K232565 · Copan Italia Spa · JSM · Oct 11, 2024 · Microbiology
Device Facts
Record ID
K232565
Device Name
UriSponge
Applicant
Copan Italia Spa
Product Code
JSM · Microbiology
Decision Date
Oct 11, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.2390
Device Class
Class 1
Indications for Use
Copan UriSponge™ - Urine Collection, Transport, and Preservation System is intended for the collection, transport and preservation of urine specimens from the collection site to the testing laboratory. In the laboratory, UriSponge™ specimens are processed using standard clinical laboratory operating procedures for the cultivation of uropathogenic bacteria and yeasts.
Device Story
UriSponge is a urine collection, transport, and preservation system consisting of a self-standing plastic tube containing a plastic stick with a hydrophilic polyurethane sponge. The sponge is impregnated with chemical preservatives (Sodium Propionate and Potassium Sorbate). The device is used by healthcare professionals to collect urine specimens at the point of care; the specimen is absorbed into the sponge for transport to a laboratory. In the laboratory, the device is centrifuged to release the urine for standard microbiological cultivation of uropathogenic bacteria and yeasts. The system maintains specimen stability for up to 48 hours at 2-25°C. The device benefits patients by ensuring the viability of microorganisms during transport, facilitating accurate diagnostic testing.
Clinical Evidence
No clinical data. Bench testing only. Microbial recovery studies performed using 8 representative uropathogenic strains (e.g., E. coli, S. agalactiae, C. albicans) in pooled human negative clinical urine. Samples were incubated for 0, 24, and 48 hours at 2-8°C and 19-25°C. Acceptance criteria defined as ΔLog10 ≤ 1 and ≥ -1 between T=0 and T=48 hours. All tested organisms met stability criteria. Fill volume flex study confirmed no significant toxicity risk due to sponge undersaturation.
Technological Characteristics
Materials: Hydrophilic polyurethane sponges, plastic tube/cap. Preservatives: Sodium Propionate, Potassium Sorbate. Dimensions: Regular (100mm x 16mm) and Mini (80mm x 12mm). Sterilization: Irradiation per ISO 11137-1:2006 (SAL 10⁻⁶). Principle: Chemical stabilization of microbial load in absorbed urine. Connectivity: None.
Indications for Use
Indicated for the collection, transport, and preservation of urine specimens from the collection site to the testing laboratory for the cultivation of uropathogenic bacteria and yeasts. For prescription use only.
Regulatory Classification
Identification
A transport culture medium is a device that consists of a semisolid, usually non-nutrient, medium that maintains the viability of suspected pathogens contained in patient specimens while in transit from the specimen collection area to the laboratory. The device aids in the diagnosis of disease caused by pathogenic microorganisms and also provides epidemiological information on these diseases.
Predicate Devices
UriSwab™ - Urine Collection, Transport and Preservation System (K180052)
Related Devices
K180052 — UriSwab-Urine Collection, Transport and Preservation System · Copan Italia Spa · Mar 23, 2018
K220739 — Disposable Urine Collection Tube · Zhejiang Gongdong Medical Technology Co., Ltd. · Nov 17, 2022
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K232565
B Applicant
Copan Italia S.p.A.
C Proprietary and Established Names
UriSponge
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JSM | Class I, reserved | 21 CFR 866.2390 - Transport Culture Medium | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To obtain substantial equivalence determination for the Copan UriSponge for urine collection, transport and preservation of urine specimens and processing by using standard clinical laboratory operating procedures for the cultivation of uropathogenic bacteria and yeasts.
B Measurand:
Not applicable
C Type of Test:
Non-propagating urine transport device containing stabilizing reagents.
## III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
K232565 - Page 2 of 7
B Indication(s) for Use:
Copan UriSponge - Urine Collection, Transport, and Preservation System, intended for the collection, transport and preservation of urine specimens from the collection site to the testing laboratory. In the laboratory, UriSponge specimens are processed using standard clinical laboratory operating procedures for the cultivation of uropathogenic bacteria and yeasts.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
None
IV Device/System Characteristics:
A Device Description:
Copan’s UriSponge - Urine Collection, Transport, and Preservation System consists of screw cap self-standing plastic tube with conical shaped bottom. Inside the tube, the cap holds a plastic stick with sponges made of hydrophilic polyurethane. The sponges include preservative substances (Sodium Propionate, and Potassium Sorbate). Two sizes of product are available: the regular tube size (100 mm length x 16 mm diameter) plastic tube, and the mini tube size (80 mm length x 12 mm diameter) plastic tube.
B Principle of Operation:
The UriSponge applicator stick has two (mini size) or three (regular size) cylindrical shaped sponges which are contained within appropriately sized sterile plastic tube with screw cap. The sponges contain the chemical preservatives (Sodium Propionate, and Potassium Sorbate). These chemical preservatives maintain the microbial load of bacteria and yeast in absorbed urine sample during the transport until they are received and tested in the laboratory by standard culture techniques.
