following describes the organisms tested in the inclusivity studies: SA and SE (both with and without the resistance marker mecA) - 98 methicillin-resistant S. aureus (MRSA) strains (including 65 representative NARSA strains), 18 methicillinsensitive Staphylococcus aureus (SA), eight (8) borderline oxacillin-resistant Staphylococcus aureus (BORSA) strains, six (6) methicillin-resistant Staphylococcus epidermidis (MRSE), and seven (7) methicillin-sensitive Staphylococcus epidermidis (SE). E. faecium and E. faecalis (both with and without resistance markers vanA and vanB) - 79 strains including 14 vancomycin-resistant Enterococcus faecalis (11 of which were classified as E. faecalis, vanA and 3 were classified as vanB). 17 vancomvcin-sensitive Enterococcus faecalis, 33 vancomycin-resistant Enterococcus faecium (26 of which were {4}------------------------------------------------ classified as vanA and 7 were classified as vanB), and 15 vancomycin-sensitive Enterococcus faecium. Streptococcus spp., Staphylococcus spp. and Listeria spp. - 233 bacterial strains, 184 of which belonged to the Streptococcus genus, including 34 Streptococcus agalactiae, 10 Streptococcus anginosus, 60 Streptococcus pneumoniae, and 16 Streptococcus pyogenes, 12 Listeria sp., and 45 Staphylococcus strains (of which four were Staphylococcus lugdunensis). One or more strains each of 13 different Staphylococcus species (other than SA, SE, and S. lugdenensis), 18 different Streptococcus species (other than S. pneumoniae, S. agalactiae, and S. pyogenes, and 6 different Listeria species were tested in the studies. For the Streptococcus anginosus group, 10 total strains covering the three species S. anginosus, S. intermedius, and S. constellatus were tested. All strains tested gave positive results with the BC-GP assay, with the exception of Listeria grayi and BORSA strains. Two strains of L. grayi were tested and both were not detected. Subsequent in silico sequence analysis verified that this species will likely not be detected by the BC-GP assay. Therefore, the following limitation was placed in the package insert: "For Listeria species detection, L. grayi should not be detected by the test, as suggested by in silico analysis and confirmed with analytical testing." Tested BORSA strains were characterized as mecA negative and expectantly identified by the BC-GP test as mecA negative. In order to support the genus level claims, a separate in silico analysis was performed by comparing the homology of the BC-GP assay probes against available GenBank entries for Streptococcus spp. Staphylococcus sp. and Listeria sp. The analysis, combined with wettesting results, suggested that the BC-GP assay should detect 41 different species of Staphylococcus, 54 species of Streptococcus, and six species of Listeria. # c. Analytical Specificity (Exclusivity) Analytical specificity was assessed using organisms phylogenetically related to panel organisms detected by the BC-GP test as well as those commonly present as contaminants in blood culture specimens. The exclusivity samples were divided into two distinct panels of organisms. The first panel consisted of 127 "non-BC-GP panel" organisms which would not have been expected to be detected by the BC-GP test. - . Fifty-two (79) gram-positive bacteria. including 15 Micrococcus strains, two each of seven Enterococcus strains and one of an eighth, a Kytococcus, a Kocuria, a Peptostreptococcus, two each of Planococcus, Rothia, Leuconostoc, Granulicatella, Lactococcus, and Pediococcus strains - Forty (40) gram-negative bacteria ● - One (1) acid-fast bacillus (attenuated Mycobacterium tuberculosis) ● - One (1) mollicute (Mycoplasma pneumonia) ● - Six (6) yeast strains The second panel consisted of over 450 "BC-GP panel" organisms, which in total comprised the analytical inclusivity study samples. The size and characteristics of the second exclusivity sample set is as follows: {5}------------------------------------------------ - . Staphylococcus strains (178), including MRSA (98), MSSA (18), BORSA (8), MRSE (6), MSSE (7), and 45 "non-SA/SE" Staphylococcus spp - Enterococcus strains (79). including E. faecium VS (15). E. faecium VRE (33). E. . faecalis VS (17), and E. faecalis VRE (14) - Streptococcus strains (184), including S. agalactiae (34), S. pneumoniae (60), S. ● anginosus (10), and S. pyogenes (16) - . Listeria strains (12) All testing was performed in BACTEC plus bottles. Positive blood cultures were checked for purity and quantitated with final concentrations determined to be between 10° to 10° CFU/ml. Testing was performed in duplicate for each strain and depending on the organism, bottle preparation was performed in one of the following ways: - Organisms were spiked into blood culture bottles containing resins and whole blood, ● and then incubated on the blood culture system until bottle ring, followed by testing with the BC-GP assay. - . High concentrations of organisms were spiked into blood culture bottles and immediately tested with the BC-GP assay. - . Previously-isolated genomic DNA was spiked into blood culture bottles for both Yersinia enterocolitica and Cryptococcus neoformans due to the risks associated with handling live cultures of these organisms. The BC-GP test demonstrated acceptable specificity when challenged with these two panels, with no cross-reactivity observed for any of the organisms and/or strains tested with the exception of two strains of Enterococcus avium, which were identified as E. faecium, and Lactococcus lactis, which was identified as Streptococcus species. This information was presented in the package insert. ## d. Interference The potential inhibitory effects