The Sensititre Haemophilus influenzae/Streptococcus pneumoniae plates are in vitro diagnostic products for clinical susceptibility testing of Haemophilus influenzae, Streptococcus pneumoniae and Streptococcus species. Sensititie® Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC Susceptibility Plates with Dtest (containing erythromycin at 1 µg/ml and clindamycin at 0.5µg/ml) broth test for Streptococcus pneumoniae and Streptococus spp.-9-Hemolytic Group is an in vitro diagnostic product for clinical susceptibility testing. The Dtest for broth microdilution test is for manual read detection of inducible clindamycin resistance in Streptococcus pyogenes and Streptococcus agalactiae resistant to erythromycin (MICs ≥ 1 µg/mL) and susceptible or intermediate to clindamycin (MICs ≤ 0.25 µg/mL or 0.5 µg/mL), and to determine absence of inducible clindamycin resistance in Streptococus pneumoniae. The performance of this test for the detection of inducible clindamycin resistance in isolates of S. pneumoniae has not been established. With S. pyogenes and S. agalactiae, a Dtest (1/0.5 ug/ml) positive (growth) test, determined by manual read, should be reported as inducible clindamycin resistance.
Device Story
Sensititre HP Susceptibility Plate is an in vitro diagnostic device for antimicrobial susceptibility testing. It uses broth microdilution in wells containing specific concentrations of erythromycin (1 µg/mL) and clindamycin (0.5 µg/mL). Standardized organism suspensions are prepared in cation-adjusted Mueller-Hinton broth with TES buffer and lysed horse blood; inoculated using the AIM (Automated Inoculation Delivery System); sealed; and incubated at 34-36°C for 20-24 hours. Results are determined by manual visual observation of turbidity or cell deposits. Growth indicates inducible clindamycin resistance; no growth indicates absence. Used in clinical laboratories to guide antibiotic therapy for Streptococcus infections. Benefits include identification of inducible resistance that might otherwise be missed, informing clindamycin treatment decisions.
Clinical Evidence
No clinical data provided. The submission relies on bench testing for the performance of the broth microdilution and Dtest methodology for susceptibility testing.
Indicated for clinical susceptibility testing of S. pneumoniae and Streptococcus spp.-β-Hemolytic Group. Dtest is for manual detection of inducible clindamycin resistance in S. pyogenes and S. agalactiae (erythromycin resistant, clindamycin susceptible/intermediate) and determining absence of inducible clindamycin resistance in S. pneumoniae.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
K073653 — SENSITITRE 18 - 24 HOUR SUSCEPTIBILITY PLATE · Trek Diagnostic Systems, Inc. · Mar 20, 2008
K040619 — SENSITITRE HAEMOPHILUS/STREPTOCOCCUS PNEUMONIAE (HP) MIC PLATES · Trek Diagnostic Systems, Inc. · Apr 7, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract design of a stylized eagle or bird-like figure, composed of three curved lines that resemble profiles of human faces.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
THERMO FISHER SCIENTIFIC CYNTHIA KNAPP DIRECTOR OF CLINICAL OPERATIONS 1 THERMO FISHER WAY OAKWOOD VILLAGE OH 44146
August 1, 2014
Re: K133847
Trade/Device Name: Sensititre Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC Susceptibility Plates with Dtest (containing erythromycin at 1ug/mL and clindamvcin at 0.5 ug/mL) Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial susceptibility test powder Regulatory Class: II Product Code: JWY, LTT, LTW Dated: July 24, 2014 Received: July 25, 2014
Dear Ms. Knapp:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the
{1}------------------------------------------------
electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Uwe Scherf - S for
Sally Hojvat, M.Sc., PhD. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K133847
#### Device Name
Sensitite® Haemophilus influenza/Streptococus pneumoniae (HP) MIC Susceptibility plates with Dtest (containing erythromyin at 1 ug/mL and clindamycin at 0.5 ug/mL)
### Indications for Use (Describe)
The Sensititre Haemophilus influenzae/Streptococcus pneumoniae plates are in vitro diagnostic products for clinical susceptibility testing of Haemophilus influenzae, Streptococcus pneumoniae and Streptococcus species.
Sensititie® Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC Susceptibility Plates with Dtest (containing erythromycin at 1 µg/ml and clindamycin at 0.5µg/ml) broth test for Streptococcus pneumoniae and Streptococus spp.-9-Hemolytic Group is an in vitro diagnostic product for clinical susceptibility testing.
The Dtest for broth microdilution test is for manual read detection of inducible clindamycin resistance in Streptococcus pyogenes and Streptococcus agalactiae resistant to erythromycin (MICs ≥ 1 µg/mL) and susceptible or intermediate to clindamycin (MICs ≤ 0.25 µg/mL or 0.5 µg/mL), and to determine absence of inducible clindamycin resistance in Streptococus pneumoniae. The performance of this test for the detection of inducible clindamycin resistance in isolates of S. pneumoniae has not been established.
With S. pyogenes and S. agalactiae, a Dtest (1/0.5 ug/ml) positive (growth) test, determined by manual read, should be reported as inducible clindamycin resistance.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
- Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.