E. COLI PNA FISH

K082068 · Advandx, Inc. · JSS · Jun 25, 2009 · Microbiology

Device Facts

Record IDK082068
Device NameE. COLI PNA FISH
ApplicantAdvandx, Inc.
Product CodeJSS · Microbiology
Decision DateJun 25, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.2660
Device ClassClass 1

Indications for Use

E. coli PNA FISH® is a qualitative nucleic acid hybridization assay intended for the presumptive identification of Escherichia coli on smears from positive blood cultures containing gram-negative rods. The test does not distinguish between E. coli and Shigella spp. Confirm all positive results with conventional identification methods. Subculturing of positive blood cultures is necessary to recover organisms for definitive identification and susceptibility testing.

Device Story

E. coli PNA FISH is a qualitative diagnostic assay for presumptive identification of E. coli in positive blood cultures. Input: smear from positive blood culture containing gram-negative rods. Process: fluorescence in situ hybridization (FISH) using fluorescein-labeled peptide nucleic acid (PNA) probes targeting E. coli-specific ribosomal RNA. Procedure involves fixation, 90-minute hybridization at 55°C, stringent wash, and mounting. Output: visual observation of green fluorescence via fluorescence microscopy (using Dual Band Filter). Performed in clinical laboratories by trained personnel. Results provide rapid presumptive identification (2.5 hours) to assist clinical decision-making regarding antibiotic therapy while awaiting definitive subculture identification. Benefits include faster time-to-result compared to conventional culture-based methods.

Clinical Evidence

Clinical study conducted at five sites using 122 GNR-positive blood culture bottles. Performance compared to conventional identification methods (Vitek2, Chromagar/Indol, API ID 32E). Sensitivity ranged from 96.4% to 100% across different study sites/comparators; specificity ranged from 88.9% to 100%. Analytical specificity testing showed cross-reactivity with E. albertii, E. fergusonii, and Shigella spp. Detection limit is approximately 10⁵ CFU/mL.

Technological Characteristics

Fluorescence in situ hybridization (FISH) using peptide nucleic acid (PNA) probes. Requires Dual Band Filter (Cat. No. AC003) and microscope slides. Hybridization and wash steps performed at 55°C. Qualitative visual readout via fluorescence microscopy.

Indications for Use

Indicated for use as an aid in the diagnosis of E. coli bacteremia in patients with positive blood cultures containing gram-negative rods.