of substances that may be encountered in blood and associated with the blood culturing process were tested with the BC-GP test at the concentration levels listed in the table below. Testing was performed using positive blood cultures containing representative strains of SA and SE (both with and without the resistance marker mecA), E. faecium (with and without vanA and vanB), Streptococcus spp., S. agalactiae, S. anginosus group, S. pneumoniae, S. pyogenes, Staphylococcus spp., S. lugdunensis and Listeria spp. A negative blood culture sample was also tested with each interferent to assess the impact of the interferent on the processing of the BC-GP. For each organism and for the negative samples, the potential interferents (in table below) were added directly into individual positive blood cultures (or negative) and tested in triplicate. All the replicates for all the strains tested (and {6}------------------------------------------------ the negative sample controls) gave the expected results indicating that these potential interferents do not interfere with the BC-GP test. | Interferent | Reference<br>Level | Concentrat<br>ion Tested | |-------------------------------------------|----------------------------------------------------------------------|--------------------------| | Hemoglobin | 1-2 g/L | 14 g/L | | Triglyceride<br>(Intralipid) | Normal <150<br>mg/dL High<br>300-500 mg/dL | 3000 mg/dL | | Conjugated | 0.1-0.4 mg/dL | 20 mg/dL | | Unconjugated | 0.1-0.8 mg/dL | 20 mg/dL | | γ-globulin | 0.7-1.7 g/dL | 6 g/dL | | Sodium<br>Polyanetholesul<br>fonate (SPS) | 0.02%-0.05%<br>w/v<br>(anticoagulant<br>in blood culture<br>bottles) | 0.25% w/v | Blood Specimen Potential Interferents In a separate interference study, representative strains of SA and SE (both with and without the resistance marker mecA), E. faecalis and E. faecium (with and without vanA and vanB), Streptococcus spp., S. agalactiae, S. anginosus group, S. pneumoniae, S. pyogenes, Staphylococcus spp., S. lugdunensis and Listeria spp were separately grown in BACTEC/F (Plus/Aerobic; Plus/Anaerobic; Standard/10 Aerobic; and PEDS Plus) and BacT/Alert (SA Aerobic, FA FAN Aerobic and PF Pediatric FAN) blood culture bottles. As expected, bottles containing targeted organisms gave positive results and negative blood culture bottles gave negative results, demonstrating that the bottle types tested did not have an impact on the performance of the BC-GP Test. Although this study demonstrated analytically that the additional bottle types do not interfere with the BC-GP Test, the majority of clinical study specimens were from BACTEC Plus Aerobic/F and BacT/ALERT FA FAN culture bottles. Therefore the Intended Use of the device will include only these two bottle types. #### Competitive Inhibition e. In order to assess competitive inhibition with the BC-GP test, representative bacterial strains targeted by the BC-GP assay were tested at their respective LoD levels when mixed with other targeted organisms at high concentration levels. The high concentration levels were obtained by incubating blood cultures for each strain at least eight hours beyond initial bottle positivity. The bottles were then spiked with the targeted organism at LoD. The mixed {7}------------------------------------------------ specimens were tested in triplicate to assess inhibition. The organism combinations tested are listed in the tables below. | Organisms | Low Titer<br>(CFU/mL) | High Titer<br>(CFU/mL) | |-----------------------|-----------------------|------------------------| | S.aureus | 4.6 x 106 | | | S. aureus (SA) | 5.7 x 106 | | | S. epidermidis (MRSE) | 7.5 x 106 | | | S. epidermidis (SE) | 6.9 x 106 | | | S. pneumoniae | | 2.8 x 108 | | S. pyogenes | | 5.9 x 108 | | S. agalactiae | | 1.3 x 109 | | S. anginosus | | 7.9 x 108 | | E. faecium (vanA) | | 1.1 x 109 | | E. faecium (vanB) | | 8.3 x 108 | Organisms Tested for Competitive Inhibition Study (SA and SE) Organisms Tested for Competitive Inhibition Study (Enterococcus) | Organisms | Low Titer<br>(CFU/mL) | High Titer<br>(CFU/mL) | |-------------------|-----------------------|------------------------| | E. faecalis | 3.3 x 107 | | | E. faecalis, vanA | 1.8 x 107 | | | E. faecalis, vanB | 1.1 x 107 | | | E. faecium | 7.4 x 106 | | | E. faecium, vanA | 9.3 x 106 | | | E. faecium, vanB | 6.8 x 106 | | | S. aureus / MRSA | | 8.9 x 107 | | S. epidermidis | | 2.6 x 108 | | S. agalactiae | | 1.3 x 108 | | S. anginosus | | 7.9 x 108 | | S. pneumoniae | | 2.8 x 108 | | L. monocytogenes | | 8.9 x 107 | ## Organisms Tested for Competitive Inhibition Study (Other Targeted Bacteria) | Organisms | Low Titer<br>(CFU/mL) | High Titer<br>(CFU/mL) | |---------------|-----------------------|------------------------| | S. anginosus | 1.4 x 107 | | | S. agalactiae | 1.2 x 107 | | | S. pneumoniae | 1.8 x 108 | | | S. pyogenes | 9.5 x 106 | | {8}------------------------------------------------ | S. lugdunensis | 3.4 x 106 | | |---------------------|-----------|-----------| | L. monocytogenes | 1.2 x 107 | | | S. aureus / MRSA | | 2.0 x 109 | | S. epidermidis/MRSE | | 5.5 x 108 | | S. aureus/MSSA | | 4.0 x 108 | | S. epidermidis/MSSE | | 7.4 x 108 | | E. faecalis (vanB) | | 1.3 x 108 | | E. faecium (vanA) | | 8.7 x 108 | | S. anginosus | | 1.2 x 109 | | S. agalactiae | | 8.6 x 108 | | S. pneumoniae | | 9.7 x 108 | | S. pyogenes | | 1.2 x 108 | | S. lugdenensis | | 2.0 x 109 | | L. monocytogenes | | 2.0 x 109 | All replicates of the mixed samples gave expected results for both bacterial targets with the exception of one of three replicates of a high titer of S. agalactiae mixed with a low titer of S. epidermidis. For this one