DEN150059 · Clever Culture Systems AG · PPU · Oct 6, 2016 · Microbiology
Device Facts
Record ID
DEN150059
Device Name
APAS Compact with Urine Analysis Module
Applicant
Clever Culture Systems AG
Product Code
PPU · Microbiology
Decision Date
Oct 6, 2016
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 866.2190
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
DEN150059 · Oct 6, 2016
APAS Compact with Urine Analysis Module
Clever Culture Systems AG
Remnant clinical urine samples from standard-of-care culture
The clinical study evaluated the performance of the APAS Compact system by comparing its automated image assessment results against the manual interpretation of the same clinical urine culture plates by a panel of microbiologists.
Remnant clinical samples; Standard of care; Clinical performance evaluation
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective evaluation of remnant clinical samples
Patients with suspected urinary tract infection (uncomplicated cases); Sample Size: 9,224 included samples; Number of Sites: 3 (1 US, 2 ex-US)
Manual inspection of plates by a panel of three microbiologists
Clinical sensitivity and specificity for detection of microbial growth
Indications for Use
The APAS Compact is an in vitro diagnostic system comprised of an instrument for automated imaging of agar culture plates and a software analysis module for the following use: The APAS Compact, when using its urine analysis module, automates urine culture plate imaging and interpretation to detect the presence or absence of microbial growth on sheep blood and MacConkey agar culture plates that are inoculated with a 1μL sample volume. The APAS Compact, when using its urine analysis module, provides a semi-quantitative assessment of colony counts that are used as an aid in the diagnosis of urinary tract infection. All urine culture plates that are identified as positive for growth by the APAS Compact, when using its urine analysis module, must be reviewed by a trained microbiologist.
Device Story
APAS Compact is an in vitro diagnostic instrument for automated imaging and interpretation of agar culture plates; utilizes software analysis module to detect microbial growth on sheep blood and MacConkey agar plates inoculated with 1μL urine samples; provides semi-quantitative colony counts; intended for use in clinical microbiology laboratories; operated by trained personnel; output requires mandatory review by a trained microbiologist for all plates identified as positive; serves as an aid in diagnosis of urinary tract infections; automates plate screening to improve laboratory workflow efficiency.
Clinical Evidence
Clinical study conducted at three sites (1 US, 2 ex-US) with 9,224 remnant urine samples from uncomplicated UTI cases. Performance compared to a panel of three blinded microbiologists. Primary endpoint: sensitivity for detection of microbial growth. Results: 99.0% sensitivity (95% CI: 98.7-99.2%) on blood agar and 99.5% (95% CI: 99.2-99.7%) on MacConkey agar. False negative rates ranged from 0% to 2.9% (blood) and 0% to 1.3% (MacConkey).
Technological Characteristics
System includes an imaging station (CCD camera, LED illumination, light-sealed chassis) and a PC-based controller. Compatible with Remel Tryptic Soy Sheep Blood Agar and MacConkey Agar with Crystal Violet. Connectivity includes LIMS interface. Software performs image processing, colony enumeration, and morphology classification. Daily calibration required using a color calibration tool. Complies with IEC 61010-1, IEC 61326-2-6, and IEC 62304.
Indications for Use
Indicated for use as an aid in the diagnosis of urinary tract infection by automating the imaging and interpretation of urine culture plates (sheep blood and MacConkey agar) inoculated with 1μL samples to detect microbial growth and provide semi-quantitative colony counts.
Regulatory Classification
Identification
An automated image assessment system for microbial colonies on solid culture media is a system that is intended to assess the presence or absence of microbial colonies on solid microbiological culture medium, and to interpret their number, and phenotypic and morphologic characteristics through analysis of two dimensional digital images as an aid in diagnosis of infectious disease.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include a detailed description of the device, including the technology employed, components and software modules, as well as a detailed explanation of the result algorithms and any expert rules that are used to assess colony characteristics and enumerate colonies from image capture through end result.
(2) Premarket notification submissions must include detailed documentation of the analytical studies performed to characterize device performance to support the intended use, as appropriate.
(3) Premarket notification submissions must include detailed documentation from clinical studies performed on a population that is consistent with the intended use population.
(i) The clinical studies must establish the device performance based on comparison to results obtained by an acceptable reference method, as appropriate.
(ii) The clinical study documentation must include the study protocol with a predefined statistical analysis plan and the final report documenting support for the Indications for Use and the results of the statistical analysis, as appropriate.
(4) Premarket notification submissions must include detailed documentation for device software, including but not limited to software applications and hardware based components that incorporate software, and any decision-making thresholds used to generate results for the device. If a part of a Total Laboratory Automation System, the premarket notification submission must include detailed documentation addressing the instrument and software system integration.
(5) Premarket notification submissions must include detailed documentation of appropriate instructions for use regarding the intended user's device quality control procedures for the instrument system and components, as appropriate.
(6) The 21 CFR 809.10 compliant device labeling must include:
(i) Detailed user instructions to mitigate the risk of failure to operate the instrument correctly.
(ii) A detailed explanation of the interpretation of results and limitations regarding the need for review of culture plates by a qualified microbiologist, as appropriate.
(iii) A summary of performance data obtained from the analytical studies used to support device performance, as appropriate.
(iv) A summary of performance data obtained from clinical studies performed on a population that is consistent with the intended use population, as appropriate.
(7) Under 21 CFR 820.10(c) design and development, device manufacturers must, as appropriate:
(i) Conduct human factors/usability validation testing with the final version of the labeling and related materials to adequately mitigate the risk of failure to operate the instrument correctly.
(ii) Document a device training program that will be offered to the end user to adequately mitigate the risk of failure to operate the instrument correctly.
Submission Summary (Full Text)
{0}------------------------------------------------
# EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR THE AGAR PLATE ASSESSMENT SYSTEM (APAS) COMPACT
### DECISION MEMORANDUM
#### A. 510(k) Number:
DEN150059
### B. Purpose for Submission:
De novo request for evaluation of automatic class III designation for the Agar Plate Assessment System (APAS) Compact
### C. Measurand:
Digital images of microbial colonies cultured on blood and MacConkey agar plates
# D. Type of Test:
The APAS Compact when using its Urine Analysis Module (the "APAS Compact" or the "APAS Compact with Urine Analysis Module") is an in vitro diagnostic test system for automated assessment and enumeration of microbial colonies on solid culture media. The system is for use on urine cultures from suspected cases of urinary tract infection (UT).
# E. Applicant:
Clever Culture Systems AG
#### F. Proprietary and Established Names:
APAS Compact
#### G. Regulatory Information:
- 1. Regulation section:
21 CFR 866.2190 Automated image assessment system for microbial colonies on solid culture media
- 2. Classification:
Class II
- 3. Product code:
PPU
{1}------------------------------------------------
# 4. Panel:
- 83 Microbiology
# H. Indications for Use:
# 1. Indication(s) for Use:
The APAS Compact is an in vitro diagnostic system comprised of an instrument for automated imaging of agar culture plates and a software analysis module for the following use:
The APAS Compact, when using its urine analysis module, automates urine culture plate imaging and interpretation to detect the presence of microbial growth on sheep blood and MacConkey agar culture plates that are inoculated with a 1uL sample volume. The APAS Compact, when using its urine analysis module, provides a semiquantitative assessment of colony counts that are used as an aid in the diagnosis of urinary tract infection. All urine culture plates that are identified as positive for growth by the APAS Compact, when using its urine analysis module, must be reviewed by a trained microbiologist.
# 2. Special conditions for use statement(s):
For prescription use only.
The performance of APAS Compact with Urine Analysis Module has not been evaluated with urine samples from pregnant women. The APAS Compact with Urine Analysis Module will detect colonies of GBS if they are present but should not be used for the primary screening for Group B Streptococcus (GBS) carriers. Follow recommended guidelines for identification of pregnant women who are colonized with GBS.
The performance of the APAS Compact with Urine Analysis Module has not been evaluated with urine samples from suspected cases of complicated urinary tract infection e.g., those with underlying urinary tract pathology, suspected cases of persistent urinary tract infection, urine collected by invasive procedures or urine samples from immunocompromised subjects.
Slow growing organisms such as Corynebacterium spp. and Gardnerella vaginalis may not exhibit detectable growth within 18-22 hours. If infection with a slow growing species is suspected, extension of the incubation time may be required, followed by manual interpretation of the culture result. APAS Compact with Urine Analysis Module should not be used to interpret cultures incubated for greater than 22 hours.
The APAS Compact with Urine Analysis Module is qualified for use with the following culture plates manufactured by Remel: Tryptic Soy Sheep Blood Agar and MacConkey Agar with Crystal Violet.
<sup>1</sup> CDC Guidelines, 2010: Prevention of perinatal Group B Streptococcal disease.
