← Product Code [LSL](/submissions/MI/subpart-d%E2%80%94serological-reagents/LSL) · K132251

# APTIMA COMBO 2 ASSAY (PANTHER SYSTEM) (K132251)

_Hologic / Gen-Probe Incorporated · LSL · Oct 17, 2013 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/LSL/K132251

## Device Facts

- **Applicant:** Hologic / Gen-Probe Incorporated
- **Product Code:** [LSL](/submissions/MI/subpart-d%E2%80%94serological-reagents/LSL.md)
- **Decision Date:** Oct 17, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3390
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

The APTIMA Combo 2 Assay is a target amplification nucleic acid probe test that utilizes target capture for the in vitro qualitative detection and differentiation of ribosomal RNA (rRNA) from Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (GC) to aid in the diagnosis of chlamydial and/or gonococcal urogenital disease using the PANTHER System as specified. On the PANTHER System, the assay may be used to test the following specimens from symptomatic and asymptomatic individuals: clinician-collected endocervical, vaginal and male urethral swab specimens, clinician-collected gynecological specimens collected in the PreservCyt Solution, patient-collected vaginal swab specimens,¹ and male urine specimens. ¹ Patient-collected vaginal swab specimens are an option for screening women when a pelvic exam is not otherwise indicated. The vaginal swab specimen collection kit is not for home use.

## Device Story

Nucleic acid amplification test (NAAT) for qualitative detection/differentiation of Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (GC) rRNA; utilizes target capture, transcription-mediated amplification (TMA), and dual kinetic assay (DKA). Input: clinician-collected swabs (endocervical, vaginal, male urethral), PreservCyt liquid Pap, patient-collected vaginal swabs, male urine. Process: rRNA isolated via magnetic microparticles; specific target amplification; chemiluminescent DNA probes labeled with acridinium ester; luminometer measures photon signals (Relative Light Units). Operation: performed on PANTHER System in clinical labs; automated processing. Output: RLU measurements interpreted by software against established cutoffs to provide CT/GC results. Clinical decision-making: aids diagnosis of urogenital chlamydial/gonococcal disease; enables targeted patient treatment.

## Clinical Evidence

Two prospective, multicenter clinical studies evaluated performance. Study 1 included 580 males and 1332 females; Study 2 included 1492 males. Primary endpoints were sensitivity and specificity compared to an infected status algorithm using reference NAATs. For male urine, CT sensitivity was 95.2% (95% CI: 91.3-97.4%) and specificity was 99.8% (95% CI: 99.4-99.9%). GC sensitivity was 98.7% (95% CI: 92.9-99.8%) and specificity was 99.7% (95% CI: 99.3-99.9%).

## Technological Characteristics

Uses Target Capture (TC), Transcription-Mediated Amplification (TMA), and Hybridization Protection Assay (HPA). Detection via chemiluminescent acridinium ester-labeled DNA probes. Instrument: PANTHER System. Connectivity: Automated. Software: Embedded firmware for instrument control and signal processing. Sterilization: Not applicable (reagents).

## Regulatory Identification

Neisseria spp. direct serological test reagents are devices that consist of antigens and antisera used in serological tests to identify Neisseria spp. from cultured isolates. Additionally, some of these reagents consist of Neisseria spp. antisera conjugated with a fluorescent dye (immunofluorescent reagents) which may be used to detect the presence of Neisseria spp. directly from clinical specimens. The identification aids in the diagnosis of disease caused by bacteria belonging to the genus Neisseria, such as epidemic cerebrospinal meningitis, meningococcal disease, and gonorrhea, and also provides epidemiological information on diseases caused by these microorganisms. The device does not include products for the detection of gonorrhea in humans by indirect methods, such as detection of antibodies or of oxidase produced by gonococcal organisms.

## Predicate Devices

- APTIMA Combo 2® Assay (PANTHER® System) ([K111409](/device/K111409.md))

## Reference Devices

- APTIMA Combo 2 Assay ([K060652](/device/K060652.md))
- APTIMA CT Assay ([K061413](/device/K061413.md))
- APTIMA GC Assay ([K061509](/device/K061509.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
K132251

B. Purpose for Submission:
To obtain clearance for an additional specimen type, male urine on APTIMA Combo 2® Assay on the PANTHER® System

C. Measurand:
Chlamydia trachomatis and Neisseria gonorrhoeae ribosomal RNA

D. Type of Test:
Nucleic acid amplification assay

E. Applicant:
Hologic/Gen-Probe Incorporated

F. Proprietary and Established Names:
APTIMA Combo 2® Assay (PANTHER® System)
APTIMA Combo 2® Assay

G. Regulatory Information:
1. Regulation section:
21 CFR 866.3120 - Chlamydia serological reagents
21 CFR 866.3390 - Neisseria spp. direct serological test reagents
2. Classification:
Class II
3. Product code:

{1}

MKZ: DNA Probe, Nucleic Acid Amplification, Chlamydia
LSL: DNA-Reagents, Neisseria
NSU: Instrumentation for Clinical Multiplex Test Systems

4. Panel:
Microbiology (83)

H. Intended Use:

1. Intended use(s):

The APTIMA Combo 2 Assay is a target amplification nucleic acid probe test that utilizes target capture for the in vitro qualitative detection and differentiation of ribosomal RNA (rRNA) from Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (GC) to aid in the diagnosis of chlamydial and/or gonococcal urogenital disease using the PANTHER System as specified.

On the PANTHER System, the assay may be used to test the following specimens from symptomatic and asymptomatic individuals: clinician-collected endocervical, vaginal and male urethral swab specimens, clinician-collected gynecological specimens collected in the PreservCyt Solution, patient-collected vaginal swab specimens,¹ and male urine specimens.

¹ Patient-collected vaginal swab specimens are an option for screening women when a pelvic exam is not otherwise indicated. The vaginal swab specimen collection kit is not for home use.

2. Indication(s) for use:
Same as the Intended Use

3. Special conditions for use statement(s):
For prescription use only

4. Special instrument requirements:
Gen-Probe PANTHER System

I. Device Description:

The APTIMA Combo 2 Assay is a nucleic acid amplification test that qualitatively detects and differentiates ribosomal RNA (rRNA) from Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (GC) in the following specimens:
- Clinician-collected endocervical, vaginal, and male urethral swab specimens
- Male urine specimens

2

{2}

- Clinician-collected PreservCyt Solution liquid Pap specimens
- Patient-collected vaginal swab specimens

The current APTIMA Combo 2 Assay is similar to the APTIMA Combo 2 Assay originally cleared (ref: K111409) for use on the Panther System. It does not introduce any changes to the original design, method of manufacture, assay procedure, principle of operation, mechanism of action, conditions of use, hardware or software of the PANTHER instrument, or to the results interpretation for the cleared APTIMA Combo 2 Assay. There are no changes to the APTIMA Combo 2 Assay (PANTHER System) catalog numbers (303094 and 302923), accessories, or ancillary kits. The existing 'APTIMA Urine Specimen Collection Kit for Male and Female Urine Specimens' (catalog number 301040) will be added to the APTIMA Combo 2 Assay (PANTHER System) package insert as an ancillary kit. This collection kit is used with the cleared APTIMA Combo 2 Assay on the DTS System (K003395) and TIGRIS System (K032194).

The APTIMA Combo 2 Assay kit is comprised of 3 boxes:

- Refrigerated Box contains the Amplification Reagent, Enzyme Reagent, Probe Reagent, and Target Capture Reagent-B
- Room Temperature Box contains Amplification Reconstitution Solution, Enzyme Reconstitution Solution, Probe Reconstitution Solution, Selection Reagent and Target Capture Reagent
- Controls Box contains the Negative and Positive Controls

The APTIMA Combo 2 Assay on PANTHER utilizes three specimen collection kits.

