CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel: Influenza B Lineage Genotyping Kit,
K181736 · Centers For Disease Control and Prevention (CDC) · OZE · Jul 30, 2018 · Microbiology
Device Facts
Record ID
K181736
Device Name
CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel: Influenza B Lineage Genotyping Kit,
Applicant
Centers For Disease Control and Prevention (CDC)
Product Code
OZE · Microbiology
Decision Date
Jul 30, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 866.3980
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K181736 · Jul 30, 2018
CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel: Influenza B Lineage Genotyping Kit,
Centers For Disease Control and Prevention (CDC)
Retrospective clinical respiratory specimens collected during influenza seasons (2011-2012, 2017, and 2017-2018)
Retrospective clinical samples were used to evaluate the performance (positive and negative agreement) of the modified Influenza B Lineage Genotyping Kit (VER 1.1 and VER 2) compared to the cleared predicate device.
Retrospective clinical study; Study Period: 2011-2012 influenza season
Patients with upper respiratory tract specimens collected during the 2011-2012 influenza season; Sample Size: 80 (30 positive, 50 negative)
Cleared InfB assay (BHQ quenched)
Positive and negative percent agreement
Retrospective clinical study; Study Period: 2017-2018 influenza seasons
Patients with upper respiratory tract clinical samples collected during the 2017 and 2017-2018 influenza seasons; Sample Size: 187 (126 positive, 61 negative)
FDA-cleared Influenza B Lineage Genotyping Kit
Positive and negative percent agreement
Indications for Use
The Influenza B Lineage Genotyping Kit contains reagents and controls of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel and is intended for use in real-time RT-PCR (rRT-PCR) assays on an Applied Biosystems (ABI) 7500 Fast Dx Real-Time PCR instrument in conjunction with clinical and epidemiological information: . For the determination of the genetic lineage of human influenza B viruses as B/Victoria or B/Yamagata lineage from viral RNA in upper respiratory tract clinical specimens (including nasopharyngeal swabs [NPS], nasal swabs [NS], throat swabs [TS], nasal aspirates [NA], nasal washes [NW] and dual nasopharyngeal/throat swabs [NPS/TS]) from human patients with signs and symptoms of respiratory infection and/or from viral culture; . To provide epidemiologic information for surveillance of circulating influenza viruses. Performance characteristics for influenza B lineage genotyping were established during a season when influenza B/Victoria and B/Yamagata lineages were in circulation. Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease. All users, analysts, and any person reporting results from use of this device should be trained to perform and interpret the results from this procedure by a competent instructor prior to use. CDC Influenza Division will limit the distribution of this device to only those users who have successfully completed a training course provided by CDC instructors or designees.
Device Story
The CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel: Influenza B Lineage Genotyping Kit is an in vitro diagnostic tool for identifying influenza B lineages. It processes viral RNA extracted from upper respiratory tract specimens. The device uses real-time RT-PCR (rRT-PCR) on the Applied Biosystems 7500 Fast Dx instrument. It utilizes oligonucleotide primers and fluorescently labeled hydrolysis probes targeting conserved regions of the non-structural (NS) and hemagglutinin (HA) genes. The kit includes reagents for detecting influenza B, B/Victoria, B/Yamagata, and human RNase P (control). It is operated by trained laboratory personnel in clinical or public health settings. The output is a qualitative genetic lineage determination, which provides epidemiologic surveillance data. Results assist healthcare providers in understanding circulating influenza strains, though they are not the sole basis for patient management decisions.
Clinical Evidence
Clinical performance was evaluated via retrospective and prospective studies. A retrospective study (n=30 positive, 50 negative) comparing ZEN-quenched probes to BHQ-quenched probes showed 100% positive and negative agreement. A prospective study (n=579) during the 2016-2017 season demonstrated 100% sensitivity and 100% specificity for B/Victoria, and 100% sensitivity and 99.8% specificity for B/Yamagata. Additional retrospective studies (n=126 positive, 61 negative) confirmed the performance of VER 1.1 and VER 2 designs, showing 100% positive and negative agreement across all tested specimen types.
Technological Characteristics
Real-time RT-PCR assay using oligonucleotide primers and fluorescently labeled hydrolysis probes. Employs ZEN or BHQ-1 quenchers for InfB assays and BHQ Plus or BHQ-1 for lineage assays. Requires Applied Biosystems 7500 Fast Dx instrument. Compatible with various commercial RNA extraction kits (e.g., QIAamp DSP, MagNA Pure). Enzyme master mixes include Invitrogen SuperScript III Platinum or Quanta qScript One-Step. Software: SDS version 1.4.
Indications for Use
Indicated for determination of genetic lineage (B/Victoria or B/Yamagata) of human influenza B viruses in symptomatic patients with respiratory infection. Specimens include nasopharyngeal swabs, nasal swabs, throat swabs, nasal aspirates, nasal washes, and dual nasopharyngeal/throat swabs. For use in conjunction with clinical/epidemiological information and viral culture.
Regulatory Classification
Identification
A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B; (2) Influenza A subtype H1 and Influenza A subtype H3; (3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B; (4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus; (5) Human Metapneumovirus; (6) Rhinovirus; and (7) Adenovirus.
Special Controls
*Classification.* Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.
Predicate Devices
CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel: Influenza B Lineage Genotyping Kit (K172091)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 30, 2018
Centers for Disease Control and Prevention (CDC) Yon Yu, Pharm.D. Associate Director for Regulatory Affairs 1600 Clifton Road Ms E-51 Atlanta, GA 30329-4027
Re: K181736
Trade/Device Name: CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel: Influenza B Lineage Genotyping Kit Regulation Number: 21 CFR 866.3980 Regulation Name: Respiratory viral panel multiplex nucleic acid assay Regulatory Class: Class II Product Code: OZE, NSU, OOI Dated: June 29, 2018 Received: July 2, 2018
Dear Yon Yu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR
{1}------------------------------------------------
Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Steven R. Gitterman -S
for
Uwe Scherf, Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K181736
#### Device Name
CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel: Influenza B Lineage Genotyping Kit
Indications for Use (Describe)
The Influenza B Lineage Genotyping Kit contains reagents and controls of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel and is intended for use in real-time RT-PCR (rRT-PCR) assays on an Applied Biosystems (ABI) 7500 Fast Dx Real-Time PCR instrument in conjunction with clinical and epidemiological information:
For the determination of the genetic lineage of human influenza B viruses as B/Victoria or B/Yamagata lineage from · viral RNA in upper respiratory tract clinical specimens (including nasopharyngeal swabs [NS], throat swabs [TS], nasal aspirates [NA], nasal washes [NW] and dual nasopharyngeal/throat swabs [NPS/TS]) from human patients with signs and symptoms of respiratory infection and/or from viral culture;
· To provide epidemiologic information for surveillance of circulating influenza viruses.