{2}------------------------------------------------
#### 3. Special instrument requirements:
APAS Compact
# I. Device Description:
The APAS Compact with Urine Analysis Module is an instrumented system that is designed for screening of urine culture plates for the presence of microbial growth. The device comprises an imaging station for capture of digital images of the culture plates, together with software for analysis of the images, the determination and enumeration of microbial growth and reporting of results. A list of the major sub-components of the APAS Compact is depicted in Table 1. The Urine Analysis Module software is compatible with the following types of culture media:
Tryptic Soy Sheep Blood Agar MacConkey Agar with Crystal Violet
| Component | Function |
|--------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Imaging Station | Light Emitting Diode (LED) illumination of culture<br>plates and image capture using a Charge Coupled<br>Device (CCD) camera |
| APAS Controller Personal Computer (PC)<br>Urine Analysis Module Software | Image capture, storage and analysis<br>Installed on the APAS Controller PC to provide the<br>configuration and instructions for image capture and<br>analysis |
| Instrument Controller Software<br>LIMS Interface Software | User interface for operation of the APAS Compact<br>Installed on the Instrument Controller PC and used to<br>import other diagnostic information, such as<br>microscopy or chemistry results, and provide context<br>for interpretation of urine culture results.<br><br>Imported information may be applied to the system:<br>"LIMS Force Flag": automatically forces an APAS<br>result to "Review" irrespective of the growth<br>characteristics observed<br>"LIMS Complementary Test Flag": automatically<br>changes a "Negative” designation to “Review" based<br>on user defined rules applied to additional diagnostic<br>information |
| Color Calibration Tool | Multicolored disk for calibration of the system optics |
Table 1. Sub-components of the APAS Compact
The imaging station comprises a fixed CCD camera, top and bottom LED illumination for the culture plates and a plate handling mechanism, all of which are housed in a light-sealed chassis. The Instrument Controller PC provides the user interface and sends instructions to the APAS Controller PC which in turn controls the imaging station.
The plate handling mechanism of the imaging station comprises a manual lever that allows
{3}------------------------------------------------
the operator to load a single culture plate into the system and position it for imaging. The operator then initiates image capture using the Instrument Controller PC. Once an image is taken by the camera it is sent to the APAS Controller PC for assessment and storage. APAS Compact reports are viewed on a separate computer through a dedicated user interface.
# J. Standard/Guidance Document Referenced (if applicable):
IEC 61010-1: Safety requirements for electrical equipment for measurement, control and laboratory use - Part 1: General requirements, 3rd Edition.
IEC 61326-2-6: Electrical equipment for measurement, control and laboratory use - EMC Requirements - Part 2-6: Particular Requirements - In vitro diagnostic (IVD) medical Equipment, 2nd Edition.
IEC 62304 Medical device software - software lifecycle processes, 1st Edition.
ISO 14971 Medical devices - Application of Risk Management to Medical Devices, 2007.
CISPR 11 Industrial, scientific and medical equipment - radio frequency disturbance characteristics - Limits and methods of measurement, Version 5.1.
# K. Test Principle:
The APAS Compact with Urine Analysis Module is designed for the assessment of microbial colonies on urine culture plates. The device comprises an imaging station for capture of images of the culture plates, together with software for analysis of the images, the assessment and enumeration of microbial colonies and result designation. Plates with growth are designated as either "Positive" or "Review" depending on the number of colonies present and their morphologic characteristics. All samples that exhibit growth therefore require follow up according to standard laboratory practice by a trained microbiologist. Plates that are designated by the APAS Compact to have no growth may be discarded without further review.
The APAS Compact with Urine Analysis Module is indicated for screening paired sheep blood and MacConkey agar plates that are each inoculated with 1ul of urine and which are incubated at 35°C ± 2°C for 18 to 22 hours. For each urine sample, both culture plates must be processed on the APAS Compact. The system takes digital images of each plate which are analyzed automatically to determine the number of colonies present and their associated morphologies. For each plate, the APAS Compact performs the following tasks:
- a) Differentiates between areas of growth (colonies) and no growth (e.g., agar, labels, handwriting):
- b) Differentiates areas that are required for interpretation of colony morphology (e.g., α- and β-hemolysis);
- c) Assigns to each area of growth a code that corresponds to a morphology type;
- d) Estimates the number of colonies present, converts this number to an organism concentration in the original urine sample and reports it in terms of colony forming units per milliliter (CFU/mL).
{4}------------------------------------------------
The interpretation of the growth on a pair of plates follows a decision tree whereby the two plate reports are combined and a designation is made for the sample.
The colony morphologies reported by the APAS Compact with Urine Analysis Module are shown in Table 2 for sheep blood agar and Table 3 for MacConkey agar. A summary of the result logic used for automated interpretation of the colony counts and morphology is shown in Table 4. A "significant organism" at 103 CFU/mL on the plate will trigger designation of that plate as "Positive," as opposed to "Review." Plates that are designated as "Positive" or "Review" undergo further evaluation by a clinical microbiologist.
Table 2. Colony morphologies on sheep blood agar identified by APAS Compact with Urine Analysis Module
| Morphology Name | Description |
|-------------------|--------------------------------------------------------|
| Coliform | Gram-negative and coliform-like colonies |
| Cream-white | Staphylococcus and related species |
| Granular | Granular morphologies (e.g., Pseudomonas spp.) |
| Small | Enterococcus spp. and related species |
| Small α-hemolytic | Small colonies with α-hemolysis or very small colonies |
| Small ß-hemolytic | Small colonies with ß-hemolysis |
| Swarming organism | Proteus spp. and related, high-motility species |
Table 3. Colony morphologies on MacConkey agar identified by APAS Compact with Urine Analysis Module
| Morphology Name | Description |
|----------------------------------|-----------------------------------|
| Lactose fermenter | Pink/red colonies |
| Non-fermenter | Colonies without red/pink pigment |
| Non-fermenter with green pigment | Pseudomonas spp. |
| Red pigmented colonies | Pigmented Serratia marcescens |
Table 4. Result interpretation for APAS Compact with Urine Analysis Module
| Colony Count (CFU/mL) | Morphology | APAS Compact Designation |
|-----------------------|-------------------------|--------------------------|
| Not Applicable | Swarming organism | |
| $>=10^4$ | Any | Positive |
| $10^3$ | Significant organism1 | |
| $10^3$ | No significant organism | Review |
| 0 | Not Applicable | Negative |
1 Defined as growth of coliform or small colonies with B-hemolysis on blood agar or lactose fermenting colonies on MacConkey agar
{5}------------------------------------------------
### L. Performance Characteristics (if/when applicable):
#### 1. Analytical performance:
#### a. Precision/Reproducibility:
The reproducibility of colony counts performed by the APAS Compact with Urine Analysis Module was evaluated on different instruments over multiple runs. Testing was performed using logio-fold dilutions of 0.5 McFarland standard suspensions of representative uropathogenic species that produced from 0 to >100 colonies per plate. Each dilution was plated in triplicate and incubated for 18 hours at 35°C. Images of each plate were taken at 5 different orientations (0, 60, 120, 180 and 270°) on each of 3 APAS Compact instruments. The mean and standard deviation of the resulting colony counts was determined for each dilution on each instrument and overall (Tables 5 and 6). The results demonstrated acceptable reproducibility of colony counts between instruments and rotations of the same plate.
