Syphilis Health Check is a qualitative rapid membrane immunochromatographic assay for the detection of Treponema pallidum (syphilis) antibodies in human whole blood, serum or plasma. This product can be used as an initial screening test or in conjunction with a non-treponemal laboratory test and clinical findings, may aid in the diagnosis of syphilis infection. This test is not intended for use in screening blood or plasma donors.
Device Story
Syphilis Health Check is a rapid, qualitative, membrane-based immunochromatographic assay for detecting anti-Treponema pallidum antibodies. The device uses immobilized recombinant treponemal antigens on a membrane. Patient samples (whole blood, serum, or plasma) are added to the sample well; human immunoglobulins bind to antigens, forming complexes. These complexes react with colloidal gold-conjugated Protein A and anti-human immunoglobulins to produce a visible pink-rose colored band. A separate control line confirms reagent function. The test is performed at the point-of-care or in clinical settings by trained technicians. Results are read visually by the healthcare provider. The output aids in the clinical diagnosis of syphilis infection when used alongside non-treponemal tests and patient clinical findings.
Clinical Evidence
Clinical performance was evaluated using 1,292 samples (880 prospective, 412 retrospective) and 164 clinically diagnosed samples. Compared to reference treponemal tests, the device showed 98.5% positive agreement (95% CI: 97.1-99.4%) and 97.3% negative agreement (95% CI: 95.9-98.4%). In clinically diagnosed populations (primary, secondary, latent), the device showed 100% agreement. Analytical studies included precision/reproducibility across three sites, interference testing (hemoglobin, bilirubin, lipids, etc.), and cross-reactivity panels (HIV, HCV, HBV, autoimmune markers).
Technological Characteristics
Rapid immunochromatographic membrane assay. Materials: recombinant treponemal antigens, colloidal gold-conjugated Protein A, anti-human immunoglobulins. Form factor: absorbent test device. Energy source: none (passive flow). Connectivity: none. Sterilization: not specified. Software: none.
Indications for Use
Indicated for the qualitative detection of Treponema pallidum antibodies in human whole blood, serum, or plasma to aid in the diagnosis of syphilis infection. Intended for use as an initial screening test or in conjunction with non-treponemal tests and clinical findings. Not for blood or plasma donor screening. Prescription use only.
Regulatory Classification
Identification
Treponema pallidum treponemal test reagents are devices that consist of the antigens, antisera and all control reagents (standardized reagents with which test results are compared) which are derived from treponemal sources and that are used in the fluorescent treponemal antibody absorption test (FTA-ABS), the Treponema pallidum immobilization test (T.P.I.), and other treponemal tests used to identify antibodies to Treponema pallidum directly from infecting treponemal organisms in serum. The identification aids in the diagnosis of syphilis caused by bacteria belonging to the genus Treponema and provides epidemiological information on syphilis.
{0}------------------------------------------------
#### 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K102400
| Submitted By: | Diagnostics Direct, LLC<br>376 96th Street<br>Stone Harbor, NJ 08247 |
|---------------|----------------------------------------------------------------------|
| Telephone | (609) 368-6912 |
Official Correspondent: Gary Lehnus, Ph.D. Lehnus & Associates Consulting
(570) 620-0198 Telephone
FAX (570) 620-0199
August 24, 2010 Date Prepared:
Trade or Proprietary Name: Syphilis Health Check
Common or Usual Name: Rapid immunochromatographic membrane assay, Treponema pallidum antibody.
| Regulation | 21 CFR 866.3830 |
|------------|-----------------|
|------------|-----------------|
Class II (Special Controls) Classification
Product Codes LIP
Panel: Immunology and Microbiology (83)
Predicate Device: K001552 - Phoenix Bio-Tech - TREPCHEK TREPONEMAL ANTIBODY EIA
Device Description SYPHILIS HEALTH CHECK is a rapid qualitative screening test for detection of human antibodies to TP in serum, plasma or whole blood.
> The method employs an unique combination of anti-human immunoglobulins gold conjugate and highly purified TP recombinant proteins to specifically detect anti-TP antibodies. The test mainly detects IgG and IgM will also react in case of high concentrations.
