CRYPTO-CEL IF TEST
Device Facts
| Record ID | K955852 |
|---|---|
| Device Name | CRYPTO-CEL IF TEST |
| Applicant | Techlab, Inc. |
| Product Code | MHJ · Microbiology |
| Decision Date | Aug 5, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 866.3220 |
| Device Class | Class 2 |
Indications for Use
The TechLab Crypto-Cel IF Test can be used to detect Cryptosporidium oocysts in fecal specimens from persons suspected of having intestinal disease due to this pathogen.
Device Story
Immunofluorescent assay kit; detects Cryptosporidium oocysts in fecal specimens. Input: fecal samples; reagent: FITC-labeled anti-Cryptosporidium monoclonal antibody derived from mouse ascites fluid. Operation: immunofluorescent staining of specimens; visualization via fluorescence microscopy. Output: visual detection of fluorescent oocysts. Used in clinical laboratories by trained personnel. Results serve as diagnostic aid for cryptosporidiosis; informs clinical management of intestinal disease.
Clinical Evidence
Two clinical studies compared TechLab Crypto-Cel IF Test against Merifluor Cryptosporidium/Giardia Direct Immunofluorescent Detection Procedure and Crypto/Giardia-Cel IF Test. Results showed 100% correlation across all compared tests for detection of Cryptosporidium in fecal specimens.
Technological Characteristics
Immunofluorescent assay; utilizes FITC-labeled anti-Cryptosporidium monoclonal antibody (mouse ascites origin). Manual diagnostic test kit.
Indications for Use
Indicated for detection of Cryptosporidium oocysts in fecal specimens from patients suspected of intestinal cryptosporidiosis.
Regulatory Classification
Identification
Entamoeba histolytica serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Entamoeba histolytica in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identify Entamoeba histolytica directly from clinical specimens. The identification aids in the diagnosis of amebiasis caused by the microscopic protozoan parasite Entamoeba histolytica and provides epidemiological information on diseases caused by this parasite. The parasite may invade the skin, liver, intestines, lungs, and diaphragm, causing disease conditions such as indolent ulcers, an amebic hepatitis, amebic dysentery, and pulmonary lesions.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.
Reference Devices
- Merifluor Cryptosporidium/Giardia Direct Immunofluorescent Detection Procedure (Meridian Diagnostics, Inc.)
- Crypto/Giardia-Cel IF Test
Related Devices
- K955897 — CRYPTO/GIARDIA-CEL IF TEST · Techlab, Inc. · Aug 5, 1996
- K024113 — IVD CRYPTO/GIARDIA DFA · Ivd Research, Inc. · Mar 5, 2003
- K031059 — IVD CRYPTOSPORIDIUM ANTIGEN DETECTION ASSAY, MODEL CP-96 · Ivd Research, Inc. · Jul 10, 2003
- K051929 — GIARDIA/CRYPTOSPORIDIUM CHEK · Techlab, Inc. · Nov 17, 2005
- K052932 — CRYPTOSPORIDIUM II · Techlab, Inc. · Nov 4, 2005