{6}------------------------------------------------
| Species | Dilution1 | Replicate | Instrument 1 | | Instrument 2 | | Instrument 3 | | Overall | | | |
|-----------------------|-----------|-----------|--------------|-------|--------------|-------|--------------|-------|---------|-------|-------|------|
| | | | Mean | %CV | Mean | %CV | Mean | %CV | Mean | %CV | Mean | %CV |
| E. coli | 3 | 1 | 76.2 | 1.9 | 83.2 | 4.4 | 74.0 | 2.1 | 77.8 | 6.0 | 78.3 | 7.0 |
| | 3 | 2 | 85.8 | 3.6 | 82.8 | 4.2 | 82.6 | 1.8 | 83.7 | 3.6 | | |
| | 3 | 3 | 72.8 | 2.5 | 73.6 | 2.6 | 73.4 | 3.7 | 73.3 | 2.8 | | |
| | 4 | 1 | 8.6 | 10.3 | 8.2 | 5.5 | 8.2 | 5.5 | 8.3 | 7.5 | 18.6 | 39.2 |
| E. coli | 4 | 2 | 23.2 | 1.9 | 23.8 | 1.9 | 23.0 | 0.0 | 23.3 | 2.1 | | |
| | 4 | 3 | 23.8 | 1.9 | 23.8 | 6.2 | 24.4 | 2.3 | 24.0 | 3.9 | | |
| | 5 | 1 | 5.6 | 9.8 | 5.6 | 9.8 | 5.6 | 9.8 | 5.6 | 9.1 | 3.2 | 75.0 |
| | 5 | 2 | 4.0 | 0.0 | 4.0 | 0.0 | 4.0 | 0.0 | 4.0 | 0.0 | | |
| | 5 | 3 | 0.0 | NA | 0.0 | NA | 0.0 | NA | 0.0 | NA | | |
| | 3 | 1 | 207.8 | 5.54 | 242.4 | 6.16 | 165.0 | 5.7 | 205.1 | 16.9 | 202.2 | 38.0 |
| | 3 | 2 | 141.8 | 6.09 | 135.6 | 10.18 | 108.6 | 4.3 | 128.7 | 13.6 | | |
| | 3 | 3 | 345.6 | 10.60 | 297.6 | 4.23 | 175.6 | 4.0 | 272.9 | 28.2 | | |
| | 4 | 1 | 14.4 | 6.18 | 16.4 | 13.35 | 15.2 | 7.2 | 15.3 | 10.7 | 19.6 | 29.6 |
| E. coli/S. agalactiae | 4 | 2 | 18.8 | 7.87 | 19.2 | 12.97 | 14.2 | 17.5 | 17.4 | 17.9 | | |
| | 4 | 3 | 28.8 | 5.14 | 29.4 | 7.45 | 20.2 | 13.7 | 26.1 | 18.4 | | |
| | 5 | 1 | 2.8 | 46.4 | 3.4 | 53.5 | 6.0 | 26.3 | 4.1 | 50.0 | 7.8 | 79.5 |
| | 5 | 2 | 15.6 | 17.3 | 17.6 | 9.5 | 14.8 | 13.9 | 16.0 | 14.8 | | |
| | 5 | 3 | 3.4 | 33.5 | 3.0 | 23.7 | 3.6 | 50.6 | 3.3 | 37.3 | | |
| | 3 | 1 | 326.0 | 9.0 | 317.6 | 3.6 | 292.8 | 1.8 | 312.1 | 7.2 | 271.1 | 18.6 |
| | 3 | 2 | 208.0 | 3.0 | 201.6 | 3.3 | 203.2 | 1.8 | 204.3 | 2.9 | | |
| | 3 | 3 | 310.2 | 1.8 | 285.0 | 2.6 | 295.4 | 1.7 | 296.9 | 4.1 | | |
| | 4 | 1 | 33.6 | 11.0 | 33.2 | 6.5 | 24.2 | 15.3 | 30.3 | 17.9 | 36.7 | 19.3 |
| E. faecalis | 4 | 2 | 45.0 | 10.1 | 41.8 | 13.0 | 41.2 | 4.7 | 42.7 | 10.0 | | |
| | 4 | 3 | 41.6 | 11.9 | 35.6 | 14.9 | 33.8 | 5.7 | 37.0 | 14.3 | | |
| | 5 | 1 | 8.4 | 18.1 | 7.0 | 20.1 | 4.4 | 34.5 | 6.6 | 33.3 | 9.6 | 38.5 |
| | 5 | 2 | 8.4 | 27.4 | 9.4 | 17.8 | 8.8 | 14.8 | 8.9 | 19.4 | | |
| | 5 | 3 | 14.4 | 18.8 | 14.6 | 20.9 | 11.4 | 17.1 | 13.5 | 21.1 | | |
| Saline | NA | 1 | 0.0 | NA | 0.2 | 225 | 0.0 | NA | 0.1 | 260.0 | 0.0 | NA |
| Saline | NA | 2 | 0.0 | NA | 0.0 | NA | 0.0 | NA | 0.0 | NA | | |
| Saline | NA | 3 | 0.0 | NA | 0.0 | NA | 0.2 | 225.0 | 0.1 | 260.0 | | |
Table 5. Reproducibility of colony counts on blood agar obtained by the APAS Compact with Urine Analysis Module
NA: Not Applicable; %CV: Percent Coefficient of Variation
¹ log10-fold dilution of 0.5 McFarland standard suspension
¹log₁₀-fold dilution of 0.5 McFarland standard suspension
{7}------------------------------------------------
| Species | Dilution1 | Replicate | Instrument 1 | | Instrument 2 | | Instrument 3 | | Overall | |
|----------------|-----------|-----------|--------------|------|--------------|-------|--------------|------|---------|-------|
| | | | Mean | %CV | Mean | %CV | Mean | %CV | Mean | %CV |
| <i>E. coli</i> | 3 | 1 | 152.2 | 3.0 | 140.2 | 3.2 | 140.2 | 2.3 | 144.2 | 4.8 |
| | | 2 | 174.6 | 4.5 | 162.4 | 1.3 | 160.0 | 1.8 | 165.7 | 4.9 |
| | | 3 | 247.2 | 2.1 | 233.0 | 3.4 | 231.8 | 2.8 | 237.3 | 4.0 |
| <i>E. coli</i> | 4 | 1 | 48.2 | 5.4 | 41.0 | 1.7 | 42.4 | 5.9 | 43.9 | 8.6 |
| | | 2 | 48.8 | 3.9 | 49.2 | 2.2 | 47.6 | 3.2 | 48.5 | 3.3 |
| | | 3 | 44.2 | 2.5 | 44.6 | 7.9 | 43.6 | 3.5 | 44.1 | 4.9 |
| <i>E. coli</i> | 5 | 1 | 4.2 | 20.0 | 4.6 | 19.3 | 3.8 | 28.9 | 4.2 | 22.4 |
| | | 2 | 2.0 | 0.0 | 1.0 | 0.0 | 1.0 | 0.0 | 1.3 | 37.7 |
| | | 3 | 8.8 | 21.8 | 7.6 | 20.0 | 8.2 | 15.9 | 8.2 | 19.1 |
| Saline | 1 | 1 | 0.0 | NA | 0.0 | NA | 0.0 | NA | 0.0 | NA |
| | | 2 | 0.0 | NA | 0.0 | NA | 0.0 | NA | 0.0 | NA |
| | | 3 | 1.0 | 71.0 | 0.2 | 225.0 | 0.0 | NA | 0.4 | 157.5 |
| | | | | | | | | | 182.4 | 22.5 |
| | | | | | | | | | 45.5 | 7.5 |
| | | | | | | | | | 4.6 | 65.2 |
| | | | | | | | | | 0.1 | 400.0 |
Table 6. Reproducibility of colony counts on MacConkey agar obtained by the APAS Compact with Urine Analysis Module
NA: Not Applicable; %CV: Percent Coefficient of Variation
1 log10-fold dilution of 0.5 McFarland standard suspension
{8}------------------------------------------------
#### b. Linearity/assay reportable range:
To determine the accuracy of colony counts obtained using the APAS Compact with Urine Analysis Module, a study was conducted using 10-fold serial dilutions of bacterial cultures that were plated on blood and/or MacConkey agar, as appropriate to the organism/organism combination. The plates were incubated and read independently by two microbiologists who each estimated colony counts. The mean of the two manual counts was used as the reference result for each plate. Each plate was also analyzed in 5 different orientations using APAS Compact with Urine Analysis Module (0, 60, 120, 180 and 270°). Note: plates that exhibited no growth on initial reading were imaged once and were not rotated to different orientations. The individual APAS Compact colony counts obtained with each plate in each orientation were compared to those obtained by the manual reference method. For analysis of the data, counts obtained by both APAS Compact and the reference method were each grouped into four categories as shown below:
| Colonies/plate | Category (CFU/mL)1 |
|----------------|--------------------|
| 0 | No Growth |
| 1-9 | $10^3$ |
| 10-99 | $10^4$ |
| ≥100 | ≥ $10^5$ |
1 Based on a 1uL inoculum volume
The results presented in Tables 7 and 8 show that all low counts obtained by APAS Compact with Urine Analysis Module were within 1-log10 of those obtained by the manual reference method. APAS Compact did not incorrectly designate any of the cultures with growth as "Negative" and there were several cases in which APAS Compact designated plates as positive for growth (i.e., ≥10° CFU/mL) but for which no growth was reported by the reference method. Together with the results of the Clinical Study in Section L(3), these data indicate that there is low probability that the APAS Compact with Urine Analysis Module will produce false negative results due to the failure to detect visible microbial colonies. In addition, because any growth detected by the APAS Compact is subject to follow-up by a trained microbiologist. the risk to patients from incorrect assignment of colony counts by the instrument is mitigated. The results of this study are therefore acceptable and provide evidence of the accuracy of colony counts obtained using the APAS Compact with Urine Analysis Module.