> As the samples flow through the absorbent device, the anti-human immunoglobulins/protein A dye conjugate binds to the human immunoglobulins forming an antigen-antibody complex. This complex binds to the recombinant protein in the positive reaction zone and produces a pink-rose colored band. In the absence of anti TP antibodies, there is no line in the positive reaction zone. The reaction mixture continues flowing through the absorbent device past the reaction and control zones. Unbound conjugate binds to the reagents in the control zone producing a pink-rose color band, demonstrating that the reagents are functioning correctly.
Intended Use Syphilis Health Check is a qualitative rapid membrane immunochromatographic assay for the detection of Treponema pallidum (syphilis) antibodies in human
1
{1}------------------------------------------------
whole blood, serum or plasma. This product can be used as an initial screening test or in conjunction with a non-treponemal laboratory test and clinical findings, may aid in the diagnosis of syphilis infection. This test is not intended for use in screening blood or plasma donors.
## Indications for Use Syphilis Health Check is a qualitative rapid membrane
immunochromatographic assay for the detection of Treponema pallidum (syphilis) antibodies in human whole blood, serum or plasma. This product can be used as an initial screening test or in conjunction with a non-treponemal laboratory test and clinical findings, may aid in the diagnosis of syphilis infection. This test is not intended for use in screening blood or plasma donors.
## TECHNOLOGICAL CHARACTERISTICS
The following tables summarize similarities and differences between Syphilis Health Check and the current TREPCHEK Treponemal Antibody EIA (K001552):
| ITEM | PREDICATE DEVICE | NEW DEVICE |
|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name | TREPCHEK | Syphilis Health Check |
| Analyte | T. pallidum antibodies | T. pallidum Ig antibodies<br>(Detection of IgG and IgM<br>isotypes) |
| Specimens | Serum and plasma | Whole Blood, Serum, and<br>Plasma |
| Method | Enzyme ImmunoAssay | Rapid Immunochromato-<br>graphic membrane |
| Test principle | TREPCHEK is an enzyme-<br>immunoassay utilizing specific<br>recombinant treponemal<br>antigens immobilized to<br>microwells. Anti-treponemal<br>specific antibodies in the<br>patient's serum or plasma bind<br>to the antigens. Non-bound<br>proteins are removed during<br>the washing step. Horseradish<br>Peroxidase-labelled goat anti-<br>human IgG is added to bind<br>with the human antibodies in<br>the well. The unbound<br>conjugate is removed by a<br>second washing step; then the<br>chromogenic substrate TMB is<br>added. After an appropriate<br>incubation period, a stop<br>solution is added and the<br>absorbance of the resulting<br>colour is measured<br>photometrically with an EIA<br>reader (450 nm). Colour<br>intensity is proportional to the<br>amount of specific antibody<br>present in the patient's sample. | Specific, recombinant treponemal<br>antigens are immobilized on the<br>membrane. Patient samples or<br>controls are added to the sample<br>well, followed by a buffer, and as<br>the sample flows through the<br>absorbent device, the human<br>immunoglobulins bind to the<br>antigens forming an antigen-<br>antibody complex. The antigen-<br>antibody complex is further<br>reacted with a unique combination<br>of colloidal gold conjugated<br>Protein A and anti-human<br>immunoglobulins producing a pink-<br>rose colored band that can be read<br>visually indicating antibody is<br>present in the patient's sample. A<br>separate Control Line indicates<br>reagents are functioning correctly |
{2}------------------------------------------------
| ITEM | PREDICATE DEVICE | NEW DEVICE |
|---------------------------|-----------------------------------------------------------------|-------------------------------------------------|
| Antigen Used | Recombinant proteins | same |
| Detection | colorimetric detection - optical<br>density | Colloidal Gold - pink-red line<br>visually read |
| Calculation | Qualitative determination with<br>ratiometric values Calibrator | Qualitative determination from<br>the Test Line |
| Quality control | 2 Controls at different levels | Positive and Negative Controls |
| Indication for<br>use | Aid in the diagnosis of syphilis<br>disease. | same |
| Sample<br>Dilution | Diluted 100 µL per well | Undiluted 25 µL or 50µL per test<br>well |
| Incubation<br>Time | 30 min / 30 min / 15 min | 10 - 15 min. |
| Incubation<br>Temperature | Room Temperature (18 - 25°C) | Room Temp (18 - 25°C) |
| Conjugate | Anti-human IgG - HRPO | Protein A / Anti-human Ig -<br>Colloidal gold |
| Sensitivity | > 97% | >98% |
| Specificity | > 97% | >98% |
Performance: The following are performance characteristics of Syphilis Health Check
In order to evaluate the performance of the Syphilis Health Check test as series of 880 patient samples were obtained that were from prospective study sites and 412 frozen retrospective samples purchased from outside commercial vendors and blood centers. In addition, a series of 164 clinically diagnosed samples were obtained that were known primary, secondary, and latent both treated and untreated. The total of 1292 known retrospective samples (576) and prospectively collected (880) patient samples were used to demonstrate the performance of the Syphilis Health Check test to Non-treponemal RPR and other treponemal tests.
