Gold Standard Diagnostics (GSD) AIX1000 Rapid Plasma Reagin (RPR) Automated Test System
K250249 · Gold Standard Diagnostics, LLC · GMQ · Oct 10, 2025 · Microbiology
Device Facts
Record ID
K250249
Device Name
Gold Standard Diagnostics (GSD) AIX1000 Rapid Plasma Reagin (RPR) Automated Test System
Applicant
Gold Standard Diagnostics, LLC
Product Code
GMQ · Microbiology
Decision Date
Oct 10, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3820
Device Class
Class 2
Indications for Use
The Gold Standard Diagnostics (GSD) AIX1000 Rapid Plasma Reagin (RPR) Automated Test System is a non-treponemal flocculation test that can qualitatively determine the presence of reagin antibodies in human serum, automated on the AIX1000 Analyzer. It may be used to aid in the diagnosis of syphilis when used in conjunction with supplemental treponemal laboratory tests and other clinical information. This test may also be used to detect non-treponemal antibodies in samples serially diluted to establish titer information. This test is not intended for screening blood or tissue donors.
Device Story
System performs automated non-treponemal flocculation testing for syphilis. Input: human serum samples. Process: AIX1000 Analyzer dispenses serum into wells, adds modified VDRL carbon antigen, incubates with shaking; onboard camera captures high-resolution images of flocculation; proprietary software algorithm analyzes images to determine reactive/non-reactive status and titers. Output: qualitative results and titer values. Used in clinical laboratories; operated by trained technicians. Results assist clinicians in syphilis diagnosis when combined with treponemal tests and clinical history. Benefits: automated processing reduces manual labor and variability in flocculation interpretation.
Clinical Evidence
Clinical performance evaluated via prospective (n=765) and retrospective (n=2,246) studies comparing subject device to FDA-cleared comparator. Prospective PPA 95.5%, NPA 99.9%. Retrospective PPA 97.2%, NPA 99.1%. Additional studies on pregnant women (n=280) and HIV-positive individuals (n=280) showed 100% agreement. Bench testing included cross-reactivity (17 conditions), interfering substances (endogenous/drugs), carry-over, and precision/reproducibility studies. All studies confirmed performance consistent with predicate.
Technological Characteristics
Non-treponemal flocculation test using modified VDRL carbon antigen (0.03% cardiolipin, 0.9% cholesterol, 0.21% lecithin). Automated microtiter plate processor (AIX1000 Analyzer) with onboard camera for image capture. Proprietary software algorithms for result interpretation. Requires daily/weekly/monthly calibration (priming, alignment, light intensity). Complies with IEC 61010-1, ISO 14971, and IEC 60601-1-2.
Indications for Use
Indicated for qualitative determination of reagin antibodies in human serum to aid in syphilis diagnosis and for titer determination via serial dilution. Applicable to patients suspected of syphilis infection. Not for blood or tissue donor screening.
Regulatory Classification
Identification
Treponema pallidum nontreponemal test reagents are devices that consist of antigens derived from nontreponemal sources (sources not directly associated with treponemal organisms) and control sera (standardized sera with which test results are compared) used in serological tests to identify reagin, an antibody-like agent, which is produced from the reaction of treponema microorganisms with body tissues. The identification aids in the diagnosis of syphilis caused by microorganisms belonging to the genus Treponema and provides epidemiological information on syphilis.
Predicate Devices
Gold Standard Diagnostics AIX1000 Rapid Plasma Reagin (RPR) Automated Test System (K150358)
Related Devices
K150358 — Gold Standard Diagnostics AIX1000 Rapid Plasma Reagin (RPR) Automated Test System · Gold Standard Diagnostics · Nov 12, 2015
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K250249
B Applicant
Gold Standard Diagnostics, LLC.
C Proprietary and Established Names
Gold Standard Diagnostics (GSD) AIX1000 Rapid Plasma Reagin (RPR) Automated Test System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| GMQ | Class II | 21 CFR 866.3820 - Treponema Pallidum Nontreponemal Test Reagents | MI - Microbiology |
## II Submission/Device Overview:
### A Purpose for Submission:
The purpose of this submission is to add a new version of AIX1000 Analyzer with high titer function to the Gold Standard Diagnostics AIX1000 Rapid Plasma Reagin (RPR) Automated Test System (which consists of the Gold Standard Diagnostics RPR reagents and the AIX1000 Analyzer) and use of RPR diluent to prepare dilutions of highly reactive samples. This submission also includes several hardware and software changes made to the AIX1000 analyzer.