Performance characteristics for influenza B lineage genotyping were established during a season when influenza B/ Victoria and B/Yamagata lineages were in circulation.
Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule out bacterial infection with other viruses. The agent detected may not be the definite cause of disease.
All users, and any person reporting results from use of this device should be trained to perform and interpret the results from this procedure by a competent instructor prior to use. CDC Influenza Division will limit the distribution of this device to only those users who have successfully completed a training course provided by CDC instructors or designees.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 8. 510(k) Summary
#### I. GENERAL INFORMATION
Submitter: Centers for Disease Control and Prevention 1600 Clifton Road, NE Atlanta, GA 30329
Contact Person: CDR Yon Yu, Pharm. D. Associate Director for Regulatory Affairs Office of the Director National Center for Emerging and Zoonotic Infectious Diseases Centers for Disease Control and Prevention 1600 Clifton RD, MS E-51; Atlanta, GA 30329-4027 Phone: 404-639-3046 Fax: 404-498-1106 Email: fkb8@cdc.gov
# Date Prepared: June 28, 2018
#### DEVICE INFORMATION II.
| Proprietary Name: | CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel<br>Influenza B Lineage Genotyping Kit (VER 1.1) and Influenza B<br>Lineage Genotyping Kit (VER 2) |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Influenza B Lineage Genotyping Kit |
| Regulation Section: | 866.3980-Respiratory viral panel multiplex nucleic acid assay |
| Subsequent Regulation<br>Sections: | 862.2570-Instrumentation for clinical multiplex systems |
| Device Classification: | Class II |
| Product Code: | OZE |
| Subsequent Product Codes: | NSU, OOI |
| Panel: | Microbiology |
{4}------------------------------------------------
#### III. PREDICATE DEVICE
CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel: Influenza B Lineage Genotyping Kit (K172091)
#### IV. DEVICE DESCRIPTION
The CDC Human Influenza Real-Time RT-PCR Diagnostic Panel is used in real-time RT-PCR (rRT-PCR) assays on the Applied Biosystems® (ABI) 7500 Fast Dx Real-time PCR system. The panel is configured in four separate kits. Each kit consists of oligonucleotide primers. fluorescently labeled hydrolysis probes, and controls which are used in rRT-PCR assays for the in vitro qualitative detection and characterization of influenza virus RNA in respiratory specimens from patients presenting with influenza-like illness (ILI). Oligonucleotide primers and probes included in the Influenza B Lineage Genotyping Kit for detection of influenza B and the two major genetic lineages of influenza B were selected from highly conserved regions of the non-structural (NS) and HA genes, respectively. Oligonucleotide primers and probes to detect the human RNase P gene (RP) in control samples and clinical specimens are also included in the kit.
#### INTENDED USE V.
The Influenza B Lineage Genotyping Kit contains reagents and controls of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel and is intended for use in real-time RT-PCR (rRT-PCR) assays on an Applied Biosystems (ABI) 7500 Fast Dx Real-Time PCR instrument in conjunction with clinical and epidemiological information:
. For the determination of the genetic lineage of human influenza B viruses as B/Victoria or B/Yamagata lineage from viral RNA in upper respiratory tract clinical specimens (including nasopharyngeal swabs [NPS], nasal swabs [NS], throat swabs [TS], nasal aspirates [NA], nasal washes [NW] and dual nasopharyngeal/throat swabs [NPS/TS]) from human patients with signs and symptoms of respiratory infection and/or from viral culture;
- To provide epidemiologic information for surveillance of circulating influenza viruses. .
Performance characteristics for influenza B lineage genotyping were established during a season when influenza B/Victoria and B/Yamagata lineages were in circulation.
Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease.
All users, analysts, and any person reporting results from use of this device should be trained to perform and interpret the results from this procedure by a competent instructor prior to use. CDC Influenza Division will limit the distribution of this device to only those users who have successfully completed a training course provided by CDC instructors or designees.
{5}------------------------------------------------
#### VI. TECHNOLOGICAL CHARACTERISTICS
The technological characteristics of the modified CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel- Influenza B Lineage Genotyping Kit (VER 1.1. and VER 2) remain the same as the predicate device. Modifications were made primarily to address recent evolutionary changes in circulating influenza B viruses that may impact the reactivity of the current Influenza B Lineage Genotyping Kit. Two design approaches were evaluated that address specific genetic mutations in the targeted hemagglutinin (HA) gene of influenza B viruses.
CDC also evaluated the ZENTM Double-Quenched Probe technology (manufactured by Integrated DNA Technologies) as an alternate fluorescent hydrolysis probe quencher chemistry for the InfB assay of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel.
#### SUBSTANTIAL EQUIVALENCE COMPARISON VII.
The CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel, Influenza B Lineage Genotyping Kit (K172091), will serve as the predicate for the proposed change. See table 8-1 below for a detailed comparison of the modified device to the predicate.
