Liofilchem MIC Test Strip (MTS)- Linezolid 0.016 - 256ug/mL
Applicant
Liofilchem s.r.l.
Product Code
JWY · Microbiology
Decision Date
Mar 15, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1640
Device Class
Class 2
Intended Use
The Liofilchem® MIC Test Strip (MTS) is a quantitative method intended for the in vitro determination of antimicrobial susceptibility of bacteria. MTS consists of specialized paper impregnated with a pre-defined concentration gradient of an antimicrobial agent, which is used to determine inhibitory concentration (MIC) in ug/mL of antimicrobial agents against bacteria as tested on agar media using overnight incubation and manual reading procedures. The Linezolid MTS at concentrations of 0.016- 256 ug/mL should be interpreted at 16-20 hours of incubation. Linezolid has been shown to be active both clinically and in vitro agamst the non-fastidious bacteria listed below according to the FDA label: Staphylococcus aureus (including methicillin-resistant isolates) Enterococcus faecium (vancomycin resistant isolates only)
Device Story
Liofilchem MIC Test Strip (MTS) is a quantitative diagnostic tool for determining antimicrobial susceptibility. It consists of a specialized paper strip impregnated with a predefined concentration gradient of Linezolid (0.016–256 µg/mL). In a clinical laboratory setting, the strip is applied to inoculated agar media. Following 16–20 hours of overnight incubation, the user manually observes the zone of inhibition to determine the MIC value. This result assists clinicians in selecting appropriate antibiotic therapy for patients with specific bacterial infections. The device provides a standardized method for assessing bacterial resistance, supporting clinical decision-making for targeted treatment.
Clinical Evidence
Performance evaluated using 299 clinical and 77 challenge isolates (S. aureus, E. faecium). Compared to CLSI broth microdilution reference method. S. aureus (MSSA/MRSA) showed 99.0% EA and 99.5% CA. E. faecium showed 98.7% EA and 94.0% CA. Trending analysis indicated a slight one-dilution higher MIC bias compared to reference, deemed acceptable. Reproducibility study (3 sites, 10 isolates, 270 data points) demonstrated ≥95% reproducibility. QC testing with ATCC strains met expected ranges >95% of the time.
Technological Characteristics
Specialized paper strip impregnated with linezolid gradient; 15 two-fold dilutions. Manual reading of inhibition ellipse on Mueller Hinton agar. Incubation at 35°C for 16–20 hours. No electronic components, software, or connectivity. Single-use.
Indications for Use
Indicated for in vitro antimicrobial susceptibility testing of non-fastidious bacteria, specifically Staphylococcus aureus (including methicillin-resistant isolates) and Enterococcus faecium (vancomycin-resistant isolates only), to determine the Minimum Inhibitory Concentration (MIC) of Linezolid.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
Related Devices
K170892 — Liofilchem MIC Test Strip (MTS)- Tedizolid 0.002 - 32 µg/mL · Liofilchem s.r.l. · May 25, 2017
K172542 — Liofilchem MIC Test Strip (MTS)- Tedizolid 0.002 - 32 mg/mL. · Liofilchem s.r.l. · Oct 23, 2017
K163517 — Liofilchem MIC Test Strip (MTS) Telavancin 0.016 - 256.0 µg/mL · Liofilchem s.r.l. · Mar 14, 2017
K180339 — Liofilchem MIC Test Strip (MTS), Azithromycin 0.016 - 256 µg/mL · Liofilchem s.r.l. · Apr 4, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 15, 2018
Liofilchem s.r.l. % Laura Koeth President Laboratory Specialists, Inc 26214 Center Ridge Road Westlake, Ohio 44145
Re: K180330
Trade/Device Name: Liofilchem MIC Test Strip (MTS) - Linezolid 0.016 - 256 ug/mL Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial Test Powder Regulatory Class: Class II Product Code: JWY Dated: February 1, 2018 Received: February 6, 2018
Dear Laura Koeth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR
{1}------------------------------------------------
Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Ribhi Shawar -S For
Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
#### 510(k) Number (if known)
#### K180330
#### Device Name
Liofilchem MIC Test Strip (MTS), Linezolid 0.016-256 ug/mL
Indications for Use (Describe)
The Liofilchem® MIC Test Strip (MTS) is a quantitative method intended for the in vitro determination of antimicrobial susceptibility of bacteria. MTS consists of specialized paper impregnated with a pre-defined concentration gradient of an antimicrobial agent, which is used to determine inhibitory concentration (MIC) in ug/mL of antimicrobial agents against bacteria as tested on agar media using overnight incubation and manual reading procedures.
The Linezolid MTS at concentrations of 0.016- 256 ug/mL should be interpreted at 16-20 hours of incubation.
Linezolid has been shown to be active both clinically and in vitro agamst the non-fastidious bacteria listed below according to the FDA label:
Staphylococcus aureus (including methicillin-resistant isolates) Enterococcus faecium (vancomycin resistant isolates only)
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
2 Prescription Use (Part 21 CFR 801 Subpart D)
O Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"Afl agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.