V Substantial Equivalence Information:
A Predicate Device Name(s):
UriSwab-Urine Collection, Transport and Preservation System
B Predicate 510(k) Number(s):
K180052
C Comparison with Predicate(s):
| Device & Predicate Device(s): | Device: K232565 | Predicate: K180052 |
| --- | --- | --- |
| Device Trade Name | UriSponge | UriSwab |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Copan UriSponge - Urine Collection, | Copan UriSwab - Urine Collection, Transport |
{2}
| | Transport, and Preservation System is intended for the collection, transport, and preservation of urine specimens from the collection site to the testing laboratory. In the laboratory, UriSponge specimens are processed using standard clinical laboratory operating procedures for the cultivation of uropathogenic bacteria and yeasts. | and Preservation System is intended for the collection, transport and preservation of urine specimens from the collection site to the testing laboratory. In the laboratory, UriSwab specimens are processed using standard clinical laboratory operating procedures for the cultivation of uropathogenic bacteria and yeasts. |
| --- | --- | --- |
| Single Use device | Yes | Same |
| Device storage temperature (prior to use) | 2-25°C | Same |
| Specimen storage temperature after collection | 2-25°C | Same |
| Urine Specimen Stability at Room Temperature | Up to 48 hours | Same |
| Sterile | Yes | Same |
| General Device Characteristic Differences | | |
| Preservative ingredients | Sodium Propionate, and Potassium Sorbate | Boric Acid, and Sodium Formate |
VI Standards/Guidance Documents Referenced:
1. ISO 11137-1:2006 (including Amendment 1:2013 and Amendment 2:2018), Sterilization of Health Care Products- Radiation-Part1: Requirements for development, validation, and routine control of a sterilization process for medical devices.
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Not applicable
2. Linearity:
Not applicable
K232565 - Page 3 of 7
{3}
K232565 - Page 4 of 7
3. Analytical Specificity/Interference:
Not applicable
4. Assay Reportable Range:
Not applicable
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Shelf Life: A real-time shelf-life stability testing was conducted. Testing was conducted with 3 different lots of the mini version of the devices (80 mm length x 12 mm diameter) at each timepoint (at within 1 month after manufacture, at approximately 5-6 months after manufacture, at approximately 12 months after manufacture and at 13 months after manufacture). Physical, mechanical stability as well as sponge absorption and release volumes were evaluated using 30 devices from each lot. To determine the physical and the mechanical stability, each lot was inspected for device's appearance and integrity through the intended use workflow. Sponge absorption and release volume were also tested to determine physical stability. Preservative content of UriSponge was tested at each timepoint by high performance liquid chromatography (HPLC) method. All results met the study acceptance criteria to support a shelf-life stability of 12 months when stored at 2-25°C in appropriate storage environment.
Sterilization: UriSponge tubes are sterilized using irradiation, with the dosage set following ISO 11137-1:2006 (including Amendment 1:2013 and Amendment 2:2019), Sterilization of Health Care Products- Radiation-Part1: Requirements for development, validation, and routine control of a sterilization process for medical devices. The acceptable sterility assurance level (SAL) for the sterilized UriSponge product was determined to be 10⁻⁶ or greater.
6. Detection Limit:
Performance Testing - Recovery Studies:
A culture-based recovery study was performed to support that ability of the UriSponge device to keep microbial load stable in the collected urine up to 48 hours at both 2-8 °C and 19-25 °C. Recovery of Escherichia coli (ATCC 25922), Streptococcus agalactiae (ATCC 13813), Enterococcus Faecalis (ATCC 29212), Pseudomonas aeruginosa (ATCC 27853), Proteus mirabilis (ATCC 7002), Staphylococcus saprophyticus (ATCC 15305), Enterobacter Clocae (ATCC 13047), Klebsiella pneumoniae (ATCC 13883) and Candida albicans (ATCC 24433) were tested as representative strains for urine specimen. Pooled human negative clinical urine samples were used as the best representative of the intended use sample type for transport and downstream analysis. In brief, the samples were prepared by suspending each of above-indicated freshly cultured organisms into human negative clinical urine. A minimum of three different dilutions of stock culture were used to obtain a colony count of 25-250 per plate and the appropriate dilution factor was considered in determining the viable count in CFU/mL. Sponge applicator of UriSponge devices of three lots (new, middle aged, and recently expired) were immersed into contrived urine sample for 5 seconds and returned to UriSponge tube and incubated for 0 hrs. (less than 20-minutes), 24 hrs., and 48 hrs. at 2-8 °C and 19-25°C. At the end of each incubation period, the devices were centrifuged, and the
{4}
absorbed sample were released, and aliquots of sample were spread onto appropriate agar plates and incubated in appropriate environment for 24-48 hours. Following incubation, colony forming units (CFU) were counted. A colony count of 25-250 per plate for at least one dilution and the $\Delta \mathrm{Log}_{10} \leq 1$ and $\geq -1$ between the average CFU/plate values at time zero $(T = 0$ hrs.) and at specific specimen incubation time (e.g., 24 hrs., 48 hrs., etc.) were considered acceptable to support a specimen stability claim for each target organism.
