K980598 · Clark Laboratories, Inc. · LLM · May 1, 1998 · Microbiology
Device Facts
Record ID
K980598
Device Name
EBNA-1 IGM ELISA TEST SYSTEM
Applicant
Clark Laboratories, Inc.
Product Code
LLM · Microbiology
Decision Date
May 1, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3235
Device Class
Class 1
Indications for Use
The Epstein Barr Nuclear Antigen 1 (EBNA-1) IgM kit is an Enzyme-Linked Immunosorbent Assay (ELISA) for the qualitative determination of IgM antibodies in human serum to EBNA-1 antigen. The Clark anti-EBNA-1 IgM assay may be used in conjunction with other Epstein-Barr serologies (VCA IgG, VCA IgM, EA-D IgG, EBNA-1 IgG and heterophile) as an aid in the diagnosis of infectious mononucleosis in adult populations.
Device Story
The EBNA-1 IgM ELISA is an in vitro diagnostic test for qualitative detection of IgM antibodies to EBNA-1 antigen in human serum. Recombinant EBNA-1 antigen is immobilized on microtiter wells; patient serum is added; specific IgM antibodies bind to the antigen. After washing, enzyme-labeled anti-human IgM is added, followed by a substrate solution. The resulting color change is measured photometrically, providing an indirect measurement of specific antibody levels. The device is intended for use in clinical laboratories by trained personnel. Results are interpreted by healthcare providers alongside other EBV serologies (VCA IgG, VCA IgM, EA-D IgG, EBNA-1 IgG, heterophile) to assist in diagnosing infectious mononucleosis. The test aids clinical decision-making by characterizing the patient's EBV infection status.
Clinical Evidence
Clinical evaluation tested 166 serum samples characterized as seronegative, acute, or seropositive. Relative sensitivity for acute cases was 51.4% (95% CI: 34.9%-67.8%). Relative specificity was 100% (95% CI: 89.4%-100%) for seronegative and 94.5% (95% CI: 89.7%-99.3%) for seropositive samples. Relative agreement was 85.3% (95% CI: 79.6%-90.9%). Precision was evaluated across two sites with coefficients of variation ranging from 4.45% to 26.36% for various serum samples. Cross-reactivity testing showed no interference from Varicella Zoster Virus or rheumatoid factor, but noted some cross-reactivity with Herpes Simplex Virus and Cytomegalovirus.
Technological Characteristics
Enzyme-Linked Immunosorbent Assay (ELISA) utilizing recombinant EBNA-1 antigen immobilized on solid-phase microtiter wells. Detection via enzyme-labeled anti-human IgM conjugate and substrate-induced colorimetric reaction measured photometrically. Standalone diagnostic kit.
Indications for Use
Indicated for adult populations as an aid in the diagnosis of infectious mononucleosis via qualitative detection of IgM antibodies to EBNA-1 in human serum, used in conjunction with other EBV serologies.
Regulatory Classification
Identification
Epstein-Barr virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Epstein-Barr virus in serum. The identification aids in the diagnosis of Epstein-Barr virus infections and provides epidemiological information on diseases caused by these viruses. Epstein-Barr viruses are thought to cause infectious mononucleosis and have been associated with Burkitt's lymphoma (a tumor of the jaw in African children and young adults) and postnasal carcinoma (cancer).
