The Helicobacter pylori IgG ELISA test is an enzyme-linked immunosorbent assay (ELISA) for the qualitative detection of human IgG class antibodies to Helicobacter pylori in serum and in EDTA or heparin-treated plasma. The test is intended to aid in the diagnosis of Helicobacter pylori infection in adult patients with clinical symptoms of gastritis. FOR IN VITRO DIAGNOSTIC USE.
Device Story
The device is an ELISA-based immunoassay for qualitative detection of human IgG antibodies to H. pylori in serum or plasma. It utilizes a microplate coated with partially purified H. pylori antigen. Patient samples are incubated in the wells; if H. pylori antibodies are present, they bind to the antigen. An HRP-conjugated anti-human IgG antibody is added, followed by a TMB substrate. The enzyme reaction produces a blue color, which turns yellow upon addition of a stop solution. The microplate is read using a 450 nm microplate reader. Results are interpreted against a cutoff value (38.5 EIU) to determine H. pylori status. The test is intended for use in clinical laboratory settings by trained personnel to assist physicians in diagnosing H. pylori infection in symptomatic adults.
Clinical Evidence
Clinical performance was evaluated against gastric mucosa culture as the gold standard. In a study of 264 samples, the device demonstrated a clinical sensitivity of 91.5% (95% CI: 86.3-95.0%) and a clinical specificity of 65.1% (95% CI: 60.8-68.0%). Positive Predictive Value was 67.9% and Negative Predictive Value was 90.5%. Analytical performance included precision studies (inter-assay CVs 1.5-7.0%), cross-reactivity testing with related microorganisms (Yersinia, E. coli, Campylobacter, Salmonella), and interference testing for triglycerides, hemoglobin, and bilirubin, showing no significant impact.
Technological Characteristics
Enzyme-linked immunosorbent assay (ELISA) using partially purified H. pylori antigen adsorbed on polystyrene microplate wells. Detection via HRP-conjugated monoclonal anti-human IgG and TMB substrate. Requires 450 nm microplate reader. Includes calibrator and positive/negative controls. Reagents include phosphate-based washing and diluent buffers. Qualitative output based on EIU units.
Indications for Use
Indicated for qualitative detection of human IgG antibodies to Helicobacter pylori in serum or EDTA/heparin plasma to aid diagnosis of H. pylori infection in adult patients with clinical symptoms of gastritis.
Regulatory Classification
Identification
Campylobacter fetus serological reagents are devices that consist of antisera conjugated with a fluorescent dye used to identify Campylobacter fetus from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by this bacterium and provides epidemiological information on these diseases. Campylobacter fetus is a frequent cause of abortion in sheep and cattle and is sometimes responsible for endocarditis (inflammation of certain membranes of the heart) and enteritis (inflammation of the intestines) in humans.
Predicate Devices
IMMULITE® Helicobacter pylori IgG (k000463)
Related Devices
K110745 — HELICOBACTER PYLORI ELISA IGG TEST KIT · Gold Standard Diagnostics · Mar 2, 2012
K991820 — HELICOBACTER PYLORI IGG ELISA TEST SYSTEM. AN ENZYME LINKED IMMUNOSORBENT ASSAY (ELISA)FOR THE DETECTION OF IGG ANTIBODI · Zeus Scientific, Inc. · Aug 12, 1999
K033067 — PYLORI IGG · Trinity Biotech USA · Nov 26, 2003
K961221 — PYLORI IGG ELISA TEST SYSTEM · Armkel, LLC · Jul 30, 1996
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
K041238
B. Purpose for Submission:
Premarket notification
C. Measurand:
Human IgG antibodies to *Helicobacter pylori*
D. Type of Test:
Enzyme linked immunosorbent (ELISA) assay for qualitative detection of IgG antibodies to *Helicobacter pylori*
E. Applicant:
Biohit Plc
F. Proprietary and Established Names:
*Helicobacter pylori* IgG ELISA Test
G. Regulatory Information:
1. Regulation section:
866.3110
2. Classification:
I
3. Product code:
LYR - Campylobacter pylori
4. Panel:
{1}
83 Microbiology
H. Intended Use:
1. Intended use(s):
The *Helicobacter pylori* IgG ELISA Test is an enzyme linked immunosorbent (ELISA) assay for qualitative detection of human IgG class antibodies to *Helicobacter pylori* (H. pylori) in EDTA or heparin treated plasma and in serum.
2. Indication(s) for use:
The test is indicated for use as an aid in the diagnosis of H. pylori infection in adult patients with clinical symptoms of gastritis.
3. Special conditions for use statement(s):
Prescription use
4. Special instrument requirements:
Microplate reader, 450 nm
I. Device Description:
The H. pylori IgG antibodies ELISA test is an enzyme immunoassay technique utilizing a partially purified H. pylori antigen adsorbed on a polystyrene coated strip on a microplate and a detection antibody (Conjugate Solution) labeled with horseradish peroxidase (HRP) to detect antibodies to H. pylori in plasma or serum. The Substrate Solution contains an tetramethylbenzidine (TMB) which is oxygenated by the enzyme in the Conjugate to produce a blue end-product. The Stop Solution (with sulfuric acid) stops the enzyme reaction, and turns the substrate yellow if antibody is present in the patient serum. The kit also includes a Washing Buffer (phosphate), Diluent Buffer (phosphate) and four Incubation Covers (plastic sheets) to cover the microplate during incubation.
A Calibrator, containing 1.5 mL of human serum-based H. pylori IgG, is used as a functioning calibrator, to determine if the assay is functioning appropriately. Negative and Positive Controls made of human serum-based H. pylori IgG are included. The kit contains 12 microplates with 8 coated strips in a frame. The reagents are sufficient for 96 wells.