{9}------------------------------------------------
| E. coli | | Reference Method (CFU/mL) | | | | |
|------------------------------------------------------|-----------------|---------------------------|------|------|------|-------|
| | | No Growth | 103 | 104 | ≥105 | Total |
| | No Growth | 0 | 0 | 0 | 0 | 0 |
| | 103 | 3 | 37 | 2 | 0 | 42 |
| APAS<br>(CFU/mL) | 104 | 0 | 8 | 50 | 5 | 63 |
| | ≥105 | 0 | 0 | 13 | 110 | 123 |
| | Total | 3 | 45 | 65 | 115 | 228 |
| | % APAS < Manual | NA | 0 | 3.1 | 4.3 | |
| | % APAS = Manual | 0 | 82.2 | 76.9 | 95.7 | |
| | % APAS > Manual | 100 | 17.8 | 20.0 | NA | |
| E. faecalis | | Reference Method (CFU/mL) | | | | |
| | | No Growth | 103 | 104 | ≥105 | Total |
| | No Growth | 0 | 0 | 0 | 0 | 0 |
| | 103 | 14 | 43 | 0 | 0 | 57 |
| APAS<br>(CFU/mL) | 104 | 0 | 37 | 10 | 0 | 47 |
| | ≥105 | 0 | 0 | 45 | 105 | 150 |
| | Total | 14 | 80 | 55 | 105 | 254 |
| | % APAS < Manual | NA | 0 | 0 | 0 | |
| | % APAS = Manual | 0 | 53.8 | 18.2 | 100 | |
| | % APAS > Manual | 100 | 46.3 | 81.8 | NA | |
| P. vulgaris | | Reference Method (CFU/mL) | | | | |
| | | No Growth | 103 | 104 | ≥105 | Total |
| | No Growth | 0 | 0 | 0 | 0 | 0 |
| | 103 | 8 | 54 | 2 | 0 | 64 |
| APAS<br>(CFU/mL) | 104 | 3 | 1 | 73 | 30 | 107 |
| | ≥105 | 0 | 0 | 0 | 80 | 80 |
| | Total | 11 | 55 | 75 | 110 | 251 |
| | % APAS < Manual | NA | 0 | 2.7 | 27.3 | |
| | % APAS = Manual | 0 | 98.2 | 97.3 | 72.7 | |
| | % APAS > Manual | 100 | 1.8 | 0.0 | NA | |
| P. aeruginosa | | Reference Method (CFU/mL) | | | | |
| | | No Growth | 103 | 104 | ≥105 | Total |
| | No Growth | 0 | 0 | 0 | 0 | 0 |
| | 103 | 0 | 38 | 0 | 0 | 38 |
| APAS<br>(CFU/mL) | 104 | 0 | 21 | 32 | 0 | 53 |
| | ≥105 | 0 | 0 | 23 | 125 | 148 |
| | Total | 0 | 59 | 55 | 125 | 239 |
| | % APAS < Manual | NA | 0 | 0 | 0 | |
| | % APAS = Manual | NA | 64.4 | 58.2 | 100 | |
| <i>S. saprophyticus</i> | | Reference Method (CFU/mL) | | | | |
| | | No Growth | 103 | 104 | ≥105 | Total |
| APAS<br>(CFU/mL) | No Growth | 0 | 0 | 0 | 0 | 0 |
| | 103 | 10 | 79 | 0 | 0 | 89 |
| | 104 | 5 | 1 | 50 | 5 | 61 |
| | ≥105 | 0 | 0 | 0 | 95 | 95 |
| Total | | 15 | 80 | 50 | 100 | 245 |
| % APAS < Manual | | NA | 0 | 0 | 5.0 | |
| % APAS = Manual | | 0 | 98.8 | 100 | 95.0 | |
| % APAS > Manual | | 100 | 1.3 | 0 | NA | |
| <i>E. coli/E. faecalis</i> 1 | | Reference Method (CFU/mL) | | | | |
| | | No Growth | 103 | 104 | ≥105 | Total |
| APAS<br>(CFU/mL) | No Growth | 0 | 0 | 0 | 0 | 0 |
| | 103 | 9 | 47 | 5 | 0 | 61 |
| | 104 | 0 | 21 | 65 | 8 | 94 |
| | ≥105 | 0 | 0 | 0 | 92 | 92 |
| Total | | 9 | 68 | 70 | 100 | 245 |
| % APAS < Manual | | NA | 0 | 7.1 | 8.0 | |
| % APAS = Manual | | 0 | 69.1 | 92.9 | 92.0 | |
| % APAS > Manual | | 100 | 30.9 | 0 | NA | |
| <i>E. coli/P. vulgaris/</i><br><i>G. vaginalis</i> 2 | | Reference Method (CFU/mL) | | | | |
| | | No Growth | 103 | 104 | ≥105 | Total |
| APAS<br>(CFU/mL) | No Growth | 0 | 0 | 0 | 0 | 0 |
| | 103 | 5 | 61 | 0 | 0 | 66 |
| | 104 | 0 | 14 | 55 | 15 | 84 |
| | ≥105 | 0 | 0 | 0 | 85 | 85 |
| Total | | 5 | 75 | 55 | 100 | 235 |
| % APAS < Manual | | NA | 0 | 0 | 15.0 | |
| % APAS = Manual | | 0 | 81.3 | 100 | 85.0 | |
| % APAS > Manual | | 100 | 18.7 | 0 | NA | |
Table 7. Correlation of colony count categories obtained on blood agar using the APAS Compact with Urine Analysis Module and the manual reference method
{10}------------------------------------------------
1. I we spection of culture plates determined the presence of colonies of both 2. coli and be ratio of
equal numbers (1:1 ratio). However, no colonies with morphology consistent with G. vaginalis were observed.
{11}------------------------------------------------
| E. coli | | Reference Method (CFU/mL) | | | | |
|------------------|-----------|---------------------------|------|------|------|-------|
| | | No Growth | 103 | 104 | ≥105 | Total |
| APAS<br>(CFU/mL) | No Growth | 0 | 0 | 0 | 0 | 0 |
| | 103 | 0 | 56 | 0 | 0 | 56 |
| | 104 | 0 | 19 | 65 | 0 | 84 |
| | ≥105 | 0 | 0 | 5 | 105 | 110 |
| | Total | 0 | 75 | 70 | 105 | 250 |
| % APAS < Manual | | NA | 0 | 0 | 0 | |
| % APAS = Manual | | NA | 74.7 | 92.9 | 100 | |
| % APAS > Manual | | NA | 25.3 | 7.1 | NA | |
| P. aeruginosa | | Reference Method (CFU/mL) | | | | |
| | | No Growth | 103 | 104 | ≥105 | Total |
| APAS<br>(CFU/mL) | No Growth | 0 | 0 | 0 | 0 | 0 |
| | 103 | 0 | 29 | 0 | 0 | 29 |
| | 104 | 0 | 26 | 40 | 5 | 71 |
| | ≥105 | 0 | 0 | 35 | 69 | 104 |
| | Total | 0 | 55 | 75 | 74 | 204 |
| % APAS < Manual | | NA | 0 | 0 | 6.8 | |
| % APAS = Manual | | NA | 52.7 | 53.3 | 93.2 | |
| % APAS > Manual | | NA | 47.3 | 46.7 | NA | |
| P. vulgaris | | Reference Method (CFU/mL) | | | | |
| | | No Growth | 103 | 104 | ≥105 | Total |
| APAS<br>(CFU/mL) | No Growth | 0 | 0 | 0 | 0 | 0 |
| | 103 | 0 | 64 | 9 | 0 | 73 |
| | 104 | 0 | 1 | 61 | 7 | 69 |
| | ≥105 | 0 | 0 | 0 | 103 | 103 |
| | Total | 0 | 65 | 70 | 110 | 245 |
| % APAS < Manual | | NA | 0 | 12.9 | 6.4 | |
| % APAS = Manual | | NA | 98.5 | 87.1 | 93.6 | |
| % APAS > Manual | | NA | 1.5 | 0 | NA | |
Table 8. Correlation of colony count categories obtained on MacConkey agar using the APAS Compact with Urine Analysis Module and the manual reference method
#### c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The instructions for use indicate that the APAS Compact requires daily calibration using a dedicated optical tool that is provided with the instrument. Daily calibration of the instruments was performed throughout the analytical and clinical studies that were conducted to evaluate the performance of the device. None of the instruments failed calibration during the course of these studies.