Prospective Studies were conducted at five clinical study sites which compared the Syphilis Health Check to RPR, a non-treponemal test, and treponemal tests such as TPPA, TPHA, or ELISA, using specimens from patients coming into four STD clinics and one hospital clinic. The patients enrolled in the study were identified by medical associates as suspected positive for syphilis and exhibiting symptoms.
Initial evaluations were performed at a university clinic and a hospital clinic to assess the performance of the Syphilis Health-Check test versus RPR and their reference treponemal tests -FTA and TPHA. Only gender and age were collected from these patients. A more comprehensive study was performed at three study sites to collect further patient history information in order to identify a broader range of STD related patients. The information and data collected from these sites is presented below.
The Syphilis Health Check assay demonstrated 95.6% and 98.5% Percent Positive Agreement versus the non-treponemal test and treponemal tests respectively, and 90.5% and 97.3% Percent Negative Agreement respectively.
# CUMULATIVE COMPARISON RESULTS
The results obtained for the prospective samples vields the following results compared to Non-treponemal RPR and treponemal tests.
{3}------------------------------------------------
TOTAL Non-Treponemal Comparison - total combined sample results are presented from the 5 prospective sites and the frozen known and suspected positive samples.
| | RPR - Non-Treponemal | | | |
|--------------------------|----------------------|----------|-------|------|
| | Positive | Negative | Total | |
| Syphilis Health<br>Check | Positive | 473 | 76 | 549 |
| | Negative | 22 | 721 | 743 |
| | Total | 495 | 797 | 1292 |
RPR - Non-Treponemal
Percent Positive Agreement: 473/495 = 95.6% (95% C.I. = 93.4 - 97.2%) Percent Negative Agreement: 721/797 = 90.5% (95% C.I. = 88.2 - 92.4%) Percent Overall Agreement: 1194/1292 = 92.4% (95% C.I. = 90.8 - 93.8%)
TOTAL Treponemal Comparison - total combined sample results are presented from the 5 sites and the frozen known and suspected positive samples
| | | Reference Treponemal tests | | |
|--------------------------|----------|----------------------------|----------|-------|
| | | Positive | Negative | Total |
| Syphilis Health<br>Check | Positive | 531 | 20 | 551 |
| Syphilis Health<br>Check | Negative | 8 | 733 | 741 |
| | Total | 539 | 753 | 1292 |
Percent Positive Agreement: 531/539 = 98.5% (95% C.I. = 997.1 - 99.4%) Percent Negative Agreement: 733/753 = 97.3% (95% C.I. = 95.9 - 98.4%) Percent Overall Agreement: 1264/1292 = 97.8% (95% C.I. = 96.9 - 98.6%)
#### PROSPECTIVE STUDIES
#### University Clinic site
| | | RPR | | |
|-----------------|-------|-----|-----|-------|
| | | pos | neg | Total |
| Syphilis | pos | 32 | 1 | 33 |
| Health<br>Check | neg | 0 | 6 | 6 |
| | Total | 32 | 7 | 39 |
Percent Positive Agreement = 100.0% (95% C.I. = 89.1 - 100%) Percent Negative Agreement = 85.7% (95% C.I. = 42.1 - 99.6%) Percent Overall Agreement = 97.4% (95% C.I. = 86.5 - 99.9%)
| FTA | | |
|-----|-----|-------|
| pos | neg | Total |
| 27 | 6 | 33 |
| 0 | 6 | 6 |
| | | |
| 27 | 12 | 39 |
100% (95% C.I. = 87.2 - 100%) 50.0% (95% C.I. = 21.1 - 78.9%) 84.6% (95% C.I. = 69.5 - 94.1%)
{4}------------------------------------------------
## Hospital Clinic site
| | | RPR | | |
|-----------------|-------|-----|-----|-------|
| | | pos | neg | Total |