2
{2}
3
J. Substantial Equivalence Information:
1. Predicate device name(s):
IMMULITE® Helicobacter pylori IgG
2. Predicate 510(k) number(s):
k000463
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended use: | For detection of H. pylori IgG antibodies | Same |
| Assay | Qualitative | Same |
| Sample | Serum and/or Plasma | Serum |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Technology | ELISA | Chemiluminescent EIA |
| Incubation | 90 minutes | 65 minutes |
| Reader | Microplate Reader 450 nm | Luminometer |
K. Standard/Guidance Document Referenced (if applicable):
FDA document: Review Criteria for Assessment of Laboratory Tests for the Detection of Antibodies to Helicobacter pylori
L. Test Principle:
The H. pylori IgG antibodies ELISA test is based on an enzyme immunoassay technique utilizing a partially purified H. pylori antigen adsorbed on a microplate and a detection antibody labeled with horseradish peroxidase (HRP). If H. pylori antibodies are present in the sample they bind with the purified H. pylori antigen coated on the polystyrene wells of the microplate. Wells are then washed to remove residual particles. An HRP-conjugated monoclonal anti-human IgG is added to bind the H. pylori antibodies, the wells are washed and a TMB substrate is added. The substrate is oxygenated by the enzyme and a blue colored end product is produced. The enzyme reaction is terminated with the stop solution. The plates are read on a microplate reader. H. pylori positive samples turn yellow with EIU units of > 42.
{3}
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Reproducibility was established at three sites using a panel consisting of a high positive, 3 borderline positives, and 2 negative samples. Inter-assay and intra-assay variation was determined. For serum samples, % CV for inter-assay ranged from 1.5 -6.7%. EDTA-plasma% CV for inter-assay ranged from 2.5 - 7.0%, and heparin from 1.3 - 6.6%. For intra-assay, % CV ranged from 2.5 - 9.8 for serum, 1.3 - 11.8% for EDTA-plasma, and 1.3 - 19.2% for heparin. These values are similar to other ELISA H. pylori assays.
b. Linearity/assay reportable range:
Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Stability studies were conducted to stress the assay and samples.
Positive and Negative Controls are included in the kit and were evaluated during the studies. The Calibrator included is only for verifying that the assay is functioning properly. A value ≥ 1.000 is assigned
d. Detection limit:
The detection limit was determined by evaluating serial dilutions performed on serum, EDTA-plasma, and heparin-plasma samples. Values obtained gave a 10% CV limit and EIU values for serum of 4.0, for < 1.3 for EDTA-plasma, and 4.2 for heparin-plasma samples. Additionally, 24 zero blank replicates were assayed and gave values within a 2 standard deviation range. The mean value was 2.3 EIU.
e. Analytical specificity:
Cross-reactivity: To determine if the presence of closely related microorganisms would affect the specificity of the assay, a panel consisting of 4 strains of Yersinia, E. coli, Campylobacter jejuni, and 2 strains of Salmonella, all at concentrations >108 CFU/ml. There were no false positive or false negative results observed.
Interference Study: Assay was evaluated for triglycerides using ten samples spiked with triglyceride, and hemoglobin and bilirubin interference using 20
4
{4}
spiked samples. Reactivity or non-reactivity was not affected by increasing concentrations of none of these substances.
f. Assay cut-off:
The cut-off was determined in a case control study using 29 patient samples that had histology results available for normal stomach mucosa without evidence of H. pylori infection, or non-atrophic or superficial gastritis without H. pylori infection. The cut-off was further challenged in a study using histology and culture as the "gold" standard and another ELISA assay for comparison.
2. Comparison studies:
a. Method comparison with predicate device:
| Biohit H. pylori | Histology | |
| --- | --- | --- |
| | Pos | Neg |
| Pos | 144 | 79 |
| Neg | 34 | 247 |
| Percent Positive Agreement = 80.9% Percent Negative Agreement = 75.8% | | |
b. Matrix comparison:
Anticoagulant studies were conducted to determine if there were significant differences between when using serum, EDTA-plasma, and heparin-plasma specimens. No significant differences were observed between serum and EDTA; however heparin-plasma samples showed wider variation in results.
3. Clinical studies:
a. Clinical Sensitivity:
The Biohit H. pylori IgG antibodies ELISA test was evaluated for clinical sensitivity using a total of 264 samples. The samples were compared to of biopsied gastric mucosa culture. The results were as follows:
Culture
| Biohit H.P. Pos. Neg | 108 | 51 |
| --- | --- | --- |
| | 10 | 95 |
{5}
$$
\text{Clinical Sensitivity} = 91.5\%
$$
$$
(95\% \mathrm{CI} = 86.3 - 95.0\%)
$$
b. Clinical specificity:
$$
\text{Clinical Specificity} = 65.1\%
$$
$$
(95\% \mathrm{CI} - 60.8 - 68.0\%)
$$
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
The assay cutoff was validated during the clinical studies. Only patients with histology and culture procedures available were used in validating the reactive cutoff. The optimal cutoff was obtained using a Receiver-Operating Curve (ROC) analysis for maximizing the sum of the sensitivity and specificity of the assay. The cutoff was determined to be 38.5 EIU.
5. Expected values/Reference range:
In the studies conducted to demonstrate that this device is substantially equivalent to the predicate device, when compared to culture, the assay had a Predictive Positive Value of 67.9% and a Predictive Negative Value of 90.5%. However, antibody levels do not necessarily correlate to infection. H. pylori positive samples are expected to give values of >42 EIU.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
{6}
7
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.