K181379 · Techlab, Inc. · LYR · Aug 21, 2018 · Microbiology
Device Facts
Record ID
K181379
Device Name
H. PYLORI QUIK CHEK
Applicant
Techlab, Inc.
Product Code
LYR · Microbiology
Decision Date
Aug 21, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3110
Device Class
Class 1
Indications for Use
The TECHLAB® H. PYLORI QUIK CHEK™ test is a rapid membrane enzyme immunoassay for the qualitative detection of Helicobacter pylori specific antigen in a single use cassette. It is intended for use with human fecal specimens to aid in the diagnosis of H. pylori infection and to demonstrate loss of H. pylori antigen following treatment. The test can be used with unpreserved fecal specimens and fecal specimens preserved in transport media from patients suspected of H. pylori infection. Testing of patients to demonstrate loss of H. pylori antigen following treatment should be performed no sooner than 4 weeks after completion of the treatment regimen. Test results should be taken into consideration by the physician in conjunction with the patient history and symptoms.
Device Story
Rapid membrane enzyme immunoassay for qualitative detection of H. pylori antigen in human fecal specimens. Device uses single-use cassette with reaction window containing immobilized antibodies. Input: unpreserved or transport-media-preserved stool samples. Principle: lateral flow immunochromatographic assay; H. pylori antigen binds to HRP-coupled antibodies; visual appearance of blue lines on 'C' (control) and 'T' (test) sides indicates presence of antigen. Used in clinical settings; operated by laboratory personnel. Results interpreted visually by clinician to aid diagnosis of H. pylori infection or confirm eradication post-treatment. Benefits include rapid, point-of-care-compatible diagnostic confirmation of infection status.
Clinical Evidence
Prospective clinical study (n=122) compared device to composite reference method (CRM: histology/rapid urease test). Sensitivity 97.0% (95% CI: 84.7-99.5%), specificity 100% (95% CI: 95.9-100%). Post-therapy study (n=9) showed 100% sensitivity. Retrospective study (n=200) compared to FDA-cleared ELISA, showing 98.9% PPA and 97.2% NPA. Bench testing confirmed 100% reproducibility and no prozone/hook effect at 10x clinical concentrations.
Technological Characteristics
Rapid membrane enzyme immunoassay (ELISA). Components: single-use cassette, HRP-conjugated antibodies, tetramethylbenzidine substrate. Analyte: H. pylori fecal antigen. Detection: visual colorimetric (blue lines). Storage: 2-8°C. Analytical sensitivity (LoD): 16.08 ng/mL (unpreserved), 13.01-19.96 ng/mL (transport media). No cross-reactivity with common intestinal pathogens.
Indications for Use
Indicated for patients suspected of H. pylori infection, including those with dyspepsia, gastritis, or peptic ulcer symptoms. Applicable for initial diagnosis and post-treatment monitoring (at least 4 weeks post-therapy). No specific age or gender contraindications.
Regulatory Classification
Identification
Campylobacter fetus serological reagents are devices that consist of antisera conjugated with a fluorescent dye used to identify Campylobacter fetus from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by this bacterium and provides epidemiological information on these diseases. Campylobacter fetus is a frequent cause of abortion in sheep and cattle and is sometimes responsible for endocarditis (inflammation of certain membranes of the heart) and enteritis (inflammation of the intestines) in humans.
K153661 — ImmunoCard STAT! HpSA · Meridian Bioscience, Inc. · Mar 14, 2016
K032222 — IMMUNOCARDSTAT HPSA · Meridian Bioscience, Inc. · Dec 5, 2003
K181400 — H. PYLORI CHEK · Techlab, Inc. · Aug 21, 2018
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
K181379
B. Purpose for Submission:
To obtain a substantial equivalence determination for the detection of *Helicobacter pylori* antigens in human stool.
C. Measurand:
*H. pylori* antigen
D. Type of Test:
Qualitative membrane enzyme immunoassay
E. Applicant:
TechLab Inc.