### B Measurand:
Serum antibodies (cardiolipin and lecithin) against rapid plasma reagin
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
{1}
C Type of Test:
Non-treponemal macroscopic flocculation test
## Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The Gold Standard Diagnostics (GSD) AIX1000 Rapid Plasma Reagin (RPR) Automated Test System is a non-treponemal flocculation test that can qualitatively determine the presence of reagin antibodies in human serum, automated on the AIX1000 Analyzer. It may be used to aid in the diagnosis of syphilis when used in conjunction with supplemental treponemal laboratory tests and other clinical information. This test may also be used to detect non-treponemal antibodies in samples serially diluted to establish titer information. This test is not intended for screening blood or tissue donors.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
The GSD AIX1000 RPR Automated Test System uses the AIX1000 Analyzer. The AIX instrument automates sample preparation and results interpretation.
## Device/System Characteristics:
A Device Description:
The GSD AIX1000 RPR Automated Test System is a non-treponemal test designed for the qualitative determination of reagin antibodies in human serum to aid in the diagnosis of syphilis. The test is also used to detect non-treponemal antibodies in samples that are serially diluted to establish titer information. The system consists of the AIX1000 Analyzer and RPR test reagents, where the AIX1000 Analyzer delivers serum from collection tubes into test wells, adds the antigen suspension, and then incubates the test wells while shaking them. An onboard camera creates a high-resolution image that is analyzed by proprietary software algorithms to interpret the results.
The RPR test reagents consist of a reactive control, a non-reactive control, and the antigen suspended in a carbon solution. When the antigen is mixed with sera, antibodies, if present, will bind to the antigen and form black flocculants due to the presence of carbon particles, while in
K250249 - Page 2 of 10
{2}
the absence of antibodies, the carbon particles remain evenly distributed. The antigen used in the GSD AIX1000 RPR Automated Test System is a modified VDRL carbon antigen with the same formulation established by the Centers for Disease Control and Prevention (CDC), containing 0.03% cardiolipin, 0.9% cholesterol, and 0.21% lecithin. The kit also includes untreated sterile 48-well reaction plates, a reactive control, a non-reactive control, and RPR diluent.
## B Principle of Operation:
The GSD AIX1000 RPR Automated Test System is a flocculation test kit intended to be used with the AIX1000 Automated Analyzer. The instrument is intended to duplicate manual analytical procedures of a flocculation test by automating all necessary procedural steps. This is a non-treponemal macroscopic flocculation test that uses image capture and analysis to detect the presence of reagin. When reagin antibodies are present in a sample, they bind to their lipid antigens. Charcoal particles added to the solution co-agglutinate with these complexes and form black clumps that are macroscopically visible.
## C Instrument Description Information:
1. Instrument Name:
AIX1000 Analyzer
2. Specimen Identification:
Specimens are identified by scanning a barcode or by manual entry.
3. Specimen Sampling and Handling:
Sample processing is automated by the AIX1000 Analyzer. The AIX1000 Analyzer is a fully automated microtiter plate processor that is able to completely perform sample processing steps, including dilutions, dispenses, and incubations.
4. Calibration:
Daily, weekly and monthly calibration and maintenance is required by the end user. These actions include instrument priming, instrument alignment, camera alignment, wash pump calibration, light intensity and camera integration time.
5. Quality Control:
The kit contains an external control set, consisting of a non-reactive control (human serum) and a reactive control (human serum reactive for syphilis)
## V Substantial Equivalence Information:
## A Predicate Device Name(s):
K250249 - Page 3 of 10
{3}
Gold Standard Diagnostics AIX1000 Rapid Plasma Reagin (RPR) Automated Test System
B Predicate 510(k) Number(s):
K150358
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K250249 | K150358 |
| --- | --- | --- |
| Device Trade Name | Gold Standard Diagnostics AIX1000 Rapid Plasma Reagin (RPR) Automated Test System | Gold Standard Diagnostics AIX1000 Rapid Plasma Reagin (RPR) Automated Test System |
| General Device Characteristics | | |
| Intended Use/Indications For Use | The Gold Standard Diagnostics (GSD) AIX1000 Rapid Plasma Reagin (RPR) Automated Test System is a non-treponemal flocculation test that can qualitatively determine the presence of reagin antibodies in human serum, automated on the AIX1000 Analyzer. It may be used to aid in the diagnosis of syphilis when used in conjunction with supplemental treponemal laboratory tests and other clinical information. This test may also be used to detect non-treponemal antibodies in samples serially diluted to establish titer information. This test is not intended for screening blood or tissue donors. | The Gold Standard Diagnostics AIX1000 Rapid Plasma Reagin (RPR) Automated Test System is a non-treponemal flocculation test that can qualitatively determine the presence of reagin antibodies in human serum. It may be used to aid in the diagnosis of syphilis when used in conjunction with supplemental treponemal laboratory tests and other clinical information. This test may also be used to detect non-treponemal antibodies in samples serially diluted to establish titer information. This test is not intended for screening blood or tissue donors. |
K250249 - Page 4 of 10
{4}
| Assay Format | Reports qualitative results and titer of non-treponemal antibodies in serially diluted samples. | Same |
| --- | --- | --- |
| Technology | Flocculation test | Same |
| Antigen | Modified VDRL carbon antigen | Same |
| Reported Results | Reactive, non-reactive, titer results | Same |
| Interpretation | Automated | Same |
| Sample processing | Automated | Same |
| Titer Diluent | PBS for low and moderate titer.