{6}------------------------------------------------
| Table 8-1: Device Comparison | | |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Item | Predicate Device | Proposed Device |
| | CDC Human Influenza Virus Real-Time RT-PCR<br>Diagnostic Panel, Influenza B Lineage Genotyping<br>Kit [K172091] | CDC Human Influenza Virus Real-Time RT-<br>PCR Diagnostic Panel, Influenza B Lineage<br>Genotyping Kit (VER 1.1 and VER 2) |
| | The Influenza B Lineage Genotyping Kit contains<br>reagents and controls of the CDC Human Influenza<br>Virus Real-Time RT-PCR Diagnostic Panel and is<br>intended for use in real-time RT-PCR (rRT-PCR)<br>assays on an Applied Biosystems (ABI) 7500 Fast Dx<br>Real-Time PCR instrument in conjunction with<br>clinical and epidemiological information: | Same |
| | For the determination of the genetic lineage of<br>human influenza B viruses as B/Victoria or<br>B/Yamagata lineage from viral RNA in upper<br>respiratory tract clinical specimens (including<br>nasopharyngeal swabs [NPS], nasal swabs [NS],<br>throat swabs [TS], nasal aspirates [NA], nasal<br>washes [NW] and dual nasopharyngeal/throat<br>swabs [NPS/TS]) from human patients with<br>signs and symptoms of respiratory infection<br>and/or from viral culture; To provide epidemiologic information for<br>surveillance of circulating influenza viruses. | |
| Intended Use | Performance characteristics for influenza B lineage<br>genotyping were established during a season when<br>influenza B/Victoria and B/Yamagata lineages were<br>found in approximately equal proportion.<br><br>Negative results do not preclude influenza virus<br>infection and should not be used as the sole basis for<br>treatment or other patient management decisions.<br>Conversely, positive results do not rule out bacterial<br>infection or co-infection with other viruses. The agent<br>detected may not be the definite cause of disease. | |
| | All users, analysts, and any person reporting results from use of this device should be<br>trained to perform and interpret the results from this procedure by a competent<br>instructor prior to use. CDC Influenza Division will limit the distribution of this device<br>to only those users who have successfully completed a training course provided by<br>CDC instructors or designees. | |
| Organism<br>Detected | Influenza B virus, lineages B/Victoria and<br>B/Yamagata | Same |
| Specimen Types | Nasopharyngeal swabs, nasal swabs, throat swabs,<br>nasal aspirates, nasal washes and dual<br>nasopharyngeal/throat swabs from human patients<br>with signs and symptoms of respiratory infection<br>and/or from viral culture | Same |
| Technological<br>Characteristics | Real-time RT-PCR based assay | Same |
| Nucleic Acid<br>Extraction | QIAamp® DSP Viral RNA Mini Kit, QIAGEN MagNA Pure Compact –Nucleic Acid Isolation Kit I,<br>Roche MagNA Pure Compact - RNA Isolation Kit, Roche MagNA Pure LC - Total Nucleic Acid Kit, Roche QIAcube - QIAamp® DSP Viral RNA Mini Kit, QIAGEN NucliSENS® easyMAG®, bioMérieux EZ1 Advanced XL – EZ1 DSP Virus Kit and EZ1 RNA<br>Tissue Mini Kit, QIAGEN MagNA Pure 96 - DNA and Viral NA Small Volume Kit,<br>Roche | Same |
| Enzyme Master<br>Mix | Invitrogen SuperScript™ III Platinum® One-Step<br> | Same |
| | Quanta BioSciences qScript™ One-Step qRT-PCR<br>Kit, Low ROX | |
| Required<br>Instrumentation | Applied Biosystems 7500 Fast Dx Real-Time PCR<br>Instrument with SDS software version 1.4 | Same |
| Probe Quenching<br>Molecule | Black Hole Quencher Probe® (BHQ-1) [InfB assay]<br>Black Hole Quencher Plus (BHQ <i>Plus</i> ) [VIC, YAM assays] | ZEN™ or BHQ-1 (InfB assay)<br>BHQ <i>Plus</i> or BHQ-1 (VIC, YAM assays) |
# Table 8-1: Device Comparison
{7}------------------------------------------------
# VIII. ANALYTICAL PERFORMANCE EVALUATION
#### Analytical Sensitivity - Influenza B Assay
A range finding study was performed to demonstrate LOD equivalency between the currently cleared InfB assay probe quenched with BHQ and the same probe quenched with ZEN. Two characterized influenza vaccine reference viruses of a known 50% egg infectious dose titer (EID50/mL) were extracted using the Roche MagNA Pure Compact RNA Isolation Kit. The RNA was serially diluted and tested (n=3 replicates) in order to determine the apparent endpoint range using both enzymes cleared for use with the CDC Human Influenza Real-Time RT-PCR Diagnostic Panel. The acceptance criterion for LOD equivalency was defined as demonstrating 100% positivity (3 out of 3 replicates) at either the same endpoint LOD concentration or within a 5-fold dilution of each other. All assays demonstrated similar reactivity (Tables 8-2 and 8-3).
| Influenza<br>B/Nevada/03/2011<br>Titer (EID50/mL) | Invitrogen<br>Superscript™ | | Quanta qScript™ | |
|---------------------------------------------------|----------------------------|------------|-------------------|------------|
| | InfB (BHQ)<br>IVD | InfB (ZEN) | InfB (BHQ)<br>IVD | InfB (ZEN) |
| 104.2 | 3/3*<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| 103.5 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| 102.8 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| 102.1 | 1/3<br>(+) | 1/3<br>(+) | 0/3<br>(-) | 0/3<br>(-) |
| 101.4 | 2/3<br>(+) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) |
| 100.7 | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) |
| 100.0 | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) |
### Table 8-2: Influenza B (InfB) Assay LOD Equivalencv - B/Nevada/03/2011
Indicates number of positive replicates out of three total.
{8}------------------------------------------------
| Influenza<br>B/Wisconsin/1/2010<br>Titer (EID50/mL) | Invitrogen Superscript™ | | Quanta qScript™ | |
|-----------------------------------------------------|-------------------------|------------|-------------------|------------|
| | InfB (BHQ)<br>IVD | InfB (ZEN) | InfB (BHQ)<br>IVD | InfB (ZEN) |
| 104.2 | 3/3*<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| 103.5 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| 102.8 | 1/3<br>(+) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) |
| 102.1 | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) |
| 101.4 | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) |
| 100.7 | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) |
| 100.0 | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) |
Table 8-3: Influenza B (InfB) Assay LOD Equivalency - B/Wisconsin/1/2010
Indicates number of positive replicates out of three total.
### Analytical Sensitivity- Influenza B Lineage Genotyping Assay Update
A range finding study was performed with the currently cleared assays of the Influenza B Lineage Genotyping Kit and with modified VIC and YAM assays (VER 1.1) as well as newly designed VIC and Y AM assays (VER 2). These studies were performed to demonstrate the LOD equivalency and improved reactivity of the VER 1.1 and VER 2 VIC and YAM assays with one historical benchmark strain and one current strain of both the B/Victoria and B/Yamagata lineages of influenza B virus. The test samples consisted of serially diluted RNA extracted with the Roche MagNA Pure Compact RNA Isolation Kit from characterized virus stocks of known 50% egg infectious dose titer (EID50mL). Three replicates per dilution were tested to determine the apparent endpoint range (Tables 8-4 to 8-7) using both enzymes cleared for use with the CDC Human Influenza Real-Time RT-PCR Diagnostic Panel. The acceptance criterion for LOD equivalency was defined as demonstrating 100% positivity (3 out of 3 replicates) at either the same endpoint LOD concentration or within a 5-fold dilution of each other. All assays demonstrated similar reactivity with apparent LOD endpoints within one 5-fold dilution of each other.