Related Devices
K982348 — EBNA-1 IGM ELISA TEST SYSTEM · Columbia Bioscience, Inc. · Nov 25, 1998
K982352 — VCA IGM ELISA TEST SYSTEM · Columbia Bioscience, Inc. · Nov 25, 1998
K020707 — EBV-EBNA IGG ELISA KIT, MODEL EBG-100 · Pan Bio Pty. , Ltd. · Jun 13, 2002
K981829 — IS EBV-EBNA-1 IGG ELISA TEST SYSTEM · Diamedix Corp. · Feb 16, 1999
K973941 — IMMUNOWELL EBNA IGG TEST · Genbio · Aug 10, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
K980598
# Summary of Safety and Effectiveness Information EBNA-1 IgM ELISA Test Kit
### 1 1998 MAY
- I. Trinity Biotech US PO Box 1059 Jamestown, NY 14702-1059 Contact person: Ron Cruver Telephone: 716-483-3851 Date of preparation: Feb 11,1998
II. Description of Device
The Wampole Epstein-Barr Virus Nuclear Antigen 1 (EBNA-1) IgM kit is an Enzyme-Linked Immunosorbent Assays (ELISA) for the qualitative determination of IgM antibodies in human serum to EBNA-1 antigen. The Wampole anti-EBNA-1 IgM assay may be used in conjunction with other Epstein-Barr serologies (VCA IgG, VCA IgM, EA-D IgG, EBNA-1 IgG and heterophile) as an aid in the diagnosis of infectious mononucleosis.
## For In Vitro Diagnostic Use Only.
The EBNA-1 IgM ELISA test is an enzyme linked immunosorbent assay to detect IgM antibodies to Epstein-Barr Nuclear Antigen -1. Recombinant EBNA-1 antigen is attached to a solid phase microtiter well. Diluted test sera is added to each well. If the antibodies are present that recognize the antigen, they will bind to the antigen in the well. After incubation the wells are washed to remove unbound antibody. An enzyme labeled anti-human IgM is added to each well. If antibody is present it will bind to the antibody attached to the antigen on the well. After incubation the wells are washed to remove unbound conjugate. A substrate solution is added to each well. If enzyme is present the substrate will undergo a color change. After an incubation period the reaction is stopped and the color intensity is measured photometrically, producing an indirect measurement of specific antibody in the patient specimen.
#### III. Predicate Device
The EBNA-1 IgM ELISA test is substantially equivalent to EBV serology. Equivalence is demonstrated by the following comparative results:
{1}------------------------------------------------
# Performance Characteristics
# 1. Sensitivity and Specificity Based on Serum Characterization
One hundred and sixty six selected serum were tested at a clinical lab. The serum from the study were characterized as seronegative ( no serological evidence of past or present EBV infection), acute (VCA IgM and heterophile antibody present, EBNA IgG absent), or seropositive (presence of VCA IgG antibodies and EBNA IgG, no evidence of VCA IgM or heterophile antibody, indicative of past infection). The sensitivity, specificity and agreement of the assay was determined based on this characterization. It was assumed that the EBNA-1 IgM response should be negative for seronegative, and convalescent serum, and positive for acute serum. The results are summarized in Table 1.
Table 1
| | | Acute<br>VCA IgM+<br>EBNA IgG -<br>Heterophile + | Seropositive<br>VCA IgG+<br>EBNA IgG+<br>VCA IgM-<br>Heterophile - | Seronegative<br>VCA IgG-<br>EBNA IgG -<br>VCA IgM -<br>Heterophile - |
|---------------------------------|----------|--------------------------------------------------|--------------------------------------------------------------------|----------------------------------------------------------------------|
| Wampole<br>EBNA-1 IgM Equivocal | Positive | 19 | 5 | 0 |
| | | 2 | 8 | 0 |
| | Negative | 18 | 86 | 28 |
| | Total | 39 | 99 | 28 |
| Relative Sensitivity (Acute) | = 19/37 = 51.4% | 95% Confidence Interval = 34.9%-67.8% |
|-------------------------------------|-------------------|---------------------------------------|
| Relative Specificity (Seronegative) | = 28/28 = 100% | 95% Confidence Interval = 89.4%-100% |
| Relative Specificity (Seropositive) | = 86/91 = 94.5% | 95% Confidence Interval = 89.7%-99.3% |
| Relative Agreement | = 133/156 = 85.3% | 95% Confidence Interval = 79.6%-90.9% |
Equivocal results were not included in the calculations.
Equivocal results were not retested. They were reported as equivocal.
The 95% confidence intervals were calculated using the normal method.
*The 95% confidence interval was calculated assuming one false positive.
{2}------------------------------------------------
2. Precision. The Wampole EBNA-1 IgM EIA was evaluated for precision by testing six sera six times each at site one and ten times each at the second site on three different days. The results are summarized in the table below.