| Syphilis | pos | 3 | 3 | 6 |
| Health<br>Check | neg | 0 | 44 | 44 |
| | | | | |
| | Total | 3 | 47 | 50 |
| TPHA | | |
|------|-----|-------|
| pos | neg | Total |
| 6 | 0 | 6 |
| 0 | 44 | 44 |
| | | |
| 6 | 44 | 50 |
Percent Positive Agreement = 100.0% (95% C.I. = 29.2 - 100.0% (95% C.I. = 54.1 - 100%) Percent Negative Agreement = 93.6% (95% C.I. = 82.5 - 98.7%) 100.0% (95% C.I. = 92.0 - 100%) Percent Overall Agreement = 94.0% (95% C.I. = 83.5 - 98.7%) 100.0% (95% C.I. = 92.9 - 100%)
#### Study Site 1
| | | RPR | | |
|-----------------------------|-------|-----|-----|-------|
| | | pos | neg | Total |
| Syphilis<br>Health<br>Check | pos | 13 | 15 | 28 |
| | neg | 2 | 370 | 372 |
| | | | | |
| | Total | 15 | 385 | 400 |
| TPPA | | |
|------|-----|-------|
| pos | neg | Total |
| 21 | 8 | 29 |
| 6 | 365 | 371 |
| | | |
| 27 | 373 | 400 |
Percent Positive Agreement = 86.7% (95% C.I.= 59.5 - 98.3) Percent Negative Agreement = 96.1% (95% C.I.= 93.7 - 97.8) Percent Overall Agreement = 95.8% (95% C.I.= 93.3 - 97.5)
77.8% (95% C.I.= 57.7 - 91.4) 97.9%(95% C.I.= 95.8 - 99.1) 96.5% (95% C.I.= 94.2 - 98.1)
It should be noted that two (2) of the false negative samples vs TPPA were only positive for TPPA but negative with the other two reference methods (RPR and FTA). Three samples were positive for TPPA and FTA, but negative for Syphilis Health Check and RPR.
#### Study Site 2
| | RPR | | | |
|-----------------------------|-------|-----|-----|-------|
| Syphilis<br>Health<br>Check | | pos | neg | Total |
| | pos | 2 | 2 | 4 |
| | neg | 0 | 85 | 85 |
| | Total | 2 | 87 | 89 |
Percent Positive Agreement = 100.0% (95% C.I. = 15.8 - 100) Percent Negative Agreement = 97.7% (95%C.I. = 91.9 - 99.7) Percent Overall Agreement = 97.8% (95% C.I. = 92.1 - 100)
| TPPA | | |
|------|-----|-------|
| pos | neg | Total |
| 4 | 0 | 4 |
| 0 | 85 | 85 |
| 4 | 85 | 89 |
100.0% (95% C.I. = 39.8 - 100%) 100.0% (95% C.I. = 95.8 - 100%) 100.0% (95% C.I. = 95.9 - 100%)
{5}------------------------------------------------
#### Study Site 3
| | RPR | | |
|-----------------------------|-----|-----|-------|
| | pos | neg | Total |
| Syphilis<br>Health<br>Check | 6 | 4 | 10 |
| | 0 | 195 | 195 |
| Total | 6 | 199 | 205 |
P P A = 100.0% (95% C.I.= 54.1 - 100) P N A = 98.0% (95% C.I.= 94.9 - 99.4) P O A = 98.0% (95% C.I.= 95.1 - 99.5)
| EIA | | |
|-----|-----|-------|
| pos | neg | Total |
| 9 | 2 | 11 |
| 1 | 193 | 194 |
| | | |
| 10 | 195 | 205 |
90.0% (95% C.I.= 55.5 - 99.7%) 99.0% (95% C.I.= 96.3 - 99.9%) 98.5% (95% C.I.= 95.8 - 99.7%)
## RETROSPECTIVE STUDIES
#### Suspected and Known Positive Syphilis Samples
A series of 412 total samples were purchased from serum and blood center suppliers consisting of 149 banked RPR and treponemal reactive serum samples and 28 serum samples that were requested to be Primary or Secondary Patients, treated, but exhibiting a Syphilitictype lesion or rash were purchased from a serum supplier, and 138 frozen serum and plasma samples were obtained from a blood center. The samples were found to be RPR and treponemal reactive and having mixed titers. Another series of 97 samples being highly suspected of having a syphilis infection were obtained from a serum supplier that were obtained from various laboratories around the U.S., and were submitted to the laboratories for testing. The samples were tested by RPR. TPPA. and MHA-TP as reference methods for comparison to Syphilis Health Check test.
