K080107 · bioMerieux, Inc. · LON · Feb 28, 2008 · Microbiology
Device Facts
Record ID
K080107
Device Name
VITEK 2 GRAM NEGATIVE PIPERACILLIN/TAZOBACTAM
Applicant
bioMerieux, Inc.
Product Code
LON · Microbiology
Decision Date
Feb 28, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1645
Device Class
Class 2
Intended Use
VITEK® 2 Gram Negative Piperacillin/tazobactam is designed for antimicrobial susceptibility testing of gram-negative organisms and is intended for use with the VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK 2 Gram Negative Piperacillin/tazobactam is a quantitative test. Piperacillin/ tazobactam has been shown to be active against most strains of the organisms listed below according to the FDA label for the antimicrobial. Active in vitro and in clinical infections Klebsiella pneumoniae Acinetobacter baumanii Escherichia coli Pseudomonas aeruginosa Active in vitro but their clinical significance is uknown: Proteus vulgaris Providencia rettgeri Providencia stuartii Citrobacter koseri Morqanella morganii Proteus mirabilis Salmonella enterica Serratia marcescens The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 and VITEK® 2 Compact Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, and S. pneumoniae.
Device Story
VITEK® 2 AST system performs automated antimicrobial susceptibility testing. Input: standardized organism suspension (McFarland 0.5) inoculated into 64-microwell test cards. Operation: instrument vacuum-fills cards; seals; incubates at 35.5°C; performs optical scanning every 15 minutes. Algorithm: quantitative growth-based detection using optical light monitoring; determines susceptibility based on predetermined growth thresholds. Output: automated quantitative or qualitative susceptibility report. Used in clinical laboratories by technicians/lab staff. Benefits: rapid, standardized determination of bacterial susceptibility to Piperacillin/tazobactam, aiding clinical antibiotic selection.
Clinical Evidence
Bench testing only. Performance evaluated using 483 clinical isolates (269 fresh, 214 stock) and 79 challenge isolates. Comparison against broth microdilution reference method. Combined results showed 96.6% Essential Agreement and 96.6% Category Agreement. Reproducibility >95% across 3 sites. Quality control testing performed with E. coli and P. aeruginosa strains, meeting expected ranges >95% of the time.
Technological Characteristics
VITEK® 2 AST card with 64 microwells containing premeasured antibiotic and culture medium. Optical sensing principle; 35.5°C incubation. Automated vacuum filling and sealing. Connectivity: VITEK® 2 and VITEK® 2 Compact Systems. Software-based growth detection algorithm.
Indications for Use
Indicated for in vitro susceptibility testing of Acinetobacter baumanii, Escherichia coli, Klebsiella pneumoniae, Pseudomonas aeruginosa, Citrobacter koseri, Morganella morganii, Proteus mirabilis, Proteus vulgaris, Providencia stuartii, Providencia rettgeri, Salmonella enterica, and Serratia marcescens to Piperacillin/tazobactam (concentrations 4/4, 16/4, 32/4, 64/4; calling range ≤4 - ≥128 μg/mL) in clinical laboratories.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
K161510 — VTEK 2 AST-GN Ceftolozane/Tazobactam (<=0.25 - >= 32 mcg/mL) · bioMerieux, Inc. · Feb 13, 2017
K073083 — VITEK 2 GRAM NEGATIVE PIPERACILLIN · bioMerieux, Inc. · Dec 10, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for bioMerieux. The logo consists of a stylized globe with vertical lines on the left side and a solid black fill on the right side. The company name "BIOMERIEUX" is printed in a stylized font below the globe.
K080107
.. ( )
### 510(k) SUMMARY
## VITEK® 2 Gram Negative Piperacillin/Tazobactam
#### 510(k) Submission Information:
| Submitter's Name: | bioMérieux, Inc. |
|----------------------|-----------------------------------------------------------------------------------------------------|
| Address: | 595 Anglum Road<br>Hazelwood, MO 63042 |
| Contact Person: | Nancy Weaver<br>Associate Director, Regulatory Affairs |
| Phone Number: | 314-731-8695 |
| Fax Number: | 314-731-8689 |
| Date of Preparation: | January 11, 2008 |
| Device Name: | |
| Formal/Trade Name: | VITEK® 2 Gram Negative Piperacillin/Tazobactam<br>(≤ 4 - ≥ 128 µg/ml) |
| Classification Name: | Fully Automated Short-Term Incubation Cycle Antimicrobial<br>Susceptibility Device, 21 CFR 866.1645 |
| Common Name: | VITEK 2 AST-GN Piperacillin/Tazobactam |
| Predicate Device: | VITEK 2 Gram Negative Levofloxacin (K072038). |
#### D. 510(k) Summary:
B.
