K131275 · Siemens Helathcare Diagnostics, Inc. · LTT · Nov 8, 2013 · Microbiology
Device Facts
Record ID
K131275
Device Name
MICROSCAN DRIED GRAM-POSTIVE MIC/COMBO PANELS
Applicant
Siemens Helathcare Diagnostics, Inc.
Product Code
LTT · Microbiology
Decision Date
Nov 8, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1640
Device Class
Class 2
Intended Use
The MicroScan® Dried Gram-Positive MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-positive bacteria. After inoculation, panels are incubated for 16 - 20 hours at 35°C +/- 1°C in a non-CO2 incubator, and read either visually or with MicroScan instrumentation, according to the Package Insert. This particular submission is for the addition of the antimicrobial Tigecycline at concentrations of 0.12 - 8 mcg/ml to the test panel. MicroScan® Dried Gram-Positive Tigecycline is a qualitative test. Tigecycline has been shown to be active against the organisms listed below according to the FDA label for the antimicrobial. Active in vitro and in clinical infections: Enterococcus faecalis (vancomycin-susceptible isolates) Staphylococcus aureus (methicillin-susceptible and -resistant isolates) Active in vitro but clinical significance unknown: Staphylococcus epidermidis (methicillin-susceptible and -resistant isolates)
Device Story
MicroScan Dried Gram-Positive MIC/Combo Panels are miniaturized, dehydrated broth microdilution susceptibility tests. Panels contain ten doubling dilutions of Tigecycline (0.015–8 mcg/mL). Inoculated with standardized organism suspensions; rehydrated with water; incubated 16–20 hours at 35°C in non-CO2 incubator. Read manually (MicroSCAN Microdilution Viewer) or via automated instrumentation (autoSCAN-4 or WalkAway). Tigecycline inhibits protein synthesis by preventing t-RNA binding to ribosomes. Used in clinical microbiology laboratories by technicians/microbiologists. Output is MIC value or qualitative susceptibility (Susceptible). Results assist clinicians in selecting appropriate antibiotic therapy for bacterial infections. Turbidity inoculation method is preferred for accuracy; Prompt inoculation system is an alternative but carries limitations for staphylococci.
Clinical Evidence
Bench testing only. Performance evaluated using 474 clinical and challenge isolates (S. aureus, E. faecalis, S. epidermidis). Comparison against CLSI broth microdilution reference method. Overall categorical agreement (CA) for turbidity inoculation/manual read was 99.4%. Essential agreement (EA) was 78.0%. Reproducibility study showed 100% agreement within ±1 dilution across all sites and methods. Potential major/very major errors noted with Prompt inoculation method for S. aureus; turbidity method identified as preferred.
Technological Characteristics
Dried microdilution broth panels; miniaturized broth dilution method. Rehydrated with organism suspension. Incubation 16-20 hours at 35°C. Compatible with autoSCAN-4 and WalkAway instrumentation. Qualitative test for Tigecycline (0.12-8 mcg/ml).
Indications for Use
Indicated for determining quantitative and/or qualitative antimicrobial susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-positive bacteria, specifically Enterococcus faecalis (vancomycin-susceptible), Staphylococcus aureus (methicillin-susceptible and -resistant), and Staphylococcus epidermidis (methicillin-susceptible and -resistant).
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
{0}------------------------------------------------
K/3/275
# 510(k) Summary
**NOV 08 2013**
### 510(k) Submission Information:
| Device Manufacturer: | Siemens Healthcare Diagnostics |
|----------------------|-------------------------------------------------------------------------------------|
| Contact name: | Elisabeth Warriner, Regulatory Technical Specialist |
| Phone: | 916-374-3244 |
| Fax: | 916-372-6418 |
| Date prepared: | August 27, 2013 |
| Product Name: | Microdilution Minimum Inhibitory Concentration (MIC) Panels |
| Trade Name: | MicroScan® Dried Gram-Positive MIC/Combo Panels with Tigecycline (0.12-8<br>mcg/ml) |
| Intended Use: | To determine antimicrobial agent susceptibility |
| Classification: | Class II |
| Product Code: | LTT |
| 510(k) Notification: | New antimicrobial - Tigecycline |
| Predicate device: | MicroScan Dried Gram-Positive MIC/Combo Panels - Linezolid (K003619) |
#### 510(k) Summary:
MicroScan Dried Gram-Positive MIC/Combo Panels are designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-positive bacteria.
The antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test that have been diluted in broth and dehydrated. Various antimicrobial agents are diluted in broth to concentrations bridging the range of clinical interest. Panels are rehydrated with water after inoculation with a standardized suspension of the organism. After incubation in a non-CO2 incubator for 16-20 hours, the minimum inhibitory concentration (MIC) for the test organism is read by determining the lowest antimicrobial concentration showing inhibition of growth.
The proposed MicroScan Dried Gram-Positive MIC/Combo Panel with Tigecvcline demonstrated substantially equivalent performance when compared with an CLSI frozen Reference Panel, as defined in the FDA document "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA", dated August 28, 2009. The Premarket Notification (510|k|) presents data in support of the MicroScan Dried Gram-Positive MIC/Combo Panel with Tigecycline. MicroScan Dried Gram-Positive Tigecycline is a qualitative test.
The external evaluation was conducted with fresh and stock Efficacy isolates and stock Challenge strains. The external evaluations were designed to confirm the acceptability of the proposed Dried Gram-Positive Panel by comparing its performance with a CLSI frozen Reference panel. Challenge strains were compared to Expected Results determined prior to the evaluation. The Dried Gram-Positive Panel demonstrated acceptable performance with an overall Categorical Agreement of 99.4% for Tigecycline when compared with the frozen Reference panel.
Inoculum and instrument reproducibility testing demonstrated acceptable reproducibility and precision with Tigecycline, regardless of which inoculum method (i.e., Turbidity and Prompt™). or instrument (autoSCAN®-4 and WalkAway®) was used.
Quality Control testing demonstrated acceptable results for Tigecycline.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for the Department of Health & Human Services (DHHS). The logo consists of a stylized caduceus symbol, which features a staff with a serpent entwined around it. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Homoshim Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
SIEMENS HEALTHCARE DIAGNOSTICS, INC. ELISABETH WARRINER REGULATORY TECHNICAL SPECIALIST 1584 ENTERPRISE BLVD. WEST SACRAMENTO CA 95691
November 8. 2013
Re: K131275
Trade/Device Name: MicroScan Dried Gram-Positive MIC/Combo Panels with Tigecycline (0.12 - 8mcg/ml) Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial susceptibility test powder Regulatory Class: II Product Code: LTT, JWY, LRG, LTW Dated: September 27, 2013 Received: September 30, 2013
Dear Ms. Warriner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2-Ms. Warriner
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometries/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Uwe Scherf -Sior
Sally Hojvat, M.Sc., PhD. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known):__K131275
Device Name: _ MicroScan® Dried Gram-Positive MIC/Combo Panels with Tigecycline (0.12 - 8 mcg/ml)
Indications For Use:
The MicroScan® Dried Gram-Positive MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-positive bacteria. After inoculation, panels are incubated for 16 - 20 hours at 35°C +/- 1°C in a non-CO2 incubator, and read either visually or with MicroScan instrumentation, according to the Package Insert.
This particular submission is for the addition of the antimicrobial Tigecycline at concentrations of 0.12 - 8 mcg/ml to the test panel. MicroScan® Dried Gram-Positive Tigecycline is a qualitative test.
Tigecycline has been shown to be active against the organisms listed below according to the FDA label for the antimicrobial.
Active in vitro and in clinical infections:
Enterococcus faecalis (vancomycin-susceptible isolates) Staphylococcus aureus (methicillin-susceptible and -resistant isolates)
Active in vitro but clinical significance unknown: Staphylococcus epidermidis (methicillin-susceptible and -resistant isolates)
Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH; Office of in Vitro Diagnostics and Radiological Health (OIR)
Image /page/3/Picture/15 description: The image shows the name "Ribhi Sha" in bold font on the top left. Below the name, the date "2013.11.07" is printed, followed by the time "08:12:19" and the timezone offset "-05'00'". There is some decorative graphic in the middle of the image.
Page 1 of _1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches