MICROSCAN SYNERGIES PLUS GRAM NEGATIVE MIC/COMBO PANELS WITH IMIPENEM (0.25, 1-32 UG/ML)
K020246 · Dade Behring, Inc. · LON · Feb 6, 2002 · Microbiology
Device Facts
Record ID
K020246
Device Name
MICROSCAN SYNERGIES PLUS GRAM NEGATIVE MIC/COMBO PANELS WITH IMIPENEM (0.25, 1-32 UG/ML)
Applicant
Dade Behring, Inc.
Product Code
LON · Microbiology
Decision Date
Feb 6, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1645
Device Class
Class 2
Intended Use
MicroScan® Synergies plus™ Gram-Negative MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic Gram-Negative bacilli (Enterobacteriaceae, glucose non-fermenters, and non-Enterobacteriaceae glucose non-fermenters). After inoculation, panels are read on the WalkAway® SI System or equivalent (upgraded WalkAway® 40 or WalkAway® 96) according to the Package Insert.
Device Story
MicroScan Synergies plus Gram-Negative MIC/Combo Panels are miniaturized broth dilution susceptibility tests; antimicrobial agents are dried in micro-titer wells with Mueller-Hinton Broth. Panels are inoculated with standardized organism suspensions; incubated at 35°C for 4.5–18 hours in WalkAway SI System or equivalent (WalkAway 40/96). System performs automated reading of growth inhibition; determines Minimum Inhibitory Concentration (MIC). Used in clinical microbiology laboratories by trained personnel. Output provides quantitative/qualitative susceptibility data to guide antibiotic therapy for Gram-negative infections.
Clinical Evidence
Bench testing only. Performance evaluated using fresh and stock clinical isolates and challenge strains compared to NCCLS frozen reference panels. Overall Essential Agreement for Imipenem was >96%. Reproducibility and precision testing confirmed acceptable performance using the WalkAway SI System and turbidity inoculum preparation.
Technological Characteristics
Miniaturized broth microdilution panels; dried antimicrobial agents in micro-titer wells; Mueller-Hinton Broth. Automated incubation and optical reading via WalkAway SI System (or upgraded 40/96). Inoculation via standardized suspension. 4.5–18 hour incubation at 35°C.
Indications for Use
Indicated for antimicrobial susceptibility testing of rapidly growing aerobic and facultative anaerobic Gram-Negative bacilli (e.g., Enterobacteriaceae, Pseudomonas, Acinetobacter, etc.) from solid media. For use with Imipenem. Prescription use only.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
Predicate Devices
MicroScan Dried Gram Negative MIC/Combo Panels
Related Devices
K020209 — MICROSCAN SYNERGIES PLUS GRAM NEGATIVE MIC/COMBO PANELS WITH AMIKACIN (0.25-128 UG/ML) · Dade Behring, Inc. · Feb 6, 2002
K020182 — MICROSCAN SYNERGIES PLUS GRAM NEGATIVE MIC/COMBO PANELS WITH NETILMICIN (0.12-32 UG/ML) · Dade Behring, Inc. · Feb 6, 2002
K020319 — MICROSCAN SYNERGIES PLUS GRAM NEGATIVE MIC/COMBO PANELS WITH AMPICILLIN/SULBACTAM (0.5/0.25-64/34 UG/ML) · Dade Behring, Inc. · Apr 17, 2002
K020173 — MICROSCAN SYNERGIES PLUS GRAM NEGATIVE MIC/COMBO PANELS WITH TETRACYCLINE (0.12-128 UG/ML) · Dade Behring, Inc. · Feb 6, 2002
K020398 — MICROSCAN SYNERGIES PLUS GRAM NEGATIVE MIC/COMBO PANELS WITH AMPICILLIN (0.25-128 UG/ML) · Dade Behring, Inc. · May 16, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with its wings spread and head turned to the right. The bird is composed of three curved lines that suggest feathers or wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
Food and Drug Administration
AUG 3 0 2004
2098 Gaither Road Rockville MD 20850
Mr. Robert Eusebio Manager Regulatory Affairs Dade MicroScan, Inc. 1584 Enterprise Boulevard West Sacramento, CA 95691
Re: k020246 Trade/Device Name: MicroScan® Synergies plus™ Gram-Negative MIC/Combo Panels with Imipenem (0.25, 1-32 µg/ml) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: July 30, 2004 Received: August 3, 2004
Dear Mr. Eusebio:
This letter corrects our substantially equivalent letter of February 6, 2002, regarding the trade This letter of to MicroScan® Synergies Plus to better reflect the intended use of the device.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent [(for the indications for use stated in the enclosure)] to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your de roo is classified and controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-3084. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at their toll free number (800) 638-2041 or at (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Freddie W. Poole for
Sally A. Hojvat, M.Sc., Ph Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use Statement
## 510(k) Number (if known): K020246
MicroScan® Synergies plus™ Gram-Negative MIC/Combo Panels with Imipenem Device Name: (0.25, 1 - 32 µg/ml)
### Indications For Use:
MicroScan® Synergies plus™ Gram-Negative MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic Gram-Negative bacilli (Enterobacteriaceae, glucose non-fermenters, and non-Enterobacteriaceae glucose (Enterouters) After inoculation, panels are read on the WalkAway® S/ System or equivalent (upgraded WalkAway® 40 or WalkAway® 96) according to the Package Insert.