replicate, S. epidermidis was not detected. An additional 20 replicates of this organism mixture were tested, resulting in correct detection of both organisms in 20/20 replicates. The original result was likely related to the SE being at LoD levels. In conclusion, results of the competitive inhibition study demonstrated no evidence of competitive inhibition for organisms with concentrations near the LoD mixed with organisms at high concentrations. #### f. Assay Cutoff In order to determine the cut-off values for the GC-GP assay, a set of bacterial strains were tested to represent all the analytes detected by the BC-GP Test. Blood culture bottles were spiked with each individual organism and grown in automated blood culture instruments to "bottle ring." A minimum of four replicates for each positive blood culture specimen were tested with the BC-GP Test. The test decisions from each test were compiled to generate a data set that was used to initially determine BC-GP Assay Test cutoff values. A set of retrospective blood culture specimens were then tested with the BC-GP Test to verify the previously determined cut off values. Results were analyzed using logistic fit and ROC statistics. The results obtained from this the cutoff determination matched the results from culture-based biochemical results, thus verifying the final cutoff values. #### Carry-Over and Cross Contamination Studies g. A study was performed to assess the potential for carryover/cross-contamination with the BC-GP by alternately running 'high positive' samples followed by 'true negative' samples. All of the high positive samples yielded the expected 'Detected' results for the intended {9}------------------------------------------------ bacteria and 'Not Detected' results for the other analytes. The true negative samples gave a 'Not Detected' call for all analytes. The studies confirmed that there was no evidence of carryover/cross-contamination from the high positive samples, or any other internal or external sources, in either of the steps of the BC-GP: Sample Extraction and Verigene Hybridization. ## h. Precision/Reproducibility: The precision and reproducibility of the BC-GP test was evaluated using a 40-member sample panel consisting of two concentrations of pure isolates of various strains of organisms grown in blood culture bottles a) until bottle positivity (Concentration 1 - approx. 10 -108 CFU/mL) and b) bottle positivity plus an additional 8 hours of incubation on the blood culture instrument (Concentration 2 - approx. 10°-10°CFU/mL). The sample panel was composed of a combination of 19 organisms and strains, including two strains of MRSA, and one strain each of MRSE, MSSA, MSSE, EFC, EFC/vanA, EFC/vanA, EFC/vanB, EFL/vanA, EFL/vanB, S. agalactiae, S. anginosus, S. mitis, S. warneri, S. pneumoniae, S. pyogenes, S. lugdunensis and Listeria monocytogenes. Two negative controls were also contained in the sample panel, one containing blood culture media only and one containing the organism Corynebacterium urealyticum. ## Precision Precision was established in-house at Nanosphere by testing the 40-member sample panel in duplicate twice daily by two operators for twelve (12) non-consecutive days, generating a total of forty-eight (48) replicates per sample. Precision was evaluated across multiple reagent lots, days, operators, runs, instruments and replicates. The percent and two-sided 95%CI for the panel members combined was: 100% (576/576; 99.4%-100%) for Staphylococcus aureus and S. epidermidis 99.8% (575/576; 99.0% -100%) for Enterococcus faecalis and E. faecium 99.9% (767/768; 99.4%-100%) for Streptococcus spp., S. agalactiae, S. anginosus group, S. pneumoniae, S. pyogenes, Staphylococcus spp., S. lugdunensis, and Listeria spp. Only two samples (E. faecalis/vanA and S. lugdenensis, both at a BP concentration of ~10 CFU/mL) yielded <100% agreement (97.9%, [47/48; 88.9-99.9]). Overall, there were 23 initial no-call results observed, all but one of which repeat tested successfully; the one final no-call result accounted for the discordant S. lugdenensis panel member result. ## Reproducibility Reproducibility was determined at three external sites by testing the same 40-member sample panel in duplicate twice daily by two (2) operators for five (5) non-consecutive days, {10}------------------------------------------------ which generated a total of sixty (60) replicates per sample. The percent agreement for the panel members for all sites combined was: 100% (720/720; 99.5-100) for Staphylococcus aureus and S. epidermidis, 100% (720/720; 99.5-100) for Enterococcus faecalis and E. faecium, 100% (960/960; 99.6% -100%) for Streptococcus spp., S. agalactiae, S. anginosus grow, S. pneumoniae, S. pyogenes, Staphylococcus spp., S. lugdunensis, and Listeria spp. Overall, there were 73 initial no-call results observed, all of which repeat tested successfully. #### i. Fresh versus Frozen Study The stability of specimens subjected to up to two freeze/thaw cycles was evaluated using blood culture specimens which were tested before and after freezing (at or below -70 ℃ to allow paired comparisons. Representative strains of SA and SE (both with and without the resistance marker mecA), E. faecalis and E. faecium (with and without vanA and vanB), Streptococcus spp.. S. agalactiae, S. anginosus group, S. pneumoniae, S. pyogenes, Staphylococcus spp., S. lugdunensis and Listeria spp. were tested. Organisms were grown in blood culture bottles and incubated on a continuous monitoring blood culture system. Cultured samples were tested at baseline at the following two time points: (i) bottle positivity and (ii) bottle positivity plus eight (8) hours. Four replicates of each samples were then tested after two freeze/thaw cycles. A comparison of matched pairs of specimens demonstrated 100% agreement between the fresh and frozen samples for each analyte tested for all conditions evaluated. In summary, the results of the fresh versus frozen study demonstrated that the use of the use of frozen specimens used in the clinical study was acceptable for testing with the BC-GP test. ## BIOCOMPATIBILITY/MATERIALS N/A ## SHELF LIFE/STERILITY The BC-GP Assay, Verigene Reader and the Verigene Processor SP have no sterility requirements. The BC-GP Test will be tested to achieve a minimum of 12 weeks stability. ### ANIMAL STUDIES N/A ELECTROMAGNETIC COMPATIBILITY AND ELECTRICAL SAFETY The Verigene Reader and the Verigene Processor SP meet all applicable international standards for electromagnetic compatibility and electrical safety. ## MAGNETIC RESONANCE (MR) COMPATIBILITY {11}------------------------------------------------ ### N/A MECHANICAL SAFETY N/A SOFTWARE | Version: Verigene Reader Software Version 1.8.2b1 and Verigene Processor SP Software<br>Version 3.3b2 | | | |-------------------------------------------------------------------------------------------------------|---|---| | Level of Concern: Moderate | | | | | Y | N | | Software description: | X | | | Device Hazard Analysis: | X | | | Software Requirements Specifications: | X | | | Architecture Design Chart: | X | | | Design Specifications: | X | | | Traceability Analysis/Matrix: | X | | | Development: | X | | | Verification & Validation Testing: | X | | | Revision level history: | X | | | Unresolved anomalies: | X | | The information provided in the submission is adequate for each of the items listed above. Additionally, documentation provided for the final verification for the Verigene Reader software version 1.8.2b1 and Verigene Processor SP software version 3.3b2 is adequate. ## SUMMARY OF CLINICAL INFORMATION A method comparison study (n=1767) was conducted at five external, geographically-diverse clinical study sites to evaluate the comparative performance of BC-GP to applicable conventional biochemical, culture, and bidirectional sequencing reference methods. The total evaluable specimens tested in the studies included; 1251 fresh, 175 frozen, and 216 contrived blood culture specimens. Eligible study subjects included individuals receiving routine care requiring blood culture testing. Blood culture specimens were collected from the patients and incubated on either the BACTEC or the BacT Alert continuous monitoring blood culture systems. Bottles that were flagged positive by the instrument were gram stained and then bottles confirmed to contain gram positive organisms were then tested with the BC-GP Test. Additionally, simulated specimens were utilized for very rare organisms such as S. lugdunensis, S. agalactiae, S. anginosus group, S. pyogenes, and Listeria species. These specimens were prepared by spiking blood culture bottles containing whole blood with bacterial suspensions of bacterial isolates. Prepared blood culture bottles were then grown to positivity on the continuous monitoring {12}------------------------------------------------ blood culture system, until flagged positive. Gram stains were performed to verify the presence of gram positive bacteria, and then testing was performed with the BC-GP Test. BC-GP Test results were compared with results from traditional laboratory reference methods (i.e. culture followed by testing blood culture isolates with conventional biochemicals. Vitek2, and cefoxitin disc testing. Cefoxitin discs were used as the reference method for confirming mecA mediated resistance in S. aureus and S. epidermidis. Vancomycin resistance and the presence of vanA or vanB in E. faecium and E. faecalis was performed using vancomycin E-tests followed by bidirectional sequencing on resistant organisms. The initial No-call rate in the clinical study was 4.7% (77/1642) , and the final No-call rate was 1.1% (18/1642). The total rate of initial No-Calls and Pre-analysis (pre-AE) Errors was 6.2% (101/1642). The final No-Call and Pre-AE rate was 1.2% (20/1642). The total specimens excluded from the BC-GP clinical study dataset (n=125) are listed by site and reason for exclusion in the following table: | Site | No. Excluded | Reason for Exclusion | | | | | | | | | |--------|--------------|----------------------|------------------------------------------------------------------------|-----------------------------------------------------------------|-----------------------------|-------------------------------------------------------|----------------------------|------------------------------------|-----------------------------------------|-----------------------------------------------------------------------| | | | Training<br>Sample | Culture/RM<br>results<br>Unavailable,<br>Unidentified, or<br>No Growth | Sample not<br>tested on BC-<br>GP or tested<br>after time limit | Inadequate<br>Sample Mixing | Incorrect Sample<br>Type or Patient<br>already tested | QC or<br>protocol<br>error | Gram stain<br>not gram<br>positive | Sample used<br>in Analytical<br>Studies | Dual organism<br>grown from<br>single stock<br>(reference<br>culture) | | MCW | 32 | | 1 | 6 | 1 | 6 | 2 | | 10 | 6 | | OSU | 30 | 5 | 4 | 1 | | 9 | 3 | | | 8 | | LIJ | 15 | 4 | 4 | 2 | | 1 | | 3 | | 1 | | MN | 19 | 3 | 1 | 4 | | 7 | 1 | | 1 | 2 | | NSU | 28 | 2 | 4 | | | 13 | | | | 9 | | Frozen | 1 | | | 1 | | | | | | | | Total | 125 | | | | | | | | | | {13}------------------------------------------------ Test performance is stratified by the manner in which the blood culture media specimen was obtained