GSD RPR Diluent for high and extra high titer | PBS |
VI Standards/Guidance Documents Referenced:
- IEC 61010-1 Ed. 3.1 b:2017 Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements.
- ISO 14971:2019+A11:2021 Medical devices. Application of risk management to medical devices.
- IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
Analytical studies supporting the GSD AIX1000 RPR Automated Test System in human serum were previously submitted and reviewed under K150358. In addition to the previously conducted studies, validating testing to incorporate dilution and detection of highly reactive samples in their titer function and use of RPR diluent was performed. A summary of study results is provided below.
1. Endpoint Titer Validation with Highly Reactive Samples
K250249 - Page 5 of 10
{5}
Four serum samples were tested using the AIX1000 Analyzer to evaluate endpoint titer validation and ensure that it is not affected by the GSD RPR diluent. The samples were tested in quadruplicate, twice a day, for 10 days with randomized three operators for a total N of 80 per sample. Additionally, two highly reactive serum samples were tested on the AIX1000 with the extra high titer function. The samples were tested in duplicate, twice per day by different operators for 20 days for a total N of 80 per sample. The results from the study are summarized o in Table 1.
Table 1. Results from Endpoint Titer Validation Study:
| | | | Endpoint Titer Results | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample Reactivity | Non-reactive | Reactive | 1:16 | 1:32 | 1:64 | 1:128 | 1:256 | 1:512 | 1:1024 | ≥1:2048 |
| Sample 1 | 0 | 80 | 0 | 61 | 19 | 0 | 0 | 0 | 0 | 0 |
| Sample 2 | 0 | 80 | 0 | 0 | 14 | 47 | 19 | 0 | 0 | 0 |
| Sample 3 | 0 | 80 | 1 | 76 | 3 | 0 | 0 | 0 | 0 | 0 |
| Sample 4 | 0 | 80 | 0 | 0 | 0 | 54 | 21 | 5 | 0 | 0 |
| Sample 5 | 0 | 80 | | | | | | 54 | 23 | 3 |
| Sample 6 | 0 | 80 | | | | | | 34 | 32 | 14 |
| Reactive control | 0 | 60 | | | | | | | | |
| Non-reactive control | 20 | 0 | | | | | | | | |
The data, as presented in Table 1, demonstrates that all replicates for each highly reactive sample were within ±1 dilution for the GSD AIX1000 RPR Automated Test System using GSD RPR diluent.
2. Linearity:
N/A
3. Analytical Specificity/Interference:
K250249 - Page 6 of 10
{6}
Please refer to the decision summary of K150358.
4. Assay Reportable Range:
N/A
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Please refer to the decision summary of K150358.
6. Detection Limit:
N/A
7. Assay Cut-Off:
Please refer to the decision summary of K150358.
8. Carry-Over:
Please refer to the decision summary of K150358.
## B Comparison Studies:
Comparison Study I:
Fifty samples were tested in parallel using both the previously FDA-cleared GSD AIX1000 RPR Automated Test System (with diluent according to the FDA-cleared Instructions for Use) and the modified version (with RPR diluent) to demonstrate equivalent performance between the two diluents. Samples were tested singly on all devices using retrospectively collected samples from laboratories and the titers were compared. The results are summarized in Table 2.
