← Product Code [NOP](/submissions/MI/subpart-d%E2%80%94serological-reagents/NOP) · K041817

# WEST NILE DETECT IGM ELISA (K041817)

_Inbios Intl., Inc. · NOP · Nov 19, 2004 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/NOP/K041817

## Device Facts

- **Applicant:** Inbios Intl., Inc.
- **Product Code:** [NOP](/submissions/MI/subpart-d%E2%80%94serological-reagents/NOP.md)
- **Decision Date:** Nov 19, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3940
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

The West Nile Detect™ IgM Capture ELISA is for the qualitative detection of IgM antibodies to WNV recombinant antigens (WNRA) in serum for the presumptive clinical laboratory diagnosis of West Nile virus infection in patients with clinical symptoms consistent with meningoencephalitis. Positive results must be confirmed by Plaque Reduction Neutralization Test (PRNT), or by using the current CDC guidelines for diagnosis of this disease. Assay performance characteristics have not been established for testing cord blood, neonate, prenatal screening, general population screening without symptoms of meningoencephalitis or automated instruments. This assay is not FDA cleared or approved for testing blood or plasma donors

## Device Story

The West Nile Detect IgM Capture ELISA is an in vitro diagnostic assay used in clinical laboratories to detect IgM antibodies against West Nile Virus (WNV) in human serum. The device utilizes recombinant WNV antigens to capture specific IgM antibodies present in patient samples. The assay is performed by laboratory technicians; results are interpreted by clinicians to support the diagnosis of WNV infection in symptomatic patients. Positive results require confirmation via Plaque Reduction Neutralization Test (PRNT) or current CDC guidelines. The device aids in clinical decision-making by providing presumptive evidence of recent WNV infection, facilitating appropriate patient management.

## Clinical Evidence

No clinical data provided in the document.

## Technological Characteristics

IgM capture ELISA; microtiter plate format. Reagents: anti-human IgM coated wells, recombinant WNV antigen (WNRA), HRP-labeled WNRA-specific antibody, TMB substrate. Energy source: none (manual/instrument-read absorbance). Connectivity: none. Sterilization: N/A.

## Regulatory Identification

West Nile virus serological reagents are devices that consist of antigens and antisera for the detection of anti-West Nile virus IgM antibodies, in human serum, from individuals who have signs and symptoms consistent with viral meningitis/encephalitis. The detection aids in the clinical laboratory diagnosis of viral meningitis/encephalitis caused by West Nile virus.

## Special Controls

*Classification.* Class II (special controls). The special control is FDA's guidance entitled “Class II Special Controls Guidance Document: Serological Reagents for the Laboratory Diagnosis of West Nile Virus.” See § 866.1(e) for the availability of this guidance document.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k041817

B. Purpose for Submission:
New device clearance

C. Measurand:
West Nile Virus IgM Antibody

D. Type of Test:
Qualitative, ELISA

E. Applicant:
InBios International, Inc.

F. Proprietary and Established Names:
West Nile Detect IgM ELISA

G. Regulatory Information:

a) Regulation section:
West Nile Virus, serological reagents (21 CFR 866.3940).

b) Classification:
Class II
Product Code:
NOP

c) Panel:
83 Microbiology

H. Intended Use:

a) Intended use(s):
The West Nile Detect™ IgM Capture ELISA is for the qualitative detection of IgM antibodies to WNV recombinant antigens (WNRA) in serum for the presumptive clinical laboratory diagnosis of West Nile virus infection in patients with clinical symptoms consistent with meningoencephalitis. Positive results must be confirmed by Plaque Reduction Neutralization Test (PRNT), or by using the current CDC guidelines for diagnosis of this disease. Assay performance characteristics have not been established for testing cord blood, neonate, prenatal screening, general population screening without symptoms of

{1}

meningoencephalitis or automated instruments. This assay is not FDA cleared or approved for testing blood or plasma donors

b) Indication(s) for use:
The West Nile Detect™ IgM Capture ELISA is for the laboratory diagnosis of West Nile Virus infection in patients with clinical symptoms consistent with meningoencephalitis.