F. Proprietary and Established Names:
*H. PYLORI* QUIK CHEK
G. Regulatory Information:
1. Regulation section:
21 CFR 866.3110 *Campylobacter fetus* serological reagents
2. Classification:
Class I
3. Product code:
LYR-- *Campylobacter pylori*
4. Panel:
{1}
83-Microbiology
## H. Intended Use:
1. Intended use(s):
The TECHLAB H. PYLORI QUIK CHEK test is a rapid membrane enzyme immunoassay for the qualitative detection of *Helicobacter pylori* specific antigen in a single use cassette. It is intended for use with human fecal specimens to aid in the diagnosis of *H. pylori* infection and to demonstrate loss of *H. pylori* antigen following treatment. The test can be used with unpreserved fecal specimens and fecal specimens preserved in transport media from patients suspected of *H. pylori* infection. Testing of patients to demonstrate loss of *H. pylori* antigen following treatment should be performed no sooner than 4 weeks after completion of the treatment regimen. Test results should be taken into consideration by the physician in conjunction with the patient history and symptoms.
2. Indication(s) for use:
Same as the Intended Use.
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
None
## I. Device Description:
The *H. PYLORI QUIK CHEK* test utilizes antibodies specific for *H. pylori* antigen in human fecal samples. The Membrane Device contains a “Reaction Window” with two vertical lines of immobilized antibodies. The test line (“T”) contains antibodies specific for *H. pylori* antigen. The control line (“C”) contains antibodies to horseradish peroxidase (HRP). The “Conjugate” consists of antibodies to *H. pylori* antigen coupled to horseradish peroxidase. After incubation, the Reaction Window is examined visually for the appearance of vertical blue lines on the “C” and “T” sides of the Reaction Window. A blue line on the “T” side of the Reaction Window indicates a positive result. A positive “C” reaction, indicated by a vertical blue line on the “C” side of the Reaction Window, confirms that the sample and reagents were added correctly, the reagents were active at the time of performing the assay, and that the sample migrated properly through the Membrane Device. It also confirms the reactivity of the other reagents associated with the assay.
{2}
J. Substantial Equivalence Information:
1. Predicate device name(s):
ImmunoCard STAT! HpSA
2. Predicate 510(k) number(s):
K032222
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device: H. PYLORI QUIK CHEK (K181379) | Predicate: ImmunoCard STAT! HpSA (K032222) |
| Product Code | LYR | LYR |
| Intended Use | The TECHLAB H. PYLORI QUIK CHEK™ test is a rapid membrane enzyme immunoassay for the qualitative detection of Helicobacter pylori specific antigen in a single use cassette. It is intended for use with human fecal specimens to aid in the diagnosis of H. pylori infection and to demonstrate loss of H. pylori antigen following treatment. The test can be used with unpreserved fecal specimens and fecal specimens preserved in transport media from patients suspected of H. pylori infection. Testing of patients to demonstrate loss of H. pylori antigen following treatment should be performed no sooner than 4 weeks after completion of the treatment regimen. Test results should be taken into consideration by the physician in conjunction with the patient history and symptoms. | ImmunoCard STAT! HpSA is a rapid in vitro qualitative assay for the detection of Helicobacter pylori antigen in human stool. The stool antigen detection is intended to aid in the diagnosis of H. pylori infection and to demonstrate loss of H. pylori stool antigen following treatment. Conventional medical practice recommends that testing by any method to confirm the loss of antigen be done at least four weeks following completion of therapy. |
| Measured analyte | Detection of H. pylori stool antigen | Same |
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Device: H. PYLORI QUIK CHEK (K181379) | Predicate: ImmunoCard STAT! HpSA (K032222) |
| Target Population | Persons suspected of having H. pylori infection | Same |
| Type of Test | Qualitative | Same |
| Controls | Positive and negative control included in kit Internal Control line | Same |
| Storage | Refrigerated (2°C – 8°C) | Same |
| Reading Method | Manual/Visual | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device: H. PYLORI QUIK CHEK (K181379) | Predicate: ImmunoCard STAT! HpSA (K032222) |
| Specimen Type | Fecal Specimens in Cary-Blair and C&S Transport Media | Unpreserved Fecal Specimen |
| Time to Result | 30 minutes | 5 minutes |
| Technology | Enzyme Linked Immunoassay (ELISA) | Immunochromatographic (ICT) |
| Antibody Format | Polyclonal/Polyclonal | Monoclonal/Monoclonal |
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP07-A2 Interference Testing In Clinical Chemistry; Approved Guideline – Second Edition
CLSI EP15-A3 User Verification of Precision And Estimation Of Bias; Approved Guideline - Third Edition
CLSI EP17-A2 Evaluation Of Detection Capability For Clinical Laboratory Measurement Procedures; Approved Guideline - Second Edition
L. Test Principle:
Lateral flow immunochromatographic assay.