- APTIMA Unisex Swab Specimen Collection Kit for Endocervical and Male Urethral Swab Specimens
- APTIMA Urine Collection Kit for Male and Female Urine Specimens (APTIMA Combo 2 Assay on the PANTHER System has not been cleared for use with female urine specimens)
- APTIMA Vaginal Swab Specimen Collection Kit
- APTIMA Specimen Transfer Kit (for use with gynecologic samples collected in PreservCyt Solution)

J. Substantial Equivalence Information:

1. Predicate device name(s):

APTIMA Combo 2® Assay (PANTHER® System)

2. Predicate 510(k) number(s):

K111409

{3}

3. Comparison with predicate:

|  Item | Predicate | Device  |
| --- | --- | --- |
|  Device Class | II | II  |
|  Regulation Specialty | Microbiology | Microbiology  |
|  Qualitative /Quantitative Assay | Qualitative | Qualitative  |
|  Indications For Use / Intended Use | The APTIMA Combo 2 Assay is a target amplification nucleic acid probe test that utilizes target capture for the in vitro qualitative detection and differentiation of ribosomal RNA (rRNA) from Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (GC) to aid in the diagnosis of chlamydial and/or gonococcal urogenital disease using the PANTHER System as specified. On the PANTHER System, the assay may be used to test the following specimens from symptomatic and asymptomatic individuals: clinician-collected endocervical, vaginal and male urethral swab specimens, clinician-collected gynecological specimens collected in the PreservCyt Solution, patient-collected vaginal swab specimens.1Patient-collected vaginal swab specimens are an option for screening women when a pelvic exam is not otherwise indicated. The vaginal swab specimen collection kit is not for home use. | The APTIMA Combo 2 Assay is a target amplification nucleic acid probe test that utilizes target capture for the in vitro qualitative detection and differentiation of ribosomal RNA (rRNA) from Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (GC) to aid in the diagnosis of chlamydial and/or gonococcal urogenital disease using the PANTHER System as specified. On the PANTHER System, the assay may be used to test the following specimens from symptomatic and asymptomatic individuals: clinician-collected endocervical, vaginal and male urethral swab specimens, clinician-collected gynecological specimens collected in the PreservCyt Solution, patient- collected vaginal swab specimens1, and male urine specimens.1Patient-collected vaginal swab specimens are an option for screening women when a pelvic exam is not otherwise indicated. The vaginal swab specimen collection kit is not for home use.  |

{4}

|  Specimen Types | Female specimens: •Vaginal swab •Endocervical swab •ThinPrep in PreservCyt solution Male Specimens: •Urethral Swab | Female specimens: •Vaginal swab •Endocervical swab •ThinPrep in PreservCyt solution Male Specimens: •Urethral Swab •Urine  |
| --- | --- | --- |
|  Specimen Transport/Storage | Swabs: After collection, transport and store swab in transport tube at 2-30°C and test within 60 days. If longer storage is desired, freeze at -20oC to -70oC for up to 365 days. | Swabs: Same  |
|   |  ThinPrep Liquid Pap in PreservCyt: Transport and store in PreservCyt solution at 2-30°C for up to 30 days. After transfer to APTIMA specimen transfer tube, store at 15-30°C for 14 days or store at 2-8°C for 30 days. If longer storage is desired, freeze at -20oC to -70oC for up to 365 days. | ThinPrep Liquid Pap in PreservCyt: Same  |
|   |  Urine not cleared. | Urine: After collection, transport the processed urine specimens in the APTIMA urine specimen transport tube at 2°C to 30°C and store at 2°C to 30°C until tested. Processed urine specimens should be assayed with the APTIMA Combo 2 Assay within 30 days of collection. If longer storage is needed, freeze at -20°C to -70°C for up to 12 months after collection.  |
|  Type of Assay | Nucleic Acid Amplification Test | Same  |
|  Technology | Target Capture (TC), Transcription-Mediated Amplification (TMA), Hybridization Protection Assay (HPA) | Same  |
|  Detection Format | HPA which provides relative light units (RLUs) that are assessed against an established assay cutoff | Same  |

K. Standard/Guidance Document Referenced (if applicable):

{5}

Establishing the Performance Characteristics of In Vitro Diagnostic Devices for Chlamydia trachomatis and/or Neisseria gonorrhea: Screening and Diagnostic Testing - Draft Guidance for Industry and FDA Staff, May 11, 2011.

EP5-A2, 2004 – Evaluation of Precision Performance of Quantitative Measurement Methods, CLSI Approved Guideline

EP15-A2, 2006 – User Verification of Performance for Precision and Trueness, CLSI Approved Guideline

L. Test Principle:

The APTIMA Combo 2 Assay involves the technologies of target capture, transcription-mediated amplification (TMA), and dual kinetic assay (DKA).

Specimens are collected and transferred into their respective specimen transport tubes. The transport solutions in these tubes release the rRNA targets and protect them from degradation during storage. When the APTIMA Combo 2 Assay is performed in the laboratory, the target rRNA molecules are isolated from specimens by use of capture oligomers via target capture that utilizes magnetic microparticles. The capture oligomers contain sequences complementary to specific regions of the target molecules as well as a string of deoxyadenosine residues. A separate capture oligomer is used for each target. During the hybridization step, the sequence specific regions of the capture oligomers bind to specific regions of the target molecules. The capture oligomer:target complex is then captured out of solution by decreasing the temperature of the reaction to room temperature. This temperature reduction allows hybridization to occur between the deoxyadenosine region on the capture oligomer and the poly-deoxythymidine molecules that are covalently attached to the magnetic particles. The microparticles, including the captured target molecules bound to them, are pulled to the side of the reaction vessel using magnets and the supernatant is aspirated. The particles are washed to remove residual specimen matrix that may contain amplification reaction inhibitors. After the target capture steps are completed, the specimens are ready for amplification.

Target amplification assays are based on the ability of complementary oligonucleotide primers to specifically anneal and allow enzymatic amplification of the target nucleic acid strands. The APTIMA Combo 2 Assay replicates a specific region of the 23S rRNA from Chlamydia trachomatis (CT) and a specific region of the 16S rRNA from Neisseria gonorrhea (GC) via DNA intermediates. A unique set of primers is used for each target molecule. Detection of the rRNA amplification product sequences (amplicon) is achieved using nucleic acid hybridization. Single-stranded chemiluminescent DNA probes, which are complementary to a region of each target amplicon, are labeled with different acridinium ester molecules. The labeled DNA probes combine with amplicon to form stable RNA:DNA hybrids. The Selection Reagent differentiates hybridized from unhybridized probe, eliminating the generation of signal from unhybridized probe. During the detection step, light emitted from the labeled RNA:DNA

6

{6}

hybrids is measured as photon signals in a luminometer, and are reported as Relative Light Units (RLU). In DKA, differences in the kinetic profiles of the CT and GC labeled probes allow for the differentiation of signal; kinetic profiles are derived from measurements of photon output during the detection read time. The chemiluminescent detection reaction for CT signal has very rapid kinetics and has the "flasher" kinetic type. The chemiluminescent detection reaction for GC signal is relatively slower and has the "glower" kinetic type. Assay results are determined by a cut-off based on the total RLU and the kinetic curve type.

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

#### a. Precision/Reproducibility:

**Within-Laboratory Repeatability:**
Within-laboratory repeatability was conducted previously to support the clearance of K111409. Please refer to decision summary of K111049.

**Reproducibility:**
Reproducibility of the APTIMA Combo 2 Assay on the PANTHER System was evaluated in two different studies using panel members created with Specimen Transport Medium in Reproducibility Study One and using panel members created with clinical urine specimens in Reproducibility Study Two.

**Reproducibility Study One**
APTIMA Combo 2 Assay reproducibility was evaluated with panel members created using Specimen Transport Medium at three external US laboratories using the PANTHER System. Testing was performed using one lot of assay reagents and a total of six operators (two at each site). Testing was performed over at least 10 days at each site. The negative panel member consisted of Specimen Transport Medium and positive panel members were created by spiking Specimen Transport Medium with lysate from CT and/or GC organisms to result in panel members with expected targeted concentrations. The table below shows the CT and GC concentrations for each panel member and the mean, standard deviation (SD), and coefficient of variation (CV) of the RLU data for each panel member between-sites, between-operators, between-days, between-runs, within-runs, and overall. Percent agreement with expected results is also shown. Only samples with valid results were included in the analyses.

