K032788 · bioMerieux, Inc. · LON · Nov 4, 2003 · Microbiology
Device Facts
Record ID
K032788
Device Name
VITEK 2 GRAM NEGATIVE GATIFLOXACIN
Applicant
bioMerieux, Inc.
Product Code
LON · Microbiology
Decision Date
Nov 4, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1645
Device Class
Class 2
Attributes
AI/ML
Intended Use
VITEK® 2 Gram Negative Gatifloxacin is designed for antimicrobial susceptibility testing of Escherichia coli, Klebsiella pneumoniae, Proteus mirabilis, Acinetobacter lwoffii, Citrobacter koseri, C. freundii, Enterobacter aerogenes, E. cloacae, Klebsiella oxytoca, Morganella morganii, and Proteus vulgaris. It is intended for use with the VITEK® 2 System as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents.
Device Story
VITEK® 2 Gram Negative Gatifloxacin test card performs automated antimicrobial susceptibility testing (AST). Input: isolated bacterial colonies suspended in saline (standardized to McFarland 0.5). Process: VITEK® 2 instrument automatically dilutes suspension, fills 64-microwell card via vacuum, seals card, and incubates at 35.5°C. Optics system performs periodic light transmittance readings every 15 minutes to measure organism growth inhibition by antibiotic. Discriminant analysis algorithm calculates Minimum Inhibitory Concentration (MIC) and provides category interpretation (S, I, R). Used in clinical laboratories by technicians. Output: quantitative MIC and qualitative susceptibility category. Benefits: rapid, automated determination of antibiotic susceptibility to guide clinical treatment decisions.
Clinical Evidence
Bench testing only. Performance evaluated by comparing VITEK 2 Gram Negative Gatifloxacin results against NCCLS agar dilution reference method using fresh/stock clinical isolates and stock challenge strains. Demonstrated 99.7% overall Essential Agreement. Reproducibility and Quality Control results were acceptable.
Technological Characteristics
Miniaturized card containing antimicrobial medium in concentrations equivalent to standard microdilution methods. Sensing principle: automated optical monitoring of bacterial growth in wells. Connectivity: integrated with VITEK 2 System. Software: automated incubation and analysis.
Indications for Use
Indicated for testing clinically significant aerobic gram-negative bacilli (Enterobacteriaceae) against Gatifloxacin at concentrations of 1, 4, and 8 ug/mL (range ≤ 0.25 to ≥ 8 ug/mL).
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
Predicate Devices
VITEK®2 Gram Negative AST Panel for cefpodoxime (N50510/S120)
Related Devices
K032711 — VITEK GRAM NEGATIVE GATIFLOXACIN · bioMerieux, Inc. · Oct 31, 2003
K191766 — VITEK 2 AST-Gram Negative Eravacycline (<=0.12 - >=4µg/mL) · bioMerieux, Inc. · Aug 5, 2019
K032861 — VITEK 2 GRAM NEGATIVE MOXIFLOXACIN · bioMerieux, Inc. · Nov 12, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
K032788
NOV = 4 2003
Image /page/0/Picture/2 description: The image shows the logo for bioMerieux. The logo consists of a stylized image of a cell or sphere that is half black and half white with lines. Below the image is the company name, "BIOMERIEUX" in a stylized font.
# 510(k) SUMMARY
# VITEK® 2 Gram Negative Gatifloxacin
### 510(k) Submission Information:
| Submitter's Name: | bioMérieux, Inc. |
|----------------------|--------------------------------------------------------------------------------------------------------|
| Address: | 595 Anglum Road<br>Hazelwood, MO 63042 |
| Contact Person: | Nancy Weaver<br>Staff Regulatory Affairs Specialist |
| Phone Number: | 314-731-8695 |
| Fax Number: | 314-731-8689 |
| Date of Preparation: | September 3, 2003 |
| B. Device Name: | |
| Formal/Trade Name: | VITEK® 2 Gram Negative Gatifloxacin |
| Classification Name: | Fully Automated Short-Term Incubation Cycle<br>Antimicrobial Susceptibility Device,<br>21 CFR 866.1645 |
| Common Name: | VITEK 2 AST-GN Gatifloxacin |
| C. Predicate Device: | VITEK 2 Gram Negative Susceptibility Test for<br>Cefpodoxime (N50510/S120) |
### D. 510(k) Summary:
VITEK® 2 Gram Negative Gatifloxacin is designed for antimicrobial susceptibility testing of Escherichia coli, Klebsiella pneumoniae, Proteus mirabilis, Acinetobacter Iwoffii, Citrobacter koseri, C. freundii, Enterobacter aerogenes, E. cloacae, Klebsiella oxytoca, Morganella morganii, and Proteus vulgaris. It is intended for use with the VITEK® 2 System as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. The antimicrobial presented in VITEK 2 AST Cards is in concentrations equivalent by efficacy to standard method concentrations in mca/ml. The VITEK 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology.
The bacterial isolate to be tested is diluted to a standardized concentration in 0.45% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 automatically fills, seals and places the card into the incubator/reader. The VITEK 2 monitors the growth of each well in the card over a defined period of time (up
{1}------------------------------------------------
to 18 hours). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card.
VITEK 2 Gram Neqative Gatifloxacin demonstrated substantially equivalent performance when compared with the NCCLS reference agar dilution method, as defined in the FDA DRAFT document "Guidance on Review Criteria for Assessment of Antimicrobial susceptibility Devices", dated March 8, 2000 and Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA. Issued Feb. 5, 2003.
The Premarket Notification (510[k]) presents data in support of VITEK 2 Gram Negative Gatifloxacin. An external evaluation was conducted with fresh and stock clinical isolates and stock challenge strains. The external evaluations were designed to confirm the acceptability of VITEK 2 Gram Negative Gatifloxacin by comparing its performance with the NCCLS agar dilution reference method. VITEK 2 Gram Negative Gatifloxacin demonstrated acceptable performance of 99.7% overall Essential Agreement when compared to the agar dilution reference method. Reproducibility and Quality Control demonstrated acceptable results.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of an eagle with three lines extending from its head.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
NOV - 4 2003
Ms. Nancy Weaver Staff Regulatory Affairs Specialist Biomerieux, Inc 595 Anglum Road Hazelwood, MO 63042-2320
Re: k032788 Trade/Device Name: VITEK® 2 Gram Negative Gatifloxacin (1-8 µg/ml) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully automated short-term incubation cycle antimicrobial susceptibility system Regulatory Class: Class II Product Code: LON Dated: September 3, 2003 Received: September 8, 2003
Dear Ms. Weaver:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use Statement
KO32788 510(k) Number (if known):
Device Name: VITEK® 2 Gram Negative Gatifloxacin (1 - 8 µg/ml)
Indications for Use:
VITEK® 2 Gram Negative Gatifloxacin is designed for antimicrobial susceptibility testing of Escherichia coli, Klebsiella pneumoniae, Proteus mirabilis, Acinetobacter lwoffii, Citrobacter koseri, C. freundii, Enterobacter aerogenes, E. cloacae, Klebsiella oxytoca, Morganella morganii, and Proteus vulgaris. It is intended for use with the VITEK® 2 System as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Fea Ott
Division Sign-Off
11/4/03
Office of In Vitro Diagnostic Device Evaluation and Safety
prescription use only.
K632788 510(k).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.