P150014 · Roche Molecular Systems, Inc. · MKT · Oct 14, 2015 · Microbiology
Device Facts
Record ID
P150014
Device Name
COBAS HBV TEST
Applicant
Roche Molecular Systems, Inc.
Product Code
MKT · Microbiology
Decision Date
Oct 14, 2015
Decision
APRL
Regulation
21 CFR 866.3174
Device Class
Class 2
Indications for Use
cobas® HBV is an in vitro nucleic acid amplification test for the quantitation of hepatitis B virus (HBV) DNA in human EDTA plasma or serum of HBV-infected individuals. This test is intended for use as an aid in the management of patients with chronic HBV infection undergoing anti-viral therapy. The test can be used to measure HBV DNA levels at baseline and during treatment to aid in assessing response to treatment. The results from cobas® HBV must be interpreted within the context of all relevant clinical and laboratory findings. The cobas® HBV is not intended for use as a screening test for the presence of HBV in blood or blood products or as a diagnostic test to confirm the presence of HBV infection.
Device Story
cobas® HBV is a quantitative in vitro nucleic acid amplification test performed on the cobas® 6800/8800 Systems. Input: human EDTA plasma or serum. Process: automated nucleic acid extraction and purification using magnetic glass particles; PCR amplification using target-specific primers and a thermostable DNA polymerase; detection via fluorescently labeled probes specific to HBV and a non-HBV DNA quantitation standard (DNA-QS). Output: HBV DNA viral load (IU/mL). Used in clinical laboratories by trained personnel. Results are reviewed by clinicians to monitor viral load, assess treatment response, and guide management of chronic HBV infection. Benefits include objective measurement of viral suppression, aiding in therapy adjustment and assessment of treatment efficacy.
Clinical Evidence
Clinical study evaluated 396 subjects with chronic HBV undergoing antiviral therapy. Primary endpoints included Virologic Response (VR), Biochemical Response (BR), and HBeAg loss at Week 96. Results showed baseline HBV DNA concentration and VR at Weeks 12, 24, and 48 were highly predictive of VR96 (PPVs 79.6%–100%). Analytical studies confirmed LoD of 6.6 IU/mL (plasma) and 3.5 IU/mL (serum), linearity from 10 to 1.00E+09 IU/mL, and precision/reproducibility across genotypes A-H.
Technological Characteristics
Quantitative real-time PCR assay. Components: cobas® 6800/8800 Systems, Assay Specific Analysis Package (ASAP) software, assay-specific reagents, and DNA-QS. Sensing: fluorescent reporter/quencher probe cleavage via 5' to 3' exonuclease activity. Connectivity: LIS/middleware/office PC interface. Sterilization: N/A (reagents). Software: automated data management and result interpretation.
Indications for Use
Indicated for quantitation of HBV DNA in human EDTA plasma or serum of patients with chronic HBV infection undergoing anti-viral therapy. Used to measure HBV DNA levels at baseline and during treatment to assess response. Not for screening blood/blood products or confirming HBV infection.
Regulatory Classification
Identification
A nucleic acid-based hepatitis B virus (HBV) assay is identified as an in vitro diagnostic device intended for prescription use in the detection of HBV nucleic acid in specimens from individuals with antibody evidence of HBV infection. In these devices, the detection of HBV nucleic acid is used as an aid in the management of HBV-infected individuals. The assay is intended for use with human serum or plasma (and other matrices as applicable) from individuals with HBV. The assay is not intended for use as a donor screening assay for the presence of HBV nucleic acids in blood, blood products, plasma, cells, or tissue donors, or as a diagnostic assay to confirm the presence of HBV infection.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Labeling required under § 809.10(b) of this chapter must include:
(i) A prominent statement that the assay is not intended for use as a screening assay for the presence of HBV DNA in blood or blood products, plasma, cells, or tissue donors, or as a diagnostic assay to confirm the presence of HBV infection.
(ii) A detailed explanation of the principles of operation and procedures for performing the assay.
(iii) A detailed explanation of the interpretation of results.
(iv) Limitations, which must be updated to reflect current clinical practice and disease presentation and/or management. These limitations must include statements that indicate:
(A) Management of patients undergoing HBV treatment should not be established on the basis of a single assay result but should be determined by a licensed healthcare professional in conjunction with the clinical presentation, history, and other diagnostic procedures,
*e.g.,* HBV serologic testing, liver function assays, liver elastography, etc.(B) The specimen types for which the device has been cleared, and that use of this assay with specimen types other than those specifically cleared for this device may result in inaccurate assay results.
(C) The results obtained with this assay may not be used interchangeably with results obtained with a different manufacturer's assay.
(2) Design verification and validation must include the following:
(i) Detailed device description, including the device components, ancillary reagents required but not provided, and an explanation of the device methodology. Additional information appropriate to the technology must be included such as design of primers and probes, rationale for the selected gene targets, specifications for amplicon size, and degree of nucleic acid sequence conservation.
(ii) For devices with assay calibrators, the design and composition of all primary, secondary, and subsequent quantitation standards used for calibration as well as their traceability to a standardized reference material that FDA has determined is appropriate (
*e.g.,* a recognized consensus standard). In addition, analytical testing must be performed following the release of a new lot of the standard material that was used for device clearance or approval, or when there is a transition to a new calibration standard.(iii) Documentation and characterization (
*e.g.,* determination of the identity, supplier, purity, and stability) of all critical reagents (including nucleic acid sequences for primers and probes) and protocols for maintaining product integrity.(iv) Risk analysis and management strategies demonstrating how risk control measures are implemented to address device system hazards, such as Failure Modes Effects Analysis and/or Hazard Analysis and Critical Control Points summaries and their impact on assay performance.
(v) Final release criteria to be used for manufactured assay lots with appropriate evidence that lots released at the extremes of the specification will meet the identified analytical and clinical performance characteristics as well as stability.
(vi) Stability studies for reagents must include documentation of an assessment of real-time stability for multiple reagent lots using the indicated specimen types and must use acceptance criteria that ensure that analytical and clinical performance characteristics are met when stability is assigned based on the extremes of the acceptance range.
(vii) All stability protocols, including acceptance criteria.
(viii) Detailed documentation of analytical performance studies conducted as appropriate to the technology, specimen types tested, and intended use of the device, including limit of detection (LoD), linearity, precision, endogenous and exogenous interferences, cross-reactivity, carryover, matrix equivalency, sample and reagents stability, and as applicable, upper and lower limits of quantitation (ULoQ and LLoQ, respectively). Samples selected for use must be from subjects with clinically relevant circulating genotypes in the United States. Cross-reactivity studies must include samples from HBV nucleic acid negative subjects with other viral or non-viral causes of liver disease, including autoimmune hepatitis, alcoholic liver disease, chronic hepatitis C virus, primary biliary cirrhosis, and nonalcoholic steatohepatitis, when applicable. The effect of each identified nucleic-acid isolation and purification procedure on detection must be evaluated.
(ix) Analytical sensitivity of the assay that is the same or better than that of other cleared or approved assays.
(x) For devices with associated software or instrumentation, documentation must include a detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review.
(xi) Detailed documentation of performance from a clinical study with a design and number of clinical samples (appropriately statistically powered) that is appropriate for the intended use of the device as well as conducted in the appropriate settings by the intended users. The samples must include prospective (sequential) samples for each claimed specimen type and, as appropriate, additional characterized clinical samples. Samples must be sourced from geographically diverse areas.
Reference Devices
COBAS TaqMan HBV Test for Use with the High Pure System (Roche)
COBAS AmpliPrep/COBAS TaqMan HBV Test, v2 (Roche)
Abbott RealTime HBV Test (Abbott)
Related Devices
P050028 — COBAS TAQMAN HBV TEST · Roche Molecular Systems, Inc. · Sep 4, 2008
K252484 — cobas HCV · Roche Molecular Systems, Inc. · Nov 3, 2025
P170025 — Aptima HBV Quant Assay · Hologic, Inc. · Jan 23, 2018
Submission Summary (Full Text)
{0}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)
I. GENERAL INFORMATION
Device Generic Name: Real-time PCR test
Device Trade Name: cobas® HBV
Device Procode: MKT
Applicant’s Name and Address: Roche Molecular Systems, Inc.
4300 Hacienda Drive
Pleasanton, CA 94588-2722
Date of Panel Recommendation: None
Premarket Approval Application (PMA) Number: P150014
Date of FDA Notice of Approval: October 14, 2015
II. INDICATIONS FOR USE
cobas® HBV is an in vitro nucleic acid amplification test for the quantitation of hepatitis B virus (HBV) DNA in human EDTA plasma or serum of HBV-infected individuals.
This test is intended for use as an aid in the management of patients with chronic HBV infection undergoing anti-viral therapy. The test can be used to measure HBV DNA levels at baseline and during treatment to aid in assessing response to treatment. The results from cobas® HBV must be interpreted within the context of all relevant clinical and laboratory findings.
The cobas® HBV is not intended for use as a screening test for the presence of HBV in blood or blood products or as a diagnostic test to confirm the presence of HBV infection.
III. CONTRAINDICATIONS
There are no known contraindications for use for this test.
IV. WARNINGS AND PRECAUTIONS
The warnings and precautions can be found in the cobas® HBV labeling.