7

{7}

Reproducibility Study One Data

|  Target Concentration |   | Agreed/N | Agmt (%) | Mean RLU (x1000) | Between-Sites |   | Between-Operators |   | Between-Days |   | Between-Runs |   | Within-Runs |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  CT (IFU/mL) | GC (CFU/mL) |   |   |   | SD (x1000) | CV (%) | SD (x1000) | CV (%) | SD (x1000) | CV (%) | SD (x1000) | CV (%) | SD (x1000) | CV (%) | SD (x1000) | CV (%)  |
|  0 | 0 | 180/180 | 100 | 6 | 1.0 | 17.5 | 0.5 | 8.1 | 0.2 | 3.7 | 0.5 | 8.2 | 1.5 | 24.4 | 1.9 | 32.4  |
|  0.25 | 0 | 180/180 | 100 | 1207 | 45.0 | 3.7 | 17.3 | 1.4 | 0.0 | 0.0 | 35.1 | 2.9 | 66.9 | 5.5 | 89.7 | 7.4  |
|  2.5 | 0 | 180/180 | 100 | 1272 | 41.3 | 3.2 | 19.2 | 1.5 | 0.0 | 0.0 | 31.0 | 2.4 | 36.8 | 2.9 | 66.3 | 5.2  |
|  25 | 0 | 180/180 | 100 | 1292 | 43.7 | 3.4 | 14.9 | 1.2 | 7.7 | 0.6 | 35.1 | 2.7 | 36.3 | 2.8 | 68.8 | 5.3  |
|  1000 | 0 | 180/180 | 100 | 1294 | 48.1 | 3.7 | 14.3 | 1.1 | 26.8 | 2.1 | 29.6 | 2.3 | 34.8 | 2.7 | 73.0 | 5.6  |
|  0 | 0.25 | 180/180 | 100 | 589 | 92.2 | 15.7 | 19.9 | 3.4 | 28.1 | 4.8 | 21.2 | 3.6 | 44.8 | 7.6 | 110.2 | 18.7  |
|  0 | 12.5 | 179/179 | 100 | 1251 | 163.5 | 13.1 | 0.0 | 0.0 | 15.1 | 1.2 | 31.5 | 2.5 | 29.8 | 2.4 | 169.8 | 13.6  |
|  0 | 125 | 180/180 | 100 | 1295 | 168.3 | 13.0 | 6.7 | 0.5 | 33.4 | 2.6 | 21.1 | 1.6 | 33.3 | 2.6 | 176.2 | 13.6  |
|  0 | 1250 | 180/180 | 100 | 1309 | 166.5 | 12.7 | 0.0 | 0.0 | 28.4 | 2.2 | 27.6 | 2.1 | 31.2 | 2.4 | 173.9 | 13.3  |
|  0 | 2500 | 179/179 | 100 | 1305 | 170.9 | 13.1 | 11.4 | 0.9 | 30.4 | 2.3 | 15.2 | 1.2 | 32.2 | 2.5 | 177.5 | 13.6  |
|  2.5 | 125 | 178/178 | 100 | 2513 | 123.9 | 4.9 | 24.6 | 1.0 | 24.0 | 1.0 | 57.5 | 2.3 | 52.4 | 2.1 | 150.3 | 6.0  |
|  2.5 | 2500 | 180/180 | 100 | 2515 | 123.5 | 4.9 | 6.5 | 0.3 | 33.8 | 1.3 | 39.3 | 1.6 | 59.4 | 2.4 | 146.6 | 5.8  |
|  1000 | 125 | 179/179 | 100 | 2524 | 117.4 | 4.6 | 35.2 | 1.4 | 52.1 | 2.1 | 28.9 | 1.1 | 54.7 | 2.2 | 146.8 | 5.8  |
|  1000 | 2500 | 180/180 | 100 | 2525 | 118.2 | 4.7 | 21.6 | 0.9 | 38.7 | 1.5 | 54.8 | 2.2 | 48.5 | 1.9 | 145.9 | 5.8  |

Agmt = agreement, CFU = colony-forming unit, CV = coefficient of variation, IFU = inclusion-forming unit, RLU = relative light unit, SD = standard deviation.
Note. Variability from some factors may be numerically negative, which can occur if the variability due to those factors is very small. When this occurs, the variability as measured with standard deviation and %CV is set to 0.

## Reproducibility Study Two

APTIMA Combo 2 Assay reproducibility was evaluated with panel members created using clinical urine specimens at two external US laboratories and in-house using the PANTHER System. Testing was performed using one lot of assay reagents and a total of six operators (two at each site). Testing was performed over at least 10 days at each site. The negative panel member consisted of negative urine and the positive panel members were created by spiking negative urine with lysate from CT and/or GC organisms to create panel members with the expected targeted concentrations. CT and GC concentrations for each panel member and the mean, SD, and CV of the RLU data for each panel member between-sites, between-operators, between-days, between-runs, within-runs, and overall are shown in the table below. Percent

{8}

agreement with expected results is also shown. Only samples with valid results were included in the analyses.

Reproducibility Study Two Data

|  Target Concentration |   | Agreed/N | Agmt (%) | Mean RLU (x1000) | Between-Sites |   | Between-Operators |   | Between-Days |   | Between-Runs |   | Within-Runs |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  CT (IFU/mL) | GC (CFU/mL) |   |   |   | SD (x1000) | CV (%) | SD (x1000) | CV (%) | SD (x1000) | CV (%) | SD (x1000) | CV (%) | SD (x1000) | CV (%) | SD (x1000) | CV (%)  |
|  0 | 0 | 178/180 | 98.9 | 6 | 1.2 | 19.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 8.2 | 131.7 | 8.3 | 133.0  |
|  0.25 | 0 | 180/180 | 100 | 1202 | 92.4 | 7.7 | 0.0 | 0.0 | 0.0 | 0.0 | 62.9 | 5.2 | 50.3 | 4.2 | 122.6 | 10.2  |
|  2.5 | 0 | 178/178 | 100 | 1185 | 90.9 | 7.7 | 0.0 | 0.0 | 0.0 | 0.0 | 53.8 | 4.5 | 34.6 | 2.9 | 111.1 | 9.4  |
|  25 | 0 | 180/180 | 100 | 1265 | 97.4 | 7.7 | 18.9 | 1.5 | 0.0 | 0.0 | 62.4 | 4.9 | 35.1 | 2.8 | 122.4 | 9.7  |
|  1000 | 0 | 180/180 | 100 | 1278 | 101.9 | 8.0 | 15.7 | 1.2 | 20.6 | 1.6 | 61.4 | 4.8 | 31.8 | 2.5 | 125.9 | 9.8  |
|  0 | 0.25 | 177/179 | 98.9 | 422 | 40.3 | 9.5 | 21.9 | 5.2 | 27.6 | 6.5 | 35.3 | 8.4 | 72.7 | 17.2 | 96.9 | 23.0  |
|  0 | 12.5 | 179/180 | 99.4 | 1142 | 11.9 | 1.0 | 0.0 | 0.0 | 44.4 | 3.9 | 37.3 | 3.3 | 75.8 | 6.6 | 96.2 | 8.4  |
|  0 | 125 | 180/180 | 100 | 1224 | 31.4 | 2.6 | 13.0 | 1.1 | 11.1 | 0.9 | 19.8 | 1.6 | 34.3 | 2.8 | 53.4 | 4.4  |
|  0 | 1250 | 180/180 | 100 | 1263 | 16.7 | 1.3 | 9.4 | 0.7 | 21.0 | 1.7 | 14.0 | 1.1 | 30.6 | 2.4 | 44.1 | 3.5  |
|  0 | 2500 | 180/180 | 100 | 1309 | 20.7 | 1.6 | 13.4 | 1.0 | 0.0 | 0.0 | 21.7 | 1.7 | 25.3 | 1.9 | 41.4 | 3.2  |
|  2.5 | 125 | 180/180 | 100 | 2468 | 71.9 | 2.9 | 31.5 | 1.3 | 21.7 | 0.9 | 64.8 | 2.6 | 44.4 | 1.8 | 113.1 | 4.6  |
|  2.5 | 2500 | 180/180 | 100 | 2453 | 76.2 | 3.1 | 30.9 | 1.3 | 0.0 | 0.0 | 62.5 | 2.5 | 51.6 | 2.1 | 115.4 | 4.7  |
|  1000 | 125 | 179/179 | 100 | 2504 | 74.0 | 3.0 | 38.5 | 1.5 | 0.0 | 0.0 | 59.1 | 2.4 | 39.1 | 1.6 | 109.4 | 4.4  |
|  1000 | 2500 | 180/180 | 100 | 2357 | 79.1 | 3.4 | 0.0 | 0.0 | 0.0 | 0.0 | 74.2 | 3.1 | 55.2 | 2.3 | 121.7 | 5.2  |

Agmt = agreement, CFU = colony-forming unit, CV = coefficient of variation, IFU = inclusion-forming unit, RLU = relative light unit, SD = standard deviation.
Note. Variability from some factors may be numerically negative, which can occur if the variability due to those factors is very small. When this occurs, the variability as measured with standard deviation and  $\% \mathrm{CV}$  is set to 0.

b. Linearity/assay reportable range:

Not Applicable

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Freeze-Thaw Stability

The purpose of this study was to evaluate the sensitivity and specificity of the APTIMA Combo 2 Assay on the PANTHER System using urine samples that were frozen and thawed prior to testing. Paired fresh (non-frozen) and frozen urine samples were evaluated. The urine specimens were diluted 1:1 with urine transport

{9}

medium (UTM) and then split into two aliquots to generate negative and positive samples:

1. One aliquot was left un-spiked (negative sample).
2. One aliquot was spiked with 0.25 IFU/mL of cultured CT serovar E and 12.5 CFU/mL of cultured GC strain 49226 (CT/GC-positive sample).

