MICROSCAN MICROSTREP PLUS PANEL CEFTRIAXONE (0.015-8MCG/ML)
K063064 · Dade Behring, Inc. · LRG · Oct 26, 2006 · Microbiology
Device Facts
Record ID
K063064
Device Name
MICROSCAN MICROSTREP PLUS PANEL CEFTRIAXONE (0.015-8MCG/ML)
Applicant
Dade Behring, Inc.
Product Code
LRG · Microbiology
Decision Date
Oct 26, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1640
Device Class
Class 2
Intended Use
The MicroScan MICroSTREP plus® Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of aerobic streptococci, including Streptococcus pneumoniae. After inoculation, panels are incubated for 20 -- 24 hours at 35°C +/- 1°C in a non-CO2 incubator, and read visually according to the Package Insert. Additionally, the panels may be incubated in and read by a MicroScan® WalkAway instrument.
Device Story
MicroScan MICroSTREP plus® Panel is a miniaturized broth dilution susceptibility test system. It uses panels containing various antimicrobial concentrations to determine the minimum inhibitory concentration (MIC) of bacteria. Input: bacterial colonies grown on solid media, inoculated into Mueller-Hinton broth supplemented with 2-5% lysed horse blood and 50 mM HEPES. Operation: panels are incubated for 20-24 hours at 35°C in a non-CO2 incubator. Output: MIC value determined either by manual visual inspection or by the MicroScan® WalkAway instrument. Used in clinical laboratories by trained personnel. The instrument-read method automates the detection of growth inhibition, providing quantitative susceptibility results to assist clinicians in selecting appropriate antibiotic therapy for streptococcal infections.
Clinical Evidence
Bench testing only. Reproducibility study (10 isolates, 3 sites, 3 days) showed >95% reproducibility between manual and automated methods. Comparison study using 70 CDC challenge streptococcal isolates showed 98.6% Essential Agreement (EA) between manual and automated reading methods. No clinical efficacy testing was required as it was previously established for the manual method.
Technological Characteristics
Miniaturized broth dilution panel. Reagents: Mueller-Hinton broth, 2-5% lysed horse blood, 50 mM HEPES. Energy source: non-CO2 incubator and MicroScan® WalkAway instrument for automated optical reading. Connectivity: WalkAway instrument integration. Software: automated growth inhibition detection algorithm.
Indications for Use
Indicated for determining antimicrobial susceptibility of aerobic streptococci, including Streptococcus pneumoniae, Streptococcus pyogenes, and viridans group streptococci, to Ceftriaxone (0.015 - 8 mcg/ml).
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
K020686 — MICROSCAN MICROSTREP PLUS PANEL NEW ANTIMICROBIAL - TRIMETHOPRIM/SULFAMETHOXAZOLE · Dade Microscan, Inc. · Apr 30, 2002
K063099 — MICROSCAN MICROSTREP PLUS PANEL CEFEPIME (O.015 - 8 MCG/ML) · Dade Behring, Inc. · Nov 1, 2006
K020820 — MICROSCAN MICROSTEP PLUS PANEL NEW ANTIMICROBIAL - CEFUROXIME · Dade Behring, Inc. · May 9, 2002
K062293 — MICROSCAN MICROSTREP PLUS PANEL CHLORAMPHENICOL (1 TO 32 MCG/ML) · Dade Behring, Inc. · Aug 24, 2006
K063053 — MICROSCAN MICROSTREP PLUS PANEL CEFUROXIME (0.12-8 MCG/ML) · Dade Behring, Inc. · Oct 26, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
Kolo30lo 4
OCT 26 2006
## 510(k) Summary Information:
| Device Manufacturer: | Dade Behring Inc. |
|----------------------|-----------------------------------------------------------------------|
| Contact name: | Maureen Mende, Group Manager, Regulatory Affairs |
| Phone/Fax: | 916-374-3174/916-374-3144 |
| Date prepared: | October 3, 2006 |
| Product Name: | Microdilution Minimum Inhibitory Concentration (MIC) Panels |
| Trade Name: | MicroScan MICroSTREP plus® Panel |
| Intended Use: | To determine bacterial susceptibility to Ceftriaxone |
| Indication for Use: | For determining antimicrobic susceptibility with aerobic streptococci |
| | including Streptococcus pneumoniae |
| Predicate device: | MicroScan MICroSTREP plus® Panel (K021110) |
## 510(k) Summary:
The MicroScan MICroSTREP plus® Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of aerobic streptococci, including Streptococcus pneumoniae. After inoculation, panels are incubated for 20 - 24 hours at 35°C +/-1°C in a non-CO2 incubator, and read according to the Package Insert.
The antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test. Various antimicrobial agents are diluted in water, buffer or minute concentrations of broth to concentrations bridging the range of clinical interest. Panels are rehydrated with 115 µl Mueller-Hinton broth supplemented with 2-5% lysed horse blood (LHB) and buffered with 50 mM HEPES, after inoculation of the broth with a standardized suspension of the organism in saline. After incubation in a non-CO2 incubator for 20-24 hours, the minimum inhibitory concentration (MIC) for the test organism is manually read by observing the lowest antimicrobial concentration showing inhibition of growth. Additionally, the panels may be incubated in and read by a MicroScan WalkAway instrument.
The proposed instrument read method for the MicroScan MICroSTREP plus® Panel demonstrated substantially equivalent performance with streptococcal isolates when compared with an expected result generated on a CLSI frozen Reference Panel, as defined in the FDA document "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA", dated February 5, 2003.
This Premarket Notification (510[k]) presents data in support of reading the MICroSTREP plus Panel with Ceftriaxone on the MicroScan® WalkAway instrument.
The evaluation was conducted with stock and CDC Challenge strains. The evaluations were designed to confirm the acceptability of the proposed instrument read method for the MICroSTREP plus Panel by comparing its performance with Expected Results determined before the evaluation. The MICroSTREP plus® Panel demonstrated acceptable performance with an overall Essential Agreement of 99% for Ceftriaxone instrument read results compared with the Expected Result.
{1}------------------------------------------------
Instrument reproducibility testing demonstrated acceptable reproducibility and precision with Ceftriaxone and the WalkAway® instrument.
Quality Control testing demonstrated acceptable results for Ceftriaxone.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
Re:
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Maureen Mende Group Manager, Regulatory Affairs Dade Behring, Inc. 1584 Enterprise Boulevard West Sacramento, CA 95691-9972
OCT 2 6 2006
k063064 Trade/Device Name: MicroScan MICroSTREP plus® Panel Cefriaxone (0.015 - 8 mcg/ml) Regulation Number: 21 CFR § 866.1640 Regulation Name: Antimicrobial susceptibility test powder Regulatory Class: II Product Code: LRG, LTT Dated: October 3, 2006 Received: October 6, 2005
Dear Ms. Mende:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcemer.ts concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Sally antrom
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indication for Use Statement
510(k) No.:
K063064 (To be assigned by FDA)
Device Name:
MicroScan MICroSTREP plus® Panel Ceftriaxone (0.015 - 8mcg/ml)
To determine bacterial antimicrobial agent susceptibility Intended Use
Indications for Use:
The MicroScan MICroSTREP plus® Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of aerobic streptococci, including Streptococcus pneumoniae. After inoculation, panels are incubated for 20 -- 24 hours at 35°C +/- 1°C in a non-CO2 incubator, and read visually according to the Package Insert. Additionally, the panels may be incubated in and read by a MicroScan® WalkAway instrument.
This particular submission is for the addition of instrument read capability of the antimicrobial Ceftriaxone, at concentrations of 0.015 to 8 mcg/ml on the MicroScan MICroSTREP plus® Panel.
The organisms which may be used for Ceftriaxone susceptibility testing in this panel are:
Streptococcus pneumoniae Streptococcus pyogenes viridans group streptococci
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
| Page 1 of 1 | Division Sign-Off |
|-------------|-------------------|
|-------------|-------------------|
Office of In Vitro Diagnostic Device Evaluation and Safety
1996) K06306Y
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.