PMA P150014: FDA Summary of Safety and Effectiveness Data
{1}
V. DEVICE DESCRIPTION
A. cobas® HBV test Description
The cobas® HBV is a quantitative test performed on the cobas® 6800 System and cob cobas® as 8800 System. The cobas® HBV allows for the detection and quantitation of HBV DNA in EDTA plasma or serum of infected patients for use in clinical laboratories for routine clinical practice in the management of patients with HBV. A single probe is used to detect and quantify, but not distinguish genotypes A-H. The viral load is quantified against a non-HBV DNA quantitation standard (DNA-QS), which is introduced into each specimen during sample preparation. The DNA-QS also functions to monitor the entire sample preparation and PCR amplification process. In addition, the test utilizes three external controls: a high titer positive, a low titer positive, and a negative control.
The cobas® HBV test system consists of:
- the cobas® 6800/8800 Systems
- the cobas® HBV Assay Specific Analysis Package (ASAP) software
- cobas® HBV in cassettes
- cobas® HBV/HCV/HIV-1 Control Kit (HPC and LPC) in cassettes
- cobas® NHP Negative Control Kit in cassettes
- Specimen preparation reagents (cobas® omni Reagents)
The cobas® HBV is based on automated sample preparation (nucleic acid extraction and purification) followed by PCR amplification and detection. The cobas® 6800/8800 Systems consist of the sample supply module, the transfer module, the processing module, and the analytic module. Automated data management is performed by the cobas® 6800/8800 Systems Software which assigns test results for all tests. Results can be reviewed directly on the system screen, exported, or printed as a report.
B. Principle of Procedure
1. Sample Preparation (Nucleic Acid Extraction and Purification)
Nucleic acid from patient samples, external controls and added DNA-QS molecules are simultaneously extracted. Viral nucleic acid is released by addition of proteinase and lysis reagent to the sample. The released nucleic acid binds to the silica surface of magnetic glass particles. Unbound substances and impurities, such as denatured proteins, cellular debris and potential PCR inhibitors are removed with subsequent wash reagent steps and purified nucleic acid is eluted from the magnetic glass particles with elution buffer at elevated temperature.
2. Nucleic Acid Amplification
Selective amplification of target nucleic acid from the sample is achieved by the use of target virus-specific forward and reverse primers which are selected from
PMA P150014: FDA Summary of Safety and Effectiveness Data
{2}
highly conserved regions of the HBV genome. Selective amplification of DNA-QS is achieved by the use of sequence-specific forward and reverse primers which are selected to have no homology with the HBV genome. A thermostable DNA polymerase enzyme is used for amplification.
The master mix includes deoxyuridine triphosphate (dUTP), instead of deoxythymidine triphosphate (dTTP), which is incorporated into the newly synthesized DNA (amplicon). Any contaminating amplicons from previous PCR runs are eliminated by the AmpErase enzyme, which is included in the PCR mix, during the first thermal cycling step. However, newly formed amplicons are not eliminated since the AmpErase enzyme is inactivated once exposed to temperatures above 55 °C.
## 3. Nucleic Acid Detection
The cobas® HBV master mix contains detection probes which are specific for the HBV target sequences and the DNA-QS nucleic acid, respectively. The specific HBV and DNA-QS detection probes are each labeled with one of two unique fluorescent dyes which act as reporters. Each probe also has a second dye which acts as a quencher. The two reporter dyes are measured at defined wavelengths, thus permitting simultaneous detection and discrimination of the amplified HBV target and the DNA-QS.
When not bound to the target sequence, the fluorescent signals of the intact probes are suppressed by a quencher dye. During the PCR amplification step, hybridization of the probes to the specific single-stranded DNA template results in cleavage of the probe by the 5' to 3' exonuclease activity of the DNA polymerase resulting in separation of the reporter and quencher dyes and the generation of a fluorescent signal. With each PCR cycle, increasing amounts of cleaved probe is generated and the cumulative signal of the reporter dye increases concomitantly. Since the two specific reporter dyes are measured at defined wavelengths, simultaneous detection and discrimination of the amplified HBV target and the DNA-QS are possible.
## 4. Controls
- Normal Human Plasma Negative Control - Normal human plasma, non-reactive by licensed tests for antibody to HCV, antibody to HIV-1/2, HBsAg, antibody to HBc; HIV-1 RNA, HIV-2 RNA, HCV RNA, HBV DNA, HEV RNA, WNV RNA, and CMV DNA not detectable by PCR methods.
- HBV/HCV/HIV-1 Low Positive Control - < 0.001% HIV-1 Group M RNA encapsulated in MS2 bacteriophage coat protein armored, < 0.001% synthetic plasmid) HBV DNA encapsulated in Lambda bacteriophage coat protein, < 0.001% synthetic (armored) HCV RNA encapsulated in MS2 bacteriophage coat protein, normal human plasma.
- HBV/HCV/HIV-1 High Positive Control - < 0.001% synthetic (armored) HIV-1 Group M RNA encapsulated in MS2 bacteriophage coat protein, < 0.001% synthetic (plasmid) HBV DNA encapsulated in Lambda
PMA P150014: FDA Summary of Safety and Effectiveness Data
Page 3
{3}
bacteriophage coat protein, < 0.001% synthetic (armored) HCV RNA encapsulated in MS2 bacteriophage coat protein, normal human plasma.
- Interpretation of Control Results: The cobas® 6800/8800 software determines whether control results are valid or invalid and therefore whether the run is valid or invalid.
## C. Instrumentation and Software
The cobas® 6800/8800 platform is configured as two instrument versions: the cobas® 6800 System, and the cobas® 8800 System.
Each system is comprised of a cobas® 6800 or cobas® 8800 instrument, system software, Assay Specific Analysis Packages (ASAP), and a sample source unit, which can be connected to a conveyor system for automated transport of samples to and from the system. The test kits consist of assay-specific reagents and omni reagents (common reagents) which can be used with any of the cobas® 6800/8800 assays, and on either the cobas® 6800 or the cobas® 8800 instrument.
In addition, the omni reagents and consumables, such as such as the P-plates, racks, AD-plates, waste bags, pipette tips, and secondary tubes, can be used by any of the cobas® 6800/8800 System assays, and on either the cobas® 6800 or the cobas® 8800 instrument.
Either system can be interfaced to an uninterruptible power supply (UPS), a Laboratory Information System (LIS), or middleware, and/or office PCs for remote viewing and messaging functionalities.
## D. Interpretation of Results
Table 1: Target Result Interpretation
| Results | Interpretation |
| --- | --- |
| Target Not Detected | HBV DNA not detected.
Report results as “HBV not detected.” |
| < Titer Min | Calculated titer is below the Lower Limit of Quantitation (LLoQ) of the assay.
Report results as “HBV detected, less than (Titer Min).”
Titer min = 10 IU/mL |
| Titer | Calculated titer is within the Linear Range of the assay – greater than or equal to Titer Min and less than or equal to Titer Max.
Report results as “(Titer) of HBV detected”. |
| > Titer Maxa | Calculated titer is above the Upper Limit of Quantitation (ULoQ) of the assay.
Report results as “HBV detected, greater than (Titer Max).”
Titer max = 1.00E+09 IU/mL |
a Sample result > Titer Max refers to HBV positive samples detected with titers above the upper limit of quantitation (ULoQ). If a quantitative result is desired, the original sample should be diluted with HBV-negative EDTA plasma or serum, depending on the type of the original sample, and the test should be repeated. Multiply the reported result by the dilution factor.
PMA P150014: FDA Summary of Safety and Effectiveness Data
Page 4
{4}
VI. ALTERNATIVE PRACTICES AND PROCEDURES
There are currently three (3) FDA approved alternative in vitro nucleic acid amplification tests for the quantitation of HBV:
- COBAS TaqMan HBV Test for Use with the High Pure System (Roche)
- COBAS AmpliPrep/COBAS TaqMan HBV Test, v2 (Roche)
- Abbott RealTime HBV Test (Abbott)
Each alternative has its own advantages and disadvantages. The patient and physician should fully discuss alternatives to select the best method.
VII. MARKETING HISTORY
The test is currently available in the following countries listed in Table 2. To date, there have been no adverse incidents or potentially-critical complaints reported.
Table 2: Marketed Countries
| Austria | Greece | Norway |
| --- | --- | --- |
| Belgium | Hungary | Poland |
| Bulgaria | Iceland | Portugal |
| Croatia | Ireland | Romania |
| Cyprus | Italy | Slovakia |
| Czech Republic | Latvia | Slovenia |
| Denmark | Liechtenstein | Spain |
| Estonia | Lithuania | Sweden |
| Finland | Luxembourg | Switzerland |
| France | Malta | Turkey |
| Germany | Netherlands | United Kingdom |
VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
Below is a list of the potential adverse effects (e.g., complications) associated with the use of the device.
The primary adverse events from use of the device are caused by an inaccurate measurement result of HBV DNA viral load. Failure of the cobas® HBV to perform as indicated, or human error in the use of the test or the interpretation of the test result, may result in an incorrect test result that is too low or too high.
An erroneous low test result or a false negative result may lead to inappropriate patient management decisions, a premature discontinuation of antiviral therapy, or may instill a false sense of security in a patient or clinician.
PMA P150014: FDA Summary of Safety and Effectiveness Data
{5}
An erroneous high test result, or a false positive result, may contribute to a change in therapy, unnecessary treatment, prolonged duration of therapy, or create anxiety in the patient.
Inaccurate results would be most significant between 2,000 – 20,000 IU/mL, where treatment may be considered under different scenarios; however, serial measurements are common, and clinicians are likely to repeat measurements when there is consideration of treatment but a deciding factor is a low HBV value. The risks from ‘overtreatment’ (i.e., treatment initiated due to an unexpectedly high HBV viral load) are relatively small absent the cost of medication.
## IX. SUMMARY OF PRE-CLINICAL STUDIES
### A. Traceability to the WHO International Standard
Several standards and controls have been used during development of this test to provide traceability to the WHO standard (2nd WHO International Standard for Hepatitis B Virus DNA for Nucleic Acid Amplification Techniques (NIBSC 97/750). The standards used during development of the test include the HBV WHO Standard, the RMS HBV Secondary Standard, and the RMS HBV Calibration Panel. The Standards and the Calibration Panel were tested. The concentration range tested for the HBV WHO Standard was from 2.00E+01 IU/mL to 2.00E+03 IU/mL (1.30 - 3.30 log₁₀ IU/mL), the RMS HBV Secondary Standard was tested at 3.89E+02 IU/mL (2.59 log₁₀ IU/mL), and the RMS HBV Calibration Panel was tested from 5.00E+01 to 2.00E+07 IU/mL (1.70 - 7.30 log₁₀ IU/mL).
All materials performed similarly and demonstrated co-linear dilution performance across the linear range of cobas® HBV (Figure 1). Based on these results, the calibration and standardization process of cobas® HBV provides quantitation values for the calibration panel, the RMS HBV Secondary Standard, and the HBV WHO Standard that are similar to the expected values with deviation of not more than 0.14 log₁₀ IU/mL. The maximum deviation was obtained around the test LLoQ using combined regression analyses for the RMS HBV Calibration Panel, the RMS HBV Secondary Standard and the HBV WHO Standard.
PMA P150014: FDA Summary of Safety and Effectiveness Data
Page 6
{6}

Figure 1: Traceability to WHO International Standard (mean observed $\log_{10}$ titer versus $\log_{10}$ WHO standard based titer) Using cobas® HBV
# B. Analytical Sensitivity - Limit of Detection (LOD)
The limit of detection (LoD) of cobas® HBV was determined with the WHO International Standard (genotype A), genotypes B through H, and the predominant precore mutant. The LoD was determined to be 6.6 IU/mL for EDTA plasma and 3.5 IU/mL for serum by PROBIT analysis. The same LoD was verified at the same level for genotypes B through H and for the predominant precore mutant.
# WHO International Standard
The limit of detection of cobas® HBV was determined by analysis of serial dilutions of the WHO International Standard for Hepatitis B Virus DNA for Nucleic Acid Amplification Technology Assays (2nd WHO International Standard) genotype A obtained from NIBSC, in HBV-negative human EDTA plasma and serum. Panels of eight concentration levels plus a negative were tested over three lots of cobas® HBV test reagents, multiple runs, days, operators, and instruments.
The study demonstrates that cobas® HBV detected HBV DNA for the WHO International Standard at a concentration of 2.7 IU/mL for LoD by PROBIT with a hit rate of $95\%$ in EDTA plasma. For serum, the study demonstrates that cobas® HBV detected HBV DNA at a concentration of 2.4 IU/mL for LoD by PROBIT with a hit rate of $95\%$ .
# C. Analytical Sensitivity - Limit of Detection (LOD) for Genotypes
The limit of detection of cobas® HBV was determined by analysis of serial dilutions for seven different genotypes (B, C, D, E, F, G, H) and the predominant precore mutant (G1896A; C1858T). Dilution of clinical specimens or contrived plasmids were made in HBV-negative human EDTA plasma and serum. Panels of eight
PMA P150014: FDA Summary of Safety and Effectiveness Data
{7}
concentration levels plus a negative were tested using three lots of cobas® HBV test reagents, over multiple runs, days, operators, and instruments.
The results for EDTA plasma and serum are shown in Table 3 and Table 4, respectively. The study demonstrates that cobas® HBV detected all HBV genotypes tested with a similar LoD as that for HBV genotype A.
Table 3: LoD for HBV DNA genotypes B through H and the Predominant Precore Mutant in EDTA plasma (500 μL)
| Genotype | 95% LoD by PROBIT | 95% Confidence Interval |
| --- | --- | --- |
| GT B | 3.5 IU/mL | 2.95 IU/mL - 4.32 IU/mL |
| GT C | 4.1 IU/mL | 3.32 IU/mL - 5.82 IU/mL |
| GT D | 4.5 IU/mL | 3.59 IU/mL - 6.49 IU/mL |
| GT E | 3.2 IU/mL | 2.76 IU/mL - 3.98 IU/mL |
| GT F | 1.9 IU/mL | 1.66 IU/mL - 2.24 IU/mL |
| GT G | 2.5 IU/mL | 2.17 IU/mL - 3.02 IU/mL |
| GT H | 6.6 IU/mL | 5.33 IU/mL - 8.77 IU/mL |
| precore mutant | 2.4 IU/mL | 2.08 IU/mL - 2.90 IU/mL |
Table 4: LoD for HBV DNA Genotypes B through H and the Predominant Precore in Serum (500 μL)
| Genotype | 95% LoD by PROBIT | 95% Confidence Interval |
| --- | --- | --- |
| GT B | 3.3 IU/mL | 2.76 IU/mL - 4.30 IU/mL |
| GT C | 3.3 IU/mL | 2.83 IU/mL - 4.23 IU/mL |
| GT D | 2.6 IU/mL | 2.17 IU/mL - 3.42 IU/mL |
| GT E | 2.7 IU/mL | 2.25 IU/mL - 3.49 IU/mL |
| GT F | 2.0 IU/mL | 1.72 IU/mL - 2.45 IU/mL |
| GT G | 2.1 IU/mL | 1.75 IU/mL - 2.66 IU/mL |
| GT H | 3.5 IU/mL | 2.89 IU/mL - 4.60 IU/mL |
| precore mutant | 1.6 IU/mL | 1.43 IU/mL - 2.03 IU/mL |
## D. Linear Range & Lower Limit of Quantitation
The linearity study for cobas® HBV was performed with a dilution series consisting of 15 panel members spanning the intended linear range for the predominant genotype (GT A). High titer panel members were prepared from a high titer HBV plasmid DNA stock whereas the lower titer panel members were prepared from a clinical sample. The linearity panel was designed to have an approximate $2\log_{10}$ titer overlap between the two material sources. The linearity panel was designed to range from one concentration below LLoQ (e.g., 7.5 IU/mL) to one concentration level
PMA P150014: FDA Summary of Safety and Effectiveness Data
Page 8
{8}
above ULoQ (e.g., $2.0\mathrm{E} + 09$ IU/mL) and to include medical decision points. Moreover, the linearity panel was designed to partly support steps of $1.0\log_{10}$ throughout the linear range. For each panel member the nominal concentration in $\mathrm{IU / mL}$ and the source of the HBV DNA were given. The data were analyzed and the best fitting polynomial regression fit and the LLoQ were calculated.
The cobas® HBV is linear for EDTA plasma and serum from 10 IU/mL to 1.00E+09 IU/mL and shows an absolute deviation from the better fitting non-linear regression of less than $\pm 0.2\log_{10}$ . Across the linear range, the accuracy of the test was within $\pm 0.24\log_{10}$ .
See Figure 2 and Figure 3 for representative results.

Figure 2: Linear Range Determination in EDTA Plasma

Figure 3: Linear Range Determination in Serum
PMA P150014: FDA Summary of Safety and Effectiveness Data
{9}
# E. Linear Range - Genotypes
The dilution series used to verify linearity of cobas® HBV for genotypes consisted of 10 panel members spanning the intended linear range. High titer panel members were prepared from a high titer plasmid DNA stock whereas the lower titer panel members were made from a high titer clinical sample. The linearity panel was designed to have an approximate $2\log_{10}$ titer overlap between the two material sources. The linear range of cobas® HBV spanned from below the LLoQ (10 IU/mL) to the ULoQ (1.00E+09 IU/mL) and included at least one medical decision point. Twenty-one replicates were tested across three lots of cobas® HBV reagent for each level in EDTA plasma and serum.
The linearity within the linear range of cobas® HBV was verified for all seven genotypes (B, C, D, E, F, G, H) and predominant precore mutant (G1896A; C1858T). The maximum deviation between the linear regression and the better fitting nonlinear regression was equal to or less than $\pm 0.2\log_{10}$ .
# F. Precision - within Laboratory
Precision of cobas® HBV was determined by analysis of serial dilutions of clinical HBV (Genotype A) samples (CS) or of HBV plasmid DNA in HBV negative EDTA plasma or in serum. Ten to 12 dilution levels were tested in 48 replicates for each level across three lots of cobas® HBV test reagents using three instruments and three operators over 12 days. Each sample was carried through the entire cobas® HBV test procedure on the cobas® 6800/8800 Systems. The results are shown in Table 5 and Table 6.