c) Special condition for use statement(s):
The device is for prescription use only

d) Special instrument requirements:
NA

I. Device Description:
IgM Capture ELISA

J. Substantial Equivalence Information:

a) Predicate device name(s):
Focus Technologies West Nile Virus IgM Capture ELISA

b) Predicate K number(s):
K031952

Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Same indications for use. Same target population. Same ELISA methodology | West Nile Detect™ IgM Capture ELISA (K041817) | Focus West Nile Virus IgM Capture ELISA (K031952)  |
|   |  Test persons having symptoms of meningoencephalitis | Test persons having symptoms of meningoencephalitis  |
|   |  IgM Capture ELISA | IgM Capture ELISA  |
|  Differences  |   |   |
|  Item | Device | Predicate  |
|  Assay Procedure | West Nile Detect™ IgM Capture ELISA (K041817) | Focus West Nile Virus IgM Capture ELISA (K031952)  |
|   |  Does specimen testing in duplicate | Does specimen testing in singlet  |

{2}

# K. Standard/Guidance Document Referenced (if applicable):

Class II Special Controls Guidance Document: Serological Reagents for the Laboratory Diagnosis of West Nile Virus. October 30, 2003

# L. Test Principle:

The West Nile Detect™ IgM Capture ELISA consists of one enzymatically amplified "two-step" sandwich-type immunoassay. In this assay, controls and unknown serum samples are incubated in microtiter wells which have been coated with anti-human IgM antibodies, followed by incubation with West Nile Virus derived recombinant WNRA protein and a control preparation (NCA) separately. The serum samples may be directly mixed with sample dilution buffer for WN IgM added in the wells. After one hour incubation and washing, the wells are treated with a WNRA-specific antibody labeled with the enzyme horseradish peroxidase (HRP). After a second incubation and washing step, the wells are incubated with the tetramethylbenzidine (TMB) substrate. An acidic stopping solution is then added and the degree of enzymatic turnover of the substrate is determined by absorbance measurement at 450 nanometers. Above a certain threshold, the ratio of the absorbance of the WNRA and the control wells presumptively determines whether antibodies to WNV are present. A set of positive and negative controls is provided in order to monitor the integrity of the kit components.

# M. Performance Characteristics (if/when applicable):

# Analytical performance:

# a) Precision/Reproducibility:

The reproducibility of the West Nile Detect IgM Capture ELISA was evaluated at three sites. One site was InBios International. Ten serum specimens using clinical specimens diluted into an analyte-negative matrix was used. The ten serum specimens (not including positive and negative controls) included specimens that were below the cutoff values (negative samples) and above the cutoff value (positive and weak positive or borderline samples). The serum dilutions selected also ensured that the analyte concentration in the specimens represented a clinically relevant range. The results are shown in the table below.

Reproducibility Results from three sites after deleting 2 outlying data points from Site #3

|   |   |   | Intra-Assay |   | Between Day |   | Between Lab |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Sample ID | n | Mean | *S.D. | %CV | *S.D. | %CV | *S.D. | %CV | *S.D. | %CV  |
|  1 | 27 | 1.18 | 0.07 | 5.8% | 0.14 | 11.4% | 0.29 | 24.5% | 0.33 | 27.6%  |
|  2 | 27 | 9.39 | 1.04 | 11.0% | 3.10 | 33.1% | 1.46 | 15.5% | 3.58 | 38.2%  |
|  3 | 27 | 17.98 | 1.50 | 8.3% | 4.00 | 22.2% | 4.18 | 23.3% | 5.98 | 33.2%  |
|  4 | 27 | 6.48 | 1.04 | 16.1% | 1.98 | 30.5% | 2.47 | 38.2% | 3.33 | 51.4%  |
|  5 | 26* | 21.07 | 2.54 | 12.0% | 5.53 | 26.3% | 7.56 | 35.9% | 9.70 | 46.1%  |
|  6 | 27 | 7.92 | 0.74 | 9.4% | 2.33 | 29.5% | 3.46 | 43.7% | 4.24 | 53.5%  |
|  7 | 26* | 12.75 | 1.46 | 11.4% | 2.21 | 17.3% | 3.85 | 30.2% | 4.67 | 36.6%  |
|  8 | 27 | 5.94 | 0.64 | 10.8% | 1.62 | 27.3% | 1.87 | 31.5% | 2.56 | 43.1%  |

{3}

4

|  9 | 27 | 24.81 | 2.53 | 10.2% | 8.86 | 35.7% | 4.23 | 17.0% | 10.1  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  10 | 27 | 1.14 | 0.07 | 5.8% | 0.04 | 3.2% | 0.16 | 14.3% | 0.18  |

All values are calculated as WNRA/NCA ratios
SD = Standard Deviation; %CV = % Coefficient of Variation
26*: 1 statistically outlying (&gt;5.5 x Standard Deviation of previous run) data point was removed.