{4}
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
The reproducibility of the *H. PYLORI QUIK CHEK* test was determined using an eight member masked fecal specimen panel. The panel consisted of 2 negative, 2 high negative (just below C5), 2 low positive (just above LoD), and 2 moderate positive (3-4x higher than the C95) specimens. Each fecal specimen was spiked using a known concentration of *H. pylori* purified flagellar antigen from whole organism to achieve the desired concentration. Testing was performed at 2 independent laboratories and on-site at TECHLAB, Inc. The Specimens were tested twice a day in triplicate over a five day period by multiple technicians at each site using two different kit lots. Positive and negative controls were run with each sample panel of masked specimens. The result from each laboratory were submitted to TECHLAB Inc. and compared. The results for the positive control, negative control, true negative, low positive, and moderate positive, was consistent among all three locations with 100% reproducibility for all test panels. The high negative provided a positive result 63 out of 180 observations.
b. Linearity/assay reportable range:
Not applicable.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Sample stability study:
The effect of specimen storage on antigen stability was evaluated for both fresh stool samples and stool samples in transport media. The following transport media were used for the study: Thermo Scientific *Protocol* Cary Blair media and Thermo Scientific *Protocol* C&S media. For the analysis, a total of 32 fecal specimens were tested with the *H. PYLORI QUIK CHEK* test. The samples were prepared using a negative fecal matrix and spiked with *H. pylori* antigen (ATCC strain 43526). The panel consisted of 2 negative, 5 high negative (C₅), 10 low positive (1-2x C₉₅), and 15 positive specimens covering the range of the test (For fresh and preserved specimens this range was 50 ng/mL – 1200 ng/mL).
Fresh samples were stored at refrigerated temperatures (between 2°C and 8°C) and room temperatures (between 20°C and 25°C) and were tested at 0, 24, 48, 72 and 96 hours. Positive and negative controls were also tested at each timepoint.
Fecal specimens added to Cary Blair and C&S media were transported as recommended in their respective package inserts; samples were stored at refrigerated temperatures (between 2°C and 8°C) and room temperatures (between 20°C and 25°C) and were tested at 24 hour intervals from 0 to 96 hours.
5
{5}
For fresh samples stored at refrigerated and room temperatures, the positive and negative samples gave the expected results 100% of the time. Storage in Cary Blair and C&S transport media did not affect the stability of the samples. Based on these results, the recommended storage time for fresh, Cary Blair, and C&S stored samples is up to 96 hours when stored between 2°C and 8°C and between 20°C and 25°C.
## Frozen sample stability study:
Stability of frozen fecal matrix samples compared to fresh samples was established using the 32 masked fecal specimen panel prepared as described for the storage stability study. Samples were not diluted into transport media for this study. Samples were stored at ≤ -10°C or ≤ -70°C for 14 days. Specimens were tested at 0, 5, 10, and 14 days. The results showed that all positive samples remained positive and negative samples remained negative throughout the study.
## Freeze/Thaw study:
A study was conducted to determine stability after 3 freeze/thaw cycles using the 32-masked fecal specimen panel described for the storage stability study. Samples were not diluted into transport media for this study. The results showed that up to 3 freeze/thaw cycles neither enhanced nor impaired the performance of the H. PYLORI QUIK CHEK test. Therefore, the package insert will indicate that if samples are not tested fresh, frozen stool samples may be tested and may undergo up to no more than 2 freeze/thaw cycles.
