COPALIS ONE IMMUNOASSAY SYSTEM/TOXOPLASMA GONDII/RUBELLA/CMV TOTAL ANTIBODY ASSAYS
K961784 · Sienna Biotech, Inc. · LQN · Oct 31, 1996 · Microbiology
Device Facts
Record ID
K961784
Device Name
COPALIS ONE IMMUNOASSAY SYSTEM/TOXOPLASMA GONDII/RUBELLA/CMV TOTAL ANTIBODY ASSAYS
Applicant
Sienna Biotech, Inc.
Product Code
LQN · Microbiology
Decision Date
Oct 31, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3510
Device Class
Class 2
Indications for Use
The Copalis™ TORC, Toxo, Rubella, and CMV Total Antibody Assays use Coupled Particle Light Scattering (Copalis) technology in microparticle agglutination-based immunoassays for the qualitative detection of total antibodies (IgG and IgM) to Toxoplasma gondii, rubella and/or cytomegalovirus (CMV) in human serum using the Copalis™ I Immunoassay System. The presence of antibodies is indicative of current or prior infection with the suspected organism. The results of these assays on a single serum specimen are used to determine the patient's immune status for rubella and to determine the patient's immunological experience for Toxoplasma gondii and CMV. When evaluating properly paired sera, the results of these assays are used to demonstrate seroconversion as evidence of recent infection. Both specimens should be tested simultaneously (see Interpretation of Results).
Device Story
System uses Coupled Particle Light Scattering (Copalis) technology for immunoassay. Input: human serum samples. Process: latex microparticles coated with inactivated T. gondii, rubella, and CMV antigens aggregate in presence of specific antibodies; 10-minute agitation; detector measures changes in light scattering, particle size, and count of reacted/unreacted particles. Output: qualitative antibody detection based on aggregation levels relative to cutoff. Used in clinical laboratories; operated by laboratory personnel. Results assist clinicians in determining patient immune status or recent infection via seroconversion. Benefits: rapid, simultaneous detection of multiple pathogens from single reagent.
Clinical Evidence
Clinical trials at 3 sites (n=769 serum samples) compared Copalis TORC to predicate assays. Relative sensitivity: T. gondii 95.4%, Rubella 95.0%, CMV 93.0%. Relative specificity: T. gondii 93.9%, Rubella 91.7%, CMV 97.1%. Agreement following resolution of discordants: 97.9% (T. gondii), 97.8% (Rubella), 99.7% (CMV). Reproducibility studies showed low %CV across sites and lots. CDC serum panels showed 100% agreement for Rubella and CMV. WHO/CDC reference standards used to establish quantitative sensitivity.
Technological Characteristics
Uses Coupled Particle Light Scattering (Copalis) technology. Employs latex microparticles coated with inactivated viral/protozoal antigens. Principle: antibody-dependent particle aggregation measured by light scattering changes. System includes Copalis One Immunoassay System analyzer. Connectivity/software details not specified. Sterilization method not specified.
Indications for Use
Indicated for qualitative detection of total antibodies (IgG and IgM) to Toxoplasma gondii, rubella, and cytomegalovirus (CMV) in human serum. Used to determine immune status for rubella and immunological experience for T. gondii and CMV. Used to demonstrate seroconversion in paired sera. Not cleared for screening blood or plasma donors.
Regulatory Classification
Identification
Rubella virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to rubella virus in serum. The identification aids in the diagnosis of rubella (German measles) or confirmation of a person's immune status from past infections or immunizations and provides epidemiological information on German measles. Newborns infected in the uterus with rubella virus may be born with multiple congenital defects (rubella syndrome).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) National Committee for Clinical Laboratory Standards':
(i) 1/LA6 “Detection and Quantitation of Rubella IgG Antibody: Evaluation and Performance Criteria for Multiple Component Test Products, Speciment Handling, and Use of the Test Products in the Clinical Laboratory, October 1997,”
(ii) 1/LA18 “Specifications for Immunological Testing for Infectious Diseases, December 1994,”
(iii) D13 “Agglutination Characteristics, Methodology, Limitations, and Clinical Validation, October 1993,”
(iv) EP5 “Evaluation of Precision Performance of Clinical Chemistry Devices, February 1999,” and
(v) EP10 “Preliminary Evaluation of the Linearity of Quantitive Clinical Laboratory Methods, May 1998,”
(2) Centers for Disease Control's:
(i) Low Titer Rubella Standard,
(ii) Reference Panel of Well Characterized Rubella Sera, and
(3) World Health Organization's International Rubella Standard.
Predicate Devices
BioWhittaker Toxostat
Becton Dickinson And Co. Rubascan
Becton Dickinson And Co. CMVscan
Abbott IMx
Related Devices
K970931 — COPALIS TORC TOTAL ANTIBODY ASSAY · Sienna Biotech, Inc. · Apr 24, 1997
K080008 — BIOPLEX 2200 SYSTEM TORC IGG KIT, CALIBRATOR SET AND CONTROL SET · Bio-Rad Laboratories, Inc. · Feb 23, 2009
K955799 — COPALIS CMV TOTAL ANTIBODY ASSAY · Sienna Biotech, Inc. · Jul 10, 1996
K170509 — BioPlex 2200 ToRC IgM, BioPlex 2200 ToRC IgM Calibrator Set, BioPlex 2200 ToRC IgM Control Set · Bio-Rad Laboratories · May 19, 2017
Submission Summary (Full Text)
{0}
August 27, 1996
# SUMMARY OF SAFETY AND EFFECTIVENESS
SUBMITTED BY:
Judith J. Smith
Sienna Biotech, Inc.