In addition to daily optical calibration, the instructions for use indicate that operators should also perform a daily functional quality control check with cultures of Enterococcus faecalis and Escherichia coli grown on blood and MacConkey agar, respectively. In order to pass, the results from both culture media must meet the prescribed acceptance criteria for colony count and morphological designation. In
{12}------------------------------------------------
addition, uninoculated plates of both types should exhibit no growth. Table 9 shows the results obtained from Quality Controls runs performed during the APAS Compact clinical studies. The expected results were obtained with all (100%) positive controls on both blood and MacConkey agar. For the negative controls, the expected results were obtained with 98.4% on blood agar and 100% on MacConkey agar.
| Site | Instrument | Positive Control | | Negative Control | |
|------|--------------|------------------|----------------------------------------|------------------|----------------------------------------|
| | | Tested | Expected Result (%)<br>Blood MacConkey | Tested | Expected Result (%)<br>Blood MacConkey |
| 1 | 11 | 50 | 50 (100) 50 (100) | 50 | 49 (98.0)3 50 (100) |
| 1 | 22 | 49 | 49 (100) 49 (100) | 49 | 49 (100) 49 (100) |
| 2 | 3 | 13 | 13 (100) 13 (100) | Note Done4 | Not Done Not Done |
| 3 | 4 | 29 | 29 (100) 29 (100) | 30 | 29 (96.7)3 30 (100) |
| | <b>Total</b> | <b>141</b> | <b>141 (100) 141 (100)</b> | <b>129</b> | <b>127 (98.4) 129 (100)</b> |
Table 9. Results of Quality Control testing performed during the APAS Compact clinical studies
1 Negative Control for blood agar and 1 for MacConkey agar gave positive results due to artifacts on the plate. Positive and Negative QC runs for both days were repeated successfully.
2 Negative Controls for blood agar gave positive results due to artifacts on the plate. Positive and Negative QC runs for that day were repeated successfully.
3 Both Negative Control failures on blood agar were due to the presence of a single contaminating colony.
4 The study at Site 2 was completed prior to the implementation of daily Negative Control testing.
#### d. Detection Limit:
Detection of colonies by the APAS Compact is, in part, dependent upon colony size. A study was therefore conducted using a representative panel of UTI pathogens to determine the effect of colony size of the ability to detect growth. The organisms in the study were selected to include species that exhibit each of the characteristic morphologies used by the Urine Analysis Module software for designation of results.
Each organism in the study was diluted to different concentrations and inoculated onto blood and/or MacConkey agar plates which were incubated at 35±2°C until pin point colonies were visible to the naked eye. Images of the plates were then captured by the APAS Compact and the operator digitally labeled multiple isolated colonies to measure their diameter. The image of each plate was then analyzed using the Urine Analysis Module software to generate the corresponding APAS Compact result. The measurements of the colonies that were successfully detected by the APAS Compact were used to calculate the minimum colony size for each organism that could be detected with 95% probability. Results of the study are presented in Tables 10 and 11 and demonstrate the limits of colony size that can be detected reliably by the APAS Compact with Urine Analysis Module.
{13}------------------------------------------------
Table 10. Limits of detection (LOD) of the APAS Compact with Urine Analysis Module for organisms cultured on blood agar
| Detected | | | Not Detected | | | LOD in mm<br>(95% CI) | |
|---------------------------------|-----|----------------------|--------------|-----|----------------------|-----------------------|-------------------------|
| | N | Colony Diameter (mm) | | N | Colony Diameter (mm) | | |
| | | Min | Max | | Min | Max | |
| Aerococcus viridans | 141 | 0.272 | 1.268 | 81 | 0.272 | 1.178 | 1.258<br>(1.049, 1.692) |
| Enterococcus faecalis | 225 | 0.181 | 1.178 | 165 | 0.181 | 1.087 | 0.856<br>(0.776, 0.980) |
| Escherichia coli | 240 | 0.181 | 1.721 | 123 | 0.181 | 0.906 | 0.899<br>(0.821, 1.024) |
| Proteus mirabilis | 140 | 0.634 | 2.808 | 281 | 0.181 | 2.717 | 2.501<br>(2.168, 3.091) |
| Pseudomonas aeruginosa | 238 | 0.453 | 3.442 | 153 | 0.181 | 2.355 | 4.037<br>(2.881, 7.121) |
| Staphylococcus<br>saprophyticus | 121 | 0.362 | 0.996 | 98 | 0.181 | 0.906 | 0.877<br>(0.784, 1.049) |
| Streptococcus agalactiae | 120 | 0.362 | 0.725 | 138 | 0.181 | 0.725 | 0.625<br>(0.584, 0.697) |
N: number of colonies; Min: minimum; Max: maximum
Table 11. Limits of detection (LOD) of the APAS Compact with Urine Analysis Module for organisms cultured on MacConkey agar
| | | Detected | | | Not Detected | | | |
|------------------------|-----|----------|----------------------|-----|--------------|----------------------|--|-------------------------|
| | N | | Colony Diameter (mm) | | | Colony Diameter (mm) | | LOD in mm<br>(95% CI) |
| | | Min | Max | N | Min | Max | | |
| Escherichia coli | 112 | 0.996 | 2.264 | 136 | 0.362 | 1.178 | | 1.253<br>(1.214, 1.324) |
| Morganella morganii | 170 | 0.634 | 1.721 | 170 | 0.272 | 1.268 | | 1.262<br>(1.164, 1.421) |
| Pseudomonas aeruginosa | 48 | 0.634 | 2.989 | 78 | 0.181 | 1.087 | | 1.170<br>(1.020, 1.596) |
| Serratia marcescens | 76 | 0.725 | 1.087 | 296 | 0.181 | 1.087 | | 1.613<br>(1.398, 2.093) |
N: number of colonies; Min: minimum; Max: maximum
# e. Analytical Specificity:
The ability of the APAS Compact with Urine Analysis Module to assign the appropriate colony morphology to different clinically relevant organisms associated with UTIs was evaluated by testing serial dilutions of both pure and mixed cultures. Dilutions of each organism or organism combination were inoculated in triplicate onto the appropriate culture plates which were incubated at 35±2℃ for 18 hours. The APAS Compact was used to capture images of plates that exhibited 1-100 individual colonies or areas of confluent growth and >100 isolated colonies. Each plate was read on the APAS Compact instrument in 3 different orientations (0, 120 and 270°). The plates were then re-incubated and imaged again after 20 and 22 hours. Results for assignment of colony morphology are shown for pure and mixed cultures below in Tables 12-15. The percentage of plate images with at least one colony of the expected morphology is reported as agreement.
{14}------------------------------------------------
#### Pure Cultures
With pure cultures on blood agar, the APAS Compact with Urine Analysis Module reported detection of the expected colony morphology on each image for all the species tested at each time point except with Aerococcus urinae, Lactobacillus rhamnosus, Pseudomonas aeruginosa and Staphylococcus epidermidis (Table 12). For all species except A. urinae and L. rhamnosus, the APAS Compact designation of each plate at each time point was either "Positive" or "Review."
For S. marcescens on MacConkey agar, the expected morphology was observed on all plates after 20 and 22 hours of incubation but not after 18 hours (Table 13). All images of cultures on MacConkey agar were designated as "Positive" by the APAS Compact.