## d. Detection limit:
The limit of detection (LoD) for the H. PYLORI QUIK CHEK test was determined by spiking purified flagellar antigen into unpreserved (raw stool) and preserved (Cary Blair and C&S media) stool samples. The concentrations are reported in ng/mL and by factoring in the dilutions and the final volume used in the assay. The LoD is the concentrations of the antigen that yields a positive result 95% of the time and a negative result 5% of the time. Table 1 lists the LoD values for antigen spiked in negative fecal matrix, Cary Blair, and C&S transport media.
Table 1. LoD Values for H. PYLORI QUIK CHEK
| ng/mL | ng/test |
| --- | --- |
| Fecal Matrix | |
| 16.1 | 0.24 |
| Cary Blair | |
| 13.0 | 0.20 |
| C & S | |
| 20.0 | 0.31 |
{6}
e. Analytical specificity:
The H. PYLORI QUIK CHEK test was evaluated for cross-reactivity with the bacteria, fungi, and viruses listed below. H. pylori purified flagellar antigen (ATCC strain 43526) was spiked in at 2-3 x LoD. H. pylori was spiked into clinical matrix that was negative for H. pylori. Bacteria was spiked at concentration of >10⁸ CFU/mL and viruses at a range from 10³.³ to 10⁸.²⁵ TCID₅₀ units per 0.2 mL. Due to safety concerns for Listeria monocytogenes, freeze-dried biomaterial from ATCC was reconstituted at >1×10⁴ CFU/mL in PBS and spiked with H. pylori purified flagellar antigen (ATCC strain 43526) at 2-3X LoD.
Acinetobacter baumannii
Bacillus cereus
Campylobacter coli
Campylobacter concisus
Campylobacter hyointestinalis
Campylobacter jejuni
Campylobacter upsaliensis
Clostridium bifermentans
Clostridium perfringens
Enterobacter cloacae
Escherichia coli
Escherichia coli EPEC
Escherichia coli O157:H7 (non-toxigenic)
Haemophilus influenzae
Listeria monocytogenes
Porphyromonas asaccharolytica
Pseudomonas fluorescens
Staphylococcus aureus
Streptococcus agalactiae
Bacillus subtilis
Borrelia burgdorferi
Campylobacter fetus
Campylobacter fetus
Campylobacter helveticus
Campylobacter lari
Candida albicans
Clostridium difficile
Edwardsiella tarda
Enterococcus faecalis
Escherichia coli EIEC
Escherichia coli ETEC
Escherichia coli O157:H7 (toxigenic)
Lactobacillus acidophilus
Peptostreptococcus anaerobius
Pseudomonas aeruginosa
Salmonella typhimurium
Staphylococcus aureus (Cowan's)
Yersinia enterocolitica
Adenovirus, 2, 40,
Coxsackievirus B1, B2, B3, B6
Human Rotavirus*
* tested at 10¹.⁷⁵
Coronavirus
Echovirus 9, 22, 70
The H. PYLORI QUIK CHEK test was evaluated for interfering substances with the substances and concentrations listed in Table 2. None of the substances where shown to interfere with the performance of the H. PYLORI QUIK CHEK test.
{7}
Table 2. Interfering Substances
| Barium sulfate (5% w/v) | Mylanta (4.2 mg/mL) |
| --- | --- |
| Benzalkonium Chloride (1% w/v) | Naproxen Sodium (5% w/v) |
| Ciprofloxacin (0.25% w/v) | Nonoxynol-9 (1% w/v) |
| Ethanol (1% w/v) | Nystatin (1% w/v) |
| Hog gastric mucin (3.5% w/v) | Palmitic acid (fecal fat) (40% w/v) |
| Human blood (40% w/v) | Pepto-Bismol (5% v/v) |
| Hydrocortisone (1% w/v) | Phenylephrine (1% w/v) |
| Imodium (5% v/v) | Prilosec OTC (5 ug/mL) |
| Kaopectate (5%v/v) | Sennosides (1% w/v) |
| Leukocytes (0.05% v/v) | Simethicone (10% w/v) |
| Maalox Advanced (5% v/v) | Steric acid (fecal fat) (40% w/v) |
| Mesalazine (10% w/v) | Tagamet (5 ug/mL) |
| Metronidazole (0.25% w/v) | TUMS (50 ug/mL) |
| MiraLAX (7% w/v) | Human Urine (5% v/v) |
| Mineral Oil (10% w/v) | Vancomycin (0.25% w/v) |
# Inclusivity study
The reactivity of six $H.$ pylori strains spanning the 3 major clades (hpEastAsia, hpEurope, and hpAfrica1) was evaluated. Samples were prepared by spiking negative fecal matrix with purified flagellar antigen from each $H.$ pylori strain at 2-3x LoD. All results were read visually/manually and were positive, demonstrating that the $H.$ PYLORI QUIK CHEK test can detect antigen from strains representing the major $H.$ pylori clades.
