← Product Code [LSR](/submissions/MI/subpart-d%E2%80%94serological-reagents/LSR) · K971169

# CAMBRIDGE BIOTECH HUMAN LYME IGG WESTERN BLOT (90111) (K971169)

_Cambridge Biotech Corp. · LSR · Feb 17, 1998 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/LSR/K971169

## Device Facts

- **Applicant:** Cambridge Biotech Corp.
- **Product Code:** [LSR](/submissions/MI/subpart-d%E2%80%94serological-reagents/LSR.md)
- **Decision Date:** Feb 17, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3830
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

The Cambridge Biotech Human Lyme B. burgdorferi IgG Western Blot is intended for use in testing human serum samples which have demonstrated positive or equivocal responses using EIA or IFA test procedures to provide supportive evidence of infection with Borrelia burgdorferi. The Cambridge Biotech Human Lyme B. burgdorferi IgG Western Blot can be used at any time following onset of symptoms. It should also be used for follow up when (1) only IgM antibodies were originally detected, (2) IgG antibodies were detected but were not considered significant, or (3) previously tested seronegative individuals are shown to develop antibodies by EIA or IFA test procedures. The assay is not intended for use as a screening assay, nor should it be used for analysis of serum from patients who have not demonstrated symptoms of Lyme Disease.

## Device Story

Western Blot assay for qualitative detection of human IgG antibodies to Borrelia burgdorferi. Input: human serum samples. Process: Borrelia proteins fractionated via SDS-PAGE, transferred to nitrocellulose membranes; strips incubated with patient serum; bound IgG detected via anti-human IgG alkaline phosphatase conjugate and chemical substrate cleavage. Output: visualized colored bands on nitrocellulose strips. Used in clinical laboratories; operated by trained technicians. Results compared against Band Identifying and Band Intensity Controls to determine reactivity. Provides supportive evidence of Lyme disease infection; aids clinical diagnosis in symptomatic patients.

## Clinical Evidence

Clinical trial evaluated 1062 samples for specificity and 287 samples for sensitivity. Specificity was 100% (95% CI: 99.8-100.0%). Sensitivity varied by disease stage and draw time: 21% for pre-treatment (<1 month), 19% for post-treatment (<1 month), and 88% for late-stage Lyme (2-12 months). Precision and reproducibility testing across multiple sites showed 100% agreement for IgG disposition and 94% agreement for band scoring.

## Technological Characteristics

Western Blot assay using Polyacrylamide Gel Electrophoresis (SDS-PAGE) and nitrocellulose membrane transfer. Detection via alkaline phosphatase-conjugated anti-human IgG and insoluble colored substrate. Qualitative visual interpretation based on band intensity controls.

## Regulatory Identification

Treponema pallidum treponemal test reagents are devices that consist of the antigens, antisera and all control reagents (standardized reagents with which test results are compared) which are derived from treponemal sources and that are used in the fluorescent treponemal antibody absorption test (FTA-ABS), the Treponema pallidum immobilization test (T.P.I.), and other treponemal tests used to identify antibodies to Treponema pallidum directly from infecting treponemal organisms in serum. The identification aids in the diagnosis of syphilis caused by bacteria belonging to the genus Treponema and provides epidemiological information on syphilis.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

R971169

Section 12 510(k) SUMMARY

Submitted By:

Cambridge Biotech Corporation
1500 East Gude Drive
Rockville, Maryland 20850-5307
(301) 251-0800, extension 145
(301) 762-1327 (fax)

FEB 17 1998

Contact Person:

Rebecca Leaper, Vice President - Operations, Responsible Head
(301) 251-0800, extension 105

Date of Preparation:

27 March 1997
Revised on 4 February, 1998

Product Name and Information

1. Name and Address of Owner/Operator, and Manufacturer

Owner/Operator:

Cambridge Biotech Corporation
A Wholly Owned Subsidiary of bioMérieux Vitek, Inc.
1500 East Gude Drive
Rockville, MD 20850-5307

Manufacturer:

Cambridge Biotech Corporation
A Wholly Owned Subsidiary of bioMérieux Vitek, Inc.
1500 East Gude Drive
Rockville, MD 20850-5307

{1}

2. Product Name

Trade Name: Cambridge Biotech Human Lyme Borrelia burgdorferi IgG Western Blot kit

Common Name: Lyme IgG Western Blot kit

Classification Name: Reagent, Borrelia Serological Reagent

3. Claim of Substantial Equivalence

The characterized samples used for the establishment of Substantial Equivalence have a clinical diagnosis of Lyme Disease based on the probability of exposure/infection (tick bite and/or patient presence in potential tick habitats in an endemic region within 30 days prior to the onset of EM (erythema migrans)), *Borrelia* isolation by culture (where possible), or, for non-EM patients, the presentation of Late Lyme clinical manifestations (e.g., cardiac, joint-involvement, or neurological symptoms).

