BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-PIPERCILLIN (GN) 0.5-128 UG/ML

K041572 · Becton, Dickinson and Co. · LON · Aug 12, 2004 · Microbiology

Device Facts

Record IDK041572
Device NameBD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-PIPERCILLIN (GN) 0.5-128 UG/ML
ApplicantBecton, Dickinson and Co.
Product CodeLON · Microbiology
Decision DateAug 12, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2

Intended Use

The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic gram-negative and gram-positive bacteria of human origin.

Device Story

The BD Phoenix System is an automated microbiology platform for rapid identification and antimicrobial susceptibility testing (AST) of bacterial isolates. The system utilizes sealed, self-inoculating polystyrene panels containing biochemicals and antimicrobial agents. Inputs consist of pure bacterial cultures inoculated into Phoenix ID or AST broth. The system performs broth microdilution testing by monitoring redox indicator changes and bacterial turbidity via continuous incubation at 35°C; readings are taken every 20 minutes. The system outputs MIC values and categorical interpretations (S, I, or R). Used in clinical microbiology laboratories by trained technicians; results assist clinicians in selecting appropriate antibiotic therapy for patients with bacterial infections.

Clinical Evidence

Performance evaluated via method comparison against NCCLS broth dilution reference panels using 1,781 clinical and challenge isolates. Combined results showed 94.3% Essential Agreement (EA) and 93.8% Category Agreement (CA). Reproducibility demonstrated >95% across three sites. Quality control testing using NCCLS strains confirmed system accuracy. No clinical sensitivity/specificity studies performed as this is an in vitro diagnostic AST system.

Technological Characteristics

Automated colorimetric oxidation-reduction growth-based system. Materials: molded polystyrene trays with 136 micro-wells. Energy: 35°C incubation. Connectivity: barcode scanning, automated software-driven analysis. Software: Expert system using NCCLS-derived rules. Dimensions: multi-well tray format. Sterilization: not specified.

Indications for Use

Indicated for in vitro quantitative determination of antimicrobial susceptibility by MIC for gram-negative aerobic and facultative anaerobic bacteria (Enterobacteriaceae, Non-fermenters) and gram-positive bacteria (Staphylococcus, Enterococcus) from pure culture. Specifically for Piperacillin (0.5-128 µg/mL).

