K032861 · bioMerieux, Inc. · LON · Nov 12, 2003 · Microbiology
Device Facts
Record ID
K032861
Device Name
VITEK 2 GRAM NEGATIVE MOXIFLOXACIN
Applicant
bioMerieux, Inc.
Product Code
LON · Microbiology
Decision Date
Nov 12, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1645
Device Class
Class 2
Attributes
AI/ML
Intended Use
The VITEK®2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK®2 System for the automated quantitative or qualitative susceptibility testing of isolated colonies for most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, and S. pneumoniae. The VITEK®2 Gram Negative Susceptibility Card is intended for use with the VITEK®2 System in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic gram-negative bacilli to antimicrobial agents when used as instructed in the Online Product Information.
Device Story
VITEK®2 Gram Negative Moxifloxacin card is an in vitro diagnostic test for antimicrobial susceptibility. Input: isolated bacterial colonies suspended in 0.45-0.5% saline. Process: VITEK®2 instrument automatically dilutes suspension, fills 64-microwell cards via vacuum, seals cards, and incubates at 35.5°C. Optics system performs periodic light transmittance measurements every 15 minutes to monitor organism growth. Discriminant analysis algorithm processes growth characteristics to calculate Minimum Inhibitory Concentration (MIC) and determine susceptibility category. Output: MIC value and category interpretation (Susceptible, Intermediate, Resistant). Used in clinical laboratories by technicians. Results assist healthcare providers in selecting appropriate antimicrobial therapy for patients with gram-negative infections.
Clinical Evidence
Performance evaluated via clinical and challenge studies (n=682 total isolates). Comparison against NCCLS reference agar dilution method. Results: 98.7% Essential Agreement (EA) and 97.8% Category Agreement (CA). No very major (vmj) or major (maj) errors observed; 2.2% minor (min) error rate. Reproducibility >95% across three sites. Bench testing confirmed comparability between automated and manual inoculation methods.
Technological Characteristics
Miniaturized AST card containing antimicrobial medium in doubling dilution concentrations. Automated incubation and optical growth monitoring. System-based processing. No specific material standards or software architecture details provided.
Indications for Use
Indicated for testing moxifloxacin at concentrations of 0.25, 1, and 4 ug/ml for reporting results between ≤0.25 – ≥8 ug/ml for Enterobacteriaceae using the VITEK®2 System.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
Predicate Devices
VITEK®2 Gram Negative Susceptibility Card for Cefpodoxime (N50510/S120)
Related Devices
K032788 — VITEK 2 GRAM NEGATIVE GATIFLOXACIN · bioMerieux, Inc. · Nov 4, 2003
K091899 — VITEK 2 GRAM NEGATIVE MEROPENEM · bioMerieux, Inc. · Aug 14, 2009
K032711 — VITEK GRAM NEGATIVE GATIFLOXACIN · bioMerieux, Inc. · Oct 31, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
K032861
NOV 1 2 2003
Image /page/0/Picture/2 description: The image shows the logo for bioMerieux. The logo consists of the company name "bioMerieux" in a stylized font, with a vertical line running through the middle of the text. Above the text is a circular graphic with a textured pattern on the left side and a solid black fill on the right.
# 510(k) SUMMARY
## VITEK® 2 Gram Negative Moxifloxacin
### 510(k) Submission Information:
| | Submitter's Name: | bioMérieux, Inc. |
|-----------------|----------------------|--------------------------------------------------------------------------------------------------------|
| | Address: | 595 Anglum Road<br>Hazelwood, MO 63042 |
| | Contact Person: | Nancy Weaver<br>Staff Regulatory Affairs Specialist |
| | Phone Number: | 314-731-8695 |
| | Fax Number: | 314-731-8689 |
| | Date of Preparation: | September 8, 2003 |
| B. Device Name: | | |
| | Formal/Trade Name: | VITEK® 2 Gram Negative Moxifloxacin (0.25 - 4.0<br>µg/ml) |
| | Classification Name: | Fully Automated Short-Term Incubation Cycle<br>Antimicrobial Susceptibility Device,<br>21 CFR 866.1645 |
| | Common Name: | VITEK 2 AST-GN Moxifloxacin |
| | C. Predicate Device: | VITEK 2 Gram Negative Susceptibility Test for<br>Cefpodoxime (N50510/S120) |
### D. 510(k) Summary:
VITEK® 2 Gram Negative Moxifloxacin is designed for antimicrobial susceptibility testing of, Klebsiella pneumoniae, Citrobacter freundii, Enterobacter cloacae, Escherichia coli Klebsiella oxytoca and Proteus mirabilis. It is intended for use with the VITEK® 2 System as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. The antimicrobial presented in VITEK 2 AST Cards is in concentrations equivalent by efficacy to standard method concentrations in mcg/ml. The VITEK 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology.
The bacterial isolate to be tested is diluted to a standardized concentration in 0.45% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 automatically fills, seals and places the card into the incubator/reader. The VITEK 2 monitors the growth of each well in the card over a defined period of time (up
{1}------------------------------------------------
to 18 hours). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card.
VITEK 2 Gram Negative Moxifloxacin demonstrated substantially equivalent performance when compared with the NCCLS reference agar dilution method, as defined in the FDA DRAFT document "Guidance on Review Criteria for Assessment of Antimicrobial susceptibility Devices", dated March 8, 2000 and Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA. Issued Feb. 5, 2003.
The Premarket Notification (510[k]) presents data in support of VITEK 2 Gram Negative Moxifloxacin. An external evaluation was conducted with fresh and stock clinical isolates and stock challenge strains. The external evaluations were designed to confirm the acceptability of VITEK 2 Gram Negative Moxifloxacin by comparing its performance with the NCCLS agar dilution reference method. VITEK 2 Gram Negative Moxifloxacin demonstrated acceptable performance of 98.7% overall Essential Agreement when compared to the agar dilution reference method. Reproducibility and Quality Control demonstrated acceptable results.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
NOV 1 2 2003
Ms. Nancy Weaver Staff Regulatory Affairs Specialist BioMérieux, Inc. 595 Anglum Road Hazelwood, MO 63042-2320
Re: k032861
Trade/Device Name: VITEK 2® Gram Negative Moxifloxacin (0.25-4 ug/ml) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: September 8, 2003 Received: September 12, 2003
Dear Ms. Weaver:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
#### Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use Statement
510(k) Number (if known): K032861
Device Name: VITEK 2° Gram Negative Moxifloxacin (0.25 - 4 µg/ml)
Indications for Use:
The VITEK 2 Gram Negative Moxifloxacin is designed for antimicrobial susceptibility testing of Klebsiella pneumoniae, Citrobacter freundii, Enterobacter cloacae, Escherichia coli, Klebsiella oxytoca and Proteus mirabilis. It is intended for use with the VITEK 2 System as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents.
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K032861
X
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.