PASCO MIC AND MIC/ID PANELS, GEMIFLOXACIN (STREP)

K032518 · Bd Diagnostic Systems · JWY · Sep 15, 2003 · Microbiology

Device Facts

Record IDK032518
Device NamePASCO MIC AND MIC/ID PANELS, GEMIFLOXACIN (STREP)
ApplicantBd Diagnostic Systems
Product CodeJWY · Microbiology
Decision DateSep 15, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1640
Device ClassClass 2

Indications for Use

Pasco MIC and MIC/ID panels are used for quantitatively measuring the susceptibility of rapidly growing aerobic and facultative anaerobic bacterial pathogens to a battery of antimicrobial agents and determining the biochemical identification of these organisms. The MIC Supplemental III panel contains antimicrobial agents fully diluted to concentrations appropriate for testing S. pneumoniae and other Streptococcus spp. which require the use of SP Blood Supplement for inoculation. This panel also contains antimicrobial agents at concentrations appropriate for testing various non-fastidious organisms which use routine inoculation procedure.

Device Story

Pasco MIC and MIC/ID panels are frozen microdilution plates containing two-fold dilutions of antimicrobial agents. Panels are thawed, inoculated with test organisms (Streptococcus spp.) using a direct turbidity method, and incubated for 16-24 hours at 35°C in a non-CO2 incubator. The device operates on the principle of broth microdilution; visible growth or color change is observed manually using an indirect lighted background viewer. The minimum inhibitory concentration (MIC) is defined as the lowest concentration of antibiotic preventing visible bacterial growth. Results are used by clinicians to determine antimicrobial susceptibility, aiding in the selection of appropriate antibiotic therapy for patients with bacterial infections.

Clinical Evidence

Bench testing only. Performance evaluated via method comparison against NCCLS broth reference panels using 570 clinical and challenge isolates of Streptococcus pneumoniae. Combined results showed 99.6% Essential Agreement and 98.9% Category Agreement. Reproducibility testing across three sites over three days showed >95% agreement. Quality control testing using ATCC 49619 and Pasco 4223 isolates demonstrated results within recommended ranges.

Technological Characteristics

Microdilution panels containing antimicrobial agents in two-fold dilutions. Principle of operation: broth microdilution for MIC determination and biochemical substrate analysis for identification. Incubation period: 16-24 hours. Manual or visual interpretation of growth/color change.

Indications for Use

Indicated for testing Streptococcus pneumoniae and Streptococcus spp. other than S. pneumoniae for susceptibility to the antimicrobial Gemifloxacin at concentrations of 0.015–4 ug/ml.

