DRIED GRAM-NEGATIVE MIC/COMBO PANELS
Device Facts
| Record ID | K970676 |
|---|---|
| Device Name | DRIED GRAM-NEGATIVE MIC/COMBO PANELS |
| Applicant | Dade Intl., Inc. |
| Product Code | JWY · Microbiology |
| Decision Date | May 2, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 866.1640 |
| Device Class | Class 2 |
Indications for Use
To determine antimicrobial agent susceptibility
Device Story
MicroScan Dried Gram-Negative MIC/Combo Panels determine minimum inhibitory concentrations (MIC) of Levofloxacin for Gram-negative bacteria; used in clinical microbiology laboratories. Device utilizes microdilution method; panels contain dried antimicrobial agents. Inoculation performed via Turbidity or Prompt methods; panels incubated and read using autoScan-4 or WalkAway instruments. Output provides MIC values to guide antibiotic therapy selection. Performance compared against NCCLS frozen reference panels to ensure clinical efficacy.
Clinical Evidence
Bench testing only. Evaluation used fresh and stock efficacy isolates and stock challenge strains. Compared MicroScan panels to NCCLS frozen reference panels. Reported 100% Essential Agreement. Reproducibility and precision testing conducted across Turbidity and Prompt inoculum methods and autoScan-4/WalkAway instruments. Quality control performance deemed acceptable.
Technological Characteristics
Microdilution panels with dried antimicrobial agents. Compatible with autoScan-4 and WalkAway automated reading instruments. Inoculation via Turbidity or Prompt methods.
Indications for Use
Indicated for determining antimicrobial susceptibility of Gram-negative bacteria to Levofloxacin.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
Predicate Devices
- NCCLS Frozen Levofloxacin Reference Panels
Related Devices
- K972688 — MICROSCAN DRIED GRAM POSITIVE MIC/COMBO PANELS LEVOFLOXACIN · Dade Microscan, Inc. · Sep 8, 1997
- K972180 — DRIED GRAM-NEGATIVE MIC/COMBO PANELS · Dade Microscan, Inc. · Aug 1, 1997
- K083262 — MICROSCAN DRIED GRAM -NEGATIVE MIC/COMBO PANELS WITH TIGECYCLINE (0.015- 32 MCG/ML) · Siemens Healthcare Diagnostics · Feb 18, 2009
- K183127 — MicroScan Dried Gram Negative MIC/Combo Panels with Meropenem/Vaborbactam (0.03/8-64/8ug/mL) · Beckman Coulter, Inc. · Feb 12, 2019