for clinical testing; i.e.; prospectively-collected fresh or stored frozen prior to testing, designated as the "Specimen Type" in the following tables. | Specimen<br>Type | n= | % Agreement (95% CI) | | Reference<br>Method | Specimen<br>Type | n= | % Agreement (95% CI) | | Reference<br>Method | | |----------------------------------------|---------------------------------|-----------------------------------|-----------------------------------|-------------------------------------------------|-----------------------------------|---------------------------|---------------------------------|----------------------------------|-------------------------------------------------|--| | <i>Staphylococcus aureus (SA)</i> | | Positive | Negative | Culture and Conventional Biochemical and Vitek2 | <i>Staphylococcus species</i> | | Positive | Negative | Culture and Conventional Biochemical and Vitek2 | | | | Fresh 1251 | 99.1%<br>322/325<br>(97.3-99.8) | 100%<br>926/926<br>(99.6-100) | | | Fresh 1251 | 97.9%<br>895/914<br>(96.8-98.7) | 99.4%<br>335/337<br>(97.9-99.9) | | | | | Frozen 175 | 100%<br>10/10<br>(69.2-100) | 100%<br>165/165<br>(97.8-100) | | | Frozen 175 | 100%<br>30/30<br>(88.4-100) | 99.3%<br>144/145<br>(96.2-99.9) | | | | | Total 1426 | 99.1%<br>332/335<br>(97.4-99.8) | 100%<br>1091/1091<br>(99.7-100) | | | Total 1426 | 98.0%<br>925/944<br>(96.9-98.8) | 99.4%<br>479/482<br>(98.2-99.9) | | | | | Simulated 216 | | 100%<br>216/216<br>(98.3-100) | | Simulated 216 | | 100%<br>25/25<br>(86.3-100) | 100%<br>191/191<br>(98.1-100) | | | | <i>Staphylococcus epidermidis (SE)</i> | | | | Culture and Conventional Biochemical and Vitek2 | <i>Staphylococcus lugdunensis</i> | | | | Culture and Conventional Biochemical and Vitek2 | | | | Fresh 1251 | 93.0%<br>294/316<br>(89.6-95.6) | 98.7%<br>923/935<br>(97.8-99.3) | | | Fresh 1251 | 87.5%<br>7/8<br>(47.4-99.7) | 100%<br>1243/1243<br>(99.7-100) | | | | | Frozen 175 | 100%<br>2/2<br>(15.8-100) | 100%<br>173/173<br>(97.9-100) | | | Frozen 175 | 100%<br>12/12<br>(73.5-100) | 100%<br>163/163<br>(97.8-100) | | | | | Total 1426 | 93.1%<br>296/318<br>(89.7-95.6) | 98.9%<br>1096/1108<br>(98.1-99.4) | | | Total 1426 | 95.0%<br>19/20<br>(75.1-99.9) | 100%<br>1406/1406<br>(99.7-99.9) | | | | | Simulated 216 | | 100%<br>2/2<br>(15.8-100) | 100%<br>214/214<br>(98.3-100) | | Simulated 216 | | 100%<br>20/20<br>(83.2-100) | 99.5%<br>195/196<br>(97.2-99.9) | | | <i>mecA</i> | | | | Culture and Cefoxitin Disk Diffusion | <i>Listeria species</i> | | | | Culture and Cefoxitin Disk Diffusion | | | Fresh 1251 | 94.1%<br>366/389<br>(91.2-96.2) | 97.8%<br>843/862<br>(96.6-98.7) | Fresh 1251 | | | 100%<br>3/3<br>(29.2-100) | 100%<br>1248/1248<br>(99.7-100) | | | | | Frozen 175 | 100%<br>9/9<br>(66.4-100) | 100%<br>166/166<br>(97.8-100) | Frozen 175 | | | | 100%<br>175/175<br>(97.9-100) | | | | | Total 1426 | 94.2%<br>375/398<br>(91.5-96.3) | 98.2%<br>1009/1028<br>(97.1-98.9) | Total 1426 | | | 100%<br>3/3<br>(29.2-100) | 100%<br>1423/1423<br>(99.7-100) | | | | | | Simulated 216 | | 100%<br>216/216<br>(98.3-100) | | Simulated 216 | | 100%<br>34/34<br>(89.7-100) | 100%<br>182/182<br>(98.0-100) | | | Summary of Clinical Test Performance versus Reference Method(s) – Staphylococcus and Listeria {14}------------------------------------------------ | Specimen<br>Type | n= | % Agreement (95% CI) | | Reference<br>Method | Specimen<br>Type | n= | % Agreement (95% CI) | | Reference<br>Method | |-----------------------------|------|-------------------------------|-----------------------------------|-------------------------------------------------|------------------|------|-------------------------------|-----------------------------------|--------------------------------------| | | | Positive | Negative | | | | Positive | Negative | | | Enterococcus faecalis (EFL) | | | | Culture and Conventional Biochemical and Vitek2 | vanA | | | | Culture and Bidirectional Sequencing | | Fresh | 1251 | 96.0%<br>72/75<br>(88.8-99.2) | 99.9%<br>1175/1176<br>(99.5-99.9) | | Fresh | 1251 | 91.9%<br>34/37<br>(78.1-98.3) | 100%<br>1214/1214<br>(99.7-100) | | | Frozen | 175 | 100%<br>21/21<br>(83.9-100) | 100%<br>154/154<br>(97.6-100) | | Frozen | 175 | 96.9%<br>31/32<br>(83.8-99.9) | 97.9%<br>140/143<br>(94.0-99.6) | | | Total | 1426 | 96.9%<br>93/96<br>(91.1-99.4) | 99.9%<br>1329/1330<br>(99.6-99.9) | | Total | 1426 | 94.2%<br>65/69<br>(85.8-98.4) | 99.8%<br>1354/1357<br>(99.4-99.9) | | | Simulated | 216 | 92.3%<br>12/13<br>(64.0-99.8) | 100%<br>203/203<br>(98.2-100) | | Simulated | 216 | 100%<br>15/15<br>(78.2-100) | 100%<br>201/201<br>(98.2-100) | | | Enterococcus faecium (EFC) | | | | Culture and Conventional Biochemical and Vitek2 | vanB | | | | Culture and Bidirectional Sequencing | | Fresh | 1251 | 94.4%<br>34/36<br>(81.3-99.3) | 100%<br>1215/1215<br>(99.7-100) | | Fresh | 1251 | 0/0 | 100%<br>1251/1251<br>(99.7-100) | | | Frozen | 175 | 100%<br>32/32<br>(89.1-100) | 100%<br>143/143<br>(97.5-100) | | Frozen | 175 | 100%<br>3/3<br>(29.2-100) | 100%<br>172/172<br>(97.9-100) | | | Total | 1426 | 97.1%<br>66/68<br>(89.8-99.6) | 100%<br>1358/1358<br>(99.7-100) | | Total | 1426 | 100%<br>3/3<br>(29.2-100) | 100%<br>1423/1423<br>(99.7-100) | | | Simulated | 216 | 100%<br>46/46<br>(92.3-100) | 99.4%<br>169/170<br>(96.8-99.9) | | Simulated | 216 | 97.3%<br>36/37<br>(85.8-99.9) | 100%<br>179/179<br>(98.0-100) | | Summary of Clinical Test Performance versus Reference Method(s) – Enterococcus ## Summary of Clinical Test Performance versus Reference Method(s) – Streptococcus | Specimen Type | n= | % Agreement (95% CI) | | Reference Method | Specimen Type | n= | % Agreement (95% CI) | | Reference Method | |---------------------------|------|---------------------------------|-----------------------------------|-------------------------------------------------|----------------------|------|-------------------------------|-----------------------------------|-------------------------------------------------| | <i>Streptococcus</i> spp. | 1251 | Positive | Negative | Culture and Conventional Biochemical and Vitek2 | <i>S. pneumoniae</i> | 1251 | Positive | Negative | Culture and Conventional Biochemical and Vitek2 | | Fresh | 1251 | 91.7%<br>143/156<br>(86.2-95.5) | 99.5%<br>1090/1095<br>(98.9-99.9) | | Fresh | 1251 | 100%<br>25/25<br>(86.3-100) | 99.6%<br>1221/1226<br>(99.1-99.9) | | | Frozen | 175 | 98.4%<br>63/64<br>(91.6-99.9) | 100%<br>111/111<br>(96.7-100) | | Frozen | 175 | 100%<br>13/13<br>(75.3-100) | 100%<br>162/162<br>(97.8-100) | | | Total | 1426 | 93.6%<br>206/220<br>(89.6-96.5) | 99.6%<br>1201/1206<br>(99.0-99.9) | | Total | 1426 | 100%<br>38/38<br>(90.8-100) | 99.6%<br>1383/1388<br>(99.2-99.9) | | | Simulated | 216 | 100%<br>92/92<br>(96.1-100) | 99.2%<br>123/124<br>(95.6-99.9) | Culture and Conventional Biochemical and Vitek2 | Simulated | 216 | 100%<br>8/8<br>(63.1-100) | 100%<br>208/208<br>(98.2-100) | Culture and Conventional Biochemical and Vitek2 | | <i>S. agalactiae</i> | 1251 | Positive | Negative | | <i>S. pyogenes</i> | 1251 | Positive | Negative | | | Fresh | 1251 | 97.5%<br>39/40<br>(86.8-99.9) | 100%<br>1211/1211<br>(99.7-100) | | Fresh | 1251 | 100%<br>10/10<br>(69.2-100) | 100%<br>1241/1241<br>(99.7-100) | | | Frozen | 175 | 100%<br>31/31<br>(88.8-100) | 100%<br>144/144<br>(97.5-100) | | Frozen | 175 | 92.9%<br>13/14<br>(66.1-99.8) | 100%<br>161/161<br>(97.7-100) | | | Total | 1426 | 98.6%<br>70/71<br>(92.4-99.9) | 100%<br>1355/1355<br>(99.7-100) | Culture and Conventional Biochemical and Vitek2 | Total | 1426 | 95.8%<br>23/24<br>(78.9-99.9) | 100%<br>1402/1402<br>(99.7-100) | Culture and Conventional Biochemical and Vitek2 | | Simulated | 216 | 100%<br>6/6<br>(54.1-100) | 100%<br>210/210<br>(98.3-100) | | Simulated | 216 | 98.2%<br>53/54<br>(90.1-99.9) | 100%<br>162/162<br>(97.8-100) | | {15}------------------------------------------------ | S. anginosus group | | | | | |--------------------|--------|-----------------------------|---------------------------------|-----------------------------------| | Prospective | Fresh | 1251 | 100%<br>9/9<br>(66.4-100) | 99.8%<br>1239/1242<br>(99.3-99.9) | | | Frozen | 175 | 100%<br>3/3<br>(29.2-100) | 100%<br>172/172<br>(97.9-100) | | | Total | 1426 | 100%<br>12/12<br>(73.5-100) | 99.8%<br>1411/1414<br>(99.4-99.9) | | Simulated | 216 | 100%<br>23/23<br>(85.2-100) | 99.5%<br>192/193<br>(97.2-99.9) | | The following tables provide the clinical performance of BC-GP for mecA detection with S. aureus and S. epidermidis (as determined by the reference method) and vanA/vanB detection with E. faecalis and E. faecium (as determined by the reference method). In the study protocol, Staphylococcus species (other than S. aureus and S. epidermidis) were not tested for methicillin resistance by cefoxitin disk analysis. However, 12 Staphylococcus species (as determined by the reference method) were positive for S. epidermidis and mecA by the BC-GP test and have been counted as false positives for mecA. These 12 specimens are accounted for in the total negative agreement calculation for the mecA target. | Organism | Resistance<br>Marker | n= | % Agreement (95% CI) with Reference Method | | |---------------------------------------------------------------------------------|---------------------------------|-----|--------------------------------------------|------------------------------------------------------------------------------------------------| | | | | Positive<br>[Cefoxitin (+)] | Negative<br>[Cefoxitin (-) or no cefoxitin<br>testing, as organism not isolated<br>on culture] | | S. aureus | <i>mecA</i> | 335 | 97.5%<br>157/161<br>(93.8-99.3) | 98.8%<br>172/174<br>(95.9-99.9) | | S. epidermidis<br>(including 12 specimens<br>identified as SE by BC-GP<br>only) | 94.2%<br>375/398<br>(91.5-96.3) | 330 | 92.0%<br>219/238<br>(87.8-95.1) | 81.5%<br>75/92<br>(72.1-88.9) | Detection of mecA with S. aureus and S. evidermidis, as Determined by Reference Method ### Detection of vanA/vanB with E. faecalis and E. faecium, as Determined by Reference Method | Organism | Resistance<br>Marker | n= | % Agreement (95% CI) with Reference Method | | |-------------|-------------------------------|-----|------------------------------------------------|------------------------------------------------| | | | | Positive<br>(Bi-directional Sequencing<br>(+)) | Negative<br>(Bi-directional Sequencing<br>(-)) | | E. faecalis | vanA | 109 | 85.7%<br>12/14<br>(57.2-98.2) | 100%<br>95/95<br>(96.2-100) | | E. faecium | 95.2%<br>80/84<br>(88.3-98.7) | 114 | 97.2%<br>69/71<br>(90.2-99.7) | 93.0%<br>40/43<br>(80.9-98.5) | | E. faecalis | vanB | 109 | 100%<br>7/7<br>(59.0-100) | 100%<br>102/102<br>(96.5-100) | | E. faecium | 97.5%<br>39/40<br>(86.8-99.9) | 114 | 97.0%<br>32/33<br>(84.2-99.9) | 100%<br>81/81<br>(95.6-100) | {16}------------------------------------------------ The following table contains additional genus/group-level specific BC-GP clinical and analytical performance data stratified by individual species within each genus (i.e.; Staphylococcus sp. (other than SA, SE, S. lugdunensis), Streptococcus sp. (other than S. agalactiae, S. pyogenes, S. pneumoniae) Listeria sp. and Streptococcus anginosus group. | Staphylococcus Genus | | | Streptococcus Genus | | | Streptococcus anginosus Group | | | |-----------------------------------|---------------------------------|-----------------------------|-----------------------------------------------|---------------------------------|-------------------------------|---------------------------------------------------------------|-----------------------------|-----------------------------| | Organism | % (+) Agreement (95% CI) | Analytical | Organism | % (+) Agreement (95% CI) | Analytical | Organism | % (+) Agreement (95% CI) | | | | Clinical | | | Clinical | | | Clinical | Analytical | | Combined<br>Staphylococcus spp. | 98.0%<br>950/969<br>(97.0-98.8) | 100%<br>37/37<br>(90.5-100) | Combined<br>Streptococcus spp. | 95.5%<br>298/312<br>(92.6-97.5) | 100%<br>183/183<br>(98.0-100) | Streptococcus.