BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - DAPTOMYCIN 0.125 - 32 UG/ML

K051689 · Becton, Dickinson & CO · LON · Aug 5, 2005 · Microbiology

Device Facts

Record IDK051689
Device NameBD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - DAPTOMYCIN 0.125 - 32 UG/ML
ApplicantBecton, Dickinson & CO
Product CodeLON · Microbiology
Decision DateAug 5, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2

Intended Use

The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram- positive bacteria of human origin.

Device Story

The BD Phoenix System is an automated microbiology platform for rapid identification and antimicrobial susceptibility testing (AST) of bacterial isolates. The system utilizes sealed, self-inoculating polystyrene panels containing biochemicals and antimicrobial agents. Input consists of bacterial isolates prepared in Phoenix ID or AST broth. The system continuously incubates panels and performs optical readings every 20 minutes using a redox indicator to detect organism growth and turbidity changes in the presence of antimicrobial agents. The system software processes these readings to determine minimum inhibitory concentration (MIC) values and categorical interpretations (S, I, R, or N). Used in clinical microbiology laboratories by trained personnel, the system provides automated results that assist clinicians in selecting appropriate antibiotic therapy for patients with bacterial infections.

Clinical Evidence

Clinical performance was evaluated by comparing the BD Phoenix System results for clinical, stock, and challenge isolates against the CLSI reference broth microdilution method. Testing included Gram-positive isolates. Results demonstrated 97.7% Essential Agreement (EA) and 98.8% Category Agreement (CA) for Daptomycin. Intra- and inter-site reproducibility exceeded 95%.

Technological Characteristics

Automated microbiology system using sealed, 136-well molded polystyrene panels with dried reagents. Employs broth microdilution method with redox indicator for growth detection. Continuous incubation at nominal temperature with optical readings every 20 minutes. System includes instrument, software, ID/AST broth, and AST indicator solution.

