Sensititre 18-24 hour MIC or Breakpoint Susceptibility System with Cefiderocol in the dilution range of 0.03-64 µg/ml
Applicant
Thermo Fisher Scientific
Product Code
JWY · Microbiology
Decision Date
Mar 12, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1640
Device Class
Class 2
Intended Use
The Sensititre MIC and Breakpoint Susceptibility system is an in vitro diagnostic product for clinical susceptibility testing of non-fastidious gram negative isolates comprising of Enterobacteriaceae, Pseudomonas aeruginosa and other non-Enterobacteriaceae and of non-fastidious gram positive isolates, comprising of Staphylococcus spp., Enterococcus spp., and Beta-hemolytic Streptococci other than S. pneumoniae.
Device Story
Sensititre 18-24 hour MIC or Breakpoint Susceptibility System performs antimicrobial susceptibility testing (AST). System utilizes microbroth dilution method to determine minimum inhibitory concentrations (MIC) or breakpoint results for Cefiderocol. Input consists of bacterial colonies isolated from patient samples; these are inoculated into test plates containing varying concentrations of antimicrobial agents. Plates are incubated for 18-24 hours. Growth or inhibition is assessed to determine susceptibility. Used in clinical microbiology laboratories by trained technicians. Output provides quantitative MIC values or qualitative susceptibility categories (e.g., susceptible, resistant) to guide antibiotic therapy selection. Benefits include standardized, accurate susceptibility data to inform clinical decision-making for bacterial infections.
Clinical Evidence
Performance evaluated via bench testing comparing the Sensititre system with Cefiderocol against reference broth microdilution methods. Testing assessed essential agreement (EA) and categorical agreement (CA) across a range of Gram-negative isolates. Results demonstrated performance met acceptance criteria for accuracy and reproducibility. No clinical prospective trials were required for this specific 510(k) clearance as the device relies on established AST performance standards.
Technological Characteristics
Microbroth dilution susceptibility testing system. Consists of multi-well plates containing dehydrated antimicrobial agents (Cefiderocol). Dimensions/form factor: standard microtiter plate format. Energy source: incubator-dependent. Connectivity: standalone or integrated with laboratory information systems. Sterilization: N/A (disposable test plates). Software: rule-based analysis for growth detection and MIC calculation.
Indications for Use
Indicated for determining antimicrobial susceptibility of non-fastidious and fastidious Gram-negative bacterial isolates to Cefiderocol (0.03 - 64 ug/mL). For use by trained laboratory personnel in clinical settings.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
Predicate Devices
Sensititre 18-24 hour MIC or Breakpoint Susceptibility System with Omadacycline (K183033)
Related Devices
K203741 — Sensititre 18-24 hour MIC or Breakpoint Susceptibility System with Cefiderocol in the dilution range of 0.03-64 ug/ml · Thermo Fisher Scientific · Dec 20, 2021
K193024 — Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint, Susceptibility System with Lefamulin in the dilution range of 0.008-16ug/mL · Thermo Fisher Scientific · Jan 8, 2020
K192729 — Sensititre 18-24 Hour MIC or Breakpoint Susceptibility System with Lefamulin in the dilution range of 0.008-16ug/ml · Thermo Fisher Scientific · Nov 6, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 12, 2020
Thermo Fisher Scientific Cynthia Knapp Director of AST/PHARMA R&D 1 Thermo Fisher Way Oakwood Village, Ohio 44014-0006
## Re: K193538
Trade/Device Name: Sensititre 18-24 hour MIC or Breakpoint Susceptibility System with Cefiderocol in the dilution range of 0.03 - 64 ug/mL Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial Susceptibility Test Powder Regulatory Class: Class II Product Code: JWY, LRG, LTT Dated: December 18, 2019 Received: December 20, 2019
## Dear Cynthia Knapp:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ribhi Shawar, Ph.D. (ABMM) Chief General Bacteriology and Antimicrobial Susceptibility Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.