<br>anginosus group | 100%<br>35/35<br>(90.0-100) | 100%<br>10/10<br>(69.2-100) | | Staphylococcus<br>arlettae | 100%<br>1/1<br>(2.5-100) | NT* | Streptococcus<br>bovis | 100%<br>1/1<br>(2.5-100) | 100%<br>4/4<br>(39.8-100) | Streptococcus<br>anginosus | 100%<br>7/7<br>(59.0-100) | 100%<br>5/5<br>(47.8-100) | | Staphylococcus<br>auricularis | 100%<br>16/16<br>(79.4-100) | 100%<br>2/2<br>(15.8-100) | Streptococcus<br>dysgalactiae | 100%<br>5/5<br>(47.8-100) | 100%<br>2/2<br>(15.8-100) | Streptococcus<br>constellatus | 100%<br>9/9<br>(66.4-100) | 100%<br>3/3<br>(29.2-100) | | Staphylococcus<br>capitis | 96.4%<br>54/56<br>(87.7-99.6) | 100%<br>2/2<br>(15.8-100) | Streptococcus<br>dysgalactiae<br>equisimilis | 100%<br>2/2<br>(15.8-100) | 100%<br>3/3<br>(29.2-100) | Streptococcus<br>intermedius | 100%<br>8/8<br>(63.1-100) | 100%<br>2/2<br>(15.8-100) | | Staphylococcus<br>caprae | 100%<br>5/5<br>(47.8-100) | 100%<br>2/2<br>(15.8-100) | Streptococcus<br>equi | NT | 100%<br>2/2<br>(15.8-100) | Streptococcus.<br>anginosus group<br>(not further Identified) | 100%<br>11/11<br>(71.5-100) | NT | | Staphylococcus cohnii | NT | 100%<br>2/2<br>(15.8-100) | Streptococcus<br>equinus | NT | 100%<br>2/2<br>(15.8-100) | | | | | Staphylococcus<br>chromogenes | 100%<br>2/2<br>(15.8-100) | NT | Streptococcus<br>gallolyricus | NT | 100%<br>3/3<br>(29.2-100) | Listeria Genus | | | | Staphylococcus<br>haemolyticus | 91.4%<br>32/35<br>(76.9-98.2) | 100%<br>2/2<br>(15.8-100) | Streptococcus<br>gallolyticus<br>pasteurianus | NT | 100%<br>1/1<br>(2.5-100) | Combined<br>Listeria spp. | 100%<br>37/37<br>(90.5-100) | 100%<br>12/12<br>(73.5-100) | | Staphylococcus<br>hominis | 98.5%<br>133/135<br>(94.8-99.8) | 100%<br>3/3<br>(29.2-100) | Streptococcus<br>gordonii | 100%<br>1/1<br>(2.5-100) | 100%<br>2/2<br>(15.8-100) | Listeria grayi | NT | 0%<br>0/2<br>(15.8-100) | | Staphylococcus<br>intermedius | 100%<br>3/3<br>(29.2-100) | 100%<br>1/1<br>(2.5-100) | Streptococcus<br>infantarius<br>infantarius | NT | 100%<br>1/1<br>(2.5-100) | Listeria innocua | 100%<br>8/8<br>(63.1-100) | 100%<br>1/1<br>(2.5-100) | | Staphylococcus<br>muscae | NT | 100%<br>1/1<br>(2.5-100) | Streptococcus<br>infantarius subsp.<br>coli | NT | 100%<br>1/1<br>(2.5-100) | Listeria ivanovii | 100%<br>2/2<br>(15.8-100) | 100%<br>1/1<br>(2.5-100) | | Staphylococcus<br>pasteuri | NT | 100%<br>2/2<br>(15.8-100) | Streptococcus<br>infantis | NT | 100%<br>2/2<br>(15.8-100) | Listeria<br>monocytogenes | 100%<br>24/24<br>(85.8-100) | 100%<br>6/6<br>(54.1-100) | | Staphylococcus<br>saccharolyticus | NT | 100%<br>4/4<br>(39.8-100) | S. mitis, S. oralis,<br>or S. mitis/oralis | 95.2%<br>20/22<br>(70.8-98.9) | 100%<br>19/19<br>(82.4-100) | Listeria seeligeri | 100%<br>1/1<br>(2.5-100) | 100%<br>1/1<br>(2.5-100) | | Staphylococcus<br>saprophyticus | 100%<br>4/4<br>(39.8-100) | 100%<br>2/2<br>(15.8-100) | Streptococcus<br>mutans | 50%<br>1/2<br>(1.3-98.7) | 100%<br>2/2<br>(15.8-100) | Listeria welshimeri | 100%<br>1/1<br>(2.5-100) | 100%<br>1/1<br>(2.5-100) | | Staphylococcus<br>schleiferi | 100%<br>1/1<br>(2.5-100) | 100%<br>2/2<br>(15.8-100) | Streptococcus<br>parasanguinis | 100%<br>4/4<br>(39.8-100) | 100%<br>4/4<br>(39.8-100) | Listeria spp.<br>unidentified | 100%<br>1/1<br>(2.5-100) | NT | | Staphylococcus sciuri | NT | 100%<br>2/2<br>(15.8-100) | Streptococcus<br>peroris | NT | 100%<br>1/1<br>(2.5-100) | | | | | Staphylococcus<br>simulans | 100%<br>5/5<br>(47.8-100) | 100%<br>2/2<br>(15.8-100) | Streptococcus<br>salivarius | 50%<br>4/8<br>(15.7-84.3) | 100%<br>5/5<br>(47.8-100) | | | | Summary of Genus/Group-level Test Performance versus Reference Method(s) – Stratified by Species {17}------------------------------------------------ | Staphylococcus<br>warneri | 90.0%<br>18/20<br>(68.3-98.8) | 100%<br>2/2<br>(15.8-100) | Streptococcus<br>sanguinis | 75%<br>3/4<br>(19.4-99.4) | 100%<br>3/3<br>(29.2-100) | |----------------------------------|-------------------------------|---------------------------|-------------------------------|-------------------------------|---------------------------| | Staphylococcus<br>xylosus | 100%<br>2/2<br>(15.8-100) | 100%<br>2/2<br>(15.8-100) | Streptococcus<br>thoraltensis | NT | 100%<br>1/1<br>(2.5-100) | | Staphylococcus<br>species (CONS) | 100%<br>29/29<br>(88.1-100) | NT | Streptococcus<br>viridans | 83.3%<br>20/24<br>(62.6-95.3) | NT | | | | | Streptococcus<br>uberis | 100%<br>1/1<br>(2.5-100) | NT | *NT = "Not Tested Mixed Culture Results: In total, there were 98 mixed specimens that were detected either by BC-GP, the reference culture methods or both. The two tables below list the distinct mixed specimen combinations detected by BC-GP in the clinical study and the additional distinct mixed specimen combinations detected by the reference/comparator methods, but not detected by BC-GP. In the clinical study, there were a total of six discrepant mixed specimens for which the BC-GP test detected organisms that were not detected by the reference culture (false positives). Additionally, there