9115 Guilford Rd. Suite 180
Columbia, MD 21046
K961784
OCT 31 1996
NAME OF DEVICES:
Trade Name:
Copalis One Immunoassay System; Copalis TORC Total Antibody Assay; Copalis Toxoplasma gondii Total Antibody Assay; Copalis Rubella Total Antibody Assay Copalis CMV Total Antibody Assay
Common Names/Descriptions:
Immunoassay Analyzer; Immunoassay for the Detection of Total Antibodies to Toxoplasma gondii, Rubella and Cytomegalovirus
Classification Names:
Nephelometer for clinical use; Toxoplasma gondii serological reagents; Rubella virus serological reagents; Cytomegalovirus serological reagents
PREDICATE DEVICES:
BioWhittaker Toxostat; Becton Dickinson And Co. Rubascan and CMVscan; Abbott IMx
DEVICE DESCRIPTION:
INTENDED USE: The Copalis™ TORC, Toxo, Rubella, and CMV Total Antibody Assays use Coupled Particle Light Scattering (Copalis) technology in microparticle agglutination-based immunoassays for the qualitative detection of total antibodies (IgG and IgM) to Toxoplasma gondii, rubella and/or cytomegalovirus (CMV) in human serum using the Copalis™ I Immunoassay System. The presence of antibodies is indicative of current or prior infection with the suspected organism. The results of these assays on a single serum specimen are used to determine the patient's immune status for rubella and to determine the patient's immunological experience for Toxoplasma gondii and CMV. When evaluating properly paired sera, the results of these assays are used to demonstrate seroconversion as evidence of recent infection. Both specimens should be tested simultaneously (see Interpretation of Results).
{1}
These assays has not been FDA cleared or approved for the screening of blood or plasma donors.
The assay will also be offered as separate microparticle immunoassays for the qualitative detection of total antibodies (IgG and IgM) to *Toxoplasma gondii*, rubella and cytomegalovirus (CMV) in human serum using the Copalis™ One Immunoassay System. The intended use of the individual assays will be specific to the individual antibodies detected but, other than that, will remain the same as the combination assay.
**KIT DESCRIPTION:** Coupled Particle Light Scattering (Copalis) technology provides a rapid method for the measurement of antibodies to specific viral or protozoal pathogens.
The Copalis™ TORC, Toxo, Rubella, and CMV Total Antibody Assays are based on the principle of antibody-dependent particle aggregation as detected by measurement of changes in light scattering. Due to the unique measuring system, a sample can be tested for antibodies to *Toxoplasma gondii*, rubella and CMV using a single TORC reagent and obtain results for the individual antibodies. Sized latex microparticles coated with inactivated *Toxoplasma gondii*, rubella and CMV antigens aggregate in the presence of antibodies to these infectious agents. After 10 minutes of agitation, the levels of aggregation are determined by discrimination of particle sizes (for the TORC assay) and measurement of the number of reacted and unreacted particles as they flow past a detector. Reactivity is assessed by the level of aggregation per particle size relative to a cutoff value. The Copalis TORC and individual Total Antibody Assays detect the presence of both IgM and IgG antibodies. Two levels of controls are used to monitor system performance.
## PERFORMANCE DATA:
**Comparison:** A comparison study was performed of patient specimens and controls using the Copalis TORC Total Antibody Assay and the Copalis individual assays. Analyses of the data by One Way Analysis of Variance showed that the differences in mean values between the component of the TORC assay and the corresponding individual assay are not great enough to exclude the possibility that the difference is due to random sampling variability; there is not a statistically significant difference (p = 0.849).
In addition, a reproducibility study was conducted to compare the components of the Copalis TORC Total Antibody Assay and the Copalis individual assays. The difference in mean CTRs and %CVs of the component of the TORC assay and the individual assay were not clinically significant. Therefore, it can be concluded that the clinical performance of the TORC and individual assays are equivalent.
Clinical trials were conducted at 3 sites (2 clinical laboratories and Sienna Biotech laboratory) to evaluate the performance of the Copalis TORC, Toxo,
{2}
Rubella, and CMV Total Antibody Assays in detecting antibodies to Toxoplasma gondii, rubella and CMV on the modified Copalis One Immunoassay System. The assay performance was compared to the BioWhittaker ToxoStat (ToxoStat) and the Becton Dickinson and Co. Rubascan (Rubascan) and CMVscan (CMVscan) assays.
A total of 769 serum samples were tested, 20% of which were fresh samples. Combined site testing results for each component of the assay are presented in Table A.