a. Linearity/assay reportable range:
NA

b. Traceability, Stability, Expected values (controls, calibrators, or method):
NA

c. Detection limit:
NA

d. Analytical specificity:
Two hundred and seventy-one sera that tested positive for other potentially cross-reactive pathogens were tested with the West Nile Detect IgM Capture ELISA test to determine the potential for cross-reactivity. The table below summarizes the results of this study.

|  Disease | Number of Samples | West Nile Detect IgM ELISA |   | Total of Positive and Equivocal  |
| --- | --- | --- | --- | --- |
|   |   |  Equivocal | Positive  |   |
|  Eastern Equine encephalitis | 17 | 0 | 0 | 0/17  |
|  Japanese encephalitis | 2 | 0 | 0 | 0/2  |
|  Saint Louis encephalitis | 32 | 1 | 16 | 17/32  |
|  La Crosse Virus | 6 | 0 | 0 | 0/6  |
|  Dengue virus | 7 | 0 | 2 | 2/7  |
|  Epstein-Barr virus | 15 | 0 | 0 | 0/15  |
|  Hepatitis A virus | 10 | 0 | 0 | 0/10  |
|  Hepatitis B virus | 49 | 0 | 0 | 0/42  |
|  Hepatitis C virus | 30 | 0 | 0 | 0/20  |
|  Herpes simplex virus | 32 | 0 | 0 | 0/32  |
|  California Encephalitis (CE) | 1 | 0 | 0 | 0/1  |
|  HIV | 20 | 0 | 0 | 0/20  |
|  Syphilis | 5 | 0 | 0 | 0/5  |
|  Cytomegalovirus | 12 | 0 | 0 | 0/12  |
|  Varicella zoster virus | 10 | 0 | 0 | 0/10  |
|  Coxsackievirus B 1-6 | 1 | 0 | 0 | 0/1  |
|  Echovirus 16 | 1 | 0 | 0 | 0/1  |
|  Measles | 1 | 0 | 0 | 0/1  |
|  Mumps | 1 | 0 | 0 | 0/1  |
|  Polio Blend | 1 | 0 | 0 | 0/1  |
|  Legionaries' disease | 3 | 0 | 0 | 0/3  |
|  Rheumatoid factor | 5 | 0 | 0 | 0/5  |

{4}

|  Anti-nuclear antibody | 10 | 0 | 0 | 0/10  |
| --- | --- | --- | --- | --- |
|  Total | 271 | 1 | 18 | 19/271  |

Caution: IgM assay cross-reactivity has been noted with some West Nile IgM assays testing specimens containing antibody to enteroviruses. Reactive results reported from children must contain a caution statement regarding possible cross-reactivity with enteroviruses.

# e. Assay cut-off:

The cut-off was selected using sera from an endemic population in the United States. The 282 samples consisted of 163 positive samples and 119 negative samples characterized by the CDC IgM Antibody Capture ELISA. The cutoff was determined by two-graph receiver operating characteristic analysis (TG-ROC).

f. Comparison studies:

a. Method comparison with predicate device:

The West Nile Detect™ IgM Capture ELISA was compared to two reference assays: The plaque-reduction neutralization test (PRNT) and the CDC IgM Antibody Capture ELISA

b. Matrix comparison: NA
c. Clinical sensitivity: NA
d. Clinical specificity: NA
e. Other clinical supportive data (when a and b are not applicable):

# PERFORMANCE CHARACTERISTICS

# a. Clinical Studies:

# Table 1 - Study Site 1

A clinical laboratory located in the mid-western U.S. tested 50 retrospective samples with clinically and laboratory confirmed cases of WNV  $(n = 50)$  or undetermined flavivirus (positive for both WNV and SLE;  $n = 2$ ). The samples were suspected to have come from

{5}

6

patients that had exhibited signs or symptoms of WN but specific clinical information could not be confirmed. In addition, 125 retrospective sequential endemic samples were tested. The sera were sequentially submitted to the laboratory, archived, and masked. Two were confirmed with undetermined flavivirus or WNV by PRNT.

|  Clinical Category | Positive | Negative | Equivocal | Total  |
| --- | --- | --- | --- | --- |
|  PRNT Positive | 50 | 2 | 0 | 52  |
|  Negative | 1 | 121 | 1 | 123  |
|  Total | 51 | 123 | 1 | 175  |

WN Virus Positive:
Serological Sensitivity = 50/52 = 96.2%  95% Confidence Interval: 87.0 – 98.9%