C.
VITEK® 2 Gram Negative Piperacillin/Tazobactam is designed for antimicrobial susceptibility testing of Acinetobacter baumanii, Escherichia coli, Klebsiella pneumoniae, Pseudomonas aeruginosa, Citrobacter koseri, Morganii, Proteus mirabilis, Proteus vulgaris, Providencia rettgeri, Providencia stuartii, Salmonella enterica and Serratia marcescens. It is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. The antimicrobial presented in VITEK 2 AST Cards is in concentrations equivalent by efficacy to standard method concentrations in mod/ml. The VITEK 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology.
The bacterial isolate to be tested is diluted to a standardized concentration in 0.45% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 System automatically fills, seals and places the card into the incubator/reader. The VITEK 2 Compact has a manual filling and sealing operation. The VITEK 2 monitors the growth of each well in the card over a defined period of time (up to 18 hours). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card.
{1}------------------------------------------------
VITEK 2 Gram Negative Piperacillin/Tazobactam demonstrated substantially equivalent VTFEN Z Grant Negalife I speracilling reference method, as defined in the penomanoo when somple Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA, Issued Feb. 5, 2003.*
The Premarket Notification (510[k]) presents data in support of VITEK 2 Gram Negative The Tremans. Notinodion (orlogy) provinced with fresh and stock clinical r liperaoliin Fuzubuotam - xn external evaluations were designed to confirm the lished stool onalience or and setive Piperacillin/Tazobactam by comparing its performance with the CLSI broth microdilution reference method. The data is representative of performance on with the SEOF broth misroaliation rolonat instrument platforms, as evidenced in the AST equivalency study presented in the VITEK 2 Compact 510(k), K050002. VITEK 2 Gram Negative oqundler your claus procession demonstrated acceptable performance of 96.6% Essential Agreement, 96.6% Category Agreement and 100% reproducibility. Quality Control results were acceptable.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
Public Health Service
FFB 28 200
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Nancy Weaver Associate Director Regulatory Affairs bioMérieux 595 Anglum Road Hazelwood, MO 63042
k080107 Re:
Trade/Device Name: VITEK® 2 Gram Negative Piperacillin/tazobactam (≤ 4 - ≥ 128 ug/ml) Regulation Number: 21 CFR § 866.1645 Regulation Name: Antimicrobial Susceptibility Test System -- Short Incubation Regulatory Class: II Product Code: LON Dated: January 11, 2008 Received: January 15, 2008
Dear Ms. Weaver:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at 240-276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sally attayna
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indication for Use
510(k) Number (if known): | (ひ80/07
Device Name: VITEK® 2 Gram Negative Piperacillin/tazobactam (≤ 4 - ≥ 128 µg/ml)
Indication For Use:
VITEK® 2 Gram Negative Piperacillin/tazobactam is designed for antimicrobial susceptibility testing of gram-negative organisms and is intended for use with the VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK 2 Gram Negative Piperacillin/tazobactam is a quantitative test. Piperacillin/ tazobactam has been shown to be active against most strains of the organisms listed below according to the FDA label for the antimicrobial.
Active in vitro and in clinical infections Klebsiella pneumoniae Acinetobacter baumanii Escherichia coli Pseudomonas aeruginosa
Active in vitro but their clinical significance is uknown:
Proteus vulgaris
Providencia rettgeri
Providencia stuartii
Citrobacter koseri Morqanella morganii Proteus mirabilis
Salmonella enterica Serratia marcescens
The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 and VITEK® 2 Compact Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, and S. pneumoniae.
Prescription Use X (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Fueddie Lu. Root
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K08 0107
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.