Table 2: Comparison Results
| Sample ID | Cleared Version | Modified version |
| --- | --- | --- |
| Sample 1 | 1:32 | 1:32 |
| Sample 2^{#} | >= 1:256 | >= 1:256 |
| Sample 3 | 1:16 | 1:32 |
| Sample 4 | 1:64 | 1:64 |
| Sample 5 | 1:32 | 1:32 |
| Sample 6 | 1:32 | 1:64 |
| Sample 7 | 1:32 | 1:32 |
| Sample 8 | 1:128 | 1:64 |
| Sample 9 | 1:64 | 1:64 |
| Sample 10 | 1:128 | 1:128 |
K250249 - Page 7 of 10
{7}
| Sample 11 | 1:16 | 1:32 |
| --- | --- | --- |
| Sample 12 | 1:32 | 1:32 |
| Sample 13 | 1:128 | 1:128 |
| Sample 14 | 1:32 | 1:64 |
| Sample 15 | 1:32 | 1:64 |
| Sample 16 | 1:128 | 1:128 |
| Sample 17 | 1:32 | 1:32 |
| Sample 18 | 1:64 | 1:64 |
| Sample 19 | 1:64 | 1:64 |
| Sample 20 | 1:32 | 1:64 |
| Sample 21a | >= 1:256 | >= 1:256 |
| Sample 22 | 1:32 | 1:32 |
| Sample 23 | 1:32 | 1:32 |
| Sample 24 | 1:128 | >= 1:256 |
| Sample 25 | 1:32 | 1:64 |
| Sample 26a | >= 1:256 | 1:128 |
| Sample 27a | 1:128 | >= 1:256 |
| Sample 28 | 1:128 | 1:64 |
| Sample 29 | 1:32 | 1:32 |
| Sample 30 | 1:64 | 1:64 |
| Sample 31 | 1:64 | 1:64 |
| Sample 32 | 1:64 | 1:64 |
| Sample 33 | 1:32 | 1:32 |
| Sample 34 | 1:16 | 1:32 |
| Sample 35a | >=1:256 | >= 1:256 |
| Sample 36 | 1:64 | 1:64 |
| Sample 37 | 1:32 | 1:32 |
| Sample 38 | 1:64 | 1:64 |
| Sample 39 | 1:32 | 1:64 |
| Sample 40 | 1:32 | 1:64 |
| Sample 41 | 1:64 | 1:128 |
| Sample 42 | 1:16 | 1:32 |
| Sample 43 | 1:16 | 1:32 |
| Sample 44 | 1:32 | 1:32 |
| Sample 45 | 1:32 | 1:32 |
| Sample 46 | 1:32 | 1:32 |
| Sample 47 | 1:32 | 1:64 |
| Sample 48 | 1:128 | 1:128 |
| Sample 49 | 1:16 | 1:32 |
| Sample 50 | 1:128 | 1:128 |
${}^{a}$ The samples with results $\geq 1 : {256}$ on the modified AIX1000 analyzer did not undergo higher titer testing,as the comparator lacked the extended high titer functionality that is only available on the new analyzer.
K250249 - Page 8 of 10
{8}
Titers obtained using the modified GSD AIX1000 RPR Automated Test System are within ± one two-fold dilution of the titers determined by the previously cleared version.
## Comparison Study II:
Five highly reactive serum samples were tested on the modified versions of the GSD AIX1000 RPR Automated Test System (using the RPR diluent) with the high titer and extra high titer function and on the FDA cleared RPR assay. The samples were tested in triplicate on both devices and the titers were compared. The results using highly reactive samples are summarized in Table 3.
Table 3: Comparison Using Highly Reactive Samples
| Sample ID | Comparator Result | AIX1000 Result |
| --- | --- | --- |
| Sample 1 | 1:256 | 1:512 |
| | 1:512 | 1:512 |
| | 1:512 | 1:1024 |
| Sample 2 | 1:256 | 1:512 |
| | 1:256 | 1:512 |
| | 1:512 | 1:512 |
| Sample 3 | 1:512 | 1:512 |
| | 1:512 | 1:512 |
| | 1:512 | 1:512 |
| Sample 4 | 1:512 | 1:512 |
| | 1:1024 | 1:512 |
| | 1:512 | 1:512 |
| Sample 5 | 1:256 | 1:512 |
| | 1:512 | 1:512 |
| | 1:512 | 1:1024 |
All results from the GSD AIX1000 RPR Automated Test System are within ±1 two-fold dilution of the results determined using the comparator assay.
## C Clinical Studies:
Please refer to the decision summary of K150358.
## D Clinical Cut-Off:
Please refer to the decision summary of K150358.
## E Expected Values/Reference Range:
Please refer to the decision summary of K150358.
## F Other Supportive Instrument Performance Characteristics Data:
K250249 - Page 9 of 10
{9}
N/A
## VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
## IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K250249 - Page 10 of 10
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.