AXSYM TOXO IGG ANTIBODY ASSAY

K954575 · Abbott Laboratories · LGD · Aug 16, 1996 · Microbiology

Device Facts

Record IDK954575
Device NameAXSYM TOXO IGG ANTIBODY ASSAY
ApplicantAbbott Laboratories
Product CodeLGD · Microbiology
Decision DateAug 16, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3780
Device ClassClass 2

Indications for Use

The AxSYM Toxo IgG Antibody assay is a Microparticle Enzyme Immunoassay (MEIA) for the quantitative measurement of IgG antibodies to Toxoplasma gondii in human serum or plasma (EDTA, heparin or sodium citrate) to aid in the determination of immune status. This product is not FDA cleared for use in testing blood or plasma donors.

Device Story

Microparticle Enzyme Immunoassay (MEIA) for quantitative detection of IgG antibodies to Toxoplasma gondii; utilizes human serum or plasma (EDTA, heparin, sodium citrate) as input. Device employs polystyrene microparticles coated with T. gondii antigen (RH strain) and goat anti-human IgG conjugated to alkaline phosphatase; 4-Methylumbelliferyl Phosphate (MUP) serves as enzyme substrate. Performed on automated AxSYM instrument. Output provides quantitative IgG antibody levels to assist clinicians in determining patient immune status regarding toxoplasmosis. Intended for clinical laboratory use.

Clinical Evidence

Clinical study conducted at two U.S. sites and one European site using 1400 specimens from pregnant women and low-risk individuals. Compared AxSYM Toxo IgG to VIDAS Toxo IgG. Discordant results resolved via IMx Toxo IgG, Platelia Toxo IgG, and Sabin-Feldman dye test. Results: relative sensitivity 99.66%, relative specificity 99.10%, relative agreement 99.34%. Precision (CVs) on positive panel members and controls ranged from 7.6% to 12.7%.

Technological Characteristics

MEIA technology; polystyrene microparticles coated with T. gondii antigen (RH strain); goat anti-human IgG conjugated to alkaline phosphatase; 4-Methylumbelliferyl Phosphate (MUP) substrate. Automated immunoassay analyzer platform. Quantitative measurement.

Indications for Use

Indicated for the quantitative measurement of IgG antibodies to Toxoplasma gondii in human serum or plasma to aid in the determination of immune status. Not for use in screening blood or plasma donors.