{9}------------------------------------------------
| Influenza<br>B/Nevada/03/2011<br>Titer (EID50/mL) | Invitrogen Superscript™ | | | | Quanta qScript™ | | | |
|---------------------------------------------------|-------------------------|------------|----------------|--------------|-----------------|------------|----------------|--------------|
| | InfB<br>IVD | VIC<br>IVD | VIC<br>VER 1.1 | VIC<br>VER 2 | InfB<br>IVD | VIC<br>IVD | VIC<br>VER 1.1 | VIC<br>VER 2 |
| 104.2 | 3/3* | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 |
| | (+) | (+) | (+) | (+) | (+) | (+) | (+) | (+) |
| 103.5 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 |
| | (+) | (+) | (+) | (+) | (+) | (+) | (+) | (+) |
| 102.8 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 |
| | (+) | (+) | (+) | (+) | (+) | (+) | (+) | (+) |
| 102.1 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 |
| | (+) | (+) | (+) | (+) | (+) | (+) | (+) | (+) |
| 101.4 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 0/3 | 3/3 | 3/3 |
| | (+) | (+) | (+) | (+) | (+) | (+) | (+) | (+) |
| 100.7 | 1/3 | 0/3 | 2/3 | 3/3 | 0/3 | 0/3 | 1/3 | 0/3 |
| | (+) | (-) | (+) | (+) | (-) | (+) | (+) | (-) |
| 100.0 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| | (-) | (-) | (-) | (-) | (-) | (-) | (-) | (-) |
| 10-0.7 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| | (-) | (-) | (-) | (-) | (-) | (-) | (-) | (-) |
Table 8-4: B/Victoria Assay LOD Equivalency- VER 1.1 and VER 2 Designs - B/Nevada/03/2011
* Indicates number of positive replicates out of three total.
| Table 8-5: B/Victoria Assay LOD Equivalency- VER 1.1 and VER 2 Designs - B/Marryand/15/2016 | | |
|---------------------------------------------------------------------------------------------|--|--|
| Influenza<br>B/Maryland/15/2016<br>Titer (EID50/mL) | Invitrogen Superscript™ | | | | Quanta qScript™ | | | |
|-----------------------------------------------------|-------------------------|------------|----------------|--------------|-----------------|------------|----------------|--------------|
| | InfB<br>IVD | VIC<br>IVD | VIC<br>VER 1.1 | VIC<br>VER 2 | InfB<br>IVD | VIC<br>IVD | VIC<br>VER 1.1 | VIC<br>VER 2 |
| 104.5 | 3/3*<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| 103.8 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| 103.1 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| 102.4 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| 101.7 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 2/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| 101.0 | 3/3<br>(+) | 0/3<br>(-) | 3/3<br>(+) | 3/3<br>(+) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) |
| 100.3 | 1/3<br>(+) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) |
| 10-0.4 | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) |
* Indicates number of positive replicates out of three total.
{10}------------------------------------------------
| Influenza<br>B/Texas/06/2011<br>Titer (EID50/mL) | Invitrogen Superscript™ | | | | Quanta qScript™ | | | |
|--------------------------------------------------|-------------------------|------------|----------------|--------------|-----------------|------------|----------------|--------------|
| | InfB<br>IVD | YAM<br>IVD | YAM<br>VER 1.1 | YAM<br>VER 2 | InfB<br>IVD | YAM<br>IVD | YAM<br>VER 1.1 | YAM<br>VER 2 |
| 104.9 | 3/3*<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| 104.2 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| 103.5 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 2/3<br>(+) | 3/3<br>(+) |
| 102.8 | 3/3<br>(+) | 3/3<br>(+) | 0/3<br>(-) | 3/3<br>(+) | 3/3<br>(+) | 2/3<br>(+) | 0/3<br>(-) | 3/3<br>(+) |
| 102.1 | 2/3<br>(+) | 1/3<br>(+) | 1/3<br>(+) | 1/3<br>(+) | 0/3<br>(-) | 2/3<br>(+) | 0/3<br>(-) | 2/3<br>(+) |
| 101.4 | 2/3<br>(+) | 0/3<br>(-) | 0/3<br>(-) | 1/3<br>(+) | 0/3<br>(-) | 1/3<br>(+) | 0/3<br>(-) | 2/3<br>(+) |
| 100.7 | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) |
| 100.0 | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) |
Table 8-6: B/Yamagata Assay LOD Equivalency- VER 1.1 and VER 2 Designs - B/Texas/06/2011
* Indicates number of positive replicates out of three total.
| Table 8-7: B/Yamagata Assay LOD Equivalency- VER 1.1 and VER 2 Designs - B/Texas/81/2016 | | |
|------------------------------------------------------------------------------------------|--|--|
| | | |
| Influenza<br>B/Texas/81/2016<br>Titer (EID50/mL) | Invitrogen SuperscriptTM | | | | Quanta qScriptTM | | | |
|--------------------------------------------------|--------------------------|------------|----------------|--------------|------------------|------------|----------------|--------------|
| | InfB<br>IVD | YAM<br>IVD | YAM<br>VER 1.1 | YAM<br>VER 2 | InfB<br>IVD | YAM<br>IVD | YAM<br>VER 1.1 | YAM<br>VER 2 |
| 104.3 | 3/3*<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| 103.6 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| 102.9 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| 102.2 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| 101.5 | 3/3<br>(+) | 2/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 1/3<br>(+) | 2/3<br>(+) |
| 100.8 | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 1/3<br>(+) | 0/3<br>(-) | 0/3<br>(-) |
| 100.1 | 1/3<br>(+) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 2/3<br>(+) | 0/3<br>(-) | 0/3<br>(-) |
| 10-0.6 | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) | 0/3<br>(-) |
* Indicates number of positive replicates out of three total.
{11}------------------------------------------------
A confirmation of the LOD for each modified or new assay was performed using a current strain of the corresponding influenza B/Victoria or B/Yamagata lineage. The VER 1.1 and VER 2 VIC and YAM primer and probe sets were tested alongside the currently cleared InfB assay as required for interpretation by the routine testing algorithm. The confirmatory LOD testing for each primer and probe set was performed on extraction replicates (n=20) of each dilution. The lowest virus concentration where InfB and VIC or InfB and YAM primer and probes sets demonstrated ≥ 95% detection is reported as the LOD for each virus (Table 8-8).
| Influenza B Lineage | Influenza Strain Designation | Assay Design | LOD (EID50/mL) | |
|---------------------|------------------------------|--------------|--------------------------|------------------|
| | | | Invitrogen SuperScriptTM | Quanta qScriptTM |
| Victoria | B/Maryland/15/2016 | VER 1.1 | 101.7 | 101.7 |
| | | VER 2 | 101.7 | 101.7 |
| Yamagata | B/Texas/81/2016 | VER 1.1 | 102.2 | 101.5 |
| | | VER 2 | 102.2 | 101.5 |
Table 8-8: LOD Confirmation Summary
### Analytical Sensitivity - Inclusivity
Inclusivity testing was performed to demonstrate the capability of the modified VER 1.1 and VER 2 VIC and YAM primer and probe sets to detect strains of the corresponding influenza B lineage at or near the established LOD. Ten influenza B viruses of each lineage and representative of different geographic locations and phylogenetic clades were selected. Characterized stocks were serially diluted to concentrations near the LOD of the assays and extracted using the Roche MagNA Pure Compact RNA Isolation Kit. Samples were tested in triplicate with the VER 2 VIC and YAM assays with both enzyme systems cleared for use with the CDC Human Influenza Real-Time RT-PCR Diagnostic Panel. The inclusivity results are summarized in Tables 8-9 and 8-10.
{12}------------------------------------------------
| Influenza Virus<br>Strain Designation | EID50/mL or<br>TCID50/mL | Invitrogen<br>SuperScript™ | | Quanta qScript™ | |
|---------------------------------------|--------------------------|----------------------------|--------------|-----------------|--------------|
| | | VIC<br>VER 1.1 | VIC<br>VER 2 | VIC<br>VER 1.1 | VIC<br>VER 2 |
| B/Hong Kong/259/2010 | 10 1.2 | 3/3*<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| B/Bolivia/1526/2010 | 10 2.4 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| B/Laos/89/2011 | 10 1.2 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| B/Michigan/09/2011 | 10 1.5 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| B/New Jersey/01/2012 | 10 1.9 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| B/Montana/05/2012 | 10 2.4 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| B/Texas/02/2013 | 10 2.1 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| B/Florida/103/2016 | 10 0.3 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| B/Florida/76/2016 | 10 1.5 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| B/Hong Kong/269/2017 | 10 -0.8 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
Table 8-9: B/Victoria Assay Inclusivity VER 1.1 and VER 2 Designs
* Indicates number of positive replicates out of three total.
{13}------------------------------------------------
| Influenza Virus<br>Strain Designation | EID50 /mL<br>or<br>TCID50/mL | Invitrogen<br>SuperScript™ | | Quanta qScript™ | |
|---------------------------------------|------------------------------|----------------------------|--------------|-----------------|--------------|
| | | YAM<br>VER 1.1 | YAM<br>VER 2 | YAM<br>VER 1.1 | YAM<br>VER 2 |
| B/Brisbane/03/2007 | 10 2.4 | 3/3*<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| B/Pennsylvania/07/2007 | 10 2.2 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| B/Hubei-<br>Wujiagang/158/2009 | 10 2.2 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| B/Wisconsin/01/2010 | 10 2.9 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| B/Finland/39/2010 | 10 2.1 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| B/Estonia/55669/2011 | 10 2.8 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| B/Taiwan/1242/2011 | 10 3.9 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| B/Massachusetts/02/2012 | 10 2.2 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| B/Phuket/3073/2013 | 10 2.9 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
| B/Guangdong-<br>Liwan/1133/2014 | 10 3.9 | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) | 3/3<br>(+) |
Table 8-10: B/Yamagata Assay Inclusivity VER 1.1 and VER 2 Designs
Indicates number of positive replicates out of three total.
# Analytical Specificity – Cross-Reactivity
Cross-reactivity of the modified VER 1.1 and VER 2 VIC and YAM primer and probe sets was evaluated by testing each set with influenza B viruses of the opposite lineage and from diverse geographic locations. Samples were extracted from high titer preparations of viruses (≥ 106 EID50 or TCID50/mL) using the Roche MagNA Pure Compact RNA Isolation Kit. Cross-reactivity was evaluated with both enzyme systems cleared for use with the CDC Human Influenza Real-Time RT-PCR Diagnostic Panel. No cross-reactivity was detected with either primer and probe set design (Tables 8-11 and 8-12).
|--|
| Influenza Virus<br>Strain Designation<br>B/Yamagata Lineage | EID50/mL<br>or<br>TCID50/mL | Invitrogen<br>SuperScript™ | | Quanta qScript™ | |
|-------------------------------------------------------------|-----------------------------|----------------------------|--------------|-----------------|--------------|
| | | VIC<br>VER 1.1 | VIC<br>VER 2 | VIC<br>VER 1.1 | VIC<br>VER 2 |
| B/Brisbane/03/2007 | 10 8.2 | - | - | - | - |
| B/Pennsylvania/07/2007 | 10 8.4 | - | - | - | - |
| B/Hubei-Wujiagang/158/2009 | 10 6.2 | - | - | - | - |
| B/Wisconsin/01/2010 | 10 8.5 | - | - | - | - |
| B/Finland/39/2010 | 10 8.9 | - | - | - | - |
| B/Estonia/55669/2011 | 10 8.4 | - | - | - | - |
| B/Taiwan/1242/2011 | 10 9.1 | - | - | - | - |
| B/Massachusetts/02/2012 | 10 6.3 | - | - | - | - |
| B/Phuket/3073/2013 | 10 6.5 | - | - | - | - |
| B/Guangdong-Liwan/1133/2014 | 10 4.2 | - | - | - | - |
{14}------------------------------------------------
| Influenza Virus<br>Strain Designation<br>B/Victoria Lineage | EID50/mL<br>or<br>TCID50/mL | Invitrogen<br>SuperScript™ | | Quanta<br>qScript™ | |
|-------------------------------------------------------------|-----------------------------|----------------------------|--------------|--------------------|--------------|
| | | YAM<br>VER 1.1 | YAM<br>VER 2 | YAM<br>VER 1.1 | YAM<br>VER 2 |
| B/Hong Kong/259/2010 | 10 $8.4$ | - | - | - | - |
| B/Bolivia/1526/2010 | 10 $8.2$ | - | - | - | - |
| B/Laos/89/2011 | 10 $8.2$ | - | - | - | - |
| B/Michigan/09/2011 | 10 $8.9$ | - | - | - | - |
| B/New Jersey/01/2012 | 10 $8.1$ | - | - | - | - |
| B/Montana/05/2012 | 10 $7.8$ | - | - | - | - |
| B/Texas/02/2013 | 10 $9.9$ | - | - | - | - |
| B/Florida/103/2016 | 10 $9.2$ | - | - | - | - |
| B/Florida/76/2016 | 10 $9.9$ | - | - | - | - |
| B/Hong Kong/269/2017 | 10 $9.9$ | - | - | - | - |
Table 8-12: B/Yamagata Assay Cross-Reactivity VER 1.1 and VER 2 Designs
# Analytical Specificity - Exclusivity
### Exclusivity with Influenza A Viruses
Exclusivity of the VER 1.1 and VER 2 VIC and YAM assays was examined with influenza A viruses of various subtypes that circulate in humans and from animal origin that infect humans (Table 8-13 and 8-14). Samples were prepared from characterized, high titer stocks (≥106 TCID50 or EID50/mL) by extracting RNA using the Roche MagNA Pure Compact RNA Isolation Kit. Testing was performed using both of the currently cleared enzyme systems.