Table 5: Within-Laboratory Precision of cobas® HBV (EDTA Plasma Samples)*
| Nominal Concentration (IU/mL) | Assigned Concentration (IU/mL) | Source Material | EDTA Plasma | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | Lot 1 | Lot 2 | Lot 3 | All Lots |
| | | | SD | SD | SD | Pooled SD |
| 1.00E+09 | 9.32E+08 | plasmid DNA | 0.04 | 0.07 | 0.09 | 0.07 |
| 1.00E+08 | 9.32E+07 | plasmid DNA | 0.04 | 0.08 | 0.05 | 0.06 |
| 1.00E+07 | 9.32E+06 | plasmid DNA | 0.06 | 0.05 | 0.04 | 0.05 |
| 1.00E+06 | 9.32E+05 | plasmid DNA | 0.06 | 0.07 | 0.04 | 0.06 |
| 1.00E+05 | 9.32E+04 | plasmid DNA | 0.06 | 0.06 | 0.07 | 0.06 |
| 2.00E+04 | 1.71E+04 | clinical specimen | 0.05 | 0.03 | 0.03 | 0.04 |
| 2.00E+03 | 1.86E+03 | plasmid DNA | 0.05 | 0.04 | 0.07 | 0.05 |
| 1.00E+03 | 8.54E+02 | clinical specimen | 0.04 | 0.05 | 0.04 | 0.04 |
| 1.00E+03 | 9.32E+02 | plasmid DNA | 0.06 | 0.06 | 0.05 | 0.06 |
| 1.00E+02 | 8.54E+01 | clinical specimen | 0.07 | 0.08 | 0.07 | 0.07 |
| 1.00E+02 | 9.32E+01 | plasmid DNA | 0.10 | 0.08 | 0.09 | 0.09 |
| 5.00E+01 | 4.27E+01 | clinical specimen | 0.09 | 0.04 | 0.08 | 0.08 |
* Titer data are considered to be log-normally distributed and are analyzed following $\log_{10}$ transformation. Standard deviations (SD) columns present the total of the log-transformed titer for each of the three reagent lots.
PMA P150014: FDA Summary of Safety and Effectiveness Data
{10}
Table 6: Within-Laboratory Precision of cobas® HBV (Serum Samples)*
| Nominal Concentration (IU/mL) | Assigned Concentration (IU/mL) | Source Material | Serum | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | Lot 1 | Lot 2 | Lot 3 | All Lots |
| | | | SD | SD | SD | Pooled SD |
| 1.00E+09 | 5.47E+08 | plasmid DNA | 0.05 | 0.06 | 0.03 | 0.05 |
| 1.00E+08 | 5.47E+07 | plasmid DNA | 0.03 | 0.04 | 0.03 | 0.04 |
| 1.00E+07 | 5.47E+06 | plasmid DNA | 0.05 | 0.05 | 0.03 | 0.05 |
| 1.00E+06 | 5.47E+05 | plasmid DNA | 0.04 | 0.06 | 0.06 | 0.05 |
| 1.00E+05 | 5.47E+04 | plasmid DNA | 0.04 | 0.03 | 0.03 | 0.04 |
| 2.00E+04 | 1.12E+04 | clinical specimen | 0.10 | 0.07 | 0.08 | 0.08 |
| 2.00E+03 | 1.09E+03 | plasmid DNA | 0.05 | 0.05 | 0.03 | 0.05 |
| 1.00E+03 | 5.62E+02 | clinical specimen | 0.03 | 0.14 | 0.03 | 0.09 |
| 1.00E+03 | 5.47E+02 | plasmid DNA | 0.04 | 0.05 | 0.04 | 0.04 |
| 1.00E+02 | 5.62E+01 | clinical specimen | 0.09 | 0.06 | 0.07 | 0.07 |
| 1.00E+02 | 5.47E+01 | plasmid DNA | 0.05 | 0.07 | 0.04 | 0.06 |
| 5.00E+01 | 2.81E+01 | clinical specimen | 0.07 | 0.06 | 0.10 | 0.08 |
* Titer data are considered to be log-normally distributed and are analyzed following $\log_{10}$ transformation. Standard deviations (SD) columns present the total of the log-transformed titer for each of the three reagent lots.
# G. Multi-Site Reproducibility Studies
The reproducibility and lot-to-lot variability of the cobas® HBV were evaluated in separate studies with EDTA plasma on the cobas® 6800 System using a mixed model to estimate the total variance.
# Lot-to-Lot Variability
Lot-to-lot variability testing was performed for genotypes A and C at 1 test site, using three reagent lots. Two operators at the site tested each lot for 6 days. Two runs were performed each day.
Table 9 below shows attributable percentages of total variance, total precision SDs, and lognormal CVs by genotype and expected $\log_{10}$ HBV DNA concentration for the cobas® 6800 System.
Table 9: Summary of Lot-to-Lot Variability Study on the cobas® 6800 System
| | HBV DNA Concentration (log10 IU/mL) | | | Percent Contribution to Total Variance (Lognormal CV(%) | | | | | Total Precision | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Genotype | Expected | Observed Meana | No. of Testsb | Lot | Operator | Day | Run | Within-Run | SDc | Log-normal CV(%)d |
| A | 1.48 | 1.50 | 107 | 13% | 0% | 0% | 0% | 87% | 0.157 | 37.27 |
PMA P150014: FDA Summary of Safety and Effectiveness Data
{11}
| | HBV DNA Concentration (log10 IU/mL) | | | Percent Contribution to Total Variance (Lognormal CV(%) | | | | | Total Precision | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Geno-type | Expected | Observed Meana | No. of Testsb | Lot | Operator | Day | Run | Within-Run | SDc | Log-normal CV(%)d |
| | | | | (12.90) | (0.00) | (0.00) | (0.00) | (34.68) | | |
| | 2.70 | 2.72 | 108 | 52% (11.96) | 0% (0.00) | 0% (0.00) | 0% (0.00) | 48% (11.56) | 0.072 | 16.69 |
| | 3.70 | 3.64 | 108 | 60% (14.29) | 0% (0.00) | 4% (3.55) | 1% (1.57) | 36% (11.01) | 0.080 | 18.53 |
| | 4.70 | 4.65 | 107 | 47% (13.05) | 0% (0.00) | 3% (3.22) | 1% (2.32) | 49% (13.29) | 0.082 | 19.14 |
| | 5.70 | 5.67 | 107 | 53% (13.66) | 2% (2.59) | 0% (0.00) | 0% (0.00) | 45% (12.54) | 0.081 | 18.80 |
| | 6.70 | 6.71 | 105 | 50% (11.66) | 0% (0.00) | 0% (0.00) | 5% (3.82) | 44% (10.92) | 0.071 | 16.48 |
| | 7.70 | 7.41 | 108 | 55% (13.08) | 0% (0.00) | 0% (0.00) | 4% (3.59) | 40% (11.18) | 0.076 | 17.65 |
| | 8.70 | 8.41 | 107 | 51% (12.52) | 0% (0.00) | 0% (0.00) | 10% (5.61) | 38% (10.75) | 0.075 | 17.51 |
| C | 1.48 | 1.49 | 107 | 23% (13.62) | 1% (2.83) | 0% (0.00) | 0% (0.00) | 76% (25.26) | 0.124 | 29.05 |
| | 2.70 | 2.71 | 105 | 53% (13.92) | 2% (2.63) | 3% (3.48) | 0% (0.00) | 41% (12.27) | 0.082 | 19.16 |
| | 3.70 | 3.64 | 107 | 61% (11.67) | 0% (0.00) | 0% (0.80) | 0% (0.00) | 39% (9.37) | 0.065 | 15.02 |
| | 4.70 | 4.65 | 106 | 47% (11.44) | 0% (0.00) | 0% (0.00) | 0% (0.00) | 53% (12.25) | 0.073 | 16.82 |
| | 5.70 | 5.69 | 107 | 60% (14.76) | 0% (0.00) | 1% (1.51) | 0% (0.00) | 39% (11.86) | 0.082 | 19.08 |
| | 6.70 | 6.69 | 107 | 48% (11.79) | 0% (0.00) | 2% (2.31) | 0% (0.00) | 50% (12.13) | 0.074 | 17.14 |
| | 7.70 | 7.38 | 107 | 51% (11.22) | 0% (0.00) | 0% (0.00) | 1% (1.57) | 48% (10.94) | 0.068 | 15.80 |
| | 8.70 | 8.42 | 106 | 56% (13.92) | 0% (0.00) | 0% (0.00) | 4% (3.54) | 40% (11.72) | 0.080 | 18.62 |
Note: Results with detectable viral load are included in this table; Within the range of assay are results from
$1.0\mathrm{E} + 01$ IU/mL to $1.0\mathrm{E} + 09$ IU/mL.
a Calculated using the SAS MIXED procedure.
b Number of valid tests with detectable viral load.
c Calculated using the total variability from the SAS MIXED procedure.
d Lognormal $\mathrm{CV}(\%) = \mathrm{sqrt}(10^{\wedge}[\mathrm{SD}^{\wedge}2*\ln (10)] - 1)*100$
$\mathrm{CV}(\%)$ = percent coefficient of variation; DNA = deoxyribonucleic acid; HBV = hepatitis B virus; No. = number;
$\mathrm{SD} =$ standard deviation; sqrt $=$ square root.