The samples were further tested by an outside laboratory for TPPA titer and Syphills Health Check results. 19 samples were found to be Negative for TPPA and Syphilis Health Check and Reactive RPR, four samples were negative by TPPA, but Positive by Syphilis Health Check and RPR, one sample was Syphilis Health Check negative and positive RPR and TPPA, and one sample was RPR neq but TPPA and Syphilis Health Check positive. The remaining two hundred eighty-nine samples remained Positive by all four methods. In the Suspected Positive patients two samples were negative with Syphilis Health Check, with one sample low reactive with RPR, and both samples Non-Reactive with TPHA. The results of the testing are shown below.
#### Known Positives
| | | RPR | | |
|-----------------------------|-------|-----|-----|-------|
| | | pos | neg | Total |
| Syphilis<br>Health<br>Check | pos | 293 | 1 | 294 |
| | neg | 20 | 1 | 21 |
| | Total | 313 | 2 | 315 |
Percent Positive Agreement = 93.4% (95%C.I. = 89.8 - 96.0%) Percent Negative Agreement = 100.0% (95%C.I. =. 100%) Percent Overall Agreement = 93.4% (95%C.1. = 89.8 - 96.0%)
| TPPA | | |
|------|-----|-------|
| pos | neg | Total |
| 290 | 4 | 294 |
| 1 | 20 | 21 |
| 291 | 24 | 315 |
99.6% (95%C.I.= 97.9 - 100.0%) 85.7% (95%C.I.= 63.7 - 97.0%) 98.6% (95%C.1.= 96.5 - 99.6%)
6
{6}------------------------------------------------
#### Suspected Positive
| | | | RPR | | | |
|-----------------------------|-----|----|-----|-----|-------|-----|
| | | | pos | neg | Total | pos |
| Syphilis<br>Health<br>Check | pos | 62 | 25 | 87 | 87 | |
| | neg | 0 | 10 | 10 | 0 | |
| Total | | 62 | 35 | 97 | 87 | |
| TPHA | | |
|------|-----|-------|
| pos | neg | Total |
| 87 | 0 | 87 |
| 0 | 10 | 10 |
| | | |
| 87 | 10 | 97 |
PPA = 100.0% (95% C.I. = 94.2 - 100%) PNA = 28.6% (95% C.I. = 14.6 - 46.3%) POA = 74.2% (95% C.I. = 64.3 - 82.6%)
100.0% (95% C.I. = 95.8 - 100%) 100.0% (95% C.I. = 69.2 - 100%) 100.0% (95% C.I. = 96.3 - 100%)
neg Total pos 87 0 87 10 0 10 87 10 97
MHTP
100.0% (95% C.I. = 95.8 - 100%) 100.0% (95% C.l. = 69.2 - 100%) 100.0% (95% C.I. = 96.3 - 100%)
### Clinically Diagnosed
A panel of one hundred sixty-four (164) well-characterized clinically diagnosed serum samples from treated and untreated patients with primary, secondary, and latent syphilis infections were obtained from a clinic serving a population of individuals with a variety of infectious diseases. The samples were tested with reference assay tests for RPR, TPPA, and FTA-ABS. The samples were then tested with Syphilis Health Check and the results are summarized below.
| Known Clinical Status | | RPR | TPPA | FTA-<br>ABS | Syphilis<br>Health<br>Check | No. | %<br>Agree<br>-ment | 95% C.I. |
|-----------------------|-----------|-------|-------|-------------|-----------------------------|-----|---------------------|-------------|
| Untreated | Primary | React | React | React | React | 23 | 100 | 85.2 - 100% |
| | Secondary | React | React | React | React | 25 | 100 | 86.3 - 100% |
| | Latent | React | React | React | React | 22 | 100 | 84.6 - 100% |
| | | NR | React | React | React | 3 | 100 | 29.2 - 100% |
| Treated | Primary | React | React | React | React | 28 | 100 | 87.8 - 100% |
| | Secondary | React | React | React | React | 26 | 100 | 86.8 - 100% |
| | Latent | React | React | React | React | 18 | 100 | 81.5 - 100% |
| | | NR | React | React | React | 19 | 100 | 82.4 - 100% |
| Total | | | | | | 164 | 100 | 97.8 - 100% |
{7}------------------------------------------------
## Interference Study
.
