Access Toxo IgG

K242022 · Beckman Coulter, Inc. · LGD · Mar 28, 2025 · Microbiology

Device Facts

Record IDK242022
Device NameAccess Toxo IgG
ApplicantBeckman Coulter, Inc.
Product CodeLGD · Microbiology
Decision DateMar 28, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3780
Device ClassClass 2

Indications for Use

The Access Toxo IgG assay is a paramagnetic-particle, chemiluminescent immunoassay for the qualitative and quantitative determination of IgG antibodies to Toxoplasma gondii in human serum, using the Access Immunoassay Systems. The Access Toxo IgG assay aids in the diagnosis of Toxoplasma gondii infection and may be used to assess the immune status of pregnant women. This product is not FDA cleared/approved for the screening of blood or plasma donors. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients, cord blood, neonatal specimens or infants.

Device Story

Access Toxo IgG is a paramagnetic-particle, chemiluminescent immunoassay for detecting T. gondii-specific IgG antibodies in adult human serum. The device operates on the DxI 9000 Access Immunoassay Analyzer. It utilizes a two-step immunoenzymatic process: patient serum is incubated with paramagnetic particles coated with solubilized T. gondii membrane antigen; a mouse monoclonal anti-human IgG antibody-alkaline phosphatase conjugate is used for detection. The system uses Lumi-Phos PRO substrate to generate a chemiluminescent signal proportional to the antibody concentration. The assay is performed in clinical laboratories by trained personnel. Results are used by clinicians to diagnose T. gondii infection and evaluate immune status in pregnant women. The device provides qualitative and quantitative outputs (IU/mL) to guide clinical management.

Clinical Evidence

No clinical studies were performed. Evidence consists of analytical performance testing (precision, linearity, LoB/LoD/LLoQ) and a method comparison study. The method comparison between the DxI 9000 and the predicate Access 2 system (n=140) showed 100% PPA (95% CI: 91.24%–100%) and 100% NPA (95% CI: 96.26%–100%).

Technological Characteristics

Paramagnetic-particle, chemiluminescent immunoassay. Reagents: T. gondii (strain RH) membrane antigen-coated particles, mouse monoclonal anti-human IgG-alkaline phosphatase conjugate. Substrate: Lumi-Phos PRO. Instrument: DxI 9000 Access Immunoassay Analyzer. Automated, two-step immunoenzymatic method. Calibration: Stored curve, 28-day frequency. Open kit stability: 28 days at 2-10°C.

Indications for Use

Indicated for qualitative and quantitative determination of IgG antibodies to Toxoplasma gondii in human serum to aid in diagnosis of infection and assessment of immune status in pregnant women. Not for blood/plasma donor screening. Performance not established for immunocompromised/immunosuppressed patients, cord blood, neonates, or infants.

