← Product Code [MXJ](/submissions/MI/subpart-d%E2%80%94serological-reagents/MXJ) · K071511

# PLEXUS HERPESELECT 1 AND 2 IGG (WITH PLEXUS TM SOFTWARE), MODEL MP0900G (K071511)

_Focus Diagnostics, Inc. · MXJ · Oct 5, 2007 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/MXJ/K071511

## Device Facts

- **Applicant:** Focus Diagnostics, Inc.
- **Product Code:** [MXJ](/submissions/MI/subpart-d%E2%80%94serological-reagents/MXJ.md)
- **Decision Date:** Oct 5, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3305
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

The Focus Diagnostics Plexus™ HerpeSelect® HSV 1 and 2 IgG (with software) is intended for qualitatively detecting the presence or absence of human IgG antibodies to HSV-1 and HSV-2 in human sera. The test is indicated for pregnant women and sexually active adults, as an aid for presumptively diagnosing HSV-1 and HSV-2 infection. The predictive value of a positive or negative result depends on the population's prevalence and the pretest likelihood of HSV-1 and HSV-2 infection. The test is not intended for donor screening or for self-testing. The performance of this assay has not been established for use in a pediatric population, for neonatal screening, for testing of immunocompromised patients, for use by a point of care facility or for use with automated equipment. The device is for prescription use only.

## Device Story

Plexus™ HerpeSelect® HSV 1 and 2 IgG is a multiplexed immunoassay for qualitative detection of human IgG antibodies to HSV-1 and HSV-2 in human sera. Input: patient serum samples. Principle: Antigen bead suspension containing distinct gG-1 and gG-2 beads; three-step procedure involves sample incubation, addition of phycoerythrin-conjugated goat anti-human IgG, and measurement of fluorescence from beads against a cut-off calibrator. Output: qualitative presence/absence of antibodies. Used in clinical laboratories; operated by laboratory technicians. Software automates index calculation from Luminex instrument raw data, replacing manual calculation. Healthcare providers use results as an aid for presumptive diagnosis of HSV infection. Benefits include efficient multiplexed testing for specific HSV types.

## Clinical Evidence

Clinical validation included testing 300 pregnant women and 300 sexually active adults across three sites. Sensitivity/specificity for pregnant women: HSV-1 (92.2%/96.5%), HSV-2 (95.5%/94.3%). For sexually active adults: HSV-1 (91.0%/96.5%), HSV-2 (96.3%/97.4%). CDC panel (n=100) showed 100% agreement. Low prevalence population (n=77) and cross-reactivity studies (n=51) were also performed. Reproducibility studies (inter-laboratory, inter-assay, intra-assay, inter-lot) showed %CV generally ≤10% for positives.

## Technological Characteristics

Multiplexed immunoassay using antigen-coated beads (gG-1, gG-2). Fluorescence-based detection via Luminex xMAP instrument. Software: Plexus™ Multi Analyte Diagnostic Software v1.0.0.2. Connectivity: supports web-based download of lot-specific factors. Sterilization: not applicable (reagents).

## Regulatory Identification

Herpes simplex virus serological assays are devices that consist of antigens and antisera used in various serological tests to identify antibodies to herpes simplex virus in serum. Additionally, some of the assays consist of herpes simplex virus antisera conjugated with a fluorescent dye (immunofluorescent assays) used to identify herpes simplex virus directly from clinical specimens or tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by herpes simplex viruses and provides epidemiological information on these diseases. Herpes simplex viral infections range from common and mild lesions of the skin and mucous membranes to a severe form of encephalitis (inflammation of the brain). Neonatal herpes virus infections range from a mild infection to a severe generalized disease with a fatal outcome.

## Special Controls

*Classification.* Class II (special controls). The device is classified as class II (special controls). The special control for the device is FDA's revised guidance document entitled “Class II Special Controls Guidance Document: Herpes Simplex Virus Types 1 and 2 Serological Assays.” For availability of the guidance revised document, see § 866.1(e).

## Predicate Devices

- Plexus™ HerpeSelect® 1 and 2 IgG ([K062543](/device/K062543.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

A. 510(k) Number:
K071511

B. Purpose for Submission:
The 510(k) holder would like to introduce into interstate commerce optional proprietary Plexus™ software for use with the currently marketed device and the Luminex xMAP® instrument and Luminex IS 2.3 software to automate some assay functions, including calculation of results, plate mapping and downloading lot-specific factors from the web.

C. Measurand:
IgG

D. Type of Test:
Enzyme immunoassay

E. Applicant:
Focus Diagnostics, Inc.

F. Proprietary and Established Names:
Plexus™ HerpeSelect® 1 and 2 IgG (with software)

G. Regulatory Information:
1. Regulation section:
21 CFR 866.3305
2. Classification:
Class II
3. Product code:
MXJ, MYF

{1}

4. Panel:

Immunology and Microbiology Devices

H. Intended Use:

1. Intended use(s):

The Focus Diagnostics Plexus™ HerpeSelect® HSV 1 and 2 IgG (with software) is intended for qualitatively detecting the presence or absence of human IgG antibodies to HSV-1 and HSV-2 in human sera. The test is indicated for pregnant women and sexually active adults, as an aid for presumptively diagnosing HSV-1 and HSV-2 infection. The predictive value of a positive or negative result depends on the population's prevalence and the pretest likelihood of HSV-1 and HSV-2 infection. The test is not intended for donor screening or for self-testing. The performance of this assay has not been established for use in a pediatric population, for neonatal screening, for testing of immunocompromised patients, for use by a point of care facility or for use with automated equipment.