Interference testing was conducted using serum. Concentrates of the compounds were prepared and diluted to multiple concentrations into eight sera with different levels of syphilis reactivity. The following results were obtained:
Hemoglobin: No effect was observed up to 1000 mg/dL of hemoglobin;
Bilirubin (total): No effect was observed up to 40 mg/dL of total bilirubin;
Triglycerides: No effect was observed up to 3000 mg/dL of triglycerides.
Cholesterol (total): No effect was observed up to 400mg/dL of cholesterol
Albumin: No effect was observed up to 1000 mq/dL of albumin
Gamma-globulin: No effect as observed up to 5000 mg/dL of gamma-globulin
#### Potential Cross Reactors:
Panels of samples were obtained to evaluate potential interference from different disease conditions confirmed positive and containing different concentrations of potentially cross-reactive antibodies and were analyzed with the Syphilis Health Check test. From 151 (138 Males) prospectively collected drug users eleven (7%) were found syphilis positive by reference methods and Syphilis Health Check, with two additional samples found positive for Syphilis Health Check. Two samples from Lyme disease and HSV showed a positive result with Syphilis Health Check but could not be confirmed. One sample each from CMV positive and heterophile positive patients were found Syphilis Health Check positive, but non-reactive by reference methods . Two HPV patients were found Syphilis Health Check positive with one confirmed by reference methods. Two co-infected Chlamydia/GC patients were found Syphilis Health Check and one confirmed positive by TPPA. The number of disease condition cateqories and reactive results obtained is listed in the following table.
| Cross-reactor | Number | # positive by<br>Syphilis<br>Health Check | # positive by<br>reference<br>method(s) |
|--------------------------------|--------|-------------------------------------------|-----------------------------------------|
| Self-reported Drug<br>Users | 151 | 13 | 11 |
| ANA Positive | 24 | 0 | |
| RF Positive | 40 | 0 | |
| U.S. Lyme Disease<br>IgG & IgM | 25 | 1 | Not confirmed |
| HSV | 24 | 1 | Not confirmed |
| CMV | 10 | 2 | 1 |
| EBV | 21 | 0 | |
| HAV | 25 | 0 | |
{8}------------------------------------------------
| HIV 1 & 2 | 14 | 1 | 1 |
|----------------------------------------------------|----|---|---|
| HTLV | 14 | 1 | 1 |
| Heterophile | 32 | 1 | 0 |
| HCV | 24 | 1 | 1 |
| Anti-HBs | 25 | 2 | 2 |
| Other STD - GC,<br>Chlamydia, HPV,<br>Trichomonas, | 78 | 4 | 2 |
### Syphilis Positive HIV / HCV / HBV Patients
A series of 13 patient EDTA Plasma samples from a blood center were identified that were screened RPR positive in patients that were also confirmed positive for HIV. HCV and/or Hepatitis B virus. Three of these patients were further identified as previously testing positive for syphilis and treated at that time. All of the samples were tested with TPPA and Syphilis Health Check to evaluate syphilis reactivity. All 13 samples were positive for RPR. TPPA, and Syphilis Health Check.
To further evaluate the influence of HIV on Syphilis Health Check results, a series of 24 banked serum containing high levels of viral load for HIV were tested with Syphilis Health Check. The HIV positive patient samples were purchased from a commercial serum supplier that were known to be RPR and/or Treponema positive. The samples were tested by an outside lab with the Syphilis Health Check test to assess reactivity. Six of these samples were Nonreactive by RPR but reactive with TPHA. One sample was RPR positive, but TPHA nonreactive. All 24 samples were Syphilis Health Check Positive.