PMA P150014: FDA Summary of Safety and Effectiveness Data
{12}
In Table 10 below, the negative percent agreement (NPA) for the cobas® 6800 System using negative panel member results was 100%.
Table 10: Negative Percent Agreement Using the Negative Panel Member (Lot-to-Lot)
| Expected HBV DNA Concentration | No. of Valid Tests | Positive Results | Negative Results | Negative Percent Agreement^{a} | 95% CI^{b} |
| --- | --- | --- | --- | --- | --- |
| Negative | 106 | 0 | 106 | 100.00 | (96.58, 100.00) |
a NPA = (number of negative results / total number of valid tests in negative panel member) * 100.
b Calculated using the Clopper-Pearson exact binomial confidence interval method.
CI = confidence interval; DNA = deoxyribonucleic acid; HBV = hepatitis B virus; No. = number; NPA = negative percent agreement.
## Reproducibility
Reproducibility testing was performed at three sites for genotypes A and C, using one reagent lot. Two operators at each site tested for 6 days. Two runs were performed each day.
Table 11 below shows attributable percentages of total variance, total precision SDs, and lognormal CVs by genotype and expected log₁₀ HBV DNA concentration on the cobas® 6800 System.
Table 11: Summary of Reproducibility Study on the cobas® 6800
| | HBV DNA Concentration (log_{10} IU/mL) | | | Percent Contribution to Total Variance (Lognormal CV(%) | | | | | Total Precision | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Genotype | Expected | Observed Mean^{a} | No. of Tests^{b} | Site | Operator | Day | Run | Within-Run | SD^{c} | Log-normal CV(%)^{d} |
| A | 1.48 | 1.48 | 107 | 1% (4.21) | 0% (0.00) | 5% (7.75) | 1% (3.56) | 93% (34.98) | 0.153 | 36.41 |
| | 2.70 | 2.66 | 108 | 34% (9.53) | 0% (0.00) | 0% (0.00) | 16% (6.40) | 50% (11.52) | 0.070 | 16.33 |
| | 3.70 | 3.60 | 108 | 34% (7.49) | 2% (1.90) | 7% (3.42) | 0% (0.00) | 56% (9.58) | 0.055 | 12.80 |
| | 4.70 | 4.62 | 107 | 13% (5.40) | 0% (0.00) | 0% (0.00) | 12% (5.28) | 75% (13.05) | 0.065 | 15.12 |
| | 5.70 | 5.63 | 107 | 37% (7.82) | 1% (1.26) | 0% (0.00) | 0% (0.00) | 62% (10.04) | 0.055 | 12.81 |
| | 6.70 | 6.67 | 106 | 20% (5.99) | 3% (2.16) | 4% (2.57) | 15% (5.16) | 60% (10.48) | 0.059 | 13.59 |
| | 7.70 | 7.37 | 108 | 3% (2.70) | 2% (2.06) | 0% (0.00) | 0% (0.00) | 95% (15.12) | 0.067 | 15.50 |
| | 8.70 | 8.36 | 107 | 12% (4.32) | 0% (0.00) | 0% (0.00) | 2% (1.53) | 86% (11.46) | 0.053 | 12.36 |
PMA P150014: FDA Summary of Safety and Effectiveness Data
Page 13
{13}
PMA P150014: FDA Summary of Safety and Effectiveness Data
Page 14
| | HBV DNA Concentration (log10 IU/mL) | | | Percent Contribution to Total Variance (Lognormal CV(%)) | | | | | Total Precision | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Genotype | Expected | Observed Mean^{a} | No. of Tests^{b} | Site | Operator | Day | Run | Within-Run | SD^{c} | Log-normal CV(%)^{d} |
| C | 1.48 | 1.48 | 107 | 2% (11.79) | 1% (7.06) | 0% (0.00) | 0% (0.00) | 97% (84.30) | 0.324 | 86.20 |
| | 2.70 | 2.67 | 105 | 19% (5.94) | 3% (2.22) | 0% (0.00) | 0% (0.00) | 79% (12.27) | 0.060 | 13.84 |
| | 3.70 | 3.61 | 107 | 14% (4.49) | 0% (0.00) | 7% (3.15) | 0% (0.00) | 78% (10.48) | 0.051 | 11.84 |
| | 4.70 | 4.62 | 106 | 24% (6.45) | 0% (0.00) | 0% (0.00) | 0% (0.00) | 76% (11.59) | 0.057 | 13.29 |
| | 5.70 | 5.65 | 107 | 18% (5.96) | 0% (0.00) | 3% (2.29) | 0% (0.00) | 80% (12.68) | 0.061 | 14.22 |
| | 6.70 | 6.65 | 107 | 23% (6.35) | 6% (3.26) | 0% (0.00) | 1% (1.33) | 70% (11.10) | 0.057 | 13.29 |
| | 7.70 | 7.34 | 106 | 0% (0.00) | 3% (2.38) | 0% (0.00) | 13% (5.12) | 84% (13.11) | 0.062 | 14.30 |
| | 8.70 | 8.36 | 107 | 4% (2.24) | 0% (0.00) | 16% (4.35) | 10% (3.46) | 70% (9.09) | 0.047 | 10.91 |
Note: Results with detectable viral load are included in this table; Within the range of assay are results from 1.0E+0 1 IU/mL to 1.0E+09 IU/mL.
a Calculated using the SAS MIXED procedure.
b Number of valid tests with detectable viral load.
c Calculated using the total variability from the SAS MIXED procedure.
d Lognormal CV(%) = sqrt(10^[SD^2 * ln(10)] - 1) * 100.
CV(%) = percent coefficient of variation; HBV = hepatitis B virus; DNA = deoxyribonucleic acid; No. = number; SD = standard deviation; sqrt = square root.
The NPA was 100% (106/106; 95% CI: 96.58% to 100%) using negative panel member tests on the cobas® 6800 System as presented in Table 12 below.
Table 12: Negative Percent Agreement Using the Negative Panel Member (Reproducibility) on the cobas® 6800 System
| Expected HBV DNA Concentration | No. of Tests | Positive Results | Negative Results | Negative Percent Agreement^{a} | 95% CI^{b} |
| --- | --- | --- | --- | --- | --- |
| Negative | 106 | 0 | 106 | 100.00 | (96.58, 100.00) |
a NPA = (number of negative results/total number of valid tests in negative panel member) * 100.
b Calculated using the Clopper-Pearson exact binomial confidence interval method.
CI = confidence interval; DNA = deoxyribonucleic acid; HBV = hepatitis B virus; No. = number; NPA = negative percent agreement.
{14}
# H. Performance with HBV DNA-Negative Samples
The performance of the cobas® HBV assay with HBV DNA-negative samples was determined by analyzing HBV negative EDTA plasma and serum samples from individual donors. Three hundred individual EDTA plasma and 300 individual serum samples (600 total results) were tested with two lots of cobas® HBV reagents. All samples tested negative for HBV DNA. In the test panel, the results of all specimens tested with cobas® HBV was $100\%$ "Target Not Detected" (with a two-sided $95\%$ confidence interval of $99.4\% - 100\%$ ).
# I. Analytical Specificity
The analytical specificity of cobas® HBV was evaluated by diluting a panel of microorganisms with HBV DNA positive and HBV DNA negative EDTA plasma (Table 7). The microorganisms were added to negative human EDTA plasma and tested with and without HBV DNA. None of the non-HBV pathogens interfered with test performance. Negative results were obtained with cobas® HBV for all microorganism samples without HBV target and positive results were obtained on all of the microorganism samples with HBV target. Furthermore, the mean $\log_{10}$ titer of each of the positive HBV samples containing potentially cross-reacting organisms was within $\pm 0.3\log_{10}$ of the mean $\log_{10}$ titer of the respective positive spike control.
Table 7: Microorganisms Tested for Cross-Reactivity
| Viruses | | Bacteria | Yeast |
| --- | --- | --- | --- |
| Adenovirus type 5 | West Nile Virus | Propionibacterium Acnes | Candida albicans |
| Cytomegalovirus | St. Louis encephalitis Virus | Staphylococcus aureus | |
| Hepatitis A Virus | Dengue virus types 1, 2, 3, and 4 | | |
| Hepatitis C Virus | FSME virus (strain HYPR) | | |
| Hepatitis D Virus | Yellow Fever Virus | | |
| Human Immunodeficiency Virus-1 | Human Papillomavirus | | |
| Human T-Cell Lymphotropic Virus types 1 and 2 | Varizella-Zoster Virus | | |
| Human Herpes Virus Type-6 | Influenza A | | |
| Herpes Simplex Virus Type-1and 2 | Zika Virus | | |
# J. Analytical Specificity - Interfering Substances
EDTA-plasma samples with and without HBV DNA were spiked with elevated levels of triglycerides $(34.5\mathrm{g / L})$ , conjugated bilirubin $(0.25\mathrm{g / L})$ , unconjugated bilirubin $(0.25\mathrm{g / L})$ , albumin $(58.7\mathrm{g / L})$ , hemoglobin $(2.9\mathrm{g / L})$ and human DNA $(2\mathrm{mg / L})$ and tested with cobas® HBV. The tested potentially interfering substances were shown not to interfere with the test performance of cobas® HBV.
