← Product Code [GQG](/submissions/MI/subpart-d%E2%80%94serological-reagents/GQG) · K070747

# QUICK VUE RSV TEST, MODELS 20193, 20199 (K070747)

_Quidel Corp. · GQG · Apr 23, 2007 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/GQG/K070747

## Device Facts

- **Applicant:** Quidel Corp.
- **Product Code:** [GQG](/submissions/MI/subpart-d%E2%80%94serological-reagents/GQG.md)
- **Decision Date:** Apr 23, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3480
- **Device Class:** Class 1
- **Review Panel:** Microbiology
- **Attributes:** Pediatric

## Indications for Use

The QuickVue RSV test is a dipstick immunoassay, which allows for the rapid, qualitative detection of respiratory syncytial virus (RSV) antigen (viral fusion protein) directly from nasopharyngeal swab, nasopharyngeal aspirate, or nasal/nasopharyngeal wash specimens for symptomatic pediatric patients (eighteen years of age and younger). The test is intended for use as an aid in the diagnosis of acute respiratory syncytial viral infections. It is recommended that negative test results be confirmed by cell culture. Negative results do not preclude RSV infection and it is recommended that they not be used as the sole basis for treatment or other management decisions. The test is intended for professional and laboratory use.

## Device Story

Lateral-flow immunoassay; detects RSV viral fusion protein antigen. Input: nasopharyngeal swab, aspirate, or wash specimens. Process: specimen placed in extraction reagent to disrupt virus; test strip inserted into extraction tube for 15 minutes. Output: visual test line (pink-to-red) indicates presence of RSV antigen; blue procedural control line indicates valid test. Used in professional/laboratory settings by healthcare personnel. Results aid clinical diagnosis of acute respiratory syncytial viral infections; negative results require confirmation by cell culture and should not be the sole basis for management decisions.

## Clinical Evidence

Multi-center field clinical study conducted to evaluate performance characteristics. Sensitivity, specificity, and overall accuracy were calculated using nasal/nasopharyngeal wash specimens compared to viral culture as the reference method.

## Technological Characteristics

Lateral-flow immunoassay dipstick; visual readout; procedural control line; extraction reagent-based specimen preparation; professional/laboratory use; no electronic components or software.

## Regulatory Identification

Respiratory syncytial virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to respiratory syncytial virus in serum. Additionally, some of these reagents consist of respiratory syncytial virus antisera conjugated with a fluorescent dye (immunofluorescent reagents) and used to identify respiratory syncytial viruses from clinical specimens or from tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of respiratory syncytial virus infections and provides epidemiological information on diseases caused by these viruses. Respiratory syncytial viruses cause a number of respiratory tract infections, including the common cold, pharyngitis, and infantile bronchopneumonia.

## Predicate Devices

- Quidel QuickVue® RSV Test ([K061008](/device/K061008.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
K070747

B. Purpose for Submission:
New specimen matrix

C. Measurand:
Respiratory syncytial virus (RSV) antigen

D. Type of Test:
Dipstick immunoassay for the rapid, qualitative detection of RSV antigen

E. Applicant:
Quidel Corporation

F. Proprietary and Established Names:
QuickVue® RSV Test

G. Regulatory Information:
1. Regulation section: 21CFR 866.3480; Respiratory syncytial virus serological reagents
2. Classification: Class: I
3. Product code: GQG, Antigens, CF (including CF controls), respiratory syncytial virus.
4. Panel: 83 Microbiology

H. Intended Use:
The QuickVue RSV test is a dipstick immunoassay, which allows for the rapid, qualitative detection of respiratory syncytial virus (RSV) antigen (viral fusion protein) directly from nasopharyngeal swab, nasopharyngeal aspirate, or nasal/nasopharyngeal wash specimens for symptomatic pediatric patients (eighteen years of age and younger). The test is intended for use as an aid in the diagnosis of acute respiratory syncytial viral infections. It is recommended that negative test

{1}

results be confirmed by cell culture. Negative results do not preclude RSV infection and it is recommended that they not be used as the sole basis for treatment or other management decisions. The test is intended for professional and laboratory use.

2. Indication(s) for use: Same as intended use
3. Special conditions for use statement(s):

Prescription Use Only

4. Special instrument requirements:

None

I. Device Description:

The QuickVue RSV test is a dipstick immunoassay that allows the capture and visual detection of RSV antigen (viral fusion protein). The patient specimen is placed in the Extraction Tube containing the Extraction Reagent, enhancing the exposure of the viral fusion protein antigen. After extraction, the Test Strip is placed in the Extraction Tube where the RSV fusion proteins in the specimen will react with the reagents in the Test Strip.

If the extracted specimen contains RSV antigens, a pink-to-red Test Line, along with a blue procedural Control Line, will appear on the Test Strip indicating a positive result. If RSV type antigens are not present, or are present at very low levels, only a blue procedural Control Line will appear.