WN Virus Negative:
Serological Specificity = 121/123 = 98.4%  95% Confidence Interval: 94.3 - 99.6%

Table 2 – Study Site 2

A State Department of Health laboratory located in Midwestern U.S. tested 88 retrospective samples clinically and laboratory confirmed cases of WNV and/or SLE and confirmed by PRNT. Seven patient samples were suspected of having either viral encephalitis or viral meningitis. The remaining patient samples had signs or symptoms of WN fever and headache. In addition, 130 retrospective, sequential endemic samples were tested. The sera were sequentially submitted to the laboratory, archived, and masked. Fourteen (14) were confirmed with SLE and/or WNV by PRNT.

|  Clinical Category | Positive | Negative | Equivocal | Total  |
| --- | --- | --- | --- | --- |
|  PRNT Positive | 99 | 2 | 1 | 102  |
|  Negative | 1 | 115 | 0 | 116  |
|  Total | 100 | 117 | 1 | 218  |

WN Virus Positive:
Clinical Sensitivity = 99/102 = 97.1%  95% Confidence Interval: 91.7 – 99.0%

WN Virus Negative:
Clinical Specificity = 115/116 = 99.1%  95% Confidence Interval: 95.3 - 99.9%

Table 3 – Study Site 3

A state department of health laboratory located in Southeastern U.S. tested 150 retrospective samples clinically and laboratory confirmed cases of WNV by PRNT. In addition, 150 retrospective, sequential endemic samples were tested. The sera were sequentially submitted to the laboratory, archived, and masked. Twenty-three (23) were confirmed with SLE and/or WNV by CDC ELISA.

|  Clinical Category | Positive | Negative | Equivocal | Total  |
| --- | --- | --- | --- | --- |
|  PRNT Positive | 172 | 1 | 0 | 173  |
|  Negative | 0 | 127 | 0 | 127  |
|  Total | 172 | 128 | 0 | 300  |

{6}

7

WN Virus Positive:
Serological Sensitivity = 172/173 = 99.4%  95% Confidence Interval: 96.8 – 99.9%

WN Virus Negative:
Serological Specificity = 127/127 = 100.0%  95% Confidence Interval: 97.1 - 100%

## Study Site 4

A state department of health laboratory located in Northeastern U.S. tested 210 retrospective, sequential endemic samples with the West Nile Detect IgM Capture ELISA and with the CDC MAC ELISA. The sera were sequentially submitted to the laboratory, archived, and masked. None of the samples gave a positive result with both tests.

|  Clinical Category | Positive | Negative | Equivocal | Total  |
| --- | --- | --- | --- | --- |
|  CDC MAC ELISA Positive | 0 | 0 | 0 | 0  |
|  Negative | 0 | 210 | 0 | 210  |
|  Total | 0 | 210 | 0 | 210  |

Negative Presumptive Agreement
210/210 = 100.0%  95% Confidence Interval: 98.2 – 100.0%

b. Clinical cut-off:
NA

c. Expected values/Reference range:

## Expected Values

West Nile virus infection is generally recognized by the presence of IgM antibodies within one week from the beginning of symptoms. Detectable levels of IgM may be low in early infection. Two hundred samples prospectively collected from Florida, Texas and Pennsylvania during March 2004. The distribution of females was 50% (100/200) and males were 50% (100/200). The data in Table 1 illustrates the prevalence of IgM antibodies in different age groups when using the West Nile Detect IgM Capture ELISA Test.

Of the 200 normal sera, one was positive and one was equivocal. The latter specimen was repeated in duplicate and remained equivocal on the West Nile Detect IgM Capture ELISA Test. The positive and equivocal sera were from Pennsylvania. Of the 200 sera, 66 were from Pennsylvania, resulting in a 3.0% prevalence (2/66) in Pennsylvania.

Table 1

|  Age | Total | Equivocal | Positive | Prevalence  |
| --- | --- | --- | --- | --- |
|  10-20 | 12 | 0 | 0 | 0.0%  |
|  21-30 | 68 | 1 | 0 | 1.5%  |
|  31-40 | 63 | 0 | 0 | 0.0%  |
|  41-50 | 47 | 0 | 1 | 2.1%  |
|  51-60 | 10 | 0 | 0 | 0.0%  |
|  Total | 200 | 1 | 1 | 1.0%  |

{7}

8

N. Proposed Labeling:
The labeling is sufficient and satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted material in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/NOP/K041817](https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/NOP/K041817)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