Indications for Use

Indicated for the addition of daptomycin (0.125 – 32 µg/mL) to the gram-positive susceptibility panel for Staphylococcus and Enterococcus species. For prescription use only.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ AUG 5 - 2005 K051689 ## CONFIDENTIAL AND PROPRIETARY # 510(k) SUMMARY | SUBMITTED BY: | Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152 Phone: 410-316-4287 Fax: 410-316-4499 | |------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | CONTACT NAME: | Monica Giguere Regulatory Affairs Specialist | | DATE PREPARED: | June 21, 2005 | | DEVICE TRADE NAME: | BD Phoenix™ Automated Microbiology System – Daptomycin 0.125-32 µg/mL | | DEVICE COMMON NAME: | Antimicrobial susceptibility test system-short incubation | | DEVICE CLASSIFICATION: | Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Device, 21 CFR 866.1645 | | PREDICATE DEVICES: | VITEK® System (PMA No. N50510) and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002). | | INTENDED USE: | The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram- positive bacteria of human origin. | ### DEVICE DESCRIPTION: The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components: - BD Phoenix instrument and software. . - BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . or AST determinations. - BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. . - BD Phoenix AST Broth used for performing AST tests only. . . - BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth . determination. {1}------------------------------------------------ The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells The Phoenix paller is a scaled and sell first and sen interacting must be a purch issues and containing dried reagents. Organisms for susceptime isolate. For each isolate. For each isolation premimarily idontified as a cistandard is prepared in Phoenix ID broth. The Phoenix AST method is a broth based microdilution test. The Phoenix System uilizes a redox I he Phoenix AST memor is a broaism growth in the presence of an antimicrobial agent. Indicator for the detection of organism growth in the presence of an the de indicator for the delection of organism growth in als prosectial turbidity are used in the determination Measurements of Changes to the marcator as were as ontains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations. The instrument houses the panels where they are continuously incubated at a nominal temperature of I he unstrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, R or N (susceptible, intermediate, resistant or not susceptible). ## DEVICE COMPARISON: The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the CLSI reference broth microdilution method. This premarket periormance when compared with and of the BD Phoenix™ Automated Microbiology System nottheanon provides and sapp only Phoenix panels with this antimicrobial agent. # SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING: The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the CLSI reference broth microdilution method (AST parels performance whole volipare to the system has been evaluated as defined in the FDA Draff propared according to NOOD 2117) Periew Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000. ### Site Reproducibility Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was millial at three sites using a panel of Gram-positive isolates. Each site tested the isolates in evaluated at three different days using one lot of Gram Positive Phoenix panels containing this antimicrobial agent and associated reagents. The results of the study demonstrate for the this antimicrobial agent there was an overall intra-site r no results of the states and an overall inter-site reproducibility greater than 95% for the Gram-positive isolates tested. {2}------------------------------------------------ ### Clinical Studies Clinical, stock and challenge isolates were tested across multiple geographically diversitilities and Offical, slock and chancilly were word word of the Phoenix antimicrobial susceptibility test with the Gram Positive Phoenix Panel format containing this antimicrobial agent. Phoenix System results the Chain Fostive I hould I and Former on the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the CLSI reference broth microdilution method. The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and The performatice of the I nooms Gysten wad assesses of the isolates tested. Essential Calegorical Agreement (Cr) to the enploanix™ Automated Microbiology System agress exactly of Agreement (EA) occurs which the DD sference result. Category Agreement (CA) occurs when the BD within I one two 1018 anaton is agy System agrees with the reference method with respect to the I hounds - Tratemated Microsoftop, - 3 FDA categorical interpretive criteria (susceptible, intermediate, resistant or not susceptible). Table 1 summarizes the performance for the isolates tested in this study. #### Performance of BD Phoenix System for Gram-Positive Organisms by Drug Table 1: | 1 | Acres of a la propose | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | EA (%) | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | (%) | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------| | Antimicrobial | Concentration | CA (n' | | CA (n) | CA | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>Daptomycin | CALL LE CANNO CHAND<br>0.125-32 ug/mL | 1568 | 07 / | 1 268 | 98.8 | ### Conclusions Drawn from Substantial Equivalence Studies The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA draft guidance document, "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002). {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus-like symbol, which is a modern interpretation of the Rod of Asclepius, a traditional symbol associated with medicine and healing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the symbol. AUG 5 - 2005 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Ms. Monica E. Giguere Regulatory Affairs Specialist BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152 Re: k051689 Trade/Device Name: BD PhoenixTM Automated Microbiology System Daptomycin (0.125-32 µg/mL) - Gram-Positive ID/AST Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: June 21, 2005 Received: June 23, 2005 Dear Ms. Giguere: We have reviewed your Section 510(k) premarket notification of intent to market the device we ferenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for associated in the May 28, 1976, the enactment date of the Medical Device Amendments, or to conninered pror to may 2011 11:12 in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou mayy arovisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it your device to such additional controls. Existing major regulations affecting your device it may 60 babject to accurations (CFR), Parts 800 to 895. In additions (CFR), Parts 800 to 895 may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean I hat FDA has made a determination that your device complies with other requirements of the Act that I Directeral statutes and regulations administered by other Federal agencies. You must or any I vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {4}------------------------------------------------ ### Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to organ mailing of substantial equivalence of your device to a legally premaired predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or you destre specific internation and advertising of your device, please contact the Office of In of questions on the promotion and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the I ou may obtain other general ers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html Sincerely yours, Sales a For Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # INDICATIONS FOR USE 510(k) Number (if known): 1.021689 Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent Daptomycin (0.125-32 ug/mL) - Gram positive ID/AST or AST only Phoenix Panels. Indications for Use: The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-positive bacterial isolates from pure culture belonging to the genera Staphylococcus and Enterococcus. This premarket notification is for the addition of the antimicrobial agent Daptomycin at concentrations of 0.125-32 ug/mL to Gram Positive ID/AST or AST only Phoenix panels. Daptomycin has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent. ### Active In Vitro and in Clinical Infections Against: Enterococcus faecalis (vancomycin-susceptible strains only) Staphylococcus aureus (including methicillin-resistant strains) ### Active In Vitro Against: Enterococcus faecalis (vancomycin-resistant strains) Enterococcus faecium (including vancomycin-resistant strains) Staphylococcus epidermidis (including methicillin-resistant strains) Staphylococcus haemolyticus V Prescription Use _ (Part 21 CFR 801 Subpart D) AND/OR Over-the-Counter Use (Part 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) HaseAts **Division Sign-Off** Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K051689 Page 9
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...