This particular submission is for the antimicrobial Imipenem on the Synergies plus™ Gram-Negative MIC/Combo Panels.
The Gram-Negative organisms which may be used for Imipenem susceptibility testing in this panel are:
Achromobacter spp Acinetobacter spp Aeromonas hydrophila Alcaligenes spp Citrobacter spp Escherichia coli Enterobacter spp Hafnia alvei Klebsiella spp Morganella morganii Pasteurella multocida Proteus vulgaris Providencia rettgeri Providencia stuartii Pseudomonas aeruginosa Serratia spp Shigellia spp Salmonella spp Yersinia spp
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
Page 1 of _ 1
510(k) 020246
{3}------------------------------------------------
# FEB 0 6 2002
### 510(k) Summary
### 510(k) Submission Information:
| Device Manufacturer: | Dade Behring Inc. |
|----------------------|-------------------------------------------------------------|
| Contact name: | Maureen Mende, Group Manager Regulatory Affairs |
| Fax: | 916-374-3144 |
| Date prepared: | January 22, 2002 |
| Product Name: | Microdilution Minimum Inhibitory Concentration (MIC) Panels |
| Trade Name: | MicroScan® rapID/S plus™ Gram-Negative MIC/Combo panels |
| Intended Use: | To determine antimicrobial agent susceptibility |
| 510(k) Notification: | Antimicrobials: Imipenem |
| Predicate device: | MicroScan Dried Gram Negative MIC/Combo Panels |
#### 510(k) Summary:
MicroScan® rapID/S plus™ Gram-Negative MIC/Combo Panels are designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-negative bacilli. The MicroScan® rapID/S plus™ Cram-Negative MIC/Combo Panels are read on the WalkAway® S7 System or equivalent (upgraded WalkAway® 40 or WalkAway® 96 instruments).
The antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test that have been diluted in Mueller-Hinton Broth to concentrations bridging the range of clinical interest and are presented in micro-titer wells in dried form. rapID/S plus™ panels are inoculated with a standardized suspension of the organism and incubated at 35℃ in the WalkAway® SI System or equivalent for 4.5 - 18 hours. The minimum inhibitory concentration (MIC) for the test organism is determined by the lowest antimicrobial concentration showing inhibition of growth.
The proposed MicroScan® rapID/S plus™ Gram-Negative MIC/Combo Panel demonstrated substantially equivalent performance when compared with an NCCLS frozen Reference Panel, as defined in the FDA DRAFT document "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", dated March 8, 2000. The Premarket Notification (510[k]) presents data in support of the MicroScan® rapID/S plus™ Gram-Negative MIC/Combo Panel Imipenem.
The external evaluation was conducted with fresh and stock Efficacy isolates and stock Challenge strains. The external evaluations were designed to confirm the acceptability of the proposed rapID/S plas™ Gram-Negative Panel by comparing its performance with an NCCLS frozen Reference panel. Challenge strains were compared to Expected Results determined prior to the evaluation. The rapID/S plus™ Gram-Negative Panel demonstrated acceptable performance with an overall Essential Agreement of greater than 96% for Imipenem when compared with the frozen Reference panel.
Instrument reproducibility testing demonstrated acceptable reproducibility and precision with Turbidity inoculum preparation method and the WalkAway® S7 System or equivalent (upgraded WalkAway® 40 or WalkAway® 96 instruments).
Quality Control testing demonstrated acceptable results for Imipenem.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.