EBNA IGG EIA TEST SYSTEM

K951549 · Immunoprobe, Inc. · LLM · Apr 29, 1996 · Microbiology

Device Facts

Record IDK951549
Device NameEBNA IGG EIA TEST SYSTEM
ApplicantImmunoprobe, Inc.
Product CodeLLM · Microbiology
Decision DateApr 29, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3235
Device ClassClass 1

Indications for Use

The Epstein-Barr Virus Nuclear Antigen 1 (EBNA-1) IgG kit is an Enzyme-Linked Immunosorbent Assays (ELISA) for the semi-quantitative determination of IgG antibodies in human serum to EBNA-1 antigen. The ImmunoProbe anti-EBNA-1 IgG assay may be used in conjunction with other Epstein-Barr serologies (VCA IgG, VCA IgM, EA (R&D), and heterophile) as an aid in the diagnosis of infectious mononucleosis. For In Vitro Diagnostic Use Only.

Device Story

The EBNA-1 IgG EIA is an ELISA-based diagnostic test. Input: human serum samples. Principle: purified recombinant EBNA-1 antigen is immobilized on microtiter wells; patient serum is added; if anti-EBNA-1 IgG is present, it binds to the antigen. After washing, enzyme-labeled anti-human IgG is added, followed by a substrate solution. The resulting color change is measured photometrically. Output: semi-quantitative index value representing antibody concentration. Used in clinical laboratory settings by trained technicians. Results are interpreted by healthcare providers to assist in the diagnosis of infectious mononucleosis by identifying past EBV infection status.

Clinical Evidence

Clinical performance evaluated across three sites (two commercial R&D labs, one large commercial lab). Relative sensitivity 97.8% (95% CI: 96.1%-99.6%) and relative specificity 100% (95% CI: 96.0%-100%) compared to a commercial EIA. Precision studies (n=360) showed inter-site CVs ranging from 7.70% to 10.78% for positive samples. Linearity confirmed via serial dilution (r=0.979-0.989). No cross-reactivity observed with HSV-1, HSV-2, CMV, or VZV. Paired sera procedure demonstrated 100% sensitivity in detecting four-fold antibody increases.

Technological Characteristics

Enzyme-Linked Immunosorbent Assay (ELISA) using purified recombinant EBNA-1 antigen immobilized on solid-phase microtiter wells. Detection via enzyme-labeled anti-human IgG conjugate and photometric measurement of substrate color change. Semi-quantitative output based on index values.

Indications for Use

Indicated for semi-quantitative detection of IgG antibodies to EBNA-1 in human serum as an aid in diagnosing infectious mononucleosis. Used in conjunction with other EBV serologies (VCA IgG, VCA IgM, EA, heterophile). No specific age or gender restrictions provided.