| | Table 8-13: Exclusivity with Influenza A Viruses- VER 1.1 and VER 2 -Invitrogen SuperScript™ | | | |
|--|----------------------------------------------------------------------------------------------|--|--|--|
| | | | | |
| Influenza Virus<br>Strain Designation | Origin | Subtype | EID50/mL<br>or<br>TCID50/mL | Invitrogen SuperScript™ | | | |
|----------------------------------------------|--------|------------------|-----------------------------|-------------------------|--------------|----------------|--------------|
| | | | | VIC<br>VER 1.1 | VIC<br>VER 2 | YAM<br>VER 1.1 | YAM<br>VER 2 |
| A/Michigan/45/2015 | Human | A(H1N1)<br>pdm09 | $10^{8.3}$ | - | - | - | - |
| A/Hong Kong/4801/2014 | Human | A(H3N2) | $10^{7.9}$ | - | - | - | - |
| A/Ohio/35/2017 | Swine | A(H1N2)v | $10^{6.9}$ | - | - | - | - |
| A/Ohio/13/2017 | Swine | A(H3N2)v | $10^{6.9}$ | - | - | - | - |
| A/gyrfalcon/<br>Washington/<br>41088-6/2014 | Avian | A(H5N8) | $10^{9.75}$ | - | - | - | - |
| A/Northern pintail/Washington/<br>40964/2014 | Avian | A(H5N2) | $10^{9.4}$ | - | - | - | - |
| A/Bangladesh/<br>0994/2011 | Avian | A(H9N2) | $10^{10.5}$ | - | - | - | - |
| A/Anhui/01/2013 | Avian | A(H7N9) | $10^{10.9}$ | - | - | - | - |
{15}------------------------------------------------
| Influenza Virus<br>Strain Designation | Origin | Subtype | EID50/mL<br>or<br>TCID50/mL | Quanta qScript™ | | | |
|----------------------------------------------|--------|------------------|-----------------------------|-----------------|--------------|----------------|--------------|
| | | | | VIC<br>VER 1.1 | VIC<br>VER 2 | YAM<br>VER 1.1 | YAM<br>VER 2 |
| A/Michigan/45/<br>2015 | Human | A(H1N1)<br>pdm09 | 108.3 | - | - | - | - |
| A/Hong Kong/4801/2014 | Human | A(H3N2) | 107.9 | - | - | - | - |
| A/Ohio/35/2017 | Swine | A(H1N2)v | 106.9 | - | - | - | - |
| A/Ohio/13/2017 | Swine | A(H3N2)v | 106.9 | - | - | - | - |
| A/gyrfalcon/<br>Washington/<br>41088-6/2014 | Avian | A(H5N8) | 109.75 | - | - | - | - |
| A/Northern pintail/Washington/<br>40964/2014 | Avian | A(H5N2) | 109.4 | - | - | - | - |
| A/Bangladesh/<br>0994/2011 | Avian | A(H9N2) | 1010.5 | - | - | - | - |
| A/Anhui/01/2013 | Avian | A(H7N9) | 1010.9 | - | - | - | - |
Table 8-14: Exclusivity with Influenza A Viruses- VER 1.1 and VER 2 -Quanta qScript™
## Exclusivity with Non-Influenza Respiratory Pathogens
Exclusivity of the VER 2 VIC and YAM assays was examined with non-influenza human respiratory viruses, bacteria, and yeast (Table 8-15). Nucleic acids were extracted using the Roche MagNA Pure Compact RNA Isolation Kit from 36 organisms (16 viruses, 19 bacteria, and 1 yeast) representing common respiratory pathogens or flora commonly present in specimens collected from the nasopharynx region. All bacteria, yeast, and non-influenza viruses were from CDC repositories or acquired from American Type Culture Collection (ATCC, Manassas, VA) and tested at high titers, typically ≥ 100 TCID50 or EID50/mL, ≥ 10° CFU/mL, or as high as culture allowed. Testing was performed using both of the currently cleared enzyme systems.
{16}------------------------------------------------
| Organism Tested | | | SuperScript | | qScript | |
|------------------------------|----------------------|-----------------------------|-------------|-------------|-------------|-------------|
| Bacteria and Yeast | Strain | cfu / mL | VIC<br>VER2 | YAM<br>VER2 | VIC<br>VER2 | YAM<br>VER2 |
| Bordetella pertussis | A639 | 10 8.3 | - | - | - | - |
| Candida albicans | 2001-21-196 | 10 8.8 | - | - | - | - |
| Chlamydia pneumoniae1 | TW183 | 40 IFU/mL | - | - | - | - |
| Corynebacterium diphtheriae | NA2 | 10 10 | - | - | - | - |
| Escherichia coli | K12 | 10 9.6 | - | - | - | - |
| Haemophilus influenzae | M15709 | 10 6.4 | - | - | - | - |
| Lactobacillus plantarum | NA | 10 8.8 | - | - | - | - |
| Legionella pneumophila | NA | 10 10.3 | - | - | - | - |
| Moraxella catarrhalis | M15757 | 10 9.5 | - | - | - | - |
| Mycobacterium tuberculosis3 | H37Rv | 95 ng/ µL | - | - | - | - |
| Mycoplasma pneumoniae | MI-29 | 10 7.7 | - | - | - | - |
| Neisseria elongata | NA | 10 8.6 | - | - | - | - |
| Neisseria meningitidis | M2578 | 10 7.9 | - | - | - | - |
| Pseudomonas aeruginosa | NA | 10 10.5 | - | - | - | - |
| Staphylococcus epidermidis | NA | 10 10.5 | - | - | - | - |
| Staphylococcus aureus | NA | 10 10.7 | - | - | - | - |
| Streptococcus pneumoniae | 249-06<br>(Thailand) | 10 6.6 | - | - | - | - |
| Streptococcus pyogenes | 7790-06 | 10 7.5 | - | - | - | - |
| Streptococcus salivarius | SS1672 | 10 8.4 | - | - | - | - |
| Viruses | Strain | EID50/mL<br>or<br>TCID50/mL | VIC<br>VER2 | YAM<br>VER2 | VIC<br>VER2 | YAM<br>VER2 |