were 25 discrepant mixed specimens for which the reference culture detected organisms that were not detected by the BC-GP Test (false negatives). The majority (22/25) of the organisms not detected by the BC-GP Test in the false negative mixed specimens were organisms that can sometimes be considered as potential blood culture contaminants (Staphylococcus sp. (not SA), S. viridans, and S. salivarius.) For any mixed blood culture, there may be different rates of growth for each organism present, and therefore the organism present in lower concentrations (below "bottle ring") may not be detected. The following limitation was placed in the package insert regarding detection of organism by the BC-GP Test in mixed blood cultures: - . In mixed cultures containing Gram positive bacteria and other organisms, BC-GP may not identify all the detectable organisms in the specimen, depending upon the concentration of each target present. {18}------------------------------------------------ | Analyte 1 | Analyte 2 | Ob<br>ser<br>ved<br>Fre<br>que<br>ncy | No. of<br>Specime<br>ns with<br>Discrepa<br>nt Co-<br>analytes | Discrepant Analyte(s)1 | |--------------------------------------------------------|--------------------------------------------------------|---------------------------------------|----------------------------------------------------------------|------------------------| | E. faecalis and vanA | E. faecium | 1 | 0 | | | E. faecalis | Staphylococcus species | 3 | 0 | | | E. faecalis | Streptococcus species | 1 | 1 | Streptococcus species | | E. faecium and vanA | Staphylococcus species | 1 | 0 | | | Listeria | S. anginosus group | 1 | 1 | S. anginosus group | | S. lugdunensis | Listeria species | 1 | 0 | | | Staphylococcus species | E. faecalis | 2 | 1 | Staphylococcus species | | Staphylococcus species and S.<br>epidermidis and mecA+ | E. faecalis | 1 | 0 | | | Staphylococcus species and S.<br>aureus | Staphylococcus species and S. epidermidis and<br>mecA+ | 1 | 0 | | | Streptococcus species and<br>S. agalactiae | Staphylococcus species and<br>S. aureus | 1 | 0 | | | Streptococcus species and<br>S. pneumoniae | Streptococcus species and<br>S. anginosus group | 1 | 1 | S. anginosus group | | Streptococcus species and S.<br>pyogenes | E. faecalis | 1 | 1 | E. faecalis | | Streptococcus species | Staphylococcus species | 2 | 1 | Staphylococcus species | | | TOTAL | 17 | 6 | | Clinical Mixed Specimen Combinations Detected by BC-GP, but not by Reference Methods 1 Defined as an analyte that was detected by BC-GP, but not detected by the reference methods. # Additional Clinical Mixed Specimens Detected by Reference Methods, but not Detected by BC-GP | Analyte 1 | Analyte 2 | Analyt<br>e 3 | Observed<br>Frequency | No. Specimens<br>w/ Discrepant<br>Co-analytes | Discrepant<br>Analyte(s)¹ | |----------------------------------|-------------------------------------|-------------------|-----------------------|-----------------------------------------------|---------------------------| | E. faecalis (vanS) | S. haemolyticus | | 2 | 2 | S. haemolyticus | | E. faecalis (vanS) | S. aureus (methR) | | 1 | 1 | S. aureus<br>(methR) | | E. faecalis (vanS) | GNR | | 2 | 0 | - | | E. faecalis (vanS) | S. epidermidis (methR) | | 1 | 1 | S. epidermidis<br>(methR) | | E. faecalis (vanS) | yeast | S.<br>warneri | 1 | 1 | S. warneri | | E. faecalis (vanS) | Bacillus species (not<br>anthracis) | | 1 | 0 | - | | E. faecalis (vanR) | S. epidermidis (methR) | | 2 | 2 | S. epidermidis<br>(methR) | | E. faecalis (vanR) | S. agalactiae | | 1 | 1 | (vanR) | | E. faecium (vanS) | S. aureus (methR) | | 1 | 1 | S. aureus<br>(methR) | | E. faecium | GNR | | 1 | 0 | - | | E. faecium (vanR) | S. warneri | | 1 | 1 | S. warneri | | E. faecium (vanR) | S. capitis | | 1 | 1 | S. capitis | | E. faecium (vanR) | S. epidermidis (methR) | | 1 | 1 | S. epidermidis<br>(methR) | | E. faecium (vanR) | GNR | | 1 | 0 | - | | E. faecium (vanR) | Corynebacterium | | 1 | 0 | - | | M. lylae | M. luteus | | 1 | 0 | - | | Streptococcus viridans<br>group | GNR | | 1 | 1 | Strep viridans<br>group | | K. rosea | K. kristinae | | 1 | 0 | - | | E. casseliflavus | GNR | | 1 | 0 | - | | Leuconostoc<br>mesenteroides | S. salivarius | | 1 | 1 | S. salivarius | | S. hominis | S. warneri | | 1 | 0 | - | | S. capitis | S. epidermidis (methS) | | 2 | 1 | S. epidermidis<br>(methS) | | Analyte 1 | Analyte 2 | Analyt e 3 | Observed<br>Frequency | No. Specimens<br>w/ Discrepant<br>Co-analytes | Discrepant<br>Analyte(s)¹ | | S. haemolyticus | K. kristinae | | 1 | 0 | - | | S. hominis | S. salivarius | | 1 | 1 | Streptococcus<br>spp. | | S. simulans | S. hominis | | 1 | 0 | - | | S. epidermidis (methR) | S. hominis | S.<br>auricularis | 1 | 1 | S. epidermidis<br>(methR) | | S. simulans | S. warneri | | 1 | 0 | - | | S. epidermidis (methS) | S. hominis | | 3 | 2 | S. epidermidis<br>(methS) | | S. hominis | Staphylococcus species<br>(CoNS) | | 2 | 0 | - | | S. hominis | S. schleiferi | | 1 | 0 | - | | S. haemolyticus | S. capitis | | 1 | 0 | - | | S. haemolyticus | S. hominis | | 1 | 0 | - | | S. auricularis | GNR | | 1 | 0 | - | | S. hominis | S. epidermidis (methR) | | 6 | 2 | S. epidermidis<br>(methR) | | S. hominis | S. capitis | | 2 | 0 | - | | S. viridans group | S. hominis | | 1 | 1 | Streptococcus<br>species | | S. epidermidis (methS) | S. warneri | | 1 | 0 | - | | S. epidermidis | Staphylococcus species<br>(CoNS) | | 1 | 0 | - | | Corynebacterium | S. epidermidis (methR) | | 1 | 0 | - | | S. capitis | S. epidermidis (methR) | | 4 | 0 | - | | S. epidermidis (methR) | S. hominis | | 5 | 0 | - | | S. hominis…
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...