Regulatory Classification

Identification

Epstein-Barr virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Epstein-Barr virus in serum. The identification aids in the diagnosis of Epstein-Barr virus infections and provides epidemiological information on diseases caused by these viruses. Epstein-Barr viruses are thought to cause infectious mononucleosis and have been associated with Burkitt's lymphoma (a tumor of the jaw in African children and young adults) and postnasal carcinoma (cancer).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} APR 29 1996 K951549 Summary of Safety and Effectiveness Information EBNA-1 IgG EIA Test Kit I. Immuno Probe Inc. 1306 Bailes Lane, Suite F Frederick, Maryland 21701 Contact person: William Boteler Telephone: 301-695-7920 Date of preparation: March 28, 1995 II. Description of Device The Epstein-Barr Virus Nuclear Antigen 1 (EBNA-1) IgG kit is an Enzyme-Linked Immunosorbent Assays (ELISA) for the semi-quantitative determination of IgG antibodies in human serum to EBNA-1 antigen. The ImmunoProbe anti-EBNA-1 IgG assay may be used in conjunction with other Epstein-Barr serologies (VCA IgG, VCA IgM, EA (R&D), and heterophile) as an aid in the diagnosis of infectious mononucleosis. For In Vitro Diagnostic Use Only. The EBNA-1 IgG EIA test is an enzyme linked immunosorbent assay to detect IgG antibodies to EBNA-1. Purified recombinant EBNA-1 is attached to a solid phase microtiter well. Diluted test sera is added to each well. If the antibodies are present that recognize the antigen, they will bind to the antigen in the well. After incubation the wells are washed to remove unbound antibody. An enzyme labeled anti-human IgG is added to each well. If antibody is present it will bind to the antibody attached to the antigen on the well. After incubation the wells are washed to remove unbound conjugate. A substrate solution is added to each well. If enzyme is present the substrate will undergo a color change. After an incubation period the reaction is stopped and the color intensity is measured photometrically, producing an indirect measurement of specific antibody in the patient specimen. III. Predicate Device The EBNA-1 IgG EIA test is substantially equivalent to BioWhittaker’s EBNA STAT test. Equivalence is demonstrated by the following comparative results: {1} # Performance Characteristics 1. Relative sensitivity and specificity. Three different sites compared the EBNA-1 IgG EIA test relative to a commercially available EIA assay. Two of the sites were R&D laboratories at commercial companies located in Maryland and New York. The other site was a large commercial lab located in Pennsylvania. The results of the studies are compiled and summarized in Table 1. None of the performance characteristics were established with specimens from patients having nasopharyngeal carcinoma or Burkitt's lymphoma. ## Table 1 ### EBNA-1 IgG Relative Sensitivity and Specificity | | | ImmunoProbe EBNA-1 G EIA | | | | | --- | --- | --- | --- | --- | --- | | | | + | eq | - | Total | | Alternate EBNA-1 G ELISA | + | 270 | 5 | 6 | 281 | | | eq | 0 | 0 | 1 | 1 | | | - | 0 | 1 | 74 | 75 | | | Total | 270 | 6 | 81 | 357 | | Relative Sensitivity = 270/276 = 97.8 | | | 95% Confidence interval = 96.1% - 99.6% | | | | Relative Specificity = 74/74 = 100% | | | 95% Confidence interval = 96.0% - 100% | | | | Relative Agreement = 344/350 = 98.3% | | | 95% Confidence interval = 96.9% - 99.7% | | | Equivocals were not included in the above calculations. The 95% confidence interval for relative specificity was calculated assuming one false positive The 95% confidence intervals were calculated using the normal method. {2} 2. Precision. Four different sera (High Positive, Mid Positive, Low Positive, and Negative) were assayed at three different sites to determine the precision of the assay. Each sera was tested three times each, twice a day for twenty days at each of the three study sites. The inter-site coefficient of variation (CV) for each serum is presented in table 2. Table 2 Inter Site Precision n= 360 | | Mean | Std Dev | CV% | | --- | --- | --- | --- | | High Positive | 4.95 | 0.381 | 7.70% | | Mid Positive | 2.89 | 0.229 | 7.92% | | Low Positive | 1.41 | 0.152 | 10.78% | | Negative | 0.01 | 0.022 | not applicable | | Calibrator (n=240) | 4.17 | 0.113 | 2.71% | | High Positive Control (n=120) | 6.90 | 0.419 | 6.07% | 3. Linearity. The data in table 3 illustrates EBNA-1 index values for serially two fold diluted sera. The index values are compared to log2 of dilution by standard linear regression. The data indicates that the antibody can be semi-quantitated by using a single serum dilution. Table 3 EBNA-1 G Linearity | Serum # | Neat | 1:2 | 1:4 | 1:8 | 1:16 | 1:32 | 1:64 | SLOPE | Y INTERCEPT | r | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | 1 | 8.6 | 5.96 | 3.50 | 1.96 | 0.87 | | | 1.95 | -1.66 | 0.985 | | 2 | 11.4 | 9.30 | 6.90 | 4.10 | 2.50 | 1.30 | 0.42 | 1.91 | -2.49 | 0.985 | | 3 | 6.90 | 4.90 | 2.7 | 1.5 | 0.69 | | | 1.58 | -1.41 | 0.983 | | 4 | 8.7 | 6.5 | 4.07 | 2.32 | 1.20 | 0.55 | | 1.67 | -1.95 | 0.979 | | 5 | 9.4 | 7.17 | 4.8 | 3.2 | 1.5 | 0.70 | | 1.77 | -1.75 | 0.989 | Linear regression compared Index Values to $\log_2$ of dilution. {3} 4. Cross-Reactivity. Serum containing IgG antibody detectable by ELISA to Herpes Simplex Virus I & II, Cytomegalovirus, and Varicella Zoster Virus were assayed. The data summarized in Table 4 indicates that antibodies to these Herpes Viruses do not cross-react with the EBNA-1 IgG EIA kit. | SERUM | EBNA-1 G | HSV1 G | HSV2 G | CMV | VZV | | --- | --- | --- | --- | --- | --- | | 1 | 0.67 - | 2.43 + | 0.60 - | 1.69 + | 2.28 + | | 2 | 0.36 - | 0.32 + | 0.08 - | 0.18 - | 3.17 + | | 3 | 0.73 - | 3.67 + | 0.86 - | 1.11 + | 2.45 + | | 4 | 0.29 - | 0.30 - | 0.39 - | 0.18 - | 3.21 + | | 5 | 0.34 - | 6.18 + | 5.98 + | 0.18 - | 2.14 + | Sera ≥ 1.10 were considered positive. Sera ≤ 0.90 were considered negative. 5. Evaluation of paired sera. To validate the sensitivity of the paired sera procedure 20 high positive sera were serially two fold diluted and run on the assay. From these dilutions, there were 68 pairs that had a four fold dilution where the acute sera had a value of less than 3.70. All 68 pairs demonstrated a >46% rise in ISR value, showing a significant rise in antibody. Therefore the paired sera procedure demonstrated 100% sensitivity in being able to detect a four fold increase in antibody level when the acute sera has a value of <3.70. 6. Sensitivity and Specificity Based on Serum Characterization The serum from the first study site were characterized as seronegative (no serological evidence of past or present EBV infection), acute (VCA IgM present), or seropositive (presence of VCA IgG antibodies, no evidence of VCA IgM, indicative of past infection). The sensitivity, specificity and accuracy of the assay was determined based on this characterization. It was assumed that the EBNA-1 IgG response should be negative for seronegative and acute serum, and positive for seropositive serum. The results are summarized in Table 5. {4} Table 5 | | | Seropositive VCA IgG + VCA IgM - | Acute VCA IgM+ | Seronegative VCA IgG - VCA IgM - | Total | | --- | --- | --- | --- | --- | --- | | IPI EBNA-1 IgG | Positive | 89 | 1 | 1 | 91 | | | Negative | 6 | 23 | 30 | 59 | | | Total | 95 | 24 | 31 | 150 | | Relative Sensitivity = 89/95 = 93.7% | | | 95% Confidence Interval = 88.7%-98.7% | | | | Relative Sensitivity = 53/55 = 96.4% | | | 95% Confidence Interval = 91.3%-100% | | | | Relative Agreement= 142/150 = 94.7% | | | 95% Confidence Interval = 91.0%-98.3% | | | Eight equivocal results were not included in the calculations. The 95% confidence intervals were calculated using the normal method.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...