Each of the clinical trial sites provided specimens that were well-characterized by the site using Lyme-specific serological analyses, including EIA and Western Blot testing.

Substantial equivalence of this device is based on the assessment of performance of the device in these clinical trials in which the well-characterized, archived Lyme Disease specimens, the Centers for Disease Control Lyme Disease Serum Panel, normal donor specimens (from endemic and non-endemic regions), and samples from diverse disease conditions were analyzed.

4. Description

The device is a Western Blot assay. Proteins and other antigenic components of the *Borrelia* spirochete are fractionated by Polyacrylamide Gel Electrophoresis in the presence of Sodium Dodecylsulfate. The separated proteins are electrophoretically transferred from the gel to nitrocellulose membranes, which are subsequently blocked to minimize non-specific binding and cut into strips. These nitrocellulose strips with *Borrelia burgdorferi* antigens are then reacted with diluted serum and controls (positive and negative sera of defined reactivity) during an incubation period.

During the incubation period, human antibodies specific to the *B. burgdorferi* antigens, if present in the sample or control, will bind to the

{2}

antigen to which they have affinity. Unbound serum and non-specific antibodies are washed from the strip. Detection of bound IgG antibodies is accomplished by reacting and incubating the strips with a solution containing anti-human IgG antibodies conjugated with alkaline phosphatase. Unbound conjugate antibodies are removed by washing. The qualitative assessment of the detected IgG antibodies is then accomplished by the reaction of the alkaline phosphatase with a chemical substrate, which is cleaved into a colored, insoluble product that can be visualized. The determination of the reactivity of each unknown specimen is accomplished by comparison of the identified, visualized bands to the Band Identifying and Band Intensity Controls.

## 5. Intended Use

The Cambridge Biotech Human Lyme B. burgdorferi IgG Western Blot is intended for use in testing human serum samples which have demonstrated positive or equivocal responses using EIA or IFA test procedures to provide supportive evidence of infection with Borrelia burgdorferi.

The Cambridge Biotech Human Lyme B. burgdorferi IgG Western Blot can be used at any time following onset of symptoms. It should also be used for follow up when (1) only IgM antibodies were originally detected, (2) IgG antibodies were detected but were not considered significant, or (3) previously tested seronegative individuals are shown to develop antibodies by EIA or IFA test procedures.

## 6. Performance Summary

The report of the complete clinical trial for the Cambridge Biotech Human Lyme IgG kit is contained in this section. Data for IgG and IgM have not been interpreted together, but separately, as will be required in clinical settings.

From a summary of the clinical trial data, the following performance characteristics are described:

### Specificity

Specificity of the device was determined from analysis of testing results of normal donor (from endemic and non-endemic regions) and disease specimens (1062 total samples) and was shown to be 100%, with 95% confidence intervals of 99.8% to 100.0%.

{3}

# Sensitivity

Sensitivity of the device was determined from analysis of test results of characterized Lyme disease specimens (287 total samples) that were drawn at different times after onset of disease:

## Sensitivity of the Cambridge Biotech Human Lyme B. burgdorferi IgGWestern Blot Relative to Lyme Disease Clinical Diagnosis and Treatment including Results by Draw Time

|  Disease Presentation | Draw Time (months) | Total Number of Specimens | Number By Draw Time | Number of Specimens Positive | Specimens Positive By Draw Time | Sensitivity | 95% CL  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  Before Treatment | <1 | 97 | 95 | 20 | 18 | 21% | 12.6-28.7%  |
|   |  1-2 |   | 2 |   | 2  |   |   |
|   |  2-12 |   | 0 |   | 0  |   |   |
|  After Treatment | <1 | 126 | 40 | 24 | 4 | 19% | 12.2-25.9%  |
|   |  1-2 |   | 48 |   | 13  |   |   |
|   |  2-12 |   | 38 |   | 7  |   |   |
|  Lyme Late | <1 | 64 | 0 | 56 | 0 | 88% | 78.8-96.2%  |
|   |  1-2 |   | 0 |   | 0  |   |   |
|   |  2-12 |   | 64 |   | 56  |   |   |

The Draw Time is the time from disease onset to specimen collection. The specimens were taken from 169 individual patients. Second draws were done on 92 of the 169 patients, third draws were done on 36 of those patients. The specimens from patients under antibiotic treatment may interfere with positive results. Ten specimens that were tested for IgM, were not tested for IgG due to insufficient sample.