2. Indication(s) for use:

Same as intended use.

The device is for prescription use only.

3. Special conditions for use statement(s):

Not applicable.

4. Special instrument requirements:

The assay is intended to be used with the Luminex xMAP instrument with Luminex IS 2.3 software or the Bio-Rad BioPlex® 200 System.

I. Device Description:

The Focus Diagnostics Plexus™ HerpeSelect® HSV 1 and 2 IgG (with software) is a multiplexed immunoassay for qualitatively detecting and differentiating human IgG antibodies to HSV-1 and HSV-2. The optional Plexus™ Multi Analyte Diagnostic Software version 1.0.0.2 automates the calculation of results using the Luminex xMAP reader with Luminex IS 2.3 software.

2

{2}

3

J. Substantial Equivalence Information:

1. Predicate device name(s):
Plexus™ HerpeSelect® 1 and 2 IgG

2. Predicate 510(k) number(s):
K062543

3. Comparison with predicate:
|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Assay, Calibrator and Controls | Same as predicate |   |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Optional Plexus™ Multi Analyte Diagnostic Software for automatic calculations of results for use with the Luminex xMAP instrument and Luminex IS 2.3 software | Optional | Manual calculations  |

K. Standard/Guidance Document Referenced (if applicable):

Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices – 05/11/2005.

Guidance for Off-the-Shelf Software Use in Medical Devices – 09/09/1999.

L. Test Principle:

The Focus Diagnostics HerpeSelect®1 and 2 Plexus IgG uses an Antigen Bead suspension that contains two distinct HSV antigen bead types that fluoresce at different wavelengths and/or intensities: gG-1 beads and gG-2 beads. The Focus Diagnostics HerpeSelect®1 and 2 Multiplex IgG is a three step procedure.

1. Patient sera are diluted, and the diluted sera are incubated with Antigen Beads. If HSV antibodies are present, then the antibodies bind to the corresponding antigen beads.
2. Phycoerythrin—conjugated goat anti-human IgG, (Conjugate) is added, and the

{3}

Conjugate binds to the bound HSV antibody (if present), and forms a Conjugate-HSV antibody-antigen bead sandwich.

3. Fluorescence from each distinct HSV antigen bead type is measured and compared against a Cut-off Calibrator.

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

a. Precision/Reproducibility:
Not applicable

b. Linearity/assay reportable range:
Not applicable

c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Not applicable

d. Detection limit:
Not applicable

e. Analytical specificity:
Not applicable

f. Assay cut-off:
Not applicable

### 2. Comparison studies:

a. Method comparison with predicate device:
Studies to Validate Plexus™ HerpeSelect® HSV 1 and 2 IgG (with software)

Testing was done at two external sites and one internal site to confirm that assay results obtained using Plexus Software are identical to results calculated manually using the instrument (Luminex) raw data.

4

{4}

5

|  Agreement between Index Calculated with Plexus Software and Index Calculated Manually (n = 600)  |   |   |
| --- | --- | --- |
|  Sample Type | n | % Agreement; 95% Confidence Interval  |
|  HSV-1 Pos | 161/161 | 100%; 95% CI: 0.977 – 1.00  |
|  HSV-1 Eqv | 4/4 | 100%; 95% CI: 0.398 – 1.00  |
|  HSV-1 Neg | 134/134 | 100%; 95% CI: 0.973 – 1.00  |
|  HSV-1 Invalid | 1/1 | -  |
|  HSV-2 Pos | 139/139 | 100%; 95% CI: 0.974 – 1.00  |
|  HSV-2 Eqv | 7/7 | 100%; 95% CI: 0.590 – 1.00  |
|  HSV-2 Neg | 153/153 | 100%; 95% CI: 0.976 – 1.00  |
|  HSV-1 Invalid | 1/1 | -  |

100% of HSV-1 indexes and 100% of HSV-2 indexes calculated using the Plexus software matched the indexes calculated manually.

b. Matrix comparison:
Not applicable

3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable

4. Clinical cut-off:
Not applicable

{5}

5. Expected values/Reference range:

Not applicable

N. Instrument Name:

The Plexus™ HerpeSelect® HSV 1 and 2 IgG (with software) device is intended to be used with the Luminex xMAP instrument or the Bio-Rad BioPlex® 200 System. The optional Plexus™ Multi Analyte Diagnostic Software version 1.0.0.2 is intended for use with the Luminex xMAP instrument and Luminex IS 2.3 software.

O. System Descriptions:

1. Modes of Operation:

The optional Plexus™ Multi Analyte Diagnostic Software version 1.0.0.2 is used with the Luminex xMAP instrument and the Luminex IS 2.3 software.

2. Software:

FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:

Yes ☐ X ☐ or No ☐

3. Specimen Identification:

Serum

4. Specimen Sampling and Handling:

Collect blood samples aseptically using approved venipuncture techniques by qualified personnel. Allow blood samples to clot at room temperature prior to centrifugation. Aseptically transfer serum to a tightly closing sterile container for storage. Separated serum should remain at 22°C for no longer than 8 hours. If the assay will not be completed within 8 hours, refrigerate the sample at 2 to 8°C. If the assay will not be completed within 48 hours, or for shipment of samples, freeze at –20°C or colder. Thaw and mix samples well prior to use.

5. Calibration:

IgG Cutoff Calibrator is supplied with the assay reagents.

6. Quality Control:

Positive and Negative IgG controls are supplied with the assay reagents.

6

{6}

P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:

Not applicable

Q. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

R. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

7

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/MXJ/K071511](https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/MXJ/K071511)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