Table 12. Detection of expected colony morphology by the APAS Compact with Urine Analysis Module: pure cultures on blood agar
| Species | Expected Colony<br>Morphology | Agreement1 | |
|------------------------------|-------------------------------|--------------|---------|
| | | Number/Total | Percent |
| Aerococcus urinae | Small, alpha hemolytic | 53/54 | 98.12 |
| Aerococcus viridans | Small, alpha hemolytic | 54/54 | 100 |
| Candida albicans | Small | 54/54 | 100 |
| Enterobacter cloacae | Coliform | 54/54 | 100 |
| Escherichia coli (strain 1) | Coliform | 54/54 | 100 |
| Escherichia coli (strain 2) | Coliform | 54/54 | 100 |
| Enterococcus faecalis | Small | 54/54 | 100 |
| Klebsiella pneumoniae | Coliform | 52/52 | 100 |
| Lactobacillus rhamnosus | Small | 6/27 | 22.23 |
| Morganella morganii | Coliform | 54/54 | 100 |
| Proteus mirabilis | Swarming | 54/54 | 100 |
| Pseudomonas aeruginosa | Granular | 50/51 | 98.04 |
| Staphylococcus aureus | Cream white | 54/54 | 100 |
| Staphylococcus epidermidis | Cream white | 38/54 | 70.45 |
| Staphylococcus saprophyticus | Cream white | 51/51 | 100 |
| Streptococcus agalactiae | Beta-hemolytic | 54/54 | 100 |
| Streptococcus dysgalactiae | Beta-hemolytic | 51/51 | 100 |
Note: The APAS Compact designation of each plate at each time point was either "Positive" for all species except A. urinae and L. rhamnosus
Number of plate images with at least one colony of the expected morphology (all time points and plate rotations combined)/total
ର Expected morphology detected: 18 hours: 18/18 (88.9%): 20 hours: 18/18 (100%): 22 hours: 17/18 (94.4%)
3 Expected morphology detected: 18 hours: 0/9 (0%); 20 hours: 0/9 (0%); 22 hours: 6/9 (66.7%)
4 Expected morphology detected: 18 hours: 18/18 (100%); 20 hours: 17/18 (94.4%); 22 hours: 15/15 (100%)
్ Expected morphology detected: 18 hours: 9/18 (50.0%); 20 hours: 11/18 (61.1%); 22 hours: 18/18 (100%)
{15}------------------------------------------------
Table 13. Detection of expected colony morphology by the APAS Compact with Urine Analysis Module: pure cultures on MacConkey agar
| Species | Expected Colony<br>Morphology | Agreement1 | |
|------------------------------|-------------------------------|--------------|---------|
| | | Number/Total | Percent |
| <i>Citrobacter koseri</i> | Non-fermenter | 53/53 | 100 |
| <i>Escherichia coli</i> | Lactose fermenter | 54/54 | 100 |
| <i>Klebsiella pneumoniae</i> | Lactose fermenter | 54/54 | 100 |
| <i>Morganella morganii</i> | Non-fermenter | 54/54 | 100 |
| <i>Serratia marcescens</i> 2 | Red pigment | 21/27 | 77.8 |
Note: All plate images of pure cultures on MacConkey agar were designated by APAS Compact as "Positive"
Number of plate images with at least one colony of the expected morphology (all time points and plate rotations combined)/total
2 Expected morphology detected: 3/9 (33.3%); 20 hours: 9/9 (100%); 22 hours: 9/9 (100%)
#### Mixed Cultures
For blood agar plates inoculated with two organisms at a 1:1 ratio, the APAS Compact reported detection of both the expected colony types at each time point in all but two images (Table 14). In both cases, the APAS Compact failed to detect the expected morphology for S. agalactiae in the presence of E. faecalis, although both images were reported as "Positive" indicating of the need for microbiological followup. With plates inoculated using a 1:10 ratio of S. agalactiae:E. faecalis there was a higher proportion of images in which colonies with the characteristic morphology of S. agalactiae were not detected. Nevertheless, all the images were still reported as "Positive."
For mixed cultures plated on MacConkey agar, both the expected colony types were detected in all but 1 image, although this image was still reported "Positive" by the instrument (Table 15).
Note: Because any growth detected by the APAS Compact with Urine Analysis Module is subject to follow-up by a trained microbiologist, the risk to patients from incorrect assignment of colony morphology by the instrument is mitigated.
Instructions for appropriate laboratory follow-up of any growth detected by the APAS Compact with Urine Analysis Module are included in the device labeling.
{16}------------------------------------------------
Table 14. Detection of expected colony morphology by the APAS Compact with Urine Analysis Module: mixed cultures on blood agar
| Species Name & Expected Morphology | | | | | Agreement (%) 2 | |
|------------------------------------|--------------------|---------------------------------|----------------|-----------|-------------------|----------------|
| Species 1 | | Species 2 | | Mixture 1 | Species 1 | Species 2 |
| Escherichia coli | Coliform | Enterococcus faecalis | Small | 1:1 | 63/63<br>(100) | 63/63<br>(100) |
| | | | | 1:10 | 81/81<br>(100) | 81/81<br>(100) |
| Streptococcus agalactiae | Beta-<br>hemolytic | Enterococcus faecalis | Small | 1:1 | 81/81<br>(100) | 81/81<br>(100) |
| | | | | 1:10 | 28/63<br>(44.4) 3 | 63/63<br>(100) |
| Escherichia coli | Coliform | Staphylococcus<br>saprophyticus | Cream<br>white | 1:1 | 54/54<br>(100) | 54/54<br>(100) |
| | | | | 1:10 | 81/81<br>(100) | 81/81<br>(100) |
| Escherichia coli | Coliform | Aerococcus viridans | Alpha- | 1:1 | 81/81<br>(100) | 81/81<br>(100) |
| | | | hemolytic | 1:10 | 81/81<br>(100) | 81/81<br>(100) |
Note: All plate images of mixed cultures on blood agar were designated by APAS Compact as "Positive"
Approximate ratio of concentration, Species 1:Species 2
ରା Number of plate images with at least one colony of the expected morphology (all time points and plate rotations combined)/total
3 Expected morphology detected: 18 hours: 12/21 (57.1%); 20 hours: 10/21 (47.6%); 22 hours: 6/21 (28.6%)
### Table 15. Detection of expected colony morphology by the APAS Compact with Urine Analysis Module: mixed cultures on MacConkey agar
| Species Name & Expected Morphology | | | Agreement (%) 2 | | | |
|------------------------------------|----------------------|------------------------|-------------------|-----------|-------------------|----------------|
| Species 1 | | Species 2 | | Mixture 1 | Species 1 | Species 2 |
| Escherichia coli | Lactose<br>Fermenter | Morganella morganii | Non-<br>fermenter | 1:1 | 53/53<br>(100) | 53/53<br>(100) |
| Escherichia coli | Lactose<br>Fermenter | Morganella morganii | Non-<br>fermenter | 1:10 | 79/80<br>(98.8) 3 | 80/80<br>(100) |
| Escherichia coli | Lactose<br>Fermenter | Pseudomonas aeruginosa | Non-<br>fermenter | 1:1 | 81/81<br>(100) | 81/81<br>(100) |
| Escherichia coli | Lactose<br>Fermenter | Pseudomonas aeruginosa | Non-<br>fermenter | 1:10 | 72/72<br>(100) | 72/72<br>(100) |
Instances of agreement with the expected morphology <100% are shaded
Note: All images of mixed cultures on MacConkey agar were designated by APAS Compact as "Positive"
Approximate ratio of concentration, Species 1:Species 2
2 Number of plate images with at least one colony of the expected morphology (all time points and plate rotations combined)/total
ന Expected morphology detected: 18 hours: 25/26 (96.2%); 20 hours: 27/27 (100%); 22 hours: 27/27 (100%)
- Assay Cut-off: .f.
Not applicable.
#### 2. Comparison studies:
- a. Method comparison with predicate device:
Not applicable. Refer to Section L(3), Clinical Studies.
{17}------------------------------------------------
- b. Matrix comparison:
Not applicable.
#### 3. Clinical Studies:
- a. Clinical Sensitivity:
To evaluate the performance of the APAS Compact with Urine Analysis Module, clinical studies were conducted at three sites (1 US and 2 ex-US) using remnant urine samples that were leftover from standard of care culture for suspected UTI. Each urine sample was inoculated onto one sheep blood agar and one MacConkey agar culture plate (both Remel) that were incubated for 18 hours at 35±2℃ prior to analysis by the APAS Compact and by an independent panel of three microbiologists. Results were reported by the Urine Analysis Module software as "Negative," "Review" or "Positive" as described in Table 4, above. For the analysis of performance, APAS Compact designations of "Review" and "Positive" were regarded as positive. For the reference method, each microbiologist was trained to read the urine culture plates in a standard fashion and was blinded to the results from the other panel members and to those obtained by the APAS Compact.
Only urine samples from "uncomplicated" cases of suspected UTI were included in the analysis of performance (i.e., those not obtained from subjects with known underlying urinary tract pathology, suspected cases of persistent urinary tract infection, urine collected by invasive procedures or urine samples from immunocompromised subjects). A summary of the reasons for exclusion of samples from the performance calculations is shown in Table 16.
| Description | Site1 | | | Total |
|-----------------------|-------|------|------|-------|
| | 1 | 2 | 3 | |
| Enrolled | 5835 | 2117 | 2148 | 10100 |
| Included | 5634 | 1769 | 1821 | 9224 |
| Excluded | 201 | 348 | 327 | 876 |
| Reasons for exclusion | | | | |
| Complicated UTI | 29 | 0 | 0 | 29 |
| Read time1 | 156 | 0 | 235 | 391 |
| Incomplete record | 16 | 6 | 92 | 114 |
| LIMS Flag2 | 0 | 342 | 0 | 342 |
Table 16. Number of urine samples enrolled at each clinical site and reasons for exclusion from the analysis of performance
1 Site 1: US; Sites 2 and 3: ex-US
2 Culture plates not read within specified timeframe after incubation
2 Additional data from Laboratory Information Management System (LIMS) used in algorithm for sample disposition (e.g., white cell count)
Tables 17 and 18 summarize the demographic characteristics of the subjects enrolled at each clinical site. The majority of urine samples were from female donors, which is consistent with the higher prevalence of UTIs in the female population. Samples were
{18}------------------------------------------------
enrolled from subjects ranging from <20 to >80 years of age, with the majority from subjects between 20 and 79 years of age.