Moreover, the presence of autoimmune diseases such as systemic lupus erythematosus (SLE), rheumatoid arthritis (RA) and antinuclear antibody were tested.
PMA P150014: FDA Summary of Safety and Effectiveness Data
{15}
In addition, drug compounds listed in Table 8 below were tested at three times the Cmax in the presence and absence of HBV DNA.
All potentially interfering substances have been shown to not interfere with the test performance. Negative results were obtained with cobas® HBV for all samples without HBV target and positive results were obtained on all of the samples with HBV target. Furthermore, the mean $\log_{10}$ titer of each of the positive HBV samples containing potentially interfering substances was within $\pm 0.5\log_{10}$ of the mean $\log_{10}$ titer of the respective positive spike control.
Table 8: Drug Compounds Tested for Interference with the Quantitation of HBV DNA by cobas® HBV
| Class of Drug | Generic Drug Name | |
| --- | --- | --- |
| Immune Modulator | Peginterferon α-2a Ribavirin | Peginterferon α-2b |
| HIV entry inhibitor | Maraviroc | |
| HIV Integrase Inhibitor | Elvitegravir/Cobicistat | Raltegravir |
| Non-nucleoside HIV Reverse Transcriptase Inhibitor | Efavirenz Etravirine | Nevirapine Rilpivirine |
| HIV protease inhibitor | Atazanavir | Lopinavir |
| | Tipranavir | Nelfinavir |
| | Darunavir | Ritonavir |
| | Fosamprenavir | Saquinavir |
| HCV Protease Inhibitor | Boceprevir | Telaprevir |
| | Simeprevir | |
| Reverse transcriptase or DNA polymerase inhibitors | Abacavir | Tenofovir |
| | Emtricitabine | Adefovir dipivoxil |
| | Entecavir | Telbivudine |
| | Foscarnet | Zidovudine |
| | Cidofovir | Aciclovir |
| | Lamivudine | Valganciclovir |
| | Ganciclovir | Sofosbuvir |
| Compounds for Treatment of Opportunistic Infections | Azithromycin | Pyrazinamide |
| | Clarithromycin | Rifabutin |
| | Ethambutol | Rifampicin |
| | Fluconazole | Sulfamethoxazole |
| | Isoniazid | Trimethoprim |
# K. Matrix Equivalency - EDTA Plasma Versus Serum
Fifty paired EDTA plasma and serum samples were analyzed for matrix equivalency. The HBV positive samples covered most genotypes and had titers across the entire linear range.
Matrix equivalency was shown in the tested samples with a mean titer deviation of $0.05\log_{10}$ (Figure 4).
PMA P150014: FDA Summary of Safety and Effectiveness Data
{16}

Figure 4: Matrix Equivalency Performance Between EDTA Plasma and Serum
# L. Cross Contamination
The cross-contamination rate for cobas® HBV was determined by testing 240 replicates of a normal, virus-negative (HIV, HCV and HBV) human EDTA-plasma sample and 225 replicates of a high titer HBV sample at $1.00\mathrm{E} + 09$ IU/mL. In total, five runs were performed with positive and negative samples in a checkerboard configuration.
All 240 replicates of the negative sample were non-reactive, resulting in a cross-contamination rate of $0\%$ . The two-sided $95\%$ exact confidence interval was $0\% - 1.53\%$ .
# M. Specimen Stability
The specimen stability of HBV viral target in whole blood specimens collected in EDTA plasma or serum preparation tubes and/or plasma and serum samples was evaluated using the cobas® HBV test for use on the cobas® 6800/8800 Systems simulating handling, transportation and storage conditions of patient samples prior to testing in diagnostic laboratories. The specimens were handled and stored under the conditions that simulated the handling, transporting and processing of blood specimens prior to testing separated plasma and serum with the cobas® HBV test for use on the cobas® 6800/8800 Systems. In addition the freeze/thaw stability of the HBV viral target in plasma and serum samples for up to 4 freeze/thaw cycles was evaluated.
PMA P150014: FDA Summary of Safety and Effectiveness Data
{17}
The results of this study demonstrated acceptable specimen stability under the following storage conditions for whole blood and separated EDTA plasma and serum samples:
- Whole blood collected in EDTA plasma or Serum preparation tubes may be stored for up to 24 hours at 2 °C to 25 °C before further processing and matrix separation
Additionally plasma and serum samples are stable for:
- up to 6 days at 2°C to 8°C or up to 6 months at -15°C to -80°C
- or 6 days at 2 to 8°C followed by 6 months at -75°C ± 15°C after matrix separation.
Plasma and serum samples are stable for up to 4 freeze/thaw cycles when frozen between -18°C to -80°C.
## N. Real-Time Reagent Stability:
Expiration dating for the cobas® HBV reagents has been established and approved at 16 months when stored at 2-8°C.
## X. SUMMARY OF CLINICAL STUDIES
### A. Study Design
The study was designed to evaluate the ability of the assay to predict clinical outcome.
Residual specimens were obtained from 396 subjects who were either:
- Participating in a randomized pharmaceutical clinical trial to receive treatment for 100 weeks with entecavir plus tenofovir or entecavir monotherapy, or
- Receiving standard treatment (tenofovir monotherapy) from routine clinical practice
Table 13: Treatment Groups
| Clinical Study | HBeAg Status | Treatment | Treatment Arm |
| --- | --- | --- | --- |
| Pharmaceutical Clinical Trial^{21} | HBeAg (+) | entecavir monotherapy | Arm I |
| | | entecavir + tenofovir | Arm II |
| | HBeAg (-) | entecavir monotherapy | Arm III (includes up to 17 subjects from clinical practice) |
| | | entecavir + tenofovir | Arm IV |
| Clinical Practice | HBeAg (-) | tenofovir monotherapy | Arm V |
HBeAg = Hepatitis B e antigen.
PMA P150014: FDA Summary of Safety and Effectiveness Data
Page 18
{18}
Testing with cobas® HBV was performed at 3 test sites, using 3 kit lots of reagents; each sample was tested with 1 kit lot of reagent. Each test site was equipped with 1 cobas® 6800 system.
The samples were labeled to ensure the test site personnel were blinded to any identifying information and reduce bias. Each sample had a unique identification number and a visit number. The samples were evenly distributed among the 3 test sites. Samples from each individual subject were tested at a single test site.
## B. Study Population Demographics and Baseline Parameters
The demographic distribution of the subjects in this study was consistent with that of patients with chronic HBV in the US, $^{2,3}$ HBeAg (+) and HBeAg (-) subjects were enrolled in this study, and data for these populations were analyzed separately (Table 14).
Table 14: Demographics and Baseline Characteristics of Subjects
| Characteristics | Statistics |
| --- | --- |
| Total, N | 396 |
| Age Category (years), n (%) | |
| < 40 | 186 (47.0%) |
| >= 40 | 210 (53.0%) |
| Age (years) | |
| Mean ± SD | 42 ± 15.2 |
| Median | 42 |
| Range | 17 - 81 |
| Gender, n (%) | |
| Male | 276 (69.7%) |
| Female | 120 (30.3%) |
| Race, n (%) | |
| Asian | 204 (51.5%) |
| Black / African American | 14 (3.5%) |
| White / Caucasian | 169 (42.7%) |
| Other | 9 (2.3%) |
| Genotype, n (%) | |
| A | 64 (16.2%) |
| A & G | 1 (0.3%) |
| B | 62 (15.7%) |
| C | 74 (18.7%) |
| D | 105 (26.5%) |
| E | 4 (1.0%) |
| F | 10 (2.5%) |
PMA P150014: FDA Summary of Safety and Effectiveness Data
Page 19
{19}
| Characteristics | Statistics |
| --- | --- |
| Mixed | 1 (0.3%) |
| Unknown | 75 (18.9%) |
| Normal ALT at Baseline, n (%) | |
| Yes | 23 (5.8%) |
| No | 361 (91.2%) |
| Unknown | 12 (3.0%) |
| Baseline ALT (IU/L) | |
| Mean ± SD | 140 ± 169.9 |
| Median | 96 |
| Range | 14 - 1583 |
| HBV DNA (log_{10} IU/mL) at Baseline | |
| Mean ± SD | 6.6 ± 2.38 |
| Median | 7.4 |
| Range | -0.0 - 10.1 |
| HBV DNA Category, n (%) | |
| < 2.0 x 10^{3} IU/mL | 41 (10.4%) |
| 2.0 x 10^{3} to 2.0 x 10^{4} IU/mL | 13 (3.3%) |
| > 2.0 x 10^{4} IU/mL | 330 (83.3%) |
| Unknown | 12 (3.0%) |
ALT = alanine aminotransferase; HBV = hepatitis B virus; DNA = deoxyribonucleic acid; SD = standard deviation.