Regulatory Classification

Identification

A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: K082068 B. Purpose for Submission: New submission for SE determination C. Measurand: E. coli specific ribosomal RNA sequences D. Type of Test: Fluorescence in situ Hybridization (FISH) using protein nucleic acid (PNA) probes E. Applicant: AdvanDx, Inc F. Proprietary and Established Names: E. coli PNA FISH® G. Regulatory Information: 1. Regulation section: 866.2660 2. Classification: Class I 3. Product code: JSS 4. Panel: 83 Microbiology {1} H. Intended Use: 1. Intended use(s): E. coli PNA FISH® is a qualitative nucleic acid hybridization assay intended for the presumptive identification of Escherichia coli on smears from positive blood cultures containing gram-negative rods. The test does not distinguish between E. coli and Shigella spp. Confirm all positive results with conventional identification methods. Subculturing of positive blood cultures is necessary to recover organisms for definitive identification and susceptibility testing. 2. Indication(s) for use: E. coli PNA FISH® is indicated for use as an aid in the diagnosis of E. coli bacteremia. 3. Special conditions for use statement(s): Prescription use only 4. Special instrument requirements: Dual Band Filter (Cat. No. AC003) Microscope Slides (Cat. No. AC001) I. Device Description: E. coli PNA FISH® is a fluorescence in situ hybridization (FISH) method using PNA probes hybridizing to E. coli-specific ribosomal RNA sequences. PNA FISH is performed directly on smears fixed onto microscope slides. Hybridization is performed at 55°C for 90 min. The coverslip is removed and a post-hybridization wash at 55°C for 30 min. with a stringent wash solution to remove unbound PNA probe. The smear is finally mounted with Mounting Medium for examination with fluorescence microscopy (Dual Band Filter). While maintaining the morphology of the cells, presumptive E. coli cells show green fluorescence by binding of the fluorescein-labeled PNA probes. E. coli PNA FISH® is provided in a kit box with the following kit components: GN Fixation Solution E. coli PNA {2} Wash Solution Mounting Medium ## J. Substantial Equivalence Information: 1. Predicate device name(s): S. aureus PNA FISH™ 2. Predicate 510(k) number(s): K060099 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Technology | Fluorescence In Situ Hybridization (FISH) using protein nucleic acid (PNA) probe | Same | | Sample | Positive blood culture | Same | | Time to Result | 2.5 hours | Same | | Interpretation of Results | Qualitative Fluorescence microscope | Same | | | | | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Function | Presumptive identification of Escherichia coli | Identification of S. aureus | | Control organisms | Pos control: E. coli ATCC 11775 Neg control: P. aeruginosa ATCC 10145 | Pos control: S. aureus Neg control: S. epidermidis | | | | | ## K. Standard/Guidance Document Referenced (if applicable): Not applicable ## L. Test Principle: {3} The PNA FISH® technology uses species-specific peptide nucleic acid (PNA) probes in a fluorescence in situ hybridization format. A mixture of a fluorescein-labeled E.coli specific PNA probe is added to a smear prepared from a positive blood culture. Hybridization is performed at 55°C for 90 minutes. The hybridization is followed by a post-hybridization wash at 55°C for 30 min with a stringent Wash Solution. Finally, the smear is mounted with Mounting Medium and examined by fluorescence microscopy. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: #### a. Precision/Reproducibility: A Reproducibility study for E. coli PNA FISH® assay was performed by using ten reference isolates of Gram negative rods, once per day with positive and negative controls, over a period of three days at three different sites, by an operator at each site performing and reading the assay according to the package insert. The study is acceptable for a qualitative assay as only positive or negative results are provided #### b. Linearity/assay reportable range: Not applicable #### c. Traceability, Stability, Expected values (controls, calibrators, or methods): Positive and negative control slides were performed at each testing site with expected results. #### d. Detection limit: The detection limit was determined to be approximately 10⁵ CFU/mL by serial dilutions of E. coli positive cultures. Serial dilutions of exponentially growing cultures were prepared. Then 0.01 mL was plated on growth media and 0.01 mL was used for preparation of smears. The smears were run through PNA FISH and scored positive or negative. The following day, colonies were counted on the plates and the average number of colonies per dilution was calculated. The data sets showed a minimum of 10⁵ CFU/mL to produce a positive result for the E. coli PNAFISH® assay. #### e. Analytical reactivity: The analytical reactivity study included 21 E. coli of which all were correctly identified. #### f. Analytical specificity: {4} The analytical specificity of the $E.$ coli PNA FISH® was determined by BLAST search and sequence alignments and experimentally by testing well characterized laboratory and reference strains comprising of 21 $E.$ coli, 67 Gram-negative organisms, 11 Gram-positive organisms and 5 yeast strains. Escherichia coli provided expected positive results. However, Escherichia albertii, Escherichia fergusonii, and Shigella species (serogroup A, B, C, or D) also provided positive results with $E.$ coli PNA FISH®. A limitation statement has been included to address the cross reactivity issue. # g. Interference Study: A study consisting of 15 Gram negative rods were tested on BACTEC Plus blood culture bottles for the interference of resin. No interferences were observed. Current peer-reviewed publications indicate that sodium polyanetholesulfonate (SPS) or charcoal does not cause interference with PNA FISH assays. # g. Assay cut-off: Not applicable # 2. Comparison studies: a. Method comparison of device to conventional methods, as the reference method: A Clinical Study was conducted at five sites in the United States and Europe, directly on blood culture bottles containing Gram negative rods (GNR). A total of 122 GNR-positive blood bottles, from two commercial continuously monitoring blood culture systems (BacT/Alert and BACTEC) were included in the study. Performance results compared to routine identifications, based on conventional methods following subculture, are summarized below. Table 1 BACTEC - verses Vitek2 (bioMérieux) | Study Site C+E | Sensitivity | Specificity | Total | | --- | --- | --- | --- | | Positive | 27 | 0 | | | Negative | 1 | 64 | | | | 96.4% (27/28) | 100% (64/64) | 92 | | | 95% CI (81.7-99.9) | 95% CI (95.4-100) | | {5} Table 2 BACTEC - verses Chromagar and Indol Spot: | Study Site D | Sensitivity | Specificity | Total | | --- | --- | --- | --- | | Positive | 56 | 1 | | | Negative | 0 | 50 | | | | 100% (56/56) | 98.0% (50/51) | 107 | | | 95% CI (94.8-100) | 95% CI (89.6-99.9) | | Table 3 BacT/Alert SA/SN - verses API ID 32E (bioMérieux) | Study Site A | Sensitivity | Specificity | Total | | --- | --- | --- | --- | | Positive | 8 | 0 | | | Negative | 0 | 9 | | | | 100% (8/8) | 100% (9/9) | 17 | | | 95% CI (68.8-100) | 95% CI (71.7-100) | | Table 4 BacT/Alert SA/SN - verses Vitek2 (bioMérieux) | Study Site B | Sensitivity | Specificity | Total | | --- | --- | --- | --- | | Positive | 5 | 1 | | | Negative | 0 | 8 | | | | 100% (5/5) | 88.9% (8/9) | 14 | | | 95% CI (54.9-100) | 95% CI (51.8-99.7) | | Table 5 BacT/Alert FA/FN - API ID32E (bioMérieux) | Study Site A | Sensitivity | Specificity | Total | | --- | --- | --- | --- | | Positive | 6 | 0 | | | Negative | 0 | 12 | | | | 100% (6/6) | 100% (12/12) | 18 | | | 95% CI (60.7-100) | 95% CI (77.9-100) | | Table 6 BacT/Alert FA - Vitek2 (bioMérieux) | Study Site B | Sensitivity | Specificity | Total | | --- | --- | --- | --- | | Positive | 17 | 0 | | | Negative | 0 | 10 | | | | 100% (17/17) | 100% (10/10) | 27 | | | 95% CI (83.8-100) | 95% CI (74.1-100) | | b. Matrix comparison: Not applicable {6} 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Positive: green fluorescence Negative: non fluorescence N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 7
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...