MICROSCAN SYNERGIES PLUS GRAM POSITIVE MIC/COMBO PANELS - RIFAMPIN

K051478 · Dade Behring, Inc. · LON · Jul 27, 2005 · Microbiology

Device Facts

Record IDK051478
Device NameMICROSCAN SYNERGIES PLUS GRAM POSITIVE MIC/COMBO PANELS - RIFAMPIN
ApplicantDade Behring, Inc.
Product CodeLON · Microbiology
Decision DateJul 27, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2

Intended Use

The MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-positive staphylococci and enterococci. After inoculation, panels are incubated for 4.5 - 18 hours at 35℃ +/- 1℃, in a WalkAway® SI, or equivalent, and read by the MicroScan® Instrumentation. Additionally, the panels may be incubated in a non-CO2 incubator and the Antimicrobial Susceptibility Testing (AST) portions can be read visually, according to the Package Insert. This particular submission is for the addition of the antimicrobial Rifampin, at concentrations of 0.25 to 8 ug/ml, to the test panel.

Device Story

MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panels determine antimicrobial susceptibility of gram-positive cocci and Listeria. Panels contain dehydrated antibiotics and control wells; rehydrated with bacterial suspension prepared via turbidity method. System uses WalkAway® -SI instrument with colorimetric optics (lamp/color wheel/photosensor) to detect bacterial growth. Input: isolated colonies from solid media. Transformation: growth-based detection via optical density changes over time (4.5 to 18 hours). Output: quantitative MIC values and qualitative susceptibility interpretations. Used in clinical microbiology labs; operated by technicians. Results assist clinicians in selecting appropriate antibiotic therapy for patients with bacterial infections.

Clinical Evidence

Performance evaluated using fresh and stock efficacy isolates and stock challenge strains. Compared against frozen reference panel. Results demonstrated >98% Essential Agreement for Rifampin. Reproducibility and precision testing performed using WalkAway® SI system. Quality control testing met acceptance criteria.

Technological Characteristics

Miniaturized broth dilution panels; dehydrated antimicrobial agents in wells. Rehydrated with Synergies plus™ Post Broth. Incubation at 35°C ± 1°C. Automated reading via WalkAway® SI system or visual reading. Supports aerobic/facultative anaerobic gram-positive organisms.

Indications for Use

Indicated for testing rifampin at 0.25 – 8 µg/mL for Staphylococcus spp. using gram-positive test panels; incubation 4.5-16 hours or 16-20 hours for overnight reading.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K051478 ## 510(k) Summary ## 510(k) Submission Information: | Device Manufacturer: | Dade Behring Inc. | |----------------------|--------------------------------------------------------------------------------------------------------| | Contact name: | Robert Eusebio, Regulatory Affairs Manager | | Fax: | 916-374-3144 | | Date prepared: | April 25, 2005 | | Product Name: | Microdilution Minimum Inhibitory Concentration (MIC) Panels | | Trade Name: | MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panels | | Intended Use: | To determine antimicrobial agent susceptibility | | 510(k) Notification: | New antimicrobial - Rifampin | | Predicate device: | MicroScan® Synergies plus™ Gram-Negative MIC/Combo Panels<br>and MicroScan® Dried Gram-Positive Panels | ### 510(k) Summary: MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panels, utilizing both the MicroScan® Rapid Fluorogenic Identification and Dried Overnight Antimicrobial Susceptibility Testing (AST) technologies, are designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-positive staphylococci and enterococci. The antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test that have been diluted in water and dehydrated. Various antimicrobial agents are diluted in water, buffer, or minute concentrations of broth, to concentrations bridging the range of clinical interest. Panels are rehydrated with Synergies plus™ Post Broth, after inoculation with a standardized suspension of the organism. After incubation in the WalkAway® SZ System, or equivalent, for 4.5 - 18 hours, the minimum inhibitory concentration (MIC) for the test organism is read by determining the lowest antimicrobial concentration showing inhibition of growth. The proposed MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panel demonstrated substantially equivalent performance when compared with a frozen Reference Panel, as defined in the FDA document "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA", dated February 5, 2003. The Premarket Notification (510[k]) presents data in support of the MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panel with Rifampin. The external evaluation was conducted with fresh and stock Efficacy isolates and stock Challenge strains. The external evaluations were designed to confirm the acceptability of the proposed Synergies plus™ Gram-Positive Panel by comparing its performance with a frozen Reference panel. Challenge strains were compared to Expected Results determined prior to the evaluation. The Synergies plus™ Gram-Positive Panel {1}------------------------------------------------ demonstrated acceptable performance with an overall Essential Agreement of >98% for Rifampin when compared with the frozen Reference panel. Instrument reproducibility testing demonstrated acceptable reproducibility and precision filled and reproduction inoculum preparation method and the WalkAway® SI System or equivalent. Quality Control testing demonstrated acceptable results for Rifampin. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized representation of a human figure, with three abstract shapes forming the body and head. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 JUL 27 2005 Mr. Robert Eusebio Regulatory Affairs Manager Dade Behring, Inc. 1584 Enterprise Boulevard West Sacramento, CA 95691-9972 k051478 Re: Trade/Device Name: MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panels Rifampin (0.25-8 µg/ml) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully automated short-term incubation cycle antimicrobial susceptibility system. Regulatory Class: Class II Product Code: LON, LRG, JWY, LTT Dated: June 1, 2005 Received: June 29, 2005 Dear Mr. Eusebio: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreated of the enactment date of the Medical Device Amendments, or to conniner or prior to 1125 20, 2017 11:11 accordance with the provisions of the Federal Food, Drug, de nees mat have been require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, mercere, mains of the Act include requirements for annual registration, listing of general controls pro risevring practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device it may of cacions in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean I loase oc advised that I DT brosantes over device complies with other requirements of the Act that I Dri has made a astad regulations administered by other Federal agencies. You must or any I cacal statutes and regaraments, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ #### Page 2 – This letter will allow you to begin marketing your device as described in your Section 510(k) I mis leter will anon you wo ought finding of substantial equivalence of your device to a legally premaired notification: ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the I va may overnit ones generational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html Sincerely yours, Sales, a Form Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): 4051478 Device Name: MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panels with Rifampin (0.25 - 8 µg/ml) ## Indications For Use: The MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-positive staphylococci and enterococci. After inoculation, panels are incubated for 4.5 - 18 hours at 35℃ +/- 1℃, in a WalkAway® SI, or equivalent, and read by the MicroScan® Instrumentation. Additionally, the panels may be incubated in a non-CO2 incubator and the Antimicrobial Susceptibility Testing (AST) portions can be read visually, according to the Package Insert. This particular submission is for the addition of the antimicrobial Rifampin, at concentrations of 0.25 to 8 ug/ml, to the test panel. The gram-positive organisms which may be used for Rifampin susceptibility testing in this panel are: > Staphylococcus aureus (including methicillin-resistant strains/MRSA) Staphylococcus epidermidis Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) | | Hall doty | |--|-----------| |--|-----------| | | Division Sign-Off | |--|-------------------| |--|-------------------| | Rif 510k_a.doc<br>6/1/2005 | Office of In Vitro Diagnostic Device<br>Evaluation and Safety | Page 1 of 1 | |----------------------------|---------------------------------------------------------------|-------------| |----------------------------|---------------------------------------------------------------|-------------| KO 51478 510(k)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...