The Platelia™ Toxo IgM kit is an in-vitro diagnostic test kit allowing the qualitative detection of anti-Toxoplasma gondii in human serum or plasma (EDTA, Heparin, Citrate)
Device Story
Platelia™ Toxo IgM is an in-vitro diagnostic test kit for qualitative detection of anti-Toxoplasma gondii IgM antibodies in human serum or plasma. Used in clinical laboratory settings by trained personnel. Operates as an immunoassay to detect specific IgM antibodies; results are interpreted by healthcare providers in conjunction with IgG antibody assay results to support diagnosis of acute, recent, or reactivated Toxoplasma gondii infection. Provides presumptive diagnostic information to assist clinical decision-making regarding patient infection status.
Clinical Evidence
No clinical data provided in the document.
Technological Characteristics
In-vitro diagnostic immunoassay kit for qualitative detection of anti-Toxoplasma gondii IgM antibodies. Compatible with human serum or plasma (EDTA, Heparin, Citrate).
Indications for Use
Indicated for the qualitative detection of IgM antibodies to Toxoplasma gondii in human serum.
Regulatory Classification
Identification
Toxoplasma gondii serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Toxoplasma gondii in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identify Toxoplasma gondii from clinical specimens. The identification aids in the diagnosis of toxoplasmosis caused by the parasitic protozoan Toxoplasma gondii and provides epidemiological information on this disease. Congenital toxoplasmosis is characterized by lesions of the central nervous system, which if undetected and untreated may lead to brain defects, blindness, and death of an unborn fetus. The disease is characterized in children by inflammation of the brain and spinal cord.
K242095 — Access Toxo IgM II · Beckman Coulter, Inc. · Oct 11, 2024
K971672 — SERAQUEST TOXOPLASMA IGM · Quest Intl., Inc. · Oct 21, 1997
K031506 — ACCESS IMMUNOASSAY SYSTEM TOXO IGM II ASSAY · Beckman Coulter, Inc. · Jun 2, 2003
Submission Summary (Full Text)
{0}
SPECIAL 510(k): Device Modification
ODE Review Memorandum (Decision Making Document is Attached)
To: THE FILE
RE: DOCUMENT NUMBER K090847,
Bio-Rad Platelia™ Toxo IgM
This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary):
1. The name and 510(k) number of the SUBMITTER'S previously cleared device.
Platelia Toxo IgM TMB kit, K013837
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use).
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification.
3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
This change was for a product name change, catalogue number, sample volume, sample dilution, reagent composition, preservative change, quality control criteria, procedure and stability.
The fundamental scientific technology of the modified device has not changed. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed.
4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, and test performance.
Table: Summary of Design Modifications and Required Validation Activities
| Design Change Category | Description of Change | Design Validation |
| --- | --- | --- |
| Composition / Concentration | Preservative change: replaced thimerosal with Proclin™300 in the following reagents: - Wash Solution - Negative Control - Cutoff Control - Pos Control - Conjugate - Diluent | - Precision Study - Interference Study - Determination of the Cut-Off Value - Cross Reactivity Study - Clinical Evaluation - Wash Solution 10x vs 20x Study - Robustness Study - Stability Studies |
| | Ready-To-Use reagent: the Chromogen / TMB Substrate solution has been prepared as a ready to use reagent and no longer requires preparation by the user prior to use. | |
{1}
2
| | Concentration change: the re-constitution instructions have been changed for the following reagents.
- Wash Solution (formerly 10x; now 20x)
- Conjugate (formerly 50x; now 101x) | |
| --- | --- | --- |
| Design Change Category | Description of Change | Design Validation |
| Procedure | Sample dilution change: the sample dilution has been changed from 1/101 to 1/21. The procedure formerly required 10 μl of sample and now requires 15 μl of sample. | - Precision Study
- Interference Study
- Determination of the Cut-Off Value
- Cross Reactivity Study
- Clinical Evaluation
- Wash Solution 10x vs 20x Study
- Robustness Study
- Stability Studies |
| | Washing procedure changes: two washing procedures have changed as follows:
- Pre-washing of the microplate prior to the addition of samples is no longer required.
- Four washing cycles are now required before addition of the Conjugate, where three cycles were required previously. | |
| | Quality control criteria change: the quality control criteria have been changed as follows:
- The negative control criteria has changed from ≤ 0.100 OD to ≤ 0.150 OD
- The positive control criteria have been replaced with a mean calibrator OD criteria requiring the mean calibrator OD to be ≥ 0.300 OD.
- The ratio and corresponding acceptance criteria for the mean cut-off control OD vs. the negative control OD have been inverted.
The following quality control criteria remain unchanged:
- The requirement that the individual OD of each replicate of the calibrator must not differ more than 20% of the cutoff value.
- The ratio of the positive control OD to the mean calibrator OD must be ≥ 1.80. | |
{2}
3
| Design Change Category | Description of Change | Design Validation |
| --- | --- | --- |
| Procedure (cont...) | Stability information changes: the reagent stability claims have changed as follows:
- The stability claim for the microplate strips has been increased from one month to 8 weeks when stored at 2 – 8°C.
- The stability claims for the negative control, calibrator, positive control, conjugate, diluent, and chromogen solution have clarified to be stable for up to 8 weeks once opened and stored at 2 – 8°C.
- The stability claim for the wash solution has been clarified to be stable for up to 2 weeks once diluted and stored at 2 – 30°C. Once opened the concentrated washing solution is stable until the expiration date indicated on the label when stored at 2 – 30°C.
- The stability claim for diluted T. gondii antigen was removed as the reconstituted antigen solution is for immediate use.
- The stability claim for diluted conjugate has been clarified to be stable after reconstitution for 8 hours at 18 – 30°C and 4 weeks at 2 – 8°C. | - Precision Study
- Interference Study
- Determination of the Cut-Off Value
- Cross Reactivity Study
- Clinical Evaluation
- Wash Solution 10x vs 20x Study
- Robustness Study
- Stability Studies |
5. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied
{3}
4
c) A declaration of conformity with design controls. The declaration of conformity should include:
i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and
ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review.
The sponsor submitted all the Design Control studies and they are acceptable. The submitter has provided the design control information as specified in The New 510(k) Paradigm.
6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.