# H. pylori strains
ATCC 700392 (hpEurope)
ATCC 43526 (hpEurope)
ATCC 700824 (hpAfrica1)
ATCC 43504 (clade unknown)
ATCC 43579 (clade unknown)
JP26 (hpEastAsia)
f. Assay cut-off:
Not applicable.
g. Prozone/Hook Effect:
The purpose of this study was to demonstrate that a high concentration of $H.$ pylori (antigen) does not interfere with a positive reaction in the $H.$ PYLORI QUIK CHEK test. High concentration samples were prepared by spiking negative fecal matrix with $H.$ pylori antigen at $12\mathrm{ug / mL}$ (i.e. $10\mathrm{x}$ the highest concentration observed in a positive clinical sample). A total of 5 different dilutions were prepared from the contrived sample. Samples were prepared by spiking the following concentrations of
{8}
antigen into a negative fecal pool: 12 ug/mL, 6 ug/mL, 2.4 ug/mL, 1.2 ug/mL, 0.6 ug/mL and 0.12 ug/mL. Testing was performed in triplicate according to the Package Insert instructions. The results demonstrated that elevated levels of antigen did not affect the test results.
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable.
b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical Sensitivity:
Prospective study
Initial Diagnosis
The performance of the H. PYLORI QUIK CHEK test was evaluated by conducting a multi-site prospective clinical study. Six independent sites were included, five of which collected specimens from patients suspected of H. pylori infection. The sixth site, TECHLAB, did not conduct specimen collection. Of these five collection sites, three conducted testing using both the H. PYLORI QUIK CHEK test and the composite reference method (CRM). The remaining two sites performed CRM testing and sent samples to TECHLAB for testing using the H. PYLORI QUIK CHEK test. The CRM for diagnosis of H. pylori infection is based on endoscopy obtained gastric biopsy. The biopsied tissue was tested for the presence of H. pylori by histology, or rapid urease test. The sensitivity and specificity for the H. PYLORI QUIK CHEK test was determined using the CRM.
Prospective testing consisted of 205 stool specimens collected from the patients with symptoms of dyspepsia, gastritis, or peptic ulcer who were scheduled to undergo endoscopy with gastric biopsy as part of routine care (Initial Diagnosis Group). Of these, 83 patients were excluded either because they were on a treatment regimen [i.e., proton-pump inhibitors (PPIs), or antibiotics] or had samples with CRM results that were rapid urease positive but histology negative. The remaining 122 patients who were not taking PPIs or antibiotics at the time of specimen collection were considered for final analysis. These specimens were tested at the following five sites: Carilion Clinic, International Centre for Diarrhoeal Disease Research Bangladesh, Mayo Clinic, University of Virginia, Kliniken Essen-Mitte, and TECHLAB (internal).
9
{9}
The ages of patients ranged from less than 19 years to 82 years with 100% of the specimens coming from patients were ≥ 18 years. Of the 122 patients tested 64% were female and 34% were male. No difference in test performance was observed based on patient age or gender. The results are provided in Table 3 which shows the clinical performance of the H. PYLORI QUIK CHEK test for all 6 test sites combined. The results of the study show that the H. PYLORI QUIK CHEK test exhibited a sensitivity of 97.0%, and a specificity of 100% compared to the CRM.
