BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - CEFAZOLIN (GP) 2-32 UG/ML
Applicant
Becton, Dickinson & CO
Product Code
LON · Microbiology
Decision Date
Apr 14, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1645
Device Class
Class 2
Intended Use
The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.
Device Story
The BD Phoenix System is an automated microbiology instrument for identification and antimicrobial susceptibility testing (AST) of bacterial isolates. The system uses sealed, molded polystyrene panels containing 136 micro-wells with various antimicrobial concentrations. Inputs consist of bacterial isolates from pure culture, prepared as an inoculum in Phoenix ID or AST broth. The system utilizes a redox indicator and turbidity measurements to monitor bacterial growth in the presence of antimicrobial agents. The instrument continuously incubates panels at a nominal temperature and takes readings every 20 minutes. Software interprets these readings to provide organism identification, MIC values, and categorical interpretations (S, I, or R). Used in clinical microbiology laboratories by trained technicians to guide antimicrobial therapy decisions. The system provides rapid results compared to traditional manual methods, aiding clinicians in selecting appropriate antibiotic treatments for patients.
Clinical Evidence
Performance evaluated via multi-site clinical study comparing Phoenix System results to NCCLS reference broth microdilution method. Study included clinical, stock, and challenge isolates. For Cefazolin (2-32 µg/mL) with Gram-positive panels, 507 isolates were tested, achieving 100% Essential Agreement and 100% Category Agreement. Intra- and inter-site reproducibility exceeded 95% for Gram-positive isolates.
Technological Characteristics
Automated microdilution system using sealed, molded polystyrene panels (136 micro-wells). Sensing principle: redox indicator and turbidity measurement. Energy source: electrical (instrument). Connectivity: integrated system with proprietary software. Sterilization: not specified. Software: automated interpretation of growth kinetics.
Indications for Use
Indicated for in vitro rapid identification and antimicrobial susceptibility testing of gram-positive bacteria (Staphylococcus, Enterococcus, other cocci/bacilli) from pure culture. Cefazolin testing is indicated for oxacillin-susceptible staphylococci only.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
Predicate Devices
VITEK® System (PMA No. N50510)
BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321)
BD Phoenix™ Automated Microbiology System with Ofloxacin (K020323)
BD Phoenix™ Automated Microbiology System with Levofloxacin (K020322)
Related Devices
K082852 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-NITROFURANTOIN (GP) 4-128 UG/ML · Becton, Dickinson & CO · Jan 29, 2009
K173252 — BD Phoenix Automated Microbiology System - GN Ceftolozane/tazobactam (0.25/4-32/4 ug/mL) · Becton, Dickinson and Company · Jan 5, 2018
K061869 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-PIPERACILLIN-TAZOBACTAM (GP) 1/4-128/4 UG/ML · Becton, Dickinson & CO · Aug 18, 2006
K033560 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - CEFTAZIDIME 0.5-64 UG/ML-GRAM-NEGATIVE ID/AST OR AST ONLY PHOENIX PANELS · Becton, Dickinson & CO · Jan 14, 2004
{0}------------------------------------------------
# 510(k) SUMMARY
| SUBMITTED BY: | Becton, Dickinson and Company<br>7 Loveton Circle<br>Sparks, MD 21152<br>Phone: 410.316-4287<br>Fax: 410.316-4499 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT NAME: | Monica Giguere<br>RA Specialist, Regulatory Affairs |
| DATE PREPARED: | April 11, 2005 |
| DEVICE TRADE NAME: | BD Phoenix™ Automated Microbiology System -<br>Cefazolin 2-32 µg/mL |
| DEVICE COMMON NAME: | Antimicrobial susceptibility test system-short incubation |
| DEVICE CLASSIFICATION: | Fully Automated Short-Term Incubation Cycle Antimicrobial<br>Susceptibility Device, 21 CFR 866.1645 |
| PREDICATE DEVICES: | VITEK® System (PMA No. N50510) and BD Phoenix™<br>Automated Microbiology System with Gatifloxacin (K020321,<br>May 23, 2002), Ofloxacin (K020323, April 14, 2002), and<br>Levofloxacin (K020322, March 27, 2002). |
| INTENDED USE: | The BD Phoenix™ Automated Microbiology System is<br>intended for the rapid identification and in vitro antimicrobial<br>susceptibility testing of isolates from pure culture of most<br>aerobic and facultative anaerobic Gram-negative and Gram-<br>positive bacteria of human origin. |
#### DEVICE DESCRIPTION:
The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the The BD Phoenix Automated Microbiology of Steeting (AST) of clinically relevant bacterial isolates. The system includes the following components:
- BD Phoenix instrument and software. .