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ | AUG 1 2 2004 | 510(K) SUMMARY | |------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | SUBMITTED BY: | Becton, Dickinson and Company<br>7 Loveton Circle<br>Sparks, MD 21152<br>Phone: 410-316-4287<br>Fax: (410)-316-4499 | | CONTACT NAME: | Monica Evelyn Giguere<br>Regulatory Affairs Specialist | | DATE PREPARED: | June 10, 2004 | | DEVICE TRADE NAME: | BD Phoenix™ Automated Microbiology System -<br>Piperacillin 0.5-128 µg/mL | | DEVICE COMMON NAME: | Antimicrobial susceptibility test system-short incubation | | DEVICE CLASSIFICATION: | Fully Automated Short-Term Incubation Cycle Antimicrobial<br>Susceptibility Device, 21 CFR 866.1645 | | PREDICATE DEVICES: | VITEK® System (PMA No. N50510) and BD Phoenix™<br>Automated Microbiology System with Gatifloxacin (K020321,<br>May 23, 2002), Ofloxacin (K020323, April 14, 2002), and<br>Levofloxacin (K020322, March 27, 2002). | | INTENDED USE: | The BD Phoenix™ Automated Microbiology System is<br>intended for the rapid identification and in vitro antimicrobial<br>susceptibility testing of isolates from pure culture of most<br>aerobic and facultative anaerobic gram-negative and gram-<br>positive bacteria of human origin. | ## DEVICE DESCRIPTION: The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components: - BD Phoenix instrument and software. . - BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . for AST determinations. - BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. . - BD Phoenix AST Broth used for performing AST tests only. . - BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth . determination. {1}------------------------------------------------ The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a gram-negative or gram-positive isolate. For each isolate, an inoculation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID Broth. The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination Macterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations. The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant). # DEVICE COMPARISON: The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the NCCLS reference broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System gram-negative ID/AST or AST only Phoenix panels with this antimicrobial agent. # SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING: The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the NCCLS reference broth microdilution method (AST panels prepared according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document, "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000. # Site Reproducibility Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel of gram-negative isolates. Each site tested the isolates in triplicate on three different days using one lot of gram-negative Phoenix panels containing this antimicrobial agent and associated reagents. The results of the study demonstrate for the this antimicrobial agent there was an overall intra-site reproducibility of greater than 90% and an overall inter-site reproducibility greater than 95% for the gram-negative isolates tested. {2}------------------------------------------------ # Clinical Studies Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the United States to delivinate the parcomtaining this antimicrobial agent. Phoenix System results me grant-hogan ve i noomir pared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the NCCLS reference broth microdilution method. The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and The performance of the Procedireference results for all isolates tested. Essential Agreement Category Agreement (CTT) to enported Microbiology System agrees exactly or within ± (EA) occurs when and 22 22 and 11 Category Agreement (CA) occurs when the BD one two Total unution to the obiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant). Table 1 summarizes the performance for the isolates tested in this study. #### Performance of BD Phoenix System for Gram-Negative Organisms by Drug Table 1: | Antimicrobial | Comments of the control of the control of the control of the contribution of the contribution of the contribution of the contribution of the contribution of the contribution<br>Concentration | A market of the contract and the contract and the comments of the comments of the comments of the comments of the contribution of the comments of the contribution of the cont<br>EA (n) | and the same of the management of the minutes of the many of<br>EA (%) | CA (n) | CA (%) | |---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|--------------|--------| | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>A<br>and the comments of the first of the first of the first of<br>.<br>Pıperacılın | · Please with a comments of the promote of the province of the proposes of the control of the contribution of the contribution of the contribution of the contribution of the<br>0.5-128 ug/mL | Acres of the Lands Labor Labor one the v<br>1781 | 94.3 | 1701<br>10 - | 93.8 | # Conclusions Drawn from Substantial Equivalence Studies The data collected from the substantial equivalence studies demonstrate that testing on the BD The cata concected from als about with this antimicrobial agent is substantially equivalent i nooms - Pratemate document, "Class II Special Controls Guidance Document: as outined in Susceptibility Test (AST) Systems; Guidance for Industry and FDA", February 5, 1 connectorial barceptions of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002). {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows a logo with a circular design on the left and a stylized bird-like figure on the right. The circular part contains text that appears to be the company name, though the letters are not very clear. The bird-like figure is composed of three curved lines that converge at the bottom, creating a sense of movement or flight. The logo is simple and uses a monochrome color scheme. Public Health Service AUG 1 2 2004 Food and Drug Administratio 2098 Gaither Road Rockville MD 20850 Ms. Monica Evelyn Giguere Regulatory Affairs Specialist BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152 Re: k041572 Trade/Device Name: BD Phoenix™ Automated Microbiology Systems Piperacillin (0.5-128 µg/mL) - Gram-Negative ID/AST Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: June 10, 2004 Received: June 14, 2004 Dear Ms. Giguere: We have reviewed your Section 510(k) premarket notification of intent to market the device w 6 have a revea and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for ass stated on to May 28, 1976, the enactment date of the Medical Device Amendments, or to conninered pror to 1125 2011 11 accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmetic Prer (110) that the device, subject to the general controls provisions of the Act. The I ou may, aterest, mains of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean r that FDA has made a determination that your device complies with other requirements of the Act man regulations and regulations administered by other Federal agencies. You must or any I odetar statuated and equirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {4}------------------------------------------------ ### Page 2 This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Salazar Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### Indications for Use 510(k) Number: K041572 -------- Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent Device Name: 5-128 ug/mL) - Gram-Negative ID/AST or AST only Phoenix panels. Indications for Use: The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Nonand ractive and most gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus and Enterococcus. This premarket notification is for the addition of the antimicrobial agent Piperacillin at concentrations of I this premarks normounter is to the SST only Phoenix panels. Piperacillin has been shown to 0.5-120 pg miro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent. # Active In Vitro and in Clinical Infections Against: Acinetobacter species Enterobacter species Escherichia coli Klebsiella species Morganella morganii Pseudomonas aeruginosa Serratia species # Active In Vitro Against: Citrobacter koseri Citrobacter freundii Pseudomonas fluorescens Yersinia enterocoloitica Prescription Use V (Part 21 CFR 801 Subpart D) Over-the-Counter Use (21 CFR 807 Subpart C) | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) | | |--------------------------------------------------------------------------|--| |--------------------------------------------------------------------------|--| | Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) | | |-------------------------------------------------------------------|--------------------------------------| | Division Sign-Off | Page 1 of 1 | | | Office of In Vitro Diagnostic Device | Office of In Vitro Diagnostic Device Evaluation and Safety
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...