| Enterovirus | Echo 6 | 10 6.9 | - | - | - | - |
| Human Adenovirus, type 1 | Ad.71 | 10 9.2 | - | - | - | - |
| Human Adenovirus, type 7a | S-1058 | 10 7.1 | - | - | - | - |
| Human Coronavirus virus3 | OC43 | 50.4 ng/µL | - | - | - | - |
| Human Coronavirus virus3 | 299E | 31.6 ng/µL | - | - | - | - |
| Human Rhinovirus A | 1A | 10 5.8 | - | - | - | - |
| Human Parainfluenza 1 virus3 | NA | 3.0 ng/ µL | - | - | - | - |
| Human Parainfluenza 2 virus | Greer | 10 3.1 | - | - | - | - |
| Human Parainfluenza 3 virus | C-243 | 10 7.9 | - | - | - | - |
| Respiratory Syncytial virus | CH93-18b | 10 6.8 | - | - | - | - |
| Herpes Simplex Virus | KOS | 10 8.4 | - | - | - | - |
| Varicella-zoster Virus | AV92-3 | 10 4.4 | - | - | - | - |
| Epstein Barr Virus3 | B95-8 | 1.7 ng/µL | - | - | - | - |
| Measles Virus | Edmonston | 10 5.2 | - | - | - | - |
| Mumps Virus | Enders | 10 7.2 | - | - | - | - |
| Cytomegalovirus | AD-169 | 10 6.9 | - | - | - | - |
## Table 8-15: Assay Exclusivity with Respiratory Viruses, Bacteria, and Yeast- VER 2 Design
1 Organism quantified by Infectious Forming Units (IFU)
² NA = not applicable
3 Organism quantified by spectrophotometry (ng/μL)
{17}------------------------------------------------
#### CLINICAL PERFORMANCE EVALUATION IX.
#### Influenza B Assay -Retrospective Study
The clinical performance of the InfB oligonucleotide probe quenched with ZEN was evaluated using retrospective clinical samples collected during the 2011-2012 influenza season that were previously determined to be positive or negative for influenza B virus. A total of 30 positive and 50 negative upper respiratory tract samples were evaluated with the cleared InfB assay using either the existing oligonucleotide probe quenched with BHQ or the investigational probe quenched with ZEN. Testing was performed with both enzymes cleared for use with the CDC Human Influenza Real-Time RT-PCR Diagnostic Panel and one of the currently cleared extraction methods. Result interpretation followed the instructions of the Package Insert for the cleared Influenza B Lineage Genotyping Kit. The InfB assay containing the investigational probe quenched with ZEN demonstrated 100% positive and negative agreement with the cleared oligonucleotide probe quenched with BHQ (Tables 8-16 and 8-17).
| | | Invitrogen SuperScript™ | Quanta qScript™ | |
|---------------|--------------------|----------------------------------|--------------------|----------------------------------|
| Specimen Type | # of<br>Positives1 | % Positive Agreement<br>(95% CI) | # of<br>Positives1 | % Positive Agreement<br>(95% CI) |
| NPS, NS | 18/18 | 100.0 (82.4-100.0) | 18/18 | 100.0 (82.4-100.0) |
| NPS/TS | 11/11 | 100.0 (74.1-100.0) | 11/11 | 100.0 (74.1-100.0) |
| NW | 1/1 | 100.0 (20.7-100.0) | 1/1 | 100.0 (20.7-100.0) |
| Total | 30/30 | 100.0 (88.7-100.0) | 30/30 | 100.0 (88.7-100.0) |
#### Table 8-16: InfB Assay-Retrospective Positive Clinical Study Results
1 Proportion of positive samples correctly identified versus the comparator.
Table 8-17: InfB Assay-Retrospective Negative Clinical Study Results
| | Invitrogen SuperScript™ | | Quanta qScript™ | |
|---------------|-------------------------|----------------------------------|--------------------|----------------------------------|
| Specimen Type | # of<br>Negatives1 | % Negative Agreement<br>(95% CI) | # of<br>Negatives1 | % Negative Agreement<br>(95% CI) |
| NPS, NS | 48/48 | 100.0 (92.6-100.0) | 48/48 | 100.0 (92.6-100.0) |
| TS | 2/2 | 100.0 (34.2-100.0) | 2/2 | 100.0 (34.2-100.0) |
| Total | 50/50 | 100.0 (92.9-100.0) | 50/50 | 100.0 (92.9-100.0) |
1 Proportion of negative samples correctly identified versus the comparator.
{18}------------------------------------------------
Influenza B Lineage Genotyping Kit (VER 2) - Prospective Study
A prospective clinical investigation was conducted at 3 U.S. public health laboratories using upper respiratory tract specimens collected during the 2016-2017 influenza season. Samples were taken from specimens collected for routine influenza testing at each site from individuals symptomatic for influenza-like illness. The range of patient ages and specimen types for the total of 592 samples collected are represented in Table 8-18. A total of 13 were excluded for reasons of unspecified specimen type, inconclusive result of the comparator, or technician testing error. A total of 579 upper respiratory tract specimens were included in the data analysis.
| Patient Age | Totals |
|----------------|--------|
| 0-16 | 149 |
| 17-54 | 192 |
| ≥55 | 249 |
| Not Reported | 2 |
| Specimen Type1 | Totals |
| NPS | 374 |
| NPS/TS | 44 |
| NA | 12 |
| NW | 34 |
| TS | 13 |
| NS | 114 |
| Not Reported | 1 |
### Table 8-18: Specimen Information
1 NPS=nasopharyngeal swab, NPS/TS=dual nasopharyngeal and throat swab, NA=nasal aspirate, NW=nasal wash, TS=throat swab, NS=nasal swab
Specimens were tested with assays from the FDA-cleared CDC Human Influenza Real-Time RT-PCR Diagnostic Panel and the investigational Influenza B Lineage Genotyping Kit (VER 2). The performance is summarized in Tables 8-19 and 8-20. Due to the low prevalence of influenza B/Victoria lineage viruses in specimens collected during the prospective study, the performance of the Influenza B Lineage Genotyping Kit (VER 2) was also evaluated in a retrospective study.
{19}------------------------------------------------
| Specimen Type1 | # of<br>Positives2 | % Sensitivity (95% CI) | # of Negatives3 | % Specificity (95%<br>CI) |
|----------------|--------------------|------------------------|-----------------|---------------------------|
| NPS, NS | 5/5 | 100.0 (56.6 – 100.0) | 472/472 | 100.0 (99.2 - 100.0) |
| NPS/TS | 0/0 | NA4 | 44/44 | 100.0 (92.0 - 100.0) |
| TS | 0/0 | NA | 13/13 | 100.0 (77.2 – 100.0) |
| NA, NW | 0/0 | NA | 45/45 | 100.0 (92.1 – 100.0) |
| Total5 | 5/5 | 100.0 (56.6 – 100.0) | 574/574 | 100.0 (99.3 – 100.0) |
Table 8-19: B/Victoria Assay - VER 2 Design- Prospective Study Results
1 NPS=nasopharyngeal swab. NPS/TS=dual nasopharyngeal and throat swab. NA=nasal wash. TS=throat swab, NS=nasal swab
2Proportion of positive samples correctly identified versus the comparator.