Regulatory Classification

Identification

Toxoplasma gondii serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Toxoplasma gondii in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identify Toxoplasma gondii from clinical specimens. The identification aids in the diagnosis of toxoplasmosis caused by the parasitic protozoan Toxoplasma gondii and provides epidemiological information on this disease. Congenital toxoplasmosis is characterized by lesions of the central nervous system, which if undetected and untreated may lead to brain defects, blindness, and death of an unborn fetus. The disease is characterized in children by inflammation of the brain and spinal cord.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ## I Background Information: A 510(k) Number K242022 B Applicant Beckman Coulter Inc. C Proprietary and Established Names Access Toxo IgG D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | LGD | Class II | 21 CFR 866.3780 - Toxoplasma Gondii Serological Reagents | MI - Microbiology | ## II Submission/Device Overview: A Purpose for Submission: Modification of a previously cleared device - Clearance of Access Toxo IgG assay to add the DxI 9000 Access Immunoassay Analyzer B Measurand: IgG antibodies to Toxoplasma gondii C Type of Test: Chemiluminescent Immunoassay Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {1} III Intended Use/Indications for Use: A Intended Use(s): See Indications for Use below. B Indication(s) for Use: The Access Toxo IgG assay is a paramagnetic-particle, chemiluminescent immunoassay for the qualitative and quantitative determination of IgG antibodies to *Toxoplasma gondii* in human serum using the Access Immunoassay Systems. The Access Toxo IgG assay aids in the diagnosis of *Toxoplasma gondii* infection and may be used to assess the immune status of pregnant women. This product is not FDA cleared/approved for the screening of blood or plasma donors. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients, cord blood, neonatal specimens or infants. C Special Conditions for Use Statement(s): Rx - For Prescription Use Only D Special Instrument Requirements: DxI 9000 Access Immunoassay Analyzer IV Device/System Characteristics: A Device Description: The Access Toxo IgG assay is a paramagnetic-particle chemiluminescent immunoassay for the qualitative and quantitative detection of *Toxoplasma gondii*-specific IgG antibody in adult human serum using the Access Immunoassay Systems. The Access Toxo IgG assay consists of the reagent pack, calibrators, and quality controls (QCs), packaged separately. Other items needed to run the assay include substrate and wash buffer. B Principle of Operation: The Access Toxo IgG assay is an enzyme immunoassay using an indirect technique. A sample is added to a reaction vessel with paramagnetic particles coated with *Toxoplasma gondii* membrane antigen. Specific antibodies present in the sample bind to the antigen. After incubation in a reaction vessel, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Alkaline phosphatase-conjugated monoclonal anti-human IgG antibody is then added and attaches to the IgG antibodies captured on the particles. A second separation and wash steps remove unbound conjugate. Then, the chemiluminescent substrate is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is directly proportional to the concentration of Toxo IgG antibody in the sample. The amount of analyte in the sample is determined from a stored, multi-point calibration curve. K242022 - Page 2 of 9 {2} The Access Toxo IgG assay results interpretation is as follows: | Results | Results Interpretation | Comment | | --- | --- | --- | | < 7.5 IU/mL | Non-reactive | Samples are considered non-reactive for the presence of Toxoplasma gondii IgG antibodies. A non-reactive result may indicate that an individual has not been infected with Toxoplasma gondii and is thus susceptible to infection. A non-reactive result, however, does not rule out an acute infection. If exposure to the pathogen is suspected despite an initial non-reactive result, a second sample should be collected and tested at least two to three weeks later. | | ≥ 7.5 IU/mL to < 10.5 IU/mL | Equivocal | Samples are considered equivocal for the presence of Toxoplasma gondii IgG antibodies. Specimens