The un-spiked and spiked urine samples were split into two aliquots to test fresh (non-frozen) and frozen sample storage conditions:

1. Fresh samples: One spiked and one un-spiked aliquot of each urine sample was stored at  $2^{\circ}\mathrm{C}$  to  $30^{\circ}\mathrm{C}$  until tested.
2. Frozen samples: One spiked and one un-spiked aliquot of each sample was frozen and stored at  $\leq -20^{\circ}\mathrm{C}$  for at least 72 hours (to ensure samples were completely frozen) until tested.

The results from this study demonstrate equivalency between urine samples tested fresh and paired samples tested after being frozen and thawed.

# Quality Control Results and Acceptability

The Positive Control, CT / Negative Control, GC and the Positive Control, GC / Negative Control, CT act as controls for the target capture, amplification, and detection steps of the assay. The Positive Control, CT / Negative Control, GC serves as the negative control for the GC test results. The Positive Control, GC / Negative Control, CT serves as the negative control for the CT test results. If desired, a dual negative control furnished by the user can be added to monitor assay background. Correct preparation of specimens is confirmed visually by the presence of a single APTIMA collection swab in a swab specimen transport tube or the absence of a swab in an APTIMA specimen transfer tube for PreservCyt liquid Pap specimens.

The APTIMA Combo 2 Assay Controls must produce the following test results:

|  Control | Total RLU (x1000) | CT Result | GC Result  |
| --- | --- | --- | --- |
|  Positive Control, CT /Negative Control, GC | ≥ 100 and < 3,000 | Positive | Negative  |
|  Positive Control, GC /Negative Control, CT | ≥ 150 and < 3,000 | Negative | Positive  |

{10}

d. Detection limit:

The Limit of Detection (LoD) study included a series of dilution panels of CT (serovars E and G) and GC (ATCC strain 49226 and 19424) prepared in urine and in specimen transport medium (STM) to achieve several test concentrations. The positivity, with the two-sided 95% Score confidence interval, was calculated for each test concentration of each panel. Probit analysis was performed on the positivity results of each panel to determine the predicted 95% detection limit. The LoDs in both STM and urine sample matrices are well below the assay sensitivity claims for CT (1 IFU/assay) and GC (50 cells/assay).

The LoD of the APTIMA Combo 2 Assay on the PANTHER system in both urine and STM sample matrices was verified for 12 serovars of CT and 30 strains of GC. Dilutions of less than 2.5 IFU/mL tested positive in the APTIMA Combo 2 Assay for the following 12 serovars: D, E, F, G, H, I, J, K, L1, L2, L2a, and L3 (≥95% positivity was observed in samples containing CT concentrations of 1.89 IFU/mL). Dilutions of less than 125 CFU/mL tested positive in the APTIMA Combo 2 Assay for 30 different strains of GC (≥95% positivity was observed in samples containing GC concentrations of 0.36 cells/mL).

e. Analytical specificity:

Please refer to the decision summary of K111049 for cross reactivity data.

f. Interference:

Please refer to the decision summary of K111049.

g. Carryover Study:

The purpose of this study was to determine the carryover rate of the APTIMA Combo 2 Assay on the PANTHER System when high titer positive samples were alternately processed with negative samples in a checkerboard format. Six runs were performed on one PANTHER instrument using one reagent lot for three days. The first run comprised of 135 negative samples was performed to establish the baseline false positive rate. The five subsequent runs were comprised of 54 samples (GC-positive and GC-negative samples) alternated in checkerboard format. High titer positive samples for this study were made by spiking GC rRNA into specimen transport media (STM) to give a final concentration equivalent to ~2 x 10⁵ cells/mL. The false positive rate from the baseline run without any positive samples (Run 1) was 0% (0/135; 100% specificity). The false positive rate from the runs with positive samples in the checkerboard design (Runs 2-6) was 0.74% (1/135; 99.26% specificity). The carryover rate was 0.74% with a 95% CI of -2.03 to 4.10.

11

{11}

# h. Assay cut-off:

Assay test results are automatically interpreted by the APTIMA Assay software, using the APTIMA Combo 2 protocol. A test result may be a negative, equivocal, positive, or invalid as determined by the kinetic type and total RLU in the detection step. A test result may be invalid due to a parameter outside the normal expected ranges. Initial equivocal and invalid test results should be retested.

|  Kinetic Type | Total RLU (x1000) to give CT Result  |   |   |
| --- | --- | --- | --- |
|   |  Negative | Equivocal | Positive  |
|  CT only | 1 to < 25 | 25 to < 100 | 100 to < 4,500  |
|  CT and GC | 1 to < 85 | 85 to < 250 | 250 to < 4,500  |
|  CT indeterminate | 1 to < 85 | 85 to < 4,500 | N/A  |
|  Kinetic Type | Total RLU (x1000) to give GC Result  |   |   |
| --- | --- | --- | --- |
|   |  Negative | Equivocal | Positive  |
|  GC only | 1 to < 60 | 60 to < 150 | 150 to < 4,500  |
|  GC and CT | 1 to < 85 | 85 to < 250 | 250 to < 4,500  |
|  GC indeterminate | 1 to < 85 | 85 to < 4,500 | N/A  |

# 2. Comparison studies:

a. Method comparison with predicate device:

The comparison studies were conducted in two clinical studies with prospectively collected samples (see the Clinical studies section below).

b. Matrix comparison:

Not applicable

# 3. Clinical studies:

Two clinical studies were performed. APTIMA Combo 2 Assay clinical performance was estimated with male urethral swab, vaginal swab, PreservCyt Solution liquid Pap, and endocervical swab specimens in Clinical Study One, and with male urine specimens in Clinical Study Two.

{12}

# Clinical Study One: Vaginal Swab, PreservCyt Solution Liquid Pap, Female Endocervical Swab, and Male Urethral Swab Specimen Clinical Study

A prospective, multi-center clinical study was conducted to establish the performance characteristics of the APTIMA Combo 2 Assay on the PANTHER System. Specimens were collected from symptomatic and asymptomatic men (n=580) and women (n=1332) enrolled from seven geographically and ethnically diverse US clinical sites, including obstetrics and gynecology, family planning, public health, and STD clinics. Subjects were classified as symptomatic if symptoms were reported by the subject. Subjects were classified as asymptomatic if the subject did not report symptoms. Of the 580 male subjects, none were &lt;18 years of age, 72 were 18 to 20 years of age, 201 were 21 to 25 years of age, and 307 were &gt;25 years of age. Of the 1332 female subjects, 11 were 14 to 15 years of age, 59 were 16 to 17 years of age, 319 were 18 to 20 years of age, 401 were 21 to 25 years of age, and 542 were &gt;25 years of age.

Up to two specimens were collected from each male subject (one urethral swab and one first-catch urine, in that order) and up to four specimens were collected from each female subject (one first-catch urine, one vaginal swab, one PreservCyt Solution liquid Pap specimen, and one endocervical swab, in that order). All specimens were clinician-collected except urine specimens and approximately half of the vaginal swab specimens, which were collected by the subject at the clinic. Approximately half of the PreservCyt Solution liquid Pap specimens were collected with a broom-type device and half were collected with a spatula and cytobrush. Samples were prepared for APTIMA testing in accordance with the appropriate APTIMA specimen collection kit package insert instructions.

