The Sensititre® YeastOne® Susceptibility System is an in vitro diagnostic product for clinical susceptibility testing of Candida spp. This 510(k) is for voriconazole in the dilution range of 0.008 - 16μg/ml for testing Candida spp. on the Sensititre® YeastOne® Susceptibility system. The approved primary "Indications for Use" and clinical significance of voriconazole is for: Candida albicans Candida krusei Candida parapsilosis Candida tropicalis
Device Story
Sensititre® YeastOne® Susceptibility system is a micro-version of broth dilution susceptibility test. System uses microdilution plates containing serially diluted antifungal agents and a colorimetric growth-indicating compound. After inoculation with yeast samples, plates are sealed and incubated at 35°C for 24 hours. Growth is determined by colorimetric change indicating REDOX state, rather than turbidity. System is used in clinical laboratories; manual reading of color change is performed by laboratory personnel. Output is quantitative MIC value. Results assist clinicians in determining antifungal susceptibility for Candida species, guiding therapeutic decisions.
Clinical Evidence
No clinical data provided. The submission relies on bench testing to demonstrate performance characteristics for the susceptibility testing of Candida spp. against voriconazole.
Technological Characteristics
Microbroth dilution susceptibility testing system. Consists of multi-well plates containing dehydrated antimicrobial agents (voriconazole). Operates via colorimetric or visual growth detection in liquid media. Standardized for clinical laboratory use.
Indications for Use
Indicated for testing C. albicans, C. krusei, C. parapsilosis, and C. tropicalis. Prescription use only. Alternative procedure required for C. glabrata. Resistance detection for Voriconazole is unknown.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
Related Devices
K211539 — Sensititre YeastOne Susceptibility System with Voriconazole in the dilution range of 0.008 - 8 ug/mL · Thermo Fisher Scientific · Dec 15, 2021
K090968 — SENSITITRE YEASTONE SUSCEPTIBILITY PLATES · Trek Diagnostic Systems, Inc. · Jun 4, 2009
K133038 — SENSITITRE YEASTONE SUSCEPTIBILITY PLATES · Trek Diagnostic Systems · Jan 31, 2014
K231433 — The Sensititre YeastOne Susceptibility System with Rezafungin in the dilution range of 0.008-8ug/mL · Thermo Fisher Scientific · Aug 31, 2023
K243738 — The Sensititre YeastOne Susceptibility System with Micafungin in the dilution range of 0.008-16 µg/mL · Thermo Fisher Scientific · Feb 27, 2025
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k081063
B. Purpose for Submission:
Addition of Voriconazole to the Sensititre® YeastOne® Susceptibility System
C. Measurand:
Voriconazole 0.008 – 16 µg/mL
D. Type of Test:
Quantitative Antimicrobial Susceptibility Test (AST) growth based
E. Applicant:
TREK Diagnostic Systems, Inc.
F. Proprietary and Established Names:
Sensititre® YeastOne® Susceptibility plates
G. Regulatory Information:
1. Regulation section:
866.1640 Antimicrobial Susceptibility Test Powder
2. Classification:
Class II
3. Product code:
NGZ Susceptibility Testing - antifungal
4. Panel:
83 Microbiology
{1}
H. Intended Use:
1. Intended use(s):
Sensititre® YeastOne® Susceptibility plates are designed for *in vitro* use in determining quantitative antifungal susceptibilities (MIC) of *Candida* species.
2. Indication(s) for use:
This 510k application is intended for the addition of Voriconazole in the dilution range of 0.008 – 16 µg/mL on the Sensititre® YeastOne® Susceptibility plates for testing *C. albicans*, *C. krusei*, *C. parapsilosis*, and *C. tropicalis*.
3. Special conditions for use statement(s):
Prescription use only
An alternative procedure should be performed if voriconazole with *C. glabrata* is identified.
The ability of the Sensititre® YeastOne® to detect resistance to Voriconazole is unknown because resistant strains were not available at the time of comparative testing.
4. Special instrument requirements:
Manual Readings only
I. Device Description:
The Sensititre® YeastOne® Susceptibility system is a micro-version of the broth dilution susceptibility test. Various antifungal agents are serially diluted to concentrations bridging the range of the clinical interest in autoclaved diluent, which contains a colorimetric growth indicating compound. Each microdilution plate is individually packaged in foil. After inoculation, plates are sealed with an adhesive seal, incubated at 35°C for 24 hours and examined for growth.
J. Substantial Equivalence Information:
1. Predicate device name(s):
VITEK® AST-YS Fluconazole
2. Predicate 510(k) number(s):
k061945
{2}
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Susceptibility testing for colonies of Candida | Susceptibility testing for colonies of Candida |
| Incubation | 35°C | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Technology | Broth micro dilution – growth based with growth colorimetric indicator for manual readings | Automated growth based Broth micro dilution -detection using growth based with attenuation of light colorimetric growth measured by optical indicator for manual scanner. |
| Format | Micro tray with dried antifungals
Medium | VITEK® 2 AST test card with dried antifungals |
| Medium | Sensititre® yeast susceptibility inoculum broth | VITEK® 2 Yeast Base broth |
| Antifungal | voriconazole | fluconazole |
| Time to Results | 24 hours | 10 to 36 hours |
K. Standard/Guidance Document Referenced (if applicable):
"Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test Systems; Guidance for Industry and FDA"; CLSI M27 S3 "Reference Method for Broth Dilution Antifungal Susceptibility Testing of Yeasts; Approved Standard - Third Edition."