J. Substantial Equivalence Information:

1. Predicate device name(s):

Viral cell culture, Quidel QuickVue® RSV Test

2. Predicate 510(k) number(s): k061008

Comparison with predicate:

{2}

Table 1: Summary of Device Similarities and Differences

|  Features | QuickVue® RSV test (Proposed) | QuickVue® RSV test (K061008)  |
| --- | --- | --- |
|  Intended Use | The QuickVue RSV test is a dipstick immunoassay which allows for the rapid, qualitative detection of respiratory syncytial virus (RSV) antigen (viral fusion protein) directly from nasopharyngeal swab, nasopharyngeal aspirate, or nasal/nasopharyngeal wash specimens for symptomatic pediatric patients (eighteen years of age and younger). The test is intended for use as an aid in the diagnosis of acute respiratory syncytial viral infections. It is recommended that negative test results be confirmed by cell culture. Negative results do not preclude RSV infection and it is recommended that they not be used as the sole basis for treatment or other management decisions. The test is intended for professional and laboratory use. | The QuickVue RSV test is a dipstick immunoassay which allows for the rapid, qualitative detection of respiratory syncytial virus (RSV) antigen (viral fusion protein) directly from nasopharyngeal swab or nasopharyngeal aspirate specimens for symptomatic pediatric patients (eighteen years of age and younger). The test is intended for use as an aid in the diagnosis of acute respiratory syncytial viral infections. It is recommended that negative test results be confirmed by cell culture. Negative results do not preclude RSV infection and it is recommended that they not be used as the sole basis for treatment or other management decisions. The test is intended for professional and laboratory use.  |
|  Specimen Types | Nasopharyngeal swab, Nasopharyngeal aspirate, and Nasal/nasopharyngeal wash | Nasopharyngeal swab and Nasopharyngeal aspirate  |
|  Extract / Elute | Extraction reagent used for swab/aspirate/wash | Extraction reagent used for swab/aspirate  |
|  Read Result Time | 15 Minutes | 15 Minutes  |
|  Format | Lateral-flow immunoassay dipstick | Lateral-flow immunoassay dipstick  |
|  Control Features | Procedural Control Line Clearing of background | Procedural Control Line Clearing of background  |
|  External Controls | Positive RSV swab RSV negative swab coated with Streptococcus C antigen | Positive RSV swab RSV negative swab coated with Streptococcus C antigen  |

# K. Standard/Guidance Document Referenced (if applicable):

Not applicable

# L. Test Principle:

{3}

Lateral flow immunocapture assay

## M. Performance Characteristics (if/when applicable):

1. **Analytical performance:**
a. **Precision/Reproducibility:**
Not applicable
b. **Linearity/assay reportable range:**
Not applicable
c. **Traceability, Stability, Expected values (controls, calibrators, or methods):**
Not applicable
d. **Detection limit:**
Not applicable
e. **Analytical specificity:**
Not applicable
f. **Assay cut-off:**
Not applicable

2. **Comparison studies:**
a. **Method comparison with gold standard:**
See clinical studies.
b. **Matrix comparison:**
Not applicable

3. **Clinical studies:**
Guidel performed the following studies to establish the sensitivity and specificity of the QuickVue RSV Test with nasal/nasopharyngeal specimens:

During the 2006/2007 flu season, the performance of the QuickVue RSV test was

4

{4}

compared to viral cell culture methods and DFA in a multi-center clinical study during the RSV season in the United States. This study was performed by professional health care personnel at two pediatric clinics and two hospital emergency departments in various geographical regions within the United States. In this multi-center, point-of-care (POC) field trial, nasal/nasopharyngeal wash specimens were collected from two hundred eighty-nine (289) patients. All clinical samples were collected from symptomatic patients less than six years of age. 60% were male and 40% were female.

On-site testing of a portion of nasal/nasopharyngeal wash was performed by physician office personnel with the QuickVue RSV test. All samples were freshly collected and tested within one hour. No samples were frozen prior to testing. The remaining sample was placed in viral transport media and transported to a reference laboratory for culture, where cells were inoculated with the specimen, incubated at 36°C for 48 hours, and then removed from culture and tested for RSV by direct fluorescent antibody (DFA) staining.

## a. Clinical Sensitivity and specificity:

## Results with Fresh Nasal/Nasopharyngeal Wash Specimens

Nasal/nasopharyngeal wash specimens from two hundred eighty-nine (289) patients were tested in QuickVue RSV and in cell culture. The QuickVue RSV test correctly identified 83% (100/121) RSV culture-positive specimens and 90% (152/168) RSV culture-negative specimens. These results are shown in Table 3.

Table 3 QuickVue RSV Nasal/Nasopharyngeal Wash Results versus Culture (&lt;6 years of age)

|   | + | -  |
| --- | --- | --- |
|  QV Pos | 100 | 16  |
|  QV Neg | 21 | 152  |

$$
\text{Sensitivity} = 100/121 = 83\%
$$

$$
(95\% \text{ C.I. } 75-88\%)
$$

$$
152/168 = 90\%
$$

$$
\text{Specificity} = (95\% \text{ C.I. } 85-94\%)
$$

$$
\text{PPV} = 100/116 = 86\%
$$

$$
\text{NPV} = 152/173 = 88\%
$$

{5}

Clinical sensitivity: 83% (100/121)
(95% CI 75-88%)

Clinical specificity: 90%. (152/168)
(95% CI 85-94%)

c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable

4. Clinical cut-off:
Not applicable

5. Expected values/Reference range:
Not applicable

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

6

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/GQG/K070747](https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/GQG/K070747)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
