MTS Imipenem-relebactam 0.002/4-32/4 µg/mL

K191809 · Liofilchem s.r.l. · JWY · Aug 23, 2019 · Microbiology

Device Facts

Record IDK191809
Device NameMTS Imipenem-relebactam 0.002/4-32/4 µg/mL
ApplicantLiofilchem s.r.l.
Product CodeJWY · Microbiology
Decision DateAug 23, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1640
Device ClassClass 2

Intended Use

The MTS (MIC Test Strip) is a quantitative method intended for the in vitro determination of antimicrobial susceptibility of non-fastidious Gram-negative organisms. MIC are used to determine the minimum inhibitory concentration (MIC) of antimicrobial agents that inhibit the growth of bacteria and thus to aid in determining the susceptibility of microorganisms to the given antimicrobial agent. The MTS Imipenem-relebactam 0.002/4-32/4 ug/mL is a quantitative method intended for the in vitro determination of antimicrobial susceptibility of non-fastidious Gram-negative organisms. The MTS is used to determine the minimum inhibitory concentration (MIC) of Imipenem-relebactam that inhibit the growth of bacteria and thus to aid in determining the susceptibility of microorganisms to the given antimicrobial agent.

Device Story

MTS Imipenem-relebactam is a quantitative antimicrobial susceptibility test; consists of specialized paper strip impregnated with predefined exponential concentration gradient of imipenem/relebactam (0.002/4–32/4 µg/mL). Used in clinical microbiology laboratories; manual procedure. Strip applied to inoculated Mueller-Hinton agar; incubated 16-20 hours. Antibiotic diffuses into agar; forms symmetrical inhibition ellipse. User manually reads MIC value at intersection of ellipse edge and strip. Results guide clinicians in selecting appropriate antibiotic therapy for bacterial infections. Benefits include standardized, quantitative determination of MIC for non-fastidious Gram-negative organisms.

Clinical Evidence

Bench testing only. Performance was evaluated by comparing MIC results obtained with the MTS Imipenem-relebactam to a reference broth microdilution method. Testing included essential agreement and categorical agreement metrics across a range of non-fastidious Gram-negative organisms to demonstrate performance equivalence.

Technological Characteristics

Specialized paper strip impregnated with 15 two-fold dilutions of imipenem/relebactam. Manual reading of inhibition ellipse. Media: Mueller-Hinton agar. Incubation: 35±2°C for 16-20 hours. No electronic components, software, or connectivity.

Indications for Use

Indicated for in vitro determination of antimicrobial susceptibility of non-fastidious Gram-negative organisms to Imipenem-relebactam. Used to determine minimum inhibitory concentration (MIC) to aid in susceptibility assessment.

Regulatory Classification

Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo, which features a stylized caduceus. To the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. August 23, 2019 Liofilchem s. r. 1. % Anne Windau Supervisor Laboratory Specialists, Inc 26214 Center Ridge Road Westlake, Ohio 44145 Re: K191809 Trade/Device Name: MTS Imipenem-relebactam 0.002/4-32/4 ug/mL Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial Susceptibility Test Powder Regulatory Class: Class II Product Code: JWY Dated: July 3, 2019 Received: July 5, 2019 Dear Ms. Windau: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Ribhi Shawar, Ph.D. (ABMM) Chief. General Bacteriology and Antimicrobial Susceptibility Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...