All evaluable samples (567 male urethral swab, 580 male urine, 1319 vaginal swab, 1330 PreservCyt Solution liquid Pap, and 1310 endocervical swab samples) were tested with the APTIMA Combo 2 Assay on the PANTHER System in accordance with package insert instructions. The samples were split among three laboratories (two external laboratories and in-house). Samples with initial invalid, equivocal, or error results were retested. Eighteen (18) male urethral swab, 25 vaginal swab, one PreservCyt Solution liquid Pap, and 37 endocervical swab samples had final invalid results and were excluded from the analyses. Most of the invalid results were due to insufficient sample volume. One vaginal swab, and one endocervical swab had final CT equivocal results and one PreservCyt Solution liquid Pap sample, and one endocervical swab had final GC equivocal results and were excluded from the analyses.

Male urethral swab, male and female urine, and PreservCyt Solution liquid Pap samples were tested with cleared nucleic acid amplification tests (NAATs) to establish the infected status. The infected status algorithm used results from two specimen types and two reference NAATs. Subjects were categorized as infected if a positive result occurred in each of the two reference NAATs (see Tables 8, 9, 11, and 12 for the infected status algorithms). For female subjects, if the positive NAAT results occurred only in the urine specimens and not in the PreservCyt Solution liquid Pap specimens, the subject was categorized as infected; however, for the evaluation of the non-urine specimen types, the

13

{13}

specimens were considered non-infected. Subjects that could not be categorized as infected or not infected were excluded from the performance analyses.

In addition, male urine samples tested with the APTIMA Combo 2 Assay on the PANTHER System were excluded from the performance analyses due to the low prevalence of GC in the study population, particularly in the asymptomatic subjects.

## Clinical Study Two: Male Urine Specimen Clinical Study

A prospective, multi-center clinical study was conducted to establish the performance characteristics of the APTIMA Combo 2 Assay on the PANTHER System in male urine specimens. Specimens were collected from symptomatic and asymptomatic men (n=1492) enrolled from 13 geographically and ethnically diverse US clinical research sites, and family planning, public health, men's health, and STD clinics. Subjects were classified as symptomatic if symptoms were reported by the subject. Subjects were classified as asymptomatic if the subject did not report symptoms. Of the 1492 subjects enrolled, 14 were withdrawn.

Two specimens were collected from each subject (one urethral swab and one first-catch urine, in that order). The urethral swab specimens were clinician-collected, and urine specimens were collected by the subject at the clinic. Urine specimens from each subject were processed into multiple samples for CT/GC testing with different NAATs in accordance with the instructions in the appropriate specimen collection kit package insert. The male urine samples for APTIMA Combo 2 Assay testing on the PANTHER System were split among three external laboratories.

All 1478 male urine samples from non-withdrawn subjects were tested with the APTIMA Combo 2 Assay on the PANTHER System in accordance with the APTIMA Combo 2 Assay package insert instructions. Samples with initial invalid, equivocal, or error results were retested. One male urine sample had a final invalid result and was excluded from the analyses. The invalid result was due to insufficient sample volume. Of the remaining 1477 evaluable male subjects, 46 were 16 to 17 years of age, 155 were 18 to 20 years of age, 524 were 21 to 30 years of age, 279 were 31 to 40 years of age, and 473 were &gt;40 years of age.

Male urethral swab and urine samples were tested with cleared NAATs to establish the infected status (see Tables 10 and 13 for the infected status algorithms). The infected status algorithm used urethral swab and urine sample results from one reference CT and GC NAAT and urine sample results from two additional reference CT and GC NAATs to generate four reference results for each analyte. Subjects were categorized as infected if a positive result occurred in at least two of the reference NAATs. Subjects that could not be categorized as infected or not infected were excluded from the performance analyses; 1 subject had an indeterminate CT infected status and was excluded from the performance analyses for detection of CT.

14

{14}

# Chlamydia trachomatis Performance Results

Performance characteristics of the APTIMA Combo 2 Assay for CT detection were estimated for each specimen type and are displayed in Tables 4 and 5 combining data from the two clinical studies. Performance was calculated by comparing PANTHER System results to an infected status algorithm, which differed between the two clinical studies (see Tables 8 through 10 for the CT infected status algorithms). Table 4 shows the sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of the APTIMA Combo 2 Assay for CT detection and the prevalence of CT (based on the infected status) in each specimen type.

Table 4: Performance Characteristics of the APTIMA Combo 2 Assay for CT Detection

|  Specimen Type1 | n | TP | FP | TN | FN | Prev % | Sensitivity % (95% CI)1 | Specificity % (95% CI)1 | PPV % (95% CI)1 | NPV % (95% CI)1  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  CVS/PVS | 1274 | 104 | 18 | 1149 | 3 | 8.4 | 97.2 (92.1-99.0) | 98.5 (97.6-99.0) | 85.2 (78.8-90.5) | 99.7 (99.3-99.9)  |
|  PCyt | 1311 | 112 | 0 | 1197 | 2 | 8.7 | 98.2 (93.8-99.5) | 100 (99.7-100) | 100 (96.9-100) | 99.8 (99.4-100)  |
|  FS | 1254 | 104 | 8 | 1139 | 3 | 8.5 | 97.2 (92.1-99.0) | 99.3 (98.6-99.6) | 92.9 (87.1-96.7) | 99.7 (99.3-99.9)  |
|  MS | 549 | 100 | 4 | 445 | 0 | 18.2 | 100 (96.3-100) | 99.1 (97.7-99.7) | 96.2 (90.8-98.9) | 100 (99.2-100)  |
|  MU | 1799 | 197 | 3 | 1589 | 10 | 11.5 | 95.2 (91.3-97.4) | 99.8 (99.4-99.9) | 98.5 (95.8-99.7) | 99.4 (98.9-99.7)  |

CI = confidence interval, CVS = clinician-collected vaginal swab, FN = false negative, FP = false positive, FS = female endocervical swab, MS = male urethral swab, MU = male urine, NPV = negative predictive value, PCyt = PreservCyt Solution liquid Pap, PPV = positive predictive value, Prev = prevalence, PVS = patient-collected vaginal swab, TN = true negative, TP = true positive.
1 Male urethral swab, vaginal swab, PreservCyt Solution liquid Pap, and endocervical swab sample results are from Clinical Study One, symptomatic male urine sample results are from Clinical Study Two, and asymptomatic male urine sample results are from both studies.
Score CI.
1 PPV  $95\%$  CI computed from the exact  $95\%$  CI for the positive likelihood ratio, NPV  $95\%$  CI computed from the exact  $95\%$  CI from the negative likelihood ratio.

Table 5 shows the sensitivity, specificity, PPV, and NPV of the APTIMA Combo 2 Assay for CT detection and the prevalence of CT (based on the infected status) in each specimen type by symptom status. CT prevalence was higher in symptomatic men and women.

Table 5: Performance Characteristics of the APTIMA Combo 2 Assay for CT Detection by Symptom Status

|  Specimen Type1 | Symptom Status | n | TP | FP | TN | FN | Prev % | Sensitivity % (95% CI)2 | Specificity % (95% CI)2 | PPV % (95% CI)3 | NPV % (95% CI)3  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  CVS/PVS | Sym | 810 | 73 | 8 | 729 | 0 | 9.0 | 100 (95.0-100) | 98.9 (97.9-99.4) | 90.1 (82.3-95.5) | 100 (99.5-100)  |
|   |  Asym | 464 | 31 | 10 | 420 | 3 | 7.3 | 91.2 (77.0-97.0) | 97.7 (95.8-98.7) | 75.6 (63.1-86.2) | 99.3 (98.1-99.8)  |

{15}

Asym = asymptomatic, CI = confidence interval, CVS = clinician-collected vaginal swab, FN = false negative, FP = false positive, FS = female endocervical swab, MS = male urethral swab, MU = male urine, NPV = negative predictive value, PCyt = PreservCyt Solution liquid Pap, PPV = positive predictive value, Prev = prevalence, PVS = patient-collected vaginal swab, Sym = symptomatic, TN = true negative, TP = true positive.

Male urethral swab, vaginal swab, PreservCyt Solution liquid Pap, and endocervical swab sample results are from Clinical Study One, symptomatic male urine sample results are from Clinical Study Two, and asymptomatic male urine sample results are from both studies Score CI.
3 PPV  $95\%$  CI computed from the exact  $95\%$  CI for the positive likelihood ratio, NPV  $95\%$  CI computed from the exact  $95\%$  CI from the negative likelihood ratio.

