SENSITITRE HAEMOPHILUS/STREPTOCOCCUS PNEUMONIAE(HP) SUSCEPTIBILITY PLATES

K140985 · Thermo Fisher Scientific · JWY · Jun 10, 2014 · Microbiology

Device Facts

Record IDK140985
Device NameSENSITITRE HAEMOPHILUS/STREPTOCOCCUS PNEUMONIAE(HP) SUSCEPTIBILITY PLATES
ApplicantThermo Fisher Scientific
Product CodeJWY · Microbiology
Decision DateJun 10, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1640
Device ClassClass 2

Intended Use

The Sensititre Haemophilus influenzae/Streptococcus pneumoniae plates are in vitro diagnostic products for clinical susceptibility testing of Haemophilus influenzae,Streptococcus pneumoniae and Streptococcus species. This 510(k) is for the addition of Clindamycin in the dilution range of 0.015-1ug/ml to the Sensititre HP MIC susceptibility plate for testing Streptococcus species. The approved primary "Indications for Use" and clinical significance of Clindamycin is for: Faculative Gram-Positive Fastidious Microorganisms: Streptococcus pyogenes Streptococcus agalactiae Streptococcus anginosus

Device Story

Sensititre HP MIC Susceptibility Plate is an in vitro diagnostic tool used in clinical microbiology laboratories. It utilizes microbroth dilution to determine the minimum inhibitory concentration (MIC) of antimicrobial agents against specific fastidious bacteria. The device consists of plates containing pre-dispensed, dried antimicrobial agents (specifically Clindamycin) at various concentrations. Laboratory technicians inoculate the plates with patient bacterial isolates; plates are incubated; bacterial growth is observed to determine the MIC. The resulting MIC values assist clinicians in selecting appropriate antibiotic therapy for patients with infections caused by susceptible organisms. The device provides standardized, quantitative susceptibility results to guide clinical decision-making and improve patient outcomes.

Clinical Evidence

Performance evaluated using 225 clinical and 60 challenge isolates of S. agalactiae, S. pyogenes, and S. anginosus. Testing compared the Sensititre system (manual and AutoRead) against CLSI M07-A9 frozen broth microdilution reference panels. Combined results showed 99.6% Essential Agreement (EA) and 100% Category Agreement (CA) for both reading methods. Reproducibility studies demonstrated 100% performance. Quality control using S. pneumoniae ATCC 49619 met expected ranges.

Technological Characteristics

Microbroth dilution susceptibility plate; contains dried antimicrobial agents (Clindamycin) in specific dilution ranges (0.015-1 ug/ml). Designed for in vitro diagnostic use in clinical laboratories. Standardized plate format for testing fastidious Gram-positive organisms.

Indications for Use

Indicated for in vitro clinical susceptibility testing of Haemophilus influenzae, Streptococcus pneumoniae, and Streptococcus species (specifically S. pyogenes, S. agalactiae, and S. anginosus) to determine minimum inhibitory concentrations (MIC) of Clindamycin.

Regulatory Classification

Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines. # DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 THERMO FISHER SCIENTIFIC CYNTHIA KNAPP DIRECTOR OF CLINICAL OPERATIONS I THERMO FISHER WAY OAKWOOD VILLAGE OH 44146 June 10, 2014 Re: K140985 Trade/Device Name: Sensititre Haemophilus influenzae/Streptococcus pneumoniae Susceptibility Plates Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial susceptibility test powder Regulatory Class: II Product Code: JWY, LRG, LTT, LTW Dated: April 15, 2014 Received: April 16, 2014 Dear Ms. Knapp: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Page 2-Ms. Knapp If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Uwe Scherf -S for Sally Hojvat, M.Sc., PhD. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use #### 510(k) Number (if known) K140985 ### Device Name Sensititre HP MIC Susceptibility Plate Clindamycin (0.0 l 5-1 ug/m)) Indications for Use (Describe) The Sensititre Haemophilus influenzae/Streptococcus pneumoniae plates are in vitro diagnostic products for clinical susceptibility testing of Haemophilus influenzae,Streptococcus pneumoniae and Streptococcus species. This 510(k) is for the addition of Clindamycin in the dilution range of 0.015-1ug/ml to the Sensititre HP MIC susceptibility plate for testing Streptococcus species. The approved primary "Indications for Use" and clinical significance of Clindamycin is for: Faculative Gram-Positive Fastidious Microorganisms: Streptococcus pyogenes Streptococcus agalactiae Streptococcus anginosus Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) # PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. ## FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Ribhi Shawar -S 2014.06.10 16:22:10 -04'00' This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...