that are equivocal may contain low levels of anti-Toxo IgG. As any cutoff is a statistically determined value, it is recommended that results close to the cutoff are interpreted cautiously and further investigated. If exposure to the pathogen is suspected despite an initial equivocal result, a second sample should be collected and tested at least two to three weeks later. | | ≥ 10.5 IU/mL | Reactive | Samples are considered reactive for the presence of Toxoplasma gondii IgG antibodies. A reactive result is generally indicative of recent or past exposure to the pathogen. | # V Substantial Equivalence Information: # A Predicate Device Name(s): Access Toxo Igg, Access Toxo Igg Calibrators And Access Toxo Igg Qc # B Predicate 510(k) Number(s): K080869 # C Comparison with Predicate(s): <table><tr><td>Candidate &amp; Predicate Devices:</td><td>Candidate K242022</td><td>Predicate K080869</td></tr><tr><td>Device Trade Name</td><td>Same</td><td>Access Toxo IgG</td></tr><tr><td>General Device Characteristic Similarities</td><td></td><td></td></tr><tr><td>Intended Use/Indications For Use</td><td>Same</td><td>The Access Toxo IgG assay is a paramagnetic-particle, chemiluminescent immunoassay for the qualitative and quantitative determination of IgG antibodies to Toxoplasma gondii in human serum using the Access Immunoassay Systems. The Access Toxo IgG assay aids in the diagnosis of</td></tr><tr><td></td><td></td><td>in human serum using the Access Immunoassay Systems. The Access Toxo IgG assay is a non-reactive test for the presence of Toxoplasma gondii IgG antibodies. The use of the Access Toxo IgG assay is a non-reactive test for the presence of Toxoplasma gondii IgG antibodies. The use of the Access Toxo IgG assay is a non-reactive test for the presence of Toxoplasma gondii IgG antibodies. The use of the Access Toxo IgG assay is a non-reactive test for the presence of Toxoplasma gondii IgG antibodies. The use of the Access Toxo IgG assay is a non-reactive test for the presence of Toxoplasma gondii IgG antibodies. The use of the Access Toxo IgG assay is a non-reactive test for the presence of Toxoplasma gondii IgG antibodies. The use of the Access Toxo IgG assay is a non-reactive test for the presence of Toxoplasma gondii IgG antibodies. The use of the Access Toxo IgG assay is a non-reactive test for the presence of Toxoplasma gondii IgG antibodies. The use of the Access Toxo IgG assay is a non-reactive test for the presence of Toxoplasma gondii IgG antibodies. The use of the Access Toxo IgG assay is a non-reactive test for the presence of Toxoplasma gondii IgG antibodies. The use of the Access Toxo IgG assay is a non-reactive test for the presence of Toxoplasma gondii IgG antibodies. The use of the Access Toxo IgG assay is a non-reactive test for the presence of Toxoplasma gondii IgG antibodies. The use of the Access Toxo IgG assay is a non-reactive test for the presence of Toxoplasma gondii IgG antibodies. The use of the Access Toxo IgG assay is a non-reactive test for the presence of Toxoplasma gondii IgG antibodies. The {3} | Candidate & Predicate Devices: | Candidate K242022 | Predicate K080869 | | --- | --- | --- | | | | Toxoplasma gondii infection and may be used to assess the immune status of pregnant women. This product is not FDA cleared/approved for the screening of blood or plasma donors. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients, cord blood, neonatal specimens or infants. | | Analyte | Same | IgG antibody to T. gondii | | Technology | Same | Two-step immunoenzymatic assay | | Format | Same | Chemiluminescent | | Method | Same | Automated | | Calibration | Same | Utilizes a stored calibration curve | | Calibration Frequency | Same | 28 days | | Sample Type | Same | Serum | | Analytical Measuring Interval | 3.2 – 450 IU/mL | 0 – 450 IU/mL | | Results Interpretation | Same | <7.5 IU/mL: Non-Reactive ≥7.5 - <10.5 IU/mL: Equivocal ≥10.5 IU/mL: Reactive | | Capture Reagent | Same | Paramagnetic particles coated with T. gondii (strain RH) solubilized membrane antigen | | Detection Antibody | Same | Mouse monoclonal anti-human IgG antibody- alkaline phosphatase (bovine) conjugate | | Open Kit Stability | Same | 28 days after opening, 2-10°C | | General Device Characteristic Differences | | | | Substrate | Lumi-Phos