| Site1 | Male | Female | Not Known | Total | % Male | % Female |
|-------|------|--------|-----------|-------|--------|----------|
| 1 | 972 | 4659 | 3 | 5634 | 17.3 | 82.7 |
| 2 | 702 | 1066 | 1 | 1769 | 39.7 | 60.3 |
| 3 | 810 | 1007 | 4 | 1821 | 44.6 | 55.4 |
| Total | 2484 | 6732 | 8 | 9224 | 27.0 | 73.0 |
Table 17. Gender distribution of donors of urine samples enrolled in the clinical studies
- Site 1: US; Sites 2 and 3: ex-US
| Site1 | | | | | |
|-------------|------|------|------|-------|------|
| Age (years) | 1 | 2 | 3 | Total | % |
| ≤20 | 747 | 154 | 97 | 998 | 10.8 |
| 20-39 | 1611 | 415 | 458 | 2384 | 25.9 |
| 40-59 | 1247 | 440 | 450 | 2137 | 23.2 |
| 60-79 | 1460 | 548 | 614 | 2622 | 28.4 |
| ≥80 | 567 | 212 | 300 | 1079 | 11.7 |
| Not Known | 2 | 0 | 2 | 4 | 0.0 |
| | 5634 | 1769 | 1821 | 9224 | 100 |
Table 18. Age distribution of urine sample donors in the clinical studies
1 Site 1: US; Sites 2 and 3: ex-US
A summary of the APAS Compact designation for samples from the clinical studies in relation to colony counts as determined by the reference microbiologist panel is shown in Tables 19-21. Colony counts by both methods were categorized as described in Section L(1)(b). Across all sites combined, the APAS Compact correctly designated 99.0% (95% confidence interval: 98.7-99.2%) of cultures on blood agar and 99.5% (95% CI: 99.2-99.7%) of cultures on MacConkey agar as positive for growth. Depending on the colony count observed by the microbiologist panel, the false negative rate for detection of microbial growth on blood agar ranged from 0% to 2.9% and on MacConkey agar from 0% to 1.3%.
False-negative results with the APAS Compact for the detection of growth could lead to discard of culture plates without further review. The results of this study demonstrate that the likelihood of such results is acceptably low.
Tables 22, 23 and 24 show the correlation of colony count categories obtained by the APAS Compact and by the reference method. In general, the colony count categories reported by the APAS Compact were the same or higher than those obtained by the reference method, and there were multiple instances in which APAS Compact designated plates that exhibited "No Growth" according to the reference method as "Positive" or "Review." This is acceptable because all cultures designated by APAS Compact as "Positive" or "Review" are subject to further microbiological follow-up and any such incorrect designations do not increase risks to patients.
{19}------------------------------------------------
| Site 1 | Reference Result | APAS Designation | | | Total | Correct 2 | % Correct<br>(95% CI) |
|--------|------------------|------------------|------|------|-------|-----------|-----------------------|
| 1 | No Growth | 1220 | 14 | 256 | 1490 | 1220 | 81.9 (79.8, 83.8) |
| 1 | 103 CFU/mL | 37 | 393 | 845 | 1275 | 1238 | 97.1 (96.0, 97.9) |
| 1 | 104 CFU/mL | 4 | 1089 | 74 | 1167 | 1163 | 99.7 (99.1, 99.9) |
| 1 | ≥105 CFU/mL | 0 | 1700 | 2 | 1702 | 1702 | 100 (99.8, 100) |
| 1 | Any Growth | 41 | 3182 | 921 | 4144 | 4103 | 99.0 (98.7, 99.3) |
| 2 | No Growth | 438 | 18 | 57 | 513 | 438 | 85.4 (82.1, 88.2) |
| 2 | 103 CFU/mL | 13 | 152 | 308 | 473 | 460 | 97.3 (95.4, 98.4) |
| 2 | 104 CFU/mL | 2 | 461 | 25 | 488 | 486 | 99.6 (98.5, 99.9) |
| 2 | ≥105 CFU/mL | 0 | 295 | 0 | 295 | 295 | 100 (98.7, 100) |
| 2 | Any Growth | 15 | 908 | 333 | 1256 | 1241 | 98.8 (98.0, 99.3) |
| 3 | No Growth | 574 | 2 | 61 | 637 | 574 | 90.1 (87.5, 92.2) |
| 3 | 103 CFU/mL | 8 | 101 | 262 | 371 | 363 | 97.8 (95.8, 98.9) |
| 3 | 104 CFU/mL | 2 | 355 | 23 | 380 | 378 | 99.5 (98.1, 99.9) |
| 3 | ≥105 CFU/mL | 3 | 422 | 8 | 433 | 430 | 99.3 (98.0, 99.8) |
| 3 | Any Growth | 13 | 878 | 293 | 1184 | 1171 | 98.9 (98.1, 99.4) |
| Total | No Growth | 2232 | 34 | 374 | 2640 | 2232 | 84.5 (83.1, 85.9) |
| Total | 103 CFU/mL | 46 | 646 | 1415 | 2119 | 2061 | 97.3 (96.5, 97.9) |
| Total | 104 CFU/mL | 8 | 1905 | 122 | 2035 | 2027 | 99.6 (99.2, 99.8) |
| Total | ≥105 CFU/mL | 3 | 2417 | 10 | 2430 | 2427 | 99.9 (99.6, 100) |
| Total | Any Growth | 69 | 4968 | 1547 | 6584 | 6515 | 99.0 (98.7. 99.2) |
Table 19. Performance of the APAS Compact with Urine Analysis Module for detection of growth on blood agar
CI: 95% score Confidence Interval
Site 1: US; Sites 2 and 3: ex-US
4 For the purposes of data analysis, APAS Compact designations of "Positive" were both considered indicative of microbial growth.
For a Reference Result of No Growth, an APAS Compact designation of "Negative" was considered correct.
For Reference Results of 10, 10, ≥10 CFU/nL or Any Growth, an APAS Compact designation of "Positive" or "Review" was considered correct.
{20}------------------------------------------------
| Site 1 | Reference Result | APAS Designation | | | Total | Correct | % Correct 2 (95% CI) |
|--------|------------------|------------------|------|-----|-------|---------|----------------------|
| 1 | No Growth | 3274 | 23 | 37 | 3334 | 3274 | 98.2 (97.7, 98.6) |
| 1 | 103 CFU/mL | 6 | 406 | 67 | 479 | 473 | 98.7 (97.3, 99.4) |
| 1 | 104 CFU/mL | 1 | 469 | 8 | 478 | 477 | 99.8 (98.8, 100) |
| 1 | ≥105 CFU/mL | 2 | 1341 | 0 | 1343 | 1341 | 99.9 (99.5, 100) |
| 1 | Any Growth | 9 | 2216 | 75 | 2300 | 2291 | 99.6 (99.3, 99.8) |
| 2 | No Growth | 1257 | 4 | 4 | 1265 | 1257 | 99.4 (98.8, 99.7) |
| 2 | 103 CFU/mL | 2 | 136 | 48 | 186 | 184 | 98.9 (96.2, 99.7) |
| 2 | 104 CFU/mL | 1 | 152 | 1 | 154 | 153 | 99.4 (96.4, 99.9) |
| 2 | ≥105 CFU/mL | 0 | 164 | 0 | 164 | 164 | 100 (97.7, 100) |
| 2 | Any Growth | 3 | 452 | 49 | 504 | 501 | 99.4 (98.3, 99.8) |
| 3 | No Growth | 1241 | 2 | 0 | 1243 | 1241 | 99.8 (99.4, 100) |
| 3 | 103 CFU/mL | 3 | 105 | 39 | 147 | 144 | 98.0 (94.2, 99.3) |
| 3 | 104 CFU/mL | 1 | 144 | 1 | 146 | 145 | 99.3 (96.2, 99.9) |
| 3 | ≥105 CFU/mL | 0 | 285 | 0 | 285 | 285 | 100 (98.7, 100) |
| 3 | Any Growth | 4 | 534 | 40 | 578 | 574 | 99.3 (98.2, 99.7) |
| Total | No Growth | 5772 | 29 | 41 | 5842 | 5772 | 98.8 (98.5, 99.1) |
| Total | 103 CFU/mL | 11 | 647 | 154 | 812 | 801 | 98.6 (97.6, 99.2) |
| Total | 104 CFU/mL | 3 | 765 | 10 | 778 | 775 | 99.6 (98.9, 99.9) |
| Total | ≥105 CFU/mL | 2 | 1790 | 0 | 1792 | 1790 | 99.9 (99.6, 100) |
| Total | Any Growth | 16 | 3202 | 164 | 3382 | 3366 | 99.5 (99.2, 99.7) |
Table 20. Performance of the APAS Compact with Urine Analysis Module for detection of growth on MacConkey agar
CI: 95% score Confidence Interval ¹ Site 1: US; Sites 2 and 3: ex-US
2 For the purposes of data analysis, APAS Compact designations of "Review" and "Positive" were both considered indicative of microbial growth.