## C. Safety and Effectiveness Results
### Definitions:
- Week 12 virologic response (VR) = HBV DNA 2 log₁₀ decrease from baseline
- Week 24 VR = HBV DNA < 2000 IU/mL (HBeAg (+)) or < 50 IU/mL (HBeAg (-))
- Week 48 VR = HBV DNA < 2000 IU/mL (HBeAg (+)) or < 50 IU/mL (HBeAg (-))
- Week 96 VR = HBV DNA < 50 IU/mL (VR endpoint)
- No-VR endpoint = HBV DNA > 50 IU/mL at Week 96
- Biochemical Response (BR) = normalization of ALT compared to baseline; for male ALT < 30 IU/L and for female ALT < 19 IU/L
- HBeAg loss = conversion from HBeAg (+) to HBeAg (-) status during therapy
### Predicting Virological Response at Week 96
In this study, baseline HBV DNA concentration and VRs at Weeks 12, 24, and 48 of treatment were used to evaluate the ability to predict outcome (VR, BR, or HBeAg
PMA P150014: FDA Summary of Safety and Effectiveness Data
{20}
loss) at Week 96 of therapy. VR96 (HBV DNA $< 50$ IU/mL) was assessed using HBV DNA results from an approved test.
When cobas® HBV was used to measure HBV DNA, a low baseline HBV DNA concentration $(< 10^{8}\mathrm{IU / mL})$ and VRs at Weeks 12, 24, and 48 were shown to be highly predictive of VR96 in both the HBeAg (+) and HBeAg (-) populations and for all the treatment arms in this study (PPVs $79.6\%$ to $100\%$ ) (Table 15 and Table 16).
Table 15: Probability of Achieving Virological Response at Week 96 Given Baseline HBV DNA $< {10}^{8}\mathrm{{IU}}/\mathrm{{mL}}$ by Treatment Arm
| | | | PPV (%) | | NPV (%) | | OR |
| --- | --- | --- | --- | --- | --- | --- | --- |
| On-Treatment Visit | Treatment Arm | Evaluable Subjects | Estimate (95% CI) | n / N | Estimate (95% CI) | n / N | Estimate (95% CI) |
| Baseline | Arm I | 103 | 93.5 (82.5, 97.8) | 43 / 46 | 31.6 (21.0, 44.5) | 18 / 57 | 6.62 (1.81, 24.20) |
| | Arm II | 102 | 96.2 (87.0, 98.9) | 50 / 52 | 4.0 (1.1, 13.5) | 2 / 50 | 1.04 (0.14, 7.69) |
| | Arm III | 49 | 100.0 (92.1, 100.0) | 46 / 46 | 25.0 (4.6, 69.9) | 1 / 4 | 30.00 (0.83, 1087.42) |
| | Arm IV | 48 | 97.9 (88.9, 99.6) | 46 / 47 | 100.0 (20.7, 100.0) | 1 / 1 | 92.00 (1.81, 4686.43) |
| | Arm V | 30 | 90.0 (74.4, 96.5) | 27 / 30 | NC | 0 | 9.00 (0.15, 541.69) |
Table 16: Probability of Achieving Virologic Response at Week 96 Given Virologic Response at a Specific On-Treatment Visit by Treatment Arm
| | | | PPV (%) | | NPV (%) | | OR |
| --- | --- | --- | --- | --- | --- | --- | --- |
| On-Treatment Visit | Treatment Arm | Eligible Subjects | Estimate (95% CI) | n / N | Estimate (95% CI) | n / N | Estimate (95% CI) |
| Week 12 | Arm I | 103 | 79.6 (70.8, 86.3) | 82 / 103 | NC | 0 | 3.90 (0.08, 202.63) |
| | Arm II | 100 | 97.0 (91.5, 99.0) | 97 / 100 | NC | 0 | 32.33 (0.54, 1921.79) |
| | Arm III | 48 | 97.8 (88.7, 99.6) | 45 / 46 | 0.0 (0.0, 65.8) | 0 / 2 | 11.25 (0.28, 445.33) |
| | Arm IV | 48 | 95.8 (86.0, 98.8) | 46 / 48 | NC | 0 | 23.00 (0.36, 1485.21) |
| | Arm V | 21 | 85.7 (48.7, 97.4) | 6 / 7 | 7.1 (1.3, 31.5) | 1 / 14 | 0.46 (0.02, 8.69) |
| Week 24 | Arm I | 103 | 96.1 (89.2, 98.7) | 74 / 77 | 69.2 (50.0, 83.5) | 18 / 26 | 55.50 (13.37, 230.39) |
PMA P150014: FDA Summary of Safety and Effectiveness Data
{21}
| | | | PPV (%) | | NPV (%) | | OR |
| --- | --- | --- | --- | --- | --- | --- | --- |
| On-Treatment Visit | Treatment Arm | Eligible Subjects | Estimate (95% CI) | n / N | Estimate (95% CI) | n / N | Estimate (95% CI) |
| | Arm II | 102 | 96.7 (90.8, 98.9) | 89 / 92 | 10.0 (1.8, 40.4) | 1 / 10 | 3.30 (0.31, 35.08) |
| | Arm III | 47 | 100.0 (89.8, 100.0) | 34 / 34 | 7.7 (1.4, 33.3) | 1 / 13 | 5.67 (0.18, 179.94) |
| | Arm IV | 49 | 97.7 (87.9, 99.6) | 42 / 43 | 16.7 (3.0, 56.4) | 1 / 6 | 8.40 (0.45, 156.19) |
| | Arm V | 20 | 94.1 (73.0, 99.0) | 16 / 17 | 33.3 (6.1, 79.2) | 1 / 3 | 8.00 (0.35, 184.38) |
| Week 48 | Arm I | 101 | 89.9 (81.9, 94.6) | 80 / 89 | 91.7 (64.6, 98.5) | 11 / 12 | 97.78 (11.28, 847.86) |
| | Arm II | 97 | 95.9 (89.9, 98.4) | 93 / 97 | NC | 0 | 23.25 (0.41, 1328.83) |
| | Arm III | 46 | 100.0 (91.6, 100.0) | 42 / 42 | 25.0 (4.6, 69.9) | 1 / 4 | 28.00 (0.77, 1015.78) |
| | Arm IV | 48 | 97.8 (88.4, 99.6) | 44 / 45 | 33.3 (6.1, 79.2) | 1 / 3 | 22.00 (0.98, 494.79) |
| | Arm V | 28 | 92.3 (75.9, 97.9) | 24 / 26 | 50.0 (9.5, 90.5) | 1 / 2 | 12.00 (0.53, 273.05) |
Notes (Table 14 & 15): Positive Predictive Value (PPV) = TP / (TP + FP) or the probability of being a VR96 given the subject was a virologic responder at a specific visit.
Negative Predictive Value (NPV) = TN / (FN + TN) or the probability of not being a VR96 given the subject was not a virologic responder at a specific visit.
Odds Ratio (OR) = (TP • TN) / (FP • FN).
95% CIs for PPV and NPV are calculated based on Wilson Score CI.
0.5 was added to empty cells (TP, TN, FP, or FN = 0), prior to calculation of OR and corresponding 95% CI.
VR96 is achieved if the subject has HBV DNA < 50 IU/mL from the COBAS® TaqMan® HBV Test For Use with the High Pure System at Week 96.
Week 12 VR = HBV DNA > 2 log₁₀ decrease from baseline; Week 24 VR = HBV DNA < 2000 IU/mL (HBeAg (+)) or < 50 IU/mL (HBeAg (-)); Week 48 VR = HBV DNA < 2000 IU/mL (HBeAg (+)) or < 50 IU/mL (HBeAg (-)).
## Predicting Biochemical Response at Week 96
The probability of achieving a biochemical response at Week 96 given an on-treatment VR at Week 12, Week 24 or Week 48 is summarized in Table 17.
The value of VR at Week 12, Week 24, or Week 48 as a predictor of VR96 varied by VR week, treatment Arm and baseline HBeAg status.
PMA P150014: FDA Summary of Safety and Effectiveness Data
{22}
Table 17: Probability of Achieving Biochemical Response at Week 96 Given Virologic Response at a Specific On-Treatment Visit by Treatment Arm
| | | | PPV (%) | | NPV (%) | | OR |
| --- | --- | --- | --- | --- | --- | --- | --- |
| On-Treatment Visit | Treatment Arm | Eligible Subjects | Estimate (95% CI) | n / N | Estimate (95% CI) | n / N | Estimate (95% CI) |
| Week 12 | Arm I | 101 | 62.4 (52.6, 71.2) | 63 / 101 | NC | 0 | 1.66 (0.03, 85.30) |
| | Arm II | 100 | 43.0 (33.7, 52.8) | 43 / 100 | NC | 0 | 0.75 (0.01, 38.79) |
| | Arm III | 49 | 50.0 (36.1, 63.9) | 23 / 46 | 66.7 (20.8, 93.9) | 2 / 3 | 2.00 (0.17, 23.62) |
| | Arm IV | 49 | 32.7 (21.2, 46.6) | 16 / 49 | NC | 0 | 0.48 (0.01, 25.57) |
| | Arm V | 21 | 40.0 (16.8, 68.7) | 4 / 10 | 90.9 (62.3, 98.4) | 10 / 11 | 6.67 (0.60, 74.51) |
| Week 24 | Arm I | 102 | 66.2 (55.1, 75.8) | 51 / 77 | 60.0 (40.7, 76.6) | 15 / 25 | 2.94 (1.16, 7.45) |
| | Arm II | 103 | 44.6 (34.8, 54.7) | 41 / 92 | 81.8 (52.3, 94.9) | 9 / 11 | 3.62 (0.74, 17.68) |
| | Arm III | 51 | 47.2 (32.0, 63.0) | 17 / 36 | 33.3 (15.2, 58.3) | 5 / 15 | 0.45 (0.13, 1.57) |
| | Arm IV | 50 | 38.6 (25.7, 53.4) | 17 / 44 | 100.0 (61.0, 100.0) | 6 / 6 | 7.56 (0.40, 144.09) |
| | Arm V | 24 | 42.1 (23.1, 63.7) | 8 / 19 | 80.0 (37.6, 96.4) | 4 / 5 | 2.91 (0.27, 31.22) |
| Week 48 | Arm I | 100 | 65.2 (54.8, 74.3) | 58 / 89 | 81.8 (52.3, 94.9) | 9 / 11 | 8.42 (1.71, 41.41) |
| | Arm II | 97 | 43.3 (33.9, 53.2) | 42 / 97 | NC | 0 | 0.76 (0.01, 39.29) |
| | Arm III | 49 | 52.3 (37.9, 66.2) | 23 / 44 | 40.0 (11.8, 76.9) | 2 / 5 | 0.73 (0.11, 4.81) |
| | Arm IV | 49 | 37.0 (24.5, 51.4) | 17 / 46 | 100.0 (43.9, 100.0) | 3 / 3 | 3.52 (0.17, 74.51) |
| | Arm V | 28 | 33.3 (18.0, 53.3) | 8 / 24 | 75.0 (30.1, 95.4) | 3 / 4 | 1.50 (0.13, 16.82) |
Predicting HBeAg loss
HBeAg loss could only be evaluated in subjects who were HBeAg (+) at baseline.