# Precision

Six IgG Controls were tested in duplicate on each of three days at three test sites, totaling 18 replicates per control for all sites. All three sites were in 100% agreement for the disposition of the six IgG Controls. Additionally, all three sites were in 92% agreement as to the presentation of the IgG diagnostic band.

{4}

# Reproducibility

Ten positive and negative specimens from the CDC 47-member panel were tested at four sites. There was 100% agreement for the IgG disposition scores with the 20 specimens at all four sites. Eighty seven point five percent (87.5%) of the ten diagnostic bands of all 20 specimens were scored identically at all four test sites. In addition, a 94% overall agreement was shown between the scoring of positive IgG bands by the sites and the expected IgG band score results.

## Agreement with Expected Results Across Four Sites

|  Site | Number of Correct Bands | Percentage of Correct Bands | Correct Number of Interpretations | Percentage of Correct Interpretations  |
| --- | --- | --- | --- | --- |
|  Site 1 | 86/101 | 85% | 20/20 | 100%  |
|  Site 2 | 98/101 | 97% | 20/20 | 100%  |
|  Site 3 | 97/101 | 96% | 20/20 | 100%  |
|  Site 4 | 97/101 | 96% | 20/20 | 100%  |
|  Total | 378/404 | 94% | 80/80 | 100%  |

## 7. Determination of Threshold Intensity

The threshold determination was originally performed by analysis of IgG sensitivity and specificity panels. The intensity of weakly reactive Lyme antibody-positive samples (n = 5) were compared to the intensities of highly reactive Lyme antibody-negative samples (n = 5) in preclinical testing to demonstrate that the threshold appropriately differentiated positive and negative specimens. The negative specimens included normal donor samples only. Subsequent lots of the Control have been approved based on continued demonstration of this characteristic.

## 8. Conclusions

Based on the clinical performance, this device has been shown to be safe and effective for the intended use in the qualitative detection of human Immunoglobulin G (IgG) antibodies in serum or plasma to *Borrelia burgdorferi* antigens, and as a supplemental, more specific, test to aid in the diagnosis of infection or exposure to *Borrelia burgdorferi*, the causative agent of Lyme disease.

{5}

DEPARTMENT OF HEALTH &amp; HUMAN SERVICES
Public Health Service
Food and Drug Administration
2098 Gaither Road
Rockville MD 20850

Mr. Wole Edwin
Director of Quality Assurance
and Regulatory Affairs
Cambridge Biotech
1500 East Gude Drive
Rockville, MD 20850-5307
FEB 17 1998

Re: K971169
Trade Name: Cambridge Biotech Human Lyme Borrelia burgdorferi IgG Western Blot
Regulatory Class: II
Product Code: LSR
Dated: December 3, 1997
Received: December 3, 1997

Dear Mr. Edwin:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for under the Electronic Product Radiation Control provisions, or other Federal law regulations.

{6}

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven I. Gutman, M.D., M.B.A.
Director
Division of Clinical Laboratory Devices
Office of Device Evaluation
Center for Devices and Radiological Health

Enclosure

{7}

Section 8 Statement of Indications for Use

510(k) Number (if known): K971169

Device Name: Cambridge Biotech Human Lyme Borrelia burgdorferi IgG Western Blot

Indications For Use:

The Cambridge Biotech Human Lyme B. burgdorferi IgG Western Blot is intended for use in testing human serum samples which have demonstrated positive or equivocal responses using EIA or IFA test procedures to provide supportive evidence of infection with Borrelia burgdorferi.

The Cambridge Biotech Human Lyme B. burgdorferi IgG Western Blot can be used at any time following onset of symptoms. It should also be used for follow up when (1) only IgM antibodies were originally detected, (2) IgG antibodies were detected but were not considered significant, or (3) previously tested seronegative individuals are shown to develop antibodies by EIA or IFA test procedures.

The assay is not intended for use as a screening assay, nor should it be used for analysis of serum from patients who have not demonstrated symptoms of Lyme Disease.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

J. Ammo
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K971169

Concurrence of CDRH, Office Device Evaluation (ODE)

Prescription Use ☑
(Per 21 CFR 801.109)

OR

Over-The-Counter Use ☐

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/LSR/K971169](https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/LSR/K971169)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