Table 2 below summarize the microbial recovery performance of the UriSponge device when the specimen containing devices were stored at $2 - 8^{\circ}\mathrm{C}$ and $19 - 25^{\circ}\mathrm{C}$ .
Table 2: Microbial recovery results summary of Urisponge device.
| Organism | Organism concentration at T = 0 hrs. (CFU/mL) | Incubation Temperature | Logarithmic difference in microbial recovery from the baseline (T= 0 hrs.) (-ve indicates reduction) | |
| --- | --- | --- | --- | --- |
| | | | T = 24 hrs. | T = 48 hrs. |
| C. albicans (ATCC 24433) | 5x102 | 2-8°C | -0.02 | -0.03 |
| | | 19-25°C | 0.18 | 0.38 |
| E. coli (ATCC 25922) | 1.5x103 | 2-8°C | -0.11 | -0.20 |
| | | 19-25°C | -0.11 | -0.14 |
| E. faecalis (ATCC 29212) | 7.5x102 | 2-8°C | -0.19 | -0.25 |
| | | 19-25°C | -0.04 | -0.07 |
| P. aeruginosa (ATCC 27853) | 1.5x103 | 2-8°C | -0.10 | -0.16 |
| | | 19-25°C | -0.20 | -0.30 |
| P. mirabilis (ATCC 7002) | 7.5x102 | 2-8°C | -0.08 | -0.05 |
| | | 19-25°C | -0.10 | -0.06 |
| S. saprophyticus (ATCC 15305) | 1.5x103 | 2-8°C | -0.17 | -0.21 |
| | | 19-25°C | -0.16 | -0.07 |
| E. cloacae (ATCC 13047) | 1.5x103 | 2-8°C | -0.23 | -0.26 |
| | | 19-25°C | -0.27 | -0.29 |
| K. pneumoniae (ATCC 13883) | 1.5x103 | 2-8°C | -0.11 | -0.06 |
| | | 19-25°C | -0.21 | -0.13 |
| S. agalactiae (ATCC 13813) | 1.5x103 | 2-8°C | -0.35 | -0.45 |
| | | 19-25°C | -0.39 | -0.53 |
Microbial recovery study results support the ability of UriSponge device to maintain the recovery of the tested microorganisms in urine samples up to 48 hours when stored at $2 - 8^{\circ}\mathrm{C}$ or at $19 - 25^{\circ}\mathrm{C}$ .
# Fill Volume Flex Study: Impact of undersaturation of UriSponge Device:
The applicator sponge in the Urisponge device contains preservative to stabilize the urine sample. In situations when the applicator sponge is not fully saturated with the absorbed urine due to sponge undersaturation, the concentrated preservative may result in the creation
K232565 - Page 5 of 7
{5}
of a toxic environment to the intended use organisms present in the urine sample. To determine the impact of sponge undersaturation on microbial recovery, a fill volume flex study was conducted with 3 newly manufactured lots and with three strains (E. coli ATCC 25922, P. aeruginosa ATCC 27853 and S. agalactiae ATCC 13813) that exhibited the highest log reduction at 72 hours incubation. First, a preliminary fill volume study was conducted by immersing the applicator sponge in urine sample at different depths and for different durations of time. Based on the minimum acceptable release volume $(\geq 200\ \mu l)$, the worst-case scenario for undersaturation was determined. Finally, a comparative microbial recovery evaluation (as described above) was conducted between the intended use workflow and the worst-case scenario. For both scenarios, a colony count of 25-250 per plate for at least one dilution and the $\Delta \mathrm{Log}_{10} \leq 1$ and $\geq -1$ between the average CFU/plate values between time zero $(T = 0\ \mathrm{hrs.})$ and at the end of final incubation time, were considered acceptable to support specimen stability. The study results for both the intended use workflow and the worst-case scenario met the study acceptance criteria. The results from the fill volume flex study thus indicate that there is no significant risk of toxicity to the intended use organism in the urine sample due to undersaturation of UriSponge device.
7. Assay Cut-Off:
Not applicable
## B Comparison Studies:
1. Method Comparison with Predicate Device:
Not applicable
2. Matrix Comparison:
Not applicable
## C Clinical Studies:
1. Clinical Sensitivity:
Not applicable
2. Clinical Specificity:
Not applicable
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable
K232565 - Page 6 of 7
{6}
D Clinical Cut-Off:
Not applicable
E Expected Values/Reference Range:
Not applicable
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K232565 - Page 7 of 7
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.