Table A
Comparison of the Copalis™ TORC Total Antibody Assay with Commercially Available Assays - Combined Site Results
| | Toxoplasma gondii Antibody | Rubella Antibody | CMV Antibody |
| --- | --- | --- | --- |
| Relative Sensitivity (95% Confidence Interval) | 95.4% | 95.0% | 93.0% |
| | (92.0 - 97.6%) | (92.9 - 96.5%) | (90.3 - 95.2%) |
| Relative Specificity (95% Confidence Interval) | 93.9% | 91.7% | 97.1% |
| | (91.5 - 95.8%) | (87.6 - 94.8%) | (94.6 - 98.6%) |
| Initial Agreement | 94.3% | 93.9% | 94.7% |
| Agreement Following Resolution of Discordants | 97.9% | 97.8% | 99.7% |
## Reproducibility:
Reproducibility studies were performed at the 3 sites using one lot of TORC tests. Assay component reproducibility was determined by testing 4 samples - 1 negative, 1 low positive (near the components' cutoffs) and 2 high positive (near the upper limit of the components' ranges). Samples were tested in triplicate twice a day for 5 days. The results are summarized in Table B.
Table B
Reproducibility Results For Copalis Torc Total Antibody Assay - Combined Sites
| SAMPLE | ANTIBODY TO TOXOPLASMA GONDII | | | ANTIBODY TO RUBELLA | | | ANTIBODY TO CMV | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | MEAN CTR | WITHIN RUN %CV | TOTAL %CV | MEAN CTR | WITHIN RUN %CV | TOTAL %CV | MEAN CTR | WITHIN RUN %CV | TOTAL %CV |
| RP1 (n = 89) | 103 | 1.8 | 2.0 | 105 | 1.9 | 2.0 | 102 | 2.1 | 2.8 |
| RP2 (n = 86) | 129 | 3.0 | 4.4 | 116 | 2.1 | 2.7 | 132 | 3.9 | 4.6 |
| RP3 (n = 86) | 157 | 4.1 | 5.9 | 128 | 3.0 | 5.2 | 166 | 4.5 | 5.8 |
| RP4 (n = 85) | 177 | 4.7 | 7.3 | 150 | 3.6 | 7.3 | 207 | 4.8 | 7.7 |
{3}
During the clinical trials, the Negative and Low Positive Controls were assayed each day of use at each site for the length of the trial. Results are summarized in Table C.
| Control | | Component | | |
| --- | --- | --- | --- | --- |
| | | Toxoplasma gondii | Rubella | CMV |
| Negative | Mean | 102 | 102 | 101 |
| | % CV | 2.6 | 3.1 | 2.8 |
| | n | 36 | 36 | 35 |
| Low Positive | Mean | 124 | 120 | 124 |
| | % CV | 4.5 | 4.5 | 4.1 |
| | n | 95 | 95 | 95 |
Lot-to-lot reproducibility was established using 3 test cup lots of Copalis Rubella Total Antibody Assay. A summary of the data is presented in Table D.
| Level | | Lot #1 | Lot #2 | Lot #3 |
| --- | --- | --- | --- | --- |
| Negative | Mean | 100 | 100 | 102 |
| | % CV | 1.3 | 3.0 | 1.5 |
| Low Positive | Mean | 131 | 125 | 128 |
| | % CV | 4.7 | 2.9 | 2.4 |
| Mid Positive | Mean | 162 | 150 | 156 |
| | % CV | 7.6 | 6.7 | 4.8 |
| High Positive | Mean | 204 | 187 | 197 |
| | % CV | 5.1 | 2.9 | 7.1 |
**CDC SERUM PANELS**: The following information is from Rubella and CMV Serum Panels obtained from the CDC and tested by Sienna Biotech, Inc. The results are presented as a means to convey further information of the performance of this assay with masked, characterized serum panels. This does not imply an endorsement of the assay by the CDC.
The Rubella Serum Panel consists of 82% positive and 18% negative samples. The Copalis TORC Total Antibody Assay demonstrated 100% total agreement with the CDC rubella antibody results.
The CMV Serum Panel consists of 66% positive and 34% negative samples. The Copalis TORC Total Antibody Assay demonstrated 100% total agreement with the CDC CMV antibody results.
**WHO AND CDC REFERENCE STANDARD TESTING**: The WHO 3rd International Standards for Toxoplasma gondii (1000 IU/mL) and rubella (1700 IU/mL) and the CDC Rubella Standard (21 IU/mL) were tested on the Copalis TORC Total Antibody Assay. Each was diluted and tested in replicates of 5 to
{4}
establish the quantitative sensitivity of the assay components' cutoffs and the reproducibility at those cutoffs. Results are summarized in Table E.
Table E
WHO and CDC Reference Standard Testing Results
| | WHO 3rd I.S. Toxoplasma gondii (1000 IU/mL)
25 IU/mL Dilution | WHO 3rd I.S. Rubella (1700 IU/mL)
10 IU/mL Dilution | CDC Rubella (21 IU/mL)
10 IU/mL Dilution |
| --- | --- | --- | --- |
| Mean (CTR) | 116 | 117 | 116 |
| %CV | 0.9% | 2.1% | 1.3% |
| Range (CTR) | 115 - 118 | 114 - 120 | 115 - 118 |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.