Regulatory Classification

Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: k032518 B. Analyte: Gemifloxacin at 0.015-4 ug/ml C. Type of Test: Quantitative – broth based growth detected by turbidity D. Applicant: BD Diagnostic Systems – Pasco Laboratories E. Proprietary and Established Names: Pasco MIC and MIC/ID Panels F. Regulatory Information: 1. Regulation section: 866.1640 - Antimicrobial Susceptibility Test Powder 2. Classification: II 3. Product Code: JWY - Manual Antimicrobial Test Systems 4. Panel: 83 - Microbiology G. Intended Use: 1. Intended use(s): Pasco MIC and MIC/ID panels are used for quantitatively measuring the susceptibility of rapidly growing aerobic and facultative anaerobic bacterial pathogens to a battery of antimicrobial agents and determining the biochemical identification of these organisms. The MIC Supplemental III panel contains antimicrobial agents fully diluted to concentrations appropriate for testing S. pneumoniae and other Streptococcus spp. which require the use of SP Blood Supplement for inoculation. This panel also contains antimicrobial agents at concentrations appropriate for testing various non-fastidious organisms which use routine inoculation procedure. 2. Indication(s) for use: The indication is for the addition of the antimicrobial Gemifloxacin at concentrations of 0.015–4 ug/ml to Pasco Panels for use in testing Streptococcus pneumoniae and Streptococcus spp. other than S. pneumoniae. 3. Special condition for use statement(s): Direct turbidity method of inoculation. 4. Special instrument Requirements: Not Applicable {1} Page 2 of 4 H. Device Description: Various concentrations of antimicrobial agents (usually in two-fold dilutions) are dispensed into the Pasco microdilution panels and the panels are then frozen. Panels are thawed prior to use, inoculated with the test organisms, incubated the traditional 16-24 hours at 35° in a non-CO₂ incubator and panels are then observed for visible growth or color changes (ID portion). The lowest concentration of each antimicrobial agent with no apparent visible growth of the test organism is recorded as the minimum inhibitory concentration (MIC). Only manual readings are performed using an indirect lighted background viewer. I. Substantial Equivalence Information: 1. Predicate device name(s): Pasco MIC Panels- gatifloxacin 2. Predicate K number(s): K032259 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Type panel | 100 μl/well frozen | 100 μl/well frozen | | Inoculum | 5 μl | 5 μl | | Inoculation method | Direct equated to a 0.5 McFarland | Direct equated to a 0.5 McFarland | | Incubation | 16-24 hours | 16-24 hours | | Reading method | Visual growth | Visual growth | | Differences | | | | Item | Device | Predicate | | Antibiotic | Gemifloxacin at predefined 2-fold dilutions | Gatifloxacin at predefined 2-fold dilutions | J. Standard/Guidance Document Referenced (if applicable): "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA", NCCLS Standard M7 Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically; Approved Standard; M100 Performance Standards for Antimicrobial Susceptibility Testing. K. Test Principle: The test panels are dependent on the growth of the organisms in the presence of the antibiotics. The lowest concentration of each antimicrobial agent with no apparent visible growth of the test organism is recorded as the minimum inhibitory concentration (MIC). L. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Ten on-scale Streptococcus spp. were tested at three sites on three separate days in triplicate. The reproducibility was acceptable at >95%. {2} Page 3 of 4 b. Linearity/assay reportable range: Not applicable c. Traceability (controls, calibrators, or method): The recommended Quality Control (QC) isolate was tested a sufficient number of times with acceptable results with the reference method. An additional isolate was selected that would provide on scale results all the time. The Pasco results demonstrate that the system can produce QC results in the recommended range. | ORGANISM | Conc ug/ml | Reference | Pasco turbidity | | --- | --- | --- | --- | | | | | | | S. pneumoniae ATCC 49619 Expected Range 0.008-0.03 ug/ml | ≤ 0.015 | 71 | 74 | | | 0.03 | 3 | | | | 0.06 | | | | | | | | | | | | | | | | | | | S. pneumoniae Pasco 4223 Expected Range 0.25-2 ug/ml | 0.12 | | | | | 0.25 | 3 | | | | 0.5 | 38 | 40 | | | 1 | | 1 | | | 2 | | | Inoculum Density Check- An internal study was performed to verify the colony counts (CC) that would be obtained with each method of inoculation. Clinical site inoculum density checks were also performed on QC isolates, reproducibility isolates and a subset of the clinical isolates and demonstrated the appropriate inoculum was used. d. Detection limit: Not applicable e. Analytical specificity: Not applicable f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: Broth reference panels supplemented with 2-5% lysed horse blood prepared according to the recommendations of the NCCLS M7 were used to compare to the Pasco results. Testing was performed at 3 sites and included fresh and stock clinical isolates and a set of challenge organisms. The selection of S. pneumoniae included both penicillin resistant and penicillin susceptible. The comparison resulted in the following performance evaluations. {3} Page 4 of 4 | | total | EA | %EA | Total evaluable | EA of evaluable | %EA | CA | %CA | #R | Min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Clinical | 470 | 468 | 99.6 | 194 | 192 | 99 | 464 | 98.7 | 8 | 6 | 0 | 0 | | Challenge | 100 | 100 | 100 | 30 | 30 | 100 | 100 | 100 | 5 | 0 | 0 | 0 | | Combined | 570 | 568 | 99.6 | 224 | 222 | 99 | 564 | 98.9 | 13 | 6 | 0 | 0 | EA-Essential Agreement CA-Category Agreement R-resistant isolates maj-major discrepancies vmj-very major discrepancies min- minor discrepancies EA is when there is agreement between the reference method and the Pasco panel within plus or minus one serial two-fold dilution of antibiotic. CA is when the interpretation of the reference method agrees exactly with the interpretation of the Pasco result. The %EA and CA are acceptable with acceptable discrepancy rates when compared to the reference method as described in the FDA guidance document, “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA”. b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a and b are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: $\leq 0.12(\mathrm{S}), 0.25(\mathrm{I}), \geq 0.5(\mathrm{R})$ These interpretative criteria are the same as those in the FDA approved gemifloxacin pharmaceutical package insert and the NCCLS M100 Standard. The recommended QC is also the same and appears in the package insert. M. Conclusion: Data analysis when analyzed as recommended in the “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA” demonstrates that the Pasco System is substantially equivalent to the predicate.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...