# Neisseria gonorrhoeae Performance Results

Performance characteristics of the APTIMA Combo 2 Assay for GC detection were estimated for each specimen type and are displayed in Tables 6 and 7 combining data from the two clinical studies. The infected status algorithm differed between the two clinical studies (see Tables 11 through 13 for the GC infected status algorithms). Table 6 shows the sensitivity, specificity, PPV, and NPV of the APTIMA Combo 2 Assay for GC detection and the prevalence of GC (based on the infected status) in each specimen type.

Table 6: Performance Characteristics of the APTIMA Combo 2 Assay for GC Detection

|  Specimen Type1 | N | TP | FP | TN | FN | Prev % | Sensitivity % (95% CI)2 | Specificity % (95% CI)2 | PPV % (95% CI)3 | NPV % (95% CI)3  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  CVS/PVS | 1258 | 42 | 5 | 1210 | 1 | 3.4 | 97.7 (87.9-99.6) | 99.6 (99.0-99.8) | 89.4 (78.6-96.1) | 99.9 (99.6-100)  |
|  PCyt | 1293 | 43 | 0 | 1250 | 0 | 3.3 | 100 (91.8-100) | 100 (99.7-100) | 100 (92.1-100) | 100 (99.7-100)  |
|  FS | 1238 | 42 | 2 | 1194 | 0 | 3.4 | 100 (91.6-100) | 99.8 (99.4-100) | 95.5 (85.4-99.4) | 100 (99.7-100)  |
|  MS | 546 | 34 | 0 | 512 | 0 | 6.2 | 100 (89.8-100) | 100 (99.3-100) | 100 (90.2-100) | 100 (99.3-100)  |
|  MU | 1797 | 75 | 5 | 1716 | 1 | 4.2 | 98.7 (92.9-99.8) | 99.7 (99.3-99.9) | 93.8 (86.7-97.8) | 99.9 (99.7-100)  |

{16}

CI = confidence interval, CVS = clinician-collected vaginal swab, FN = false negative, FP = false positive, FS = female endocervical swab, MS = male urethral swab, MU = male urine, NPV = negative predictive value, PCyt = PreservCyt Solution liquid Pap, PPV = positive predictive value, Prev = prevalence, PVS = patient-collected vaginal swab, TN = true negative, TP = true positive.

Vaginal swab, PreservCyt Solution liquid Pap, endocervical swab, and male urethral swab sample results are from Clinical Study One, symptomatic male urine sample results are from Clinical Study Two, and asymptomatic male urine sample results are from both studies.
2 Score CI.
3 PPV  $95\%$  CI computed from the exact  $95\%$  CI for the positive likelihood ratio, NPV  $95\%$  CI computed from the exact  $95\%$  CI from the negative likelihood ratio.

Table 7 shows the sensitivity, specificity, PPV, and NPV of the APTIMA Combo 2 Assay for GC detection and the prevalence of GC (based on the infected status) in each specimen type by symptom status. GC prevalence was higher in symptomatic men but similar in symptomatic and asymptomatic women.

Table 7: Performance Characteristics of the APTIMA Combo 2 Assay for GC Detection by Symptom Status

|  Specimen Type1 | Symptom Status | n | TP | FP | TN | FN | Prev % | Sensitivity % (95% CI)2 | Specificity % (95% CI)2 | PPV % (95% CI)3 | NPV % (95% CI)3  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  CVS/PVS | Sym | 802 | 27 | 4 | 771 | 0 | 3.4 | 100 (87.5-100) | 99.5 (98.7-99.8) | 87.1 (72.6-96.1) | 100 (99.6-100)  |
|   |  Asym | 456 | 15 | 1 | 439 | 1 | 3.5 | 93.8 (71.7-98.9) | 99.8 (98.7-100) | 93.8 (74.0-99.8) | 99.8 (98.9-100)  |
|  PCyt | Sym | 829 | 27 | 0 | 802 | 0 | 3.3 | 100 (87.5-100) | 100 (99.5-100) | 100 (88.0-100) | 100 (99.6-100)  |
|   |  Asym | 464 | 16 | 0 | 448 | 0 | 3.4 | 100 (80.6-100) | 100 (99.1-100) | 100 (81.3-100) | 100 (99.3-100)  |
|  FS | Sym | 785 | 26 | 1 | 758 | 0 | 3.3 | 100 (87.1-100) | 99.9 (99.3-100) | 96.3 (82.4-99.9) | 100 (99.5-100)  |
|   |  Asym | 453 | 16 | 1 | 436 | 0 | 3.5 | 100 (80.6-100) | 99.8 (98.7-100) | 94.1 (74.3-99.8) | 100 (99.3-100)  |
|  MS | Sym | 236 | 31 | 0 | 205 | 0 | 13.1 | 100 (89.0-100) | 100 (98.2-100) | 100 (89.5-100) | 100 (98.3-100)  |
|   |  Asym | 310 | 3 | 0 | 307 | 0 | 1.0 | 100 (43.9-100) | 100 (98.8-100) | 100 (44.4-100) | 100 (99.3-100)  |
|  MU | Sym | 497 | 66 | 1 | 430 | 0 | 13.3 | 100 (94.5-100) | 99.8 (98.7-100) | 98.5 (92.3-100) | 100 (99.2-100)  |
|   |  Asym | 1300 | 9 | 4 | 1286 | 1 | 0.8 | 90.0 (59.6-98.2) | 99.7 (99.2-99.9) | 69.2 (45.6-91.7) | 99.9 (99.7-100)  |

Asym = asymptomatic, CI = confidence interval, CVS = clinician-collected vaginal swab, FN = false negative, FP = false positive, FS = female endocervical swab, MS = male urethral swab, MU = male urine, NPV = negative predictive value, PCyt = PreservCyt Solution liquid Pap, PPV = positive predictive value, Prev = prevalence, PVS = patient-collected vaginal swab, Sym = symptomatic, TN = true negative, TP = true positive.
1 Vaginal swab, PreservCyt Solution liquid Pap, endocervical swab, and male urethral swab sample results are from Clinical Study One, symptomatic male urine sample results are from Clinical Study Two, and asymptomatic male urine sample results are from both studies.
2 Score CI.
3 PPV  $95\%$  CI computed from the exact  $95\%$  CI for the positive likelihood ratio, NPV  $95\%$  CI computed from the exact  $95\%$  CI from the negative likelihood ratio.

{17}

# Chlamydia trachomatis Infected Status Tables

The frequency of test outcomes from reference NAAT and investigational PANTHER System testing is summarized in Tables 8 through 10 for CT.

Table 8: Clinical Study One. CT Infected Status for Performance Evaluation in Female Vaginal Swab, PreservCyt Solution Liquid Pap, and Endocervical Swab Samples

|  CT Infected Status | Assay Results |   |   |   |   |   |   | Symptom Status  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  AC2 TIGRIS |   | ACT TIGRIS |   | AC2 PANTHER  |   |   |   |   |
|   |  PCyt | FU | PCyt | FU | CVS/PVS | PCyt | FS | Sym | Asym  |
|  Infected | + | + | + | + | + | + | + | 62 | 26  |
|  Infected | + | + | + | + | + | + | - | 0 | 1  |
|  Infected | + | + | + | + | + | + | NA | 3 | 0  |
|  Infected | + | + | + | + | + | - | + | 0 | 2  |
|  Infected | + | + | + | + | - | + | + | 0 | 1  |
|  Infected | + | + | + | + | NA | + | + | 1 | 1  |
|  Infected | + | + | + | + | NA | + | NA | 2 | 1  |
|  Infected | + | - | + | + | + | + | + | 4 | 1  |
|  Infected | + | - | + | + | NA | + | NA | 0 | 1  |
|  Infected | + | - | + | - | + | + | + | 4 | 0  |
|  Infected | + | - | + | - | - | + | - | 0 | 1  |
|  Infected | + | - | + | - | NA | + | + | 0 | 1  |
|  Infected | + | NA | + | NA | + | + | + | 0 | 1  |
|  Infected | + | NA | + | NA | - | + | - | 0 | 1  |
|  Infected1 | - | + | - | + | + | - | + | 1 | 0  |
|  Infected1 | - | + | - | + | + | - | - | 2 | 0  |
|  Infected1 | - | + | - | + | - | - | - | 1 | 1  |
|  Not Infected | + | - | - | - | - | - | - | 0 | 2  |
|  Not Infected | - | + | - | - | - | - | - | 1 | 0  |
|  Not Infected | - | - | + | - | + | - | + | 0 | 1  |
|  Not Infected | - | - | + | - | - | - | - | 5 | 0  |
|  Not Infected | - | - | - | + | + | - | - | 0 | 1  |
|  Not Infected | - | - | - | + | + | - | NA | 0 | 1  |
|  Not Infected | - | - | - | + | - | - | - | 1 | 3  |