3 Proportion of negative samples correctly identified versus the comparator.
4NA=not applicable
| Specimen Type1 | # of<br>Positives2 | % Sensitivity (95% CI) | # of Negatives3 | % Specificity (95%<br>CI) |
|----------------|--------------------|------------------------|-----------------|---------------------------|
| NPS, NS | 31/31 | 100.0 (89.0 – 100.0) | 446/446 | 100.0 (99.1 - 100.0) |
| NPS/TS | 2/2 | 100.0 (34.2 – 100.0) | 42/42 | 100.0 (91.6 - 100.0) |
| TS | 0/0 | NA4 | 13/13 | 100.0 (77.2 - 100.0) |
| NA, NW | 5/5 | 100.0 (56.6 - 100.0) | 39/40 | 97.5 (87.1 - 99.6) |
| Total5 | 38/38 | 100.0 (90.8 - 100.0) | 540/541 | 99.8 (99.0 - 100.0) |
Table 8-20: B/Yamagata Assay - VER 2 Design - Prospective Study Results
1 NPS=nasopharyngeal swab, NPS/TS=dual nasopharyngeal and throat swab, NA=nasal wash, TS=throat swab, NS=nasal swab
2Proportion of positive samples correctly identified versus the comparator.
3 Proportion of negative samples correctly identified versus the comparator.
4NA=not applicable
#### Influenza B Lineage Genotyping Kit VER 1.1 and VER 2 - Retrospective Study Results
A retrospective study was performed using the VER 1.1 and YER 2 VIC and Y AM assays to evaluate their clinical performance. Upper respiratory tract clinical samples were collected during the 2017 and 2017-2018 influenza seasons and determined to be positive for influenza B/Victoria or B/Yamagata viruses using the FDA-cleared Influenza B Lineage Genotyping Kit. In the current study, specimens were tested using the cleared InfB assay and the investigative VER 1.1 and VER 2 VIC and Y AM assays. A total of 126 specimens positive for either influenza B/Victoria or influenza B/Yamagata viruses and 61 specimens negative for influenza B viruses were tested. Four lung tissue specimens were included in the testing, but were used for informational purposes only and not included in calculations of assay performance (Tables 8-21 to 8-24). Result interpretation followed the instructions of the Package Insert for the cleared Influenza B Lineage Genotyping Kit.
{20}------------------------------------------------
| Specimen Type | # of<br>Positives1 | % Positive Agreement<br>(95% CI) | # of<br>Negatives2 | % Negative Agreement<br>(95% CI) |
|---------------|--------------------|----------------------------------|--------------------|----------------------------------|
| NPS, NS | 29/29 | 100.0 (88.3 - 100.0) | 156/156 | 100.0 (97.6 - 100.0) |
| TS | 1/1 | 100.0 (20.7 - 100.0) | 1/1 | 100.0 (20.7 - 100.0) |
| Total | 30/30 | 100.0 (88.7 - 100.0) | 157/157 | 100.0 (97.6 - 100.0) |
Table 8-21: VIC VER 1.1 Design-Retrospective Study Results
1Proportion of positive samples correctly identified versus the comparator.
2Proportion of negative samples correctly identified versus the comparator.
| Specimen Type | # of<br>Positives1 | % Positive Agreement<br>(95% CI) | # of<br>Negatives2 | % Negative Agreement<br>(95% CI) |
|---------------|--------------------|----------------------------------|--------------------|----------------------------------|
| NPS, NS | 95/95 | 100.0 (96.1 – 100.0) | 90/90 | 100.0 (95.9 – 100.0) |
| TS | 1/1 | 100.0 (20.7 - 100.0) | 1/1 | 100.0 (20.7 - 100.0) |
| Total | 96/96 | 100.0 (96.2 - 100.0) | 91/91 | 100.0 (96.0 – 100.0) |
### Table 8-22: YAM VER 1.1 Design-Retrospective Study Results
1Proportion of positive samples correctly identified versus the comparator.
2Proportion of negative samples correctly identified versus the comparator.
| Specimen Type | # of<br>Positives1 | % Positive Agreement<br>(95% CI) | # of<br>Negatives2 | % Negative Agreement<br>(95% CI) |
|---------------|--------------------|----------------------------------|--------------------|----------------------------------|
| NPS, NS | 29/29 | 100.0 (88.3 – 100.0) | 156/156 | 100.0 (97.6 – 100.0) |
| TS | 1/1 | 100.0 (20.7 - 100.0) | 1/1 | 100.0 (20.7 - 100.0) |
| Total | 30/30 | 100.0 (88.7 – 100.0) | 157/157 | 100.0 (97.6 – 100.0) |
### Table 8-23: VIC VER 2 Design-Retrospective Study Results
1Proportion of positive samples correctly identified versus the comparator.
2Proportion of negative samples correctly identified versus the comparator.
| Specimen Type | # of<br>Positives1 | % Positive Agreement<br>(95% CI) | # of<br>Negatives2 | % Negative Agreement<br>(95% CI) |
|---------------|--------------------|----------------------------------|--------------------|----------------------------------|
| NPS, NS | 95/95 | 100.0 (96.1 – 100.0) | 90/90 | 100.0 (95.9 – 100.0) |
| TS | 1/1 | 100.0 (20.7 - 100.0) | 1/1 | 100.0 (20.7 - 100.0) |
| Total | 96/96 | 100.0 (96.2 - 100.0) | 91/91 | 100.0 (96.0 – 100.0) |
## Table 8-24: YAM VER 2 Design-Retrospective Study Results
1Proportion of positive samples correctly identified versus the comparator.
2Proportion of negative samples correctly identified versus the comparator.
#### X. CONCLUSION
The modification of the Influenza B Lineage Genotyping Kit of the CDC Human Influenza Virus rRT-PCR Diagnostic Panel to ensure comprehensive detection of influenza B viruses does not
{21}------------------------------------------------
change the fundamental scientific technology of the device. Analytical and clinical data demonstrate that the performance of the modified device to detect and characterize influenza B viruses is accomplished with high positive and negative percent agreement in a manner substantially equivalent to the predicate. The indications for use remain the same.