For a Reference Result of No Growth, an APAS Compact designation of "Negative" was considered correct.
For Reference Results of 10, 10, ≥10 CFU/nL or Any Growth, an APAS Compact designation of "Positive" or "Review" was considered correct.
{21}------------------------------------------------
Table 21. Performance of the APAS Compact with Urine Analysis Module for detection of growth by sample (blood and MacConkey agars combined)
| Site 1 | Reference Result | APAS Designation | | | Total | Correct | % Correct 2<br>(95% CI) |
|--------|------------------|------------------|------|------|-------|---------|-------------------------|
| 1 | No Growth | 1166 | 25 | 261 | 1452 | 1166 | 80.3 (78.2, 82.3) |
| | 103 CFU/mL | 37 | 463 | 780 | 1280 | 1243 | 97.1 (96.0, 97.9) |
| 1 | 104 CFU/mL | 4 | 1104 | 64 | 1172 | 1168 | 99.7 (99.1, 99.9) |
| | ≥105 CFU/mL | 0 | 1729 | 1 | 1730 | 1730 | 100 (99.8, 100) |
| | Any Growth | 41 | 3296 | 845 | 4182 | 4141 | 99.0 (98.7, 99.3) |
| 2 | No Growth | 429 | 19 | 57 | 505 | 429 | 85.0 (81.6, 87.8) |
| | 103 CFU/mL | 13 | 166 | 300 | 479 | 466 | 97.3 (95.4, 98.4) |
| 2 | 104 CFU/mL | 1 | 464 | 25 | 490 | 489 | 99.8 (98.9, 100) |
| | ≥105 CFU/mL | 0 | 295 | 0 | 295 | 295 | 100 (98.7, 100) |
| | Any Growth | 14 | 925 | 325 | 1264 | 1250 | 98.9 (98.1, 99.3) |
| 3 | No Growth | 555 | 3 | 60 | 618 | 555 | 89.8 (87.2, 92.0) |
| | 103 CFU/mL | 9 | 115 | 262 | 386 | 377 | 97.7 (95.6, 98.8) |
| 3 | 104 CFU/mL | 2 | 356 | 23 | 381 | 379 | 99.5 (98.1, 99.9) |
| | ≥105 CFU/mL | 3 | 426 | 7 | 436 | 433 | 99.0 (98.0, 99.8) |
| | Any Growth | 14 | 897 | 292 | 1203 | 1189 | 98.8 (98.1, 99.3) |
| Total | No Growth | 2150 | 47 | 378 | 2575 | 2150 | 83.5 (96.7, 99.1) |
| | 103 CFU/mL | 59 | 744 | 1342 | 2145 | 2086 | 97.2 (96.5, 97.9) |
| Total | 104 CFU/mL | 7 | 1924 | 112 | 2043 | 2036 | 99.7 (99.3, 99.8) |
| | ≥105 CFU/mL | 3 | 2450 | 8 | 2461 | 2458 | 99.9 (99.6, 100) |
| | Any Growth | 69 | 5118 | 1462 | 6649 | 6580 | 99.0 (98.7, 99.2) |
CI: 95% score Confidence Interval
¹ Site 1: US; Sites 2 and 3: ex-US
2 For the purposes of data analysis, APAS Compact designations of "Review" and "Positive" were both considered indicative of microbial growth.
For a Reference Result of No Growth, an APAS Compact designation of "Negative" was considered correct.
For Reference Results of 10, 10, ≥10 CFU/nL or Any Growth, an APAS Compact designation of "Positive" or "Review" was considered correct.
{22}------------------------------------------------
| Site 1 (US) | | Reference Method (CFU/mL) | | | | |
|----------------------------------------------|-----------|---------------------------|------|------|------|-------|
| | | No Growth | 103 | 104 | ≥105 | Total |
| APAS<br>(CFU/mL) | No Growth | 1220 | 37 | 4 | 0 | 1261 |
| | 103 | 262 | 1049 | 87 | 2 | 1400 |
| | 104 | 7 | 182 | 828 | 27 | 1044 |
| | ≥105 | 1 | 7 | 248 | 1673 | 1929 |
| | Total | 1490 | 1275 | 1167 | 1702 | 5634 |
| % APAS < Reference | | NA | 2.9 | 7.8 | 1.7 | |
| % APAS = Reference | | 81.9 | 82.3 | 71.0 | 98.3 | |
| % APAS > Reference | | 18.1 | 14.8 | 21.3 | NA | |
| Site 2 (ex-US) | | Reference Method (CFU/mL) | | | | |
| | | No Growth | 103 | 104 | ≥105 | Total |
| APAS<br>(CFU/mL) | No Growth | 438 | 13 | 2 | 0 | 453 |
| | 103 | 68 | 388 | 31 | 1 | 488 |
| | 104 | 7 | 63 | 368 | 21 | 459 |
| | ≥105 | 0 | 9 | 87 | 273 | 369 |
| | Total | 513 | 473 | 488 | 295 | 1769 |
| % APAS < Reference | | NA | 2.7 | 6.8 | 7.5 | |
| % APAS = Reference | | 85.4 | 82.0 | 75.4 | 92.5 | |
| % APAS > Reference | | 14.6 | 15.2 | 17.8 | NA | |
| Site 3 (ex-US) | | Reference Method (CFU/mL) | | | | |
| | | No Growth | 103 | 104 | ≥105 | Total |
| APAS<br>(CFU/mL) | No Growth | 574 | 8 | 2 | 3 | 587 |
| | 103 | 63 | 312 | 27 | 8 | 410 |
| | 104 | 0 | 45 | 245 | 12 | 302 |
| | ≥105 | 0 | 6 | 106 | 410 | 522 |
| | Total | 637 | 371 | 380 | 433 | 1821 |
| % APAS < Reference | | NA | 2.2 | 7.6 | 5.3 | |
| % APAS = Reference | | 90.1 | 84.1 | 64.5 | 94.7 | |
| % APAS > Reference | | 9.9 | 13.7 | 27.9 | NA | |
| All Sites Combined | | Reference Method (CFU/mL) | | | | |
| | | No Growth | 103 | 104 | ≥105 | Total |
| APAS<br>(CFU/mL) | No Growth | 2232 | 58 | 8 | 3 | 2301 |
| | 103 | 393 | 1749 | 145 | 11 | 2298 |
| | 104 | 14 | 290 | 1441 | 60 | 1805 |
| | ≥105 | 1 | 22 | 441 | 2356 | 2820 |
| | Total | 2640 | 2119 | 2035 | 2430 | 9224 |
| % APAS < Reference | | NA | 2.7 | 7.5 | 3.0 | |
| % APAS = Reference | | 84.5 | 82.5 | 70.8 | 97.0 | |
| Site 1 (US) | | Reference Method (CFU/mL) | | | | |
| | | No Growth | 103 | 104 | ≥105 | Total |
| APAS<br>(CFU/mL) | No Growth | 3274 | 6 | 1 | 2 | 3283 |
| | 103 | 56 | 440 | 11 | 0 | 507 |
| | 104 | 4 | 33 | 407 | 38 | 482 |
| | ≥105 | 0 | 0 | 59 | 1303 | 1362 |
| | Total | 3334 | 479 | 478 | 1343 | 5634 |
| % APAS < Reference | | NA | 1.3 | 2.5 | 3.0 | |
| % APAS = Reference | | 98.2 | 91.9 | 85.1 | 97.0 | |
| % APAS > Reference | | 1.8 | 6.9 | 12.3 | NA | |
| Site 2 (ex-US) | | Reference Method (CFU/mL) | | | | |
| | | No Growth | 103 | 104 | ≥105 | Total |
| APAS<br>(CFU/mL) | No Growth | 1257 | 2 | 1 | 0 | 1260 |
| | 103 | 8 | 168 | 4 | 0 | 180 |
| | 104 | 0 | 16 | 136 | 8 | 160 |
| | ≥105 | 0 | 0 | 13 | 156 | 169 |
| | Total | 1265 | 186 | 154 | 164 | 1769 |
| % APAS < Reference | | NA | 1.1 | 3.2 | 4.9 | |
| % APAS = Reference | | 99.4 | 90.3 | 88.3 | 95.1 | |
| % APAS > Reference | | 0.6 | 8.6 | 8.4 | NA | |
| Site 3 (ex-US) | | Reference Method (CFU/mL) | | | | |
| | | No Growth | 103 | 104 | ≥105 | Total |
| APAS<br>(CFU/mL) | No Growth | 1241 | 3 | 1 | 0 | 1245 |
| | 103 | 2 | 127 | 3 | 0 | 132 |
|…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.