PRO substrate | Access Substrate | | Instrument | DxI 9000 Access Immunoassay Analyzer | Access 2 Immunoassay System | VI Standards/Guidance Documents Referenced: The following Clinical and Laboratory Standards Institute (CLSI) guidelines were used: - CLSI EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures K242022 - Page 4 of 9 {4} - CLSI EP06-Ed 2: Evaluation of the Linearity of Quantitative Measurement Procedures - CLSI EP09c-Ed 3: Measurement Procedure Comparison and Bias Estimation Using Patient Samples - CLSI EP12-Ed3, Evaluation of Qualitative, Binary Output Examination Performance, Third Edition # VII Performance Characteristics (if/when applicable): # A. Analytical Performance: # 1. Precision/Reproducibility: a) Within-Laboratory Precision: A 20-day within-laboratory precision study was conducted using 3 lots of the Access Toxo IgG reagents, 3 lots of the Access Toxo IgG calibrators and 3 DxI 9000 Access Immunoassay Analyzer systems. Six human serum samples were tested in 2 replicates at 2 runs per day over 20 days using 3 reagent lot/calibrator lot combinations, where a unique reagent lot and calibrator lot are paired. The data were analyzed for repeatability (within-run), between-run, between-day, and within-laboratory precision. Within-laboratory precision data summary is shown in Table 1. Table 1: Access Toxo IgG 20-day Within-Laboratory Precision | Sample | N | Mean (IU/mL) | Within-Run (Repeatability) | | Between-Run | | Between-Day | | Between Lot and Instrumenta | | Overall Precisionb | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | Sample 1 | 240 | 2.7 | 0.15 | 5.3 | 0.16 | 5.8 | 0.10 | 3.6 | 0.29 | 10.8 | 0.38 | 13.9 | | Sample 2 | 240 | 8.3 | 0.36 | 4.3 | 0.24 | 2.9 | 0.43 | 5.2 | 0.61 | 7.3 | 0.86 | 10.4 | | Sample 3 | 240 | 13.1 | 0.56 | 4.3 | 0.54 | 4.2 | 0.35 | 2.7 | 1.34 | 10.2 | 1.59 | 12.1 | | Sample 4 | 240 | 69.3 | 3.03 | 4.4 | 3.17 | 4.6 | 2.17 | 3.1 | 6.56 | 9.5 | 8.18 | 11.8 | | Sample 5 | 240 | 174.2 | 7.29 | 4.2 | 6.80 | 3.9 | 6.28 | 3.6 | 14.01 | 8.0 | 18.30 | 10.5 | | Sample 6 | 240 | 361.0 | 20.45 | 5.7 | 12.55 | 3.5 | 20.38 | 5.6 | 14.07 | 3.9 | 34.48 | 9.6 | a Access Toxo IgG reagent lot, Access Toxo IgG calibrator lot and DxI 9000 instruments are confounded, and the confounding effect is represented by between-lot and instrument. b Overall within-laboratory variability includes within-run, between-run, between-day, and between-lot variance components. b) Reproducibility (between-Instrument Precision): An additional precision study was conducted over 5-day by testing six serum samples on 3 DxI 9000 Access Immunoassay Analyzers in an internal site. The samples were tested with 3 lots of Access Toxo IgG reagents, and 1 lot of Access Toxo IgG calibrator on each instrument with 5 replicates per run and 1 run per day over 5 days. Between-instrument reproducibility data is shown in Table 2. K242022 - Page 5 of 9 {5} Table 2: Access Toxo IgG Assay Reproducibility (Between-Instrument Precision) | Sample | N | Mean (IU/mL) | Within-Run (Repeatability) | | Between-Day/Runsa | | Between Reagent Lot | | Between-Instruments | | Reproducibilityb | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | Sample 1 | 225 | 2.5 | 0.16 | 6.4 | 0.17 | 7.0 | 0.11 | 4.5 | 0.14 | 5.6 | 0.29 | 11.9 | | Sample 2 | 225 | 7.9 | 0.40 | 5.0 | 0.42 | 5.3 | 0.25 | 3.1 | 0.35 | 4.5 | 0.72 | 9.1 | | Sample 3 | 225 | 12.4 | 0.57 | 4.6 | 0.51 | 4.1 | 0.58 | 4.7 | 0.55 | 4.4 | 1.11 | 8.9 | | Sample 4 | 225 | 67.4 | 3.17 | 4.7 | 2.96 | 4.4 | 3.66 | 5.4 | 2.90 | 4.3 | 6.37 | 9.4 | | Sample 5 | 225 | 169.7 | 7.83 | 4.6 | 9.04 | 5.3 | 9.64 | 5.7 | 6.24 | 3.7 | 16.58 | 9.8 | | Sample 6 | 225 | 377.5 | 25.58 | 6.8 | 30.77 | 8.2 | 22.70 | 6.0 | 21.45 | 5.7 | 50.76 | 13.4 | a Days and runs are confounded. b Reproducibility includes within-run, between-run, between-day, and between-lot variance components. c) Reproducibility (between-site precision): Refer to K080869. # 2. Linearity: Linearity of the Access Toxo IgG on the DxI 9000 Access Immunoassay Analyzer was evaluated following CLSI Guideline EP06, 2nd Edition "Evaluation of Linearity of Quantitative Measurement Procedures". A tiled approach was used to evaluate the linearity of five segments of the measuring range. Five sample panels were generated by diluting