VITEK 2 GRAM NEGATIVE DOXYCYCLINE

K121546 · bioMerieux, Inc. · LON · Nov 20, 2012 · Microbiology

Device Facts

Record IDK121546
Device NameVITEK 2 GRAM NEGATIVE DOXYCYCLINE
ApplicantbioMerieux, Inc.
Product CodeLON · Microbiology
Decision DateNov 20, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2

Intended Use

VITEK® 2 Gram Negative Doxycycline is designed for antimicrobial susceptibility testing of Gram negative bacilli. VITEK® 2 Gram Negative Doxycycline is a quantitative test intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. Doxycycline has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections Acinetobacter species, Enterobacter aerogines, Escherichia coli, Klebsiella species, and Shigella species. The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast.

Device Story

VITEK® 2 AST card; miniaturized, automated version of doubling dilution technique. Input: bacterial isolate suspension (standardized to 0.5 McFarland). Process: card wells contain premeasured antibiotics in culture media; system automatically fills, seals, and incubates (35.5°C); optical scanning every 15 minutes monitors growth. Output: MIC value and result interpretation. Used in clinical laboratories; operated by technicians. Output aids clinicians in determining in vitro susceptibility to antimicrobial agents, informing antibiotic treatment decisions.

Clinical Evidence

Performance evaluated using 409 clinical isolates and 98 challenge isolates. Compared against broth microdilution reference method. Combined performance for indicated species showed 97.3% Essential Agreement (EA) and 96.5% Category Agreement (CA). Combined performance for all species showed 97.6% EA and 95.1% CA. Reproducibility >95% across three sites. No clinical sensitivity/specificity data required for AST systems.

Technological Characteristics

Miniaturized broth microdilution card containing antimicrobial medium. Automated incubation and optical growth monitoring. Compatible with VITEK 2 and VITEK 2 Compact systems. Quantitative MIC determination. No specific materials or software architecture details provided.

Indications for Use

Indicated for antimicrobial susceptibility testing of isolated colonies of clinically significant aerobic Gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp., S. pneumoniae, and clinically significant yeast. Specifically for Doxycycline, indicated for Acinetobacter species, Enterobacter aerogenes, Escherichia coli, Klebsiella species, and Shigella species.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K12/546 Image /page/0/Picture/1 description: The image shows the logo for bioMerieux. The logo consists of the word "BIOMERIEUX" in a sans-serif font, with a stylized graphic above it. The graphic is a circle that is divided into two halves, with one half filled with horizontal lines and the other half left blank. A curved line runs vertically through the center of the circle and extends above and below it. NOV 2 0 2012 1 ## 510(k) SUMMARY ### VITEK® 2 Gram Negative Doxycycline #### 510(k) Submission Information: | Submitter's Name: | bioMérieux, Inc. | |-------------------|----------------------------------------| | Address: | 595 Anglum Road<br>Hazelwood, MO 63042 | | Contact Person: | Jennifer Chia-Hsuan Lin | Regulatory Affairs Specialist Phone Number: Fax Number: 314-731-8689 314 -731-8639 May 24, 2012 #### B. Device Name: Formal/Trade Name: Date of Preparation: Classification Name: 21 CFR 866.1645 Fully Automated Short-Term Incubation Cycle Antimicrobial Common Name: VITEK® 2 AST-GN Doxycycline Susceptibility System Product Code LON VITEK® 2 Gram Negative Doxycycline C. Predicate Device: VITEK® 2 Gram Negative Imipenem (K103752) #### D. 510(k) Summary: VITEK® 2 Gram Negative Doxycycline is designed for antimicrobial susceptibility testing of Gramnegative bacilli. VITEK 2 GN Doxycycline is a quantitative test intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. Doxycycline has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections Acinetobacter species Enterobacter aerogenes Escherichia coli Klebsiella species Shigella species The VITEK 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK 2 and VITEK 2 Compact Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Streptococcus spp., S. pneumoniae, and clinically significant yeast. ## bioMérieux, Inc. Phone: 314/731-8500 800/638-4835 595 Anglum Road, Hazelwood, Missouri 63042-2320, USA Fax: 314/731-8700 http://www.biomerieux-usa.com {1}------------------------------------------------ The antimicrobial presented in VITEK® 2 AST-GN Cards is in concentrations equivalent by efficacy to standard method concentrations in ug/mL. The VITEK® 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology. The isolate to be tested is diluted to a standardized concentration in 0.45 - 0.50% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK® 2 automatically fills, seals and places the card into the incubator/reader. The VITEK® 2 Compact has a manual filling and sealing operation. The VITEK® 2 monitors the growth of each well in the card over a defined period of time (up to 18 hours). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antimicrobial contained on the card. VITEK® 2 GN Doxycycline demonstrated substantially equivalent performance when compared with the CLSI broth microdilution reference method, as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA. Issued August 28, 2009. The Premarket Notification (510[k]) presents data in support of VITEK® 2 GN Doxycycline. An external evaluation was conducted with fresh and stock clinical isolates, as well as a set of challenge strains. The external evaluations were designed to confirm the acceptability of VITEK® 2 GN Doxycycline by comparing its performance with the CLSI broth microditution reference method incubated at 24 hrs. The data is representative of performance on both the VITEK® 2 and VITEK® 2 Compact instrument platforms. VITEK® 2 GN Doxycycline demonstrated acceptable performance of 97.6% overall Essential Agreement and 95.1% overall Category Agreement with the reference method. Reproducibility and Quality Control demonstrated acceptable results using both the VITEK 2 and VITEK 2 Compact instrument systems. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three curved shapes that resemble a human figure. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 bioMerieux. Inc. C/O Jennifer Chia-Hsuan Lin. Regulatory Affairs Specialist 595 Anglum Road Hazelwood MO 63042 NOV 2 0 2012 Re: K121546 VITEK®2 Gram Negative Doxycycline Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-term Incubation Cycle Antimicrobial Susceptibility System Regulatory Class: Class II Product Code: LON Dated: October 19, 2012 Received: October 22, 2012 Dear Ms. Lin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, # Sally A. Hojvat Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known):_ K12 1846 / S00 1 Device Name: VITEK® 2 Gram Negative Doxycycline (<0.5 ->16 ug/mL) Indications For Use: VITEK® 2 Gram Negative Doxycline is designed for antimicrobial susceptibility testing of Gramnegative bacilli. VITEK 2 Gram Negative Doxycycline is a quantitative test intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. Doxycycline has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections Acinetobacter species Enterobacter aerogenes Escherichia coli Klebsiella species Shigella species The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 System for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp., S. pneumoniae, and clinically significant yeast. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Jayart Division Sign-Off Division Sign-C Office of In Vitro Diagnostic Device Evaluation and Safety 510(K) KIZ 1 244 Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...