Absence of VR at Week 24 was highly predictive of persistence of HBeAg (NPVs were $\geq 80.0\%$ for both Arms I and II), and absence of VR at Week 48 also predicted HBeAg persistence in Arm I (NPV was $100\%$ ) (Table 18). As all subjects on the
PMA P150014: FDA Summary of Safety and Effectiveness Data
{23}
combination regimen (Arm II) had achieved VRs by Week 48, it was not possible to calculate an NPV at this time point for this group.
Table 18: Probability of HBeAg Loss at Week 96 Given Virologic Response at a Specific On-Treatment Visit by Treatment Arm
| | | | PPV (%) | | NPV (%) | | OR |
| --- | --- | --- | --- | --- | --- | --- | --- |
| On-Treatment Visit | Treatment Arm | Eligible Subjects | Estimate (95% CI) | n / N | Estimate (95% CI) | n / N | Estimate (95% CI) |
| Week 12 | Arm I | 102 | 46.1 (36.7, 55.7) | 47 / 102 | NC | 0 | 0.85 (0.02, 43.91) |
| | Arm II | 101 | 41.6 (32.5, 51.3) | 42 / 101 | NC | 0 | 0.71 (0.01, 36.60) |
| Week 24 | Arm I | 103 | 52.6 (41.6, 63.3) | 41 / 78 | 80.0 (60.9, 91.1) | 20 / 25 | 4.43 (1.51, 13.00) |
| | Arm II | 104 | 44.1 (34.4, 54.2) | 41 / 93 | 81.8 (52.3, 94.9) | 9 / 11 | 3.55 (0.73, 17.33) |
| Week 48 | Arm I | 101 | 51.1 (41.0, 61.2) | 46 / 90 | 100.0 (74.1, 100.0) | 11 / 11 | 23.00 (1.31, 403.28) |
| | Arm II | 98 | 40.8 (31.6, 50.7) | 40 / 98 | NC | 0 | 0.69 (0.01, 35.48) |
Conclusion
The results demonstrated that cobas® HBV is effective for monitoring the HBV DNA viral load in subjects with chronic HBV infection at the start of and during antiviral treatment. This study demonstrated that HBV DNA concentration measurement at baseline, a decrease in HBV DNA concentration at Week 12, or HBV DNA concentrations below specific thresholds at Weeks 24 or 48 during treatment predicted response to therapy; the study identified subjects who achieved Virologic Response, Biochemical Response, or loss of HBeAg at Week 96 of therapy.
D. Financial Disclosure
The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The pivotal clinical study included three investigators. None of the clinical investigators had disclosable financial interests/arrangements as defined in sections 54.2(a), (b), (c), and (f). The information provided does not raise any questions about the reliability of the data.
XI. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION
In accordance with the provisions of section 515(c)(2) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Microbiology Advisory
PMA P150014: FDA Summary of Safety and Effectiveness Data
Page 24
{24}
Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.
## XII. CONCLUSIONS DRAWN FROM PRECLINICAL AND CLINICAL STUDIES
### A. Effectiveness Conclusions
The effectiveness of the cobas® HBV has been demonstrated when used for the quantitation of hepatitis B virus (HBV) DNA in human EDTA plasma or serum of HBV-infected individuals. Study results indicate this test is effective when used as an aid in the management of patients with chronic HBV infection undergoing antiviral therapy. The test can be used to measure HBV DNA levels at baseline and during treatment to aid in assessing response to treatment.
### B. Safety Conclusions
Based on the results of the analytical and clinical laboratory studies, the cobas® HBV, when used according to the provided directions and in conjunction with other laboratory results and clinical information, should be safe and pose minimal risk to the patient due to false test results.
### C. Risk/Benefit Analysis
The probable benefits of the device are based on data collected in a clinical study conducted to support PMA approval as described above. As an in vitro nucleic acid amplification test, the cobas® HBV involves removal of blood from an individual for testing purposes. The benefits to chronically HBV-infected individuals undergoing antiviral therapy tested by the assay outweigh any potential adverse event or risk to the patient or user due to assay malfunction or operator error. Additionally, the direct benefit of the measurement of HBV DNA measurement is to determine response to anti-viral therapy and adjust dosages and durations of these critical medications accordingly.
- Additional factors to be considered in determining probable risks and benefits for the cobas® HBV include:
- As a quantitative assay, risks are primarily from inaccurate results rather than false positive or false negative results, although the latter could be considered at the very low end of measurement where levels below 50 IU are considered a virological response. In essence, when used as a quantitative assay, there are several important decisions points, i.e., high levels (> 10⁸) where combination therapy may be considered, and at very low levels used to confirm complete response. Within this range, effects of inaccurate measurements are likely to have limited clinical consequences; perhaps the greatest potential for an adverse effect would be if a measurement was inaccurately elevated in the setting of treatment such that treatment response (or viral resistance) was inferred. However, in this circumstance, other markers (such as ALT) might suggest inaccuracy, and repeat measurement is likely. There is also substantial biological variability in HBV viral load (in the absence of treatment), and inaccurate measurements will be interpreted in the context of expected variability.
PMA P150014: FDA Summary of Safety and Effectiveness Data
Page 25
{25}
- Inaccurate results would be most significant between 2000 – 20,000 IU/mL, where treatment may be considered under different scenarios; however, serial measurements are common, and clinicians are likely to repeat measurements when there is consideration of treatment but a deciding factor is a low HBV value. The risks from ‘overtreatment’ (i.e., treatment initiated due to an unexpectedly high HBV viral load) are relatively small absent the cost of medication.
- There are other FDA-approved devices for this specific use. Performance of this device is similar to existing devices. Studies with other devices have arrived at conclusion similar to the clinical performance study contained in this submission.
- Chronic Hepatitis B infection is a significant illness that is potentially fatal over time; approximately 1- 5% of immunocompetent adults with acute Hepatitis B infection develop chronic infection; the rate of chronic infection is far higher for patients exposed at birth by infected mothers or exposed early in childhood. Excluding neonatal exposure, transmission of Hepatitis B is by exposure to infected blood or by sexual contact with a chronically infected partner.
- There are multiple FDA-approved drugs for treatment, primarily nucleoside/nucleotide analogs as well as Interferon. Treatment is generally very effective but less so in patients with greater degrees of cirrhosis and may be complicated by the emergence of viral resistance. Well adopted published treatment guidelines exist.
- In conclusion, given the available information above, the data support that the cobas® HBV, when used for the quantitation of hepatitis B virus (HBV) DNA in human EDTA plasma or serum of HBV-infected individuals, the probable benefits outweigh the probable risks.
D. Overall Conclusions
The data in this application support the reasonable assurance of safety and effectiveness of this device when used in accordance with the indications for use. The data from the preclinical studies demonstrated acceptable analytical sensitivity, traceability, linearity, precision, and analytical specificity of the cobas® HBV when used according to the instructions for use as stated in the labeling, the warnings and precautions, and limitations sections of the labeling. The clinical usefulness study and the statistical analysis of clinical data in this application has shown that the results obtained with the cobas® HBV are informative for use as an aid in the management of patients with chronic HBV infection undergoing anti-viral therapy, and that the test is safe and effective when used according to the directions for use in the labeling.
XIII. CDRH DECISION
CDRH issued an approval order on October 14, 2015. The final conditions of approval can be found in the approval order.
The applicant’s manufacturing facilities have been inspected and found to be in compliance with the device Quality System (QS) regulation (21 CFR 820).
PMA P150014: FDA Summary of Safety and Effectiveness Data
Page 26
{26}
PMA P150014: FDA Summary of Safety and Effectiveness Data
Page 27
# XIV. APPROVAL SPECIFICATIONS
Directions for use: See device labeling.
Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events in the device labeling.
Post-approval Requirements and Restrictions: See approval order.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.