unique pairs of high and low Toxo IgG antibody positive serum samples in various ratios. Each panel consisted of 7 - 9 dilutions in addition to high and low concentration samples. The table below indicates the number of dilutions and concentration ranges covered by each sample panel. | Panel | Concentration Range (IU/mL) | Number of Samples per Panel | | --- | --- | --- | | 1 | 3.08 – 128.59 | 11 | | 2 | 77.44 – 196.66 | 9 | | 3 | 176.17 – 323.84 | 9 | | 4 | 248.91 – 399.72 | 9 | | 5 | 363.24 – 475.99 | 9 | This study was run on one DxI 9000 Access Immunoassay Analyzer, using one reagent lot and one calibrator lot. Low sample was run in replicates of eight, and all other samples were run in replicates of six. Two quality controls were run in replicates of two. The linearity of the Access Toxo IgG assay was considered acceptable if the data set was determined to be linear and if the deviation from linearity within or equal to $\pm 10\%$ for samples with concentrations $\geq 3.2$ IU/mL. This assay is linear across the analytical measuring interval of 3.2 to $450.0~\mathrm{IU / mL}$ . K242022 - Page 6 of 9 {6} ![img-0.jpeg](img-0.jpeg) Regression equation: $y = 0.9685x - 0.02095$ 3. Analytical Specificity: Interference and Cross-Reactivity: Refer to K080869. 4. Assay Reportable Range: 3.2 IU/mL - 450.0 IU/mL. See linearity above. 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): Refer to K080869. The Access Toxo IgG Calibrators are referenced to the Third International Standard Preparation for Anti-Toxoplasma Serum (NIBSC code: TOXM). 6. Detection Limit: The Limit of Blank (LoB), Limit of Detection (LoD) and Lower Limit of Quantitation (LLoQ) for the Access Toxo IgG assay on Dxl 9000 Access Immunoassay Analyzer were established following CLSI guideline EP17-A2 (Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline - Second Edition). In order to obtain at least 60 replicates of each sample level, the LoB was established with 5 zero-analyte samples tested in replicates of 5 over 3 days; the LoD with 6 low-analyte samples tested in replicates of 9 over 5 days; and the LLoQ with 11 low-analyte samples tested in replicates of 9 over 5 days. Detection Limit for the Access Toxo IgG assay on the Dxl 9000 Access Immunoassay Analyzer are: $\mathrm{LoB} = 0.8 \mathrm{IU} / \mathrm{mL}$ . $\mathrm{LoD} = 1.5 \mathrm{IU} / \mathrm{mL}$ . $\mathrm{LLoQ} = 3.2 \mathrm{IU} / \mathrm{mL}$ . K242022 - Page 7 of 9 {7} 7. Assay Cut-Off: Refer to K080869. 8. Carry-Over: Not applicable. ## B. Comparison Studies: a. Comparison of Access Toxo IgG assay on the Access 2 Immunoassay Analyzer to the DxI 9000 Access Immunoassay Analyzer: A method comparison study was conducted to compare the performance of the Access Toxo IgG Assay on DxI 9000 Access Immunoassay Analyzer to the Access Toxo IgG Assay on the Access 2 Immunoassay System. A total of 140 native serum samples were tested. Positive percent agreement (PPA) and negative percent agreement (NPA) between the DxI 9000 Access Immunoassay Analyzer and the Access 2 Immunoassay system was calculated for the Access Toxo IgG assay and shown in Table 3. Table 3: Performance Agreement of Access Toxo IgG assay on the Access 2 Immunoassay Analyzer and the DxI 9000 Access Immunoassay Analyzer (n=140) | Access Toxo IgG | Access 2 Immunoassay Analyzer | | | | | | --- | --- | --- | --- | --- | --- | | | | Reactive | Equivocal | Non-reactive | Total | | DxI 9000 Access Immunoassay Analyzer | Reactive | 40 | 0 | 0 | 40 | | | Equivocal | 0 | 1 | 0 | 1 | | | Non-Reactive | 0 | 0 | 99 | 99 | | | Total | 40 | 1 | 99 | 140 | | | PPA^{a} | 100.00% (40/40) | 95% CI = 91.24% to 100.00% | | | | | NPA^{b} | 100.00% (99/99) | 95% CI = 96.26% to 100.00% | | | a, b 95% CI for PPA and NPA were estimated using the Wilson score method. K242022 - Page 8 of 9 {8} b. Matrix Comparison: Not applicable. C Clinical Studies: 1. Clinical Sensitivity: Not applicable. 2. Clinical Specificity: Not applicable. 3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): For clinical agreement study, refer to K080869. D Clinical Cut-Off: Not applicable. E Expected Values/Reference Range: Refer to K080869. VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. K242022 - Page 9 of 9
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...