#### Pregnant Women
A series of sixty-nine (69) pregnant female samples were purchased from a vendor that had known trimester, age, and ethnicity. An additional set of 93 pregnant women serum samples were obtained from a commercial source that were identified as syphilis positive by RPR screening and further tested with TPPA and Syphilis Health Check. Age and trimester were known, but ethnicity was not identified. Three samples out of the 162 total samples were RPR low positive but nonreactive by the treponemal methods.
#### Pregnant Women
Summary
| | | RPR | | |
|-----------------------------|-------|-----|-----|-------|
| | | pos | neg | Total |
| Syphilis<br>Health<br>Check | pos | 91 | 0 | 91 |
| | neg | 3 | 68 | 71 |
| | Total | 94 | 68 | 162 |
| TPPA | | |
|------|-----|-------|
| pos | neg | Total |
| 94 | 0 | 94 |
| 0 | 68 | 68 |
| | | |
| 94 | 68 | 162 |
9
{9}------------------------------------------------
| Percent Positive Agreement = 96.8% (95% C.I.= 91.0 - 99.3%) | 100.0% (95% C.I.= 96.2 - 100%) |
|-------------------------------------------------------------|--------------------------------|
| Percent Negative Agreement = 100.0% (95% C.I.= 94.7 - 100%) | 100.0% (95% C.I.= 94.7 - 100%) |
| Percent Overall Agreement = 98.1% (95% C.I.= 94.7 - 99.6%) | 100.0% (95% C.I.= 97.7 - 100%) |
#### PRECISION and REPRODUCIBILITY
Studies were performed to demonstrate the Intra-Assay, and Inter-Lot reproducibility of the Syphilis Health Check Test kit.
To demonstrate the reproducibility of the Syphilis Health Check test three laboratory sites using two trained technicians at each site to perform the testing for within-Run and between day reproducibility using a panel of 6 pooled samples. Each testing site conducted reproducibility studies using a supplied panel ranging from non-reactive to highly reactive, i.e. one nonreactive, one borderline, one borderline reactive, one moderate reactive, and two mid-high to high reactive in addition to the kit controls. Each site ran these panel member solutions for at least 5 days, twice per day. Each site performed one Within-Run assay by each of the operators each run for 10 times on one day. The one negative and three low to high positive samples had 100% agreement. The two borderline samples near the cut-off vielded the following results.
The following summary demonstrates the Reproducibility of the two critical borderline samples obtained by the Technicians.
Intra-Run
Panel D (Borderline Reactive)
| | Positive | Negative | Expected | % Agree | Discrepant |
|------------------------|----------|----------|-------------|---------|------------|
| Site 1 (2 operators) | 57 | 3 | 60 positive | 95.0 | 3 |
| Site 2 (2 operators) | 58 | 2 | 60 positive | 96.7 | 2 |
| In-house (2 operators) | 59 | 1 | 60 positive | 98.3 | 1 |
Panel E (Borderline Nonreactive)
| | Positive | Negative | Expected | % Agree | Discrepant |
|----------|----------|----------|-------------|---------|------------|
| Site 1 | 1 | 59 | 60 negative | 98.3 | 1 |
| Site 2 | 2 | 58 | 60 negative | 96.7 | 2 |
| In-House | 0 | 60 | 60 negative | 100.0 | 0 |
{10}------------------------------------------------
# Inter-Day (5 days)
# Panel D (Borderline Reactive)
| | Positive | Negative | Expected | % Agree | Discrepant |
|------------------------|----------|----------|-------------|---------|------------|
| Site 1 (2 operators) | 57 | 3 | 60 positive | 95.0 | 3 |
| Site 2 (2 operators) | 58 | 2 | 60 positive | 96.7 | 2 |
| In-house (2 operators) | 57 | 3 | 60 positive | 95.0 | 3 |
Panel E (Borderline Nonreactive)
| | Positive | Negative | Expected | % Agree | Discrepant |
|----------|----------|----------|-------------|---------|------------|
| Site 1 | 2 | 58 | 60 negative | 96.7 | 2 |
| Site 2 | 2 | 58 | 60 negative | 96.7 | 2 |
| In-House | 1 | 59 | 60 negative | 98.3 | 1 |
## Inter-Lot
Studies were performed to demonstrate the Inter-Lot assay reproducibility of the Syphilis Health Check Test kits. The Inter-Lot study used four dilutions of a positive pool and negative sera samples. These five samples were run in duplicate on 5 different lots to demonstrate lot-lot reproducibility run on the same day by the same technician.