Table 3. Clinical Performance of the H. PYLORI QUIK CHEK
| Initial diagnosis | CRM Positive | CRM Negative |
| --- | --- | --- |
| H. PYLORI QUIK CHEK
Positive | 32 | 0 |
| H. PYLORI QUIK CHEK
Negative | 1* | 89 |
| Sensitivity (95% C.I.) | 97.0% (84.7% - 99.5%) |
| --- | --- |
| Specificity (95% C.I.) | 100%(95.9% - 100%) |
*Additional testing with an FDA cleared H. pylori stool antigen test provided an antigen negative result.
## Post Therapy Diagnosis
Eradication (post-therapy) evaluation was conducted on patients enrolled prospectively at 3 sites. A total of 9 specimens were collected at least 4 weeks after completion of the treatment regimen. Post therapy evaluation was conducted using a two-step algorithm of patient analysis. First, patients were screened for the continued presence of H. pylori using an FDA-cleared stool antigen test. Positive patients were reflexed to a follow-up endoscopy and biopsy analysis by rapid urease test and histology. The results are provided in Table 4 which shows the clinical performance of the H. PYLORI QUIK CHEK test for all 6 test sites combined. The results of the study show that the H. PYLORI QUIK CHEK test exhibited a sensitivity of 100.0% compared to the CRM.
Table 4. Clinical Performance of the H. PYLORI QUIK CHEK
| Post treatment diagnosis | CRM Positive | CRM Negative |
| --- | --- | --- |
| H. PYLORI QUIK CHEK
Positive | 9 | 0 |
| H. PYLORI QUIK CHEK
Negative | 0 | 0 |
| Sensitivity (95% C.I.) | 100% (70.1% - 100%) |
| --- | --- |
The ages of patients ranged from 33 years to 72 Years. Of the 9 patients tested, 6
{10}
were female and 3 were male.
# Retrospective study
A retrospective study was conducted to evaluate performance and to augment the prospective clinical study. Testing of retrospective samples was conducted at TECHLAB.
To provide assurance that the retrospective samples are representative of wide range of OD readings, CRM positives samples and retrospective positive samples were both tested by an FDA cleared ELISA. The distribution and mean OD values obtained from the retrospective study, $n = 200$ samples, were compared to those values obtained from CRM positive samples, $n = 46$ , to ensure that the use of retrospective sample results reflects the OD distribution of CRM positive samples. This analysis demonstrated that the use of retrospective and prospective clinical samples has similar distributions and no concern of bias was noted.
The performance of the H. PYLORI QUIK CHEK test was evaluated by testing retrospective samples at TECHLAB. Retrospective testing consisted of 200 frozen stool specimens (94 positive and 106 negative by an FDA Cleared ELISA) obtained from sample repositories. Positive percent agreement (PPA) and negative percent agreement (NPA) for the H. PYLORI QUIK CHEK test was determined by comparing to an FDA Cleared ELISA that was tested concurrently. The results of the study show that the H. PYLORI QUIK CHEK test exhibited a PPA of $98.9\%$ and a NPA of $97.2\%$ with an FDA Cleared ELISA. The results are shown in Table 5.
Table 5. Clinical performance of the H. PYLORI QUIK CHEK test on retrospective specimens.
| N = 200 | FDA Cleared ELISA (Positive) | FDA Cleared ELISA (Negative) |
| --- | --- | --- |
| H. PYLORI QUIK CHEK Positive | 93 | 3** |
| H. PYLORI QUIK CHEK Negative | 1* | 103 |
| Performance | | 95% CI |
| Positive Percent Agreement | 98.9% | 94.2%-99.8% |
| Negative Percent Agreement | 97.2% | 92.0%-99.0% |
* H. pylori DNA was amplified from the samples with PCR
**No H. pylori DNA was amplified from the sample with PCR
{11}
b. Clinical specificity:
See section M3a. above.
c. Other clinical supportive data (when a. and b. are not applicable):
See section M3a. above.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
Not applicable.
N. Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
12
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.