L. Test Principle:
The Sensititre® YeastOne® panel measures growth by colorimetric determination of REDOX state produced by the test organism rather than measuring the turbidity of the test medium. In the Sensititre panel as in most broth dilution tests the difference between wells in which growth occurs and those in which growth is inhibited is distinctive. However, the Sensititre panel produces an easily distinguished color change when light growth occurs – a circumstance which is often difficult to interpret correctly using turbidity readings.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Reproducibility was conducted at three sites on ten yeast isolates with testing performed over three days in triplicate. The testing was performed using both
{3}
manual and automated inoculation methods. The reproducibility was >95% with ±1 one dilution observations.
b. Linearity/assay reportable range:
Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Quality Control (QC) testing was performed on each day of clinical testing on the QC isolates, C. parapsilosis ATCC 22019, and C. krusei ATCC 6258, as recommended in the CLSI standard M27. The following table represents the frequency of the results in both the reference method and the Sensititre® YeastOne® Susceptibility plates and the acceptable range. Quality Control was also performed at three sites using both manual (electronic pipette) and autoinoculation methods.
Quality Control Table
| ORGANISM | Conc ug/mL | Automated Inoculation Method | | | Manual Inoculation Method | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Sensititre® YeastOne® | Reference | | Sensititre® YeastOne® | Reference | |
| | | | 24h | 48h | | 24h | 48h |
| | | | | | | | |
| C. parapsilosis ATCC 22019
Expected Range :
0.016 – 0.12 μg/mL | 0.016 | | | | | 1 | |
| | 0.03 | 4 | 30 | 4 | 4 | 33 | 6 |
| | 0.06 | 47 | 31 | 55 | 27 | 9 | 22 |
| | 0.12 | 10 | | 2 | 12 | | 15 |
| | | | | | | | |
| C. krusei ATCC 6258
Expected Range :
0.06 – 0.25 μg/mL | 0.06 | 0 | 1 | | | 1 | |
| | 0.12 | 19 | 39 | 27 | 19 | 24 | |
| | 0.25 | 42 | 21 | 33 | 23 | 16 | 11 |
| | 0.5 | | | 1 | 1 | 2 | 32 |
| | | | | | | | |
Nephelometer was used at each site to standardize the inoculum. Colony counts from QC ATCC source were performed using direct inoculum method.
d. Detection limit:
Not Applicable
e. Analytical specificity:
Not Applicable
f. Assay cut-off:
{4}
Not Applicable
2. Comparison studies:
a. Method comparison with predicate device:
Performance was established on the Sensititre® YeastOne® Susceptibility System for Candida spp. at three clinical sites. The CLSI reference method as described in the CLSI document M27 "Method for Broth Dilution Antifungal Susceptibility Testing of Yeasts" was compared to the Sensititre® YeastOne® Susceptibility System. A total of 367 isolates of Candida species were tested on the Sensititre® YeastOne® using both manual and automated inoculation methods. The no growth rate is <10%.
EA was calculated when the results for the reference method and the Sensititre® YeastOne® were within +/- two doubling dilutions of the antifungal drug. The following tables present the performance of the Sensititre® YeastOne® Susceptibility System when read at 24 hours using both manual and automated inoculation methods as compared to the reference method when read at 48 hours.
Summary Table using the AUTOINOCULATOR (24hour vs 48hour)
| | EA Tot | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | CA N | CA % | #R | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Clinical | 174 | 171 | 98.3 | 130 | 127 | 97.7 | 171 | 98.2 | 6 | 3 | 0 | 0 |
| Challenge | 193 | 184 | 95.3 | 172 | 162 | 94.1 | 176 | 91.2 | 2 | 17 | 0 | 0 |
| Combined | 367 | 355 | 9637 | 302 | 289 | 35.7 | 347 | 94.5 | 8 | 20 | 0 | 0 |
*There were 14 min errors with C. glabrata with the autoinoculation method with a min error rate of 20.2%. Because of the high number of errors, a limitation has been included in the technical insert to perform an alternate method for testing when this drug/bug combination is isolated.
Summary Table using MANUAL Dilution (24hour vs 48hour)
| | EA Tot | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | CA N | CA % | #R | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Clinical | 268 | 261 | 97.3 | 211 | 204 | 96.7 | 256 | 95.5 | 9 | 12 | 0 | 0 |
| Challenge | 101 | 98 | 97.0 | 85 | 82 | 96.4 | 92 | 91.1 | 2 | 9 | 0 | 0 |
| Combined | 369 | 359 | 97.2 | 296 | 286 | 96.6 | 348 | 94.3 | 11 | 21 | 0 | 0 |
*There were 10 min errors with C. glabrata with the manual method of inoculation with a min error rate of 19.6%. Because of the high number of errors, a limitation has been included in the technical insert to perform an alternate method for testing when this drug/bug combination is isolated.
EA - Essential Agreement
CA - Category Agreement
R-resistant isolates
maj-major discrepancies
vmj-very major discrepancies
min- minor discrepancies
Category agreement (CA) is when the Sensititre® panel result interpretation agrees exactly with the reference panel result interpretation. Evaluable EA is when the
{5}
MIC result is on scale for both the Sensititre® and the reference and have on-scale EA.
b. Matrix comparison:
Not Applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable
4. Clinical cut-off:
Not Applicable
5. Expected values/Reference range:
Candida species <1 (S), 2 (I), ≥4 (R)
N. Proposed Labeling:
The expected value range, interpretive criteria and QC are included in the package insert. The labeling is sufficient and satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
6
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.