Regulatory Classification

Identification

Toxoplasma gondii serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Toxoplasma gondii in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identify Toxoplasma gondii from clinical specimens. The identification aids in the diagnosis of toxoplasmosis caused by the parasitic protozoan Toxoplasma gondii and provides epidemiological information on this disease. Congenital toxoplasmosis is characterized by lesions of the central nervous system, which if undetected and untreated may lead to brain defects, blindness, and death of an unborn fetus. The disease is characterized in children by inflammation of the brain and spinal cord.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} AUG 16 1996 Abbott Laboratories AxSYM Toxo IgG Antibody Assay [510(k)] Submission K954575 # ATTACHMENT A [510(k)] Summary of Safety and Effectiveness Information Supporting A Substantially Equivalent Determination ATTACH-A. page 1 of 5 0038 {1} Abbott Laboratories AxSYM Toxo IgG Antibody Assay [510(k)] Submission # [510(k)] Summary of Safety and Effectiveness Information Supporting a Substantially Equivalent Determination The following information presented in the [510(k)] notification for the AxSYM Toxo IgG antibody assay constitutes data supporting a substantially equivalent determination: # [510(k)] Summary of Device Performance The AxSYM Toxo IgG Antibody assay is a Microparticle Enzyme Immunoassay (MEIA) for the quantitative measurement of IgG antibodies to Toxoplasma gondii in human serum or plasma (EDTA, heparin or sodium citrate) to aid in the determination of immune status. This product is not FDA cleared for use in testing blood or plasma donors. The predicate device for determination of substantial equivalence is the bioMerieux VIDAS Toxo IgG assay. The VIDAS Toxo IgG (TXG) assay is intended for use with a VIDAS (Vitek ImmunoDiagnostic Assay System) instrument as a semi-quantitative automated enzyme-linked fluorescent immunoassay (ELFA). It is intended for use in determination of Toxoplasma gondii immunological experience from a single serum sample, or as an aid in the diagnosis of T. gondii recent infection or reactivation through evaluation of paired sera for a significant increase in T. gondii-specific IgG. It is not intended for use in testing (screening) blood or plasma donors. In two U.S. sites and one European site, the AxSYM Toxo IgG Antibody assay was compared to the VIDAS Toxo IgG Antibody assay using 1400 specimens from pregnant women and random low risk individuals. Discordant results were resolved by testing with IMx Toxo IgG Antibody assay and Platelia Toxo IgG, both of which are legally marketed devices. Discordant results were also resolved by the Sabin-Feldman dye test, a reference laboratory method for the determination of antibody to Toxoplasma gondii in human serum. This method was first developed in 1948 and is still considered the most sensitive and specific method for the detection of IgG antibodies to toxoplasma. Based on this study, the AxSYM Toxo IgG showed relative sensitivity of 99.66%, relative specificity of 99.10% and relative agreement of 99.34%. Specimens giving equivocal results were not included in the calculation of relative agreement, relative sensitivity or relative specificity. Percent CV's on positive panel members and positive control were 7.6% to 12.7%. ATTACH-A. page 2 of 5 0039 {2} Abbott Laboratories AxSYM Toxo IgG Antibody Assay [510(k)] Submission In conclusion, the AxSYM Toxo IgG antibody assay is substantially equivalent to the bioMerieux VIDAS Toxo IgG antibody assay for the detection of IgG antibodies to toxoplasmosis in human serum and plasma (EDTA, heparin, or sodium citrate) samples. ATTACH-A page 3 of 5 0040 {3} Abbott Laboratories AxSYM Toxo IgG Antibody Assay [510(k)] Submission # [510(k)] Summary Of Technological Comparison The AxSYM Toxo IgG Antibody Assay and the VIDAS Toxo IgG Antibody assay are substantially equivalent in that: A. Both are in vitro immunologic test methods. B. Both are intended for use in the detection of IgG antibody to Toxoplasmosis gondii in human serum. C. Both are based on the formation of immune complexes between Toxoplasmosis gondii antigens and antibody. D. Both are quantitative assays. E. Both use automated immunoassay analyzers. F. Both use a polystyrene solid phase. G. Both use an anti-human IgG antibody conjugated to alkaline phosphatase. H. Both use 4-Methylumbelliferyl Phosphate (MUP) as the enzyme substrate. The AxSYM Toxo IgG Antibody Assay and VIDAS Toxo IgG Antibody assay differ in that: 1. The solid phase in the AxSYM Toxo IgG assay is polystyrene microparticles coated with Toxoplasmosis gondii antigen (RH strain derived from HeLa cell cultures). The solid phase for the VIDAS Toxo IgG assay consists of Toxoplasmosis gondii antigen coated to the inside of the SPR (solid phase receptacle). 2. The conjugate in the AxSYM Toxo IgG assay uses goat anti-human IgG antibody. The conjugate in the VIDAS Toxo IgG assay uses mouse monoclonal anti-human IgG antibody. 3. Plasma (EDTA, heparin, sodium citrate) specimens may be tested in the AxSYM Toxo IgG assay. The use of plasma specimens has not been validated in the VIDAS Toxo IgG assay. ATTACH-A page 4 of 5 0041 {4} Abbott Laboratories AxSYM Toxo IgG Antibody Assay [510(k)] Submission # Comparison of Methods | Assay Characteristics | AxSYM Toxo IgG Quantitative | VIDAS Toxo IgG Quantitative | | --- | --- | --- | | Assay Type | | | | Antibody Measured | Specific IgG | Specific IgG | | Assay Principle | MEIA | EIA(ELFA) | | Solid Phase | polystyrene microparticles | polystyrene solid phase receptacle | | Solid Phase Coating | Toxo antigen, RH strain derived from HeLa cell cultures | Toxo antigen, RH strain | | Conjugate | goat anti-human IgG conjugated to alkaline phosphatase | mouse monoclonal anti-human IgG conjugated to alkaline phosphatase | | Specimen | Human serum and plasma (EDTA, heparin, sodium citrate) | Human serum | | Automation | Performed on an automated instrument | Performed on an automated instrument | | Relative Specificity | 99.1% (European & U S ) | 97.8% (European study) 97.9% (U.S. study) | | Relative Sensitivity (final interpretation) | 99.6% (European & U S ) | 98.4% (European study) 96.0% (U.S. study) | ATTACH-A. page 5 of 5 0042
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...