{18}

Table 9: Clinical Study One. CT Infected Status for Performance Evaluation in Male Urethral Swab Samples

|  CT Infected Status | Assay Results |   |   |   |   |   | Symptom Status  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  AC2 DTS |   | ACT TIGRIS |   | AC2 PANTHER  |   |   |   |
|   |  MS | MU | MS | MU | MS | Sym | Asym |   |
|  Infected | + | + | + | + | + | 50 | 37 |   |
|  Infected | + | + | + | + | NA | 4 | 1 |   |
|  Infected | + | + | + | - | + | 2 | 0 |   |
|  Infected | + | - | + | + | + | 4 | 2 |   |
|  Infected | + | - | + | - | + | 3 | 2 |   |
|  Not Infected | + | + | - | - | - | 0 | 1 |   |
|  Not Infected | + | - | - | - | + | 0 | 1 |   |
|  Not Infected | + | - | - | - | - | 1 | 1 |   |
|  Not Infected | - | - | + | - | - | 3 | 2 |   |
|  Not Infected | - | - | - | + | - | 1 | 1 |   |

{19}

Table 10: Clinical Study One and Clinical Study Two. CT Infected Status for Performance Evaluation in Male Urine Samples

|  CT Infected Status | Assay Results |   |   |   |   |   |   | Symptom Status  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  AC21 |   | ACT TIGRIS2 |   | NAAT 13 |   | NAAT 23  |   |   |
|   |  MS | MU | MS | MU | MU | MU | MU | Sym | Asym  |
|  Clinical Study One  |   |   |   |   |   |   |   |   |   |
|  Infected | + | + | + | + |  |  | + |  | 38  |
|  Infected | + | - | + | + |  |  | + |  | 2  |
|  Infected | + | - | + | - |  |  | - |  | 2  |
|  Clinical Study Two  |   |   |   |   |   |   |   |   |   |
|  Infected | + | + |  |  | + | + | + | 73 | 66  |
|  Infected | + | + |  |  | + | + | - | 2 | 1  |
|  Infected | + | + |  |  | + | - | + | 0 | 1  |
|  Infected | + | + |  |  | + | NA | + | 0 | 1  |
|  Infected | + | + |  |  | - | + | + | 3 | 0  |
|  Infected | + | + |  |  | - | + | - | 0 | 1  |
|  Infected | + | - |  |  | + | + | + | 4 | 0  |
|  Infected | + | - |  |  | + | + | - | 3 | 0  |
|  Infected | + | = |  |  | - | + | - | 0 | 1  |
|  Infected | - | + |  |  | + | + | + | 5 | 4  |
|  Clinical Study One  |   |   |   |   |   |   |   |   |   |
|  Not Infected | + | + | - | - |  |  | - |  | 1  |
|  Not Infected | + | - | - | - |  |  | - |  | 2  |
|  Not Infected | - | - | + | - |  |  | - |  | 2  |
|  Not Infected | - | - | - | + |  |  | + |  | 1  |

{20}

# Neisseria gonorrhoeae Infected Status Tables

The frequency of test outcomes from reference NAAT and investigational PANTHER System testing is summarized in Tables 11 through 13 for GC.

Table 11: Clinical Study One. GC Infected Status for Performance Evaluation in Female Vaginal Swab, PreservCyt Solution Liquid Pap, and Endocervical Swab

|  GC Infected Status | Assay Results |   |   |   |   |   |   | Symptom Status  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  AC2TIGRIS |   | AGCTIGRIS |   | AC2PANTHER  |   |   |   |   |
|   |  PCyt | FU | PCyt | FU | CVS/PVS | PCyt | FS | Sym | Asym  |
|  Infected | + | + | + | + | + | + | + | 22 | 10  |
|  Infected | + | + | + | + | + | + | NA | 1 | 0  |
|  Infected | + | + | + | - | + | + | + | 1 | 0  |
|  Infected | + | + | + | = | + | + | + | 0 | 1  |
|  Infected | + | - | + | - | + | + | + | 3 | 3  |

{21}

Table 12: Clinical Study One. GC Infected Status for Performance Evaluation in Male Urethral Swab Samples

|  GC Infected Status | Assay Results |   |   |   |   |   | Symptom Status  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  AC2 DTS |   | AGC DTS |   | AC2 PANTHER  |   |   |   |
|   |  MS | MU | MS | MU | MS | Sym | Asym |   |
|  Infected | + | + | + | + | + | 30 | 2 |   |
|  Infected | + | + | + | + | NA | 0 | 1 |   |
|  Infected | + | - | + | - | + | 1 | 1 |   |
|  Infected | NA | + | NA | + | NA | 1 | 0 |   |
|  Not Infected | - | - | NA | NA | - | 205 | 307 |   |
|  Not Infected | - | - | NA | NA | NA | 14 | 9 |   |

AC2 = APTIMA Combo 2 Assay, AGC = APTIMA GC Assay, Asym = asymptomatic, CVS = clinician-collected vaginal swab, FS = female endocervical swab, FU = female urine, NA = result not available, PANTHER = PANTHER System, PCyt = PreservCyt Solution liquid Pap, PVS = patient-collected vaginal swab, Sym = symptomatic, TIGRIS = TIGRIS DTS System. The equal symbol (=) represents an equivocal result on repeat testing.

{22}

Table 13: Clinical Study One and Clinical Study Two. GC Infected Status for Performance Evaluation in Male Urine Samples

|  GC Infected Status | Assay Results |   |   |   |   |   |   | Symptom Status  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  AC21 |   | AGC DTS2 |   | NAAT 13 |   | NAAT 23  |   |   |
|   |  MS | MU | MS | MU | MU | MU | MU | Sym | Asym  |
|  Clinical Study One |  |  |  |  |  |  |  |  |   |
|  Infected | + | + | + | + |  |  | + |  | 3  |
|  Infected | + | - | + | - |  |  | - |  | 1  |
|  Clinical Study Two |  |  |  |  |  |  |  |  |   |
|  Infected | + | + |  |  | + | + | + | 63 | 4  |
|  Infected | + | + |  |  | + | NA | + | 1 | 1  |
|  Infected | - | + |  |  | + | - | + | 0 | 1  |
|  Infected | NA | + |  |  | + | + | + | 2 | 0  |
|  Clinical Study One |  |  |  |  |  |  |  |  |   |
|  Not Infected | - | - | NA | NA |  |  | + |  | 2  |
|  Not Infected | - | - | NA | NA |  |  | - |  | 314  |
|  Clinical Study Two |  |  |  |  |  |  |  |  |   |
|  Not Infected | + | - |  |  | - | - | - | 2 | 4  |
|  Not Infected | - | + |  |  | - | - | + | 0 | 1  |
|  Not Infected | - | - |  |  | + | - | - | 6 | 2  |
|  Not Infected | - | - |  |  | - | + | - | 1 | 0  |
|  Not Infected | - | - |  |  | - | - | + | 1 | 1  |
|  Not Infected | - | - |  |  | - | - | - | 407 | 945  |
|  Not Infected | - | - |  |  | - | NA | - | 9 | 19  |
|  Not Infected | - | - |  |  | NA | - | - | 1 | 2  |
|  Not Infected | - | NA |  |  | - | - | - | 2 | 0  |
|  Not Infected | NA | - |  |  | - | - | - | 2 | 0  |

AC2 = APTIMA Combo 2 Assay, AGC = APTIMA GC Assay, Asym = asymptomatic, DTS = DTS Systems, MS = male urethral swab, MU = male urine, NA = result not available, PANTHER = PANTHER System, Sym = symptomatic.
1 Male urethral swab and male urine samples were tested with the APTIMA Combo 2 Assay on the DTS Systems in Clinical Study One and on the TIGRIS DTS System in Clinical Study Two.
2 Male urethral swab and male urine samples were tested with the APTIMA GC Assay on the DTS Systems in Clinical Study One.
3 Male urine samples were tested with two FDA-cleared GC NAATs in Clinical Study Two.
Note. Data from asymptomatic men in Clinical Study One are combined with data from Clinical Study Two.