The following table demonstrates the reproducibility of the Syphilis Health Check Test over 5 lots of kits.
| Dilution | Results after 10 min. | | | | | |
|----------------|-----------------------|-------|--------|---------|--------|-------|
| | Expected<br>results | Lot I | Lot II | Lot III | Lot IV | Lot V |
| Negative serum | - | - | - | - | - | - |
| 1/104 | - | - | - | - | - | - |
| 1/103 | +/- | +/- | +/- | +/- | +/- | +/- |
| 1/102 | + | + | + | + | + | + |
| 1/10 | + | + | + | + | + | + |
{11}------------------------------------------------
## EXPECTED VALUES
To assess the normal range (Expected Values) of the Syphilis Health Check test a series of ninety-eight (98) samples were obtained from various hospital laboratories in different U.S. geographical locations. The samples came from a general "presumed" healthy normal population (ages 20 - 66) whose serum or plasma were collected by hospital laboratories for routine serology testing, not related to STD.
All of the samples were tested with RPR, TPPA and Syphilis Health Check tests. Two samples were found positive by the treponemal tests with both samples non-reactive with RPR. These 2 samples were confirmed with MHA-TP.
The results of the testing of a presumed normal population obtained the following results for the Syphilis Health Check test: This resulted in a Overall Agreement of 100%. Therefore from this study of 98 presumed normal samples 2 were confirmed positive resulting in a 2% positivity rate.
In a prospectively collected population of drug users visiting STD clinics, of 138 males and 13 females (ages 18 - 61 yrs) thirteen (8.6%) were found to be Syphilis Health Check positive with eleven of those also reference method positive.
In a series of 69 pregnant women, ages 20 - 40 yrs, that were tested for routine screening, one sample (1.4%) was found to be Syphilis Health Check positive and confirmed by reference methods.
Eight hundred eighty (880) patients were prospectively collected in a population of individuals visiting STD and a hospital clinics and POC sites complaining and/or exhibiting signs and symptoms of STD infections, ages 16 - 81 vrs, and having 53% to 47% male to female ratio. It was found that a range of 3% - 6% of the patients identified as suspected syphilis were found positive by Syphilis Health Check and yielded a Percent positive agreement to RPR of 98.3% and 95.7% to treponemal tests. Percent Negative Agreement to RPR and Treponemal test was 93.4% and 97.8% respectively. Overall Agreement to RPR and Treponemal tests was 94.1% and 97.4% respectively.
These results are consistent with published rates for prevalence of antibody in the adult population. Prevalence may vary depending on a variety of factors such as geography, age, socio-economic status, ethnic background, type of test employed, specimen collection and handling procedures, clinical and epidemiological history.
{12}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/12/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
Diagnostics Direct, LLC c/o Gary Lehnus President Lehnus & Associates Consulting 150 Cherry Lane Road East Stroudsburg, PA 18301
AUG - 1 2011
Re: K102400
Trade/Device Name: Syphilis Health Check Treponemal Antibody Test; Rapid immunochromatographic membrane assay, Treponema pallidum Regulation Number: 21 CFR 866.3830 Regulation Name: Treponema pallidum treponemal test reagents Regulatory Class: Class II Product Code: LIP Dated: July 21, 2011 Received: July 27, 2011
Dear Mr. Lehnus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000
{13}------------------------------------------------
1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Sef atten
Sally A. Hoivat, M.S Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{14}------------------------------------------------
# . Indications for Use
510(k) Number (if known):
Device Name: Syphilis Health Check Test
Indications For Use:
Syphilis Health Check is a qualitative rapid membrane immunochromatographic assay for the detection of Treponema pallidum (syphilis) antibodies in human whole blood, serum or plasma. This product can be used as an initial screening test or in conjunction with a non-treponemal laboratory test and clinical findings to aid in the diagnosis of syphilis infection. This test is not intended for use in screening blood or plasma donors.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (OIVD)
Division Sign-Off
**Division Sign-Off**
**Office of In Vitro Diagnostic Device Evaluation and Safety**
510(K) K102400.
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.