- BD Flioenix instrument and solvirals for organism ID testing and antimicrobial agents . or AST determinations.
- of AST deceminations:
BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. .
- BD Phoenix AST Broth used for performing AST tests only. .
- DD Fhoenix AST Brom assurrer posseded to the AST Broth to aid in bacterial growth . determination.
{1}------------------------------------------------
The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells I The Filocins panel is a secured and over for susceptibility testing must be a pure culture and containing urica reagens. Organisms ion or Gram-positive isolate. For each isolate, an inoculation premimatiry tooners of the Farland standard is prepared in Phoenix ID broth.
The Phoenix AST method is a broth based microdilution test. The Phoenix system utilizes a redox I he I noonia AST method if a ganism growth in the presence of an antimicrobial agent. Indicator for the detection of organism go as well as bacterial turbidity are used in the determination Measurents of thanges to the marcates ageral antimicrobial agents with a wide range of two-fold doubling dilution concentrations.
The instrument houses the panels where they are continuously incubated at a nominal temperature of The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).
### DEVICE COMPARISON:
The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the NCCLS reference broth microdilution method. This premarket performatics when somporting the use of the BD Phoenix™ Automated Microbiology System hourieanon provides and cappermise hoenix panels with this antimicrobial agent.
# SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:
The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the NCCLS reference broth microdilution method (AST panels performance with compared to the research has been evaluated as defined in the FDA Draft prepared according to FFOOD 117). The of Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000.
#### Site Reproducibility
Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was millia- and micr-site reproductions) of Gram-positive isolates. Each site tested the isolates in e valuated at three different days using one lot of Gram Positive Phoenix panels containing this antimicrobial agent and associated reagents.
The results of the study demonstrate for the this antimicrobial agent there was an overall intra-site The results of the study demonentals to: an overall inter-site reproducibility greater than 95% for the Gram-positive isolates tested.
{2}------------------------------------------------
#### Clinical Studies
Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the Gram Positive Phoenix Panel format containing this antimicrobial agent. Phoenix System results the Challenge set isolates were compared to the expected results. Phoenix System results for clinical ior Charlenge bot bolates nor results obtained from the NCCLS reference broth microdilution method.
The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and The performance of the I no expected/reference results for all isolates tested. Essential Agreement Calegory Agreement (OT) to expected Microbiology System agrees exactly or within ± one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD oncenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant).
Table 1 summarizes the performance for the isolates tested in this study.
#### Performance of BD Phoenix System for Gram-Positive Organisms by Drug Table 1:
| a different of the same of the said the same of the same of the same of the<br>Antimicrobial | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>The contract and particular procession in accession in accession in accession in accession in accession in accession in accession in accession in accession in accession in ac<br>Concentration | The first of the many of the comments of the first of the first of<br>EA (n) | Free of Canadian Cases on and Concession of Concession of Concession of Concession of Concessional Concessional Concession of Concessional of Concessional of Concessional of<br><br>EA (%)<br>. | House and can and and<br><br>CA (n)<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | The family of the first of the first<br>CA (%)<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Carren Comments of Children Comments of Children<br>The Comments and College and College of Concess of<br>A<br>Cefazolin | Concession from on an every of they in form<br>THE LEAST LE LEASE LEASE LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST<br>And Concession Company of Children<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>2-32 ug/mL | Children assess as compare of concession in any<br>507 | 00 5<br>1 - 1 - 1 - | 507 | 00 7 |
## Conclusions Drawn from Substantial Equivalence Studies
The data collected from the substantial equivalence studies demonstrate that testing on the BD The udia conceded from the odiolany System with this antimicrobial agent is substantially equivalent as outlined in the FDA draft guidance document, "Guidance on Review Criteria for Assessment of as outined in the I Dri drit garames" March 8, 2000. Technological characteristics of this system Amilificantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with I Dri under 1 (K020321, May 23, 2002), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002).