{23}

24

# RLU Distribution of APTIMA Combo 2 Controls

The distribution of the RLU values for the APTIMA Combo 2 controls is presented in Table 14 for all valid PANTHER System runs performed during Clinical Study One and Clinical Study Two.

Table 14: Clinical Study One and Clinical Study Two. RLU Distribution of APTIMA Combo 2 Controls

|  Control | Statistic | Total RLU (x1000)  |   |
| --- | --- | --- | --- |
|   |   |  Clinical Study One | Clinical Study Two  |
|  Positive Control, CT/Negative Control, GC | N | 66 | 23  |
|   |  Maximum | 1335 | 1258  |
|   |  Median | 1081.5 | 1135.0  |
|   |  Minimum | 624 | 910  |
|   |  CV% | 11.2 | 7.5  |
|  Positive Control, GC/Negative Control, CT | N | 66 | 23  |
|   |  Maximum | 1241 | 1311  |
|   |  Median | 1172.0 | 1174.0  |
|   |  Minimum | 1063 | 1082  |
|   |  CV% | 3.2 | 4.9  |

4. Clinical cut-off:

Not Applicable

5. Expected values/Reference range:

A summary of the prevalence of CT and GC, by specimen type, as determined by the APTIMA Combo 2 Assay on the PANTHER System for two multi-center clinical studies is shown in the following tables.

Clinical Study One: Prevalence of CT and GC Infections as Determined by the APTIMA Combo 2 Assay in Male Urethral Swab, Vaginal Swab, PreservCyt Solution Liquid Pap, and Endocervical Swab Samples by Clinical Site

|  Prevalence % (# positive/# tested with valid results)  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  Site | MS | CVS/PVS | PCyt | FS |   |
|   | CT+/GC- CT-/GC+CT+/GC+ | CT+/GC- CT-/GC+ CT+/GC+ | CT+/GC- CT-/GC+ CT+/GC+ | CT+/GC- CT-/GC+ CT+/GC+ |   |

{24}

|  1 | 0 (-) | 0 (-) | 0 (-) | 9.9 (21/212) | 3.3 (7/212) | 3.8 (8/212) | 8.9 (20/225) | 2.7 (6/225) | 3.1 (7/225) | 10.4 (20/193) | 3.1 (6/193) | 3.6 (7/193)  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  2 | 13.9 (28/202) | 5.9 (12/202) | 3.0 (6/202) | 8.3 (19/230) | 3.9 (9/230) | 1.3 (3/230) | 8.8 (21/239) | 4.6 (11/239) | 0.8 (2/239) | 8.2 (19/231) | 4.8 (11/231) | 0.9 (2/231)  |
|  3 | 1.3 (1/76) | 1.3 (1/76) | 0.0 (0/76) | 2.7 (6/222) | 0.5 (1/222) | 0.0 (0/222) | 3.1 (7/226) | 0.4 (1/226) | 0.0 (0/226) | 2.7 (6/223) | 0.4 (1/223) | 0.0 (0/223)  |
|  4 | 24.4 (33/135) | 1.5 (2/135) | 4.4 (6/135) | 11.7 (40/342) | 1.5 (5/342) | 1.2 (4/342) | 10.2 (35/342) | 1.5 (5/342) | 0.9 (3/342) | 11.3 (38/337) | 1.8 (6/337) | 0.9 (3/337)  |
|  5 | 0 (-) | 0 (-) | 0 (-) | 4.5 (1/22) | 0.0 (0/22) | 0.0 (0/22) | 4.8 (1/21) | 0.0 (0/21) | 0.0 (0/21) | 4.3 (1/23) | 0.0 (0/23) | 0.0 (0/23)  |
|  6 | 21.5 (28/130) | 5.4 (7/130) | 0.8 (1/130) | 11.9 (13/109) | 3.7 (4/109) | 0.9 (1/109) | 8.7 (10/115) | 1.7 (2/115) | 0.9 (1/115) | 8.8 (10/114) | 1.8 (2/114) | 0.9 (1/114)  |
|  7 | 16.7 (1/6) | 0.0 (0/6) | 0.0 (0/6) | 3.2 (5/157) | 2.5 (4/157) | 0.6 (1/157) | 2.5 (4/161) | 2.5 (4/161) | 0.6 (1/161) | 2.6 (4/152) | 2.6 (4/152) | 0.7 (1/152)  |
|  All | 16.6 (91/549) | 4.0 (22/549) | 2.4 (13/549) | 8.1 (105/1294) | 2.3 (30/1294) | 1.3 (17/1294) | 7.4 (98/1329) | 2.2 (29/1329) | 1.1 (14/1329) | 7.7 (98/1273) | 2.4 (30/1273) | 1.1 (14/1273)  |

CVS = clinician-collected vaginal swab, FS = female endocervical swab, MS = male urethral swab, PCyt = PreservCyt Solution liquid Pap, PVS = patient-collected vaginal swab.

Clinical Study One and Clinical Study Two. Prevalence of CT and GC Infections as Determined by the APTIMA Combo 2 Assay in Male Urine Samples by Clinical Site

|  Site | Prevalence % (# positive/# tested with valid results)  |   |   |
| --- | --- | --- | --- |
|   |  CT+/GC- | CT-/GC+ | CT+/GC+  |
|  1 | 6.0(6/100) | 0.0(0/100) | 0.0(0/100)  |
|  2 | 3.0(2/67) | 3.0(2/67) | 0.0(0/67)  |
|  3 | 0.0(0/109) | 0.9(1/109) | 0.0(0/109)  |
|  4 | 13.0(13/100) | 3.0(3/100) | 1.0(1/100)  |
|  5 | 13.6(17/125) | 5.6(7/125) | 0.0(0/125)  |
|  6 | 15.1(43/284) | 7.0(20/284) | 2.1(6/284)  |
|  7 | 1.4(3/212) | 0.9(2/212) | 0.0(0/212)  |
|  8 | 1.3(1/75) | 0.0(0/75) | 0.0(0/75)  |
|  9 | 16.7(42/251) | 5.2(13/251) | 3.2(8/251)  |

{25}

|  10 | 20.5
(17/83) | 1.2
(1/83) | 0.0
(0/83)  |
| --- | --- | --- | --- |
|  11 | 4.1
(6/146) | 0.7
(1/146) | 0.7
(1/146)  |
|  12 | 14.3
(16/112) | 4.5
(5/112) | 2.7
(3/112)  |
|  13 | 8.9
(10/112) | 2.7
(3/112) | 2.7
(3/112)  |
|  14 | 7.7
(2/26) | 0.0
(0/26) | 0.0
(0/26)  |
|  All | 9.9
(178/1802) | 3.2
(58/1802) | 1.2
(22/1802)  |

Note. CT and GC prevalence was estimated using symptomatic male urine samples from Clinical Study Two and asymptomatic male urine samples from both studies.

## N. Instrument Name:

The PANTHER System

## O. System Descriptions:

1. **Modes of Operation:**

Batch, random access

2. **Software:**

FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:

Yes ☐ X ☐ or No ☐

3. **Specimen Identification:**

By hand-held barcode reader and positional checks

4. **Specimen Sampling and Handling:**

Fully automated

5. **Calibration:**

Gen-Probe Field Service Engineers perform a luminometer calibration on the PANTHER System every 12 months as part of the Preventive Maintenance. Also, there are process

{26}

controls and calibration checks on all of the dispensers, thermal devices, and the vacuum system.

6. Quality Control:

In addition to the assay controls that are specific to each assay, the PANTHER System contains process controls that employ both hardware and software components. The process controls include, but are not limited to:

- Verification that the sequence of assay processing steps is correct for each reaction.
- Verification that the reaction incubation times and temperatures are correct.
- Verification that reagents and fluids are appropriately dispensed.

P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:

Not Applicable

Q. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

R. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

27

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/LSL/K132251](https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/LSL/K132251)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