## Inter Site Precision Data
| Inter Site Precision (n=48) | | | |
|-----------------------------|------|-------|---------|
| Serum# | X | S.D. | C.V. |
| 1 | 1.87 | 0.161 | 8.60% |
| 2 | 1.56 | 0.166 | 10.63% |
| 3 | 2.38 | 0.156 | 6.57% |
| 4 | 1.30 | 0.132 | 10.18% |
| 5 | 0.47 | 0.050 | 10.52% |
| 6 | 0.19 | 0.051 | 26.36% |
| HPC* | 3.28 | 0.146 | 4.45% |
| CAL** | 2.22 | 0.144 | 6.48% |
| NC* | 0.01 | 0.012 | 181.66% |
X = Mean ISR Value S.D. = Standard Deviation C.V. = Coefficient of Variation
* HPC and NC - n=6 ** CAL - n=18
{3}------------------------------------------------
3. Cross-Reactivity. Sera containing IgM antibody detectable by ELISA to Herpes Simplex Virus I & II, Cytomegalovirus, and Varicella Zoster Virus were assayed. Sera containing rhuematoid factor (RF) were also assayed. The data summarized in Table 3 indicates that antibodies to Varicella and sera containing RF do not cross-react with the EBNA-1 IgM EIA kit. There is some cross reactivity to Herpes Simplex Virus and to Cytomegalovirus.
| Specificity | EBNA-1 IgM | Alternate Assay | |
|-------------|------------|-----------------|---|
| RF + | 0.15 | 1.87 | + |
| RF + | 0.09 | 1.82 | + |
| RF + | 0.01 | 1.73 | + |
| RF + | 0.10 | 1.80 | + |
| RF + | 0.10 | 1.85 | + |
| VZV M + | 0.44 | 3.28 | + |
| VZV M + | 0.17 | 5.46 | + |
| VZV M + | 0.10 | 4.98 | + |
| VZV M + | 0.29 | 2.34 | + |
| VZV M + | 0.20 | 2.18 | + |
| HSV 1 M + | 2.53 | 2.53 | + |
| HSV 1 M + | 0.27 | 1.65 | + |
| HSV 1 M + | 0.05 | 1.34 | + |
| HSV 1 M + | 0.08 | 1.32 | + |
| HSV 2 M + | 0.53 | 1.76 | + |
| HSV 2 M + | 0.08 | 1.60 | + |
| HSV 2 M + | 0.03 | 2.09 | + |
| HSV 2 M + | 0.62 | 1.96 | + |
| CMV M + | 1.57 | 1.23 | + |
| CMV M + | 0.98 | 1.92 | + |
| CMV M + | 2.00 | 3.83 | + |
| CMV M + | 0.38 | 1.32 | + |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized eagle with its head turned to the left and its wings forming three curved lines. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
1 1998 MAY
Clark Laboratories, Inc. c/o William L. Boteler, Jr. Immuno Probe, Inc. 1306 Bailes Lane, Suite F Frederick, Maryland 21701
K980598 Re: Trade Name: EBNA-1 IgM ELISA Regulatory Class: I Product Code: LLM Dated: February 13, 1998 Received: February 17, 1998
Dear Mr. Boteler:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
### Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page 1 of 1
510(k) Number. Not Known
Device Name: EBNA-1 IgM ELISA
Indications For Use: The Epstein Barr Nuclear Antigen 1 (EBNA-1) IgM kit is an Enzyme-Linked Immunosorbent Assay (ELISA) for the qualitative determination of IgM antibodies in human serum to EBNA-1 antigen. The Clark anti-EBNA-1 IgM assay may be used in conjunction with other Epstein-Barr serologies (VCA IgG, VCA IgM, EA-D IgG, EBNA-1 IgG and heterophile) as an aid in the diagnosis of infectious mononucleosis in adult populations.
# PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The Counter Use (Optional Format 1-2-96) --
Alo H
(Division Sign-Off) Division of Clinical Laboratory 510(k) Number_
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.