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three curved lines representing its wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
APR 1 4 2005
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Monica Giguere RA Specialist, Regulatory Affairs BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152
k050555 Re: K050335
Trade/Device Name: BD Phoenix™ Automated Microbiology System Cefazolin (2-32 µg/mL) - Gram-Postive ID/AST or AST Regulation Number: 21 CFR 866.1645 Regulation Name: 21 CFTC 000.1015
Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: March 1, 2005 Received: March 3, 2005
Dear Ms. Giguere:
We have reviewed your Section 510(k) premarket notification of intent to market the indication We have reviewed your Section >10(x) premation is suivalent (for the indications
referenced above and have determined the device is subscreen in interstate referenced above and have delemined the devices marketed in interstate
for use stated in the enclosure) to legally marketed president device Amendments for use stated in the enclosure) to legally manced production Device Amendments, or to
commerce prior to May 28, 1976, the enactment as ansieral of the Federal Food. Drug commerce prior to May 28, 1970, the chaculteners with the provisions of the Federal Food, Drug, devices that have been recuire approval of a proval at a proval application (PMA).
and Cosmetic Act (Act) that do not require approval of a provisions of the Act . The and Cosmetic Act (Act) that do not require approval controls provisions of the Act. The Act. The You may, therefore, market the device, sayler to tire gents for annual registration, listing of
general controls provisions of the Act include required withing and general controls provisions of the receined requirements of the management misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), If your device is classified (See adove) ntrols. Existing major regulations affecting your device.
It may be subject to such additional controls. Existing major regulation it may be subject to such additional controls. Existing mayor regals.800 to 895. In addition, FDA
can be found in Title 21, Code of Federal Regulations (CFR), Parts Registe can be found in Title 21, Code of Pederal Regaralises (1) (1) = ) = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = =
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issualited on a substitution with other requirements of the Act
that FDA has made a determination that your device and two Edgerains. You must that FDA has made a determination inal your devices by other Federal agencies. You must
or any Federal statutes and regulations administered by other registration and listi or any Federal statures and regilations annings, but not lied to: registration and listing (21
comply with all the Act's requirements, including, on anonomonomonmonmonmentice comply with all the Act s requirements, including, and good manufacturing practice
CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice CFR Part 807); labeling (21 CFR Parts 601 and 607);
requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
#### Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k)
the same of the secure of alience for betweenes of your device to a legal This letter will allow you to begin marketing your advice of your device of your device to a legally
premarket notification. The FDA finding of substantial equivale and this premarket notification. The IDA inding of Justillation for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, a If you desire spectile information and advertising of your device, please contact the Office of In
or questions on the promotion and advertising of your device, please pote t or questions on the promotion and advertising of Jour 2015-0484. Also, please note the Vitro Diagnostic Device Evaluation and Barey at (210-210).
regulation entitled, "Misbranding by reference to premarket notification weder the Ast from the regulation entitled, "Misoranding of reichers on your responsibilities under the Act from the You may obtain other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address (000) 050-2012 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11
Sincerely yours,
Sales a form
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
510(k) Number: _ K 05 05 5 5
Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent Device Name. " DD Frioomira - Gram positive ID/AST or AST only Phoenix panels.
Indications for Use:
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of The BD Phoenix™ Automatics witchoology Bystem in ation (MIC) of most Gram-negative aerobic
antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Nega and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaeae and Nonand facultative anactoric bacteria isolates from pure culture belonging to the genera Staphylococcus and Enterococcus.
This premarket notification is for the addition of the antimicrobial agent oefarzolin at concentrations of This premarket nothcation is for the addition of the animalized and only of the many of 2-32 µgmL to Grain Fostive D7701 of NOT off Processions is described in the FDA-approved package insert for this antimicrobial agent.
# Active In Vitro and in Clinical Infections Against:
Staphylococcus aureus (including penicillinase-producing strains) Staphylococcus epidermidis
Prescription Use _ ブ (Per 21 CFR 801.109) Over-the-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Division Signature | |
|--------------------|--|
|--------------------|--|
Office of *In Vitro* Diagnostic Device Evaluation and Safety
Page 1
| BD Diagnostic Systems<br>Becton, Dickinson and Company | 510(k) K05 0555 |
|--------------------------------------------------------|-----------------|
|--------------------------------------------------------|-----------------|
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.